How long a blood sample keeps

How long a sample can wait before it's measured depends on the analyte, the tube, whether it was spun and separated, and the temperature. This collects what published studies found: 160 analytes, 1434 findings from 51 studies, each with the study's own test for “stable” and the table it came from.

25-Hydroxyvitamin D (25-OH D)34 findings
Specimen, tubeConditionTemperatureStable forCriterionStudy
plasma-edta, BD Vacutainer EDTA (2 mL)Storageseparated−20 °C7 daysstable for the longest time testedFriedman repeated measures ANOVA vs zero time, P <= 0.05Colak et al. 2013, Biochem Med (Zagreb)Table 2
plasma-edta, BD Vacutainer EDTA (2 mL)Storageseparated−80 °C3 monthsstable for the longest time testedFriedman repeated measures ANOVA vs zero time, P <= 0.05Colak et al. 2013, Biochem Med (Zagreb)Table 2
plasma-edta, BD Vacutainer EDTA (2 mL)Storageseparated2-8 °C24 hstable for the longest time testedFriedman repeated measures ANOVA vs zero time, P <= 0.05Colak et al. 2013, Biochem Med (Zagreb)Table 1
plasma-edta, BD Vacutainer EDTA (2 mL)StorageseparatedRT4 hstable for the longest time testedFriedman repeated measures ANOVA vs zero time, P <= 0.05Colak et al. 2013, Biochem Med (Zagreb)Table 1
serum, BD Vacutainer SSTFreeze–thawseparated−70 °C, 1 month; thawed at room temperature1 cyclestable for the longest time testedmean % change (bias) from fresh baseline within total allowable error (TEa, desirable bias)Abraham et al. 2021, J Lab PhysiciansTable 2 and Table 3, row 'Vitamin D (ng/mL)'
serum, not statedStorageseparated−20 °C7 monthsstable for the longest time testedFriedman test P < 0.05 with Bonferroni-adjusted post hoc vs baseline (#)Ko et al. 2025, Sci RepTable 1, overall, −20 °C
serum, not statedStorageseparated−20 °C7 monthsstable for the longest time testedFriedman test P < 0.05 with Bonferroni-adjusted post hoc vs baseline (#)Ko et al. 2025, Sci RepTable 2, overall, −20 °C
serum, BD Vacutainer serum separator tube with clot activator (8.5 mL)Storageseparated−20 °C7 daysstable for the longest time testedFriedman repeated measures ANOVA vs zero time, P <= 0.05Colak et al. 2013, Biochem Med (Zagreb)Table 2
serum, BD plain (red top) tubeStorageseparated−20 °C1 daysthen changed, changed by 2 dayspaired t test vs day 0, P < .05Borai et al. 2020, J Clin Lab AnalTable 2, non-vortexed
serum, BD plain (red top) tubeStorageseparated−20 °C0 dayschanged at the first time point, changed by 1 dayspaired t test vs day 0, P < .05Borai et al. 2020, J Clin Lab AnalTable 2, vortexed
serum, BD plain (red top) tubeStorageseparated−20 °C30 daysstable for the longest time tested% difference vs day 0 > RCV 18% (Brescia, LIAISON)Borai et al. 2020, J Clin Lab AnalDiscussion, paragraph 5; Table 2
serum, BD plain (red top) tubeStorageseparated−20 °C30 daysstable for the longest time tested% difference vs day 0 > TEa 46% (DEQAS)Borai et al. 2020, J Clin Lab AnalDiscussion, paragraph 6; Table 2
serum, BD plain (red top) tubeStorageseparated−20 °C30 daysstable for the longest time tested% difference vs day 0 > RCV 18% (Brescia, LIAISON)Borai et al. 2020, J Clin Lab AnalDiscussion, paragraph 5; Table 2
serum, BD plain (red top) tubeStorageseparated−20 °C30 daysstable for the longest time tested% difference vs day 0 > TEa 46% (DEQAS)Borai et al. 2020, J Clin Lab AnalDiscussion, paragraph 6; Table 2
serum, not statedStorageseparated−80 °C2 weeksthen changed, changed by 7 monthsFriedman test P < 0.05 with Bonferroni-adjusted post hoc vs baseline (#)Ko et al. 2025, Sci RepTable 1, overall, −80 °C
serum, not statedStorageseparated−80 °C0 weekschanged at the first time point, changed by 2 weeksFriedman test P < 0.05 with Bonferroni-adjusted post hoc vs baseline (#)Ko et al. 2025, Sci RepTable 2, overall, −80 °C
serum, BD Vacutainer serum separator tube with clot activator (8.5 mL)Storageseparated−80 °C3 monthsstable for the longest time testedFriedman repeated measures ANOVA vs zero time, P <= 0.05Colak et al. 2013, Biochem Med (Zagreb)Table 2
serum, BD Vacutainer SST (5 mL, REF 366566)Storageseparated−80 °C48 hstable for the longest time testedpaired t-test vs immediate analysis, P < 0.05Yu et al. 2016, PLoS OneDiscussion, paragraph 3
serum, VACUETTE no-additive (4 mL, REF 454001)Storageseparated−80 °C48 hstable for the longest time testedpaired t-test vs immediate analysis, P < 0.05Yu et al. 2016, PLoS OneDiscussion, paragraph 3
serum, VACUETTE serum clot activator and gel (REF 454067)Storageseparated−80 °C48 hstable for the longest time testedpaired t-test vs immediate analysis, P < 0.05Yu et al. 2016, PLoS OneDiscussion, paragraph 3
serum, IMPROVEVACUTER serum clot activator and gel (REF U222538)Storageseparated−80 °C0 hchanged at the first time point, changed by 48 hpaired t-test vs immediate analysis, P < 0.05Yu et al. 2016, PLoS OneResults, 'Evaluation of the effects of storage temperature', paragraph 1
serum, IMPROVEVACUTER serum clot activator and gel (REF U222538)Storageseparated−80 °C48 hstable for the longest time testedno significant difference vs immediate analysis (test not named for this set; P < 0.05)Yu et al. 2016, PLoS OneDiscussion, paragraph 3
serum, BD Vacutainer serum separator tube with clot activator (8.5 mL)Storageseparated2-8 °C24 hstable for the longest time testedFriedman repeated measures ANOVA vs zero time, P <= 0.05Colak et al. 2013, Biochem Med (Zagreb)Table 1
serum, BD Vacutainer SST (5 mL, REF 366566)Storageseparated2-8 °C48 hstable for the longest time testedpaired t-test vs immediate analysis, P < 0.05Yu et al. 2016, PLoS OneResults, 'Evaluation of the effects of storage temperature', paragraph 1
serum, IMPROVEVACUTER serum clot activator and gel (REF U222538)Storageseparated2-8 °C48 hstable for the longest time testedpaired t-test vs immediate analysis, P < 0.05Yu et al. 2016, PLoS OneResults, 'Evaluation of the effects of storage temperature', paragraph 1
serum, VACUETTE no-additive (4 mL, REF 454001)Storageseparated2-8 °C48 hstable for the longest time testedpaired t-test vs immediate analysis, P < 0.05Yu et al. 2016, PLoS OneResults, 'Evaluation of the effects of storage temperature', paragraph 1
serum, VACUETTE serum clot activator and gel (REF 454067)Storageseparated2-8 °C48 hstable for the longest time testedpaired t-test vs immediate analysis, P < 0.05Yu et al. 2016, PLoS OneResults, 'Evaluation of the effects of storage temperature', paragraph 1
serum, IMPROVEVACUTER serum clot activator and gel (REF U222538)Storageseparated2-8 °C48 hstable for the longest time testedno significant difference vs immediate analysis (test not named for this set; P < 0.05)Yu et al. 2016, PLoS OneDiscussion, paragraph 3
serum, BD Vacutainer serum separator tube with clot activator (8.5 mL)StorageseparatedRT4 hstable for the longest time testedFriedman repeated measures ANOVA vs zero time, P <= 0.05Colak et al. 2013, Biochem Med (Zagreb)Table 1
whole-blood-clotted, BD Vacutainer trace element-free serum tubeBefore spinninguncentrifugedcold packs in insulated box; 9 (2) °C on arrival24 hstable for the longest time tested% difference within desirable quality specification (from intra- and interindividual variation)Jones et al. 2024, J NutrTable 2, row '25-Hydroxyvitamin D, nmol/L'; Table 1
whole-blood-clotted, BD Vacutainer SSTBefore spinninguncentrifuged> 30 °C (logger 38-40 °C)6 hthen changed, changed by 12 hmean % deviation from baseline within total change limit TCL = sqrt((2.77 CVa)^2 + (0.5 CVb)^2), CVb from Ricos (Table 2)Abraham et al. 2019, Biochem Med (Zagreb)Results, paragraph 1 (Figure 3b)
whole-blood-clotted, Sarstedt Monovette serumBefore spinninguncentrifugedchilled in transit (insulated box, 2 frozen cool packs; temperature not recorded)24 hstable for the longest time tested% difference within desirable quality specification for bias (Fraser, from intra- and interindividual variation)Jones et al. 2021, J NutrTable 2, row '25-hydroxyvitamin D', column Serum
whole-blood-edta, Sarstedt Monovette EDTABefore spinninguncentrifugedchilled in transit (insulated box, 2 frozen cool packs; temperature not recorded)24 hstable for the longest time tested% difference within desirable quality specification for bias (Fraser, from intra- and interindividual variation)Jones et al. 2021, J NutrTable 2, row '25-hydroxyvitamin D', column EDTA
whole-blood-heparin, Sarstedt Monovette lithium heparin, trace-element freeBefore spinninguncentrifugedchilled in transit (insulated box, 2 frozen cool packs; temperature not recorded)24 hstable for the longest time tested% difference within desirable quality specification for bias (Fraser, from intra- and interindividual variation)Jones et al. 2021, J NutrTable 2, row '25-hydroxyvitamin D', column LH
4-Pyridoxic acid4 findings
Specimen, tubeConditionTemperatureStable forCriterionStudy
whole-blood-clotted, Sarstedt Monovette serumBefore spinninguncentrifugedchilled in transit (insulated box, 2 frozen cool packs; temperature not recorded)24 hstable for the longest time tested% difference within desirable quality specification for bias (Fraser, from intra- and interindividual variation)Jones et al. 2021, J NutrTable 2, row 'PA (vitamin B-6)', column Serum
whole-blood-edta, Sarstedt Monovette EDTABefore spinninguncentrifugedchilled in transit (insulated box, 2 frozen cool packs; temperature not recorded)24 hstable for the longest time tested% difference within desirable quality specification for bias (Fraser, from intra- and interindividual variation)Jones et al. 2021, J NutrTable 2, row 'PA (vitamin B-6)', column EDTA
whole-blood-heparin, BD Vacutainer lithium heparinBefore spinninguncentrifugedcold packs in insulated box; 9 (2) °C on arrival24 hstable for the longest time tested% difference within desirable quality specification (from intra- and interindividual variation)Jones et al. 2024, J NutrTable 2, row 'Vitamin B6 (PA), nmol/L'; Table 1
whole-blood-heparin, Sarstedt Monovette lithium heparin, trace-element freeBefore spinninguncentrifugedchilled in transit (insulated box, 2 frozen cool packs; temperature not recorded)24 hstable for the longest time tested% difference within desirable quality specification for bias (Fraser, from intra- and interindividual variation)Jones et al. 2021, J NutrTable 2, row 'PA (vitamin B-6)', column LH
Activated Partial Thromboplastin Time10 findings
Specimen, tubeConditionTemperatureStable forCriterionStudy
plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquotsFreeze–thawseparated−80 °C1 cyclelimit stated by the authorsmax allowable mean deviation from t0: <100 s: ±10%; ≥100 s: ±20%Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability'
plasma-citrate, BD Vacutainer 2.7 mL, 3.2% buffered sodium citrate; PPP aliquoted into microcentrifuge tubesStorageseparated−20 °C36 hstable for the longest time testedchange of more than ±10% of mean from baselinePatil et al. 2022, CureusTable 2
plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquotsStorageseparated−80 °C11 monthslimit stated by the authorsmax allowable mean deviation from t0: <100 s: ±10%; ≥100 s: ±20%Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability'
plasma-citrate, BD 0.109 M sodium citrate (9:1); plasma aliquoted into nonactivating Eppendorf tubesStorageseparated2-8 °C12 hthen changed, changed by 24 hmean % change from baseline >10%, or >25% of samples with individual change >10% (van Geest-Daalderop)Feng et al. 2014, Sci RepTable 2, 'Acceptable Time (h)'
plasma-citrate, BD Vacutainer 2.7 mL, 3.2% buffered sodium citrate; PPP aliquoted into microcentrifuge tubesStorageseparated2-8 °C0 hchanged at the first time point, changed by 12 hchange of more than ±10% of mean from baselinePatil et al. 2022, CureusTable 2
plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquotsStorageseparatedRT2 hlimit stated by the authorsmax allowable mean deviation from t0: <100 s: ±10%; ≥100 s: ±20%Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability'
plasma-citrate, BD 0.109 M sodium citrate (9:1); plasma aliquoted into nonactivating Eppendorf tubesStorageseparatedRT8 hthen changed, changed by 12 hmean % change from baseline >10%, or >25% of samples with individual change >10% (van Geest-Daalderop)Feng et al. 2014, Sci RepTable 2, 'Acceptable Time (h)'
plasma-citrate, BD Vacutainer 2.7 mL, 3.2% buffered sodium citrate; PPP aliquoted into microcentrifuge tubesStorageseparatedRT0 hchanged at the first time point, changed by 12 hchange of more than ±10% of mean from baselinePatil et al. 2022, CureusTable 2
whole-blood-citrate, Greiner Vacuette sodium citrate, standard fill (blood–air interface)Before spinninguncentrifugedRT6 hthen changed, changed by 24 hmaximum acceptable bias = 0.375 × total biological variation (EFLM BV database, Model 2)Storm et al. 2026, Adv HematolTable 2, 'Standard tubes' row 'APTT'
whole-blood-citrate, Greiner Vacuette sodium citrate, completely filled (no air)Before spinninguncentrifugedRT6 hthen changed, changed by 24 hmaximum acceptable bias = 0.375 × total biological variation (EFLM BV database, Model 2)Storm et al. 2026, Adv HematolTable 2, 'Full tubes' row 'APTT'
Adenosine deaminase2 findings
Specimen, tubeConditionTemperatureStable forCriterionStudy
plasma-li-heparin, Sarstedt lithium heparin, no gel separator (primary tube)Storagecentrifuged-on-cells2-8 °C48 hstable for the longest time testedmean % difference vs time 0 exceeds total limit of change, TLC = ±√((1.65·CVa)² + (0.5·CVb)²) (EFLM BV), here 25.1%; stability limit from instability-equation regressionIllana et al. 2024, Biochem Med (Zagreb)Table 1, row 'ADA'
plasma-li-heparin, Sarstedt lithium heparin, no gel separator; plasma aliquotStorageseparated2-8 °C96 hstable for the longest time testedmean % difference vs time 0 exceeds total limit of change, TLC = ±√((1.65·CVa)² + (0.5·CVb)²) (EFLM BV), here 25.1%; stability limit from instability-equation regressionIllana et al. 2024, Biochem Med (Zagreb)Table 2, row 'ADA'
Adrenocorticotropic Hormone3 findings
Specimen, tubeConditionTemperatureStable forCriterionStudy
whole-blood-edta, BD K2-EDTABefore spinninguncentrifuged2-8 °C8 hstable for the longest time testedmean % deviation vs T0 > total change limit (TCL) 7.5%Fraissinet et al. 2024, Biochem Med (Zagreb)Table 1; Results paragraph 1
whole-blood-edta, BD K3-EDTA + aprotininBefore spinninguncentrifuged2-8 °C4 hthen changed, changed by 6 hmean % deviation vs T0 > total change limit (TCL) 7.5%Fraissinet et al. 2024, Biochem Med (Zagreb)Table 1; Results paragraph 1
whole-blood-edta, BD K3-EDTA + aprotininBefore spinninguncentrifugedRT2 hthen changed, changed by 4 hmean % deviation vs T0 > total change limit (TCL) 7.5%Fraissinet et al. 2024, Biochem Med (Zagreb)Table 1; Results paragraph 1
Alanine Aminotransferase (ALT)20 findings
Specimen, tubeConditionTemperatureStable forCriterionStudy
plasma-li-heparin, Sarstedt lithium heparin, no gel separator (primary tube)Storagecentrifuged-on-cells2-8 °C48 hstable for the longest time testedmean % difference vs time 0 exceeds total limit of change, TLC = ±√((1.65·CVa)² + (0.5·CVb)²) (EFLM BV), here 15.3%; stability limit from instability-equation regressionIllana et al. 2024, Biochem Med (Zagreb)Table 1, row 'ALT'
plasma-li-heparin, Sarstedt lithium heparin, no gel separator; plasma aliquotStorageseparated2-8 °C96 hstable for the longest time testedmean % difference vs time 0 exceeds total limit of change, TLC = ±√((1.65·CVa)² + (0.5·CVb)²) (EFLM BV), here 15.3%; stability limit from instability-equation regressionIllana et al. 2024, Biochem Med (Zagreb)Table 2, row 'ALT'
plasma-li-heparin, BD Vacutainer lithium heparin 6 mL (regular)Storagenot-stated2-8 °C72 hstable for the longest time testedmean % difference vs T0 exceeds total change limit (TCL), here ±21.75%Dupuy et al. 2018, Biochem Med (Zagreb)Table 2, BD lithium heparin tubes, row 'ALT, U/L'
plasma-li-heparin, BD Barricor lithium heparin 5.5 mL (mechanical separator)Storagecentrifuged-on-gel2-8 °C72 hstable for the longest time testedmean % difference vs T0 exceeds total change limit (TCL), here ±21.75%Dupuy et al. 2018, Biochem Med (Zagreb)Table 2, BD Barricor lithium heparin tubes, row 'ALT, U/L'
plasma-li-heparin, BD Barricor 4.5 mL (lithium heparin, mechanical separator)Storagecentrifuged-on-gel2-8 °C7 daysstable for the longest time testedmean % difference vs time 0 exceeds desirable imprecision based on biological variation (EFLM/Westgard), here 5.2%Mandić et al. 2020, Biochem Med (Zagreb)Tables 3 and 4, row 'ALT, U/L', BD Barricor columns
serum, BD Vacutainer 8.5 mL plastic serum tube; pooled serum aliquoted into 1.5 mL Eppendorf tubesFreeze–thawseparated−20 °C10 cyclesstable for the longest time testedunstable only if both P < 0.005 (Bonferroni) and % change from T0 > desirable bias (Westgard/Ricos), here 11.4%Cuhadar et al. 2013, Biochem Med (Zagreb)Table 2, row 'ALT (U/L)', columns T1d-T10d
serum, BD Vacutainer 8.5 mL plastic serum tube; pooled serum aliquoted into 1.5 mL Eppendorf tubesStorageseparated−20 °C3 monthsstable for the longest time testedunstable only if both P < 0.013 (Bonferroni) and % change from T0 > desirable bias (Westgard/Ricos), here 11.4%Cuhadar et al. 2013, Biochem Med (Zagreb)Table 2, row 'ALT (U/L)', columns T1m-T3m
serum, BD Vacutainer serum (plain), 7.5 mLStorageseparated−20 °C30 daysstable for the longest time testedclinically significant change: mean outside significant change limit (SCL = T1d mean ± 3.0 usual SD from 2 months QC)Kachhawa et al. 2017, J Lab PhysiciansDiscussion, 'Enzymes' (Table 2 image-only)
serum, Vacuette Tube Serum Gel Separator 8 mL (Greiner, 455071)Storagecentrifuged-on-gel2-8 °C1 weekstable for the longest time tested% bias of median vs baseline exceeds desirable bias (Westgard/Ricos BV), here 12.0%Cuhadar et al. 2012, Biochem Med (Zagreb)Table 3, row 'Alanine aminotransferase', gel 4°C
serum, Vacuette Tube Serum No Gel 9 mL (Greiner, 455092), plainStoragecentrifuged-on-cells2-8 °C1 weekstable for the longest time tested% bias of median vs baseline exceeds desirable bias (Westgard/Ricos BV), here 12.0%Cuhadar et al. 2012, Biochem Med (Zagreb)Table 3, row 'Alanine aminotransferase', w/o gel 4°C
serum, BD Vacutainer RST 4.0 mL (thrombin clot activator, gel)Storagecentrifuged-on-gel2-8 °C24 hstable for the longest time testedclinically significant if mean bias vs t0 exceeds desirable bias (biological variation), here 3.0%Kocijancic et al. 2014, Biochem Med (Zagreb)Table 3, row 'ALT'
serum, BD Vacutainer SST II Advance 5.0 mL (clot activator, gel)Storagecentrifuged-on-gel2-8 °C24 hstable for the longest time testedclinically significant if mean bias vs t0 exceeds desirable bias (biological variation), here 3.0%Kocijancic et al. 2014, Biochem Med (Zagreb)Table 4, row 'ALT'
serum, BD Vacutainer RST 5 mL (thrombin clot activator, gel)Storagecentrifuged-on-gel2-8 °C7 daysstable for the longest time testedmean % difference vs time 0 exceeds desirable imprecision based on biological variation (EFLM/Westgard), here 5.2%Mandić et al. 2020, Biochem Med (Zagreb)Tables 3 and 4, row 'ALT, U/L', BD RST columns
serum, Vacuette Tube Serum Gel Separator 8 mL (Greiner, 455071)Storagecentrifuged-on-gelRT1 weekstable for the longest time tested% bias of median vs baseline exceeds desirable bias (Westgard/Ricos BV), here 12.0%Cuhadar et al. 2012, Biochem Med (Zagreb)Table 3, row 'Alanine aminotransferase', gel 24°C
serum, Vacuette Tube Serum No Gel 9 mL (Greiner, 455092), plainStoragecentrifuged-on-cellsRT72 hthen changed, changed by 1 week% bias of median vs baseline exceeds desirable bias (Westgard/Ricos BV), here 12.0%Cuhadar et al. 2012, Biochem Med (Zagreb)Table 3, row 'Alanine aminotransferase', w/o gel 24°C
serum, BD Vacutainer SST (SKU 367986), gelStoragecentrifuged-on-gel30 °C (±1 °C)24 hthen changed, changed by 72 hmean bias vs paired baseline within CLIA acceptance limits AND >2/3 of individual results within limitsIacovetti et al. 2026, Pract Lab MedResults, 'Analyte stability after storage at 30 °C', paragraph 1 (Table 1 image-only)
serum, BD SST Microtainer (SKU 395967), gel; filled from venous bloodStoragecentrifuged-on-gel0 °C72 hstable for the longest time testedmean bias vs paired baseline within CLIA acceptance limits AND >2/3 of individual results within limitsIacovetti et al. 2026, Pract Lab MedResults, 'Analyte stability after separation and storage in microvolume tubes at 0 °C' (Table 2 image-only)
whole-blood-clotted, BD Vacutainer SST (SKU 367986), gelBefore spinninguncentrifuged30 °C (±1 °C)0 hchanged at the first time point, changed by 24 hmean bias vs paired baseline within CLIA acceptance limits AND >2/3 of individual results within limitsIacovetti et al. 2026, Pract Lab MedResults, 'Analyte stability after storage at 30 °C', paragraph 2 (Table 1 image-only)
whole-blood-heparin, BD Vacutainer lithium heparin, no gel (ref 368884)Before spinninguncentrifugedRT10 hstable for the longest time tested|mean % deviation 10 h vs baseline| > laboratory bias goal (Table II; mostly biological-variation based, some empirically modified)Henriksen et al. 2014, Scand J Clin Lab InvestTable II, row 'ALAT'
whole-blood-heparin, BD Vacutainer lithium heparin, no gel (ref 368884)Before spinninguncentrifugedRT10 hstable for the longest time testedCV of paired baseline/10 h results > goal CV (Table II)Henriksen et al. 2014, Scand J Clin Lab InvestTable II, row 'ALAT'
Albumin24 findings
Specimen, tubeConditionTemperatureStable forCriterionStudy
plasma-edta, BD Vacutainer EDTA, 4 mLFreeze–thawseparated−20 °C0 cycleschanged at the first time point, changed by 1 cyclessignificant change vs day 1 (repeated-measures ANOVA, Tukey HSD post hoc, p<0.05)Gojković et al. 2024, J Med BiochemResults paragraph 3 (EDTA -20°C, Figure 3)
plasma-edta, BD Vacutainer EDTA, 4 mLStorageseparated2-8 °C0 dayschanged at the first time point, changed by 1 dayssignificant change vs day 1 (repeated-measures ANOVA, Tukey HSD post hoc, p<0.05)Gojković et al. 2024, J Med BiochemResults paragraph 1 (EDTA+4°C, Figure 1)
plasma-fluoride, BD Vacutainer sodium fluoride/potassium oxalate, 4 mLFreeze–thawseparated−20 °C0 cycleschanged at the first time point, changed by 1 cyclessignificant change vs day 1 (repeated-measures ANOVA, Tukey HSD post hoc, p<0.05)Gojković et al. 2024, J Med BiochemResults paragraph 4 (NaF -20°C, Figure 4)
plasma-fluoride, BD Vacutainer sodium fluoride/potassium oxalate, 4 mLStorageseparated2-8 °C0 dayschanged at the first time point, changed by 1 dayssignificant change vs day 1 (repeated-measures ANOVA, Tukey HSD post hoc, p<0.05)Gojković et al. 2024, J Med BiochemResults paragraph 2 (NaF+4°C, Figure 2)
plasma-li-heparin, Sarstedt lithium heparin, no gel separator (primary tube)Storagecentrifuged-on-cells2-8 °C48 hstable for the longest time testedmean % difference vs time 0 exceeds total limit of change, TLC = ±√((1.65·CVa)² + (0.5·CVb)²) (EFLM BV), here 4.7%; stability limit from instability-equation regressionIllana et al. 2024, Biochem Med (Zagreb)Table 1, row 'Alb'
plasma-li-heparin, Sarstedt lithium heparin, no gel separator; plasma aliquotStorageseparated2-8 °C96 hstable for the longest time testedmean % difference vs time 0 exceeds total limit of change, TLC = ±√((1.65·CVa)² + (0.5·CVb)²) (EFLM BV), here 4.7%; stability limit from instability-equation regressionIllana et al. 2024, Biochem Med (Zagreb)Table 2, row 'Alb'
plasma-li-heparin, BD Barricor 4.5 mL (lithium heparin, mechanical separator)Storagecentrifuged-on-gel2-8 °C24 hthen changed, changed by 7 daysmean % difference vs time 0 exceeds desirable imprecision based on biological variation (EFLM/Westgard), here 1.3%Mandić et al. 2020, Biochem Med (Zagreb)Tables 3 and 4, row 'Albumin, g/L', BD Barricor columns
serum, BD Vacutainer 8.5 mL plastic serum tube; pooled serum aliquoted into 1.5 mL Eppendorf tubesFreeze–thawseparated−20 °C7 cyclesthen changed, changed by 8 cyclesunstable only if both P < 0.005 (Bonferroni) and % change from T0 > desirable bias (Westgard/Ricos), here 1.3%Cuhadar et al. 2013, Biochem Med (Zagreb)Table 2, row 'Albumin (g/L)', columns T1d-T10d
serum, BD Vacutainer SSTFreeze–thawseparated−70 °C, 1 month; thawed at room temperature1 cyclestable for the longest time testedmean % change (bias) from fresh baseline within total allowable error (TEa, desirable bias)Abraham et al. 2021, J Lab PhysiciansTable 2 and Table 3, row 'Serum albumin (mg/dL)'
serum, BD Vacutainer 8.5 mL plastic serum tube; pooled serum aliquoted into 1.5 mL Eppendorf tubesStorageseparated−20 °C2 monthsthen changed, changed by 3 monthsunstable only if both P < 0.013 (Bonferroni) and % change from T0 > desirable bias (Westgard/Ricos), here 1.3%Cuhadar et al. 2013, Biochem Med (Zagreb)Table 2, row 'Albumin (g/L)', columns T1m-T3m
serum, BD Vacutainer serum (plain), 7.5 mLStorageseparated−20 °C30 daysstable for the longest time testedclinically significant change: mean outside significant change limit (SCL = T1d mean ± 3.0 usual SD from 2 months QC)Kachhawa et al. 2017, J Lab PhysiciansDiscussion, 'Protein' (Table 2 image-only)
serum, Vacuette Tube Serum Gel Separator 8 mL (Greiner, 455071)Storagecentrifuged-on-gel2-8 °C24 hthen changed, changed by 30 h% bias of median vs baseline exceeds desirable bias (Westgard/Ricos BV), here 1.3%Cuhadar et al. 2012, Biochem Med (Zagreb)Table 3, row 'Albumin', gel 4°C
serum, Vacuette Tube Serum No Gel 9 mL (Greiner, 455092), plainStoragecentrifuged-on-cells2-8 °C6 hthen changed, changed by 12 h% bias of median vs baseline exceeds desirable bias (Westgard/Ricos BV), here 1.3%Cuhadar et al. 2012, Biochem Med (Zagreb)Table 3, row 'Albumin', w/o gel 4°C
serum, BD Vacutainer RST 5 mL (thrombin clot activator, gel)Storagecentrifuged-on-gel2-8 °C7 daysstable for the longest time testedmean % difference vs time 0 exceeds desirable imprecision based on biological variation (EFLM/Westgard), here 1.3%Mandić et al. 2020, Biochem Med (Zagreb)Tables 3 and 4, row 'Albumin, g/L', BD RST columns
serum, Vacuette Tube Serum Gel Separator 8 mL (Greiner, 455071)Storagecentrifuged-on-gelRT0 hchanged at the first time point, changed by 6 h% bias of median vs baseline exceeds desirable bias (Westgard/Ricos BV), here 1.3%Cuhadar et al. 2012, Biochem Med (Zagreb)Table 3, row 'Albumin', gel 24°C
serum, Vacuette Tube Serum No Gel 9 mL (Greiner, 455092), plainStoragecentrifuged-on-cellsRT0 hchanged at the first time point, changed by 6 h% bias of median vs baseline exceeds desirable bias (Westgard/Ricos BV), here 1.3%Cuhadar et al. 2012, Biochem Med (Zagreb)Table 3, row 'Albumin', w/o gel 24°C
serum, BD Vacutainer SST (SKU 367986), gelStoragecentrifuged-on-gel30 °C (±1 °C)72 hstable for the longest time testedmean bias vs paired baseline within CLIA acceptance limits AND >2/3 of individual results within limitsIacovetti et al. 2026, Pract Lab MedResults, 'Analyte stability after storage at 30 °C', paragraph 1 (Table 1 image-only)
serum, BD SST Microtainer (SKU 395967), gel; filled from venous bloodStoragecentrifuged-on-gel0 °C72 hstable for the longest time testedmean bias vs paired baseline within CLIA acceptance limits AND >2/3 of individual results within limitsIacovetti et al. 2026, Pract Lab MedResults, 'Analyte stability after separation and storage in microvolume tubes at 0 °C' (Table 2 image-only)
whole-blood-clotted, BD Vacutainer SSTBefore spinninguncentrifugedRT24 hstable for the longest time testedmean % deviation from baseline within total change limit TCL = sqrt((2.77 CVa)^2 + (0.5 CVb)^2), CVb from Ricos (Table 2)Abraham et al. 2019, Biochem Med (Zagreb)Discussion, paragraph 1 (Results paragraphs 1-3; Figure 3)
whole-blood-clotted, BD Vacutainer SST (SKU 367986), gelBefore spinninguncentrifuged30 °C (±1 °C)72 hstable for the longest time testedmean bias vs paired baseline within CLIA acceptance limits AND >2/3 of individual results within limitsIacovetti et al. 2026, Pract Lab MedResults, 'Analyte stability after storage at 30 °C', paragraph 2 (Table 1 image-only)
whole-blood-clotted, BD Vacutainer SSTBefore spinninguncentrifuged22-30 °C (logger 24-26 °C)24 hstable for the longest time testedmean % deviation from baseline within total change limit TCL = sqrt((2.77 CVa)^2 + (0.5 CVb)^2), CVb from Ricos (Table 2)Abraham et al. 2019, Biochem Med (Zagreb)Discussion, paragraph 1 (Results paragraphs 1-3; Figure 3)
whole-blood-clotted, BD Vacutainer SSTBefore spinninguncentrifuged> 30 °C (logger 38-40 °C)24 hstable for the longest time testedmean % deviation from baseline within total change limit TCL = sqrt((2.77 CVa)^2 + (0.5 CVb)^2), CVb from Ricos (Table 2)Abraham et al. 2019, Biochem Med (Zagreb)Discussion, paragraph 1 (Results paragraphs 1-3; Figure 3)
whole-blood-heparin, BD Vacutainer lithium heparin, no gel (ref 368884)Before spinninguncentrifugedRT10 hstable for the longest time tested|mean % deviation 10 h vs baseline| > laboratory bias goal (Table II; mostly biological-variation based, some empirically modified)Henriksen et al. 2014, Scand J Clin Lab InvestTable II, row 'ALB'
whole-blood-heparin, BD Vacutainer lithium heparin, no gel (ref 368884)Before spinninguncentrifugedRT10 hstable for the longest time testedCV of paired baseline/10 h results > goal CV (Table II)Henriksen et al. 2014, Scand J Clin Lab InvestTable II, row 'ALB'
Alkaline Phosphatase (ALP)16 findings
Specimen, tubeConditionTemperatureStable forCriterionStudy
plasma-li-heparin, Sarstedt lithium heparin, no gel separator (primary tube)Storagecentrifuged-on-cells2-8 °C48 hstable for the longest time testedmean % difference vs time 0 exceeds total limit of change, TLC = ±√((1.65·CVa)² + (0.5·CVb)²) (EFLM BV), here 10.1%; stability limit from instability-equation regressionIllana et al. 2024, Biochem Med (Zagreb)Table 1, row 'ALP'
plasma-li-heparin, Sarstedt lithium heparin, no gel separator; plasma aliquotStorageseparated2-8 °C96 hstable for the longest time testedmean % difference vs time 0 exceeds total limit of change, TLC = ±√((1.65·CVa)² + (0.5·CVb)²) (EFLM BV), here 10.1%; stability limit from instability-equation regressionIllana et al. 2024, Biochem Med (Zagreb)Table 2, row 'ALP'
plasma-li-heparin, BD Vacutainer lithium heparin 6 mL (regular)Storagenot-stated2-8 °C72 hstable for the longest time testedmean % difference vs T0 exceeds total change limit (TCL), here ±14.34%Dupuy et al. 2018, Biochem Med (Zagreb)Table 2, BD lithium heparin tubes, row 'ALP, U/L'
plasma-li-heparin, BD Barricor lithium heparin 5.5 mL (mechanical separator)Storagecentrifuged-on-gel2-8 °C72 hstable for the longest time testedmean % difference vs T0 exceeds total change limit (TCL), here ±14.34%Dupuy et al. 2018, Biochem Med (Zagreb)Table 2, BD Barricor lithium heparin tubes, row 'ALP, U/L'
plasma-li-heparin, BD Barricor 4.5 mL (lithium heparin, mechanical separator)Storagecentrifuged-on-gel2-8 °C24 hthen changed, changed by 7 daysmean % difference vs time 0 exceeds desirable imprecision based on biological variation (EFLM/Westgard), here 2.7%Mandić et al. 2020, Biochem Med (Zagreb)Tables 3 and 4, row 'ALP, U/L', BD Barricor columns
serum, BD Vacutainer serum (plain), 7.5 mLStorageseparated−20 °C30 daysstable for the longest time testedclinically significant change: mean outside significant change limit (SCL = T1d mean ± 3.0 usual SD from 2 months QC)Kachhawa et al. 2017, J Lab PhysiciansDiscussion, 'Enzymes' (Table 2 image-only)
serum, Vacuette Tube Serum Gel Separator 8 mL (Greiner, 455071)Storagecentrifuged-on-gel2-8 °C1 weekstable for the longest time tested% bias of median vs baseline exceeds desirable bias (Westgard/Ricos BV), here 6.4%Cuhadar et al. 2012, Biochem Med (Zagreb)Table 3, row 'Alkaline phosphatase', gel 4°C
serum, Vacuette Tube Serum No Gel 9 mL (Greiner, 455092), plainStoragecentrifuged-on-cells2-8 °C1 weekstable for the longest time tested% bias of median vs baseline exceeds desirable bias (Westgard/Ricos BV), here 6.4%Cuhadar et al. 2012, Biochem Med (Zagreb)Table 3, row 'Alkaline phosphatase', w/o gel 4°C
serum, BD Vacutainer RST 5 mL (thrombin clot activator, gel)Storagecentrifuged-on-gel2-8 °C7 daysstable for the longest time testedmean % difference vs time 0 exceeds desirable imprecision based on biological variation (EFLM/Westgard), here 2.7%Mandić et al. 2020, Biochem Med (Zagreb)Tables 3 and 4, row 'ALP, U/L', BD RST columns
serum, Vacuette Tube Serum Gel Separator 8 mL (Greiner, 455071)Storagecentrifuged-on-gelRT1 weekstable for the longest time tested% bias of median vs baseline exceeds desirable bias (Westgard/Ricos BV), here 6.4%Cuhadar et al. 2012, Biochem Med (Zagreb)Table 3, row 'Alkaline phosphatase', gel 24°C
serum, Vacuette Tube Serum No Gel 9 mL (Greiner, 455092), plainStoragecentrifuged-on-cellsRT1 weekstable for the longest time tested% bias of median vs baseline exceeds desirable bias (Westgard/Ricos BV), here 6.4%Cuhadar et al. 2012, Biochem Med (Zagreb)Table 3, row 'Alkaline phosphatase', w/o gel 24°C
serum, BD Vacutainer SST (SKU 367986), gelStoragecentrifuged-on-gel30 °C (±1 °C)72 hstable for the longest time testedmean bias vs paired baseline within CLIA acceptance limits AND >2/3 of individual results within limitsIacovetti et al. 2026, Pract Lab MedResults, 'Analyte stability after storage at 30 °C', paragraph 1 (Table 1 image-only)
serum, BD SST Microtainer (SKU 395967), gel; filled from venous bloodStoragecentrifuged-on-gel0 °C72 hstable for the longest time testedmean bias vs paired baseline within CLIA acceptance limits AND >2/3 of individual results within limitsIacovetti et al. 2026, Pract Lab MedResults, 'Analyte stability after separation and storage in microvolume tubes at 0 °C' (Table 2 image-only)
whole-blood-clotted, BD Vacutainer SST (SKU 367986), gelBefore spinninguncentrifuged30 °C (±1 °C)72 hstable for the longest time testedmean bias vs paired baseline within CLIA acceptance limits AND >2/3 of individual results within limitsIacovetti et al. 2026, Pract Lab MedResults, 'Analyte stability after storage at 30 °C', paragraph 2 (Table 1 image-only)
whole-blood-heparin, BD Vacutainer lithium heparin, no gel (ref 368884)Before spinninguncentrifugedRT10 hstable for the longest time tested|mean % deviation 10 h vs baseline| > laboratory bias goal (Table II; mostly biological-variation based, some empirically modified)Henriksen et al. 2014, Scand J Clin Lab InvestTable II, row 'BASP'
whole-blood-heparin, BD Vacutainer lithium heparin, no gel (ref 368884)Before spinninguncentrifugedRT10 hstable for the longest time testedCV of paired baseline/10 h results > goal CV (Table II)Henriksen et al. 2014, Scand J Clin Lab InvestTable II, row 'BASP'
Alpha-1 Antitrypsin (A1AT)2 findings
Specimen, tubeConditionTemperatureStable forCriterionStudy
whole-blood-heparin, BD Vacutainer lithium heparin, no gel (ref 368884)Before spinninguncentrifugedRT10 hstable for the longest time tested|mean % deviation 10 h vs baseline| > laboratory bias goal (Table II; mostly biological-variation based, some empirically modified)Henriksen et al. 2014, Scand J Clin Lab InvestTable II, row 'ATRYP'
whole-blood-heparin, BD Vacutainer lithium heparin, no gel (ref 368884)Before spinninguncentrifugedRT10 hstable for the longest time testedCV of paired baseline/10 h results > goal CV (Table II)Henriksen et al. 2014, Scand J Clin Lab InvestTable II, row 'ATRYP'
Alpha-carotene3 findings
Specimen, tubeConditionTemperatureStable forCriterionStudy
whole-blood-clotted, Sarstedt Monovette serumBefore spinninguncentrifugedchilled in transit (insulated box, 2 frozen cool packs; temperature not recorded)24 hstable for the longest time tested% difference within desirable quality specification for bias (Fraser, from intra- and interindividual variation)Jones et al. 2021, J NutrTable 2, row 'α-carotene', column Serum
whole-blood-edta, Sarstedt Monovette EDTABefore spinninguncentrifugedchilled in transit (insulated box, 2 frozen cool packs; temperature not recorded)24 hstable for the longest time tested% difference within desirable quality specification for bias (Fraser, from intra- and interindividual variation)Jones et al. 2021, J NutrTable 2, row 'α-carotene', column EDTA
whole-blood-heparin, Sarstedt Monovette lithium heparin, trace-element freeBefore spinninguncentrifugedchilled in transit (insulated box, 2 frozen cool packs; temperature not recorded)24 hstable for the longest time tested% difference within desirable quality specification for bias (Fraser, from intra- and interindividual variation)Jones et al. 2021, J NutrTable 2, row 'α-carotene', column LH
Amylase7 findings
Specimen, tubeConditionTemperatureStable forCriterionStudy
plasma-li-heparin, Sarstedt lithium heparin, no gel separator (primary tube)Storagecentrifuged-on-cells2-8 °C48 hstable for the longest time testedmean % difference vs time 0 exceeds total limit of change, TLC = ±√((1.65·CVa)² + (0.5·CVb)²) (EFLM BV), here 9.1%; stability limit from instability-equation regressionIllana et al. 2024, Biochem Med (Zagreb)Table 1, row 'AMY'
plasma-li-heparin, Sarstedt lithium heparin, no gel separator; plasma aliquotStorageseparated2-8 °C96 hstable for the longest time testedmean % difference vs time 0 exceeds total limit of change, TLC = ±√((1.65·CVa)² + (0.5·CVb)²) (EFLM BV), here 9.1%; stability limit from instability-equation regressionIllana et al. 2024, Biochem Med (Zagreb)Table 2, row 'AMY'
plasma-li-heparin, BD Barricor 4.5 mL (lithium heparin, mechanical separator)Storagecentrifuged-on-gel2-8 °C7 daysstable for the longest time testedmean % difference vs time 0 exceeds desirable imprecision based on biological variation (EFLM/Westgard), here 3.3%Mandić et al. 2020, Biochem Med (Zagreb)Tables 3 and 4, row 'Amylase, U/L', BD Barricor columns
serum, BD Vacutainer serum (plain), 7.5 mLStorageseparated−20 °C0 dayschanged at the first time point, changed by 7 daysclinically significant change: mean outside significant change limit (SCL = T1d mean ± 3.0 usual SD from 2 months QC)Kachhawa et al. 2017, J Lab PhysiciansDiscussion, 'Enzymes' (Table 2 image-only)
serum, BD Vacutainer RST 5 mL (thrombin clot activator, gel)Storagecentrifuged-on-gel2-8 °C7 daysstable for the longest time testedmean % difference vs time 0 exceeds desirable imprecision based on biological variation (EFLM/Westgard), here 3.3%Mandić et al. 2020, Biochem Med (Zagreb)Tables 3 and 4, row 'Amylase, U/L', BD RST columns
whole-blood-heparin, BD Vacutainer lithium heparin, no gel (ref 368884)Before spinninguncentrifugedRT10 hstable for the longest time tested|mean % deviation 10 h vs baseline| > laboratory bias goal (Table II; mostly biological-variation based, some empirically modified)Henriksen et al. 2014, Scand J Clin Lab InvestTable II, row 'AMYL'
whole-blood-heparin, BD Vacutainer lithium heparin, no gel (ref 368884)Before spinninguncentrifugedRT10 hstable for the longest time testedCV of paired baseline/10 h results > goal CV (Table II)Henriksen et al. 2014, Scand J Clin Lab InvestTable II, row 'AMYL'
Anti-Müllerian Hormone12 findings
Specimen, tubeConditionTemperatureStable forCriterionStudy
plasma-li-heparin, VACUETTE lithium heparin Plasma Sep (gel)Storageseparated−20 °C30 daysstable for the longest time testedequivalence test (TOST, paired) vs initial value, tolerance ±0.30 ng/mLVrzáková et al. 2023, Diagnostics (Basel)Results, paragraph 6; Table 1
serum, VACUETTE Z Serum Sep (gel)Freeze–thawseparated−80 °C5 cyclesstable for the longest time testedequivalence test (TOST, paired) vs initial value, tolerance ±0.30 ng/mLVrzáková et al. 2023, Diagnostics (Basel)Results, paragraph 3; Table 1
serum, VACUETTE Z Serum Sep (gel)Storageseparated−20 °C30 daysstable for the longest time testedequivalence test (TOST, paired) vs initial value, tolerance ±0.30 ng/mLVrzáková et al. 2023, Diagnostics (Basel)Results, paragraph 6; Table 1
serum, V-Tube serum separator gel tube (AB Medical)Storageseparated−20 °C0 hchanged at the first time point, changed by 48 hWilcoxon signed-rank test vs fresh control, p < 0.05Lee et al. 2023, Clin Exp Reprod MedTable 3, Rev-Gen II-AMH
serum, V-Tube serum separator gel tube (AB Medical)Storageseparated−20 °C0 yearschanged at the first time point, changed by 2 yearsWilcoxon signed-rank test vs fresh control, p < 0.05Lee et al. 2023, Clin Exp Reprod MedTable 3, Rev-Gen II-AMH
serum, V-Tube serum separator gel tube (AB Medical)Storageseparated−20 °C48 hthen changed, changed by 1 weekWilcoxon signed-rank test vs fresh control, p < 0.05Lee et al. 2023, Clin Exp Reprod MedTable 3, Access-AMH; Results paragraph 2
serum, V-Tube serum separator gel tube (AB Medical)Storageseparated−20 °C0 yearschanged at the first time point, changed by 2 yearsWilcoxon signed-rank test vs fresh control, p < 0.05Lee et al. 2023, Clin Exp Reprod MedTable 3, Access-AMH
serum, VACUETTE Z Serum Sep (gel)Storageseparated2-8 °C168 hstable for the longest time testedequivalence test (TOST, paired) vs initial value, tolerance ±0.30 ng/mLVrzáková et al. 2023, Diagnostics (Basel)Results, paragraph 5; Table 1
serum, V-Tube serum separator gel tube (AB Medical)Storageseparated2-8 °C0 hchanged at the first time point, changed by 48 hWilcoxon signed-rank test vs fresh control, p < 0.05Lee et al. 2023, Clin Exp Reprod MedTable 3, Rev-Gen II-AMH
serum, V-Tube serum separator gel tube (AB Medical)Storageseparated2-8 °C48 hthen changed, changed by 1 weekWilcoxon signed-rank test vs fresh control, p < 0.05Lee et al. 2023, Clin Exp Reprod MedTable 3, Access-AMH; Results paragraph 2
whole-blood-clotted, V-Tube serum separator gel tube (AB Medical)Before spinninguncentrifugedRT0 hchanged at the first time point, changed by 48 hWilcoxon signed-rank test vs fresh control, p < 0.05Lee et al. 2023, Clin Exp Reprod MedTable 4; Results paragraph 4
whole-blood-clotted, V-Tube serum separator gel tube (AB Medical)Before spinninguncentrifugedRT0 hchanged at the first time point, changed by 48 hWilcoxon signed-rank test vs fresh control, p < 0.05Lee et al. 2023, Clin Exp Reprod MedTable 4; Results paragraph 4
Anti-Xa activity5 findings
Specimen, tubeConditionTemperatureStable forCriterionStudy
plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquotsFreeze–thawseparated−20 °C1 cyclelimit stated by the authorsmax allowable mean deviation from t0: ≤0.45 IU/mL: ≤0.045 IU/mL;>0.45 IU/mL: ≤10%Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability'
plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquotsFreeze–thawseparated−80 °C3 cycleslimit stated by the authorsmax allowable mean deviation from t0: ≤0.45 IU/mL: ≤0.045 IU/mL;>0.45 IU/mL: ≤10%Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability'
plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquotsStorageseparated−20 °C14 dayslimit stated by the authorsmax allowable mean deviation from t0: ≤0.45 IU/mL: ≤0.045 IU/mL;>0.45 IU/mL: ≤10%Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability'
plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquotsStorageseparated−80 °C11 weekslimit stated by the authorsmax allowable mean deviation from t0: ≤0.45 IU/mL: ≤0.045 IU/mL;>0.45 IU/mL: ≤10%Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability'
plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquotsStorageseparatedRT2 hlimit stated by the authorsmax allowable mean deviation from t0: ≤0.45 IU/mL: ≤0.045 IU/mL;>0.45 IU/mL: ≤10%Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability'
Antithrombin III7 findings
Specimen, tubeConditionTemperatureStable forCriterionStudy
plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquotsFreeze–thawseparated−20 °C1 cyclelimit stated by the authorsmax allowable mean deviation from t0: ≤40.0%: ±4.0% (absolute);>40.0%: ±10% (relative)Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability'
plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquotsFreeze–thawseparated−80 °C3 cycleslimit stated by the authorsmax allowable mean deviation from t0: ≤40.0%: ±4.0% (absolute);>40.0%: ±10% (relative)Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability'
plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquotsStorageseparated−20 °C28 dayslimit stated by the authorsmax allowable mean deviation from t0: ≤40.0%: ±4.0% (absolute);>40.0%: ±10% (relative)Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability'
plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquotsStorageseparated−20 °C7 weekslimit stated by the authorsmax allowable mean deviation from t0: ≤40.0%: ±4.0% (absolute);>40.0%: ±10% (relative)Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability'
plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquotsStorageseparated−80 °C23 monthslimit stated by the authorsmax allowable mean deviation from t0: ≤40.0%: ±4.0% (absolute);>40.0%: ±10% (relative)Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability'
plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquotsStorageseparated2-8 °C14 dayslimit stated by the authorsmax allowable mean deviation from t0: ≤40.0%: ±4.0% (absolute);>40.0%: ±10% (relative)Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability'
plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquotsStorageseparatedRT2 dayslimit stated by the authorsmax allowable mean deviation from t0: ≤40.0%: ±4.0% (absolute);>40.0%: ±10% (relative)Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability'
aPTT (lupus-sensitive reagent)3 findings
Specimen, tubeConditionTemperatureStable forCriterionStudy
plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquotsFreeze–thawseparated−80 °C1 cyclelimit stated by the authorsmax allowable mean deviation from t0: <100 s: ±10%; ≥100 s: ±20%Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability'
plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquotsStorageseparated−80 °C5 monthslimit stated by the authorsmax allowable mean deviation from t0: <100 s: ±10%; ≥100 s: ±20%Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability'
plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquotsStorageseparatedRT2 hlimit stated by the authorsmax allowable mean deviation from t0: <100 s: ±10%; ≥100 s: ±20%Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability'
aPTT screen3 findings
Specimen, tubeConditionTemperatureStable forCriterionStudy
plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquotsFreeze–thawseparated−80 °C1 cyclelimit stated by the authorsmax allowable mean deviation from t0: <100 s: ±10%; ≥100 s: ±20%Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability'
plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquotsStorageseparated−80 °C11 weekslimit stated by the authorsmax allowable mean deviation from t0: <100 s: ±10%; ≥100 s: ±20%Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability'
plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquotsStorageseparatedRT2 hlimit stated by the authorsmax allowable mean deviation from t0: <100 s: ±10%; ≥100 s: ±20%Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability'
Aspartate Aminotransferase (AST)22 findings
Specimen, tubeConditionTemperatureStable forCriterionStudy
plasma-li-heparin, Sarstedt lithium heparin, no gel separator (primary tube)Storagecentrifuged-on-cells2-8 °C7.9 hthen changed, changed by 8 hmean % difference vs time 0 exceeds total limit of change, TLC = ±√((1.65·CVa)² + (0.5·CVb)²) (EFLM BV), here 8.7%; stability limit from instability-equation regressionIllana et al. 2024, Biochem Med (Zagreb)Table 1, row 'AST'
plasma-li-heparin, Sarstedt lithium heparin, no gel separator; plasma aliquotStorageseparated2-8 °C96 hstable for the longest time testedmean % difference vs time 0 exceeds total limit of change, TLC = ±√((1.65·CVa)² + (0.5·CVb)²) (EFLM BV), here 8.7%; stability limit from instability-equation regressionIllana et al. 2024, Biochem Med (Zagreb)Table 2, row 'AST'
plasma-li-heparin, BD Vacutainer lithium heparin 6 mL (regular)Storagenot-stated2-8 °C24 hthen changed, changed by 72 hmean % difference vs T0 exceeds total change limit (TCL), here ±10.83%Dupuy et al. 2018, Biochem Med (Zagreb)Table 2, BD lithium heparin tubes, row 'AST, U/L'
plasma-li-heparin, BD Barricor lithium heparin 5.5 mL (mechanical separator)Storagecentrifuged-on-gel2-8 °C24 hthen changed, changed by 72 hmean % difference vs T0 exceeds total change limit (TCL), here ±10.83%Dupuy et al. 2018, Biochem Med (Zagreb)Table 2, BD Barricor lithium heparin tubes, row 'AST, U/L'
plasma-li-heparin, BD Vacutainer LH PST II gel separator, 3.0 mL (REF 367374)Storagecentrifuged-on-gel2-8 °C24 hstable for the longest time testedmean % difference smaller than TEa (Ricos: 0.25 x (CVi^2 + CVg^2)^0.5 + 1.65 x 0.5 x CVi)Nonkes et al. 2019, Biochem Med (Zagreb)Table 1, row 'AST, U/L'
plasma-li-heparin, BD Vacutainer Barricor LH 3 mL, mechanical separatorStoragecentrifuged-on-gel2-8 °C80 hstable for the longest time testedbias vs aliquoted plasma reference tube > maximal allowable bias 5.4% (Ricos et al.); time to MAB from linear regression of bias on time to analysisGawria et al. 2020, J Clin Lab AnalTable 2, row 'AST', Barricor BD
plasma-li-heparin, BD Vacutainer PST LH 3 mL, gelStoragecentrifuged-on-gel2-8 °C0 hchanged at the first time point, changed by 3 hbias vs aliquoted plasma reference tube > maximal allowable bias 5.4% (Ricos et al.); time to MAB from linear regression of bias on time to analysisGawria et al. 2020, J Clin Lab AnalTable 2, row 'AST', Gel BD
plasma-li-heparin, Greiner Vacuette LH 3 mL Sep, gelStoragecentrifuged-on-gel2-8 °C0 hchanged at the first time point, changed by 3 hbias vs aliquoted plasma reference tube > maximal allowable bias 5.4% (Ricos et al.); time to MAB from linear regression of bias on time to analysisGawria et al. 2020, J Clin Lab AnalTable 2, row 'AST', Gel Greiner
plasma-li-heparin, BD Barricor 4.5 mL (lithium heparin, mechanical separator)Storagecentrifuged-on-gel2-8 °C0 hchanged at the first time point, changed by 24 hmean % difference vs time 0 exceeds desirable imprecision based on biological variation (EFLM/Westgard), here 4.75%Mandić et al. 2020, Biochem Med (Zagreb)Tables 3 and 4, row 'AST, U/L', BD Barricor columns
serum, BD Vacutainer 8.5 mL plastic serum tube; pooled serum aliquoted into 1.5 mL Eppendorf tubesFreeze–thawseparated−20 °C10 cyclesstable for the longest time testedunstable only if both P < 0.005 (Bonferroni) and % change from T0 > desirable bias (Westgard/Ricos), here 5.4%Cuhadar et al. 2013, Biochem Med (Zagreb)Table 2, row 'AST (U/L)', columns T1d-T10d
serum, BD Vacutainer 8.5 mL plastic serum tube; pooled serum aliquoted into 1.5 mL Eppendorf tubesStorageseparated−20 °C3 monthsstable for the longest time testedunstable only if both P < 0.013 (Bonferroni) and % change from T0 > desirable bias (Westgard/Ricos), here 5.4%Cuhadar et al. 2013, Biochem Med (Zagreb)Table 2, row 'AST (U/L)', columns T1m-T3m
serum, BD Vacutainer serum (plain), 7.5 mLStorageseparated−20 °C30 daysstable for the longest time testedclinically significant change: mean outside significant change limit (SCL = T1d mean ± 3.0 usual SD from 2 months QC)Kachhawa et al. 2017, J Lab PhysiciansDiscussion, 'Enzymes' (Table 2 image-only)
serum, Vacuette Tube Serum Gel Separator 8 mL (Greiner, 455071)Storagecentrifuged-on-gel2-8 °C0 hchanged at the first time point, changed by 6 h% bias of median vs baseline exceeds desirable bias (Westgard/Ricos BV), here 5.4%Cuhadar et al. 2012, Biochem Med (Zagreb)Table 3, row 'Aspartate aminotransferase', gel 4°C
serum, Vacuette Tube Serum No Gel 9 mL (Greiner, 455092), plainStoragecentrifuged-on-cells2-8 °C6 hthen changed, changed by 12 h% bias of median vs baseline exceeds desirable bias (Westgard/Ricos BV), here 5.4%Cuhadar et al. 2012, Biochem Med (Zagreb)Table 3, row 'Aspartate aminotransferase', w/o gel 4°C
serum, BD Vacutainer RST 4.0 mL (thrombin clot activator, gel)Storagecentrifuged-on-gel2-8 °C24 hstable for the longest time testedclinically significant if mean bias vs t0 exceeds desirable bias (biological variation), here 2.0%Kocijancic et al. 2014, Biochem Med (Zagreb)Table 3, row 'AST'
serum, BD Vacutainer SST II Advance 5.0 mL (clot activator, gel)Storagecentrifuged-on-gel2-8 °C24 hstable for the longest time testedclinically significant if mean bias vs t0 exceeds desirable bias (biological variation), here 2.0%Kocijancic et al. 2014, Biochem Med (Zagreb)Table 4, row 'AST'
serum, BD Vacutainer RST 5 mL (thrombin clot activator, gel)Storagecentrifuged-on-gel2-8 °C24 hthen changed, changed by 7 daysmean % difference vs time 0 exceeds desirable imprecision based on biological variation (EFLM/Westgard), here 4.75%Mandić et al. 2020, Biochem Med (Zagreb)Tables 3 and 4, row 'AST, U/L', BD RST columns
serum, Vacuette Tube Serum Gel Separator 8 mL (Greiner, 455071)Storagecentrifuged-on-gelRT0 hchanged at the first time point, changed by 6 h% bias of median vs baseline exceeds desirable bias (Westgard/Ricos BV), here 5.4%Cuhadar et al. 2012, Biochem Med (Zagreb)Table 3, row 'Aspartate aminotransferase', gel 24°C
serum, Vacuette Tube Serum No Gel 9 mL (Greiner, 455092), plainStoragecentrifuged-on-cellsRT6 hthen changed, changed by 12 h% bias of median vs baseline exceeds desirable bias (Westgard/Ricos BV), here 5.4%Cuhadar et al. 2012, Biochem Med (Zagreb)Table 3, row 'Aspartate aminotransferase', w/o gel 24°C
serum, BD Vacutainer SST (SKU 367986), gelStoragecentrifuged-on-gel30 °C (±1 °C)72 hstable for the longest time testedmean bias vs paired baseline within CLIA acceptance limits AND >2/3 of individual results within limitsIacovetti et al. 2026, Pract Lab MedResults, 'Analyte stability after storage at 30 °C', paragraph 1 (Table 1 image-only)
serum, BD SST Microtainer (SKU 395967), gel; filled from venous bloodStoragecentrifuged-on-gel0 °C72 hstable for the longest time testedmean bias vs paired baseline within CLIA acceptance limits AND >2/3 of individual results within limitsIacovetti et al. 2026, Pract Lab MedResults, 'Analyte stability after separation and storage in microvolume tubes at 0 °C' (Table 2 image-only)
whole-blood-clotted, BD Vacutainer SST (SKU 367986), gelBefore spinninguncentrifuged30 °C (±1 °C)24 hthen changed, changed by 72 hmean bias vs paired baseline within CLIA acceptance limits AND >2/3 of individual results within limitsIacovetti et al. 2026, Pract Lab MedResults, 'Analyte stability after storage at 30 °C', paragraph 2 (Table 1 image-only)
Basophils (Absolute)20 findings
Specimen, tubeConditionTemperatureStable forCriterionStudy
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C48 hstable for the longest time testedHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 4, row 'BA', XN-9000, 4 °C
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C48 hstable for the longest time testedBias% vs optimal bias (OP-Bias%, from biological variation; Table 1)Buoro et al. 2016, Rev Bras Hematol HemoterTable 4, row 'BA', XN-9000, 4 °C
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C48 hstable for the longest time testedHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 4, row 'BA', BC-6800, 4 °C
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C48 hstable for the longest time testedBias% vs optimal bias (OP-Bias%, from biological variation; Table 1)Buoro et al. 2016, Rev Bras Hematol HemoterTable 4, row 'BA', BC-6800, 4 °C
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C8 hstable for the longest time testedHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 5, row 'BA', XN-9000, 4 °C
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C4 hthen changed, changed by 8 hBias% vs optimal bias (OP-Bias%, from biological variation; Table 1)Buoro et al. 2016, Rev Bras Hematol HemoterTable 5, row 'BA', XN-9000, 4 °C
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C8 hstable for the longest time testedBias% vs critical difference (CD%, reference change value; Table 1)Buoro et al. 2016, Rev Bras Hematol HemoterTable 5, row 'BA', XN-9000, 4 °C
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C8 hstable for the longest time testedHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 5, row 'BA', BC-6800, 4 °C
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C8 hstable for the longest time testedBias% vs optimal bias (OP-Bias%, from biological variation; Table 1)Buoro et al. 2016, Rev Bras Hematol HemoterTable 5, row 'BA', BC-6800, 4 °C
whole-blood-edta, K2-EDTA (BD Vacutainer)Storageuncentrifuged2-8 °C48 hstable for the longest time testedRepeated-measures one-way ANOVA with Tukey post hoc vs 0 h, p < 0.05 (* in table)Unalli et al. 2021, J Med BiochemTable 1, column '4 °C' (no * at any time point)
whole-blood-edta, K3-EDTA (Becton Dickinson)StorageuncentrifugedRT48 hstable for the longest time testedHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 4, row 'BA', XN-9000, RT
whole-blood-edta, K3-EDTA (Becton Dickinson)StorageuncentrifugedRT48 hstable for the longest time testedBias% vs optimal bias (OP-Bias%, from biological variation; Table 1)Buoro et al. 2016, Rev Bras Hematol HemoterTable 4, row 'BA', XN-9000, RT
whole-blood-edta, K3-EDTA (Becton Dickinson)StorageuncentrifugedRT48 hstable for the longest time testedHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 4, row 'BA', BC-6800, RT
whole-blood-edta, K3-EDTA (Becton Dickinson)StorageuncentrifugedRT48 hstable for the longest time testedBias% vs optimal bias (OP-Bias%, from biological variation; Table 1)Buoro et al. 2016, Rev Bras Hematol HemoterTable 4, row 'BA', BC-6800, RT
whole-blood-edta, K3-EDTA (Becton Dickinson)StorageuncentrifugedRT8 hstable for the longest time testedHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 5, row 'BA', XN-9000, RT
whole-blood-edta, K3-EDTA (Becton Dickinson)StorageuncentrifugedRT8 hstable for the longest time testedBias% vs optimal bias (OP-Bias%, from biological variation; Table 1)Buoro et al. 2016, Rev Bras Hematol HemoterTable 5, row 'BA', XN-9000, RT
whole-blood-edta, K3-EDTA (Becton Dickinson)StorageuncentrifugedRT8 hstable for the longest time testedHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 5, row 'BA', BC-6800, RT
whole-blood-edta, K3-EDTA (Becton Dickinson)StorageuncentrifugedRT8 hstable for the longest time testedBias% vs optimal bias (OP-Bias%, from biological variation; Table 1)Buoro et al. 2016, Rev Bras Hematol HemoterTable 5, row 'BA', BC-6800, RT
whole-blood-edta, K2-EDTA (BD Vacutainer)StorageuncentrifugedRT8 hthen changed, changed by 12 hRepeated-measures one-way ANOVA with Tukey post hoc vs 0 h, p < 0.05 (* in table)Unalli et al. 2021, J Med BiochemTable 1, block 'BASO', row '12 h', column '23 °C'
whole-blood-edta, K2-EDTA (BD Vacutainer)Storageuncentrifuged10 °C48 hstable for the longest time testedRepeated-measures one-way ANOVA with Tukey post hoc vs 0 h, p < 0.05 (* in table)Unalli et al. 2021, J Med BiochemTable 1, column '10 °C' (no * at any time point)
Beta-carotene4 findings
Specimen, tubeConditionTemperatureStable forCriterionStudy
whole-blood-clotted, Sarstedt Monovette serumBefore spinninguncentrifugedchilled in transit (insulated box, 2 frozen cool packs; temperature not recorded)24 hstable for the longest time tested% difference within desirable quality specification for bias (Fraser, from intra- and interindividual variation)Jones et al. 2021, J NutrTable 2, row 'β -carotene', column Serum
whole-blood-edta, Sarstedt Monovette EDTABefore spinninguncentrifugedchilled in transit (insulated box, 2 frozen cool packs; temperature not recorded)24 hstable for the longest time tested% difference within desirable quality specification for bias (Fraser, from intra- and interindividual variation)Jones et al. 2021, J NutrTable 2, row 'β -carotene', column EDTA
whole-blood-heparin, BD Vacutainer lithium heparinBefore spinninguncentrifugedcold packs in insulated box; 9 (2) °C on arrival24 hstable for the longest time tested% difference within desirable quality specification (from intra- and interindividual variation)Jones et al. 2024, J NutrTable 2, row 'β-Carotene, μmol/L'; Table 1
whole-blood-heparin, Sarstedt Monovette lithium heparin, trace-element freeBefore spinninguncentrifugedchilled in transit (insulated box, 2 frozen cool packs; temperature not recorded)24 hstable for the longest time tested% difference within desirable quality specification for bias (Fraser, from intra- and interindividual variation)Jones et al. 2021, J NutrTable 2, row 'β -carotene', column LH
Beta-cryptoxanthin3 findings
Specimen, tubeConditionTemperatureStable forCriterionStudy
whole-blood-clotted, Sarstedt Monovette serumBefore spinninguncentrifugedchilled in transit (insulated box, 2 frozen cool packs; temperature not recorded)24 hstable for the longest time tested% difference within desirable quality specification for bias (Fraser, from intra- and interindividual variation)Jones et al. 2021, J NutrTable 2, row 'β-cryptoxanthin', column Serum
whole-blood-edta, Sarstedt Monovette EDTABefore spinninguncentrifugedchilled in transit (insulated box, 2 frozen cool packs; temperature not recorded)24 hstable for the longest time tested% difference within desirable quality specification for bias (Fraser, from intra- and interindividual variation)Jones et al. 2021, J NutrTable 2, row 'β-cryptoxanthin', column EDTA
whole-blood-heparin, Sarstedt Monovette lithium heparin, trace-element freeBefore spinninguncentrifugedchilled in transit (insulated box, 2 frozen cool packs; temperature not recorded)24 hstable for the longest time tested% difference within desirable quality specification for bias (Fraser, from intra- and interindividual variation)Jones et al. 2021, J NutrTable 2, row 'β-cryptoxanthin', column LH
Bicarbonate (Total CO2)6 findings
Specimen, tubeConditionTemperatureStable forCriterionStudy
plasma-li-heparin, BD Vacutainer LH PST II gel separator, 3.0 mL (REF 367374)Storagecentrifuged-on-gel2-8 °C24 hstable for the longest time testedmean % difference smaller than TEa (Ricos: 0.25 x (CVi^2 + CVg^2)^0.5 + 1.65 x 0.5 x CVi)Nonkes et al. 2019, Biochem Med (Zagreb)Table 1, row 'Bicarbonate, mmol/L'
serum, BD Vacutainer SST (SKU 367986), gelStoragecentrifuged-on-gel30 °C (±1 °C)24 hthen changed, changed by 72 hmean bias vs paired baseline within CLIA acceptance limits AND >2/3 of individual results within limitsIacovetti et al. 2026, Pract Lab MedResults, 'Analyte stability after storage at 30 °C', paragraph 1 (Table 1 image-only)
serum, BD SST Microtainer (SKU 395967), gel; filled from venous bloodStoragecentrifuged-on-gel0 °C72 hstable for the longest time testedmean bias vs paired baseline within CLIA acceptance limits AND >2/3 of individual results within limitsIacovetti et al. 2026, Pract Lab MedResults, 'Analyte stability after separation and storage in microvolume tubes at 0 °C' (Table 2 image-only)
whole-blood-clotted, BD Vacutainer SST (SKU 367986), gelBefore spinninguncentrifuged30 °C (±1 °C)24 hthen changed, changed by 72 hmean bias vs paired baseline within CLIA acceptance limits AND >2/3 of individual results within limitsIacovetti et al. 2026, Pract Lab MedResults, 'Analyte stability after storage at 30 °C', paragraph 2 (Table 1 image-only)
whole-blood-heparin, BD Vacutainer lithium heparin, no gel (ref 368884)Before spinninguncentrifugedRT10 hstable for the longest time tested|mean % deviation 10 h vs baseline| > laboratory bias goal (Table II; mostly biological-variation based, some empirically modified)Henriksen et al. 2014, Scand J Clin Lab InvestTable II, row 'CO2'
whole-blood-heparin, BD Vacutainer lithium heparin, no gel (ref 368884)Before spinninguncentrifugedRT10 hstable for the longest time testedCV of paired baseline/10 h results > goal CV (Table II)Henriksen et al. 2014, Scand J Clin Lab InvestTable II, row 'CO2'
Blood Urea Nitrogen9 findings
Specimen, tubeConditionTemperatureStable forCriterionStudy
serum, BD Vacutainer 8.5 mL plastic serum tube; pooled serum aliquoted into 1.5 mL Eppendorf tubesFreeze–thawseparated−20 °C3 cyclesthen changed, changed by 4 cyclesunstable only if both P < 0.005 (Bonferroni) and % change from T0 > desirable bias (Westgard/Ricos), here 5.5%Cuhadar et al. 2013, Biochem Med (Zagreb)Table 2, row 'BUN (mmol/L)', columns T1d-T10d
serum, BD Vacutainer 8.5 mL plastic serum tube; pooled serum aliquoted into 1.5 mL Eppendorf tubesStorageseparated−20 °C0 monthschanged at the first time point, changed by 1 monthsunstable only if both P < 0.013 (Bonferroni) and % change from T0 > desirable bias (Westgard/Ricos), here 5.5%Cuhadar et al. 2013, Biochem Med (Zagreb)Table 2, row 'BUN (mmol/L)', columns T1m-T3m
serum, Vacuette Tube Serum Gel Separator 8 mL (Greiner, 455071)Storagecentrifuged-on-gel2-8 °C30 hthen changed, changed by 36 h% bias of median vs baseline exceeds desirable bias (Westgard/Ricos BV), here 5.5%Cuhadar et al. 2012, Biochem Med (Zagreb)Table 3, row 'Urea nitrogen', gel 4°C
serum, Vacuette Tube Serum No Gel 9 mL (Greiner, 455092), plainStoragecentrifuged-on-cells2-8 °C0 hchanged at the first time point, changed by 6 h% bias of median vs baseline exceeds desirable bias (Westgard/Ricos BV), here 5.5%Cuhadar et al. 2012, Biochem Med (Zagreb)Table 3, row 'Urea nitrogen', w/o gel 4°C
serum, Vacuette Tube Serum Gel Separator 8 mL (Greiner, 455071)Storagecentrifuged-on-gelRT0 hchanged at the first time point, changed by 6 h% bias of median vs baseline exceeds desirable bias (Westgard/Ricos BV), here 5.5%Cuhadar et al. 2012, Biochem Med (Zagreb)Table 3, row 'Urea nitrogen', gel 24°C
serum, Vacuette Tube Serum No Gel 9 mL (Greiner, 455092), plainStoragecentrifuged-on-cellsRT12 hthen changed, changed by 18 h% bias of median vs baseline exceeds desirable bias (Westgard/Ricos BV), here 5.5%Cuhadar et al. 2012, Biochem Med (Zagreb)Table 3, row 'Urea nitrogen', w/o gel 24°C
serum, BD Vacutainer SST (SKU 367986), gelStoragecentrifuged-on-gel30 °C (±1 °C)72 hstable for the longest time testedmean bias vs paired baseline within CLIA acceptance limits AND >2/3 of individual results within limitsIacovetti et al. 2026, Pract Lab MedResults, 'Analyte stability after storage at 30 °C', paragraph 1 (Table 1 image-only)
serum, BD SST Microtainer (SKU 395967), gel; filled from venous bloodStoragecentrifuged-on-gel0 °C72 hstable for the longest time testedmean bias vs paired baseline within CLIA acceptance limits AND >2/3 of individual results within limitsIacovetti et al. 2026, Pract Lab MedResults, 'Analyte stability after separation and storage in microvolume tubes at 0 °C' (Table 2 image-only)
whole-blood-clotted, BD Vacutainer SST (SKU 367986), gelBefore spinninguncentrifuged30 °C (±1 °C)24 hthen changed, changed by 72 hmean bias vs paired baseline within CLIA acceptance limits AND >2/3 of individual results within limitsIacovetti et al. 2026, Pract Lab MedResults, 'Analyte stability after storage at 30 °C', paragraph 2 (Table 1 image-only)
C-Peptide10 findings
Specimen, tubeConditionTemperatureStable forCriterionStudy
plasma-edta, Sarstedt K+-EDTA 2.7 mLFreeze–thawnot-stated−80 °C5 cyclesstable for the longest time testedmean % change from baseline > 10% considered clinically significantMcDonald et al. 2012, PLoS OneTable 1
plasma-edta, Sarstedt K+-EDTA 2.7 mLStoragecentrifuged-on-cellsRT24 hstable for the longest time testedmean % change from baseline > 10% considered clinically significantMcDonald et al. 2012, PLoS OneResults 1.1, paragraph 4
serum, Sarstedt serum gel 2.7 mLFreeze–thawnot-stated−80 °C5 cyclesstable for the longest time testedmean % change from baseline > 10% considered clinically significantMcDonald et al. 2012, PLoS OneTable 1
serum, Sarstedt serum gel 2.7 mLStoragecentrifuged-on-gelRT24 hstable for the longest time testedmean % change from baseline > 10% considered clinically significantMcDonald et al. 2012, PLoS OneResults 1.1, paragraph 4
whole-blood-clotted, Sarstedt serum gel 2.7 mLBefore spinninguncentrifuged2-8 °C24 hstable for the longest time testedmean % change from baseline > 10% considered clinically significantMcDonald et al. 2012, PLoS OneResults 1.1, paragraph 5
whole-blood-clotted, Sarstedt serum gel 2.7 mLBefore spinninguncentrifuged2-8 °C120 hstable for the longest time testedmean % change from baseline > 10% considered clinically significantMcDonald et al. 2012, PLoS OneResults 1.2
whole-blood-edta, Sarstedt K+-EDTA 2.7 mLBefore spinninguncentrifuged2-8 °C24 hstable for the longest time testedmean % change from baseline > 10% considered clinically significantMcDonald et al. 2012, PLoS OneResults 1.1, paragraph 5
whole-blood-edta, Sarstedt K+-EDTA 2.7 mLBefore spinninguncentrifuged2-8 °C120 hstable for the longest time testedmean % change from baseline > 10% considered clinically significantMcDonald et al. 2012, PLoS OneResults 1.2
whole-blood-edta, Sarstedt K+-EDTA 2.7 mLBefore spinninguncentrifugedRT24 hstable for the longest time testedmean % change from baseline > 10% considered clinically significantMcDonald et al. 2012, PLoS OneResults 1.1, paragraph 4
whole-blood-edta, Sarstedt K+-EDTA 2.7 mLBefore spinninguncentrifugedRT36 hthen changed, changed by 48 hmean % change from baseline > 10% considered clinically significantMcDonald et al. 2012, PLoS OneResults 1.2; Discussion paragraph 2
C-Reactive Protein20 findings
Specimen, tubeConditionTemperatureStable forCriterionStudy
dbs, Whatman CF12 filter paper, zip-lock plastic bag; venous EDTA whole blood spottedStoragedried−20 °C31 daysstable for the longest time testedsamples remain in the same clinical CRP category (0-3, 3+, 10+ mg/L) as on day 1Schakelaar et al. 2023, Sci RepResults, 'CRP stability up to 31 days'; Figure 2; Methods 'Time and temperature stability'
dbs, Whatman CF12 filter paper, zip-lock plastic bag; venous EDTA whole blood spottedStoragedried2-8 °C31 daysstable for the longest time testedsamples remain in the same clinical CRP category (0-3, 3+, 10+ mg/L) as on day 1Schakelaar et al. 2023, Sci RepResults, 'CRP stability up to 31 days'; Figure 2; Methods 'Time and temperature stability'
dbs, Whatman CF12 filter paper, zip-lock plastic bag; venous EDTA whole blood spottedStoragedriedRT31 daysstable for the longest time testedsamples remain in the same clinical CRP category (0-3, 3+, 10+ mg/L) as on day 1Schakelaar et al. 2023, Sci RepResults, 'CRP stability up to 31 days'; Figure 2; Methods 'Time and temperature stability'
dbs, Whatman 903 card, mailed in foil pouch with desiccantStoragedriednot stated; ambient postal shipping in the US (Apr-Dec 2016)8 daysstable for the longest time testedregression coefficient for days before freezing vs 0-2 days significant (P < .05), Model 1 (DBS value, unadjusted)Crimmins et al. 2020, Am J Hum BiolTable 3, rows 'Days before freezing', column 'CRP'
dbs, Whatman 903 card, mailed in foil pouch with desiccantStoragedriednot stated; ambient postal shipping in the US (Apr-Dec 2016)8 daysstable for the longest time testedregression coefficient for days before freezing vs 0-2 days significant (P < .05), Model 2 (DBS value adjusted for venous value)Crimmins et al. 2020, Am J Hum BiolTable 3, rows 'Days before freezing', column 'CRP'
dbs, Whatman CF12 filter paper, zip-lock plastic bag; venous EDTA whole blood spottedStoragedried+ 30 °C31 daysstable for the longest time testedsamples remain in the same clinical CRP category (0-3, 3+, 10+ mg/L) as on day 1Schakelaar et al. 2023, Sci RepResults, 'CRP stability up to 31 days'; Figure 2; Methods 'Time and temperature stability'
plasma-li-heparin, BD Vacutainer lithium heparin 6 mL (regular)Storagenot-stated2-8 °C72 hstable for the longest time testedmean % difference vs T0 exceeds total change limit (TCL), here ±38.72%Dupuy et al. 2018, Biochem Med (Zagreb)Table 2, BD lithium heparin tubes, row 'CRP, mg/L'
plasma-li-heparin, BD Barricor lithium heparin 5.5 mL (mechanical separator)Storagecentrifuged-on-gel2-8 °C72 hstable for the longest time testedmean % difference vs T0 exceeds total change limit (TCL), here ±38.72%Dupuy et al. 2018, Biochem Med (Zagreb)Table 2, BD Barricor lithium heparin tubes, row 'CRP, mg/L'
plasma-li-heparin, BD Barricor 4.5 mL (lithium heparin, mechanical separator)Storagecentrifuged-on-gel2-8 °C7 daysstable for the longest time testedmean % difference vs time 0 exceeds desirable imprecision based on biological variation (EFLM/Westgard), here 16.75%Mandić et al. 2020, Biochem Med (Zagreb)Tables 3 and 4, row 'CRP, mg/L', BD Barricor columns
serum, BD Vacutainer SSTFreeze–thawseparated−70 °C, 1 month; thawed at room temperature1 cyclestable for the longest time testedmean % change (bias) from fresh baseline within total allowable error (TEa, desirable bias)Abraham et al. 2021, J Lab PhysiciansTable 2 and Table 3, row 'CRP (mg/dL)'
serum, BD Vacutainer RST 5 mL (thrombin clot activator, gel)Storagecentrifuged-on-gel2-8 °C7 daysstable for the longest time testedmean % difference vs time 0 exceeds desirable imprecision based on biological variation (EFLM/Westgard), here 16.75%Mandić et al. 2020, Biochem Med (Zagreb)Tables 3 and 4, row 'CRP, mg/L', BD RST columns
whole-blood-clotted, BD Vacutainer SSTBefore spinninguncentrifugedRT24 hstable for the longest time testedmean % deviation from baseline within total change limit TCL = sqrt((2.77 CVa)^2 + (0.5 CVb)^2), CVb from Ricos (Table 2)Abraham et al. 2019, Biochem Med (Zagreb)Discussion, paragraph 1 (Results paragraphs 1-3; Figure 3)
whole-blood-clotted, BD Vacutainer trace element-free serum tubeBefore spinninguncentrifugedcold packs in insulated box; 9 (2) °C on arrival24 hstable for the longest time tested% difference within desirable quality specification (from intra- and interindividual variation)Jones et al. 2024, J NutrTable 2, row 'C-reactive protein, mg/L'; Table 1
whole-blood-clotted, BD Vacutainer SSTBefore spinninguncentrifuged22-30 °C (logger 24-26 °C)24 hstable for the longest time testedmean % deviation from baseline within total change limit TCL = sqrt((2.77 CVa)^2 + (0.5 CVb)^2), CVb from Ricos (Table 2)Abraham et al. 2019, Biochem Med (Zagreb)Discussion, paragraph 1 (Results paragraphs 1-3; Figure 3)
whole-blood-clotted, BD Vacutainer SSTBefore spinninguncentrifuged> 30 °C (logger 38-40 °C)24 hstable for the longest time testedmean % deviation from baseline within total change limit TCL = sqrt((2.77 CVa)^2 + (0.5 CVb)^2), CVb from Ricos (Table 2)Abraham et al. 2019, Biochem Med (Zagreb)Discussion, paragraph 1 (Results paragraphs 1-3; Figure 3)
whole-blood-clotted, Sarstedt Monovette serumBefore spinninguncentrifugedchilled in transit (insulated box, 2 frozen cool packs; temperature not recorded)24 hstable for the longest time tested% difference within desirable quality specification for bias (Fraser, from intra- and interindividual variation)Jones et al. 2021, J NutrTable 2, row 'C-reactive protein', column Serum
whole-blood-edta, Sarstedt Monovette EDTABefore spinninguncentrifugedchilled in transit (insulated box, 2 frozen cool packs; temperature not recorded)24 hstable for the longest time tested% difference within desirable quality specification for bias (Fraser, from intra- and interindividual variation)Jones et al. 2021, J NutrTable 2, row 'C-reactive protein', column EDTA
whole-blood-heparin, BD Vacutainer lithium heparin, no gel (ref 368884)Before spinninguncentrifugedRT10 hstable for the longest time tested|mean % deviation 10 h vs baseline| > laboratory bias goal (Table II; mostly biological-variation based, some empirically modified)Henriksen et al. 2014, Scand J Clin Lab InvestTable II, row 'CRP'
whole-blood-heparin, BD Vacutainer lithium heparin, no gel (ref 368884)Before spinninguncentrifugedRT10 hstable for the longest time testedCV of paired baseline/10 h results > goal CV (Table II)Henriksen et al. 2014, Scand J Clin Lab InvestTable II, row 'CRP'
whole-blood-heparin, Sarstedt Monovette lithium heparin, trace-element freeBefore spinninguncentrifugedchilled in transit (insulated box, 2 frozen cool packs; temperature not recorded)24 hstable for the longest time tested% difference within desirable quality specification for bias (Fraser, from intra- and interindividual variation)Jones et al. 2021, J NutrTable 2, row 'C-reactive protein', column LH
Calcium, total15 findings
Specimen, tubeConditionTemperatureStable forCriterionStudy
plasma-li-heparin, Sarstedt lithium heparin, no gel separator (primary tube)Storagecentrifuged-on-cells2-8 °C48 hstable for the longest time testedmean % difference vs time 0 exceeds total limit of change, TLC = ±√((1.65·CVa)² + (0.5·CVb)²) (EFLM BV), here 4.6%; stability limit from instability-equation regressionIllana et al. 2024, Biochem Med (Zagreb)Table 1, row 'Ca'
plasma-li-heparin, Sarstedt lithium heparin, no gel separator; plasma aliquotStorageseparated2-8 °C96 hstable for the longest time testedmean % difference vs time 0 exceeds total limit of change, TLC = ±√((1.65·CVa)² + (0.5·CVb)²) (EFLM BV), here 4.6%; stability limit from instability-equation regressionIllana et al. 2024, Biochem Med (Zagreb)Table 2, row 'Ca'
plasma-li-heparin, BD Vacutainer LH PST II gel separator, 3.0 mL (REF 367374)Storagecentrifuged-on-gel2-8 °C24 hstable for the longest time testedmean % difference smaller than TEa (Ricos: 0.25 x (CVi^2 + CVg^2)^0.5 + 1.65 x 0.5 x CVi)Nonkes et al. 2019, Biochem Med (Zagreb)Table 1, row 'Calcium, mmol/L'
plasma-li-heparin, BD Barricor 4.5 mL (lithium heparin, mechanical separator)Storagecentrifuged-on-gel2-8 °C24 hthen changed, changed by 7 daysmean % difference vs time 0 exceeds desirable imprecision based on biological variation (EFLM/Westgard), here 1.05%Mandić et al. 2020, Biochem Med (Zagreb)Tables 3 and 4, row 'Calcium, mmol/L', BD Barricor columns
serum, BD Vacutainer 8.5 mL plastic serum tube; pooled serum aliquoted into 1.5 mL Eppendorf tubesFreeze–thawseparated−20 °C3 cyclesthen changed, changed by 4 cyclesunstable only if both P < 0.005 (Bonferroni) and % change from T0 > desirable bias (Westgard/Ricos), here 0.8%Cuhadar et al. 2013, Biochem Med (Zagreb)Table 2, row 'Calcium (mmol/L)', columns T1d-T10d
serum, BD Vacutainer 8.5 mL plastic serum tube; pooled serum aliquoted into 1.5 mL Eppendorf tubesStorageseparated−20 °C3 monthsstable for the longest time testedunstable only if both P < 0.013 (Bonferroni) and % change from T0 > desirable bias (Westgard/Ricos), here 0.8%Cuhadar et al. 2013, Biochem Med (Zagreb)Table 2, row 'Calcium (mmol/L)', columns T1m-T3m
serum, BD Vacutainer serum (plain), 7.5 mLStorageseparated−20 °C30 daysstable for the longest time testedclinically significant change: mean outside significant change limit (SCL = T1d mean ± 3.0 usual SD from 2 months QC)Kachhawa et al. 2017, J Lab PhysiciansResults paragraph 1 (Table 2 image-only)
serum, BD Vacutainer RST 4.0 mL (thrombin clot activator, gel)Storagecentrifuged-on-gel2-8 °C24 hstable for the longest time testedclinically significant if mean bias vs t0 exceeds desirable bias (biological variation), here 0.8%Kocijancic et al. 2014, Biochem Med (Zagreb)Table 3, row 'Ca'
serum, BD Vacutainer SST II Advance 5.0 mL (clot activator, gel)Storagecentrifuged-on-gel2-8 °C24 hstable for the longest time testedclinically significant if mean bias vs t0 exceeds desirable bias (biological variation), here 0.8%Kocijancic et al. 2014, Biochem Med (Zagreb)Table 4, row 'Ca'
serum, BD Vacutainer RST 5 mL (thrombin clot activator, gel)Storagecentrifuged-on-gel2-8 °C24 hthen changed, changed by 7 daysmean % difference vs time 0 exceeds desirable imprecision based on biological variation (EFLM/Westgard), here 1.05%Mandić et al. 2020, Biochem Med (Zagreb)Tables 3 and 4, row 'Calcium, mmol/L', BD RST columns
serum, BD Vacutainer SST (SKU 367986), gelStoragecentrifuged-on-gel30 °C (±1 °C)72 hstable for the longest time testedmean bias vs paired baseline within CLIA acceptance limits AND >2/3 of individual results within limitsIacovetti et al. 2026, Pract Lab MedResults, 'Analyte stability after storage at 30 °C', paragraph 1 (Table 1 image-only)
serum, BD SST Microtainer (SKU 395967), gel; filled from venous bloodStoragecentrifuged-on-gel0 °C72 hstable for the longest time testedmean bias vs paired baseline within CLIA acceptance limits AND >2/3 of individual results within limitsIacovetti et al. 2026, Pract Lab MedResults, 'Analyte stability after separation and storage in microvolume tubes at 0 °C' (Table 2 image-only)
whole-blood-clotted, BD Vacutainer SST (SKU 367986), gelBefore spinninguncentrifuged30 °C (±1 °C)24 hthen changed, changed by 72 hmean bias vs paired baseline within CLIA acceptance limits AND >2/3 of individual results within limitsIacovetti et al. 2026, Pract Lab MedResults, 'Analyte stability after storage at 30 °C', paragraph 2 (Table 1 image-only)
whole-blood-heparin, BD Vacutainer lithium heparin, no gel (ref 368884)Before spinninguncentrifugedRT10 hstable for the longest time tested|mean % deviation 10 h vs baseline| > laboratory bias goal (Table II; mostly biological-variation based, some empirically modified)Henriksen et al. 2014, Scand J Clin Lab InvestTable II, row 'CA'
whole-blood-heparin, BD Vacutainer lithium heparin, no gel (ref 368884)Before spinninguncentrifugedRT10 hstable for the longest time testedCV of paired baseline/10 h results > goal CV (Table II)Henriksen et al. 2014, Scand J Clin Lab InvestTable II, row 'CA'
Chloride11 findings
Specimen, tubeConditionTemperatureStable forCriterionStudy
plasma-li-heparin, Sarstedt lithium heparin, no gel separator (primary tube)Storagecentrifuged-on-cells2-8 °C48 hstable for the longest time testedmean % difference vs time 0 exceeds total limit of change, TLC = ±√((1.65·CVa)² + (0.5·CVb)²) (EFLM BV), here 4.8%; stability limit from instability-equation regressionIllana et al. 2024, Biochem Med (Zagreb)Table 1, row 'Cl'
plasma-li-heparin, Sarstedt lithium heparin, no gel separator; plasma aliquotStorageseparated2-8 °C96 hstable for the longest time testedmean % difference vs time 0 exceeds total limit of change, TLC = ±√((1.65·CVa)² + (0.5·CVb)²) (EFLM BV), here 4.8%; stability limit from instability-equation regressionIllana et al. 2024, Biochem Med (Zagreb)Table 2, row 'Cl'
plasma-li-heparin, BD Vacutainer PST II lithium heparin 3 mL (gel)Storagecentrifuged-on-gel2-8 °C15 hstable for the longest time testedclinically significant when the 95% CI of mean bias lies completely outside the SEQC maximum allowable bias (MAB = 1.65 x CVa), here 1%Balbás et al. 2017, Biochem Med (Zagreb)Table 2, Cl- row 'PST II sealed, 4 şC'
plasma-li-heparin, BD Barricor lithium heparin 5.5 mL (mechanical separator)Storagecentrifuged-on-gel2-8 °C15 hstable for the longest time testedclinically significant when the 95% CI of mean bias lies completely outside the SEQC maximum allowable bias (MAB = 1.65 x CVa), here 1%Balbás et al. 2017, Biochem Med (Zagreb)Table 2, Cl- row 'Barricor sealed, 4 şC'
plasma-li-heparin, BD Barricor 4.5 mL (lithium heparin, mechanical separator)Storagecentrifuged-on-gel2-8 °C7 daysstable for the longest time testedmean % difference vs time 0 exceeds desirable imprecision based on biological variation (EFLM/Westgard), here 0.55%Mandić et al. 2020, Biochem Med (Zagreb)Tables 3 and 4, row 'Chloride, mmol/L', BD Barricor columns
plasma-li-heparin, BD Vacutainer PST II lithium heparin 3 mL (gel)Storagecentrifuged-on-gelRT3 hthen changed, changed by 6 hclinically significant when the 95% CI of mean bias lies completely outside the SEQC maximum allowable bias (MAB = 1.65 x CVa), here 1%Balbás et al. 2017, Biochem Med (Zagreb)Table 2, Cl- row 'PST II open, rt'
plasma-li-heparin, BD Barricor lithium heparin 5.5 mL (mechanical separator)Storagecentrifuged-on-gelRT0 hchanged at the first time point, changed by 3 hclinically significant when the 95% CI of mean bias lies completely outside the SEQC maximum allowable bias (MAB = 1.65 x CVa), here 1%Balbás et al. 2017, Biochem Med (Zagreb)Table 2, Cl- row 'Barricor open, rt'
serum, BD Vacutainer RST 5 mL (thrombin clot activator, gel)Storagecentrifuged-on-gel2-8 °C7 daysstable for the longest time testedmean % difference vs time 0 exceeds desirable imprecision based on biological variation (EFLM/Westgard), here 0.55%Mandić et al. 2020, Biochem Med (Zagreb)Tables 3 and 4, row 'Chloride, mmol/L', BD RST columns
serum, BD Vacutainer SST (SKU 367986), gelStoragecentrifuged-on-gel30 °C (±1 °C)72 hstable for the longest time testedmean bias vs paired baseline within CLIA acceptance limits AND >2/3 of individual results within limitsIacovetti et al. 2026, Pract Lab MedResults, 'Analyte stability after storage at 30 °C', paragraph 1 (Table 1 image-only)
serum, BD SST Microtainer (SKU 395967), gel; filled from venous bloodStoragecentrifuged-on-gel0 °C72 hstable for the longest time testedmean bias vs paired baseline within CLIA acceptance limits AND >2/3 of individual results within limitsIacovetti et al. 2026, Pract Lab MedResults, 'Analyte stability after separation and storage in microvolume tubes at 0 °C' (Table 2 image-only)
whole-blood-clotted, BD Vacutainer SST (SKU 367986), gelBefore spinninguncentrifuged30 °C (±1 °C)0 hchanged at the first time point, changed by 24 hmean bias vs paired baseline within CLIA acceptance limits AND >2/3 of individual results within limitsIacovetti et al. 2026, Pract Lab MedResults, 'Analyte stability after storage at 30 °C', paragraph 2 (Table 1 image-only)
Coagulation factor II5 findings
Specimen, tubeConditionTemperatureStable forCriterionStudy
plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquotsFreeze–thawseparated−20 °C1 cyclelimit stated by the authorsmax allowable mean deviation from t0: ±15%Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability'
plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquotsFreeze–thawseparated−80 °C3 cycleslimit stated by the authorsmax allowable mean deviation from t0: ±15%Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability'
plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquotsStorageseparated−20 °C28 dayslimit stated by the authorsmax allowable mean deviation from t0: ±15%Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability'
plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquotsStorageseparated−80 °C12 monthslimit stated by the authorsmax allowable mean deviation from t0: ±15%Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability'
plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquotsStorageseparatedRT4 hlimit stated by the authorsmax allowable mean deviation from t0: ±15%Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability'
Coagulation factor IX8 findings
Specimen, tubeConditionTemperatureStable forCriterionStudy
plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquotsFreeze–thawseparated−20 °C1 cyclelimit stated by the authorsmax allowable mean deviation from t0: ≤1.25%: ±0.250% (absolute); >1.25–≤30.0%: ±20.0% (relative); >30.0%: ±15.0%Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability'
plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquotsFreeze–thawseparated−80 °C3 cycleslimit stated by the authorsmax allowable mean deviation from t0: ≤1.25%: ±0.250% (absolute); >1.25–≤30.0%: ±20.0% (relative); >30.0%: ±15.0%Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability'
plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquotsStorageseparated−20 °C14 dayslimit stated by the authorsmax allowable mean deviation from t0: ≤1.25%: ±0.250% (absolute); >1.25–≤30.0%: ±20.0% (relative); >30.0%: ±15.0%Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability'
plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquotsStorageseparated−80 °C28 dayslimit stated by the authorsmax allowable mean deviation from t0: ≤1.25%: ±0.250% (absolute); >1.25–≤30.0%: ±20.0% (relative); >30.0%: ±15.0%Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability'
plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquotsStorageseparated−80 °C24 monthslimit stated by the authorsmax allowable mean deviation from t0: ≤1.25%: ±0.250% (absolute); >1.25–≤30.0%: ±20.0% (relative); >30.0%: ±15.0%Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability'
plasma-citrate, BD 0.109 M sodium citrate (9:1); plasma aliquoted into nonactivating Eppendorf tubesStorageseparated2-8 °C4 hthen changed, changed by 6 hmean % change from baseline >10%, or >25% of samples with individual change >10% (van Geest-Daalderop)Feng et al. 2014, Sci RepTable 2, 'Acceptable Time (h)'
plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquotsStorageseparatedRT4 hlimit stated by the authorsmax allowable mean deviation from t0: ≤1.25%: ±0.250% (absolute); >1.25–≤30.0%: ±20.0% (relative); >30.0%: ±15.0%Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability'
plasma-citrate, BD 0.109 M sodium citrate (9:1); plasma aliquoted into nonactivating Eppendorf tubesStorageseparatedRT4 hthen changed, changed by 6 hmean % change from baseline >10%, or >25% of samples with individual change >10% (van Geest-Daalderop)Feng et al. 2014, Sci RepTable 2, 'Acceptable Time (h)'
Coagulation factor V4 findings
Specimen, tubeConditionTemperatureStable forCriterionStudy
plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquotsFreeze–thawseparated−80 °C1 cyclelimit stated by the authorsmax allowable mean deviation from t0: ±15%Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability'
plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquotsStorageseparated−80 °C28 dayslimit stated by the authorsmax allowable mean deviation from t0: ±15%Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability'
plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquotsStorageseparated−80 °C24 monthslimit stated by the authorsmax allowable mean deviation from t0: ±15%Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability'
plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquotsStorageseparatedRT4 hlimit stated by the authorsmax allowable mean deviation from t0: ±15%Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability'
Coagulation factor VII5 findings
Specimen, tubeConditionTemperatureStable forCriterionStudy
plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquotsFreeze–thawseparated−20 °C1 cyclelimit stated by the authorsmax allowable mean deviation from t0: ±15%Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability'
plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquotsFreeze–thawseparated−80 °C3 cycleslimit stated by the authorsmax allowable mean deviation from t0: ±15%Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability'
plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquotsStorageseparated−20 °C28 dayslimit stated by the authorsmax allowable mean deviation from t0: ±15%Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability'
plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquotsStorageseparated−80 °C24 monthslimit stated by the authorsmax allowable mean deviation from t0: ±15%Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability'
plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquotsStorageseparatedRT4 hlimit stated by the authorsmax allowable mean deviation from t0: ±15%Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability'
Coagulation factor VIII5 findings
Specimen, tubeConditionTemperatureStable forCriterionStudy
plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquotsFreeze–thawseparated−20 °C1 cyclelimit stated by the authorsmax allowable mean deviation from t0: ±15%Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability'
plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquotsStorageseparated−20 °C14 dayslimit stated by the authorsmax allowable mean deviation from t0: ±15%Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability'
plasma-citrate, BD 0.109 M sodium citrate (9:1); plasma aliquoted into nonactivating Eppendorf tubesStorageseparated2-8 °C2 hthen changed, changed by 4 hmean % change from baseline >10%, or >25% of samples with individual change >10% (van Geest-Daalderop)Feng et al. 2014, Sci RepTable 2, 'Acceptable Time (h)'
plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquotsStorageseparatedRT2 hlimit stated by the authorsmax allowable mean deviation from t0: ±15%Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability'
plasma-citrate, BD 0.109 M sodium citrate (9:1); plasma aliquoted into nonactivating Eppendorf tubesStorageseparatedRT2 hthen changed, changed by 4 hmean % change from baseline >10%, or >25% of samples with individual change >10% (van Geest-Daalderop)Feng et al. 2014, Sci RepTable 2, 'Acceptable Time (h)'
Coagulation factor X5 findings
Specimen, tubeConditionTemperatureStable forCriterionStudy
plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquotsFreeze–thawseparated−20 °C1 cyclelimit stated by the authorsmax allowable mean deviation from t0: ±10%Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability'
plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquotsFreeze–thawseparated−80 °C1 cyclelimit stated by the authorsmax allowable mean deviation from t0: ±10%Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability'
plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquotsStorageseparated−20 °C28 dayslimit stated by the authorsmax allowable mean deviation from t0: ±10%Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability'
plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquotsStorageseparated−80 °C24 monthslimit stated by the authorsmax allowable mean deviation from t0: ±10%Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability'
plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquotsStorageseparatedRT4 hlimit stated by the authorsmax allowable mean deviation from t0: ±10%Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability'
Coagulation factor XI5 findings
Specimen, tubeConditionTemperatureStable forCriterionStudy
plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquotsFreeze–thawseparated−20 °C1 cyclelimit stated by the authorsmax allowable mean deviation from t0: ±15%Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability'
plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquotsFreeze–thawseparated−80 °C3 cycleslimit stated by the authorsmax allowable mean deviation from t0: ±15%Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability'
plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquotsStorageseparated−20 °C28 dayslimit stated by the authorsmax allowable mean deviation from t0: ±15%Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability'
plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquotsStorageseparated−80 °C24 monthslimit stated by the authorsmax allowable mean deviation from t0: ±15%Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability'
plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquotsStorageseparatedRT4 hlimit stated by the authorsmax allowable mean deviation from t0: ±15%Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability'
Coagulation factor XII6 findings
Specimen, tubeConditionTemperatureStable forCriterionStudy
plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquotsFreeze–thawseparated−20 °C1 cyclelimit stated by the authorsmax allowable mean deviation from t0: ±15%Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability'
plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquotsFreeze–thawseparated−80 °C3 cycleslimit stated by the authorsmax allowable mean deviation from t0: ±15%Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability'
plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquotsStorageseparated−20 °C28 dayslimit stated by the authorsmax allowable mean deviation from t0: ±15%Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability'
plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquotsStorageseparated−80 °C28 dayslimit stated by the authorsmax allowable mean deviation from t0: ±15%Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability'
plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquotsStorageseparated−80 °C6 monthslimit stated by the authorsmax allowable mean deviation from t0: ±15%Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability'
plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquotsStorageseparatedRT4 hlimit stated by the authorsmax allowable mean deviation from t0: ±15%Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability'
Coagulation factor XIII6 findings
Specimen, tubeConditionTemperatureStable forCriterionStudy
plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquotsFreeze–thawseparated−20 °C1 cyclelimit stated by the authorsmax allowable mean deviation from t0: 1.57–20.0% FXIII antigen potency: ≤±3.00% (absolute); >20.0–109% FXIII antigen potency: ≤±15.0% (relative)Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability'
plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquotsFreeze–thawseparated−80 °C3 cycleslimit stated by the authorsmax allowable mean deviation from t0: 1.57–20.0% FXIII antigen potency: ≤±3.00% (absolute); >20.0–109% FXIII antigen potency: ≤±15.0% (relative)Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability'
plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquotsStorageseparated−20 °C28 dayslimit stated by the authorsmax allowable mean deviation from t0: 1.57–20.0% FXIII antigen potency: ≤±3.00% (absolute); >20.0–109% FXIII antigen potency: ≤±15.0% (relative)Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability'
plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquotsStorageseparated−80 °C28 dayslimit stated by the authorsmax allowable mean deviation from t0: 1.57–20.0% FXIII antigen potency: ≤±3.00% (absolute); >20.0–109% FXIII antigen potency: ≤±15.0% (relative)Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability'
plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquotsStorageseparated−80 °C24 monthslimit stated by the authorsmax allowable mean deviation from t0: 1.57–20.0% FXIII antigen potency: ≤±3.00% (absolute); >20.0–109% FXIII antigen potency: ≤±15.0% (relative)Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability'
plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquotsStorageseparatedRT4 hlimit stated by the authorsmax allowable mean deviation from t0: 1.57–20.0% FXIII antigen potency: ≤±3.00% (absolute); >20.0–109% FXIII antigen potency: ≤±15.0% (relative)Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability'
Creatine Kinase MB Isoenzyme2 findings
Specimen, tubeConditionTemperatureStable forCriterionStudy
plasma-li-heparin, BD Barricor 4.5 mL (lithium heparin, mechanical separator)Storagecentrifuged-on-gel2-8 °C24 hthen changed, changed by 7 daysmean % difference vs time 0 exceeds desirable imprecision based on biological variation (EFLM/Westgard), here 9.9%Mandić et al. 2020, Biochem Med (Zagreb)Tables 3 and 4, row 'CK-MB, U/L', BD Barricor columns
serum, BD Vacutainer RST 5 mL (thrombin clot activator, gel)Storagecentrifuged-on-gel2-8 °C24 hthen changed, changed by 7 daysmean % difference vs time 0 exceeds desirable imprecision based on biological variation (EFLM/Westgard), here 9.9%Mandić et al. 2020, Biochem Med (Zagreb)Tables 3 and 4, row 'CK-MB, U/L', BD RST columns
Creatine Kinase Total12 findings
Specimen, tubeConditionTemperatureStable forCriterionStudy
plasma-li-heparin, Sarstedt lithium heparin, no gel separator (primary tube)Storagecentrifuged-on-cells2-8 °C48 hstable for the longest time testedmean % difference vs time 0 exceeds total limit of change, TLC = ±√((1.65·CVa)² + (0.5·CVb)²) (EFLM BV), here 9.5%; stability limit from instability-equation regressionIllana et al. 2024, Biochem Med (Zagreb)Table 1, row 'CK'
plasma-li-heparin, Sarstedt lithium heparin, no gel separator; plasma aliquotStorageseparated2-8 °C96 hstable for the longest time testedmean % difference vs time 0 exceeds total limit of change, TLC = ±√((1.65·CVa)² + (0.5·CVb)²) (EFLM BV), here 9.5%; stability limit from instability-equation regressionIllana et al. 2024, Biochem Med (Zagreb)Table 2, row 'CK'
plasma-li-heparin, BD Barricor 4.5 mL (lithium heparin, mechanical separator)Storagecentrifuged-on-gel2-8 °C7 daysstable for the longest time testedmean % difference vs time 0 exceeds desirable imprecision based on biological variation (EFLM/Westgard), here 8.0%Mandić et al. 2020, Biochem Med (Zagreb)Tables 3 and 4, row 'CK, U/L', BD Barricor columns
serum, BD Vacutainer 8.5 mL plastic serum tube; pooled serum aliquoted into 1.5 mL Eppendorf tubesFreeze–thawseparated−20 °C10 cyclesstable for the longest time testedunstable only if both P < 0.005 (Bonferroni) and % change from T0 > desirable bias (Westgard/Ricos), here 11.5%Cuhadar et al. 2013, Biochem Med (Zagreb)Table 2, row 'CK (U/L)', columns T1d-T10d
serum, BD Vacutainer 8.5 mL plastic serum tube; pooled serum aliquoted into 1.5 mL Eppendorf tubesStorageseparated−20 °C3 monthsstable for the longest time testedunstable only if both P < 0.013 (Bonferroni) and % change from T0 > desirable bias (Westgard/Ricos), here 11.5%Cuhadar et al. 2013, Biochem Med (Zagreb)Table 2, row 'CK (U/L)', columns T1m-T3m
serum, Vacuette Tube Serum Gel Separator 8 mL (Greiner, 455071)Storagecentrifuged-on-gel2-8 °C1 weekstable for the longest time tested% bias of median vs baseline exceeds desirable bias (Westgard/Ricos BV), here 11.5%Cuhadar et al. 2012, Biochem Med (Zagreb)Table 3, row 'Creatine kinase', gel 4°C
serum, Vacuette Tube Serum No Gel 9 mL (Greiner, 455092), plainStoragecentrifuged-on-cells2-8 °C1 weekstable for the longest time tested% bias of median vs baseline exceeds desirable bias (Westgard/Ricos BV), here 11.5%Cuhadar et al. 2012, Biochem Med (Zagreb)Table 3, row 'Creatine kinase', w/o gel 4°C
serum, BD Vacutainer RST 5 mL (thrombin clot activator, gel)Storagecentrifuged-on-gel2-8 °C7 daysstable for the longest time testedmean % difference vs time 0 exceeds desirable imprecision based on biological variation (EFLM/Westgard), here 8.0%Mandić et al. 2020, Biochem Med (Zagreb)Tables 3 and 4, row 'CK, U/L', BD RST columns
serum, Vacuette Tube Serum Gel Separator 8 mL (Greiner, 455071)Storagecentrifuged-on-gelRT72 hthen changed, changed by 1 week% bias of median vs baseline exceeds desirable bias (Westgard/Ricos BV), here 11.5%Cuhadar et al. 2012, Biochem Med (Zagreb)Table 3, row 'Creatine kinase', gel 24°C
serum, Vacuette Tube Serum No Gel 9 mL (Greiner, 455092), plainStoragecentrifuged-on-cellsRT72 hthen changed, changed by 1 week% bias of median vs baseline exceeds desirable bias (Westgard/Ricos BV), here 11.5%Cuhadar et al. 2012, Biochem Med (Zagreb)Table 3, row 'Creatine kinase', w/o gel 24°C
whole-blood-heparin, BD Vacutainer lithium heparin, no gel (ref 368884)Before spinninguncentrifugedRT10 hstable for the longest time tested|mean % deviation 10 h vs baseline| > laboratory bias goal (Table II; mostly biological-variation based, some empirically modified)Henriksen et al. 2014, Scand J Clin Lab InvestTable II, row 'CK'
whole-blood-heparin, BD Vacutainer lithium heparin, no gel (ref 368884)Before spinninguncentrifugedRT10 hstable for the longest time testedCV of paired baseline/10 h results > goal CV (Table II)Henriksen et al. 2014, Scand J Clin Lab InvestTable II, row 'CK'
Creatinine22 findings
Specimen, tubeConditionTemperatureStable forCriterionStudy
other, BD Microtainer SST (capillary, lancet)Before spinninguncentrifugedRT3 daysstable for the longest time testedBland-Altman bias within total allowable error for creatinine (8%) and r > 0.95Woolley et al. 2023, Br J Biomed SciResults (Table 1, Figure 1) and Discussion paragraph 4
other, BD Microtainer SST (capillary, lancet)Before spinninguncentrifuged24 h at 2°C–8°C, 24 h at ambient (16°C–22°C) and 24 h 37°C3 daysstable for the longest time testedcorrelation r > 0.95 and agreement within total allowable error for creatinine (8%)Woolley et al. 2023, Br J Biomed SciTable 2 and Discussion paragraph 5
plasma-li-heparin, Sarstedt lithium heparin, no gel separator (primary tube)Storagecentrifuged-on-cells2-8 °C48 hstable for the longest time testedmean % difference vs time 0 exceeds total limit of change, TLC = ±√((1.65·CVa)² + (0.5·CVb)²) (EFLM BV), here 10.1%; stability limit from instability-equation regressionIllana et al. 2024, Biochem Med (Zagreb)Table 1, row 'CREA'
plasma-li-heparin, Sarstedt lithium heparin, no gel separator; plasma aliquotStorageseparated2-8 °C96 hstable for the longest time testedmean % difference vs time 0 exceeds total limit of change, TLC = ±√((1.65·CVa)² + (0.5·CVb)²) (EFLM BV), here 10.1%; stability limit from instability-equation regressionIllana et al. 2024, Biochem Med (Zagreb)Table 2, row 'CREA'
plasma-li-heparin, BD Barricor 4.5 mL (lithium heparin, mechanical separator)Storagecentrifuged-on-gel2-8 °C7 daysstable for the longest time testedmean % difference vs time 0 exceeds desirable imprecision based on biological variation (EFLM/Westgard), here 2.25%Mandić et al. 2020, Biochem Med (Zagreb)Tables 3 and 4, row 'Creatinine, μmol/L', BD Barricor columns
serum, BD Vacutainer 8.5 mL plastic serum tube; pooled serum aliquoted into 1.5 mL Eppendorf tubesFreeze–thawseparated−20 °C10 cyclesstable for the longest time testedunstable only if both P < 0.005 (Bonferroni) and % change from T0 > desirable bias (Westgard/Ricos), here 4.0%Cuhadar et al. 2013, Biochem Med (Zagreb)Table 2, row 'Creatinine (μmol/L)', columns T1d-T10d
serum, BD Vacutainer SSTFreeze–thawseparated−70 °C, 1 month; thawed at room temperature1 cyclestable for the longest time testedmean % change (bias) from fresh baseline within total allowable error (TEa, desirable bias)Abraham et al. 2021, J Lab PhysiciansTable 2 and Table 3, row 'Serum creatinine (mg/dL)'
serum, BD Vacutainer 8.5 mL plastic serum tube; pooled serum aliquoted into 1.5 mL Eppendorf tubesStorageseparated−20 °C3 monthsstable for the longest time testedunstable only if both P < 0.013 (Bonferroni) and % change from T0 > desirable bias (Westgard/Ricos), here 4.0%Cuhadar et al. 2013, Biochem Med (Zagreb)Table 2, row 'Creatinine (μmol/L)', columns T1m-T3m
serum, BD Vacutainer serum (plain), 7.5 mLStorageseparated−20 °C30 daysstable for the longest time testedclinically significant change: mean outside significant change limit (SCL = T1d mean ± 3.0 usual SD from 2 months QC)Kachhawa et al. 2017, J Lab PhysiciansResults paragraph 1 (Table 2 image-only)
serum, Vacuette Tube Serum Gel Separator 8 mL (Greiner, 455071)Storagecentrifuged-on-gel2-8 °C1 weekstable for the longest time tested% bias of median vs baseline exceeds desirable bias (Westgard/Ricos BV), here 4.0%Cuhadar et al. 2012, Biochem Med (Zagreb)Table 3, row 'Creatinine', gel 4°C
serum, Vacuette Tube Serum No Gel 9 mL (Greiner, 455092), plainStoragecentrifuged-on-cells2-8 °C48 hthen changed, changed by 72 h% bias of median vs baseline exceeds desirable bias (Westgard/Ricos BV), here 4.0%Cuhadar et al. 2012, Biochem Med (Zagreb)Table 3, row 'Creatinine', w/o gel 4°C
serum, BD Vacutainer RST 5 mL (thrombin clot activator, gel)Storagecentrifuged-on-gel2-8 °C7 daysstable for the longest time testedmean % difference vs time 0 exceeds desirable imprecision based on biological variation (EFLM/Westgard), here 2.25%Mandić et al. 2020, Biochem Med (Zagreb)Tables 3 and 4, row 'Creatinine, μmol/L', BD RST columns
serum, Vacuette Tube Serum Gel Separator 8 mL (Greiner, 455071)Storagecentrifuged-on-gelRT18 hthen changed, changed by 24 h% bias of median vs baseline exceeds desirable bias (Westgard/Ricos BV), here 4.0%Cuhadar et al. 2012, Biochem Med (Zagreb)Table 3, row 'Creatinine', gel 24°C
serum, Vacuette Tube Serum No Gel 9 mL (Greiner, 455092), plainStoragecentrifuged-on-cellsRT1 weekstable for the longest time tested% bias of median vs baseline exceeds desirable bias (Westgard/Ricos BV), here 4.0%Cuhadar et al. 2012, Biochem Med (Zagreb)Table 3, row 'Creatinine', w/o gel 24°C
serum, BD Vacutainer SST (SKU 367986), gelStoragecentrifuged-on-gel30 °C (±1 °C)24 hthen changed, changed by 72 hmean bias vs paired baseline within CLIA acceptance limits AND >2/3 of individual results within limitsIacovetti et al. 2026, Pract Lab MedResults, 'Analyte stability after storage at 30 °C', paragraph 1 (Table 1 image-only)
serum, BD SST Microtainer (SKU 395967), gel; filled from venous bloodStoragecentrifuged-on-gel0 °C72 hstable for the longest time testedmean bias vs paired baseline within CLIA acceptance limits AND >2/3 of individual results within limitsIacovetti et al. 2026, Pract Lab MedResults, 'Analyte stability after separation and storage in microvolume tubes at 0 °C' (Table 2 image-only)
whole-blood-clotted, BD Vacutainer SST (SKU 367986), gelBefore spinninguncentrifuged30 °C (±1 °C)0 hchanged at the first time point, changed by 24 hmean bias vs paired baseline within CLIA acceptance limits AND >2/3 of individual results within limitsIacovetti et al. 2026, Pract Lab MedResults, 'Analyte stability after storage at 30 °C', paragraph 2 (Table 1 image-only)
whole-blood-clotted, Sarstedt Monovette serumBefore spinninguncentrifugedchilled in transit (insulated box, 2 frozen cool packs; temperature not recorded)24 hstable for the longest time tested% difference within desirable quality specification for bias (Fraser, from intra- and interindividual variation)Jones et al. 2021, J NutrTable 2, row 'Creatinine', column Serum
whole-blood-edta, Sarstedt Monovette EDTABefore spinninguncentrifugedchilled in transit (insulated box, 2 frozen cool packs; temperature not recorded)24 hstable for the longest time tested% difference within desirable quality specification for bias (Fraser, from intra- and interindividual variation)Jones et al. 2021, J NutrTable 2, row 'Creatinine', column EDTA
whole-blood-heparin, BD Vacutainer lithium heparin, no gel (ref 368884)Before spinninguncentrifugedRT10 hstable for the longest time tested|mean % deviation 10 h vs baseline| > laboratory bias goal (Table II; mostly biological-variation based, some empirically modified)Henriksen et al. 2014, Scand J Clin Lab InvestTable II, row 'CREA'
whole-blood-heparin, BD Vacutainer lithium heparin, no gel (ref 368884)Before spinninguncentrifugedRT10 hstable for the longest time testedCV of paired baseline/10 h results > goal CV (Table II)Henriksen et al. 2014, Scand J Clin Lab InvestTable II, row 'CREA'
whole-blood-heparin, Sarstedt Monovette lithium heparin, trace-element freeBefore spinninguncentrifugedchilled in transit (insulated box, 2 frozen cool packs; temperature not recorded)24 hstable for the longest time tested% difference within desirable quality specification for bias (Fraser, from intra- and interindividual variation)Jones et al. 2021, J NutrTable 2, row 'Creatinine', column LH
Cystatin C2 findings
Specimen, tubeConditionTemperatureStable forCriterionStudy
dbs, Whatman 903 card, mailed in foil pouch with desiccantStoragedriednot stated; ambient postal shipping in the US (Apr-Dec 2016)8 daysstable for the longest time testedregression coefficient for days before freezing vs 0-2 days significant (P < .05), Model 1 (DBS value, unadjusted)Crimmins et al. 2020, Am J Hum BiolTable 3, rows 'Days before freezing', column 'Cystatin C'
dbs, Whatman 903 card, mailed in foil pouch with desiccantStoragedriednot stated; ambient postal shipping in the US (Apr-Dec 2016)7 daysthen changed, changed by 8 daysregression coefficient for days before freezing vs 0-2 days significant (P < .05), Model 2 (DBS value adjusted for venous value)Crimmins et al. 2020, Am J Hum BiolTable 3, rows 'Days before freezing', column 'Cystatin C'
D-Dimer7 findings
Specimen, tubeConditionTemperatureStable forCriterionStudy
plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquotsFreeze–thawseparated−20 °C1 cyclelimit stated by the authorsmax allowable mean deviation from t0: 0.2–0.5 μg FEU/mL: ±0.05 μg FEU/mL c ; >0.5 μg FEU/mL: ±10%Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability'
plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquotsFreeze–thawseparated−80 °C3 cycleslimit stated by the authorsmax allowable mean deviation from t0: 0.2–0.5 μg FEU/mL: ±0.05 μg FEU/mL c ; >0.5 μg FEU/mL: ±10%Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability'
plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquotsStorageseparated−20 °C28 dayslimit stated by the authorsmax allowable mean deviation from t0: 0.2–0.5 μg FEU/mL: ±0.05 μg FEU/mL c ; >0.5 μg FEU/mL: ±10%Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability'
plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquotsStorageseparated−20 °C11 weekslimit stated by the authorsmax allowable mean deviation from t0: 0.2–0.5 μg FEU/mL: ±0.05 μg FEU/mL c ; >0.5 μg FEU/mL: ±10%Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability'
plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquotsStorageseparated−80 °C23 monthslimit stated by the authorsmax allowable mean deviation from t0: 0.2–0.5 μg FEU/mL: ±0.05 μg FEU/mL c ; >0.5 μg FEU/mL: ±10%Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability'
plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquotsStorageseparated2-8 °C4 dayslimit stated by the authorsmax allowable mean deviation from t0: 0.2–0.5 μg FEU/mL: ±0.05 μg FEU/mL c ; >0.5 μg FEU/mL: ±10%Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability'
plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquotsStorageseparatedRT8 hlimit stated by the authorsmax allowable mean deviation from t0: 0.2–0.5 μg FEU/mL: ±0.05 μg FEU/mL c ; >0.5 μg FEU/mL: ±10%Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability'
Direct (Conjugated) Bilirubin5 findings
Specimen, tubeConditionTemperatureStable forCriterionStudy
plasma-li-heparin, Sarstedt lithium heparin, no gel separator (primary tube)Storagecentrifuged-on-cells2-8 °C48 hstable for the longest time testedmean % difference vs time 0 exceeds total limit of change, TLC = ±√((1.65·CVa)² + (0.5·CVb)²) (EFLM BV), here 20.8%; stability limit from instability-equation regressionIllana et al. 2024, Biochem Med (Zagreb)Table 1, row 'DBIL'
plasma-li-heparin, Sarstedt lithium heparin, no gel separator; plasma aliquotStorageseparated2-8 °C96 hstable for the longest time testedmean % difference vs time 0 exceeds total limit of change, TLC = ±√((1.65·CVa)² + (0.5·CVb)²) (EFLM BV), here 20.8%; stability limit from instability-equation regressionIllana et al. 2024, Biochem Med (Zagreb)Table 2, row 'DBIL'
serum, BD Vacutainer 8.5 mL plastic serum tube; pooled serum aliquoted into 1.5 mL Eppendorf tubesFreeze–thawseparated−20 °C10 cyclesstable for the longest time testedunstable only if both P < 0.005 (Bonferroni) and % change from T0 > desirable bias (Westgard/Ricos), here 14.2%Cuhadar et al. 2013, Biochem Med (Zagreb)Table 2, row 'Bilirubin, direct (μmol/L)', columns T1d-T10d
serum, BD Vacutainer 8.5 mL plastic serum tube; pooled serum aliquoted into 1.5 mL Eppendorf tubesStorageseparated−20 °C3 monthsstable for the longest time testedunstable only if both P < 0.013 (Bonferroni) and % change from T0 > desirable bias (Westgard/Ricos), here 14.2%Cuhadar et al. 2013, Biochem Med (Zagreb)Table 2, row 'Bilirubin, direct (μmol/L)', columns T1m-T3m
serum, BD Vacutainer serum (plain), 7.5 mLStorageseparated−20 °C30 daysstable for the longest time testedclinically significant change: mean outside significant change limit (SCL = T1d mean ± 3.0 usual SD from 2 months QC)Kachhawa et al. 2017, J Lab PhysiciansResults paragraph 1 (Table 2 image-only)
Eosinophils (Absolute)21 findings
Specimen, tubeConditionTemperatureStable forCriterionStudy
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C48 hstable for the longest time testedHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 4, row 'EO', XN-9000, 4 °C
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C48 hstable for the longest time testedBias% vs optimal bias (OP-Bias%, from biological variation; Table 1)Buoro et al. 2016, Rev Bras Hematol HemoterTable 4, row 'EO', XN-9000, 4 °C
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C48 hstable for the longest time testedHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 4, row 'EO', BC-6800, 4 °C
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C48 hstable for the longest time testedBias% vs optimal bias (OP-Bias%, from biological variation; Table 1)Buoro et al. 2016, Rev Bras Hematol HemoterTable 4, row 'EO', BC-6800, 4 °C
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C8 hstable for the longest time testedHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 5, row 'EO', XN-9000, 4 °C
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C4 hthen changed, changed by 8 hBias% vs optimal bias (OP-Bias%, from biological variation; Table 1)Buoro et al. 2016, Rev Bras Hematol HemoterTable 5, row 'EO', XN-9000, 4 °C
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C8 hstable for the longest time testedBias% vs critical difference (CD%, reference change value; Table 1)Buoro et al. 2016, Rev Bras Hematol HemoterTable 5, row 'EO', XN-9000, 4 °C
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C8 hstable for the longest time testedHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 5, row 'EO', BC-6800, 4 °C
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C8 hstable for the longest time testedBias% vs optimal bias (OP-Bias%, from biological variation; Table 1)Buoro et al. 2016, Rev Bras Hematol HemoterTable 5, row 'EO', BC-6800, 4 °C
whole-blood-edta, K2-EDTA (BD Vacutainer)Storageuncentrifuged2-8 °C48 hstable for the longest time testedRepeated-measures one-way ANOVA with Tukey post hoc vs 0 h, p < 0.05 (* in table)Unalli et al. 2021, J Med BiochemTable 1, column '4 °C' (no * at any time point)
whole-blood-edta, K3-EDTA (Becton Dickinson)StorageuncentrifugedRT48 hstable for the longest time testedHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 4, row 'EO', XN-9000, RT
whole-blood-edta, K3-EDTA (Becton Dickinson)StorageuncentrifugedRT48 hstable for the longest time testedBias% vs optimal bias (OP-Bias%, from biological variation; Table 1)Buoro et al. 2016, Rev Bras Hematol HemoterTable 4, row 'EO', XN-9000, RT
whole-blood-edta, K3-EDTA (Becton Dickinson)StorageuncentrifugedRT48 hstable for the longest time testedHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 4, row 'EO', BC-6800, RT
whole-blood-edta, K3-EDTA (Becton Dickinson)StorageuncentrifugedRT48 hstable for the longest time testedBias% vs optimal bias (OP-Bias%, from biological variation; Table 1)Buoro et al. 2016, Rev Bras Hematol HemoterTable 4, row 'EO', BC-6800, RT
whole-blood-edta, K3-EDTA (Becton Dickinson)StorageuncentrifugedRT8 hstable for the longest time testedHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 5, row 'EO', XN-9000, RT
whole-blood-edta, K3-EDTA (Becton Dickinson)StorageuncentrifugedRT8 hstable for the longest time testedBias% vs optimal bias (OP-Bias%, from biological variation; Table 1)Buoro et al. 2016, Rev Bras Hematol HemoterTable 5, row 'EO', XN-9000, RT
whole-blood-edta, K3-EDTA (Becton Dickinson)StorageuncentrifugedRT8 hstable for the longest time testedHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 5, row 'EO', BC-6800, RT
whole-blood-edta, K3-EDTA (Becton Dickinson)StorageuncentrifugedRT4 hthen changed, changed by 8 hBias% vs optimal bias (OP-Bias%, from biological variation; Table 1)Buoro et al. 2016, Rev Bras Hematol HemoterTable 5, row 'EO', BC-6800, RT
whole-blood-edta, K3-EDTA (Becton Dickinson)StorageuncentrifugedRT8 hstable for the longest time testedBias% vs critical difference (CD%, reference change value; Table 1)Buoro et al. 2016, Rev Bras Hematol HemoterTable 5, row 'EO', BC-6800, RT
whole-blood-edta, K2-EDTA (BD Vacutainer)StorageuncentrifugedRT4 hthen changed, changed by 8 hRepeated-measures one-way ANOVA with Tukey post hoc vs 0 h, p < 0.05 (* in table)Unalli et al. 2021, J Med BiochemTable 1, block 'EOS', row '8 h', column '23 °C'
whole-blood-edta, K2-EDTA (BD Vacutainer)Storageuncentrifuged10 °C48 hstable for the longest time testedRepeated-measures one-way ANOVA with Tukey post hoc vs 0 h, p < 0.05 (* in table)Unalli et al. 2021, J Med BiochemTable 1, column '10 °C' (no * at any time point)
Erythrocyte glutathione reductase activation coefficient (EGRAC)3 findings
Specimen, tubeConditionTemperatureStable forCriterionStudy
whole-blood-edta, Sarstedt Monovette EDTABefore spinninguncentrifugedchilled in transit (insulated box, 2 frozen cool packs; temperature not recorded)24 hstable for the longest time tested% difference within desirable quality specification for bias (Fraser, from intra- and interindividual variation)Jones et al. 2021, J NutrTable 2, row 'EGRAC (riboflavin)', column EDTA
whole-blood-heparin, BD Vacutainer lithium heparinBefore spinninguncentrifugedcold packs in insulated box; 9 (2) °C on arrival24 hstable for the longest time tested% difference within desirable quality specification (from intra- and interindividual variation)Jones et al. 2024, J NutrTable 2, row 'EGRAC (B2) [ratio]'; Table 1
whole-blood-heparin, Sarstedt Monovette lithium heparin, trace-element freeBefore spinninguncentrifugedchilled in transit (insulated box, 2 frozen cool packs; temperature not recorded)24 hstable for the longest time tested% difference within desirable quality specification for bias (Fraser, from intra- and interindividual variation)Jones et al. 2021, J NutrTable 2, row 'EGRAC (riboflavin)', column LH
Erythrocyte Sedimentation Rate5 findings
Specimen, tubeConditionTemperatureStable forCriterionStudy
whole-blood-edta, BD Vacutainer K2-EDTA 3.0 mLStorageuncentrifuged2-8 °C36 hthen changed, changed by 48 hstatistically significant difference vs T0 (p < 0.05)Lorubbio et al. 2024, Diagnostics (Basel)Results, paragraph 3
whole-blood-edta, BD Vacutainer K2 EDTAStorageuncentrifuged2-8 °C48 hstable for the longest time testedPassing-Bablok vs baseline: 95% CI of slope includes 1.00, 95% CI of intercept includes 0.00, Spearman r >= 0.90Koshy et al. 2026, Diagnostics (Basel)Table 3
whole-blood-edta, BD Vacutainer K2-EDTA 3.0 mLStorageuncentrifugedRT0 hchanged at the first time point, changed by 4 hstatistically significant difference vs T0 (p < 0.05)Lorubbio et al. 2024, Diagnostics (Basel)Results, paragraph 2
whole-blood-edta, BD Vacutainer K2 EDTAStorageuncentrifugedRT10 hthen changed, changed by 12 hPassing-Bablok vs baseline: 95% CI of slope includes 1.00, 95% CI of intercept includes 0.00, Spearman r >= 0.90Koshy et al. 2026, Diagnostics (Basel)Table 3
whole-blood-edta, BD Vacutainer K2 EDTAStorageuncentrifugedRT28 hthen changed, changed by 36 hPassing-Bablok vs baseline: 95% CI of slope includes 1.00, 95% CI of intercept includes 0.00, Spearman r >= 0.90Koshy et al. 2026, Diagnostics (Basel)Table 3
Erythrocyte transketolase activation coefficient (ETKAC)3 findings
Specimen, tubeConditionTemperatureStable forCriterionStudy
whole-blood-edta, Sarstedt Monovette EDTABefore spinninguncentrifugedchilled in transit (insulated box, 2 frozen cool packs; temperature not recorded)24 hstable for the longest time tested% difference within desirable quality specification for bias (Fraser, from intra- and interindividual variation)Jones et al. 2021, J NutrTable 2, row 'ETKAC (thiamin)', column EDTA
whole-blood-heparin, BD Vacutainer lithium heparinBefore spinninguncentrifugedcold packs in insulated box; 9 (2) °C on arrival24 hstable for the longest time tested% difference within desirable quality specification (from intra- and interindividual variation)Jones et al. 2024, J NutrTable 2, row 'ETKAC (B1) [ratio]'; Table 1
whole-blood-heparin, Sarstedt Monovette lithium heparin, trace-element freeBefore spinninguncentrifugedchilled in transit (insulated box, 2 frozen cool packs; temperature not recorded)24 hstable for the longest time tested% difference within desirable quality specification for bias (Fraser, from intra- and interindividual variation)Jones et al. 2021, J NutrTable 2, row 'ETKAC (thiamin)', column LH
Ferritin12 findings
Specimen, tubeConditionTemperatureStable forCriterionStudy
serum, VACUMED 16x100 mm, inert separator gel and clot activatorFreeze–thawseparated−80 °C0 cycleschanged at the first time point, changed by 2 cyclesStudent t test for paired samples vs fresh measurement (SF1), p < 0.05 impliedSacri et al. 2017, PLoS OneResults, paragraph 3
serum, BD Vacutainer SSTFreeze–thawseparated−70 °C, 1 month; thawed at room temperature1 cyclestable for the longest time testedmean % change (bias) from fresh baseline within total allowable error (TEa, desirable bias)Abraham et al. 2021, J Lab PhysiciansTable 2 and Table 3, row 'Ferritin (ng/mL)'
serum, VACUMED 16x100 mm, inert separator gel and clot activatorStorageseparated−80 °C0 yearschanged at the first time point, changed by 3 yearsStudent t test for paired samples vs fresh measurement (SF1), p < 0.05 impliedSacri et al. 2017, PLoS OneResults, paragraph 2
whole-blood-clotted, BD Vacutainer serum, no gel (ref 368815)Before spinninguncentrifugedRT10 hstable for the longest time tested|mean % deviation 10 h vs baseline| > laboratory bias goal (Table II; mostly biological-variation based, some empirically modified)Henriksen et al. 2014, Scand J Clin Lab InvestTable II, row 'FER'
whole-blood-clotted, BD Vacutainer serum, no gel (ref 368815)Before spinninguncentrifugedRT10 hstable for the longest time testedCV of paired baseline/10 h results > goal CV (Table II)Henriksen et al. 2014, Scand J Clin Lab InvestTable II, row 'FER'
whole-blood-clotted, BD Vacutainer SSTBefore spinninguncentrifugedRT24 hstable for the longest time testedmean % deviation from baseline within total change limit TCL = sqrt((2.77 CVa)^2 + (0.5 CVb)^2), CVb from Ricos (Table 2)Abraham et al. 2019, Biochem Med (Zagreb)Results, paragraph 2 (Figure 3f)
whole-blood-clotted, BD Vacutainer trace element-free serum tubeBefore spinninguncentrifugedcold packs in insulated box; 9 (2) °C on arrival0 hchanged at the first time point, changed by 24 h% difference within desirable quality specification (from intra- and interindividual variation)Jones et al. 2024, J NutrTable 2, row 'Ferritin, μg/L'; Table 1
whole-blood-clotted, BD Vacutainer SSTBefore spinninguncentrifuged22-30 °C (logger 24-26 °C)24 hstable for the longest time testedmean % deviation from baseline within total change limit TCL = sqrt((2.77 CVa)^2 + (0.5 CVb)^2), CVb from Ricos (Table 2)Abraham et al. 2019, Biochem Med (Zagreb)Results, paragraph 2 (Figure 3f)
whole-blood-clotted, BD Vacutainer SSTBefore spinninguncentrifuged> 30 °C (logger 38-40 °C)12 hthen changed, changed by 24 hmean % deviation from baseline within total change limit TCL = sqrt((2.77 CVa)^2 + (0.5 CVb)^2), CVb from Ricos (Table 2)Abraham et al. 2019, Biochem Med (Zagreb)Results, paragraph 2 (Figure 3f)
whole-blood-clotted, Sarstedt Monovette serumBefore spinninguncentrifugedchilled in transit (insulated box, 2 frozen cool packs; temperature not recorded)24 hstable for the longest time tested% difference within desirable quality specification for bias (Fraser, from intra- and interindividual variation)Jones et al. 2021, J NutrTable 2, row 'Ferritin', column Serum
whole-blood-edta, Sarstedt Monovette EDTABefore spinninguncentrifugedchilled in transit (insulated box, 2 frozen cool packs; temperature not recorded)24 hstable for the longest time tested% difference within desirable quality specification for bias (Fraser, from intra- and interindividual variation)Jones et al. 2021, J NutrTable 2, row 'Ferritin', column EDTA
whole-blood-heparin, Sarstedt Monovette lithium heparin, trace-element freeBefore spinninguncentrifugedchilled in transit (insulated box, 2 frozen cool packs; temperature not recorded)24 hstable for the longest time tested% difference within desirable quality specification for bias (Fraser, from intra- and interindividual variation)Jones et al. 2021, J NutrTable 2, row 'Ferritin', column LH
Fibrinogen7 findings
Specimen, tubeConditionTemperatureStable forCriterionStudy
plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquotsFreeze–thawseparated−20 °C1 cyclelimit stated by the authorsmax allowable mean deviation from t0: ±10%Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability'
plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquotsFreeze–thawseparated−80 °C1 cyclelimit stated by the authorsmax allowable mean deviation from t0: ±10%Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability'
plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquotsStorageseparated−20 °C7 weekslimit stated by the authorsmax allowable mean deviation from t0: ±10%Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability'
plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquotsStorageseparated−80 °C11 monthslimit stated by the authorsmax allowable mean deviation from t0: ±10%Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability'
plasma-citrate, BD 0.109 M sodium citrate (9:1); plasma aliquoted into nonactivating Eppendorf tubesStorageseparated2-8 °C24 hstable for the longest time testedmean % change from baseline >10%, or >25% of samples with individual change >10% (van Geest-Daalderop)Feng et al. 2014, Sci RepTable 2, 'Acceptable Time (h)'
plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquotsStorageseparatedRT4 hlimit stated by the authorsmax allowable mean deviation from t0: ±10%Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability'
plasma-citrate, BD 0.109 M sodium citrate (9:1); plasma aliquoted into nonactivating Eppendorf tubesStorageseparatedRT24 hstable for the longest time testedmean % change from baseline >10%, or >25% of samples with individual change >10% (van Geest-Daalderop)Feng et al. 2014, Sci RepTable 2, 'Acceptable Time (h)'
Fibrinogen (PT-derived)5 findings
Specimen, tubeConditionTemperatureStable forCriterionStudy
plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquotsFreeze–thawseparated−20 °C1 cyclelimit stated by the authorsmax allowable mean deviation from t0: ±10%Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability'
plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquotsFreeze–thawseparated−80 °C1 cyclelimit stated by the authorsmax allowable mean deviation from t0: ±10%Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability'
plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquotsStorageseparated−20 °C7 weekslimit stated by the authorsmax allowable mean deviation from t0: ±10%Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability'
plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquotsStorageseparated−80 °C11 monthslimit stated by the authorsmax allowable mean deviation from t0: ±10%Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability'
plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquotsStorageseparatedRT12 hlimit stated by the authorsmax allowable mean deviation from t0: ±10%Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability'
Folate (Red Blood Cell)1 finding
Specimen, tubeConditionTemperatureStable forCriterionStudy
other, BD Vacutainer EDTA; haemolysate in folate ascorbic acidStorageuncentrifuged> 30 °C (logger 38-40 °C)0 hchanged at the first time point, changed by 6 hmean % deviation from baseline within total change limit TCL = sqrt((2.77 CVa)^2 + (0.5 CVb)^2), CVb from Ricos (Table 2)Abraham et al. 2019, Biochem Med (Zagreb)Results, paragraph 1 (Figure 3d)
Folate (serum/plasma)7 findings
Specimen, tubeConditionTemperatureStable forCriterionStudy
serum, serum (tube not stated)Storageseparated−70 °C for the first year, −80 °C thereafter0 yearschanged at the first time point, changed by 1 yeart-test (two-sample, equal variances) vs t=0, P < 0.05Jansen et al. 2018, J Nutr MetabResults, section 3.2
whole-blood-clotted, BD Vacutainer serum, no gel (ref 368815)Before spinninguncentrifugedRT10 hstable for the longest time tested|mean % deviation 10 h vs baseline| > laboratory bias goal (Table II; mostly biological-variation based, some empirically modified)Henriksen et al. 2014, Scand J Clin Lab InvestTable II, row 'FOL'
whole-blood-clotted, BD Vacutainer serum, no gel (ref 368815)Before spinninguncentrifugedRT0 hchanged at the first time point, changed by 10 hCV of paired baseline/10 h results > goal CV (Table II)Henriksen et al. 2014, Scand J Clin Lab InvestTable II, row 'FOL'
whole-blood-clotted, BD Vacutainer trace element-free serum tubeBefore spinninguncentrifugedcold packs in insulated box; 9 (2) °C on arrival24 hstable for the longest time tested% difference within desirable quality specification (from intra- and interindividual variation)Jones et al. 2024, J NutrTable 2, row 'Total folate, nmol/L'; Table 1
whole-blood-clotted, Sarstedt Monovette serumBefore spinninguncentrifugedchilled in transit (insulated box, 2 frozen cool packs; temperature not recorded)24 hstable for the longest time tested% difference within desirable quality specification for bias (Fraser, from intra- and interindividual variation)Jones et al. 2021, J NutrTable 2, row 'Total folate', column Serum
whole-blood-edta, Sarstedt Monovette EDTABefore spinninguncentrifugedchilled in transit (insulated box, 2 frozen cool packs; temperature not recorded)24 hstable for the longest time tested% difference within desirable quality specification for bias (Fraser, from intra- and interindividual variation)Jones et al. 2021, J NutrTable 2, row 'Total folate', column EDTA
whole-blood-heparin, Sarstedt Monovette lithium heparin, trace-element freeBefore spinninguncentrifugedchilled in transit (insulated box, 2 frozen cool packs; temperature not recorded)24 hstable for the longest time tested% difference within desirable quality specification for bias (Fraser, from intra- and interindividual variation)Jones et al. 2021, J NutrTable 2, row 'Total folate', column LH
Free Thyroxine6 findings
Specimen, tubeConditionTemperatureStable forCriterionStudy
plasma-li-heparin, Sarstedt lithium heparin, no gel separator (primary tube)Storagecentrifuged-on-cells2-8 °C48 hstable for the longest time testedmean % difference vs time 0 exceeds total limit of change, TLC = ±√((1.65·CVa)² + (0.5·CVb)²) (EFLM BV), here 18.9%; stability limit from instability-equation regressionIllana et al. 2024, Biochem Med (Zagreb)Table 1, row 'fT4'
plasma-li-heparin, Sarstedt lithium heparin, no gel separator; plasma aliquotStorageseparated2-8 °C96 hstable for the longest time testedmean % difference vs time 0 exceeds total limit of change, TLC = ±√((1.65·CVa)² + (0.5·CVb)²) (EFLM BV), here 18.9%; stability limit from instability-equation regressionIllana et al. 2024, Biochem Med (Zagreb)Table 2, row 'fT4'
plasma-li-heparin, BD Barricor 4.5 mL (lithium heparin, mechanical separator)Storagecentrifuged-on-gel2-8 °C7 daysstable for the longest time testedmean % difference vs time 0 exceeds desirable imprecision based on biological variation (EFLM/Westgard), here 3.85%Mandić et al. 2020, Biochem Med (Zagreb)Tables 3 and 4, row 'fT4, pmol/L', BD Barricor columns
serum, BD Vacutainer RST 5 mL (thrombin clot activator, gel)Storagecentrifuged-on-gel2-8 °C7 daysstable for the longest time testedmean % difference vs time 0 exceeds desirable imprecision based on biological variation (EFLM/Westgard), here 3.85%Mandić et al. 2020, Biochem Med (Zagreb)Tables 3 and 4, row 'fT4, pmol/L', BD RST columns
whole-blood-clotted, BD Vacutainer serum, no gel (ref 368815)Before spinninguncentrifugedRT10 hstable for the longest time tested|mean % deviation 10 h vs baseline| > laboratory bias goal (Table II; mostly biological-variation based, some empirically modified)Henriksen et al. 2014, Scand J Clin Lab InvestTable II, row 'F-T4'
whole-blood-clotted, BD Vacutainer serum, no gel (ref 368815)Before spinninguncentrifugedRT10 hstable for the longest time testedCV of paired baseline/10 h results > goal CV (Table II)Henriksen et al. 2014, Scand J Clin Lab InvestTable II, row 'F-T4'
Free Triiodothyronine6 findings
Specimen, tubeConditionTemperatureStable forCriterionStudy
plasma-li-heparin, Sarstedt lithium heparin, no gel separator (primary tube)Storagecentrifuged-on-cells2-8 °C48 hstable for the longest time testedmean % difference vs time 0 exceeds total limit of change, TLC = ±√((1.65·CVa)² + (0.5·CVb)²) (EFLM BV), here 14.6%; stability limit from instability-equation regressionIllana et al. 2024, Biochem Med (Zagreb)Table 1, row 'fT3'
plasma-li-heparin, Sarstedt lithium heparin, no gel separator; plasma aliquotStorageseparated2-8 °C96 hstable for the longest time testedmean % difference vs time 0 exceeds total limit of change, TLC = ±√((1.65·CVa)² + (0.5·CVb)²) (EFLM BV), here 14.6%; stability limit from instability-equation regressionIllana et al. 2024, Biochem Med (Zagreb)Table 2, row 'fT3'
plasma-li-heparin, BD Barricor 4.5 mL (lithium heparin, mechanical separator)Storagecentrifuged-on-gel2-8 °C7 daysstable for the longest time testedmean % difference vs time 0 exceeds desirable imprecision based on biological variation (EFLM/Westgard), here 3.0%Mandić et al. 2020, Biochem Med (Zagreb)Tables 3 and 4, row 'fT3, pmol/L', BD Barricor columns
serum, BD Vacutainer RST 5 mL (thrombin clot activator, gel)Storagecentrifuged-on-gel2-8 °C0 hchanged at the first time point, changed by 24 hmean % difference vs time 0 exceeds desirable imprecision based on biological variation (EFLM/Westgard), here 3.0%Mandić et al. 2020, Biochem Med (Zagreb)Tables 3 and 4, row 'fT3, pmol/L', BD RST columns
whole-blood-clotted, BD Vacutainer serum, no gel (ref 368815)Before spinninguncentrifugedRT10 hstable for the longest time tested|mean % deviation 10 h vs baseline| > laboratory bias goal (Table II; mostly biological-variation based, some empirically modified)Henriksen et al. 2014, Scand J Clin Lab InvestTable II, row 'F-T3'
whole-blood-clotted, BD Vacutainer serum, no gel (ref 368815)Before spinninguncentrifugedRT10 hstable for the longest time testedCV of paired baseline/10 h results > goal CV (Table II)Henriksen et al. 2014, Scand J Clin Lab InvestTable II, row 'F-T3'
Gamma-Glutamyl Transferase (GGT)12 findings
Specimen, tubeConditionTemperatureStable forCriterionStudy
plasma-li-heparin, Sarstedt lithium heparin, no gel separator (primary tube)Storagecentrifuged-on-cells2-8 °C48 hstable for the longest time testedmean % difference vs time 0 exceeds total limit of change, TLC = ±√((1.65·CVa)² + (0.5·CVb)²) (EFLM BV), here 9.7%; stability limit from instability-equation regressionIllana et al. 2024, Biochem Med (Zagreb)Table 1, row 'GGT'
plasma-li-heparin, Sarstedt lithium heparin, no gel separator; plasma aliquotStorageseparated2-8 °C96 hstable for the longest time testedmean % difference vs time 0 exceeds total limit of change, TLC = ±√((1.65·CVa)² + (0.5·CVb)²) (EFLM BV), here 9.7%; stability limit from instability-equation regressionIllana et al. 2024, Biochem Med (Zagreb)Table 2, row 'GGT'
plasma-li-heparin, BD Barricor 4.5 mL (lithium heparin, mechanical separator)Storagecentrifuged-on-gel2-8 °C7 daysstable for the longest time testedmean % difference vs time 0 exceeds desirable imprecision based on biological variation (EFLM/Westgard), here 4.4%Mandić et al. 2020, Biochem Med (Zagreb)Tables 3 and 4, row 'GGT, U/L', BD Barricor columns
serum, BD Vacutainer 8.5 mL plastic serum tube; pooled serum aliquoted into 1.5 mL Eppendorf tubesFreeze–thawseparated−20 °C10 cyclesstable for the longest time testedunstable only if both P < 0.005 (Bonferroni) and % change from T0 > desirable bias (Westgard/Ricos), here 10.8%Cuhadar et al. 2013, Biochem Med (Zagreb)Table 2, row 'GGT (U/L)', columns T1d-T10d
serum, BD Vacutainer 8.5 mL plastic serum tube; pooled serum aliquoted into 1.5 mL Eppendorf tubesStorageseparated−20 °C3 monthsstable for the longest time testedunstable only if both P < 0.013 (Bonferroni) and % change from T0 > desirable bias (Westgard/Ricos), here 10.8%Cuhadar et al. 2013, Biochem Med (Zagreb)Table 2, row 'GGT (U/L)', columns T1m-T3m
serum, Vacuette Tube Serum Gel Separator 8 mL (Greiner, 455071)Storagecentrifuged-on-gel2-8 °C1 weekstable for the longest time tested% bias of median vs baseline exceeds desirable bias (Westgard/Ricos BV), here 10.8%Cuhadar et al. 2012, Biochem Med (Zagreb)Table 3, row 'γ-Glutamyltransferase', gel 4°C
serum, Vacuette Tube Serum No Gel 9 mL (Greiner, 455092), plainStoragecentrifuged-on-cells2-8 °C1 weekstable for the longest time tested% bias of median vs baseline exceeds desirable bias (Westgard/Ricos BV), here 10.8%Cuhadar et al. 2012, Biochem Med (Zagreb)Table 3, row 'γ-Glutamyltransferase', w/o gel 4°C
serum, BD Vacutainer RST 5 mL (thrombin clot activator, gel)Storagecentrifuged-on-gel2-8 °C7 daysstable for the longest time testedmean % difference vs time 0 exceeds desirable imprecision based on biological variation (EFLM/Westgard), here 4.4%Mandić et al. 2020, Biochem Med (Zagreb)Tables 3 and 4, row 'GGT, U/L', BD RST columns
serum, Vacuette Tube Serum Gel Separator 8 mL (Greiner, 455071)Storagecentrifuged-on-gelRT1 weekstable for the longest time tested% bias of median vs baseline exceeds desirable bias (Westgard/Ricos BV), here 10.8%Cuhadar et al. 2012, Biochem Med (Zagreb)Table 3, row 'γ-Glutamyltransferase', gel 24°C
serum, Vacuette Tube Serum No Gel 9 mL (Greiner, 455092), plainStoragecentrifuged-on-cellsRT1 weekstable for the longest time tested% bias of median vs baseline exceeds desirable bias (Westgard/Ricos BV), here 10.8%Cuhadar et al. 2012, Biochem Med (Zagreb)Table 3, row 'γ-Glutamyltransferase', w/o gel 24°C
whole-blood-heparin, BD Vacutainer lithium heparin, no gel (ref 368884)Before spinninguncentrifugedRT10 hstable for the longest time tested|mean % deviation 10 h vs baseline| > laboratory bias goal (Table II; mostly biological-variation based, some empirically modified)Henriksen et al. 2014, Scand J Clin Lab InvestTable II, row 'GGT'
whole-blood-heparin, BD Vacutainer lithium heparin, no gel (ref 368884)Before spinninguncentrifugedRT10 hstable for the longest time testedCV of paired baseline/10 h results > goal CV (Table II)Henriksen et al. 2014, Scand J Clin Lab InvestTable II, row 'GGT'
Gamma-tocopherol4 findings
Specimen, tubeConditionTemperatureStable forCriterionStudy
whole-blood-clotted, Sarstedt Monovette serumBefore spinninguncentrifugedchilled in transit (insulated box, 2 frozen cool packs; temperature not recorded)24 hstable for the longest time tested% difference within desirable quality specification for bias (Fraser, from intra- and interindividual variation)Jones et al. 2021, J NutrTable 2, row 'γ-tocopherol', column Serum
whole-blood-edta, Sarstedt Monovette EDTABefore spinninguncentrifugedchilled in transit (insulated box, 2 frozen cool packs; temperature not recorded)24 hstable for the longest time tested% difference within desirable quality specification for bias (Fraser, from intra- and interindividual variation)Jones et al. 2021, J NutrTable 2, row 'γ-tocopherol', column EDTA
whole-blood-heparin, BD Vacutainer lithium heparinBefore spinninguncentrifugedcold packs in insulated box; 9 (2) °C on arrival24 hstable for the longest time tested% difference within desirable quality specification (from intra- and interindividual variation)Jones et al. 2024, J NutrTable 2, row 'γ-Tocopherol, μmol/L'; Table 1
whole-blood-heparin, Sarstedt Monovette lithium heparin, trace-element freeBefore spinninguncentrifugedchilled in transit (insulated box, 2 frozen cool packs; temperature not recorded)24 hstable for the longest time tested% difference within desirable quality specification for bias (Fraser, from intra- and interindividual variation)Jones et al. 2021, J NutrTable 2, row 'γ-tocopherol', column LH
Glucose30 findings
Specimen, tubeConditionTemperatureStable forCriterionStudy
plasma-edta, BD Vacutainer EDTA, 4 mLFreeze–thawseparated−20 °C4 cyclesstable for the longest time testedsignificant change vs day 1 (repeated-measures ANOVA, Tukey HSD post hoc, p<0.05)Gojković et al. 2024, J Med BiochemResults paragraph 3 (EDTA -20°C, Figure 3)
plasma-edta, BD Vacutainer EDTA, 4 mLStorageseparated2-8 °C3 daysthen changed, changed by 4 dayssignificant change vs day 1 (repeated-measures ANOVA, Tukey HSD post hoc, p<0.05)Gojković et al. 2024, J Med BiochemResults paragraph 1 (EDTA+4°C, Figure 1)
plasma-fluoride, BD Vacutainer sodium fluoride/potassium oxalate, 4 mLFreeze–thawseparated−20 °C3 cyclesthen changed, changed by 4 cyclessignificant change vs day 1 (repeated-measures ANOVA, Tukey HSD post hoc, p<0.05)Gojković et al. 2024, J Med BiochemResults paragraph 4 (NaF -20°C, Figure 4)
plasma-fluoride, BD Vacutainer fluorideFreeze–thawseparated−70 °C, 1 month; thawed at room temperature1 cyclestable for the longest time testedmean % change (bias) from fresh baseline within total allowable error (TEa, desirable bias)Abraham et al. 2021, J Lab PhysiciansTable 2 and Table 3, row 'Glucose (mg/dL)'
plasma-fluoride, sodium fluoride/potassium oxalate (NaF/KOx)Storagenot-stated2-8 °C72 hstable for the longest time testedno significant change in mixed effects model vs first analysis (overall p = 0.25)Dibbasey et al. 2024, Pract Lab MedTable 2; Results
plasma-fluoride, BD Vacutainer sodium fluoride/potassium oxalate, 4 mLStorageseparated2-8 °C4 daysstable for the longest time testedsignificant change vs day 1 (repeated-measures ANOVA, Tukey HSD post hoc, p<0.05)Gojković et al. 2024, J Med BiochemResults paragraph 2 (NaF+4°C, Figure 2)
plasma-li-heparin, Sarstedt lithium heparin, no gel separator (primary tube)Storagecentrifuged-on-cells2-8 °C2.7 hthen changed, changed by 4 hmean % difference vs time 0 exceeds total limit of change, TLC = ±√((1.65·CVa)² + (0.5·CVb)²) (EFLM BV), here 6.3%; stability limit from instability-equation regressionIllana et al. 2024, Biochem Med (Zagreb)Table 1, row 'Glc'
plasma-li-heparin, Sarstedt lithium heparin, no gel separator; plasma aliquotStorageseparated2-8 °C96 hstable for the longest time testedmean % difference vs time 0 exceeds total limit of change, TLC = ±√((1.65·CVa)² + (0.5·CVb)²) (EFLM BV), here 6.3%; stability limit from instability-equation regressionIllana et al. 2024, Biochem Med (Zagreb)Table 2, row 'Glc'
plasma-li-heparin, BD Vacutainer LH PST II gel separator, 3.0 mL (REF 367374)Storagecentrifuged-on-gel2-8 °C24 hstable for the longest time testedmean % difference smaller than TEa (Ricos: 0.25 x (CVi^2 + CVg^2)^0.5 + 1.65 x 0.5 x CVi)Nonkes et al. 2019, Biochem Med (Zagreb)Table 1, row 'Glucose, mmol/L'
plasma-li-heparin, BD Vacutainer Barricor LH 3 mL, mechanical separatorStoragecentrifuged-on-gel2-8 °C18 hlimit stated by the authorsbias vs aliquoted plasma reference tube > maximal allowable bias 2.5% (Ricos et al.); time to MAB from linear regression of bias on time to analysisGawria et al. 2020, J Clin Lab AnalTable 2, row 'Glucose', Barricor BD
plasma-li-heparin, BD Vacutainer PST LH 3 mL, gelStoragecentrifuged-on-gel2-8 °C28 hlimit stated by the authorsbias vs aliquoted plasma reference tube > maximal allowable bias 2.5% (Ricos et al.); time to MAB from linear regression of bias on time to analysisGawria et al. 2020, J Clin Lab AnalTable 2, row 'Glucose', Gel BD
plasma-li-heparin, Greiner Vacuette LH 3 mL Sep, gelStoragecentrifuged-on-gel2-8 °C26 hlimit stated by the authorsbias vs aliquoted plasma reference tube > maximal allowable bias 2.5% (Ricos et al.); time to MAB from linear regression of bias on time to analysisGawria et al. 2020, J Clin Lab AnalTable 2, row 'Glucose', Gel Greiner
plasma-li-heparin, BD Barricor 4.5 mL (lithium heparin, mechanical separator)Storagecentrifuged-on-gel2-8 °C0 hchanged at the first time point, changed by 24 hmean % difference vs time 0 exceeds desirable imprecision based on biological variation (EFLM/Westgard), here 2.55%Mandić et al. 2020, Biochem Med (Zagreb)Tables 3 and 4, row 'Glucose, mmol/L', BD Barricor columns
serum, BD Vacutainer 8.5 mL plastic serum tube; pooled serum aliquoted into 1.5 mL Eppendorf tubesFreeze–thawseparated−20 °C10 cyclesstable for the longest time testedunstable only if both P < 0.005 (Bonferroni) and % change from T0 > desirable bias (Westgard/Ricos), here 2.2%Cuhadar et al. 2013, Biochem Med (Zagreb)Table 2, row 'Glucose (mmol/L)', columns T1d-T10d
serum, BD Vacutainer 8.5 mL plastic serum tube; pooled serum aliquoted into 1.5 mL Eppendorf tubesStorageseparated−20 °C3 monthsstable for the longest time testedunstable only if both P < 0.013 (Bonferroni) and % change from T0 > desirable bias (Westgard/Ricos), here 2.2%Cuhadar et al. 2013, Biochem Med (Zagreb)Table 2, row 'Glucose (mmol/L)', columns T1m-T3m
serum, serum separator tube (SST)Storagenot-stated2-8 °C72 hstable for the longest time testedno significant change in mixed effects model vs first analysis (overall p = 0.25)Dibbasey et al. 2024, Pract Lab MedTable 2; Results
serum, Vacuette Tube Serum Gel Separator 8 mL (Greiner, 455071)Storagecentrifuged-on-gel2-8 °C36 hthen changed, changed by 48 h% bias of median vs baseline exceeds desirable bias (Westgard/Ricos BV), here 2.2%Cuhadar et al. 2012, Biochem Med (Zagreb)Table 3, row 'Glucose', gel 4°C
serum, Vacuette Tube Serum No Gel 9 mL (Greiner, 455092), plainStoragecentrifuged-on-cells2-8 °C0 hchanged at the first time point, changed by 6 h% bias of median vs baseline exceeds desirable bias (Westgard/Ricos BV), here 2.2%Cuhadar et al. 2012, Biochem Med (Zagreb)Table 3, row 'Glucose', w/o gel 4°C
serum, BD Vacutainer RST 4.0 mL (thrombin clot activator, gel)Storagecentrifuged-on-gel2-8 °C24 hstable for the longest time testedclinically significant if mean bias vs t0 exceeds desirable bias (biological variation), here 1.5%Kocijancic et al. 2014, Biochem Med (Zagreb)Table 3, row 'Glu'
serum, BD Vacutainer SST II Advance 5.0 mL (clot activator, gel)Storagecentrifuged-on-gel2-8 °C24 hstable for the longest time testedclinically significant if mean bias vs t0 exceeds desirable bias (biological variation), here 1.5%Kocijancic et al. 2014, Biochem Med (Zagreb)Table 4, row 'Glu'
serum, BD Vacutainer RST 5 mL (thrombin clot activator, gel)Storagecentrifuged-on-gel2-8 °C7 daysstable for the longest time testedmean % difference vs time 0 exceeds desirable imprecision based on biological variation (EFLM/Westgard), here 2.55%Mandić et al. 2020, Biochem Med (Zagreb)Tables 3 and 4, row 'Glucose, mmol/L', BD RST columns
serum, Vacuette Tube Serum Gel Separator 8 mL (Greiner, 455071)Storagecentrifuged-on-gelRT18 hthen changed, changed by 24 h% bias of median vs baseline exceeds desirable bias (Westgard/Ricos BV), here 2.2%Cuhadar et al. 2012, Biochem Med (Zagreb)Table 3, row 'Glucose', gel 24°C
serum, Vacuette Tube Serum No Gel 9 mL (Greiner, 455092), plainStoragecentrifuged-on-cellsRT0 hchanged at the first time point, changed by 6 h% bias of median vs baseline exceeds desirable bias (Westgard/Ricos BV), here 2.2%Cuhadar et al. 2012, Biochem Med (Zagreb)Table 3, row 'Glucose', w/o gel 24°C
serum, BD Vacutainer SST (SKU 367986), gelStoragecentrifuged-on-gel30 °C (±1 °C)72 hstable for the longest time testedmean bias vs paired baseline within CLIA acceptance limits AND >2/3 of individual results within limitsIacovetti et al. 2026, Pract Lab MedResults, 'Analyte stability after storage at 30 °C', paragraph 1 (Table 1 image-only)
serum, BD SST Microtainer (SKU 395967), gel; filled from venous bloodStoragecentrifuged-on-gel0 °C72 hstable for the longest time testedmean bias vs paired baseline within CLIA acceptance limits AND >2/3 of individual results within limitsIacovetti et al. 2026, Pract Lab MedResults, 'Analyte stability after separation and storage in microvolume tubes at 0 °C' (Table 2 image-only)
whole-blood-clotted, BD Vacutainer SST (SKU 367986), gelBefore spinninguncentrifuged30 °C (±1 °C)0 hchanged at the first time point, changed by 24 hmean bias vs paired baseline within CLIA acceptance limits AND >2/3 of individual results within limitsIacovetti et al. 2026, Pract Lab MedResults, 'Analyte stability after storage at 30 °C', paragraph 2 (Table 1 image-only)
whole-blood-fluoride, Greiner Vacuette FC-Mix NaF-EDTA-citrate (granulated), blood aliquoted into micro-tubesBefore spinninguncentrifugedRT48 hthen changed, changed by 72 hmean bias vs reference glucose within desirable bias (< ±1.95%, biological variation)Bonetti et al. 2017, Biochem Med (Zagreb)Table 1
whole-blood-fluoride, BD Vacutainer fluorideBefore spinninguncentrifugedRT24 hstable for the longest time testedmean % deviation from baseline within total change limit TCL = sqrt((2.77 CVa)^2 + (0.5 CVb)^2), CVb from Ricos (Table 2)Abraham et al. 2019, Biochem Med (Zagreb)Discussion, paragraph 1 (Results paragraphs 1-3; Figure 3)
whole-blood-fluoride, BD Vacutainer fluorideBefore spinninguncentrifuged22-30 °C (logger 24-26 °C)24 hstable for the longest time testedmean % deviation from baseline within total change limit TCL = sqrt((2.77 CVa)^2 + (0.5 CVb)^2), CVb from Ricos (Table 2)Abraham et al. 2019, Biochem Med (Zagreb)Discussion, paragraph 1 (Results paragraphs 1-3; Figure 3)
whole-blood-fluoride, BD Vacutainer fluorideBefore spinninguncentrifuged> 30 °C (logger 38-40 °C)24 hstable for the longest time testedmean % deviation from baseline within total change limit TCL = sqrt((2.77 CVa)^2 + (0.5 CVb)^2), CVb from Ricos (Table 2)Abraham et al. 2019, Biochem Med (Zagreb)Discussion, paragraph 1 (Results paragraphs 1-3; Figure 3)
Haptoglobin2 findings
Specimen, tubeConditionTemperatureStable forCriterionStudy
whole-blood-heparin, BD Vacutainer lithium heparin, no gel (ref 368884)Before spinninguncentrifugedRT10 hstable for the longest time tested|mean % deviation 10 h vs baseline| > laboratory bias goal (Table II; mostly biological-variation based, some empirically modified)Henriksen et al. 2014, Scand J Clin Lab InvestTable II, row 'HAPT'
whole-blood-heparin, BD Vacutainer lithium heparin, no gel (ref 368884)Before spinninguncentrifugedRT10 hstable for the longest time testedCV of paired baseline/10 h results > goal CV (Table II)Henriksen et al. 2014, Scand J Clin Lab InvestTable II, row 'HAPT'
HDL Cholesterol22 findings
Specimen, tubeConditionTemperatureStable forCriterionStudy
dbs, Whatman 903 card, mailed in foil pouch with desiccantStoragedriednot stated; ambient postal shipping in the US (Apr-Dec 2016)3 daysthen changed, changed by 4 daysregression coefficient for days before freezing vs 0-2 days significant (P < .05), Model 1 (DBS value, unadjusted)Crimmins et al. 2020, Am J Hum BiolTable 3, rows 'Days before freezing', column 'HDL cholesterol'
dbs, Whatman 903 card, mailed in foil pouch with desiccantStoragedriednot stated; ambient postal shipping in the US (Apr-Dec 2016)7 daysthen changed, changed by 8 daysregression coefficient for days before freezing vs 0-2 days significant (P < .05), Model 2 (DBS value adjusted for venous value)Crimmins et al. 2020, Am J Hum BiolTable 3, rows 'Days before freezing', column 'HDL cholesterol'
serum, BD Vacutainer 8.5 mL plastic serum tube; pooled serum aliquoted into 1.5 mL Eppendorf tubesFreeze–thawseparated−20 °C10 cyclesstable for the longest time testedunstable only if both P < 0.005 (Bonferroni) and % change from T0 > desirable bias (Westgard/Ricos), here 5.2%Cuhadar et al. 2013, Biochem Med (Zagreb)Table 2, row 'HDL (mmol/L)', columns T1d-T10d
serum, BD Vacutainer SSTFreeze–thawseparated−70 °C, 1 month; thawed at room temperature1 cyclestable for the longest time testedmean % change (bias) from fresh baseline within total allowable error (TEa, desirable bias)Abraham et al. 2021, J Lab PhysiciansTable 2 and Table 3, row 'HDL (mg/dL)'
serum, serum (tube not stated)Storageseparated−20 °C2 monthsthen changed, changed by 6 monthst-test (two-sample, equal variances) vs t=0, 95% CIBeekhof et al. 2012, Lipids Health DisResults, paragraph 4
serum, BD Vacutainer 8.5 mL plastic serum tube; pooled serum aliquoted into 1.5 mL Eppendorf tubesStorageseparated−20 °C3 monthsstable for the longest time testedunstable only if both P < 0.013 (Bonferroni) and % change from T0 > desirable bias (Westgard/Ricos), here 5.2%Cuhadar et al. 2013, Biochem Med (Zagreb)Table 2, row 'HDL (mmol/L)', columns T1m-T3m
serum, Vacuette Tube Serum Gel Separator 8 mL (Greiner, 455071)Storagecentrifuged-on-gel2-8 °C24 hthen changed, changed by 30 h% bias of median vs baseline exceeds desirable bias (Westgard/Ricos BV), here 5.2%Cuhadar et al. 2012, Biochem Med (Zagreb)Table 3, row 'HDL-cholesterol', gel 4°C
serum, Vacuette Tube Serum No Gel 9 mL (Greiner, 455092), plainStoragecentrifuged-on-cells2-8 °C36 hthen changed, changed by 48 h% bias of median vs baseline exceeds desirable bias (Westgard/Ricos BV), here 5.2%Cuhadar et al. 2012, Biochem Med (Zagreb)Table 3, row 'HDL-cholesterol', w/o gel 4°C
serum, Vacuette Tube Serum Gel Separator 8 mL (Greiner, 455071)Storagecentrifuged-on-gelRT36 hthen changed, changed by 48 h% bias of median vs baseline exceeds desirable bias (Westgard/Ricos BV), here 5.2%Cuhadar et al. 2012, Biochem Med (Zagreb)Table 3, row 'HDL-cholesterol', gel 24°C
serum, Vacuette Tube Serum No Gel 9 mL (Greiner, 455092), plainStoragecentrifuged-on-cellsRT6 hthen changed, changed by 12 h% bias of median vs baseline exceeds desirable bias (Westgard/Ricos BV), here 5.2%Cuhadar et al. 2012, Biochem Med (Zagreb)Table 3, row 'HDL-cholesterol', w/o gel 24°C
serum, BD Vacutainer SST (SKU 367986), gelStoragecentrifuged-on-gel30 °C (±1 °C)72 hstable for the longest time testedmean bias vs paired baseline within CLIA acceptance limits AND >2/3 of individual results within limitsIacovetti et al. 2026, Pract Lab MedResults, 'Analyte stability after storage at 30 °C', paragraph 1 (Table 1 image-only)
serum, BD SST Microtainer (SKU 395967), gel; filled from venous bloodStoragecentrifuged-on-gel0 °C72 hstable for the longest time testedmean bias vs paired baseline within CLIA acceptance limits AND >2/3 of individual results within limitsIacovetti et al. 2026, Pract Lab MedResults, 'Analyte stability after separation and storage in microvolume tubes at 0 °C' (Table 2 image-only)
serum, serum (tube not stated)Storageseparated−70°C12 monthsstable for the longest time testedt-test (two-sample, equal variances) vs t=0, 95% CIBeekhof et al. 2012, Lipids Health DisResults, paragraph 4
serum, serum (tube not stated)Storageseparated−196°C12 monthsstable for the longest time testedt-test (two-sample, equal variances) vs t=0, 95% CIBeekhof et al. 2012, Lipids Health DisResults, paragraph 4
whole-blood-clotted, BD Vacutainer SSTBefore spinninguncentrifugedRT24 hstable for the longest time testedmean % deviation from baseline within total change limit TCL = sqrt((2.77 CVa)^2 + (0.5 CVb)^2), CVb from Ricos (Table 2)Abraham et al. 2019, Biochem Med (Zagreb)Discussion, paragraph 4
whole-blood-clotted, BD Vacutainer trace element-free serum tubeBefore spinninguncentrifugedcold packs in insulated box; 9 (2) °C on arrival24 hstable for the longest time tested% difference within desirable quality specification (from intra- and interindividual variation)Jones et al. 2024, J NutrTable 2, row 'HDL cholesterol, mmol/L'; Table 1
whole-blood-clotted, BD Vacutainer SST (SKU 367986), gelBefore spinninguncentrifuged30 °C (±1 °C)72 hstable for the longest time testedmean bias vs paired baseline within CLIA acceptance limits AND >2/3 of individual results within limitsIacovetti et al. 2026, Pract Lab MedResults, 'Analyte stability after storage at 30 °C', paragraph 2 (Table 1 image-only)
whole-blood-clotted, Sarstedt Monovette serumBefore spinninguncentrifugedchilled in transit (insulated box, 2 frozen cool packs; temperature not recorded)24 hstable for the longest time tested% difference within desirable quality specification for bias (Fraser, from intra- and interindividual variation)Jones et al. 2021, J NutrTable 2, row 'HDL cholesterol', column Serum
whole-blood-edta, Sarstedt Monovette EDTABefore spinninguncentrifugedchilled in transit (insulated box, 2 frozen cool packs; temperature not recorded)24 hstable for the longest time tested% difference within desirable quality specification for bias (Fraser, from intra- and interindividual variation)Jones et al. 2021, J NutrTable 2, row 'HDL cholesterol', column EDTA
whole-blood-heparin, BD Vacutainer lithium heparin, no gel (ref 368884)Before spinninguncentrifugedRT10 hstable for the longest time tested|mean % deviation 10 h vs baseline| > laboratory bias goal (Table II; mostly biological-variation based, some empirically modified)Henriksen et al. 2014, Scand J Clin Lab InvestTable II, row 'HDL-C'
whole-blood-heparin, BD Vacutainer lithium heparin, no gel (ref 368884)Before spinninguncentrifugedRT10 hstable for the longest time testedCV of paired baseline/10 h results > goal CV (Table II)Henriksen et al. 2014, Scand J Clin Lab InvestTable II, row 'HDL-C'
whole-blood-heparin, Sarstedt Monovette lithium heparin, trace-element freeBefore spinninguncentrifugedchilled in transit (insulated box, 2 frozen cool packs; temperature not recorded)24 hstable for the longest time tested% difference within desirable quality specification for bias (Fraser, from intra- and interindividual variation)Jones et al. 2021, J NutrTable 2, row 'HDL cholesterol', column LH
Hematocrit24 findings
Specimen, tubeConditionTemperatureStable forCriterionStudy
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C48 hstable for the longest time testedHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'HT', XN-9000, 4 °C
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C24 hthen changed, changed by 36 hBias% vs optimal bias (OP-Bias%, from biological variation; Table 1)Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'HT', XN-9000, 4 °C
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C48 hstable for the longest time testedBias% vs critical difference (CD%, reference change value; Table 1)Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'HT', XN-9000, 4 °C
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C48 hstable for the longest time testedHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'HT', BC-6800, 4 °C
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C24 hthen changed, changed by 36 hBias% vs optimal bias (OP-Bias%, from biological variation; Table 1)Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'HT', BC-6800, 4 °C
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C48 hstable for the longest time testedBias% vs critical difference (CD%, reference change value; Table 1)Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'HT', BC-6800, 4 °C
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C8 hstable for the longest time testedHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'HT', XN-9000, 4 °C
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C8 hstable for the longest time testedBias% vs optimal bias (OP-Bias%, from biological variation; Table 1)Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'HT', XN-9000, 4 °C
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C8 hstable for the longest time testedHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'HT', BC-6800, 4 °C
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C8 hstable for the longest time testedBias% vs optimal bias (OP-Bias%, from biological variation; Table 1)Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'HT', BC-6800, 4 °C
whole-blood-edta, K2-EDTA (BD Vacutainer)Storageuncentrifuged2-8 °C48 hstable for the longest time testedRepeated-measures one-way ANOVA with Tukey post hoc vs 0 h, p < 0.05 (* in table)Unalli et al. 2021, J Med BiochemResults paragraphs 1-2 and Discussion paragraph 1 (Figures 1-2; numbers in figures only)
whole-blood-edta, K3-EDTA (Becton Dickinson)StorageuncentrifugedRT36 hthen changed, changed by 48 hHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'HT', XN-9000, RT
whole-blood-edta, K3-EDTA (Becton Dickinson)StorageuncentrifugedRT8 hthen changed, changed by 24 hBias% vs optimal bias (OP-Bias%, from biological variation; Table 1)Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'HT', XN-9000, RT
whole-blood-edta, K3-EDTA (Becton Dickinson)StorageuncentrifugedRT24 hthen changed, changed by 36 hBias% vs critical difference (CD%, reference change value; Table 1)Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'HT', XN-9000, RT
whole-blood-edta, K3-EDTA (Becton Dickinson)StorageuncentrifugedRT36 hthen changed, changed by 48 hHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'HT', BC-6800, RT
whole-blood-edta, K3-EDTA (Becton Dickinson)StorageuncentrifugedRT8 hthen changed, changed by 24 hBias% vs optimal bias (OP-Bias%, from biological variation; Table 1)Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'HT', BC-6800, RT
whole-blood-edta, K3-EDTA (Becton Dickinson)StorageuncentrifugedRT24 hthen changed, changed by 36 hBias% vs critical difference (CD%, reference change value; Table 1)Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'HT', BC-6800, RT
whole-blood-edta, K3-EDTA (Becton Dickinson)StorageuncentrifugedRT8 hstable for the longest time testedHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'HT', XN-9000, RT
whole-blood-edta, K3-EDTA (Becton Dickinson)StorageuncentrifugedRT8 hstable for the longest time testedBias% vs optimal bias (OP-Bias%, from biological variation; Table 1)Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'HT', XN-9000, RT
whole-blood-edta, K3-EDTA (Becton Dickinson)StorageuncentrifugedRT8 hstable for the longest time testedHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'HT', BC-6800, RT
whole-blood-edta, K3-EDTA (Becton Dickinson)StorageuncentrifugedRT8 hstable for the longest time testedBias% vs optimal bias (OP-Bias%, from biological variation; Table 1)Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'HT', BC-6800, RT
whole-blood-edta, K2-EDTA (BD Vacutainer)StorageuncentrifugedRT24 hthen changed, changed by 48 hRepeated-measures one-way ANOVA with Tukey post hoc vs 0 h, p < 0.05 (* in table)Unalli et al. 2021, J Med BiochemResults, paragraph 2 (Figure 2B)
whole-blood-edta, K2EDTA (Becton Dickinson)Storageuncentrifuged33 °C0 hchanged at the first time point, changed by 4 hUnacceptable if median % difference > Biasmax (biological variation) and Holm-adjusted Wilcoxon P < 0.05; stable if equivalence shown (two one-tailed Wilcoxon tests vs Biasmax, Pholm < 0.05)Jain et al. 2018, Biochem Med (Zagreb)Results paragraph 1; Table 2, row 'HCT'
whole-blood-edta, K2-EDTA (BD Vacutainer)Storageuncentrifuged10 °C48 hstable for the longest time testedRepeated-measures one-way ANOVA with Tukey post hoc vs 0 h, p < 0.05 (* in table)Unalli et al. 2021, J Med BiochemResults paragraphs 1-2 and Discussion paragraph 1 (Figures 1-2; numbers in figures only)
Hemoglobin24 findings
Specimen, tubeConditionTemperatureStable forCriterionStudy
whole-blood-edta, BD Vacutainer K2EDTA, 2 mL (spray-coated)Storageuncentrifuged2-8 °C24 hstable for the longest time testedMean % change from baseline within analytical measurement uncertainty (MU ±1.2%)Al-Najdawi et al. 2025, Med Pharm RepTable II, row 'Hb'; Results
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C48 hstable for the longest time testedHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'Hb', XN-9000, 4 °C
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C0 hchanged at the first time point, changed by 2 hBias% vs optimal bias (OP-Bias%, from biological variation; Table 1)Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'Hb', XN-9000, 4 °C
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C48 hstable for the longest time testedBias% vs critical difference (CD%, reference change value; Table 1)Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'Hb', XN-9000, 4 °C
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C48 hstable for the longest time testedHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'Hb', BC-6800, 4 °C
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C48 hstable for the longest time testedBias% vs optimal bias (OP-Bias%, from biological variation; Table 1)Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'Hb', BC-6800, 4 °C
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C8 hstable for the longest time testedHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'Hb', XN-9000, 4 °C
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C8 hstable for the longest time testedBias% vs optimal bias (OP-Bias%, from biological variation; Table 1)Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'Hb', XN-9000, 4 °C
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C8 hstable for the longest time testedHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'Hb', BC-6800, 4 °C
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C8 hstable for the longest time testedBias% vs optimal bias (OP-Bias%, from biological variation; Table 1)Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'Hb', BC-6800, 4 °C
whole-blood-edta, K2-EDTA (BD Vacutainer)Storageuncentrifuged2-8 °C48 hstable for the longest time testedRepeated-measures one-way ANOVA with Tukey post hoc vs 0 h, p < 0.05 (* in table)Unalli et al. 2021, J Med BiochemResults paragraphs 1-2 and Discussion paragraph 1 (Figures 1-2; numbers in figures only)
whole-blood-edta, K3-EDTA (Becton Dickinson)StorageuncentrifugedRT48 hstable for the longest time testedHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'Hb', XN-9000, RT
whole-blood-edta, K3-EDTA (Becton Dickinson)StorageuncentrifugedRT48 hstable for the longest time testedBias% vs optimal bias (OP-Bias%, from biological variation; Table 1)Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'Hb', XN-9000, RT
whole-blood-edta, K3-EDTA (Becton Dickinson)StorageuncentrifugedRT48 hstable for the longest time testedHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'Hb', BC-6800, RT
whole-blood-edta, K3-EDTA (Becton Dickinson)StorageuncentrifugedRT48 hstable for the longest time testedBias% vs optimal bias (OP-Bias%, from biological variation; Table 1)Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'Hb', BC-6800, RT
whole-blood-edta, K3-EDTA (Becton Dickinson)StorageuncentrifugedRT8 hstable for the longest time testedHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'Hb', XN-9000, RT
whole-blood-edta, K3-EDTA (Becton Dickinson)StorageuncentrifugedRT8 hstable for the longest time testedBias% vs optimal bias (OP-Bias%, from biological variation; Table 1)Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'Hb', XN-9000, RT
whole-blood-edta, K3-EDTA (Becton Dickinson)StorageuncentrifugedRT8 hstable for the longest time testedHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'Hb', BC-6800, RT
whole-blood-edta, K3-EDTA (Becton Dickinson)StorageuncentrifugedRT8 hstable for the longest time testedBias% vs optimal bias (OP-Bias%, from biological variation; Table 1)Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'Hb', BC-6800, RT
whole-blood-edta, K2-EDTA (BD Vacutainer)StorageuncentrifugedRT48 hstable for the longest time testedRepeated-measures one-way ANOVA with Tukey post hoc vs 0 h, p < 0.05 (* in table)Unalli et al. 2021, J Med BiochemResults paragraphs 1-2 and Discussion paragraph 1 (Figures 1-2; numbers in figures only)
whole-blood-edta, K2EDTA (Becton Dickinson)Storageuncentrifuged33 °C24 hthen changed, changed by 48 hUnacceptable if median % difference > Biasmax (biological variation) and Holm-adjusted Wilcoxon P < 0.05; stable if equivalence shown (two one-tailed Wilcoxon tests vs Biasmax, Pholm < 0.05)Jain et al. 2018, Biochem Med (Zagreb)Results, last paragraph (equivalence); Table 2, row 'HGB'
whole-blood-edta, BD Vacutainer EDTAStorageuncentrifuged22-30 °C (logger 24-26 °C)6 hthen changed, changed by 12 hmean % deviation from baseline within total change limit TCL = sqrt((2.77 CVa)^2 + (0.5 CVb)^2), CVb from Ricos (Table 2)Abraham et al. 2019, Biochem Med (Zagreb)Results, paragraph 2 (Figure 3g)
whole-blood-edta, BD Vacutainer EDTAStorageuncentrifuged> 30 °C (logger 38-40 °C)0 hchanged at the first time point, changed by 6 hmean % deviation from baseline within total change limit TCL = sqrt((2.77 CVa)^2 + (0.5 CVb)^2), CVb from Ricos (Table 2)Abraham et al. 2019, Biochem Med (Zagreb)Results, paragraph 2 (Figure 3g)
whole-blood-edta, K2-EDTA (BD Vacutainer)Storageuncentrifuged10 °C48 hstable for the longest time testedRepeated-measures one-way ANOVA with Tukey post hoc vs 0 h, p < 0.05 (* in table)Unalli et al. 2021, J Med BiochemResults paragraphs 1-2 and Discussion paragraph 1 (Figures 1-2; numbers in figures only)
Hemoglobin A1c7 findings
Specimen, tubeConditionTemperatureStable forCriterionStudy
dbs, Whatman 903 card, mailed in foil pouch with desiccantStoragedriednot stated; ambient postal shipping in the US (Apr-Dec 2016)0 dayschanged at the first time point, changed by 3 daysregression coefficient for days before freezing vs 0-2 days significant (P < .05), Model 1 (DBS value, unadjusted)Crimmins et al. 2020, Am J Hum BiolTable 3, rows 'Days before freezing', column 'HbA1c'
dbs, Whatman 903 card, mailed in foil pouch with desiccantStoragedriednot stated; ambient postal shipping in the US (Apr-Dec 2016)0 dayschanged at the first time point, changed by 3 daysregression coefficient for days before freezing vs 0-2 days significant (P < .05), Model 2 (DBS value adjusted for venous value)Crimmins et al. 2020, Am J Hum BiolTable 3, rows 'Days before freezing', column 'HbA1c'
dbs, HemaSpot device (Spot on Sciences), integrated desiccant; fingerprick bloodStoragedriedposted by participants (ambient), then stored at 4°C in the laboratory4 daysstable for the longest time testedtime between sampling and testing does not significantly change the DBS-venous calibration (nested model F-test, 5% level)Hall et al. 2020, Diabet MedTable 2; Discussion paragraph 1
other, ACD (Becton Dickinson) whole blood + RPMI 1640 + DMSO in cryotube; baseline fresh from K2-EDTA (Becton Dickinson)Storageuncentrifuged−196°C7 yearsstable for the longest time testeddifference of 0.50% HbA1c considered clinically significant; observed mean bias −0.04%Abdullah et al. 2023, J Clin Lab AnalDiscussion, paragraph 1; Table 1
whole-blood-edta, BD Vacutainer EDTAStorageuncentrifugedRT24 hstable for the longest time testedmean % deviation from baseline within total change limit TCL = sqrt((2.77 CVa)^2 + (0.5 CVb)^2), CVb from Ricos (Table 2)Abraham et al. 2019, Biochem Med (Zagreb)Discussion, paragraph 1 (Results paragraphs 1-3; Figure 3)
whole-blood-edta, BD Vacutainer EDTAStorageuncentrifuged22-30 °C (logger 24-26 °C)24 hstable for the longest time testedmean % deviation from baseline within total change limit TCL = sqrt((2.77 CVa)^2 + (0.5 CVb)^2), CVb from Ricos (Table 2)Abraham et al. 2019, Biochem Med (Zagreb)Discussion, paragraph 1 (Results paragraphs 1-3; Figure 3)
whole-blood-edta, BD Vacutainer EDTAStorageuncentrifuged> 30 °C (logger 38-40 °C)24 hstable for the longest time testedmean % deviation from baseline within total change limit TCL = sqrt((2.77 CVa)^2 + (0.5 CVb)^2), CVb from Ricos (Table 2)Abraham et al. 2019, Biochem Med (Zagreb)Discussion, paragraph 1 (Results paragraphs 1-3; Figure 3)
High-fluorescence cells (HFC)8 findings
Specimen, tubeConditionTemperatureStable forCriterionStudy
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C48 hstable for the longest time testedHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 4, row 'HFC', XN-9000, 4 °C
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C48 hstable for the longest time testedHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 4, row 'HFC', BC-6800, 4 °C
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C8 hstable for the longest time testedHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 5, row 'HFC', XN-9000, 4 °C
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C8 hstable for the longest time testedHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 5, row 'HFC', BC-6800, 4 °C
whole-blood-edta, K3-EDTA (Becton Dickinson)StorageuncentrifugedRT48 hstable for the longest time testedHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 4, row 'HFC', XN-9000, RT
whole-blood-edta, K3-EDTA (Becton Dickinson)StorageuncentrifugedRT48 hstable for the longest time testedHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 4, row 'HFC', BC-6800, RT
whole-blood-edta, K3-EDTA (Becton Dickinson)StorageuncentrifugedRT8 hstable for the longest time testedHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 5, row 'HFC', XN-9000, RT
whole-blood-edta, K3-EDTA (Becton Dickinson)StorageuncentrifugedRT8 hstable for the longest time testedHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 5, row 'HFC', BC-6800, RT
High-fluorescence reticulocyte ratio (HFR)8 findings
Specimen, tubeConditionTemperatureStable forCriterionStudy
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C48 hstable for the longest time testedHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'HFR', XN-9000, 4 °C
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C48 hstable for the longest time testedHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'HFR', BC-6800, 4 °C
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C8 hstable for the longest time testedHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'HFR', XN-9000, 4 °C
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C8 hstable for the longest time testedHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'HFR', BC-6800, 4 °C
whole-blood-edta, K3-EDTA (Becton Dickinson)StorageuncentrifugedRT48 hstable for the longest time testedHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'HFR', XN-9000, RT
whole-blood-edta, K3-EDTA (Becton Dickinson)StorageuncentrifugedRT48 hstable for the longest time testedHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'HFR', BC-6800, RT
whole-blood-edta, K3-EDTA (Becton Dickinson)StorageuncentrifugedRT8 hstable for the longest time testedHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'HFR', XN-9000, RT
whole-blood-edta, K3-EDTA (Becton Dickinson)StorageuncentrifugedRT8 hstable for the longest time testedHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'HFR', BC-6800, RT
High-Sensitivity C-Reactive Protein3 findings
Specimen, tubeConditionTemperatureStable forCriterionStudy
serum, BD Vacutainer SST (SKU 367986), gelStoragecentrifuged-on-gel30 °C (±1 °C)72 hstable for the longest time testedmean bias vs paired baseline within CLIA acceptance limits AND >2/3 of individual results within limitsIacovetti et al. 2026, Pract Lab MedResults, 'Analyte stability after storage at 30 °C', paragraph 1 (Table 1 image-only)
serum, BD SST Microtainer (SKU 395967), gel; filled from venous bloodStoragecentrifuged-on-gel0 °C72 hstable for the longest time testedmean bias vs paired baseline within CLIA acceptance limits AND >2/3 of individual results within limitsIacovetti et al. 2026, Pract Lab MedResults, 'Analyte stability after separation and storage in microvolume tubes at 0 °C' (Table 2 image-only)
whole-blood-clotted, BD Vacutainer SST (SKU 367986), gelBefore spinninguncentrifuged30 °C (±1 °C)72 hstable for the longest time testedmean bias vs paired baseline within CLIA acceptance limits AND >2/3 of individual results within limitsIacovetti et al. 2026, Pract Lab MedResults, 'Analyte stability after storage at 30 °C', paragraph 2 (Table 1 image-only)
Holotranscobalamin (Active B12)4 findings
Specimen, tubeConditionTemperatureStable forCriterionStudy
whole-blood-clotted, BD Vacutainer trace element-free serum tubeBefore spinninguncentrifugedcold packs in insulated box; 9 (2) °C on arrival24 hstable for the longest time tested% difference within desirable quality specification (from intra- and interindividual variation)Jones et al. 2024, J NutrTable 2, row 'Holotranscobalamin, pmol/L'; Table 1
whole-blood-clotted, Sarstedt Monovette serumBefore spinninguncentrifugedchilled in transit (insulated box, 2 frozen cool packs; temperature not recorded)24 hstable for the longest time tested% difference within desirable quality specification for bias (Fraser, from intra- and interindividual variation)Jones et al. 2021, J NutrTable 2, row 'Holotranscobalamin', column Serum
whole-blood-edta, Sarstedt Monovette EDTABefore spinninguncentrifugedchilled in transit (insulated box, 2 frozen cool packs; temperature not recorded)24 hstable for the longest time tested% difference within desirable quality specification for bias (Fraser, from intra- and interindividual variation)Jones et al. 2021, J NutrTable 2, row 'Holotranscobalamin', column EDTA
whole-blood-heparin, Sarstedt Monovette lithium heparin, trace-element freeBefore spinninguncentrifugedchilled in transit (insulated box, 2 frozen cool packs; temperature not recorded)24 hstable for the longest time tested% difference within desirable quality specification for bias (Fraser, from intra- and interindividual variation)Jones et al. 2021, J NutrTable 2, row 'Holotranscobalamin', column LH
Homocysteine3 findings
Specimen, tubeConditionTemperatureStable forCriterionStudy
plasma-li-heparin, BD Vacutainer Barricor LH 3 mL, mechanical separatorStoragecentrifuged-on-gel2-8 °C80 hstable for the longest time testedbias vs aliquoted plasma reference tube > maximal allowable bias 7.7% (Ricos et al.); time to MAB from linear regression of bias on time to analysisGawria et al. 2020, J Clin Lab AnalTable 2, row 'Homocysteine', Barricor BD
plasma-li-heparin, BD Vacutainer PST LH 3 mL, gelStoragecentrifuged-on-gel2-8 °C80 hstable for the longest time testedbias vs aliquoted plasma reference tube > maximal allowable bias 7.7% (Ricos et al.); time to MAB from linear regression of bias on time to analysisGawria et al. 2020, J Clin Lab AnalTable 2, row 'Homocysteine', Gel BD
plasma-li-heparin, Greiner Vacuette LH 3 mL Sep, gelStoragecentrifuged-on-gel2-8 °C80 hstable for the longest time testedbias vs aliquoted plasma reference tube > maximal allowable bias 7.7% (Ricos et al.); time to MAB from linear regression of bias on time to analysisGawria et al. 2020, J Clin Lab AnalTable 2, row 'Homocysteine', Gel Greiner
Immature Reticulocyte Fraction8 findings
Specimen, tubeConditionTemperatureStable forCriterionStudy
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C48 hstable for the longest time testedHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'IRF', XN-9000, 4 °C
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C48 hstable for the longest time testedHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'IRF', BC-6800, 4 °C
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C8 hstable for the longest time testedHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'IRF', XN-9000, 4 °C
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C8 hstable for the longest time testedHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'IRF', BC-6800, 4 °C
whole-blood-edta, K3-EDTA (Becton Dickinson)StorageuncentrifugedRT8 hthen changed, changed by 24 hHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'IRF', XN-9000, RT
whole-blood-edta, K3-EDTA (Becton Dickinson)StorageuncentrifugedRT24 hthen changed, changed by 36 hHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'IRF', BC-6800, RT
whole-blood-edta, K3-EDTA (Becton Dickinson)StorageuncentrifugedRT8 hstable for the longest time testedHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'IRF', XN-9000, RT
whole-blood-edta, K3-EDTA (Becton Dickinson)StorageuncentrifugedRT8 hstable for the longest time testedHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'IRF', BC-6800, RT
Immunoglobulin A (IgA)2 findings
Specimen, tubeConditionTemperatureStable forCriterionStudy
whole-blood-heparin, BD Vacutainer lithium heparin, no gel (ref 368884)Before spinninguncentrifugedRT10 hstable for the longest time tested|mean % deviation 10 h vs baseline| > laboratory bias goal (Table II; mostly biological-variation based, some empirically modified)Henriksen et al. 2014, Scand J Clin Lab InvestTable II, row 'IgA'
whole-blood-heparin, BD Vacutainer lithium heparin, no gel (ref 368884)Before spinninguncentrifugedRT10 hstable for the longest time testedCV of paired baseline/10 h results > goal CV (Table II)Henriksen et al. 2014, Scand J Clin Lab InvestTable II, row 'IgA'
Immunoglobulin G (IgG)2 findings
Specimen, tubeConditionTemperatureStable forCriterionStudy
whole-blood-heparin, BD Vacutainer lithium heparin, no gel (ref 368884)Before spinninguncentrifugedRT10 hstable for the longest time tested|mean % deviation 10 h vs baseline| > laboratory bias goal (Table II; mostly biological-variation based, some empirically modified)Henriksen et al. 2014, Scand J Clin Lab InvestTable II, row 'IgG'
whole-blood-heparin, BD Vacutainer lithium heparin, no gel (ref 368884)Before spinninguncentrifugedRT10 hstable for the longest time testedCV of paired baseline/10 h results > goal CV (Table II)Henriksen et al. 2014, Scand J Clin Lab InvestTable II, row 'IgG'
Immunoglobulin M (IgM)2 findings
Specimen, tubeConditionTemperatureStable forCriterionStudy
whole-blood-heparin, BD Vacutainer lithium heparin, no gel (ref 368884)Before spinninguncentrifugedRT10 hstable for the longest time tested|mean % deviation 10 h vs baseline| > laboratory bias goal (Table II; mostly biological-variation based, some empirically modified)Henriksen et al. 2014, Scand J Clin Lab InvestTable II, row 'IgM'
whole-blood-heparin, BD Vacutainer lithium heparin, no gel (ref 368884)Before spinninguncentrifugedRT10 hstable for the longest time testedCV of paired baseline/10 h results > goal CV (Table II)Henriksen et al. 2014, Scand J Clin Lab InvestTable II, row 'IgM'
Insulin8 findings
Specimen, tubeConditionTemperatureStable forCriterionStudy
plasma-edta, Sarstedt K+-EDTA 2.7 mLFreeze–thawnot-stated−80 °C5 cyclesstable for the longest time testedmean % change from baseline > 10% considered clinically significantMcDonald et al. 2012, PLoS OneTable 1
plasma-edta, Sarstedt K+-EDTA 2.7 mLStoragecentrifuged-on-cellsRT24 hstable for the longest time testedmean % change from baseline > 10% considered clinically significantMcDonald et al. 2012, PLoS OneResults 1.1, paragraph 2
serum, Sarstedt serum gel 2.7 mLFreeze–thawnot-stated−80 °C5 cyclesstable for the longest time testedmean % change from baseline > 10% considered clinically significantMcDonald et al. 2012, PLoS OneTable 1
whole-blood-clotted, Sarstedt serum gel 2.7 mLBefore spinninguncentrifuged2-8 °C120 hstable for the longest time testedmean % change from baseline > 10% considered clinically significantMcDonald et al. 2012, PLoS OneResults 1.2
whole-blood-edta, Sarstedt K+-EDTA 2.7 mLBefore spinninguncentrifuged2-8 °C24 hstable for the longest time testedmean % change from baseline > 10% considered clinically significantMcDonald et al. 2012, PLoS OneResults 1.1, paragraph 3
whole-blood-edta, Sarstedt K+-EDTA 2.7 mLBefore spinninguncentrifuged2-8 °C120 hstable for the longest time testedmean % change from baseline > 10% considered clinically significantMcDonald et al. 2012, PLoS OneResults 1.2
whole-blood-edta, Sarstedt K+-EDTA 2.7 mLBefore spinninguncentrifugedRT24 hstable for the longest time testedmean % change from baseline > 10% considered clinically significantMcDonald et al. 2012, PLoS OneResults 1.1, paragraph 2
whole-blood-edta, Sarstedt K+-EDTA 2.7 mLBefore spinninguncentrifugedRT120 hstable for the longest time testedmean % change from baseline > 10% considered clinically significantMcDonald et al. 2012, PLoS OneResults 1.2
International Normalized Ratio7 findings
Specimen, tubeConditionTemperatureStable forCriterionStudy
plasma-citrate, BD Vacutainer 2.7 mL, 3.2% buffered sodium citrate; PPP aliquoted into microcentrifuge tubesStorageseparated−20 °C36 hstable for the longest time testedchange of more than ±10% of mean from baselinePatil et al. 2022, CureusTable 2
plasma-citrate, BD 0.109 M sodium citrate (9:1); plasma aliquoted into nonactivating Eppendorf tubesStorageseparated2-8 °C24 hstable for the longest time testedmean % change from baseline >10%, or >25% of samples with individual change >10% (van Geest-Daalderop)Feng et al. 2014, Sci RepTable 2, 'Acceptable Time (h)'
plasma-citrate, BD Vacutainer 2.7 mL, 3.2% buffered sodium citrate; PPP aliquoted into microcentrifuge tubesStorageseparated2-8 °C24 hthen changed, changed by 36 hchange of more than ±10% of mean from baselinePatil et al. 2022, CureusTable 2
plasma-citrate, BD 0.109 M sodium citrate (9:1); plasma aliquoted into nonactivating Eppendorf tubesStorageseparatedRT24 hstable for the longest time testedmean % change from baseline >10%, or >25% of samples with individual change >10% (van Geest-Daalderop)Feng et al. 2014, Sci RepTable 2, 'Acceptable Time (h)'
plasma-citrate, BD Vacutainer 2.7 mL, 3.2% buffered sodium citrate; PPP aliquoted into microcentrifuge tubesStorageseparatedRT0 hchanged at the first time point, changed by 12 hchange of more than ±10% of mean from baselinePatil et al. 2022, CureusTable 2
whole-blood-citrate, Greiner Vacuette sodium citrate, standard fill (blood–air interface)Before spinninguncentrifugedRT0 hchanged at the first time point, changed by 2 hmaximum acceptable bias = 0.375 × total biological variation (EFLM BV database, Model 2)Storm et al. 2026, Adv HematolTable 2, 'Standard tubes' row 'INR'
whole-blood-citrate, Greiner Vacuette sodium citrate, completely filled (no air)Before spinninguncentrifugedRT0 hchanged at the first time point, changed by 2 hmaximum acceptable bias = 0.375 × total biological variation (EFLM BV database, Model 2)Storm et al. 2026, Adv HematolTable 2, 'Full tubes' row 'INR'
Iron2 findings
Specimen, tubeConditionTemperatureStable forCriterionStudy
whole-blood-heparin, BD Vacutainer lithium heparin, no gel (ref 368884)Before spinninguncentrifugedRT10 hstable for the longest time tested|mean % deviation 10 h vs baseline| > laboratory bias goal (Table II; mostly biological-variation based, some empirically modified)Henriksen et al. 2014, Scand J Clin Lab InvestTable II, row 'Fe'
whole-blood-heparin, BD Vacutainer lithium heparin, no gel (ref 368884)Before spinninguncentrifugedRT10 hstable for the longest time testedCV of paired baseline/10 h results > goal CV (Table II)Henriksen et al. 2014, Scand J Clin Lab InvestTable II, row 'Fe'
Lactate1 finding
Specimen, tubeConditionTemperatureStable forCriterionStudy
plasma-li-heparin, BD Vacutainer LH PST II gel separator, 3.0 mL (REF 367374)Storagecentrifuged-on-gel2-8 °C24 hstable for the longest time testedmean % difference smaller than TEa (Ricos: 0.25 x (CVi^2 + CVg^2)^0.5 + 1.65 x 0.5 x CVi)Nonkes et al. 2019, Biochem Med (Zagreb)Table 1, row 'Lactate, mmol/L'
Lactate Dehydrogenase (LDH)20 findings
Specimen, tubeConditionTemperatureStable forCriterionStudy
plasma-li-heparin, Sarstedt lithium heparin, no gel separator (primary tube)Storagecentrifuged-on-cells2-8 °C2.9 hthen changed, changed by 6 hmean % difference vs time 0 exceeds total limit of change, TLC = ±√((1.65·CVa)² + (0.5·CVb)²) (EFLM BV), here 11.5%; stability limit from instability-equation regressionIllana et al. 2024, Biochem Med (Zagreb)Table 1, row 'LD'
plasma-li-heparin, Sarstedt lithium heparin, no gel separator; plasma aliquotStorageseparated2-8 °C63 hthen changed, changed by 72 hmean % difference vs time 0 exceeds total limit of change, TLC = ±√((1.65·CVa)² + (0.5·CVb)²) (EFLM BV), here 11.5%; stability limit from instability-equation regressionIllana et al. 2024, Biochem Med (Zagreb)Table 2, row 'LD'
plasma-li-heparin, BD Vacutainer lithium heparin 6 mL (regular)Storagenot-stated2-8 °C4 hthen changed, changed by 24 hmean % difference vs T0 exceeds total change limit (TCL), here ±8.82%Dupuy et al. 2018, Biochem Med (Zagreb)Table 2, BD lithium heparin tubes, row 'LD, U/L'
plasma-li-heparin, BD Barricor lithium heparin 5.5 mL (mechanical separator)Storagecentrifuged-on-gel2-8 °C72 hstable for the longest time testedmean % difference vs T0 exceeds total change limit (TCL), here ±8.82%Dupuy et al. 2018, Biochem Med (Zagreb)Table 2, BD Barricor lithium heparin tubes, row 'LD, U/L'
plasma-li-heparin, BD Vacutainer LH PST II gel separator, 3.0 mL (REF 367374)Storagecentrifuged-on-gel2-8 °C24 hstable for the longest time testedmean % difference smaller than TEa (Ricos: 0.25 x (CVi^2 + CVg^2)^0.5 + 1.65 x 0.5 x CVi)Nonkes et al. 2019, Biochem Med (Zagreb)Table 1, row 'LD, U/L'
plasma-li-heparin, BD Vacutainer Barricor LH 3 mL, mechanical separatorStoragecentrifuged-on-gel2-8 °C0 hchanged at the first time point, changed by 3 hbias vs aliquoted plasma reference tube > maximal allowable bias 4.3% (Ricos et al.); time to MAB from linear regression of bias on time to analysisGawria et al. 2020, J Clin Lab AnalTable 2, row 'LDH', Barricor BD
plasma-li-heparin, BD Vacutainer PST LH 3 mL, gelStoragecentrifuged-on-gel2-8 °C0 hchanged at the first time point, changed by 3 hbias vs aliquoted plasma reference tube > maximal allowable bias 4.3% (Ricos et al.); time to MAB from linear regression of bias on time to analysisGawria et al. 2020, J Clin Lab AnalTable 2, row 'LDH', Gel BD
plasma-li-heparin, Greiner Vacuette LH 3 mL Sep, gelStoragecentrifuged-on-gel2-8 °C0 hchanged at the first time point, changed by 3 hbias vs aliquoted plasma reference tube > maximal allowable bias 4.3% (Ricos et al.); time to MAB from linear regression of bias on time to analysisGawria et al. 2020, J Clin Lab AnalTable 2, row 'LDH', Gel Greiner
plasma-li-heparin, BD Barricor 4.5 mL (lithium heparin, mechanical separator)Storagecentrifuged-on-gel2-8 °C0 hchanged at the first time point, changed by 24 hmean % difference vs time 0 exceeds desirable imprecision based on biological variation (EFLM/Westgard), here 2.6%Mandić et al. 2020, Biochem Med (Zagreb)Tables 3 and 4, row 'LD, U/L', BD Barricor columns
serum, BD Vacutainer 8.5 mL plastic serum tube; pooled serum aliquoted into 1.5 mL Eppendorf tubesFreeze–thawseparated−20 °C5 cyclesthen changed, changed by 6 cyclesunstable only if both P < 0.005 (Bonferroni) and % change from T0 > desirable bias (Westgard/Ricos), here 4.3%Cuhadar et al. 2013, Biochem Med (Zagreb)Table 2, row 'LD (U/L)', columns T1d-T10d
serum, BD Vacutainer 8.5 mL plastic serum tube; pooled serum aliquoted into 1.5 mL Eppendorf tubesStorageseparated−20 °C3 monthsstable for the longest time testedunstable only if both P < 0.013 (Bonferroni) and % change from T0 > desirable bias (Westgard/Ricos), here 4.3%Cuhadar et al. 2013, Biochem Med (Zagreb)Table 2, row 'LD (U/L)', columns T1m-T3m
serum, Vacuette Tube Serum Gel Separator 8 mL (Greiner, 455071)Storagecentrifuged-on-gel2-8 °C1 weekstable for the longest time tested% bias of median vs baseline exceeds desirable bias (Westgard/Ricos BV), here 4.3%Cuhadar et al. 2012, Biochem Med (Zagreb)Table 3, row 'Lactate dehydrogenase', gel 4°C
serum, Vacuette Tube Serum No Gel 9 mL (Greiner, 455092), plainStoragecentrifuged-on-cells2-8 °C24 hthen changed, changed by 30 h% bias of median vs baseline exceeds desirable bias (Westgard/Ricos BV), here 4.3%Cuhadar et al. 2012, Biochem Med (Zagreb)Table 3, row 'Lactate dehydrogenase', w/o gel 4°C
serum, BD Vacutainer RST 4.0 mL (thrombin clot activator, gel)Storagecentrifuged-on-gel2-8 °C24 hstable for the longest time testedclinically significant if mean bias vs t0 exceeds desirable bias (biological variation), here 2.0%Kocijancic et al. 2014, Biochem Med (Zagreb)Table 3, row 'LD'
serum, BD Vacutainer SST II Advance 5.0 mL (clot activator, gel)Storagecentrifuged-on-gel2-8 °C24 hstable for the longest time testedclinically significant if mean bias vs t0 exceeds desirable bias (biological variation), here 2.0%Kocijancic et al. 2014, Biochem Med (Zagreb)Table 4, row 'LD'
serum, BD Vacutainer RST 5 mL (thrombin clot activator, gel)Storagecentrifuged-on-gel2-8 °C24 hthen changed, changed by 7 daysmean % difference vs time 0 exceeds desirable imprecision based on biological variation (EFLM/Westgard), here 2.6%Mandić et al. 2020, Biochem Med (Zagreb)Tables 3 and 4, row 'LD, U/L', BD RST columns
serum, Vacuette Tube Serum Gel Separator 8 mL (Greiner, 455071)Storagecentrifuged-on-gelRT1 weekstable for the longest time tested% bias of median vs baseline exceeds desirable bias (Westgard/Ricos BV), here 4.3%Cuhadar et al. 2012, Biochem Med (Zagreb)Table 3, row 'Lactate dehydrogenase', gel 24°C
serum, Vacuette Tube Serum No Gel 9 mL (Greiner, 455092), plainStoragecentrifuged-on-cellsRT72 hthen changed, changed by 1 week% bias of median vs baseline exceeds desirable bias (Westgard/Ricos BV), here 4.3%Cuhadar et al. 2012, Biochem Med (Zagreb)Table 3, row 'Lactate dehydrogenase', w/o gel 24°C
whole-blood-heparin, BD Vacutainer lithium heparin, no gel (ref 368884)Before spinninguncentrifugedRT10 hstable for the longest time tested|mean % deviation 10 h vs baseline| > laboratory bias goal (Table II; mostly biological-variation based, some empirically modified)Henriksen et al. 2014, Scand J Clin Lab InvestTable II, row 'LD'
whole-blood-heparin, BD Vacutainer lithium heparin, no gel (ref 368884)Before spinninguncentrifugedRT10 hstable for the longest time testedCV of paired baseline/10 h results > goal CV (Table II)Henriksen et al. 2014, Scand J Clin Lab InvestTable II, row 'LD'
LDL Cholesterol4 findings
Specimen, tubeConditionTemperatureStable forCriterionStudy
serum, BD Vacutainer SSTFreeze–thawseparated−70 °C, 1 month; thawed at room temperature1 cyclestable for the longest time testedmean % change (bias) from fresh baseline within total allowable error (TEa, desirable bias)Abraham et al. 2021, J Lab PhysiciansTable 2 and Table 3, row 'LDL (mg/dL)'
whole-blood-clotted, BD Vacutainer SSTBefore spinninguncentrifugedRT24 hstable for the longest time testedmean % deviation from baseline within total change limit TCL = sqrt((2.77 CVa)^2 + (0.5 CVb)^2), CVb from Ricos (Table 2)Abraham et al. 2019, Biochem Med (Zagreb)Discussion, paragraph 1 (Results paragraphs 1-3; Figure 3)
whole-blood-clotted, BD Vacutainer SSTBefore spinninguncentrifuged22-30 °C (logger 24-26 °C)24 hstable for the longest time testedmean % deviation from baseline within total change limit TCL = sqrt((2.77 CVa)^2 + (0.5 CVb)^2), CVb from Ricos (Table 2)Abraham et al. 2019, Biochem Med (Zagreb)Discussion, paragraph 1 (Results paragraphs 1-3; Figure 3)
whole-blood-clotted, BD Vacutainer SSTBefore spinninguncentrifuged> 30 °C (logger 38-40 °C)24 hstable for the longest time testedmean % deviation from baseline within total change limit TCL = sqrt((2.77 CVa)^2 + (0.5 CVb)^2), CVb from Ricos (Table 2)Abraham et al. 2019, Biochem Med (Zagreb)Discussion, paragraph 1 (Results paragraphs 1-3; Figure 3)
Lipase2 findings
Specimen, tubeConditionTemperatureStable forCriterionStudy
plasma-li-heparin, Sarstedt lithium heparin, no gel separator (primary tube)Storagecentrifuged-on-cells2-8 °C48 hstable for the longest time testedmean % difference vs time 0 exceeds total limit of change, TLC = ±√((1.65·CVa)² + (0.5·CVb)²) (EFLM BV), here 17.2%; stability limit from instability-equation regressionIllana et al. 2024, Biochem Med (Zagreb)Table 1, row 'Lip'
plasma-li-heparin, Sarstedt lithium heparin, no gel separator; plasma aliquotStorageseparated2-8 °C96 hstable for the longest time testedmean % difference vs time 0 exceeds total limit of change, TLC = ±√((1.65·CVa)² + (0.5·CVb)²) (EFLM BV), here 17.2%; stability limit from instability-equation regressionIllana et al. 2024, Biochem Med (Zagreb)Table 2, row 'Lip'
Low-fluorescence reticulocyte ratio (LFR)8 findings
Specimen, tubeConditionTemperatureStable forCriterionStudy
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C48 hstable for the longest time testedHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'LFR', XN-9000, 4 °C
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C48 hstable for the longest time testedHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'LFR', BC-6800, 4 °C
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C8 hstable for the longest time testedHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'LFR', XN-9000, 4 °C
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C8 hstable for the longest time testedHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'LFR', BC-6800, 4 °C
whole-blood-edta, K3-EDTA (Becton Dickinson)StorageuncentrifugedRT8 hthen changed, changed by 24 hHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'LFR', XN-9000, RT
whole-blood-edta, K3-EDTA (Becton Dickinson)StorageuncentrifugedRT24 hthen changed, changed by 36 hHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'LFR', BC-6800, RT
whole-blood-edta, K3-EDTA (Becton Dickinson)StorageuncentrifugedRT8 hstable for the longest time testedHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'LFR', XN-9000, RT
whole-blood-edta, K3-EDTA (Becton Dickinson)StorageuncentrifugedRT8 hstable for the longest time testedHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'LFR', BC-6800, RT
Lupus anticoagulant screen/confirm ratio3 findings
Specimen, tubeConditionTemperatureStable forCriterionStudy
plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquotsFreeze–thawseparated−80 °C1 cyclelimit stated by the authorsmax allowable mean deviation from t0: ±10.0% up to INR 1.5;±20.0% for INR ≥1.5 based on secondsHoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability'
plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquotsStorageseparated−80 °C12 weekslimit stated by the authorsmax allowable mean deviation from t0: ±10.0% up to INR 1.5;±20.0% for INR ≥1.5 based on secondsHoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability'
plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquotsStorageseparatedRT4 hlimit stated by the authorsmax allowable mean deviation from t0: ±10.0% up to INR 1.5;±20.0% for INR ≥1.5 based on secondsHoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability'
Lutein4 findings
Specimen, tubeConditionTemperatureStable forCriterionStudy
whole-blood-clotted, Sarstedt Monovette serumBefore spinninguncentrifugedchilled in transit (insulated box, 2 frozen cool packs; temperature not recorded)24 hstable for the longest time tested% difference within desirable quality specification for bias (Fraser, from intra- and interindividual variation)Jones et al. 2021, J NutrTable 2, row 'Lutein', column Serum
whole-blood-edta, Sarstedt Monovette EDTABefore spinninguncentrifugedchilled in transit (insulated box, 2 frozen cool packs; temperature not recorded)24 hstable for the longest time tested% difference within desirable quality specification for bias (Fraser, from intra- and interindividual variation)Jones et al. 2021, J NutrTable 2, row 'Lutein', column EDTA
whole-blood-heparin, BD Vacutainer lithium heparinBefore spinninguncentrifugedcold packs in insulated box; 9 (2) °C on arrival24 hstable for the longest time tested% difference within desirable quality specification (from intra- and interindividual variation)Jones et al. 2024, J NutrTable 2, row 'Lutein, μmol/L'; Table 1
whole-blood-heparin, Sarstedt Monovette lithium heparin, trace-element freeBefore spinninguncentrifugedchilled in transit (insulated box, 2 frozen cool packs; temperature not recorded)24 hstable for the longest time tested% difference within desirable quality specification for bias (Fraser, from intra- and interindividual variation)Jones et al. 2021, J NutrTable 2, row 'Lutein', column LH
LY-WX6 findings
Specimen, tubeConditionTemperatureStable forCriterionStudy
whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mLStorageuncentrifuged2-8 °C24 hstable for the longest time testedApproach 2: absolute % difference between medians (Tx vs T0) must be < 10%Roccaforte et al. 2026, Biochem Med (Zagreb)Table 1, row 'LY-WX', 4 °C, column 'Approach 2 (% difference between medians)'
whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mLStorageuncentrifuged2-8 °C24 hstable for the longest time testedEFLM regression: time at which fitted % difference line reaches MPE 10%Roccaforte et al. 2026, Biochem Med (Zagreb)Table 3, row 'LY-WX', 4 °C
whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mLStorageuncentrifugedRT24 hstable for the longest time testedApproach 2: absolute % difference between medians (Tx vs T0) must be < 10%Roccaforte et al. 2026, Biochem Med (Zagreb)Table 1, row 'LY-WX', RT, column 'Approach 2 (% difference between medians)'
whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mLStorageuncentrifugedRT24 hstable for the longest time testedEFLM regression: time at which fitted % difference line reaches MPE 10%Roccaforte et al. 2026, Biochem Med (Zagreb)Table 3, row 'LY-WX', RT
whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mLStorageuncentrifuged37 °C12 hthen changed, changed by 24 hApproach 2: absolute % difference between medians (Tx vs T0) must be < 10%Roccaforte et al. 2026, Biochem Med (Zagreb)Table 2, row 'LY-WX', 37 °C (blank in Table 1), column 'Approach 2 (% difference between medians)'
whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mLStorageuncentrifuged37 °C24 hstable for the longest time testedEFLM regression: time at which fitted % difference line reaches MPE 10%Roccaforte et al. 2026, Biochem Med (Zagreb)Table 3, row 'LY-WX', 37 °C
LY-WY4 findings
Specimen, tubeConditionTemperatureStable forCriterionStudy
whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mLStorageuncentrifuged2-8 °C24 hstable for the longest time testedApproach 2: absolute % difference between medians (Tx vs T0) must be < 10%Roccaforte et al. 2026, Biochem Med (Zagreb)Table 1, row 'LY-WY', 4 °C, column 'Approach 2 (% difference between medians)'
whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mLStorageuncentrifuged2-8 °C24 hstable for the longest time testedEFLM regression: time at which fitted % difference line reaches MPE 10%Roccaforte et al. 2026, Biochem Med (Zagreb)Table 3, row 'LY-WY', 4 °C
whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mLStorageuncentrifugedRT24 hstable for the longest time testedApproach 2: absolute % difference between medians (Tx vs T0) must be < 10%Roccaforte et al. 2026, Biochem Med (Zagreb)Table 1, row 'LY-WY', RT, column 'Approach 2 (% difference between medians)'
whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mLStorageuncentrifuged37 °C13 hlimit stated by the authorsEFLM regression: time at which fitted % difference line reaches MPE 10%Roccaforte et al. 2026, Biochem Med (Zagreb)Table 3, row 'LY-WY', 37 °C
LY-WZ5 findings
Specimen, tubeConditionTemperatureStable forCriterionStudy
whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mLStorageuncentrifuged2-8 °C24 hstable for the longest time testedApproach 2: absolute % difference between medians (Tx vs T0) must be < 10%Roccaforte et al. 2026, Biochem Med (Zagreb)Table 1, row 'LY-WZ', 4 °C, column 'Approach 2 (% difference between medians)'
whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mLStorageuncentrifuged2-8 °C24 hstable for the longest time testedEFLM regression: time at which fitted % difference line reaches MPE 10%Roccaforte et al. 2026, Biochem Med (Zagreb)Table 3, row 'LYWZ', 4 °C
whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mLStorageuncentrifugedRT24 hstable for the longest time testedApproach 2: absolute % difference between medians (Tx vs T0) must be < 10%Roccaforte et al. 2026, Biochem Med (Zagreb)Table 1, row 'LY-WZ', RT, column 'Approach 2 (% difference between medians)'
whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mLStorageuncentrifugedRT24 hstable for the longest time testedEFLM regression: time at which fitted % difference line reaches MPE 10%Roccaforte et al. 2026, Biochem Med (Zagreb)Table 3, row 'LYWZ', RT
whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mLStorageuncentrifuged37 °C7 hlimit stated by the authorsEFLM regression: time at which fitted % difference line reaches MPE 10%Roccaforte et al. 2026, Biochem Med (Zagreb)Table 3, row 'LYWZ', 37 °C
LY-X9 findings
Specimen, tubeConditionTemperatureStable forCriterionStudy
whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mLStorageuncentrifuged2-8 °C24 hstable for the longest time testedApproach 2: absolute % difference between medians (Tx vs T0) must be < 10%Roccaforte et al. 2026, Biochem Med (Zagreb)Table 1, row 'LY-X', 4 °C, column 'Approach 2 (% difference between medians)'
whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mLStorageuncentrifuged2-8 °C24 hstable for the longest time testedApproach 3: fewer than 2 of 25 samples with absolute % change from baseline > 10%Roccaforte et al. 2026, Biochem Med (Zagreb)Table 1, row 'LY-X', 4 °C, column 'Approach 3 (per-sample % difference)'
whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mLStorageuncentrifuged2-8 °C24 hstable for the longest time testedEFLM regression: time at which fitted % difference line reaches MPE 10%Roccaforte et al. 2026, Biochem Med (Zagreb)Table 3, row 'LY-X', 4 °C
whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mLStorageuncentrifugedRT24 hstable for the longest time testedApproach 2: absolute % difference between medians (Tx vs T0) must be < 10%Roccaforte et al. 2026, Biochem Med (Zagreb)Table 1, row 'LY-X', RT, column 'Approach 2 (% difference between medians)'
whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mLStorageuncentrifugedRT24 hstable for the longest time testedApproach 3: fewer than 2 of 25 samples with absolute % change from baseline > 10%Roccaforte et al. 2026, Biochem Med (Zagreb)Table 1, row 'LY-X', RT, column 'Approach 3 (per-sample % difference)'
whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mLStorageuncentrifugedRT24 hstable for the longest time testedEFLM regression: time at which fitted % difference line reaches MPE 10%Roccaforte et al. 2026, Biochem Med (Zagreb)Table 3, row 'LY-X', RT
whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mLStorageuncentrifuged37 °C24 hstable for the longest time testedApproach 2: absolute % difference between medians (Tx vs T0) must be < 10%Roccaforte et al. 2026, Biochem Med (Zagreb)Table 1, row 'LY-X', 37 °C, column 'Approach 2 (% difference between medians)'
whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mLStorageuncentrifuged37 °C12 hthen changed, changed by 24 hApproach 3: fewer than 2 of 25 samples with absolute % change from baseline > 10%Roccaforte et al. 2026, Biochem Med (Zagreb)Table 2, row 'LY-X', 37 °C (blank in Table 1), column 'Approach 3 (per-sample % difference)'
whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mLStorageuncentrifuged37 °C24 hstable for the longest time testedEFLM regression: time at which fitted % difference line reaches MPE 10%Roccaforte et al. 2026, Biochem Med (Zagreb)Table 3, row 'LY-X', 37 °C
LY-Y8 findings
Specimen, tubeConditionTemperatureStable forCriterionStudy
whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mLStorageuncentrifuged2-8 °C24 hstable for the longest time testedApproach 2: absolute % difference between medians (Tx vs T0) must be < 10%Roccaforte et al. 2026, Biochem Med (Zagreb)Table 1, row 'LY-Y', 4 °C, column 'Approach 2 (% difference between medians)'
whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mLStorageuncentrifuged2-8 °C24 hstable for the longest time testedApproach 3: fewer than 2 of 25 samples with absolute % change from baseline > 10%Roccaforte et al. 2026, Biochem Med (Zagreb)Table 1, row 'LY-Y', 4 °C, column 'Approach 3 (per-sample % difference)'
whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mLStorageuncentrifuged2-8 °C24 hstable for the longest time testedEFLM regression: time at which fitted % difference line reaches MPE 10%Roccaforte et al. 2026, Biochem Med (Zagreb)Table 3, row 'LY-Y', 4 °C
whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mLStorageuncentrifugedRT24 hstable for the longest time testedApproach 2: absolute % difference between medians (Tx vs T0) must be < 10%Roccaforte et al. 2026, Biochem Med (Zagreb)Table 1, row 'LY-Y', RT, column 'Approach 2 (% difference between medians)'
whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mLStorageuncentrifugedRT24 hstable for the longest time testedApproach 3: fewer than 2 of 25 samples with absolute % change from baseline > 10%Roccaforte et al. 2026, Biochem Med (Zagreb)Table 1, row 'LY-Y', RT, column 'Approach 3 (per-sample % difference)'
whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mLStorageuncentrifugedRT24 hstable for the longest time testedEFLM regression: time at which fitted % difference line reaches MPE 10%Roccaforte et al. 2026, Biochem Med (Zagreb)Table 3, row 'LY-Y', RT
whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mLStorageuncentrifuged37 °C12 hthen changed, changed by 24 hApproach 2: absolute % difference between medians (Tx vs T0) must be < 10%Roccaforte et al. 2026, Biochem Med (Zagreb)Table 2, row 'LY-Y', 37 °C (blank in Table 1), column 'Approach 2 (% difference between medians)'
whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mLStorageuncentrifuged37 °C8 hlimit stated by the authorsEFLM regression: time at which fitted % difference line reaches MPE 10%Roccaforte et al. 2026, Biochem Med (Zagreb)Table 3, row 'LY-Y', 37 °C
LY-Z6 findings
Specimen, tubeConditionTemperatureStable forCriterionStudy
whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mLStorageuncentrifuged2-8 °C17 hlimit stated by the authorsEFLM regression: time at which fitted % difference line reaches MPE 10%Roccaforte et al. 2026, Biochem Med (Zagreb)Table 3, row 'LY-Z', 4 °C
whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mLStorageuncentrifugedRT24 hstable for the longest time testedApproach 2: absolute % difference between medians (Tx vs T0) must be < 10%Roccaforte et al. 2026, Biochem Med (Zagreb)Table 1, row 'LY-Z', RT, column 'Approach 2 (% difference between medians)'
whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mLStorageuncentrifugedRT24 hstable for the longest time testedApproach 3: fewer than 2 of 25 samples with absolute % change from baseline > 10%Roccaforte et al. 2026, Biochem Med (Zagreb)Table 1, row 'LY-Z', RT, column 'Approach 3 (per-sample % difference)'
whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mLStorageuncentrifugedRT24 hstable for the longest time testedEFLM regression: time at which fitted % difference line reaches MPE 10%Roccaforte et al. 2026, Biochem Med (Zagreb)Table 3, row 'LY-Z', RT
whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mLStorageuncentrifuged37 °C12 hthen changed, changed by 24 hApproach 2: absolute % difference between medians (Tx vs T0) must be < 10%Roccaforte et al. 2026, Biochem Med (Zagreb)Table 2, row 'LY-Z', 37 °C (blank in Table 1), column 'Approach 2 (% difference between medians)'
whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mLStorageuncentrifuged37 °C14 hlimit stated by the authorsEFLM regression: time at which fitted % difference line reaches MPE 10%Roccaforte et al. 2026, Biochem Med (Zagreb)Table 3, row 'LY-Z', 37 °C
Lycopene4 findings
Specimen, tubeConditionTemperatureStable forCriterionStudy
whole-blood-clotted, Sarstedt Monovette serumBefore spinninguncentrifugedchilled in transit (insulated box, 2 frozen cool packs; temperature not recorded)24 hstable for the longest time tested% difference within desirable quality specification for bias (Fraser, from intra- and interindividual variation)Jones et al. 2021, J NutrTable 2, row 'Lycopene', column Serum
whole-blood-edta, Sarstedt Monovette EDTABefore spinninguncentrifugedchilled in transit (insulated box, 2 frozen cool packs; temperature not recorded)24 hstable for the longest time tested% difference within desirable quality specification for bias (Fraser, from intra- and interindividual variation)Jones et al. 2021, J NutrTable 2, row 'Lycopene', column EDTA
whole-blood-heparin, BD Vacutainer lithium heparinBefore spinninguncentrifugedcold packs in insulated box; 9 (2) °C on arrival24 hstable for the longest time tested% difference within desirable quality specification (from intra- and interindividual variation)Jones et al. 2024, J NutrTable 2, row 'Lycopene, μmol/L'; Table 1
whole-blood-heparin, Sarstedt Monovette lithium heparin, trace-element freeBefore spinninguncentrifugedchilled in transit (insulated box, 2 frozen cool packs; temperature not recorded)24 hstable for the longest time tested% difference within desirable quality specification for bias (Fraser, from intra- and interindividual variation)Jones et al. 2021, J NutrTable 2, row 'Lycopene', column LH
Lymphocytes (Absolute)21 findings
Specimen, tubeConditionTemperatureStable forCriterionStudy
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C48 hstable for the longest time testedHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 4, row 'LY', XN-9000, 4 °C
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C48 hstable for the longest time testedBias% vs optimal bias (OP-Bias%, from biological variation; Table 1)Buoro et al. 2016, Rev Bras Hematol HemoterTable 4, row 'LY', XN-9000, 4 °C
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C48 hstable for the longest time testedHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 4, row 'LY', BC-6800, 4 °C
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C8 hthen changed, changed by 24 hBias% vs optimal bias (OP-Bias%, from biological variation; Table 1)Buoro et al. 2016, Rev Bras Hematol HemoterTable 4, row 'LY', BC-6800, 4 °C
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C48 hstable for the longest time testedBias% vs critical difference (CD%, reference change value; Table 1)Buoro et al. 2016, Rev Bras Hematol HemoterTable 4, row 'LY', BC-6800, 4 °C
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C8 hstable for the longest time testedHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 5, row 'LY', XN-9000, 4 °C
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C8 hstable for the longest time testedBias% vs optimal bias (OP-Bias%, from biological variation; Table 1)Buoro et al. 2016, Rev Bras Hematol HemoterTable 5, row 'LY', XN-9000, 4 °C
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C8 hstable for the longest time testedHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 5, row 'LY', BC-6800, 4 °C
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C4 hthen changed, changed by 8 hBias% vs optimal bias (OP-Bias%, from biological variation; Table 1)Buoro et al. 2016, Rev Bras Hematol HemoterTable 5, row 'LY', BC-6800, 4 °C
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C8 hstable for the longest time testedBias% vs critical difference (CD%, reference change value; Table 1)Buoro et al. 2016, Rev Bras Hematol HemoterTable 5, row 'LY', BC-6800, 4 °C
whole-blood-edta, K2-EDTA (BD Vacutainer)Storageuncentrifuged2-8 °C48 hstable for the longest time testedRepeated-measures one-way ANOVA with Tukey post hoc vs 0 h, p < 0.05 (* in table)Unalli et al. 2021, J Med BiochemTable 1, column '4 °C' (no * at any time point)
whole-blood-edta, K3-EDTA (Becton Dickinson)StorageuncentrifugedRT48 hstable for the longest time testedHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 4, row 'LY', XN-9000, RT
whole-blood-edta, K3-EDTA (Becton Dickinson)StorageuncentrifugedRT48 hstable for the longest time testedBias% vs optimal bias (OP-Bias%, from biological variation; Table 1)Buoro et al. 2016, Rev Bras Hematol HemoterTable 4, row 'LY', XN-9000, RT
whole-blood-edta, K3-EDTA (Becton Dickinson)StorageuncentrifugedRT48 hstable for the longest time testedHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 4, row 'LY', BC-6800, RT
whole-blood-edta, K3-EDTA (Becton Dickinson)StorageuncentrifugedRT48 hstable for the longest time testedBias% vs optimal bias (OP-Bias%, from biological variation; Table 1)Buoro et al. 2016, Rev Bras Hematol HemoterTable 4, row 'LY', BC-6800, RT
whole-blood-edta, K3-EDTA (Becton Dickinson)StorageuncentrifugedRT8 hstable for the longest time testedHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 5, row 'LY', XN-9000, RT
whole-blood-edta, K3-EDTA (Becton Dickinson)StorageuncentrifugedRT8 hstable for the longest time testedBias% vs optimal bias (OP-Bias%, from biological variation; Table 1)Buoro et al. 2016, Rev Bras Hematol HemoterTable 5, row 'LY', XN-9000, RT
whole-blood-edta, K3-EDTA (Becton Dickinson)StorageuncentrifugedRT8 hstable for the longest time testedHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 5, row 'LY', BC-6800, RT
whole-blood-edta, K3-EDTA (Becton Dickinson)StorageuncentrifugedRT8 hstable for the longest time testedBias% vs optimal bias (OP-Bias%, from biological variation; Table 1)Buoro et al. 2016, Rev Bras Hematol HemoterTable 5, row 'LY', BC-6800, RT
whole-blood-edta, K2-EDTA (BD Vacutainer)StorageuncentrifugedRT8 hthen changed, changed by 12 hRepeated-measures one-way ANOVA with Tukey post hoc vs 0 h, p < 0.05 (* in table)Unalli et al. 2021, J Med BiochemTable 1, block 'LYM', row '12 h', column '23 °C'
whole-blood-edta, K2-EDTA (BD Vacutainer)Storageuncentrifuged10 °C24 hthen changed, changed by 48 hRepeated-measures one-way ANOVA with Tukey post hoc vs 0 h, p < 0.05 (* in table)Unalli et al. 2021, J Med BiochemTable 1, block 'LYM', row '48 h', column '10 °C'
Magnesium5 findings
Specimen, tubeConditionTemperatureStable forCriterionStudy
plasma-li-heparin, Sarstedt lithium heparin, no gel separator (primary tube)Storagecentrifuged-on-cells2-8 °C48 hstable for the longest time testedmean % difference vs time 0 exceeds total limit of change, TLC = ±√((1.65·CVa)² + (0.5·CVb)²) (EFLM BV), here 11.0%; stability limit from instability-equation regressionIllana et al. 2024, Biochem Med (Zagreb)Table 1, row 'Mg'
plasma-li-heparin, Sarstedt lithium heparin, no gel separator; plasma aliquotStorageseparated2-8 °C96 hstable for the longest time testedmean % difference vs time 0 exceeds total limit of change, TLC = ±√((1.65·CVa)² + (0.5·CVb)²) (EFLM BV), here 11.0%; stability limit from instability-equation regressionIllana et al. 2024, Biochem Med (Zagreb)Table 2, row 'Mg'
plasma-li-heparin, BD Vacutainer Barricor LH 3 mL, mechanical separatorStoragecentrifuged-on-gel2-8 °C75 hlimit stated by the authorsbias vs aliquoted plasma reference tube > maximal allowable bias 1.8% (Ricos et al.); time to MAB from linear regression of bias on time to analysisGawria et al. 2020, J Clin Lab AnalTable 2, row 'Magnesium', Barricor BD
plasma-li-heparin, BD Vacutainer PST LH 3 mL, gelStoragecentrifuged-on-gel2-8 °C80 hstable for the longest time testedbias vs aliquoted plasma reference tube > maximal allowable bias 1.8% (Ricos et al.); time to MAB from linear regression of bias on time to analysisGawria et al. 2020, J Clin Lab AnalTable 2, row 'Magnesium', Gel BD
plasma-li-heparin, Greiner Vacuette LH 3 mL Sep, gelStoragecentrifuged-on-gel2-8 °C0 hchanged at the first time point, changed by 3 hbias vs aliquoted plasma reference tube > maximal allowable bias 1.8% (Ricos et al.); time to MAB from linear regression of bias on time to analysisGawria et al. 2020, J Clin Lab AnalTable 2, row 'Magnesium', Gel Greiner
Mean Corpuscular Hemoglobin21 findings
Specimen, tubeConditionTemperatureStable forCriterionStudy
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C48 hstable for the longest time testedHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'MCH', XN-9000, 4 °C
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C48 hstable for the longest time testedBias% vs optimal bias (OP-Bias%, from biological variation; Table 1)Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'MCH', XN-9000, 4 °C
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C48 hstable for the longest time testedHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'MCH', BC-6800, 4 °C
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C48 hstable for the longest time testedBias% vs optimal bias (OP-Bias%, from biological variation; Table 1)Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'MCH', BC-6800, 4 °C
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C8 hstable for the longest time testedHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'MCH', XN-9000, 4 °C
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C8 hstable for the longest time testedBias% vs optimal bias (OP-Bias%, from biological variation; Table 1)Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'MCH', XN-9000, 4 °C
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C8 hstable for the longest time testedHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'MCH', BC-6800, 4 °C
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C4 hthen changed, changed by 8 hBias% vs optimal bias (OP-Bias%, from biological variation; Table 1)Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'MCH', BC-6800, 4 °C
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C8 hstable for the longest time testedBias% vs critical difference (CD%, reference change value; Table 1)Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'MCH', BC-6800, 4 °C
whole-blood-edta, K2-EDTA (BD Vacutainer)Storageuncentrifuged2-8 °C48 hstable for the longest time testedRepeated-measures one-way ANOVA with Tukey post hoc vs 0 h, p < 0.05 (* in table)Unalli et al. 2021, J Med BiochemTable 2, column '4 °C' (no * at any time point)
whole-blood-edta, K3-EDTA (Becton Dickinson)StorageuncentrifugedRT48 hstable for the longest time testedHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'MCH', XN-9000, RT
whole-blood-edta, K3-EDTA (Becton Dickinson)StorageuncentrifugedRT48 hstable for the longest time testedBias% vs optimal bias (OP-Bias%, from biological variation; Table 1)Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'MCH', XN-9000, RT
whole-blood-edta, K3-EDTA (Becton Dickinson)StorageuncentrifugedRT48 hstable for the longest time testedHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'MCH', BC-6800, RT
whole-blood-edta, K3-EDTA (Becton Dickinson)StorageuncentrifugedRT48 hstable for the longest time testedBias% vs optimal bias (OP-Bias%, from biological variation; Table 1)Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'MCH', BC-6800, RT
whole-blood-edta, K3-EDTA (Becton Dickinson)StorageuncentrifugedRT8 hstable for the longest time testedHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'MCH', XN-9000, RT
whole-blood-edta, K3-EDTA (Becton Dickinson)StorageuncentrifugedRT8 hstable for the longest time testedBias% vs optimal bias (OP-Bias%, from biological variation; Table 1)Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'MCH', XN-9000, RT
whole-blood-edta, K3-EDTA (Becton Dickinson)StorageuncentrifugedRT8 hstable for the longest time testedHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'MCH', BC-6800, RT
whole-blood-edta, K3-EDTA (Becton Dickinson)StorageuncentrifugedRT8 hstable for the longest time testedBias% vs optimal bias (OP-Bias%, from biological variation; Table 1)Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'MCH', BC-6800, RT
whole-blood-edta, K2-EDTA (BD Vacutainer)StorageuncentrifugedRT48 hstable for the longest time testedRepeated-measures one-way ANOVA with Tukey post hoc vs 0 h, p < 0.05 (* in table)Unalli et al. 2021, J Med BiochemTable 2, column '23 °C' (no * at any time point)
whole-blood-edta, K2EDTA (Becton Dickinson)Storageuncentrifuged33 °C24 hthen changed, changed by 48 hUnacceptable if median % difference > Biasmax (biological variation) and Holm-adjusted Wilcoxon P < 0.05; stable if equivalence shown (two one-tailed Wilcoxon tests vs Biasmax, Pholm < 0.05)Jain et al. 2018, Biochem Med (Zagreb)Results, last paragraph (equivalence); Table 2, row 'MCH'
whole-blood-edta, K2-EDTA (BD Vacutainer)Storageuncentrifuged10 °C48 hstable for the longest time testedRepeated-measures one-way ANOVA with Tukey post hoc vs 0 h, p < 0.05 (* in table)Unalli et al. 2021, J Med BiochemTable 2, column '10 °C' (no * at any time point)
Mean Corpuscular Hemoglobin Concentration25 findings
Specimen, tubeConditionTemperatureStable forCriterionStudy
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C0 hchanged at the first time point, changed by 2 hHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'MCHC', XN-9000, 4 °C
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C0 hchanged at the first time point, changed by 2 hBias% vs optimal bias (OP-Bias%, from biological variation; Table 1)Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'MCHC', XN-9000, 4 °C
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C48 hstable for the longest time testedBias% vs critical difference (CD%, reference change value; Table 1)Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'MCHC', XN-9000, 4 °C
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C36 hthen changed, changed by 48 hHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'MCHC', BC-6800, 4 °C
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C24 hthen changed, changed by 36 hBias% vs optimal bias (OP-Bias%, from biological variation; Table 1)Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'MCHC', BC-6800, 4 °C
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C48 hstable for the longest time testedBias% vs critical difference (CD%, reference change value; Table 1)Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'MCHC', BC-6800, 4 °C
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C8 hstable for the longest time testedHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'MCHC', XN-9000, 4 °C
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C8 hstable for the longest time testedBias% vs optimal bias (OP-Bias%, from biological variation; Table 1)Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'MCHC', XN-9000, 4 °C
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C4 hthen changed, changed by 8 hHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'MCHC', BC-6800, 4 °C
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C4 hthen changed, changed by 8 hBias% vs optimal bias (OP-Bias%, from biological variation; Table 1)Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'MCHC', BC-6800, 4 °C
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C8 hstable for the longest time testedBias% vs critical difference (CD%, reference change value; Table 1)Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'MCHC', BC-6800, 4 °C
whole-blood-edta, K2-EDTA (BD Vacutainer)Storageuncentrifuged2-8 °C48 hstable for the longest time testedRepeated-measures one-way ANOVA with Tukey post hoc vs 0 h, p < 0.05 (* in table)Unalli et al. 2021, J Med BiochemTable 2, column '4 °C' (no * at any time point)
whole-blood-edta, K3-EDTA (Becton Dickinson)StorageuncentrifugedRT2 hthen changed, changed by 4 hHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'MCHC', XN-9000, RT
whole-blood-edta, K3-EDTA (Becton Dickinson)StorageuncentrifugedRT0 hchanged at the first time point, changed by 2 hBias% vs optimal bias (OP-Bias%, from biological variation; Table 1)Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'MCHC', XN-9000, RT
whole-blood-edta, K3-EDTA (Becton Dickinson)StorageuncentrifugedRT8 hthen changed, changed by 24 hBias% vs critical difference (CD%, reference change value; Table 1)Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'MCHC', XN-9000, RT
whole-blood-edta, K3-EDTA (Becton Dickinson)StorageuncentrifugedRT8 hthen changed, changed by 24 hHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'MCHC', BC-6800, RT
whole-blood-edta, K3-EDTA (Becton Dickinson)StorageuncentrifugedRT8 hthen changed, changed by 24 hBias% vs optimal bias (OP-Bias%, from biological variation; Table 1)Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'MCHC', BC-6800, RT
whole-blood-edta, K3-EDTA (Becton Dickinson)StorageuncentrifugedRT8 hthen changed, changed by 24 hBias% vs critical difference (CD%, reference change value; Table 1)Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'MCHC', BC-6800, RT
whole-blood-edta, K3-EDTA (Becton Dickinson)StorageuncentrifugedRT8 hstable for the longest time testedHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'MCHC', XN-9000, RT
whole-blood-edta, K3-EDTA (Becton Dickinson)StorageuncentrifugedRT4 hthen changed, changed by 8 hBias% vs optimal bias (OP-Bias%, from biological variation; Table 1)Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'MCHC', XN-9000, RT
whole-blood-edta, K3-EDTA (Becton Dickinson)StorageuncentrifugedRT8 hstable for the longest time testedHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'MCHC', BC-6800, RT
whole-blood-edta, K3-EDTA (Becton Dickinson)StorageuncentrifugedRT8 hstable for the longest time testedBias% vs optimal bias (OP-Bias%, from biological variation; Table 1)Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'MCHC', BC-6800, RT
whole-blood-edta, K2-EDTA (BD Vacutainer)StorageuncentrifugedRT8 hthen changed, changed by 12 hRepeated-measures one-way ANOVA with Tukey post hoc vs 0 h, p < 0.05 (* in table)Unalli et al. 2021, J Med BiochemTable 2, block 'MCHC', row '12 h', column '23 °C'
whole-blood-edta, K2EDTA (Becton Dickinson)Storageuncentrifuged33 °C0 hchanged at the first time point, changed by 4 hUnacceptable if median % difference > Biasmax (biological variation) and Holm-adjusted Wilcoxon P < 0.05; stable if equivalence shown (two one-tailed Wilcoxon tests vs Biasmax, Pholm < 0.05)Jain et al. 2018, Biochem Med (Zagreb)Results paragraph 1; Table 2, row 'MCHC'
whole-blood-edta, K2-EDTA (BD Vacutainer)Storageuncentrifuged10 °C48 hstable for the longest time testedRepeated-measures one-way ANOVA with Tukey post hoc vs 0 h, p < 0.05 (* in table)Unalli et al. 2021, J Med BiochemTable 2, column '10 °C' (no * at any time point)
Mean Corpuscular Volume25 findings
Specimen, tubeConditionTemperatureStable forCriterionStudy
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C48 hstable for the longest time testedHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'MCV', XN-9000, 4 °C
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C24 hthen changed, changed by 36 hBias% vs optimal bias (OP-Bias%, from biological variation; Table 1)Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'MCV', XN-9000, 4 °C
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C48 hstable for the longest time testedBias% vs critical difference (CD%, reference change value; Table 1)Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'MCV', XN-9000, 4 °C
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C48 hstable for the longest time testedHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'MCV', BC-6800, 4 °C
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C36 hthen changed, changed by 48 hBias% vs optimal bias (OP-Bias%, from biological variation; Table 1)Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'MCV', BC-6800, 4 °C
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C48 hstable for the longest time testedBias% vs critical difference (CD%, reference change value; Table 1)Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'MCV', BC-6800, 4 °C
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C8 hstable for the longest time testedHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'MCV', XN-9000, 4 °C
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C8 hstable for the longest time testedBias% vs optimal bias (OP-Bias%, from biological variation; Table 1)Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'MCV', XN-9000, 4 °C
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C8 hstable for the longest time testedHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'MCV', BC-6800, 4 °C
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C8 hstable for the longest time testedBias% vs optimal bias (OP-Bias%, from biological variation; Table 1)Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'MCV', BC-6800, 4 °C
whole-blood-edta, K2-EDTA (BD Vacutainer)Storageuncentrifuged2-8 °C48 hstable for the longest time testedRepeated-measures one-way ANOVA with Tukey post hoc vs 0 h, p < 0.05 (* in table)Unalli et al. 2021, J Med BiochemTable 2, column '4 °C' (no * at any time point)
whole-blood-edta, K3-EDTA (Becton Dickinson)StorageuncentrifugedRT8 hthen changed, changed by 24 hHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'MCV', XN-9000, RT
whole-blood-edta, K3-EDTA (Becton Dickinson)StorageuncentrifugedRT2 hthen changed, changed by 4 hBias% vs optimal bias (OP-Bias%, from biological variation; Table 1)Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'MCV', XN-9000, RT
whole-blood-edta, K3-EDTA (Becton Dickinson)StorageuncentrifugedRT8 hthen changed, changed by 24 hBias% vs critical difference (CD%, reference change value; Table 1)Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'MCV', XN-9000, RT
whole-blood-edta, K3-EDTA (Becton Dickinson)StorageuncentrifugedRT8 hthen changed, changed by 24 hHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'MCV', BC-6800, RT
whole-blood-edta, K3-EDTA (Becton Dickinson)StorageuncentrifugedRT8 hthen changed, changed by 24 hBias% vs optimal bias (OP-Bias%, from biological variation; Table 1)Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'MCV', BC-6800, RT
whole-blood-edta, K3-EDTA (Becton Dickinson)StorageuncentrifugedRT8 hthen changed, changed by 24 hBias% vs critical difference (CD%, reference change value; Table 1)Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'MCV', BC-6800, RT
whole-blood-edta, K3-EDTA (Becton Dickinson)StorageuncentrifugedRT8 hstable for the longest time testedHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'MCV', XN-9000, RT
whole-blood-edta, K3-EDTA (Becton Dickinson)StorageuncentrifugedRT8 hstable for the longest time testedBias% vs optimal bias (OP-Bias%, from biological variation; Table 1)Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'MCV', XN-9000, RT
whole-blood-edta, K3-EDTA (Becton Dickinson)StorageuncentrifugedRT8 hstable for the longest time testedHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'MCV', BC-6800, RT
whole-blood-edta, K3-EDTA (Becton Dickinson)StorageuncentrifugedRT4 hthen changed, changed by 8 hBias% vs optimal bias (OP-Bias%, from biological variation; Table 1)Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'MCV', BC-6800, RT
whole-blood-edta, K3-EDTA (Becton Dickinson)StorageuncentrifugedRT8 hstable for the longest time testedBias% vs critical difference (CD%, reference change value; Table 1)Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'MCV', BC-6800, RT
whole-blood-edta, K2-EDTA (BD Vacutainer)StorageuncentrifugedRT24 hthen changed, changed by 48 hRepeated-measures one-way ANOVA with Tukey post hoc vs 0 h, p < 0.05 (* in table)Unalli et al. 2021, J Med BiochemTable 2, block 'MCV', row '48 h', column '23 °C'
whole-blood-edta, K2EDTA (Becton Dickinson)Storageuncentrifuged33 °C0 hchanged at the first time point, changed by 4 hUnacceptable if median % difference > Biasmax (biological variation) and Holm-adjusted Wilcoxon P < 0.05; stable if equivalence shown (two one-tailed Wilcoxon tests vs Biasmax, Pholm < 0.05)Jain et al. 2018, Biochem Med (Zagreb)Results paragraph 1; Table 2, row 'MCV'
whole-blood-edta, K2-EDTA (BD Vacutainer)Storageuncentrifuged10 °C48 hstable for the longest time testedRepeated-measures one-way ANOVA with Tukey post hoc vs 0 h, p < 0.05 (* in table)Unalli et al. 2021, J Med BiochemTable 2, column '10 °C' (no * at any time point)
Mean Platelet Volume27 findings
Specimen, tubeConditionTemperatureStable forCriterionStudy
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C36 hthen changed, changed by 48 hHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 4, row 'MPV', XN-9000, 4 °C
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C8 hthen changed, changed by 24 hBias% vs optimal bias (OP-Bias%, from biological variation; Table 1)Buoro et al. 2016, Rev Bras Hematol HemoterTable 4, row 'MPV', XN-9000, 4 °C
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C48 hstable for the longest time testedBias% vs critical difference (CD%, reference change value; Table 1)Buoro et al. 2016, Rev Bras Hematol HemoterTable 4, row 'MPV', XN-9000, 4 °C
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C8 hthen changed, changed by 24 hHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 4, row 'MPV', BC-6800, 4 °C
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C48 hstable for the longest time testedBias% vs optimal bias (OP-Bias%, from biological variation; Table 1)Buoro et al. 2016, Rev Bras Hematol HemoterTable 4, row 'MPV', BC-6800, 4 °C
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C48 hstable for the longest time testedBias% vs critical difference (CD%, reference change value; Table 1)Buoro et al. 2016, Rev Bras Hematol HemoterTable 4, row 'MPV', BC-6800, 4 °C
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C8 hstable for the longest time testedHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 5, row 'MPV', XN-9000, 4 °C
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C4 hthen changed, changed by 8 hBias% vs optimal bias (OP-Bias%, from biological variation; Table 1)Buoro et al. 2016, Rev Bras Hematol HemoterTable 5, row 'MPV', XN-9000, 4 °C
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C8 hstable for the longest time testedBias% vs critical difference (CD%, reference change value; Table 1)Buoro et al. 2016, Rev Bras Hematol HemoterTable 5, row 'MPV', XN-9000, 4 °C
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C8 hstable for the longest time testedHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 5, row 'MPV', BC-6800, 4 °C
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C0 hchanged at the first time point, changed by 4 hBias% vs optimal bias (OP-Bias%, from biological variation; Table 1)Buoro et al. 2016, Rev Bras Hematol HemoterTable 5, row 'MPV', BC-6800, 4 °C
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C8 hstable for the longest time testedBias% vs critical difference (CD%, reference change value; Table 1)Buoro et al. 2016, Rev Bras Hematol HemoterTable 5, row 'MPV', BC-6800, 4 °C
whole-blood-edta, K2-EDTA (BD Vacutainer)Storageuncentrifuged2-8 °C16 hthen changed, changed by 20 hRepeated-measures one-way ANOVA with Tukey post hoc vs 0 h, p < 0.05 (* in table)Unalli et al. 2021, J Med BiochemTable 2, block 'MPV', row '20 h', column '4 °C'
whole-blood-edta, K3-EDTA (Becton Dickinson)StorageuncentrifugedRT8 hthen changed, changed by 24 hHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 4, row 'MPV', XN-9000, RT
whole-blood-edta, K3-EDTA (Becton Dickinson)StorageuncentrifugedRT0 hchanged at the first time point, changed by 2 hBias% vs optimal bias (OP-Bias%, from biological variation; Table 1)Buoro et al. 2016, Rev Bras Hematol HemoterTable 4, row 'MPV', XN-9000, RT
whole-blood-edta, K3-EDTA (Becton Dickinson)StorageuncentrifugedRT48 hstable for the longest time testedBias% vs critical difference (CD%, reference change value; Table 1)Buoro et al. 2016, Rev Bras Hematol HemoterTable 4, row 'MPV', XN-9000, RT
whole-blood-edta, K3-EDTA (Becton Dickinson)StorageuncentrifugedRT2 hthen changed, changed by 4 hHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 4, row 'MPV', BC-6800, RT
whole-blood-edta, K3-EDTA (Becton Dickinson)StorageuncentrifugedRT0 hchanged at the first time point, changed by 2 hBias% vs optimal bias (OP-Bias%, from biological variation; Table 1)Buoro et al. 2016, Rev Bras Hematol HemoterTable 4, row 'MPV', BC-6800, RT
whole-blood-edta, K3-EDTA (Becton Dickinson)StorageuncentrifugedRT24 hthen changed, changed by 36 hBias% vs critical difference (CD%, reference change value; Table 1)Buoro et al. 2016, Rev Bras Hematol HemoterTable 4, row 'MPV', BC-6800, RT
whole-blood-edta, K3-EDTA (Becton Dickinson)StorageuncentrifugedRT8 hstable for the longest time testedHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 5, row 'MPV', XN-9000, RT
whole-blood-edta, K3-EDTA (Becton Dickinson)StorageuncentrifugedRT8 hstable for the longest time testedBias% vs optimal bias (OP-Bias%, from biological variation; Table 1)Buoro et al. 2016, Rev Bras Hematol HemoterTable 5, row 'MPV', XN-9000, RT
whole-blood-edta, K3-EDTA (Becton Dickinson)StorageuncentrifugedRT8 hstable for the longest time testedHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 5, row 'MPV', BC-6800, RT
whole-blood-edta, K3-EDTA (Becton Dickinson)StorageuncentrifugedRT0 hchanged at the first time point, changed by 4 hBias% vs optimal bias (OP-Bias%, from biological variation; Table 1)Buoro et al. 2016, Rev Bras Hematol HemoterTable 5, row 'MPV', BC-6800, RT
whole-blood-edta, K3-EDTA (Becton Dickinson)StorageuncentrifugedRT8 hstable for the longest time testedBias% vs critical difference (CD%, reference change value; Table 1)Buoro et al. 2016, Rev Bras Hematol HemoterTable 5, row 'MPV', BC-6800, RT
whole-blood-edta, K2-EDTA (BD Vacutainer)StorageuncentrifugedRT20 hthen changed, changed by 24 hRepeated-measures one-way ANOVA with Tukey post hoc vs 0 h, p < 0.05 (* in table)Unalli et al. 2021, J Med BiochemTable 2, block 'MPV', row '24 h', column '23 °C'
whole-blood-edta, K2EDTA (Becton Dickinson)Storageuncentrifuged33 °C0 hchanged at the first time point, changed by 4 hUnacceptable if median % difference > Biasmax (biological variation) and Holm-adjusted Wilcoxon P < 0.05; stable if equivalence shown (two one-tailed Wilcoxon tests vs Biasmax, Pholm < 0.05)Jain et al. 2018, Biochem Med (Zagreb)Results paragraph 1; Table 2, row 'MPV'
whole-blood-edta, K2-EDTA (BD Vacutainer)Storageuncentrifuged10 °C16 hthen changed, changed by 20 hRepeated-measures one-way ANOVA with Tukey post hoc vs 0 h, p < 0.05 (* in table)Unalli et al. 2021, J Med BiochemTable 2, block 'MPV', row '20 h', column '10 °C'
Medium-fluorescence reticulocyte ratio (MFR)8 findings
Specimen, tubeConditionTemperatureStable forCriterionStudy
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C48 hstable for the longest time testedHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'MFR', XN-9000, 4 °C
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C48 hstable for the longest time testedHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'MFR', BC-6800, 4 °C
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C8 hstable for the longest time testedHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'MFR', XN-9000, 4 °C
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C8 hstable for the longest time testedHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'MFR', BC-6800, 4 °C
whole-blood-edta, K3-EDTA (Becton Dickinson)StorageuncentrifugedRT8 hthen changed, changed by 24 hHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'MFR', XN-9000, RT
whole-blood-edta, K3-EDTA (Becton Dickinson)StorageuncentrifugedRT24 hthen changed, changed by 36 hHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'MFR', BC-6800, RT
whole-blood-edta, K3-EDTA (Becton Dickinson)StorageuncentrifugedRT8 hstable for the longest time testedHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'MFR', XN-9000, RT
whole-blood-edta, K3-EDTA (Becton Dickinson)StorageuncentrifugedRT8 hstable for the longest time testedHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'MFR', BC-6800, RT
MO-WX5 findings
Specimen, tubeConditionTemperatureStable forCriterionStudy
whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mLStorageuncentrifuged2-8 °C24 hstable for the longest time testedApproach 2: absolute % difference between medians (Tx vs T0) must be < 10%Roccaforte et al. 2026, Biochem Med (Zagreb)Table 1, row 'MO-WX', 4 °C, column 'Approach 2 (% difference between medians)'
whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mLStorageuncentrifuged2-8 °C24 hstable for the longest time testedEFLM regression: time at which fitted % difference line reaches MPE 10%Roccaforte et al. 2026, Biochem Med (Zagreb)Table 3, row 'MO-WX', 4 °C
whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mLStorageuncentrifugedRT24 hstable for the longest time testedApproach 2: absolute % difference between medians (Tx vs T0) must be < 10%Roccaforte et al. 2026, Biochem Med (Zagreb)Table 1, row 'MO-WX', RT, column 'Approach 2 (% difference between medians)'
whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mLStorageuncentrifugedRT24 hstable for the longest time testedEFLM regression: time at which fitted % difference line reaches MPE 10%Roccaforte et al. 2026, Biochem Med (Zagreb)Table 3, row 'MO-WX', RT
whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mLStorageuncentrifuged37 °C4 hlimit stated by the authorsEFLM regression: time at which fitted % difference line reaches MPE 10%Roccaforte et al. 2026, Biochem Med (Zagreb)Table 3, row 'MO-WX', 37 °C
MO-WY4 findings
Specimen, tubeConditionTemperatureStable forCriterionStudy
whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mLStorageuncentrifuged2-8 °C24 hstable for the longest time testedEFLM regression: time at which fitted % difference line reaches MPE 10%Roccaforte et al. 2026, Biochem Med (Zagreb)Table 3, row 'MO-WY', 4 °C
whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mLStorageuncentrifugedRT12 hthen changed, changed by 24 hApproach 2: absolute % difference between medians (Tx vs T0) must be < 10%Roccaforte et al. 2026, Biochem Med (Zagreb)Table 2, row 'MO-WY', RT (blank in Table 1), column 'Approach 2 (% difference between medians)'
whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mLStorageuncentrifugedRT17 hlimit stated by the authorsEFLM regression: time at which fitted % difference line reaches MPE 10%Roccaforte et al. 2026, Biochem Med (Zagreb)Table 3, row 'MO-WY', RT
whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mLStorageuncentrifuged37 °C5 hlimit stated by the authorsEFLM regression: time at which fitted % difference line reaches MPE 10%Roccaforte et al. 2026, Biochem Med (Zagreb)Table 3, row 'MO-WY', 37 °C
MO-WZ5 findings
Specimen, tubeConditionTemperatureStable forCriterionStudy
whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mLStorageuncentrifuged2-8 °C24 hstable for the longest time testedApproach 2: absolute % difference between medians (Tx vs T0) must be < 10%Roccaforte et al. 2026, Biochem Med (Zagreb)Table 1, row 'MO-WZ', 4 °C, column 'Approach 2 (% difference between medians)'
whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mLStorageuncentrifuged2-8 °C24 hstable for the longest time testedEFLM regression: time at which fitted % difference line reaches MPE 10%Roccaforte et al. 2026, Biochem Med (Zagreb)Table 3, row 'MO-WZ', 4 °C
whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mLStorageuncentrifugedRT24 hstable for the longest time testedApproach 2: absolute % difference between medians (Tx vs T0) must be < 10%Roccaforte et al. 2026, Biochem Med (Zagreb)Table 1, row 'MO-WZ', RT, column 'Approach 2 (% difference between medians)'
whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mLStorageuncentrifugedRT24 hstable for the longest time testedEFLM regression: time at which fitted % difference line reaches MPE 10%Roccaforte et al. 2026, Biochem Med (Zagreb)Table 3, row 'MO-WZ', RT
whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mLStorageuncentrifuged37 °C8 hlimit stated by the authorsEFLM regression: time at which fitted % difference line reaches MPE 10%Roccaforte et al. 2026, Biochem Med (Zagreb)Table 3, row 'MO-WZ', 37 °C
MO-X9 findings
Specimen, tubeConditionTemperatureStable forCriterionStudy
whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mLStorageuncentrifuged2-8 °C24 hstable for the longest time testedApproach 2: absolute % difference between medians (Tx vs T0) must be < 10%Roccaforte et al. 2026, Biochem Med (Zagreb)Table 1, row 'MO-X', 4 °C, column 'Approach 2 (% difference between medians)'
whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mLStorageuncentrifuged2-8 °C24 hstable for the longest time testedApproach 3: fewer than 2 of 25 samples with absolute % change from baseline > 10%Roccaforte et al. 2026, Biochem Med (Zagreb)Table 1, row 'MO-X', 4 °C, column 'Approach 3 (per-sample % difference)'
whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mLStorageuncentrifuged2-8 °C24 hstable for the longest time testedEFLM regression: time at which fitted % difference line reaches MPE 10%Roccaforte et al. 2026, Biochem Med (Zagreb)Table 3, row 'MO-X', 4 °C
whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mLStorageuncentrifugedRT24 hstable for the longest time testedApproach 2: absolute % difference between medians (Tx vs T0) must be < 10%Roccaforte et al. 2026, Biochem Med (Zagreb)Table 1, row 'MO-X', RT, column 'Approach 2 (% difference between medians)'
whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mLStorageuncentrifugedRT24 hstable for the longest time testedApproach 3: fewer than 2 of 25 samples with absolute % change from baseline > 10%Roccaforte et al. 2026, Biochem Med (Zagreb)Table 1, row 'MO-X', RT, column 'Approach 3 (per-sample % difference)'
whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mLStorageuncentrifugedRT24 hstable for the longest time testedEFLM regression: time at which fitted % difference line reaches MPE 10%Roccaforte et al. 2026, Biochem Med (Zagreb)Table 3, row 'MO-X', RT
whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mLStorageuncentrifuged37 °C24 hstable for the longest time testedApproach 2: absolute % difference between medians (Tx vs T0) must be < 10%Roccaforte et al. 2026, Biochem Med (Zagreb)Table 1, row 'MO-X', 37 °C, column 'Approach 2 (% difference between medians)'
whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mLStorageuncentrifuged37 °C24 hstable for the longest time testedApproach 3: fewer than 2 of 25 samples with absolute % change from baseline > 10%Roccaforte et al. 2026, Biochem Med (Zagreb)Table 1, row 'MO-X', 37 °C, column 'Approach 3 (per-sample % difference)'
whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mLStorageuncentrifuged37 °C24 hstable for the longest time testedEFLM regression: time at which fitted % difference line reaches MPE 10%Roccaforte et al. 2026, Biochem Med (Zagreb)Table 3, row 'MO-X', 37 °C
MO-Y6 findings
Specimen, tubeConditionTemperatureStable forCriterionStudy
whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mLStorageuncentrifuged2-8 °C24 hstable for the longest time testedApproach 2: absolute % difference between medians (Tx vs T0) must be < 10%Roccaforte et al. 2026, Biochem Med (Zagreb)Table 1, row 'MO-Y', 4 °C, column 'Approach 2 (% difference between medians)'
whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mLStorageuncentrifuged2-8 °C24 hstable for the longest time testedEFLM regression: time at which fitted % difference line reaches MPE 10%Roccaforte et al. 2026, Biochem Med (Zagreb)Table 3, row 'MO-Y', 4 °C
whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mLStorageuncentrifugedRT24 hstable for the longest time testedApproach 2: absolute % difference between medians (Tx vs T0) must be < 10%Roccaforte et al. 2026, Biochem Med (Zagreb)Table 1, row 'MO-Y', RT, column 'Approach 2 (% difference between medians)'
whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mLStorageuncentrifugedRT24 hstable for the longest time testedEFLM regression: time at which fitted % difference line reaches MPE 10%Roccaforte et al. 2026, Biochem Med (Zagreb)Table 3, row 'MO-Y', RT
whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mLStorageuncentrifuged37 °C12 hthen changed, changed by 24 hApproach 2: absolute % difference between medians (Tx vs T0) must be < 10%Roccaforte et al. 2026, Biochem Med (Zagreb)Table 2, row 'MO-Y', 37 °C (blank in Table 1), column 'Approach 2 (% difference between medians)'
whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mLStorageuncentrifuged37 °C24 hstable for the longest time testedEFLM regression: time at which fitted % difference line reaches MPE 10%Roccaforte et al. 2026, Biochem Med (Zagreb)Table 3, row 'MO-Y', 37 °C
MO-Z5 findings
Specimen, tubeConditionTemperatureStable forCriterionStudy
whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mLStorageuncentrifuged2-8 °C15 hlimit stated by the authorsEFLM regression: time at which fitted % difference line reaches MPE 10%Roccaforte et al. 2026, Biochem Med (Zagreb)Table 3, row 'MO-Z', 4 °C
whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mLStorageuncentrifugedRT24 hstable for the longest time testedApproach 2: absolute % difference between medians (Tx vs T0) must be < 10%Roccaforte et al. 2026, Biochem Med (Zagreb)Table 1, row 'MO-Z', RT, column 'Approach 2 (% difference between medians)'
whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mLStorageuncentrifugedRT24 hstable for the longest time testedEFLM regression: time at which fitted % difference line reaches MPE 10%Roccaforte et al. 2026, Biochem Med (Zagreb)Table 3, row 'MO-Z', RT
whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mLStorageuncentrifuged37 °C12 hthen changed, changed by 24 hApproach 2: absolute % difference between medians (Tx vs T0) must be < 10%Roccaforte et al. 2026, Biochem Med (Zagreb)Table 2, row 'MO-Z', 37 °C (blank in Table 1), column 'Approach 2 (% difference between medians)'
whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mLStorageuncentrifuged37 °C11 hlimit stated by the authorsEFLM regression: time at which fitted % difference line reaches MPE 10%Roccaforte et al. 2026, Biochem Med (Zagreb)Table 3, row 'MO-Z', 37 °C
Monocytes (Absolute)23 findings
Specimen, tubeConditionTemperatureStable forCriterionStudy
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C48 hstable for the longest time testedHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 4, row 'MO', XN-9000, 4 °C
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C24 hthen changed, changed by 36 hBias% vs optimal bias (OP-Bias%, from biological variation; Table 1)Buoro et al. 2016, Rev Bras Hematol HemoterTable 4, row 'MO', XN-9000, 4 °C
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C48 hstable for the longest time testedBias% vs critical difference (CD%, reference change value; Table 1)Buoro et al. 2016, Rev Bras Hematol HemoterTable 4, row 'MO', XN-9000, 4 °C
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C36 hthen changed, changed by 48 hHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 4, row 'MO', BC-6800, 4 °C
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C24 hthen changed, changed by 36 hBias% vs optimal bias (OP-Bias%, from biological variation; Table 1)Buoro et al. 2016, Rev Bras Hematol HemoterTable 4, row 'MO', BC-6800, 4 °C
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C48 hstable for the longest time testedBias% vs critical difference (CD%, reference change value; Table 1)Buoro et al. 2016, Rev Bras Hematol HemoterTable 4, row 'MO', BC-6800, 4 °C
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C8 hstable for the longest time testedHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 5, row 'MO', XN-9000, 4 °C
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C8 hstable for the longest time testedBias% vs optimal bias (OP-Bias%, from biological variation; Table 1)Buoro et al. 2016, Rev Bras Hematol HemoterTable 5, row 'MO', XN-9000, 4 °C
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C8 hstable for the longest time testedHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 5, row 'MO', BC-6800, 4 °C
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C4 hthen changed, changed by 8 hBias% vs optimal bias (OP-Bias%, from biological variation; Table 1)Buoro et al. 2016, Rev Bras Hematol HemoterTable 5, row 'MO', BC-6800, 4 °C
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C8 hstable for the longest time testedBias% vs critical difference (CD%, reference change value; Table 1)Buoro et al. 2016, Rev Bras Hematol HemoterTable 5, row 'MO', BC-6800, 4 °C
whole-blood-edta, K2-EDTA (BD Vacutainer)Storageuncentrifuged2-8 °C48 hstable for the longest time testedRepeated-measures one-way ANOVA with Tukey post hoc vs 0 h, p < 0.05 (* in table)Unalli et al. 2021, J Med BiochemTable 1, column '4 °C' (no * at any time point)
whole-blood-edta, K3-EDTA (Becton Dickinson)StorageuncentrifugedRT48 hstable for the longest time testedHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 4, row 'MO', XN-9000, RT
whole-blood-edta, K3-EDTA (Becton Dickinson)StorageuncentrifugedRT48 hstable for the longest time testedBias% vs optimal bias (OP-Bias%, from biological variation; Table 1)Buoro et al. 2016, Rev Bras Hematol HemoterTable 4, row 'MO', XN-9000, RT
whole-blood-edta, K3-EDTA (Becton Dickinson)StorageuncentrifugedRT36 hthen changed, changed by 48 hHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 4, row 'MO', BC-6800, RT
whole-blood-edta, K3-EDTA (Becton Dickinson)StorageuncentrifugedRT36 hthen changed, changed by 48 hBias% vs optimal bias (OP-Bias%, from biological variation; Table 1)Buoro et al. 2016, Rev Bras Hematol HemoterTable 4, row 'MO', BC-6800, RT
whole-blood-edta, K3-EDTA (Becton Dickinson)StorageuncentrifugedRT48 hstable for the longest time testedBias% vs critical difference (CD%, reference change value; Table 1)Buoro et al. 2016, Rev Bras Hematol HemoterTable 4, row 'MO', BC-6800, RT
whole-blood-edta, K3-EDTA (Becton Dickinson)StorageuncentrifugedRT8 hstable for the longest time testedHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 5, row 'MO', XN-9000, RT
whole-blood-edta, K3-EDTA (Becton Dickinson)StorageuncentrifugedRT8 hstable for the longest time testedBias% vs optimal bias (OP-Bias%, from biological variation; Table 1)Buoro et al. 2016, Rev Bras Hematol HemoterTable 5, row 'MO', XN-9000, RT
whole-blood-edta, K3-EDTA (Becton Dickinson)StorageuncentrifugedRT8 hstable for the longest time testedHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 5, row 'MO', BC-6800, RT
whole-blood-edta, K3-EDTA (Becton Dickinson)StorageuncentrifugedRT8 hstable for the longest time testedBias% vs optimal bias (OP-Bias%, from biological variation; Table 1)Buoro et al. 2016, Rev Bras Hematol HemoterTable 5, row 'MO', BC-6800, RT
whole-blood-edta, K2-EDTA (BD Vacutainer)StorageuncentrifugedRT24 hthen changed, changed by 48 hRepeated-measures one-way ANOVA with Tukey post hoc vs 0 h, p < 0.05 (* in table)Unalli et al. 2021, J Med BiochemTable 1, block 'MONO', row '48 h', column '23 °C'
whole-blood-edta, K2-EDTA (BD Vacutainer)Storageuncentrifuged10 °C48 hstable for the longest time testedRepeated-measures one-way ANOVA with Tukey post hoc vs 0 h, p < 0.05 (* in table)Unalli et al. 2021, J Med BiochemTable 1, column '10 °C' (no * at any time point)
Myoglobin4 findings
Specimen, tubeConditionTemperatureStable forCriterionStudy
plasma-li-heparin, Sarstedt lithium heparin, no gel separator (primary tube)Storagecentrifuged-on-cells2-8 °C48 hstable for the longest time testedmean % difference vs time 0 exceeds total limit of change, TLC = ±√((1.65·CVa)² + (0.5·CVb)²) (EFLM BV), here 24.9%; stability limit from instability-equation regressionIllana et al. 2024, Biochem Med (Zagreb)Table 1, row 'Myoglobin'
plasma-li-heparin, Sarstedt lithium heparin, no gel separator; plasma aliquotStorageseparated2-8 °C96 hstable for the longest time testedmean % difference vs time 0 exceeds total limit of change, TLC = ±√((1.65·CVa)² + (0.5·CVb)²) (EFLM BV), here 24.9%; stability limit from instability-equation regressionIllana et al. 2024, Biochem Med (Zagreb)Table 2, row 'Myoglobin'
plasma-li-heparin, BD Barricor 4.5 mL (lithium heparin, mechanical separator)Storagecentrifuged-on-gel2-8 °C7 daysstable for the longest time testedmean % difference vs time 0 exceeds desirable imprecision based on biological variation (EFLM/Westgard), here 7.0%Mandić et al. 2020, Biochem Med (Zagreb)Tables 3 and 4, row 'Myoglobin, μg/L', BD Barricor columns
serum, BD Vacutainer RST 5 mL (thrombin clot activator, gel)Storagecentrifuged-on-gel2-8 °C7 daysstable for the longest time testedmean % difference vs time 0 exceeds desirable imprecision based on biological variation (EFLM/Westgard), here 7.0%Mandić et al. 2020, Biochem Med (Zagreb)Tables 3 and 4, row 'Myoglobin, μg/L', BD RST columns
N-Terminal Pro-BNP2 findings
Specimen, tubeConditionTemperatureStable forCriterionStudy
plasma-li-heparin, BD Vacutainer lithium heparin 6 mL (regular)Storagenot-stated2-8 °C72 hstable for the longest time testedmean % difference vs T0 exceeds total change limit (TCL), here ±18.35%Dupuy et al. 2018, Biochem Med (Zagreb)Table 2, BD lithium heparin tubes, row 'NT-proBNP, ng/L'
plasma-li-heparin, BD Barricor lithium heparin 5.5 mL (mechanical separator)Storagecentrifuged-on-gel2-8 °C72 hstable for the longest time testedmean % difference vs T0 exceeds total change limit (TCL), here ±18.35%Dupuy et al. 2018, Biochem Med (Zagreb)Table 2, BD Barricor lithium heparin tubes, row 'NT-proBNP, ng/L'
NE-FSC7 findings
Specimen, tubeConditionTemperatureStable forCriterionStudy
whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mLStorageuncentrifuged2-8 °C24 hstable for the longest time testedEFLM regression: time at which fitted % difference line reaches MPE 10%Roccaforte et al. 2026, Biochem Med (Zagreb)Table 3, row 'NE-FSC', 4 °C
whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mLStorageuncentrifugedRT24 hstable for the longest time testedApproach 2: absolute % difference between medians (Tx vs T0) must be < 10%Roccaforte et al. 2026, Biochem Med (Zagreb)Table 1, row 'NE-FSC', RT, column 'Approach 2 (% difference between medians)'
whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mLStorageuncentrifugedRT24 hstable for the longest time testedApproach 3: fewer than 2 of 25 samples with absolute % change from baseline > 10%Roccaforte et al. 2026, Biochem Med (Zagreb)Table 1, row 'NE-FSC', RT, column 'Approach 3 (per-sample % difference)'
whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mLStorageuncentrifugedRT24 hstable for the longest time testedEFLM regression: time at which fitted % difference line reaches MPE 10%Roccaforte et al. 2026, Biochem Med (Zagreb)Table 3, row 'NE-FSC', RT
whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mLStorageuncentrifuged37 °C24 hstable for the longest time testedApproach 2: absolute % difference between medians (Tx vs T0) must be < 10%Roccaforte et al. 2026, Biochem Med (Zagreb)Table 1, row 'NE-FSC', 37 °C, column 'Approach 2 (% difference between medians)'
whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mLStorageuncentrifuged37 °C24 hstable for the longest time testedApproach 3: fewer than 2 of 25 samples with absolute % change from baseline > 10%Roccaforte et al. 2026, Biochem Med (Zagreb)Table 1, row 'NE-FSC', 37 °C, column 'Approach 3 (per-sample % difference)'
whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mLStorageuncentrifuged37 °C24 hstable for the longest time testedEFLM regression: time at which fitted % difference line reaches MPE 10%Roccaforte et al. 2026, Biochem Med (Zagreb)Table 3, row 'NE-FSC', 37 °C
NE-SFL6 findings
Specimen, tubeConditionTemperatureStable forCriterionStudy
whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mLStorageuncentrifuged2-8 °C24 hstable for the longest time testedApproach 2: absolute % difference between medians (Tx vs T0) must be < 10%Roccaforte et al. 2026, Biochem Med (Zagreb)Table 1, row 'NE-SFL', 4 °C, column 'Approach 2 (% difference between medians)'
whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mLStorageuncentrifuged2-8 °C19 hlimit stated by the authorsEFLM regression: time at which fitted % difference line reaches MPE 10%Roccaforte et al. 2026, Biochem Med (Zagreb)Table 3, row 'NE-SFL', 4 °C
whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mLStorageuncentrifugedRT24 hstable for the longest time testedApproach 2: absolute % difference between medians (Tx vs T0) must be < 10%Roccaforte et al. 2026, Biochem Med (Zagreb)Table 1, row 'NE-SFL', RT, column 'Approach 2 (% difference between medians)'
whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mLStorageuncentrifugedRT24 hstable for the longest time testedEFLM regression: time at which fitted % difference line reaches MPE 10%Roccaforte et al. 2026, Biochem Med (Zagreb)Table 3, row 'NE-SFL', RT
whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mLStorageuncentrifuged37 °C12 hthen changed, changed by 24 hApproach 2: absolute % difference between medians (Tx vs T0) must be < 10%Roccaforte et al. 2026, Biochem Med (Zagreb)Table 2, row 'NE-SFL', 37 °C (blank in Table 1), column 'Approach 2 (% difference between medians)'
whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mLStorageuncentrifuged37 °C19 hlimit stated by the authorsEFLM regression: time at which fitted % difference line reaches MPE 10%Roccaforte et al. 2026, Biochem Med (Zagreb)Table 3, row 'NE-SFL', 37 °C
NE-SSC9 findings
Specimen, tubeConditionTemperatureStable forCriterionStudy
whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mLStorageuncentrifuged2-8 °C24 hstable for the longest time testedApproach 2: absolute % difference between medians (Tx vs T0) must be < 10%Roccaforte et al. 2026, Biochem Med (Zagreb)Table 1, row 'NE-SSC', 4 °C, column 'Approach 2 (% difference between medians)'
whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mLStorageuncentrifuged2-8 °C24 hstable for the longest time testedApproach 3: fewer than 2 of 25 samples with absolute % change from baseline > 10%Roccaforte et al. 2026, Biochem Med (Zagreb)Table 1, row 'NE-SSC', 4 °C, column 'Approach 3 (per-sample % difference)'
whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mLStorageuncentrifuged2-8 °C24 hstable for the longest time testedEFLM regression: time at which fitted % difference line reaches MPE 10%Roccaforte et al. 2026, Biochem Med (Zagreb)Table 3, row 'NE-SSC', 4 °C
whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mLStorageuncentrifugedRT24 hstable for the longest time testedApproach 2: absolute % difference between medians (Tx vs T0) must be < 10%Roccaforte et al. 2026, Biochem Med (Zagreb)Table 1, row 'NE-SSC', RT, column 'Approach 2 (% difference between medians)'
whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mLStorageuncentrifugedRT24 hstable for the longest time testedApproach 3: fewer than 2 of 25 samples with absolute % change from baseline > 10%Roccaforte et al. 2026, Biochem Med (Zagreb)Table 1, row 'NE-SSC', RT, column 'Approach 3 (per-sample % difference)'
whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mLStorageuncentrifugedRT24 hstable for the longest time testedEFLM regression: time at which fitted % difference line reaches MPE 10%Roccaforte et al. 2026, Biochem Med (Zagreb)Table 3, row 'NE-SSC', RT
whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mLStorageuncentrifuged37 °C24 hstable for the longest time testedApproach 2: absolute % difference between medians (Tx vs T0) must be < 10%Roccaforte et al. 2026, Biochem Med (Zagreb)Table 1, row 'NE-SSC', 37 °C, column 'Approach 2 (% difference between medians)'
whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mLStorageuncentrifuged37 °C24 hstable for the longest time testedApproach 3: fewer than 2 of 25 samples with absolute % change from baseline > 10%Roccaforte et al. 2026, Biochem Med (Zagreb)Table 1, row 'NE-SSC', 37 °C, column 'Approach 3 (per-sample % difference)'
whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mLStorageuncentrifuged37 °C24 hstable for the longest time testedEFLM regression: time at which fitted % difference line reaches MPE 10%Roccaforte et al. 2026, Biochem Med (Zagreb)Table 3, row 'NE-SSC', 37 °C
NE-WX6 findings
Specimen, tubeConditionTemperatureStable forCriterionStudy
whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mLStorageuncentrifuged2-8 °C24 hstable for the longest time testedApproach 2: absolute % difference between medians (Tx vs T0) must be < 10%Roccaforte et al. 2026, Biochem Med (Zagreb)Table 1, row 'NE-WX', 4 °C, column 'Approach 2 (% difference between medians)'
whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mLStorageuncentrifuged2-8 °C23 hlimit stated by the authorsEFLM regression: time at which fitted % difference line reaches MPE 10%Roccaforte et al. 2026, Biochem Med (Zagreb)Table 3, row 'NE-WX', 4 °C
whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mLStorageuncentrifugedRT12 hthen changed, changed by 24 hApproach 2: absolute % difference between medians (Tx vs T0) must be < 10%Roccaforte et al. 2026, Biochem Med (Zagreb)Table 2, row 'NE-WX', RT (blank in Table 1), column 'Approach 2 (% difference between medians)'
whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mLStorageuncentrifugedRT22 hlimit stated by the authorsEFLM regression: time at which fitted % difference line reaches MPE 10%Roccaforte et al. 2026, Biochem Med (Zagreb)Table 3, row 'NE-WX', RT
whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mLStorageuncentrifuged37 °C12 hthen changed, changed by 24 hApproach 2: absolute % difference between medians (Tx vs T0) must be < 10%Roccaforte et al. 2026, Biochem Med (Zagreb)Table 2, row 'NE-WX', 37 °C (blank in Table 1), column 'Approach 2 (% difference between medians)'
whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mLStorageuncentrifuged37 °C11 hlimit stated by the authorsEFLM regression: time at which fitted % difference line reaches MPE 10%Roccaforte et al. 2026, Biochem Med (Zagreb)Table 3, row 'NE-WX', 37 °C
NE-WY6 findings
Specimen, tubeConditionTemperatureStable forCriterionStudy
whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mLStorageuncentrifuged2-8 °C24 hstable for the longest time testedApproach 2: absolute % difference between medians (Tx vs T0) must be < 10%Roccaforte et al. 2026, Biochem Med (Zagreb)Table 1, row 'NE-WY', 4 °C, column 'Approach 2 (% difference between medians)'
whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mLStorageuncentrifuged2-8 °C24 hstable for the longest time testedEFLM regression: time at which fitted % difference line reaches MPE 10%Roccaforte et al. 2026, Biochem Med (Zagreb)Table 3, row 'NEWY', 4 °C
whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mLStorageuncentrifugedRT24 hstable for the longest time testedApproach 2: absolute % difference between medians (Tx vs T0) must be < 10%Roccaforte et al. 2026, Biochem Med (Zagreb)Table 1, row 'NE-WY', RT, column 'Approach 2 (% difference between medians)'
whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mLStorageuncentrifugedRT24 hstable for the longest time testedEFLM regression: time at which fitted % difference line reaches MPE 10%Roccaforte et al. 2026, Biochem Med (Zagreb)Table 3, row 'NEWY', RT
whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mLStorageuncentrifuged37 °C12 hthen changed, changed by 24 hApproach 2: absolute % difference between medians (Tx vs T0) must be < 10%Roccaforte et al. 2026, Biochem Med (Zagreb)Table 2, row 'NE-WY', 37 °C (blank in Table 1), column 'Approach 2 (% difference between medians)'
whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mLStorageuncentrifuged37 °C5 hlimit stated by the authorsEFLM regression: time at which fitted % difference line reaches MPE 10%Roccaforte et al. 2026, Biochem Med (Zagreb)Table 3, row 'NEWY', 37 °C
NE-WZ5 findings
Specimen, tubeConditionTemperatureStable forCriterionStudy
whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mLStorageuncentrifuged2-8 °C24 hstable for the longest time testedApproach 2: absolute % difference between medians (Tx vs T0) must be < 10%Roccaforte et al. 2026, Biochem Med (Zagreb)Table 1, row 'NE-WZ', 4 °C, column 'Approach 2 (% difference between medians)'
whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mLStorageuncentrifuged2-8 °C24 hstable for the longest time testedEFLM regression: time at which fitted % difference line reaches MPE 10%Roccaforte et al. 2026, Biochem Med (Zagreb)Table 3, row 'NE-WZ', 4 °C
whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mLStorageuncentrifugedRT12 hthen changed, changed by 24 hApproach 2: absolute % difference between medians (Tx vs T0) must be < 10%Roccaforte et al. 2026, Biochem Med (Zagreb)Table 2, row 'NE-WZ', RT (blank in Table 1), column 'Approach 2 (% difference between medians)'
whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mLStorageuncentrifugedRT13 hlimit stated by the authorsEFLM regression: time at which fitted % difference line reaches MPE 10%Roccaforte et al. 2026, Biochem Med (Zagreb)Table 3, row 'NE-WZ', RT
whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mLStorageuncentrifuged37 °C9 hlimit stated by the authorsEFLM regression: time at which fitted % difference line reaches MPE 10%Roccaforte et al. 2026, Biochem Med (Zagreb)Table 3, row 'NE-WZ', 37 °C
Neutrophils (Absolute)20 findings
Specimen, tubeConditionTemperatureStable forCriterionStudy
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C48 hstable for the longest time testedHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 4, row 'NE', XN-9000, 4 °C
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C48 hstable for the longest time testedBias% vs optimal bias (OP-Bias%, from biological variation; Table 1)Buoro et al. 2016, Rev Bras Hematol HemoterTable 4, row 'NE', XN-9000, 4 °C
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C48 hstable for the longest time testedHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 4, row 'NE', BC-6800, 4 °C
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C24 hthen changed, changed by 36 hBias% vs optimal bias (OP-Bias%, from biological variation; Table 1)Buoro et al. 2016, Rev Bras Hematol HemoterTable 4, row 'NE', BC-6800, 4 °C
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C8 hstable for the longest time testedHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 5, row 'NE', XN-9000, 4 °C
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C8 hstable for the longest time testedBias% vs optimal bias (OP-Bias%, from biological variation; Table 1)Buoro et al. 2016, Rev Bras Hematol HemoterTable 5, row 'NE', XN-9000, 4 °C
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C8 hstable for the longest time testedHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 5, row 'NE', BC-6800, 4 °C
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C4 hthen changed, changed by 8 hBias% vs optimal bias (OP-Bias%, from biological variation; Table 1)Buoro et al. 2016, Rev Bras Hematol HemoterTable 5, row 'NE', BC-6800, 4 °C
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C8 hstable for the longest time testedBias% vs critical difference (CD%, reference change value; Table 1)Buoro et al. 2016, Rev Bras Hematol HemoterTable 5, row 'NE', BC-6800, 4 °C
whole-blood-edta, K2-EDTA (BD Vacutainer)Storageuncentrifuged2-8 °C48 hstable for the longest time testedRepeated-measures one-way ANOVA with Tukey post hoc vs 0 h, p < 0.05 (* in table)Unalli et al. 2021, J Med BiochemTable 1, column '4 °C' (no * at any time point)
whole-blood-edta, K3-EDTA (Becton Dickinson)StorageuncentrifugedRT48 hstable for the longest time testedHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 4, row 'NE', XN-9000, RT
whole-blood-edta, K3-EDTA (Becton Dickinson)StorageuncentrifugedRT48 hstable for the longest time testedBias% vs optimal bias (OP-Bias%, from biological variation; Table 1)Buoro et al. 2016, Rev Bras Hematol HemoterTable 4, row 'NE', XN-9000, RT
whole-blood-edta, K3-EDTA (Becton Dickinson)StorageuncentrifugedRT48 hstable for the longest time testedHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 4, row 'NE', BC-6800, RT
whole-blood-edta, K3-EDTA (Becton Dickinson)StorageuncentrifugedRT48 hstable for the longest time testedBias% vs optimal bias (OP-Bias%, from biological variation; Table 1)Buoro et al. 2016, Rev Bras Hematol HemoterTable 4, row 'NE', BC-6800, RT
whole-blood-edta, K3-EDTA (Becton Dickinson)StorageuncentrifugedRT8 hstable for the longest time testedHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 5, row 'NE', XN-9000, RT
whole-blood-edta, K3-EDTA (Becton Dickinson)StorageuncentrifugedRT8 hstable for the longest time testedBias% vs optimal bias (OP-Bias%, from biological variation; Table 1)Buoro et al. 2016, Rev Bras Hematol HemoterTable 5, row 'NE', XN-9000, RT
whole-blood-edta, K3-EDTA (Becton Dickinson)StorageuncentrifugedRT8 hstable for the longest time testedHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 5, row 'NE', BC-6800, RT
whole-blood-edta, K3-EDTA (Becton Dickinson)StorageuncentrifugedRT8 hstable for the longest time testedBias% vs optimal bias (OP-Bias%, from biological variation; Table 1)Buoro et al. 2016, Rev Bras Hematol HemoterTable 5, row 'NE', BC-6800, RT
whole-blood-edta, K2-EDTA (BD Vacutainer)StorageuncentrifugedRT8 hthen changed, changed by 12 hRepeated-measures one-way ANOVA with Tukey post hoc vs 0 h, p < 0.05 (* in table)Unalli et al. 2021, J Med BiochemTable 1, block 'NEU', row '12 h', column '23 °C'
whole-blood-edta, K2-EDTA (BD Vacutainer)Storageuncentrifuged10 °C48 hstable for the longest time testedRepeated-measures one-way ANOVA with Tukey post hoc vs 0 h, p < 0.05 (* in table)Unalli et al. 2021, J Med BiochemTable 1, column '10 °C' (no * at any time point)
Nucleated Red Blood Cells (Absolute)8 findings
Specimen, tubeConditionTemperatureStable forCriterionStudy
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C48 hstable for the longest time testedHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'NRBC', XN-9000, 4 °C
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C48 hstable for the longest time testedHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'NRBC', BC-6800, 4 °C
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C0 hchanged at the first time point, changed by 4 hHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'NRBC', XN-9000, 4 °C
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C4 hthen changed, changed by 8 hHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'NRBC', BC-6800, 4 °C
whole-blood-edta, K3-EDTA (Becton Dickinson)StorageuncentrifugedRT48 hstable for the longest time testedHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'NRBC', XN-9000, RT
whole-blood-edta, K3-EDTA (Becton Dickinson)StorageuncentrifugedRT48 hstable for the longest time testedHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'NRBC', BC-6800, RT
whole-blood-edta, K3-EDTA (Becton Dickinson)StorageuncentrifugedRT4 hthen changed, changed by 8 hHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'NRBC', XN-9000, RT
whole-blood-edta, K3-EDTA (Becton Dickinson)StorageuncentrifugedRT8 hstable for the longest time testedHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'NRBC', BC-6800, RT
Orosomucoid (alpha-1-acid glycoprotein)2 findings
Specimen, tubeConditionTemperatureStable forCriterionStudy
whole-blood-heparin, BD Vacutainer lithium heparin, no gel (ref 368884)Before spinninguncentrifugedRT10 hstable for the longest time tested|mean % deviation 10 h vs baseline| > laboratory bias goal (Table II; mostly biological-variation based, some empirically modified)Henriksen et al. 2014, Scand J Clin Lab InvestTable II, row 'OROS'
whole-blood-heparin, BD Vacutainer lithium heparin, no gel (ref 368884)Before spinninguncentrifugedRT10 hstable for the longest time testedCV of paired baseline/10 h results > goal CV (Table II)Henriksen et al. 2014, Scand J Clin Lab InvestTable II, row 'OROS'
Paraoxonase-1 (PON1) activity3 findings
Specimen, tubeConditionTemperatureStable forCriterionStudy
serum, serum (tube not stated)Storageseparated−20 °C2 monthsthen changed, changed by 6 monthst-test (two-sample, equal variances) vs t=0, 95% CIBeekhof et al. 2012, Lipids Health DisResults, paragraph 3
serum, serum (tube not stated)Storageseparated−70°C12 monthsstable for the longest time testedt-test (two-sample, equal variances) vs t=0, 95% CIBeekhof et al. 2012, Lipids Health DisResults, paragraph 3; Discussion
serum, serum (tube not stated)Storageseparated−196°C12 monthsstable for the longest time testedt-test (two-sample, equal variances) vs t=0, 95% CIBeekhof et al. 2012, Lipids Health DisResults, paragraph 3; Discussion
Parathyroid Hormone (PTH)16 findings
Specimen, tubeConditionTemperatureStable forCriterionStudy
plasma-edta, BD Vacutainer EDTA-K2Storageseparated−20 °C8 daysstable for the longest time testedTEo (% change from initial) > TEa 30% (CLIA)Khalil et al. 2021, J Clin Lab AnalTable 2
plasma-edta, BD Vacutainer EDTA-K2Storageseparated−20 °C0 dayschanged at the first time point, changed by 2 daysLSD post hoc vs initial, p < 0.001 (repeated-measures ANOVA)Khalil et al. 2021, J Clin Lab AnalTable 2
plasma-edta, BD Vacutainer EDTA-K2Storageseparated2-8 °C8 daysstable for the longest time testedTEo (% change from initial) > TEa 30% (CLIA)Khalil et al. 2021, J Clin Lab AnalTable 2
plasma-edta, BD Vacutainer EDTA-K2Storageseparated2-8 °C0 dayschanged at the first time point, changed by 2 daysLSD post hoc vs initial, p < 0.001 (repeated-measures ANOVA)Khalil et al. 2021, J Clin Lab AnalTable 2
plasma-edta, BD Vacutainer EDTA-K2StorageseparatedRT4 daysthen changed, changed by 8 daysTEo (% change from initial) > TEa 30% (CLIA)Khalil et al. 2021, J Clin Lab AnalTable 2
plasma-edta, BD Vacutainer EDTA-K2StorageseparatedRT0 dayschanged at the first time point, changed by 2 daysLSD post hoc vs initial, p < 0.001 (repeated-measures ANOVA)Khalil et al. 2021, J Clin Lab AnalTable 2
serum, BD Vacutainer plain tube (no gel, no additive)Storageseparated−20 °C8 daysstable for the longest time testedTEo (% change from initial) > TEa 30% (CLIA)Khalil et al. 2021, J Clin Lab AnalTable 2
serum, BD Vacutainer plain tube (no gel, no additive)Storageseparated−20 °C0 dayschanged at the first time point, changed by 2 daysLSD post hoc vs initial, p < 0.001 (repeated-measures ANOVA)Khalil et al. 2021, J Clin Lab AnalTable 2
serum, BD Vacutainer plain tube (no gel, no additive)Storageseparated2-8 °C8 daysstable for the longest time testedTEo (% change from initial) > TEa 30% (CLIA)Khalil et al. 2021, J Clin Lab AnalTable 2
serum, BD Vacutainer plain tube (no gel, no additive)Storageseparated2-8 °C0 dayschanged at the first time point, changed by 2 daysLSD post hoc vs initial, p < 0.001 (repeated-measures ANOVA)Khalil et al. 2021, J Clin Lab AnalTable 2
serum, BD Vacutainer plain tube (no gel, no additive)StorageseparatedRT0 dayschanged at the first time point, changed by 2 daysTEo (% change from initial) > TEa 30% (CLIA)Khalil et al. 2021, J Clin Lab AnalTable 2
serum, BD Vacutainer plain tube (no gel, no additive)StorageseparatedRT0 dayschanged at the first time point, changed by 2 daysLSD post hoc vs initial, p < 0.001 (repeated-measures ANOVA)Khalil et al. 2021, J Clin Lab AnalTable 2
whole-blood-clotted, BD Vacutainer plain tube (no gel, no additive)Before spinninguncentrifugednot stated2 hstable for the longest time testedTEo (% change from initial) > TEa 30% (CLIA)Khalil et al. 2021, J Clin Lab AnalResults, 'Effects of a two-h delay in separation'
whole-blood-clotted, BD Vacutainer plain tube (no gel, no additive)Before spinninguncentrifugednot stated0 hchanged at the first time point, changed by 2 hpaired t test vs immediate separation, p < 0.05Khalil et al. 2021, J Clin Lab AnalResults, 'Effects of a two-h delay in separation'
whole-blood-edta, BD Vacutainer EDTA-K2Before spinninguncentrifugednot stated2 hstable for the longest time testedTEo (% change from initial) > TEa 30% (CLIA)Khalil et al. 2021, J Clin Lab AnalResults, 'Effects of a two-h delay in separation'
whole-blood-edta, BD Vacutainer EDTA-K2Before spinninguncentrifugednot stated0 hchanged at the first time point, changed by 2 hpaired t test vs immediate separation, p < 0.05Khalil et al. 2021, J Clin Lab AnalResults, 'Effects of a two-h delay in separation'
Phosphate (inorganic phosphorus)12 findings
Specimen, tubeConditionTemperatureStable forCriterionStudy
plasma-li-heparin, Sarstedt lithium heparin, no gel separator (primary tube)Storagecentrifuged-on-cells2-8 °C6.2 hthen changed, changed by 8 hmean % difference vs time 0 exceeds total limit of change, TLC = ±√((1.65·CVa)² + (0.5·CVb)²) (EFLM BV), here 7.4%; stability limit from instability-equation regressionIllana et al. 2024, Biochem Med (Zagreb)Table 1, row 'Phos'
plasma-li-heparin, Sarstedt lithium heparin, no gel separator; plasma aliquotStorageseparated2-8 °C96 hstable for the longest time testedmean % difference vs time 0 exceeds total limit of change, TLC = ±√((1.65·CVa)² + (0.5·CVb)²) (EFLM BV), here 7.4%; stability limit from instability-equation regressionIllana et al. 2024, Biochem Med (Zagreb)Table 2, row 'Phos'
plasma-li-heparin, BD Vacutainer LH PST II gel separator, 3.0 mL (REF 367374)Storagecentrifuged-on-gel2-8 °C24 hstable for the longest time testedmean % difference smaller than TEa (Ricos: 0.25 x (CVi^2 + CVg^2)^0.5 + 1.65 x 0.5 x CVi)Nonkes et al. 2019, Biochem Med (Zagreb)Table 1, row 'Phosphate, mmol/L'
plasma-li-heparin, BD Vacutainer Barricor LH 3 mL, mechanical separatorStoragecentrifuged-on-gel2-8 °C68 hlimit stated by the authorsbias vs aliquoted plasma reference tube > maximal allowable bias 3.2% (Ricos et al.); time to MAB from linear regression of bias on time to analysisGawria et al. 2020, J Clin Lab AnalTable 2, row 'Phosphate', Barricor BD
plasma-li-heparin, BD Vacutainer PST LH 3 mL, gelStoragecentrifuged-on-gel2-8 °C29 hlimit stated by the authorsbias vs aliquoted plasma reference tube > maximal allowable bias 3.2% (Ricos et al.); time to MAB from linear regression of bias on time to analysisGawria et al. 2020, J Clin Lab AnalTable 2, row 'Phosphate', Gel BD
plasma-li-heparin, Greiner Vacuette LH 3 mL Sep, gelStoragecentrifuged-on-gel2-8 °C35 hlimit stated by the authorsbias vs aliquoted plasma reference tube > maximal allowable bias 3.2% (Ricos et al.); time to MAB from linear regression of bias on time to analysisGawria et al. 2020, J Clin Lab AnalTable 2, row 'Phosphate', Gel Greiner
serum, BD Vacutainer serum (plain), 7.5 mLStorageseparated−20 °C30 daysstable for the longest time testedclinically significant change: mean outside significant change limit (SCL = T1d mean ± 3.0 usual SD from 2 months QC)Kachhawa et al. 2017, J Lab PhysiciansResults paragraph 1 (Table 2 image-only)
serum, BD Vacutainer SST (SKU 367986), gelStoragecentrifuged-on-gel30 °C (±1 °C)72 hstable for the longest time testedmean bias vs paired baseline within CLIA acceptance limits AND >2/3 of individual results within limitsIacovetti et al. 2026, Pract Lab MedResults, 'Analyte stability after storage at 30 °C', paragraph 1 (Table 1 image-only)
serum, BD SST Microtainer (SKU 395967), gel; filled from venous bloodStoragecentrifuged-on-gel0 °C72 hstable for the longest time testedmean bias vs paired baseline within CLIA acceptance limits AND >2/3 of individual results within limitsIacovetti et al. 2026, Pract Lab MedResults, 'Analyte stability after separation and storage in microvolume tubes at 0 °C' (Table 2 image-only)
whole-blood-clotted, BD Vacutainer SST (SKU 367986), gelBefore spinninguncentrifuged30 °C (±1 °C)0 hchanged at the first time point, changed by 24 hmean bias vs paired baseline within CLIA acceptance limits AND >2/3 of individual results within limitsIacovetti et al. 2026, Pract Lab MedResults, 'Analyte stability after storage at 30 °C', paragraph 2 (Table 1 image-only)
whole-blood-heparin, BD Vacutainer lithium heparin, no gel (ref 368884)Before spinninguncentrifugedRT0 hchanged at the first time point, changed by 10 h|mean % deviation 10 h vs baseline| > laboratory bias goal (Table II; mostly biological-variation based, some empirically modified)Henriksen et al. 2014, Scand J Clin Lab InvestTable II, row 'P'
whole-blood-heparin, BD Vacutainer lithium heparin, no gel (ref 368884)Before spinninguncentrifugedRT0 hchanged at the first time point, changed by 10 hCV of paired baseline/10 h results > goal CV (Table II)Henriksen et al. 2014, Scand J Clin Lab InvestTable II, row 'P'
Platelet Count21 findings
Specimen, tubeConditionTemperatureStable forCriterionStudy
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C48 hstable for the longest time testedHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 4, row 'PLT', XN-9000, 4 °C
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C0 hchanged at the first time point, changed by 2 hBias% vs optimal bias (OP-Bias%, from biological variation; Table 1)Buoro et al. 2016, Rev Bras Hematol HemoterTable 4, row 'PLT', XN-9000, 4 °C
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C48 hstable for the longest time testedBias% vs critical difference (CD%, reference change value; Table 1)Buoro et al. 2016, Rev Bras Hematol HemoterTable 4, row 'PLT', XN-9000, 4 °C
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C48 hstable for the longest time testedHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 4, row 'PLT', BC-6800, 4 °C
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C48 hstable for the longest time testedBias% vs optimal bias (OP-Bias%, from biological variation; Table 1)Buoro et al. 2016, Rev Bras Hematol HemoterTable 4, row 'PLT', BC-6800, 4 °C
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C8 hstable for the longest time testedHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 5, row 'PLT', XN-9000, 4 °C
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C8 hstable for the longest time testedBias% vs optimal bias (OP-Bias%, from biological variation; Table 1)Buoro et al. 2016, Rev Bras Hematol HemoterTable 5, row 'PLT', XN-9000, 4 °C
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C8 hstable for the longest time testedHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 5, row 'PLT', BC-6800, 4 °C
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C8 hstable for the longest time testedBias% vs optimal bias (OP-Bias%, from biological variation; Table 1)Buoro et al. 2016, Rev Bras Hematol HemoterTable 5, row 'PLT', BC-6800, 4 °C
whole-blood-edta, K2-EDTA (BD Vacutainer)Storageuncentrifuged2-8 °C48 hstable for the longest time testedRepeated-measures one-way ANOVA with Tukey post hoc vs 0 h, p < 0.05 (* in table)Unalli et al. 2021, J Med BiochemResults paragraphs 1-2 and Discussion paragraph 1 (Figures 1-2; numbers in figures only)
whole-blood-edta, K3-EDTA (Becton Dickinson)StorageuncentrifugedRT48 hstable for the longest time testedHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 4, row 'PLT', XN-9000, RT
whole-blood-edta, K3-EDTA (Becton Dickinson)StorageuncentrifugedRT48 hstable for the longest time testedBias% vs optimal bias (OP-Bias%, from biological variation; Table 1)Buoro et al. 2016, Rev Bras Hematol HemoterTable 4, row 'PLT', XN-9000, RT
whole-blood-edta, K3-EDTA (Becton Dickinson)StorageuncentrifugedRT48 hstable for the longest time testedHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 4, row 'PLT', BC-6800, RT
whole-blood-edta, K3-EDTA (Becton Dickinson)StorageuncentrifugedRT48 hstable for the longest time testedBias% vs optimal bias (OP-Bias%, from biological variation; Table 1)Buoro et al. 2016, Rev Bras Hematol HemoterTable 4, row 'PLT', BC-6800, RT
whole-blood-edta, K3-EDTA (Becton Dickinson)StorageuncentrifugedRT8 hstable for the longest time testedHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 5, row 'PLT', XN-9000, RT
whole-blood-edta, K3-EDTA (Becton Dickinson)StorageuncentrifugedRT8 hstable for the longest time testedBias% vs optimal bias (OP-Bias%, from biological variation; Table 1)Buoro et al. 2016, Rev Bras Hematol HemoterTable 5, row 'PLT', XN-9000, RT
whole-blood-edta, K3-EDTA (Becton Dickinson)StorageuncentrifugedRT8 hstable for the longest time testedHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 5, row 'PLT', BC-6800, RT
whole-blood-edta, K3-EDTA (Becton Dickinson)StorageuncentrifugedRT8 hstable for the longest time testedBias% vs optimal bias (OP-Bias%, from biological variation; Table 1)Buoro et al. 2016, Rev Bras Hematol HemoterTable 5, row 'PLT', BC-6800, RT
whole-blood-edta, K2-EDTA (BD Vacutainer)StorageuncentrifugedRT48 hstable for the longest time testedRepeated-measures one-way ANOVA with Tukey post hoc vs 0 h, p < 0.05 (* in table)Unalli et al. 2021, J Med BiochemResults paragraphs 1-2 and Discussion paragraph 1 (Figures 1-2; numbers in figures only)
whole-blood-edta, K2EDTA (Becton Dickinson)Storageuncentrifuged33 °C0 hchanged at the first time point, changed by 4 hUnacceptable if median % difference > Biasmax (biological variation) and Holm-adjusted Wilcoxon P < 0.05; stable if equivalence shown (two one-tailed Wilcoxon tests vs Biasmax, Pholm < 0.05)Jain et al. 2018, Biochem Med (Zagreb)Results paragraph 1; Table 2, row 'PLT'
whole-blood-edta, K2-EDTA (BD Vacutainer)Storageuncentrifuged10 °C48 hstable for the longest time testedRepeated-measures one-way ANOVA with Tukey post hoc vs 0 h, p < 0.05 (* in table)Unalli et al. 2021, J Med BiochemResults paragraphs 1-2 and Discussion paragraph 1 (Figures 1-2; numbers in figures only)
Platelet Distribution Width8 findings
Specimen, tubeConditionTemperatureStable forCriterionStudy
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C36 hthen changed, changed by 48 hHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 4, row 'PDW', XN-9000, 4 °C
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C8 hthen changed, changed by 24 hHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 4, row 'PDW', BC-6800, 4 °C
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C8 hstable for the longest time testedHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 5, row 'PDW', XN-9000, 4 °C
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C8 hstable for the longest time testedHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 5, row 'PDW', BC-6800, 4 °C
whole-blood-edta, K3-EDTA (Becton Dickinson)StorageuncentrifugedRT8 hthen changed, changed by 24 hHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 4, row 'PDW', XN-9000, RT
whole-blood-edta, K3-EDTA (Becton Dickinson)StorageuncentrifugedRT8 hthen changed, changed by 24 hHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 4, row 'PDW', BC-6800, RT
whole-blood-edta, K3-EDTA (Becton Dickinson)StorageuncentrifugedRT8 hstable for the longest time testedHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 5, row 'PDW', XN-9000, RT
whole-blood-edta, K3-EDTA (Becton Dickinson)StorageuncentrifugedRT8 hstable for the longest time testedHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 5, row 'PDW', BC-6800, RT
Platelet Large Cell Ratio8 findings
Specimen, tubeConditionTemperatureStable forCriterionStudy
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C36 hthen changed, changed by 48 hHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 4, row 'P-LCR', XN-9000, 4 °C
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C8 hthen changed, changed by 24 hHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 4, row 'P-LCR', BC-6800, 4 °C
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C8 hstable for the longest time testedHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 5, row 'P-LCR', XN-9000, 4 °C
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C8 hstable for the longest time testedHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 5, row 'P-LCR', BC-6800, 4 °C
whole-blood-edta, K3-EDTA (Becton Dickinson)StorageuncentrifugedRT8 hthen changed, changed by 24 hHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 4, row 'P-LCR', XN-9000, RT
whole-blood-edta, K3-EDTA (Becton Dickinson)StorageuncentrifugedRT4 hthen changed, changed by 6 hHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 4, row 'P-LCR', BC-6800, RT
whole-blood-edta, K3-EDTA (Becton Dickinson)StorageuncentrifugedRT8 hstable for the longest time testedHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 5, row 'P-LCR', XN-9000, RT
whole-blood-edta, K3-EDTA (Becton Dickinson)StorageuncentrifugedRT8 hstable for the longest time testedHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 5, row 'P-LCR', BC-6800, RT
Plateletcrit8 findings
Specimen, tubeConditionTemperatureStable forCriterionStudy
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C48 hstable for the longest time testedHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 4, row 'PCT', XN-9000, 4 °C
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C48 hstable for the longest time testedHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 4, row 'PCT', BC-6800, 4 °C
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C8 hstable for the longest time testedHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 5, row 'PCT', XN-9000, 4 °C
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C8 hstable for the longest time testedHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 5, row 'PCT', BC-6800, 4 °C
whole-blood-edta, K3-EDTA (Becton Dickinson)StorageuncentrifugedRT48 hstable for the longest time testedHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 4, row 'PCT', XN-9000, RT
whole-blood-edta, K3-EDTA (Becton Dickinson)StorageuncentrifugedRT48 hstable for the longest time testedHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 4, row 'PCT', BC-6800, RT
whole-blood-edta, K3-EDTA (Becton Dickinson)StorageuncentrifugedRT8 hstable for the longest time testedHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 5, row 'PCT', XN-9000, RT
whole-blood-edta, K3-EDTA (Becton Dickinson)StorageuncentrifugedRT8 hstable for the longest time testedHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 5, row 'PCT', BC-6800, RT
Potassium25 findings
Specimen, tubeConditionTemperatureStable forCriterionStudy
plasma-li-heparin, Sarstedt lithium heparin, no gel separator (primary tube)Storagecentrifuged-on-cells2-8 °C4.7 hthen changed, changed by 6 hmean % difference vs time 0 exceeds total limit of change, TLC = ±√((1.65·CVa)² + (0.5·CVb)²) (EFLM BV), here 4.7%; stability limit from instability-equation regressionIllana et al. 2024, Biochem Med (Zagreb)Table 1, row 'K'
plasma-li-heparin, Sarstedt lithium heparin, no gel separator; plasma aliquotStorageseparated2-8 °C96 hstable for the longest time testedmean % difference vs time 0 exceeds total limit of change, TLC = ±√((1.65·CVa)² + (0.5·CVb)²) (EFLM BV), here 4.7%; stability limit from instability-equation regressionIllana et al. 2024, Biochem Med (Zagreb)Table 2, row 'K'
plasma-li-heparin, BD Vacutainer PST II lithium heparin 3 mL (gel)Storagecentrifuged-on-gel2-8 °C15 hstable for the longest time testedclinically significant when the 95% CI of mean bias lies completely outside the SEQC maximum allowable bias (MAB = 1.65 x CVa), here 4%Balbás et al. 2017, Biochem Med (Zagreb)Table 2, K+ row 'PST II sealed, 4 şC'
plasma-li-heparin, BD Barricor lithium heparin 5.5 mL (mechanical separator)Storagecentrifuged-on-gel2-8 °C15 hstable for the longest time testedclinically significant when the 95% CI of mean bias lies completely outside the SEQC maximum allowable bias (MAB = 1.65 x CVa), here 4%Balbás et al. 2017, Biochem Med (Zagreb)Table 2, K+ row 'Barricor sealed, 4 şC'
plasma-li-heparin, BD Vacutainer lithium heparin 6 mL (regular)Storagenot-stated2-8 °C4 hthen changed, changed by 24 hmean % difference vs T0 exceeds total change limit (TCL), here ±6.21%Dupuy et al. 2018, Biochem Med (Zagreb)Table 2, BD lithium heparin tubes, row 'K, mmol/L'
plasma-li-heparin, BD Barricor lithium heparin 5.5 mL (mechanical separator)Storagecentrifuged-on-gel2-8 °C72 hstable for the longest time testedmean % difference vs T0 exceeds total change limit (TCL), here ±6.21%Dupuy et al. 2018, Biochem Med (Zagreb)Table 2, BD Barricor lithium heparin tubes, row 'K, mmol/L'
plasma-li-heparin, BD Vacutainer LH PST II gel separator, 3.0 mL (REF 367374)Storagecentrifuged-on-gel2-8 °C24 hstable for the longest time testedmean % difference smaller than TEa (Ricos: 0.25 x (CVi^2 + CVg^2)^0.5 + 1.65 x 0.5 x CVi)Nonkes et al. 2019, Biochem Med (Zagreb)Table 1, row 'Potassium, mmol/L'
plasma-li-heparin, BD Vacutainer Barricor LH 3 mL, mechanical separatorStoragecentrifuged-on-gel2-8 °C40 hlimit stated by the authorsbias vs aliquoted plasma reference tube > maximal allowable bias 1.8% (Ricos et al.); time to MAB from linear regression of bias on time to analysisGawria et al. 2020, J Clin Lab AnalTable 2, row 'Potassium', Barricor BD
plasma-li-heparin, BD Vacutainer PST LH 3 mL, gelStoragecentrifuged-on-gel2-8 °C8.7 hlimit stated by the authorsbias vs aliquoted plasma reference tube > maximal allowable bias 1.8% (Ricos et al.); time to MAB from linear regression of bias on time to analysisGawria et al. 2020, J Clin Lab AnalTable 2, row 'Potassium', Gel BD
plasma-li-heparin, Greiner Vacuette LH 3 mL Sep, gelStoragecentrifuged-on-gel2-8 °C12 hlimit stated by the authorsbias vs aliquoted plasma reference tube > maximal allowable bias 1.8% (Ricos et al.); time to MAB from linear regression of bias on time to analysisGawria et al. 2020, J Clin Lab AnalTable 2, row 'Potassium', Gel Greiner
plasma-li-heparin, BD Vacutainer PST LH 3 mL, gelStoragecentrifuged-on-gel2-8 °C32 hlimit stated by the authorsbias vs aliquoted reference > 4.0% (higher quality goal used by other authors)Gawria et al. 2020, J Clin Lab AnalDiscussion, paragraph on potassium
plasma-li-heparin, Greiner Vacuette LH 3 mL Sep, gelStoragecentrifuged-on-gel2-8 °C34 hlimit stated by the authorsbias vs aliquoted reference > 4.0% (higher quality goal used by other authors)Gawria et al. 2020, J Clin Lab AnalDiscussion, paragraph on potassium
plasma-li-heparin, BD Vacutainer Barricor LH 3 mL, mechanical separatorStoragecentrifuged-on-gel2-8 °C80 hstable for the longest time testedbias vs aliquoted reference > 4.0% (higher quality goal used by other authors)Gawria et al. 2020, J Clin Lab AnalDiscussion, paragraph on potassium
plasma-li-heparin, BD Barricor 4.5 mL (lithium heparin, mechanical separator)Storagecentrifuged-on-gel2-8 °C24 hthen changed, changed by 7 daysmean % difference vs time 0 exceeds desirable imprecision based on biological variation (EFLM/Westgard), here 2.05%Mandić et al. 2020, Biochem Med (Zagreb)Tables 3 and 4, row 'Potassium, mmol/L', BD Barricor columns
plasma-li-heparin, BD Vacutainer PST II lithium heparin 3 mL (gel)Storagecentrifuged-on-gelRT6 hthen changed, changed by 9 hclinically significant when the 95% CI of mean bias lies completely outside the SEQC maximum allowable bias (MAB = 1.65 x CVa), here 4%Balbás et al. 2017, Biochem Med (Zagreb)Table 2, K+ row 'PST II open, rt'
plasma-li-heparin, BD Barricor lithium heparin 5.5 mL (mechanical separator)Storagecentrifuged-on-gelRT6 hthen changed, changed by 9 hclinically significant when the 95% CI of mean bias lies completely outside the SEQC maximum allowable bias (MAB = 1.65 x CVa), here 4%Balbás et al. 2017, Biochem Med (Zagreb)Table 2, K+ row 'Barricor open, rt'
serum, BD Vacutainer serum (plain), 7.5 mLStorageseparated−20 °C30 daysstable for the longest time testedclinically significant change: mean outside significant change limit (SCL = T1d mean ± 3.0 usual SD from 2 months QC)Kachhawa et al. 2017, J Lab PhysiciansResults paragraph 1 (Table 2 image-only)
serum, BD Vacutainer RST 4.0 mL (thrombin clot activator, gel)Storagecentrifuged-on-gel2-8 °C24 hstable for the longest time testedclinically significant if mean bias vs t0 exceeds desirable bias (biological variation), here 1.5%Kocijancic et al. 2014, Biochem Med (Zagreb)Table 3, row 'K'
serum, BD Vacutainer SST II Advance 5.0 mL (clot activator, gel)Storagecentrifuged-on-gel2-8 °C24 hstable for the longest time testedclinically significant if mean bias vs t0 exceeds desirable bias (biological variation), here 1.5%Kocijancic et al. 2014, Biochem Med (Zagreb)Table 4, row 'K'
serum, BD Vacutainer RST 5 mL (thrombin clot activator, gel)Storagecentrifuged-on-gel2-8 °C24 hthen changed, changed by 7 daysmean % difference vs time 0 exceeds desirable imprecision based on biological variation (EFLM/Westgard), here 2.05%Mandić et al. 2020, Biochem Med (Zagreb)Tables 3 and 4, row 'Potassium, mmol/L', BD RST columns
serum, BD Vacutainer SST (SKU 367986), gelStoragecentrifuged-on-gel30 °C (±1 °C)72 hstable for the longest time testedmean bias vs paired baseline within CLIA acceptance limits AND >2/3 of individual results within limitsIacovetti et al. 2026, Pract Lab MedResults, 'Analyte stability after storage at 30 °C', paragraph 1 (Table 1 image-only)
serum, BD SST Microtainer (SKU 395967), gel; filled from venous bloodStoragecentrifuged-on-gel0 °C0 hchanged at the first time point, changed by 24 hmean bias vs paired baseline within CLIA acceptance limits AND >2/3 of individual results within limitsIacovetti et al. 2026, Pract Lab MedResults, 'Analyte stability after separation and storage in microvolume tubes at 0 °C' (Table 2 image-only)
whole-blood-clotted, BD Vacutainer SST (SKU 367986), gelBefore spinninguncentrifuged30 °C (±1 °C)0 hchanged at the first time point, changed by 24 hmean bias vs paired baseline within CLIA acceptance limits AND >2/3 of individual results within limitsIacovetti et al. 2026, Pract Lab MedResults, 'Analyte stability after storage at 30 °C', paragraph 2 (Table 1 image-only)
whole-blood-heparin, BD Vacutainer lithium heparin, no gel (ref 368884)Before spinninguncentrifugedRT10 hstable for the longest time tested|mean % deviation 10 h vs baseline| > laboratory bias goal (Table II; mostly biological-variation based, some empirically modified)Henriksen et al. 2014, Scand J Clin Lab InvestTable II, row 'K'
whole-blood-heparin, BD Vacutainer lithium heparin, no gel (ref 368884)Before spinninguncentrifugedRT0 hchanged at the first time point, changed by 10 hCV of paired baseline/10 h results > goal CV (Table II)Henriksen et al. 2014, Scand J Clin Lab InvestTable II, row 'K'
Prostate-Specific Antigen, Total (PSA)2 findings
Specimen, tubeConditionTemperatureStable forCriterionStudy
whole-blood-clotted, BD Vacutainer serum, no gel (ref 368815)Before spinninguncentrifugedRT10 hstable for the longest time tested|mean % deviation 10 h vs baseline| > laboratory bias goal (Table II; mostly biological-variation based, some empirically modified)Henriksen et al. 2014, Scand J Clin Lab InvestTable II, row 'PSA'
whole-blood-clotted, BD Vacutainer serum, no gel (ref 368815)Before spinninguncentrifugedRT10 hstable for the longest time testedCV of paired baseline/10 h results > goal CV (Table II)Henriksen et al. 2014, Scand J Clin Lab InvestTable II, row 'PSA'
Protein C5 findings
Specimen, tubeConditionTemperatureStable forCriterionStudy
plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquotsFreeze–thawseparated−20 °C1 cyclelimit stated by the authorsmax allowable mean deviation from t0: 4.00–8.00%: ±0.800% (absolute);>8.00%: ±10.0% (relative)Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability'
plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquotsFreeze–thawseparated−80 °C3 cycleslimit stated by the authorsmax allowable mean deviation from t0: 4.00–8.00%: ±0.800% (absolute);>8.00%: ±10.0% (relative)Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability'
plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquotsStorageseparated−20 °C28 dayslimit stated by the authorsmax allowable mean deviation from t0: 4.00–8.00%: ±0.800% (absolute);>8.00%: ±10.0% (relative)Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability'
plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquotsStorageseparated−80 °C10 monthslimit stated by the authorsmax allowable mean deviation from t0: 4.00–8.00%: ±0.800% (absolute);>8.00%: ±10.0% (relative)Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability'
plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquotsStorageseparatedRT4 hlimit stated by the authorsmax allowable mean deviation from t0: 4.00–8.00%: ±0.800% (absolute);>8.00%: ±10.0% (relative)Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability'
Protein S5 findings
Specimen, tubeConditionTemperatureStable forCriterionStudy
plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquotsFreeze–thawseparated−20 °C1 cyclelimit stated by the authorsmax allowable mean deviation from t0: <20.0%: ≤±2.00% (absolute);≥20.0%: ≤±10.0% (relative)Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability'
plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquotsFreeze–thawseparated−80 °C3 cycleslimit stated by the authorsmax allowable mean deviation from t0: <20.0%: ≤±2.00% (absolute);≥20.0%: ≤±10.0% (relative)Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability'
plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquotsStorageseparated−20 °C12 weekslimit stated by the authorsmax allowable mean deviation from t0: <20.0%: ≤±2.00% (absolute);≥20.0%: ≤±10.0% (relative)Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability'
plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquotsStorageseparated−80 °C24 monthslimit stated by the authorsmax allowable mean deviation from t0: <20.0%: ≤±2.00% (absolute);≥20.0%: ≤±10.0% (relative)Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability'
plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquotsStorageseparatedRT24 hlimit stated by the authorsmax allowable mean deviation from t0: <20.0%: ≤±2.00% (absolute);≥20.0%: ≤±10.0% (relative)Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability'
Prothrombin Time15 findings
Specimen, tubeConditionTemperatureStable forCriterionStudy
plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquotsFreeze–thawseparated−20 °C1 cyclelimit stated by the authorsmax allowable mean deviation from t0: ±10%Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability'
plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquotsFreeze–thawseparated−20 °C1 cyclelimit stated by the authorsmax allowable mean deviation from t0: ≤4.5 INR: 100% ± 10%;>4.5 INR: 100% ± 15%Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability'
plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquotsFreeze–thawseparated−80 °C1 cyclelimit stated by the authorsmax allowable mean deviation from t0: ±10%Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability'
plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquotsFreeze–thawseparated−80 °C1 cyclelimit stated by the authorsmax allowable mean deviation from t0: ≤4.5 INR: 100% ± 10%;>4.5 INR: 100% ± 15%Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability'
plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquotsStorageseparated−20 °C7 weekslimit stated by the authorsmax allowable mean deviation from t0: ±10%Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability'
plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquotsStorageseparated−20 °C6 weekslimit stated by the authorsmax allowable mean deviation from t0: ≤4.5 INR: 100% ± 10%;>4.5 INR: 100% ± 15%Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability'
plasma-citrate, BD Vacutainer 2.7 mL, 3.2% buffered sodium citrate; PPP aliquoted into microcentrifuge tubesStorageseparated−20 °C36 hstable for the longest time testedchange of more than ±10% of mean from baselinePatil et al. 2022, CureusTable 2
plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquotsStorageseparated−80 °C11 monthslimit stated by the authorsmax allowable mean deviation from t0: ±10%Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability'
plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquotsStorageseparated−80 °C3 monthslimit stated by the authorsmax allowable mean deviation from t0: ≤4.5 INR: 100% ± 10%;>4.5 INR: 100% ± 15%Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability'
plasma-citrate, BD 0.109 M sodium citrate (9:1); plasma aliquoted into nonactivating Eppendorf tubesStorageseparated2-8 °C24 hstable for the longest time testedmean % change from baseline >10%, or >25% of samples with individual change >10% (van Geest-Daalderop)Feng et al. 2014, Sci RepTable 2, 'Acceptable Time (h)'
plasma-citrate, BD Vacutainer 2.7 mL, 3.2% buffered sodium citrate; PPP aliquoted into microcentrifuge tubesStorageseparated2-8 °C24 hthen changed, changed by 36 hchange of more than ±10% of mean from baselinePatil et al. 2022, CureusTable 2
plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquotsStorageseparatedRT12 hlimit stated by the authorsmax allowable mean deviation from t0: ±10%Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability'
plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquotsStorageseparatedRT24 hlimit stated by the authorsmax allowable mean deviation from t0: ≤4.5 INR: 100% ± 10%;>4.5 INR: 100% ± 15%Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability'
plasma-citrate, BD 0.109 M sodium citrate (9:1); plasma aliquoted into nonactivating Eppendorf tubesStorageseparatedRT24 hstable for the longest time testedmean % change from baseline >10%, or >25% of samples with individual change >10% (van Geest-Daalderop)Feng et al. 2014, Sci RepTable 2, 'Acceptable Time (h)'
plasma-citrate, BD Vacutainer 2.7 mL, 3.2% buffered sodium citrate; PPP aliquoted into microcentrifuge tubesStorageseparatedRT0 hchanged at the first time point, changed by 12 hchange of more than ±10% of mean from baselinePatil et al. 2022, CureusTable 2
Red Blood Cell Count21 findings
Specimen, tubeConditionTemperatureStable forCriterionStudy
whole-blood-edta, BD Vacutainer K2EDTA, 2 mL (spray-coated)Storageuncentrifuged2-8 °C24 hstable for the longest time testedMean % change from baseline within analytical measurement uncertainty (MU ±1.5%)Al-Najdawi et al. 2025, Med Pharm RepTable II, row 'RBC'; Recommendations
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C48 hstable for the longest time testedHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'RBC', XN-9000, 4 °C
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C48 hstable for the longest time testedBias% vs optimal bias (OP-Bias%, from biological variation; Table 1)Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'RBC', XN-9000, 4 °C
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C48 hstable for the longest time testedHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'RBC', BC-6800, 4 °C
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C24 hthen changed, changed by 36 hBias% vs optimal bias (OP-Bias%, from biological variation; Table 1)Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'RBC', BC-6800, 4 °C
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C8 hstable for the longest time testedHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'RBC', XN-9000, 4 °C
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C8 hstable for the longest time testedBias% vs optimal bias (OP-Bias%, from biological variation; Table 1)Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'RBC', XN-9000, 4 °C
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C8 hstable for the longest time testedHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'RBC', BC-6800, 4 °C
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C8 hstable for the longest time testedBias% vs optimal bias (OP-Bias%, from biological variation; Table 1)Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'RBC', BC-6800, 4 °C
whole-blood-edta, K2-EDTA (BD Vacutainer)Storageuncentrifuged2-8 °C48 hstable for the longest time testedRepeated-measures one-way ANOVA with Tukey post hoc vs 0 h, p < 0.05 (* in table)Unalli et al. 2021, J Med BiochemResults paragraphs 1-2 and Discussion paragraph 1 (Figures 1-2; numbers in figures only)
whole-blood-edta, K3-EDTA (Becton Dickinson)StorageuncentrifugedRT48 hstable for the longest time testedHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'RBC', XN-9000, RT
whole-blood-edta, K3-EDTA (Becton Dickinson)StorageuncentrifugedRT48 hstable for the longest time testedBias% vs optimal bias (OP-Bias%, from biological variation; Table 1)Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'RBC', XN-9000, RT
whole-blood-edta, K3-EDTA (Becton Dickinson)StorageuncentrifugedRT48 hstable for the longest time testedHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'RBC', BC-6800, RT
whole-blood-edta, K3-EDTA (Becton Dickinson)StorageuncentrifugedRT48 hstable for the longest time testedBias% vs optimal bias (OP-Bias%, from biological variation; Table 1)Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'RBC', BC-6800, RT
whole-blood-edta, K3-EDTA (Becton Dickinson)StorageuncentrifugedRT8 hstable for the longest time testedHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'RBC', XN-9000, RT
whole-blood-edta, K3-EDTA (Becton Dickinson)StorageuncentrifugedRT8 hstable for the longest time testedBias% vs optimal bias (OP-Bias%, from biological variation; Table 1)Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'RBC', XN-9000, RT
whole-blood-edta, K3-EDTA (Becton Dickinson)StorageuncentrifugedRT8 hstable for the longest time testedHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'RBC', BC-6800, RT
whole-blood-edta, K3-EDTA (Becton Dickinson)StorageuncentrifugedRT8 hstable for the longest time testedBias% vs optimal bias (OP-Bias%, from biological variation; Table 1)Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'RBC', BC-6800, RT
whole-blood-edta, K2-EDTA (BD Vacutainer)StorageuncentrifugedRT48 hstable for the longest time testedRepeated-measures one-way ANOVA with Tukey post hoc vs 0 h, p < 0.05 (* in table)Unalli et al. 2021, J Med BiochemResults paragraphs 1-2 and Discussion paragraph 1 (Figures 1-2; numbers in figures only)
whole-blood-edta, K2EDTA (Becton Dickinson)Storageuncentrifuged33 °C72 hstable for the longest time testedUnacceptable if median % difference > Biasmax (biological variation) and Holm-adjusted Wilcoxon P < 0.05; stable if equivalence shown (two one-tailed Wilcoxon tests vs Biasmax, Pholm < 0.05)Jain et al. 2018, Biochem Med (Zagreb)Results, last paragraph (equivalence); Table 2, row 'RBC'
whole-blood-edta, K2-EDTA (BD Vacutainer)Storageuncentrifuged10 °C48 hstable for the longest time testedRepeated-measures one-way ANOVA with Tukey post hoc vs 0 h, p < 0.05 (* in table)Unalli et al. 2021, J Med BiochemResults paragraphs 1-2 and Discussion paragraph 1 (Figures 1-2; numbers in figures only)
Red cell distribution width (CV or SD not stated)4 findings
Specimen, tubeConditionTemperatureStable forCriterionStudy
whole-blood-edta, K2-EDTA (BD Vacutainer)Storageuncentrifuged2-8 °C48 hstable for the longest time testedRepeated-measures one-way ANOVA with Tukey post hoc vs 0 h, p < 0.05 (* in table)Unalli et al. 2021, J Med BiochemTable 2, column '4 °C' (no * at any time point)
whole-blood-edta, K2-EDTA (BD Vacutainer)StorageuncentrifugedRT8 hthen changed, changed by 12 hRepeated-measures one-way ANOVA with Tukey post hoc vs 0 h, p < 0.05 (* in table)Unalli et al. 2021, J Med BiochemTable 2, block 'RDW', row '12 h', column '23 °C'
whole-blood-edta, K2EDTA (Becton Dickinson)Storageuncentrifuged33 °C0 hchanged at the first time point, changed by 4 hUnacceptable if median % difference > Biasmax (biological variation) and Holm-adjusted Wilcoxon P < 0.05; stable if equivalence shown (two one-tailed Wilcoxon tests vs Biasmax, Pholm < 0.05)Jain et al. 2018, Biochem Med (Zagreb)Results paragraph 1; Table 2, row 'RDW'
whole-blood-edta, K2-EDTA (BD Vacutainer)Storageuncentrifuged10 °C48 hstable for the longest time testedRepeated-measures one-way ANOVA with Tukey post hoc vs 0 h, p < 0.05 (* in table)Unalli et al. 2021, J Med BiochemTable 2, column '10 °C' (no * at any time point)
Red Cell Distribution Width (CV)21 findings
Specimen, tubeConditionTemperatureStable forCriterionStudy
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C8 hthen changed, changed by 24 hHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'RDW-CV', XN-9000, 4 °C
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C8 hthen changed, changed by 24 hBias% vs optimal bias (OP-Bias%, from biological variation; Table 1)Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'RDW-CV', XN-9000, 4 °C
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C48 hstable for the longest time testedBias% vs critical difference (CD%, reference change value; Table 1)Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'RDW-CV', XN-9000, 4 °C
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C48 hstable for the longest time testedHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'RDW-CV', BC-6800, 4 °C
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C48 hstable for the longest time testedBias% vs optimal bias (OP-Bias%, from biological variation; Table 1)Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'RDW-CV', BC-6800, 4 °C
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C8 hstable for the longest time testedHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'RDW-CV', XN-9000, 4 °C
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C8 hstable for the longest time testedBias% vs optimal bias (OP-Bias%, from biological variation; Table 1)Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'RDW-CV', XN-9000, 4 °C
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C8 hstable for the longest time testedHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'RDW-CV', BC-6800, 4 °C
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C4 hthen changed, changed by 8 hBias% vs optimal bias (OP-Bias%, from biological variation; Table 1)Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'RDW-CV', BC-6800, 4 °C
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C8 hstable for the longest time testedBias% vs critical difference (CD%, reference change value; Table 1)Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'RDW-CV', BC-6800, 4 °C
whole-blood-edta, K3-EDTA (Becton Dickinson)StorageuncentrifugedRT8 hthen changed, changed by 24 hHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'RDW-CV', XN-9000, RT
whole-blood-edta, K3-EDTA (Becton Dickinson)StorageuncentrifugedRT8 hthen changed, changed by 24 hBias% vs optimal bias (OP-Bias%, from biological variation; Table 1)Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'RDW-CV', XN-9000, RT
whole-blood-edta, K3-EDTA (Becton Dickinson)StorageuncentrifugedRT48 hstable for the longest time testedBias% vs critical difference (CD%, reference change value; Table 1)Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'RDW-CV', XN-9000, RT
whole-blood-edta, K3-EDTA (Becton Dickinson)StorageuncentrifugedRT8 hthen changed, changed by 24 hHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'RDW-CV', BC-6800, RT
whole-blood-edta, K3-EDTA (Becton Dickinson)StorageuncentrifugedRT2 hthen changed, changed by 4 hBias% vs optimal bias (OP-Bias%, from biological variation; Table 1)Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'RDW-CV', BC-6800, RT
whole-blood-edta, K3-EDTA (Becton Dickinson)StorageuncentrifugedRT8 hthen changed, changed by 24 hBias% vs critical difference (CD%, reference change value; Table 1)Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'RDW-CV', BC-6800, RT
whole-blood-edta, K3-EDTA (Becton Dickinson)StorageuncentrifugedRT8 hstable for the longest time testedHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'RDW-CV', XN-9000, RT
whole-blood-edta, K3-EDTA (Becton Dickinson)StorageuncentrifugedRT8 hstable for the longest time testedBias% vs optimal bias (OP-Bias%, from biological variation; Table 1)Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'RDW-CV', XN-9000, RT
whole-blood-edta, K3-EDTA (Becton Dickinson)StorageuncentrifugedRT8 hstable for the longest time testedHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'RDW-CV', BC-6800, RT
whole-blood-edta, K3-EDTA (Becton Dickinson)StorageuncentrifugedRT0 hchanged at the first time point, changed by 4 hBias% vs optimal bias (OP-Bias%, from biological variation; Table 1)Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'RDW-CV', BC-6800, RT
whole-blood-edta, K3-EDTA (Becton Dickinson)StorageuncentrifugedRT8 hstable for the longest time testedBias% vs critical difference (CD%, reference change value; Table 1)Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'RDW-CV', BC-6800, RT
Red Cell Distribution Width (SD)8 findings
Specimen, tubeConditionTemperatureStable forCriterionStudy
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C48 hstable for the longest time testedHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'RDW-SD', XN-9000, 4 °C
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C48 hstable for the longest time testedHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'RDW-SD', BC-6800, 4 °C
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C8 hstable for the longest time testedHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'RDW-SD', XN-9000, 4 °C
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C8 hstable for the longest time testedHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'RDW-SD', BC-6800, 4 °C
whole-blood-edta, K3-EDTA (Becton Dickinson)StorageuncentrifugedRT8 hthen changed, changed by 24 hHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'RDW-SD', XN-9000, RT
whole-blood-edta, K3-EDTA (Becton Dickinson)StorageuncentrifugedRT8 hthen changed, changed by 24 hHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'RDW-SD', BC-6800, RT
whole-blood-edta, K3-EDTA (Becton Dickinson)StorageuncentrifugedRT8 hstable for the longest time testedHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'RDW-SD', XN-9000, RT
whole-blood-edta, K3-EDTA (Becton Dickinson)StorageuncentrifugedRT4 hthen changed, changed by 8 hHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'RDW-SD', BC-6800, RT
Reticulocyte Count (Absolute)18 findings
Specimen, tubeConditionTemperatureStable forCriterionStudy
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C48 hstable for the longest time testedHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'RET', XN-9000, 4 °C
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C48 hstable for the longest time testedBias% vs optimal bias (OP-Bias%, from biological variation; Table 1)Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'RET', XN-9000, 4 °C
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C48 hstable for the longest time testedHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'RET', BC-6800, 4 °C
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C36 hthen changed, changed by 48 hBias% vs optimal bias (OP-Bias%, from biological variation; Table 1)Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'RET', BC-6800, 4 °C
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C48 hstable for the longest time testedBias% vs critical difference (CD%, reference change value; Table 1)Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'RET', BC-6800, 4 °C
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C8 hstable for the longest time testedHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'RET', XN-9000, 4 °C
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C8 hstable for the longest time testedBias% vs optimal bias (OP-Bias%, from biological variation; Table 1)Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'RET', XN-9000, 4 °C
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C8 hstable for the longest time testedHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'RET', BC-6800, 4 °C
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C8 hstable for the longest time testedBias% vs optimal bias (OP-Bias%, from biological variation; Table 1)Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'RET', BC-6800, 4 °C
whole-blood-edta, K3-EDTA (Becton Dickinson)StorageuncentrifugedRT48 hstable for the longest time testedHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'RET', XN-9000, RT
whole-blood-edta, K3-EDTA (Becton Dickinson)StorageuncentrifugedRT48 hstable for the longest time testedBias% vs optimal bias (OP-Bias%, from biological variation; Table 1)Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'RET', XN-9000, RT
whole-blood-edta, K3-EDTA (Becton Dickinson)StorageuncentrifugedRT48 hstable for the longest time testedHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'RET', BC-6800, RT
whole-blood-edta, K3-EDTA (Becton Dickinson)StorageuncentrifugedRT36 hthen changed, changed by 48 hBias% vs optimal bias (OP-Bias%, from biological variation; Table 1)Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'RET', BC-6800, RT
whole-blood-edta, K3-EDTA (Becton Dickinson)StorageuncentrifugedRT48 hstable for the longest time testedBias% vs critical difference (CD%, reference change value; Table 1)Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'RET', BC-6800, RT
whole-blood-edta, K3-EDTA (Becton Dickinson)StorageuncentrifugedRT8 hstable for the longest time testedHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'RET', XN-9000, RT
whole-blood-edta, K3-EDTA (Becton Dickinson)StorageuncentrifugedRT8 hstable for the longest time testedBias% vs optimal bias (OP-Bias%, from biological variation; Table 1)Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'RET', XN-9000, RT
whole-blood-edta, K3-EDTA (Becton Dickinson)StorageuncentrifugedRT8 hstable for the longest time testedHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'RET', BC-6800, RT
whole-blood-edta, K3-EDTA (Becton Dickinson)StorageuncentrifugedRT8 hstable for the longest time testedBias% vs optimal bias (OP-Bias%, from biological variation; Table 1)Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'RET', BC-6800, RT
ROTEM EXTEM clot formation time2 findings
Specimen, tubeConditionTemperatureStable forCriterionStudy
whole-blood-citrate, Greiner Vacuette sodium citrate, standard fill (blood–air interface)StorageuncentrifugedRT0 hchanged at the first time point, changed by 2 hmaximum acceptable bias = 0.375 × total biological variation (EFLM BV database, Model 2)Storm et al. 2026, Adv HematolTable 1, 'EXTEM, standard tubes' row 'CFT (s)'
whole-blood-citrate, Greiner Vacuette sodium citrate, completely filled (no air)StorageuncentrifugedRT2 hthen changed, changed by 4 hmaximum acceptable bias = 0.375 × total biological variation (EFLM BV database, Model 2)Storm et al. 2026, Adv HematolTable 1, 'EXTEM, full tubes' row 'CFT (s)'
ROTEM EXTEM clotting time2 findings
Specimen, tubeConditionTemperatureStable forCriterionStudy
whole-blood-citrate, Greiner Vacuette sodium citrate, standard fill (blood–air interface)StorageuncentrifugedRT24 hthen changed, changed by 48 hmaximum acceptable bias = 0.375 × total biological variation (EFLM BV database, Model 2)Storm et al. 2026, Adv HematolTable 1, 'EXTEM, standard tubes' row 'CT (s)'
whole-blood-citrate, Greiner Vacuette sodium citrate, completely filled (no air)StorageuncentrifugedRT4 hthen changed, changed by 6 hmaximum acceptable bias = 0.375 × total biological variation (EFLM BV database, Model 2)Storm et al. 2026, Adv HematolTable 1, 'EXTEM, full tubes' row 'CT (s)'
ROTEM EXTEM maximum clot firmness2 findings
Specimen, tubeConditionTemperatureStable forCriterionStudy
whole-blood-citrate, Greiner Vacuette sodium citrate, standard fill (blood–air interface)StorageuncentrifugedRT6 hthen changed, changed by 24 hmaximum acceptable bias = 0.375 × total biological variation (EFLM BV database, Model 2)Storm et al. 2026, Adv HematolTable 1, 'EXTEM, standard tubes' row 'MCF (millimeters)'
whole-blood-citrate, Greiner Vacuette sodium citrate, completely filled (no air)StorageuncentrifugedRT2 hthen changed, changed by 4 hmaximum acceptable bias = 0.375 × total biological variation (EFLM BV database, Model 2)Storm et al. 2026, Adv HematolTable 1, 'EXTEM, full tubes' row 'MCF (millimeters)'
ROTEM INTEM clot formation time2 findings
Specimen, tubeConditionTemperatureStable forCriterionStudy
whole-blood-citrate, Greiner Vacuette sodium citrate, standard fill (blood–air interface)StorageuncentrifugedRT48 hstable for the longest time testedmaximum acceptable bias = 0.375 × total biological variation (EFLM BV database, Model 2)Storm et al. 2026, Adv HematolTable 1, 'INTEM, standard tubes' row 'CFT (s)'
whole-blood-citrate, Greiner Vacuette sodium citrate, completely filled (no air)StorageuncentrifugedRT24 hthen changed, changed by 48 hmaximum acceptable bias = 0.375 × total biological variation (EFLM BV database, Model 2)Storm et al. 2026, Adv HematolTable 1, 'INTEM, full tubes' row 'CFT (s)'
ROTEM INTEM clotting time2 findings
Specimen, tubeConditionTemperatureStable forCriterionStudy
whole-blood-citrate, Greiner Vacuette sodium citrate, standard fill (blood–air interface)StorageuncentrifugedRT48 hstable for the longest time testedmaximum acceptable bias = 0.375 × total biological variation (EFLM BV database, Model 2)Storm et al. 2026, Adv HematolTable 1, 'INTEM, standard tubes' row 'CT (s)'
whole-blood-citrate, Greiner Vacuette sodium citrate, completely filled (no air)StorageuncentrifugedRT48 hstable for the longest time testedmaximum acceptable bias = 0.375 × total biological variation (EFLM BV database, Model 2)Storm et al. 2026, Adv HematolTable 1, 'INTEM, full tubes' row 'CT (s)'
Selenium4 findings
Specimen, tubeConditionTemperatureStable forCriterionStudy
whole-blood-clotted, BD Vacutainer trace element-free serum tubeBefore spinninguncentrifugedcold packs in insulated box; 9 (2) °C on arrival24 hstable for the longest time tested% difference within desirable quality specification (from intra- and interindividual variation)Jones et al. 2024, J NutrTable 2, row 'Selenium, μg/L'; Table 1
whole-blood-clotted, Sarstedt Monovette serumBefore spinninguncentrifugedchilled in transit (insulated box, 2 frozen cool packs; temperature not recorded)24 hstable for the longest time tested% difference within desirable quality specification for bias (Fraser, from intra- and interindividual variation)Jones et al. 2021, J NutrTable 2, row 'Selenium', column Serum
whole-blood-edta, Sarstedt Monovette EDTABefore spinninguncentrifugedchilled in transit (insulated box, 2 frozen cool packs; temperature not recorded)24 hstable for the longest time tested% difference within desirable quality specification for bias (Fraser, from intra- and interindividual variation)Jones et al. 2021, J NutrTable 2, row 'Selenium', column EDTA
whole-blood-heparin, Sarstedt Monovette lithium heparin, trace-element freeBefore spinninguncentrifugedchilled in transit (insulated box, 2 frozen cool packs; temperature not recorded)24 hstable for the longest time tested% difference within desirable quality specification for bias (Fraser, from intra- and interindividual variation)Jones et al. 2021, J NutrTable 2, row 'Selenium', column LH
Sodium16 findings
Specimen, tubeConditionTemperatureStable forCriterionStudy
plasma-li-heparin, Sarstedt lithium heparin, no gel separator (primary tube)Storagecentrifuged-on-cells2-8 °C48 hstable for the longest time testedmean % difference vs time 0 exceeds total limit of change, TLC = ±√((1.65·CVa)² + (0.5·CVb)²) (EFLM BV), here 3.0%; stability limit from instability-equation regressionIllana et al. 2024, Biochem Med (Zagreb)Table 1, row 'Na'
plasma-li-heparin, Sarstedt lithium heparin, no gel separator; plasma aliquotStorageseparated2-8 °C96 hstable for the longest time testedmean % difference vs time 0 exceeds total limit of change, TLC = ±√((1.65·CVa)² + (0.5·CVb)²) (EFLM BV), here 3.0%; stability limit from instability-equation regressionIllana et al. 2024, Biochem Med (Zagreb)Table 2, row 'Na'
plasma-li-heparin, BD Vacutainer PST II lithium heparin 3 mL (gel)Storagecentrifuged-on-gel2-8 °C15 hstable for the longest time testedclinically significant when the 95% CI of mean bias lies completely outside the SEQC maximum allowable bias (MAB = 1.65 x CVa), here 0.6%Balbás et al. 2017, Biochem Med (Zagreb)Table 2, Na+ row 'PST II sealed, 4 şC'
plasma-li-heparin, BD Barricor lithium heparin 5.5 mL (mechanical separator)Storagecentrifuged-on-gel2-8 °C12 hthen changed, changed by 15 hclinically significant when the 95% CI of mean bias lies completely outside the SEQC maximum allowable bias (MAB = 1.65 x CVa), here 0.6%Balbás et al. 2017, Biochem Med (Zagreb)Table 2, Na+ row 'Barricor sealed, 4 şC'
plasma-li-heparin, BD Vacutainer lithium heparin 6 mL (regular)Storagenot-stated2-8 °C72 hstable for the longest time testedmean % difference vs T0 exceeds total change limit (TCL), here ±3.64%Dupuy et al. 2018, Biochem Med (Zagreb)Table 2, BD lithium heparin tubes, row 'Na, mmol/L'
plasma-li-heparin, BD Barricor lithium heparin 5.5 mL (mechanical separator)Storagecentrifuged-on-gel2-8 °C72 hstable for the longest time testedmean % difference vs T0 exceeds total change limit (TCL), here ±3.64%Dupuy et al. 2018, Biochem Med (Zagreb)Table 2, BD Barricor lithium heparin tubes, row 'Na, mmol/L'
plasma-li-heparin, BD Barricor 4.5 mL (lithium heparin, mechanical separator)Storagecentrifuged-on-gel2-8 °C24 hthen changed, changed by 7 daysmean % difference vs time 0 exceeds desirable imprecision based on biological variation (EFLM/Westgard), here 0.25%Mandić et al. 2020, Biochem Med (Zagreb)Tables 3 and 4, row 'Sodium, mmol/L', BD Barricor columns
plasma-li-heparin, BD Vacutainer PST II lithium heparin 3 mL (gel)Storagecentrifuged-on-gelRT0 hchanged at the first time point, changed by 3 hclinically significant when the 95% CI of mean bias lies completely outside the SEQC maximum allowable bias (MAB = 1.65 x CVa), here 0.6%Balbás et al. 2017, Biochem Med (Zagreb)Table 2, Na+ row 'PST II open, rt'
plasma-li-heparin, BD Barricor lithium heparin 5.5 mL (mechanical separator)Storagecentrifuged-on-gelRT0 hchanged at the first time point, changed by 3 hclinically significant when the 95% CI of mean bias lies completely outside the SEQC maximum allowable bias (MAB = 1.65 x CVa), here 0.6%Balbás et al. 2017, Biochem Med (Zagreb)Table 2, Na+ row 'Barricor open, rt'
serum, BD Vacutainer serum (plain), 7.5 mLStorageseparated−20 °C30 daysstable for the longest time testedclinically significant change: mean outside significant change limit (SCL = T1d mean ± 3.0 usual SD from 2 months QC)Kachhawa et al. 2017, J Lab PhysiciansResults paragraph 1; Discussion 'Metabolites' (Table 2 image-only)
serum, BD Vacutainer RST 5 mL (thrombin clot activator, gel)Storagecentrifuged-on-gel2-8 °C24 hthen changed, changed by 7 daysmean % difference vs time 0 exceeds desirable imprecision based on biological variation (EFLM/Westgard), here 0.25%Mandić et al. 2020, Biochem Med (Zagreb)Tables 3 and 4, row 'Sodium, mmol/L', BD RST columns
serum, BD Vacutainer SST (SKU 367986), gelStoragecentrifuged-on-gel30 °C (±1 °C)24 hthen changed, changed by 72 hmean bias vs paired baseline within CLIA acceptance limits AND >2/3 of individual results within limitsIacovetti et al. 2026, Pract Lab MedResults, 'Analyte stability after storage at 30 °C', paragraph 1 (Table 1 image-only)
serum, BD SST Microtainer (SKU 395967), gel; filled from venous bloodStoragecentrifuged-on-gel0 °C72 hstable for the longest time testedmean bias vs paired baseline within CLIA acceptance limits AND >2/3 of individual results within limitsIacovetti et al. 2026, Pract Lab MedResults, 'Analyte stability after separation and storage in microvolume tubes at 0 °C' (Table 2 image-only)
whole-blood-clotted, BD Vacutainer SST (SKU 367986), gelBefore spinninguncentrifuged30 °C (±1 °C)24 hthen changed, changed by 72 hmean bias vs paired baseline within CLIA acceptance limits AND >2/3 of individual results within limitsIacovetti et al. 2026, Pract Lab MedResults, 'Analyte stability after storage at 30 °C', paragraph 2 (Table 1 image-only)
whole-blood-heparin, BD Vacutainer lithium heparin, no gel (ref 368884)Before spinninguncentrifugedRT10 hstable for the longest time tested|mean % deviation 10 h vs baseline| > laboratory bias goal (Table II; mostly biological-variation based, some empirically modified)Henriksen et al. 2014, Scand J Clin Lab InvestTable II, row 'Na'
whole-blood-heparin, BD Vacutainer lithium heparin, no gel (ref 368884)Before spinninguncentrifugedRT10 hstable for the longest time testedCV of paired baseline/10 h results > goal CV (Table II)Henriksen et al. 2014, Scand J Clin Lab InvestTable II, row 'Na'
Soluble P-selectin2 findings
Specimen, tubeConditionTemperatureStable forCriterionStudy
whole-blood-edta, Greiner Vacuette K2E EDTA, standard fillBefore spinninguncentrifugedRT2 hthen changed, changed by 4 hmaximum acceptable bias = 0.375 × total biological variation (EFLM BV database, Model 2)Storm et al. 2026, Adv HematolTable 2, 'Standard tubes' row 'sP-selectin'
whole-blood-edta, Greiner Vacuette K2E EDTA, completely filled (no air)Before spinninguncentrifugedRT0 hchanged at the first time point, changed by 2 hmaximum acceptable bias = 0.375 × total biological variation (EFLM BV database, Model 2)Storm et al. 2026, Adv HematolTable 2, 'Full tubes' row 'sP-selectin'
Soluble Transferrin Receptor3 findings
Specimen, tubeConditionTemperatureStable forCriterionStudy
whole-blood-clotted, BD Vacutainer SSTBefore spinninguncentrifugedRT24 hstable for the longest time testedmean % deviation from baseline within total change limit TCL = sqrt((2.77 CVa)^2 + (0.5 CVb)^2), CVb from Ricos (Table 2)Abraham et al. 2019, Biochem Med (Zagreb)Discussion, paragraph 1 (Results paragraphs 1-3; Figure 3)
whole-blood-clotted, BD Vacutainer SSTBefore spinninguncentrifuged22-30 °C (logger 24-26 °C)24 hstable for the longest time testedmean % deviation from baseline within total change limit TCL = sqrt((2.77 CVa)^2 + (0.5 CVb)^2), CVb from Ricos (Table 2)Abraham et al. 2019, Biochem Med (Zagreb)Discussion, paragraph 1 (Results paragraphs 1-3; Figure 3)
whole-blood-clotted, BD Vacutainer SSTBefore spinninguncentrifuged> 30 °C (logger 38-40 °C)24 hstable for the longest time testedmean % deviation from baseline within total change limit TCL = sqrt((2.77 CVa)^2 + (0.5 CVb)^2), CVb from Ricos (Table 2)Abraham et al. 2019, Biochem Med (Zagreb)Discussion, paragraph 1 (Results paragraphs 1-3; Figure 3)
Testosterone, Total2 findings
Specimen, tubeConditionTemperatureStable forCriterionStudy
dbs, PerkinElmer 226 Spot Saver RUO cardStoragedried2-8 °C30 daysstable for the longest time testedauthors' verdict vs baseline; no numeric stability limit statedGruenstein 2024, Anal Sci AdvResults i (Stability of the DBS); Table 4, row 'Refrigerated'; Table 5
dbs, PerkinElmer 226 Spot Saver RUO cardStoragedriedRT30 daysstable for the longest time testedauthors' verdict vs baseline; no numeric stability limit statedGruenstein 2024, Anal Sci AdvResults i (Stability of the DBS); Table 4, row 'Room temperature'; Table 5
Thrombin Time3 findings
Specimen, tubeConditionTemperatureStable forCriterionStudy
plasma-citrate, BD 0.109 M sodium citrate (9:1); plasma aliquoted into nonactivating Eppendorf tubesStorageseparated2-8 °C24 hstable for the longest time testedmean % change from baseline >10%, or >25% of samples with individual change >10% (van Geest-Daalderop)Feng et al. 2014, Sci RepTable 2, 'Acceptable Time (h)'
plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquotsStorageseparatedRT4 hlimit stated by the authorsmax allowable mean deviation from t0: ±10%; ±20% for heparin samplesHoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability'
plasma-citrate, BD 0.109 M sodium citrate (9:1); plasma aliquoted into nonactivating Eppendorf tubesStorageseparatedRT24 hstable for the longest time testedmean % change from baseline >10%, or >25% of samples with individual change >10% (van Geest-Daalderop)Feng et al. 2014, Sci RepTable 2, 'Acceptable Time (h)'
Thyroid Stimulating Hormone12 findings
Specimen, tubeConditionTemperatureStable forCriterionStudy
dbs, Whatman 903 Neonatal Screening Cards; heparinized venous blood (50 µL), wrapped in aluminium foilStoragedried−20 °C60 daysthen changed, changed by 180 dayssignificant change of log-TSH vs D0 (mixed effects model, P < 0.05)Magalhães et al. 2018, Arch Endocrinol MetabTable 1; Results paragraph 2
dbs, Whatman 903 Neonatal Screening Cards; heparinized venous blood (50 µL), wrapped in aluminium foilStoragedried−20 °C60 daysthen changed, changed by 180 dayssignificant change of log-TSH vs D0 (mixed effects model, P < 0.05)Magalhães et al. 2018, Arch Endocrinol MetabTable 1; Results paragraph 2
dbs, Whatman 903 Neonatal Screening Cards; heparinized venous blood (50 µL), wrapped in aluminium foilStoragedried2-8 °C60 daysthen changed, changed by 180 dayssignificant change of log-TSH vs D0 (mixed effects model, P < 0.05)Magalhães et al. 2018, Arch Endocrinol MetabTable 1; Results paragraph 2
dbs, Whatman 903 Neonatal Screening Cards; heparinized venous blood (50 µL), wrapped in aluminium foilStoragedried2-8 °C60 daysthen changed, changed by 180 dayssignificant change of log-TSH vs D0 (mixed effects model, P < 0.05)Magalhães et al. 2018, Arch Endocrinol MetabTable 1; Results paragraph 2
dbs, Whatman 903 Neonatal Screening Cards; heparinized venous blood (50 µL), wrapped in aluminium foilStoragedriedRT30 daysthen changed, changed by 60 dayssignificant change of log-TSH vs D0 (mixed effects model, P < 0.05)Magalhães et al. 2018, Arch Endocrinol MetabTable 1; Results paragraph 2
dbs, Whatman 903 Neonatal Screening Cards; heparinized venous blood (50 µL), wrapped in aluminium foilStoragedriedRT30 daysthen changed, changed by 60 dayssignificant change of log-TSH vs D0 (mixed effects model, P < 0.05)Magalhães et al. 2018, Arch Endocrinol MetabTable 1; Results paragraph 2
plasma-li-heparin, Sarstedt lithium heparin, no gel separator (primary tube)Storagecentrifuged-on-cells2-8 °C48 hstable for the longest time testedmean % difference vs time 0 exceeds total limit of change, TLC = ±√((1.65·CVa)² + (0.5·CVb)²) (EFLM BV), here 15.3%; stability limit from instability-equation regressionIllana et al. 2024, Biochem Med (Zagreb)Table 1, row 'TSH'
plasma-li-heparin, Sarstedt lithium heparin, no gel separator; plasma aliquotStorageseparated2-8 °C96 hstable for the longest time testedmean % difference vs time 0 exceeds total limit of change, TLC = ±√((1.65·CVa)² + (0.5·CVb)²) (EFLM BV), here 15.3%; stability limit from instability-equation regressionIllana et al. 2024, Biochem Med (Zagreb)Table 2, row 'TSH'
plasma-li-heparin, BD Barricor 4.5 mL (lithium heparin, mechanical separator)Storagecentrifuged-on-gel2-8 °C7 daysstable for the longest time testedmean % difference vs time 0 exceeds desirable imprecision based on biological variation (EFLM/Westgard), here 7.95%Mandić et al. 2020, Biochem Med (Zagreb)Tables 3 and 4, row 'TSH, mIU/L', BD Barricor columns
serum, BD Vacutainer RST 5 mL (thrombin clot activator, gel)Storagecentrifuged-on-gel2-8 °C7 daysstable for the longest time testedmean % difference vs time 0 exceeds desirable imprecision based on biological variation (EFLM/Westgard), here 7.95%Mandić et al. 2020, Biochem Med (Zagreb)Tables 3 and 4, row 'TSH, mIU/L', BD RST columns
whole-blood-clotted, BD Vacutainer serum, no gel (ref 368815)Before spinninguncentrifugedRT10 hstable for the longest time tested|mean % deviation 10 h vs baseline| > laboratory bias goal (Table II; mostly biological-variation based, some empirically modified)Henriksen et al. 2014, Scand J Clin Lab InvestTable II, row 'TSH'
whole-blood-clotted, BD Vacutainer serum, no gel (ref 368815)Before spinninguncentrifugedRT10 hstable for the longest time testedCV of paired baseline/10 h results > goal CV (Table II)Henriksen et al. 2014, Scand J Clin Lab InvestTable II, row 'TSH'
Total Bilirubin16 findings
Specimen, tubeConditionTemperatureStable forCriterionStudy
plasma-li-heparin, Sarstedt lithium heparin, no gel separator (primary tube)Storagecentrifuged-on-cells2-8 °C48 hstable for the longest time testedmean % difference vs time 0 exceeds total limit of change, TLC = ±√((1.65·CVa)² + (0.5·CVb)²) (EFLM BV), here 21.4%; stability limit from instability-equation regressionIllana et al. 2024, Biochem Med (Zagreb)Table 1, row 'TBIL'
plasma-li-heparin, Sarstedt lithium heparin, no gel separator; plasma aliquotStorageseparated2-8 °C96 hstable for the longest time testedmean % difference vs time 0 exceeds total limit of change, TLC = ±√((1.65·CVa)² + (0.5·CVb)²) (EFLM BV), here 21.4%; stability limit from instability-equation regressionIllana et al. 2024, Biochem Med (Zagreb)Table 2, row 'TBIL'
plasma-li-heparin, BD Barricor 4.5 mL (lithium heparin, mechanical separator)Storagecentrifuged-on-gel2-8 °C7 daysstable for the longest time testedmean % difference vs time 0 exceeds desirable imprecision based on biological variation (EFLM/Westgard), here 10.9%Mandić et al. 2020, Biochem Med (Zagreb)Tables 3 and 4, row 'Bilirubin, μmol/L', BD Barricor columns
serum, BD Vacutainer 8.5 mL plastic serum tube; pooled serum aliquoted into 1.5 mL Eppendorf tubesFreeze–thawseparated−20 °C1 cyclesthen changed, changed by 2 cyclesunstable only if both P < 0.005 (Bonferroni) and % change from T0 > desirable bias (Westgard/Ricos), here 11.4%Cuhadar et al. 2013, Biochem Med (Zagreb)Table 2, row 'Bilirubin, total (μmol/L)', columns T1d-T10d
serum, BD Vacutainer 8.5 mL plastic serum tube; pooled serum aliquoted into 1.5 mL Eppendorf tubesStorageseparated−20 °C3 monthsstable for the longest time testedunstable only if both P < 0.013 (Bonferroni) and % change from T0 > desirable bias (Westgard/Ricos), here 11.4%Cuhadar et al. 2013, Biochem Med (Zagreb)Table 2, row 'Bilirubin, total (μmol/L)', columns T1m-T3m
serum, BD Vacutainer serum (plain), 7.5 mLStorageseparated−20 °C30 daysstable for the longest time testedclinically significant change: mean outside significant change limit (SCL = T1d mean ± 3.0 usual SD from 2 months QC)Kachhawa et al. 2017, J Lab PhysiciansResults paragraph 1 (Table 2 image-only)
serum, Vacuette Tube Serum Gel Separator 8 mL (Greiner, 455071)Storagecentrifuged-on-gel2-8 °C1 weekstable for the longest time tested% bias of median vs baseline exceeds desirable bias (Westgard/Ricos BV), here 11.4%Cuhadar et al. 2012, Biochem Med (Zagreb)Table 3, row 'Bilirubin (total)', gel 4°C
serum, Vacuette Tube Serum No Gel 9 mL (Greiner, 455092), plainStoragecentrifuged-on-cells2-8 °C12 hthen changed, changed by 18 h% bias of median vs baseline exceeds desirable bias (Westgard/Ricos BV), here 11.4%Cuhadar et al. 2012, Biochem Med (Zagreb)Table 3, row 'Bilirubin (total)', w/o gel 4°C
serum, BD Vacutainer RST 5 mL (thrombin clot activator, gel)Storagecentrifuged-on-gel2-8 °C7 daysstable for the longest time testedmean % difference vs time 0 exceeds desirable imprecision based on biological variation (EFLM/Westgard), here 10.9%Mandić et al. 2020, Biochem Med (Zagreb)Tables 3 and 4, row 'Bilirubin, μmol/L', BD RST columns
serum, Vacuette Tube Serum Gel Separator 8 mL (Greiner, 455071)Storagecentrifuged-on-gelRT1 weekstable for the longest time tested% bias of median vs baseline exceeds desirable bias (Westgard/Ricos BV), here 11.4%Cuhadar et al. 2012, Biochem Med (Zagreb)Table 3, row 'Bilirubin (total)', gel 24°C
serum, Vacuette Tube Serum No Gel 9 mL (Greiner, 455092), plainStoragecentrifuged-on-cellsRT12 hthen changed, changed by 18 h% bias of median vs baseline exceeds desirable bias (Westgard/Ricos BV), here 11.4%Cuhadar et al. 2012, Biochem Med (Zagreb)Table 3, row 'Bilirubin (total)', w/o gel 24°C
serum, BD Vacutainer SST (SKU 367986), gelStoragecentrifuged-on-gel30 °C (±1 °C)24 hthen changed, changed by 72 hmean bias vs paired baseline within CLIA acceptance limits AND >2/3 of individual results within limitsIacovetti et al. 2026, Pract Lab MedResults, 'Analyte stability after storage at 30 °C', paragraph 1 (Table 1 image-only)
serum, BD SST Microtainer (SKU 395967), gel; filled from venous bloodStoragecentrifuged-on-gel0 °C72 hstable for the longest time testedmean bias vs paired baseline within CLIA acceptance limits AND >2/3 of individual results within limitsIacovetti et al. 2026, Pract Lab MedResults, 'Analyte stability after separation and storage in microvolume tubes at 0 °C' (Table 2 image-only)
whole-blood-clotted, BD Vacutainer SST (SKU 367986), gelBefore spinninguncentrifuged30 °C (±1 °C)72 hstable for the longest time testedmean bias vs paired baseline within CLIA acceptance limits AND >2/3 of individual results within limitsIacovetti et al. 2026, Pract Lab MedResults, 'Analyte stability after storage at 30 °C', paragraph 2 (Table 1 image-only)
whole-blood-heparin, BD Vacutainer lithium heparin, no gel (ref 368884)Before spinninguncentrifugedRT10 hstable for the longest time tested|mean % deviation 10 h vs baseline| > laboratory bias goal (Table II; mostly biological-variation based, some empirically modified)Henriksen et al. 2014, Scand J Clin Lab InvestTable II, row 'BILI'
whole-blood-heparin, BD Vacutainer lithium heparin, no gel (ref 368884)Before spinninguncentrifugedRT10 hstable for the longest time testedCV of paired baseline/10 h results > goal CV (Table II)Henriksen et al. 2014, Scand J Clin Lab InvestTable II, row 'BILI'
Total Cholesterol28 findings
Specimen, tubeConditionTemperatureStable forCriterionStudy
dbs, Whatman 903 card, mailed in foil pouch with desiccantStoragedriednot stated; ambient postal shipping in the US (Apr-Dec 2016)5 daysthen changed, changed by 6 daysregression coefficient for days before freezing vs 0-2 days significant (P < .05), Model 1 (DBS value, unadjusted)Crimmins et al. 2020, Am J Hum BiolTable 3, rows 'Days before freezing', column 'Total cholesterol'
dbs, Whatman 903 card, mailed in foil pouch with desiccantStoragedriednot stated; ambient postal shipping in the US (Apr-Dec 2016)5 daysthen changed, changed by 6 daysregression coefficient for days before freezing vs 0-2 days significant (P < .05), Model 2 (DBS value adjusted for venous value)Crimmins et al. 2020, Am J Hum BiolTable 3, rows 'Days before freezing', column 'Total cholesterol'
plasma-edta, BD Vacutainer EDTA, 4 mLFreeze–thawseparated−20 °C2 cyclesthen changed, changed by 3 cyclessignificant change vs day 1 (repeated-measures ANOVA, Tukey HSD post hoc, p<0.05)Gojković et al. 2024, J Med BiochemResults paragraph 3 (EDTA -20°C, Figure 3)
plasma-edta, BD Vacutainer EDTA, 4 mLStorageseparated2-8 °C2 daysthen changed, changed by 3 dayssignificant change vs day 1 (repeated-measures ANOVA, Tukey HSD post hoc, p<0.05)Gojković et al. 2024, J Med BiochemResults paragraph 1 (EDTA+4°C, Figure 1)
plasma-fluoride, BD Vacutainer sodium fluoride/potassium oxalate, 4 mLFreeze–thawseparated−20 °C0 cycleschanged at the first time point, changed by 1 cyclessignificant change vs day 1 (repeated-measures ANOVA, Tukey HSD post hoc, p<0.05)Gojković et al. 2024, J Med BiochemResults paragraph 4 (NaF -20°C, Figure 4)
plasma-fluoride, BD Vacutainer sodium fluoride/potassium oxalate, 4 mLStorageseparated2-8 °C4 daysstable for the longest time testedsignificant change vs day 1 (repeated-measures ANOVA, Tukey HSD post hoc, p<0.05)Gojković et al. 2024, J Med BiochemResults paragraph 2 (NaF+4°C, Figure 2)
plasma-li-heparin, Sarstedt lithium heparin, no gel separator (primary tube)Storagecentrifuged-on-cells2-8 °C48 hstable for the longest time testedmean % difference vs time 0 exceeds total limit of change, TLC = ±√((1.65·CVa)² + (0.5·CVb)²) (EFLM BV), here 4.7%; stability limit from instability-equation regressionIllana et al. 2024, Biochem Med (Zagreb)Table 1, row 'CHOL'
plasma-li-heparin, Sarstedt lithium heparin, no gel separator; plasma aliquotStorageseparated2-8 °C96 hstable for the longest time testedmean % difference vs time 0 exceeds total limit of change, TLC = ±√((1.65·CVa)² + (0.5·CVb)²) (EFLM BV), here 4.7%; stability limit from instability-equation regressionIllana et al. 2024, Biochem Med (Zagreb)Table 2, row 'CHOL'
serum, BD Vacutainer 8.5 mL plastic serum tube; pooled serum aliquoted into 1.5 mL Eppendorf tubesFreeze–thawseparated−20 °C10 cyclesstable for the longest time testedunstable only if both P < 0.005 (Bonferroni) and % change from T0 > desirable bias (Westgard/Ricos), here 4.0%Cuhadar et al. 2013, Biochem Med (Zagreb)Table 2, row 'Cholesterol (mmol/L)', columns T1d-T10d
serum, BD Vacutainer SSTFreeze–thawseparated−70 °C, 1 month; thawed at room temperature0 cycleschanged at the first time point, changed by 1 cyclemean % change (bias) from fresh baseline within total allowable error (TEa, desirable bias)Abraham et al. 2021, J Lab PhysiciansTable 2 and Table 3, row 'Cholesterol (mg/dL)'
serum, BD Vacutainer 8.5 mL plastic serum tube; pooled serum aliquoted into 1.5 mL Eppendorf tubesStorageseparated−20 °C3 monthsstable for the longest time testedunstable only if both P < 0.013 (Bonferroni) and % change from T0 > desirable bias (Westgard/Ricos), here 4.0%Cuhadar et al. 2013, Biochem Med (Zagreb)Table 2, row 'Cholesterol (mmol/L)', columns T1m-T3m
serum, BD Vacutainer serum (plain), 7.5 mLStorageseparated−20 °C30 daysstable for the longest time testedclinically significant change: mean outside significant change limit (SCL = T1d mean ± 3.0 usual SD from 2 months QC)Kachhawa et al. 2017, J Lab PhysiciansDiscussion, 'Lipid' (Table 2 image-only)
serum, Vacuette Tube Serum Gel Separator 8 mL (Greiner, 455071)Storagecentrifuged-on-gel2-8 °C1 weekstable for the longest time tested% bias of median vs baseline exceeds desirable bias (Westgard/Ricos BV), here 4.0%Cuhadar et al. 2012, Biochem Med (Zagreb)Table 3, row 'Cholesterol', gel 4°C
serum, Vacuette Tube Serum No Gel 9 mL (Greiner, 455092), plainStoragecentrifuged-on-cells2-8 °C1 weekstable for the longest time tested% bias of median vs baseline exceeds desirable bias (Westgard/Ricos BV), here 4.0%Cuhadar et al. 2012, Biochem Med (Zagreb)Table 3, row 'Cholesterol', w/o gel 4°C
serum, Vacuette Tube Serum Gel Separator 8 mL (Greiner, 455071)Storagecentrifuged-on-gelRT1 weekstable for the longest time tested% bias of median vs baseline exceeds desirable bias (Westgard/Ricos BV), here 4.0%Cuhadar et al. 2012, Biochem Med (Zagreb)Table 3, row 'Cholesterol', gel 24°C
serum, Vacuette Tube Serum No Gel 9 mL (Greiner, 455092), plainStoragecentrifuged-on-cellsRT1 weekstable for the longest time tested% bias of median vs baseline exceeds desirable bias (Westgard/Ricos BV), here 4.0%Cuhadar et al. 2012, Biochem Med (Zagreb)Table 3, row 'Cholesterol', w/o gel 24°C
serum, BD Vacutainer SST (SKU 367986), gelStoragecentrifuged-on-gel30 °C (±1 °C)72 hstable for the longest time testedmean bias vs paired baseline within CLIA acceptance limits AND >2/3 of individual results within limitsIacovetti et al. 2026, Pract Lab MedResults, 'Analyte stability after storage at 30 °C', paragraph 1 (Table 1 image-only)
serum, BD SST Microtainer (SKU 395967), gel; filled from venous bloodStoragecentrifuged-on-gel0 °C72 hstable for the longest time testedmean bias vs paired baseline within CLIA acceptance limits AND >2/3 of individual results within limitsIacovetti et al. 2026, Pract Lab MedResults, 'Analyte stability after separation and storage in microvolume tubes at 0 °C' (Table 2 image-only)
whole-blood-clotted, BD Vacutainer SSTBefore spinninguncentrifugedRT24 hstable for the longest time testedmean % deviation from baseline within total change limit TCL = sqrt((2.77 CVa)^2 + (0.5 CVb)^2), CVb from Ricos (Table 2)Abraham et al. 2019, Biochem Med (Zagreb)Discussion, paragraph 1 (Results paragraphs 1-3; Figure 3)
whole-blood-clotted, BD Vacutainer trace element-free serum tubeBefore spinninguncentrifugedcold packs in insulated box; 9 (2) °C on arrival24 hstable for the longest time tested% difference within desirable quality specification (from intra- and interindividual variation)Jones et al. 2024, J NutrTable 2, row 'Total cholesterol, mmol/L'; Table 1
whole-blood-clotted, BD Vacutainer SST (SKU 367986), gelBefore spinninguncentrifuged30 °C (±1 °C)72 hstable for the longest time testedmean bias vs paired baseline within CLIA acceptance limits AND >2/3 of individual results within limitsIacovetti et al. 2026, Pract Lab MedResults, 'Analyte stability after storage at 30 °C', paragraph 2 (Table 1 image-only)
whole-blood-clotted, BD Vacutainer SSTBefore spinninguncentrifuged22-30 °C (logger 24-26 °C)24 hstable for the longest time testedmean % deviation from baseline within total change limit TCL = sqrt((2.77 CVa)^2 + (0.5 CVb)^2), CVb from Ricos (Table 2)Abraham et al. 2019, Biochem Med (Zagreb)Discussion, paragraph 1 (Results paragraphs 1-3; Figure 3)
whole-blood-clotted, BD Vacutainer SSTBefore spinninguncentrifuged> 30 °C (logger 38-40 °C)24 hstable for the longest time testedmean % deviation from baseline within total change limit TCL = sqrt((2.77 CVa)^2 + (0.5 CVb)^2), CVb from Ricos (Table 2)Abraham et al. 2019, Biochem Med (Zagreb)Discussion, paragraph 1 (Results paragraphs 1-3; Figure 3)
whole-blood-clotted, Sarstedt Monovette serumBefore spinninguncentrifugedchilled in transit (insulated box, 2 frozen cool packs; temperature not recorded)24 hstable for the longest time tested% difference within desirable quality specification for bias (Fraser, from intra- and interindividual variation)Jones et al. 2021, J NutrTable 2, row 'Total cholesterol', column Serum
whole-blood-edta, Sarstedt Monovette EDTABefore spinninguncentrifugedchilled in transit (insulated box, 2 frozen cool packs; temperature not recorded)24 hstable for the longest time tested% difference within desirable quality specification for bias (Fraser, from intra- and interindividual variation)Jones et al. 2021, J NutrTable 2, row 'Total cholesterol', column EDTA
whole-blood-heparin, BD Vacutainer lithium heparin, no gel (ref 368884)Before spinninguncentrifugedRT10 hstable for the longest time tested|mean % deviation 10 h vs baseline| > laboratory bias goal (Table II; mostly biological-variation based, some empirically modified)Henriksen et al. 2014, Scand J Clin Lab InvestTable II, row 'CHOL'
whole-blood-heparin, BD Vacutainer lithium heparin, no gel (ref 368884)Before spinninguncentrifugedRT10 hstable for the longest time testedCV of paired baseline/10 h results > goal CV (Table II)Henriksen et al. 2014, Scand J Clin Lab InvestTable II, row 'CHOL'
whole-blood-heparin, Sarstedt Monovette lithium heparin, trace-element freeBefore spinninguncentrifugedchilled in transit (insulated box, 2 frozen cool packs; temperature not recorded)24 hstable for the longest time tested% difference within desirable quality specification for bias (Fraser, from intra- and interindividual variation)Jones et al. 2021, J NutrTable 2, row 'Total cholesterol', column LH
Total protein15 findings
Specimen, tubeConditionTemperatureStable forCriterionStudy
plasma-li-heparin, Sarstedt lithium heparin, no gel separator (primary tube)Storagecentrifuged-on-cells2-8 °C48 hstable for the longest time testedmean % difference vs time 0 exceeds total limit of change, TLC = ±√((1.65·CVa)² + (0.5·CVb)²) (EFLM BV), here 4.0%; stability limit from instability-equation regressionIllana et al. 2024, Biochem Med (Zagreb)Table 1, row 'TP'
plasma-li-heparin, Sarstedt lithium heparin, no gel separator; plasma aliquotStorageseparated2-8 °C96 hstable for the longest time testedmean % difference vs time 0 exceeds total limit of change, TLC = ±√((1.65·CVa)² + (0.5·CVb)²) (EFLM BV), here 4.0%; stability limit from instability-equation regressionIllana et al. 2024, Biochem Med (Zagreb)Table 2, row 'TP'
plasma-li-heparin, BD Barricor 4.5 mL (lithium heparin, mechanical separator)Storagecentrifuged-on-gel2-8 °C7 daysstable for the longest time testedmean % difference vs time 0 exceeds desirable imprecision based on biological variation (EFLM/Westgard), here 1.3%Mandić et al. 2020, Biochem Med (Zagreb)Tables 3 and 4, row 'Total protein, g/L', BD Barricor columns
serum, BD Vacutainer 8.5 mL plastic serum tube; pooled serum aliquoted into 1.5 mL Eppendorf tubesFreeze–thawseparated−20 °C0 cycleschanged at the first time point, changed by 1 cyclesunstable only if both P < 0.005 (Bonferroni) and % change from T0 > desirable bias (Westgard/Ricos), here 1.2%Cuhadar et al. 2013, Biochem Med (Zagreb)Table 2, row 'Protein, total (g/L)', columns T1d-T10d
serum, BD Vacutainer SSTFreeze–thawseparated−70 °C, 1 month; thawed at room temperature1 cyclestable for the longest time testedmean % change (bias) from fresh baseline within total allowable error (TEa, desirable bias)Abraham et al. 2021, J Lab PhysiciansTable 2 and Table 3, row 'Total protein (g/dL)'
serum, BD Vacutainer 8.5 mL plastic serum tube; pooled serum aliquoted into 1.5 mL Eppendorf tubesStorageseparated−20 °C2 monthsthen changed, changed by 3 monthsunstable only if both P < 0.013 (Bonferroni) and % change from T0 > desirable bias (Westgard/Ricos), here 1.2%Cuhadar et al. 2013, Biochem Med (Zagreb)Table 2, row 'Protein, total (g/L)', columns T1m-T3m
serum, BD Vacutainer serum (plain), 7.5 mLStorageseparated−20 °C30 daysstable for the longest time testedclinically significant change: mean outside significant change limit (SCL = T1d mean ± 3.0 usual SD from 2 months QC)Kachhawa et al. 2017, J Lab PhysiciansDiscussion, 'Protein' (Table 2 image-only)
serum, Vacuette Tube Serum Gel Separator 8 mL (Greiner, 455071)Storagecentrifuged-on-gel2-8 °C0 hchanged at the first time point, changed by 6 h% bias of median vs baseline exceeds desirable bias (Westgard/Ricos BV), here 1.2%Cuhadar et al. 2012, Biochem Med (Zagreb)Table 3, row 'Protein (total)', gel 4°C
serum, Vacuette Tube Serum No Gel 9 mL (Greiner, 455092), plainStoragecentrifuged-on-cells2-8 °C0 hchanged at the first time point, changed by 6 h% bias of median vs baseline exceeds desirable bias (Westgard/Ricos BV), here 1.2%Cuhadar et al. 2012, Biochem Med (Zagreb)Table 3, row 'Protein (total)', w/o gel 4°C
serum, BD Vacutainer RST 5 mL (thrombin clot activator, gel)Storagecentrifuged-on-gel2-8 °C7 daysstable for the longest time testedmean % difference vs time 0 exceeds desirable imprecision based on biological variation (EFLM/Westgard), here 1.3%Mandić et al. 2020, Biochem Med (Zagreb)Tables 3 and 4, row 'Total protein, g/L', BD RST columns
serum, Vacuette Tube Serum Gel Separator 8 mL (Greiner, 455071)Storagecentrifuged-on-gelRT0 hchanged at the first time point, changed by 6 h% bias of median vs baseline exceeds desirable bias (Westgard/Ricos BV), here 1.2%Cuhadar et al. 2012, Biochem Med (Zagreb)Table 3, row 'Protein (total)', gel 24°C
serum, Vacuette Tube Serum No Gel 9 mL (Greiner, 455092), plainStoragecentrifuged-on-cellsRT12 hthen changed, changed by 18 h% bias of median vs baseline exceeds desirable bias (Westgard/Ricos BV), here 1.2%Cuhadar et al. 2012, Biochem Med (Zagreb)Table 3, row 'Protein (total)', w/o gel 24°C
serum, BD Vacutainer SST (SKU 367986), gelStoragecentrifuged-on-gel30 °C (±1 °C)72 hstable for the longest time testedmean bias vs paired baseline within CLIA acceptance limits AND >2/3 of individual results within limitsIacovetti et al. 2026, Pract Lab MedResults, 'Analyte stability after storage at 30 °C', paragraph 1 (Table 1 image-only)
serum, BD SST Microtainer (SKU 395967), gel; filled from venous bloodStoragecentrifuged-on-gel0 °C72 hstable for the longest time testedmean bias vs paired baseline within CLIA acceptance limits AND >2/3 of individual results within limitsIacovetti et al. 2026, Pract Lab MedResults, 'Analyte stability after separation and storage in microvolume tubes at 0 °C' (Table 2 image-only)
whole-blood-clotted, BD Vacutainer SST (SKU 367986), gelBefore spinninguncentrifuged30 °C (±1 °C)72 hstable for the longest time testedmean bias vs paired baseline within CLIA acceptance limits AND >2/3 of individual results within limitsIacovetti et al. 2026, Pract Lab MedResults, 'Analyte stability after storage at 30 °C', paragraph 2 (Table 1 image-only)
Transferrin2 findings
Specimen, tubeConditionTemperatureStable forCriterionStudy
whole-blood-heparin, BD Vacutainer lithium heparin, no gel (ref 368884)Before spinninguncentrifugedRT10 hstable for the longest time tested|mean % deviation 10 h vs baseline| > laboratory bias goal (Table II; mostly biological-variation based, some empirically modified)Henriksen et al. 2014, Scand J Clin Lab InvestTable II, row 'TRANS'
whole-blood-heparin, BD Vacutainer lithium heparin, no gel (ref 368884)Before spinninguncentrifugedRT10 hstable for the longest time testedCV of paired baseline/10 h results > goal CV (Table II)Henriksen et al. 2014, Scand J Clin Lab InvestTable II, row 'TRANS'
Triglycerides26 findings
Specimen, tubeConditionTemperatureStable forCriterionStudy
plasma-edta, BD Vacutainer EDTA, 4 mLFreeze–thawseparated−20 °C1 cyclesthen changed, changed by 2 cyclessignificant change vs day 1 (repeated-measures ANOVA, Tukey HSD post hoc, p<0.05)Gojković et al. 2024, J Med BiochemResults paragraph 3 (EDTA -20°C, Figure 3)
plasma-edta, BD Vacutainer EDTA, 4 mLStorageseparated2-8 °C4 daysstable for the longest time testedsignificant change vs day 1 (repeated-measures ANOVA, Tukey HSD post hoc, p<0.05)Gojković et al. 2024, J Med BiochemResults paragraph 1 (EDTA+4°C, Figure 1)
plasma-fluoride, BD Vacutainer sodium fluoride/potassium oxalate, 4 mLFreeze–thawseparated−20 °C4 cyclesstable for the longest time testedsignificant change vs day 1 (repeated-measures ANOVA, Tukey HSD post hoc, p<0.05)Gojković et al. 2024, J Med BiochemResults paragraph 4 (NaF -20°C, Figure 4)
plasma-fluoride, BD Vacutainer sodium fluoride/potassium oxalate, 4 mLStorageseparated2-8 °C4 daysstable for the longest time testedsignificant change vs day 1 (repeated-measures ANOVA, Tukey HSD post hoc, p<0.05)Gojković et al. 2024, J Med BiochemResults paragraph 2 (NaF+4°C, Figure 2)
plasma-li-heparin, Sarstedt lithium heparin, no gel separator (primary tube)Storagecentrifuged-on-cells2-8 °C48 hstable for the longest time testedmean % difference vs time 0 exceeds total limit of change, TLC = ±√((1.65·CVa)² + (0.5·CVb)²) (EFLM BV), here 11.4%; stability limit from instability-equation regressionIllana et al. 2024, Biochem Med (Zagreb)Table 1, row 'TG'
plasma-li-heparin, Sarstedt lithium heparin, no gel separator; plasma aliquotStorageseparated2-8 °C96 hstable for the longest time testedmean % difference vs time 0 exceeds total limit of change, TLC = ±√((1.65·CVa)² + (0.5·CVb)²) (EFLM BV), here 11.4%; stability limit from instability-equation regressionIllana et al. 2024, Biochem Med (Zagreb)Table 2, row 'TG'
serum, BD Vacutainer 8.5 mL plastic serum tube; pooled serum aliquoted into 1.5 mL Eppendorf tubesFreeze–thawseparated−20 °C10 cyclesstable for the longest time testedunstable only if both P < 0.005 (Bonferroni) and % change from T0 > desirable bias (Westgard/Ricos), here 10.7%Cuhadar et al. 2013, Biochem Med (Zagreb)Table 2, row 'Triglyceride (mmol/L)', columns T1d-T10d
serum, BD Vacutainer SSTFreeze–thawseparated−70 °C, 1 month; thawed at room temperature1 cyclestable for the longest time testedmean % change (bias) from fresh baseline within total allowable error (TEa, desirable bias)Abraham et al. 2021, J Lab PhysiciansTable 2 and Table 3, row 'TAG (mg/dL)'
serum, BD Vacutainer 8.5 mL plastic serum tube; pooled serum aliquoted into 1.5 mL Eppendorf tubesStorageseparated−20 °C3 monthsstable for the longest time testedunstable only if both P < 0.013 (Bonferroni) and % change from T0 > desirable bias (Westgard/Ricos), here 10.7%Cuhadar et al. 2013, Biochem Med (Zagreb)Table 2, row 'Triglyceride (mmol/L)', columns T1m-T3m
serum, BD Vacutainer serum (plain), 7.5 mLStorageseparated−20 °C30 daysstable for the longest time testedclinically significant change: mean outside significant change limit (SCL = T1d mean ± 3.0 usual SD from 2 months QC)Kachhawa et al. 2017, J Lab PhysiciansDiscussion, 'Lipid' (Table 2 image-only)
serum, Vacuette Tube Serum Gel Separator 8 mL (Greiner, 455071)Storagecentrifuged-on-gel2-8 °C1 weekstable for the longest time tested% bias of median vs baseline exceeds desirable bias (Westgard/Ricos BV), here 10.7%Cuhadar et al. 2012, Biochem Med (Zagreb)Table 3, row 'Triglycerides', gel 4°C
serum, Vacuette Tube Serum No Gel 9 mL (Greiner, 455092), plainStoragecentrifuged-on-cells2-8 °C1 weekstable for the longest time tested% bias of median vs baseline exceeds desirable bias (Westgard/Ricos BV), here 10.7%Cuhadar et al. 2012, Biochem Med (Zagreb)Table 3, row 'Triglycerides', w/o gel 4°C
serum, Vacuette Tube Serum Gel Separator 8 mL (Greiner, 455071)Storagecentrifuged-on-gelRT1 weekstable for the longest time tested% bias of median vs baseline exceeds desirable bias (Westgard/Ricos BV), here 10.7%Cuhadar et al. 2012, Biochem Med (Zagreb)Table 3, row 'Triglycerides', gel 24°C
serum, Vacuette Tube Serum No Gel 9 mL (Greiner, 455092), plainStoragecentrifuged-on-cellsRT1 weekstable for the longest time tested% bias of median vs baseline exceeds desirable bias (Westgard/Ricos BV), here 10.7%Cuhadar et al. 2012, Biochem Med (Zagreb)Table 3, row 'Triglycerides', w/o gel 24°C
serum, BD Vacutainer SST (SKU 367986), gelStoragecentrifuged-on-gel30 °C (±1 °C)72 hstable for the longest time testedmean bias vs paired baseline within CLIA acceptance limits AND >2/3 of individual results within limitsIacovetti et al. 2026, Pract Lab MedResults, 'Analyte stability after storage at 30 °C', paragraph 1 (Table 1 image-only)
serum, BD SST Microtainer (SKU 395967), gel; filled from venous bloodStoragecentrifuged-on-gel0 °C72 hstable for the longest time testedmean bias vs paired baseline within CLIA acceptance limits AND >2/3 of individual results within limitsIacovetti et al. 2026, Pract Lab MedResults, 'Analyte stability after separation and storage in microvolume tubes at 0 °C' (Table 2 image-only)
whole-blood-clotted, BD Vacutainer SSTBefore spinninguncentrifugedRT24 hstable for the longest time testedmean % deviation from baseline within total change limit TCL = sqrt((2.77 CVa)^2 + (0.5 CVb)^2), CVb from Ricos (Table 2)Abraham et al. 2019, Biochem Med (Zagreb)Discussion, paragraph 1 (Results paragraphs 1-3; Figure 3)
whole-blood-clotted, BD Vacutainer trace element-free serum tubeBefore spinninguncentrifugedcold packs in insulated box; 9 (2) °C on arrival24 hstable for the longest time tested% difference within desirable quality specification (from intra- and interindividual variation)Jones et al. 2024, J NutrTable 2, row 'Triglycerides, mmol/L'; Table 1
whole-blood-clotted, BD Vacutainer SST (SKU 367986), gelBefore spinninguncentrifuged30 °C (±1 °C)72 hstable for the longest time testedmean bias vs paired baseline within CLIA acceptance limits AND >2/3 of individual results within limitsIacovetti et al. 2026, Pract Lab MedResults, 'Analyte stability after storage at 30 °C', paragraph 2 (Table 1 image-only)
whole-blood-clotted, BD Vacutainer SSTBefore spinninguncentrifuged22-30 °C (logger 24-26 °C)24 hstable for the longest time testedmean % deviation from baseline within total change limit TCL = sqrt((2.77 CVa)^2 + (0.5 CVb)^2), CVb from Ricos (Table 2)Abraham et al. 2019, Biochem Med (Zagreb)Discussion, paragraph 1 (Results paragraphs 1-3; Figure 3)
whole-blood-clotted, BD Vacutainer SSTBefore spinninguncentrifuged> 30 °C (logger 38-40 °C)24 hstable for the longest time testedmean % deviation from baseline within total change limit TCL = sqrt((2.77 CVa)^2 + (0.5 CVb)^2), CVb from Ricos (Table 2)Abraham et al. 2019, Biochem Med (Zagreb)Discussion, paragraph 1 (Results paragraphs 1-3; Figure 3)
whole-blood-clotted, Sarstedt Monovette serumBefore spinninguncentrifugedchilled in transit (insulated box, 2 frozen cool packs; temperature not recorded)24 hstable for the longest time tested% difference within desirable quality specification for bias (Fraser, from intra- and interindividual variation)Jones et al. 2021, J NutrTable 2, row 'Triglycerides', column Serum
whole-blood-edta, Sarstedt Monovette EDTABefore spinninguncentrifugedchilled in transit (insulated box, 2 frozen cool packs; temperature not recorded)24 hstable for the longest time tested% difference within desirable quality specification for bias (Fraser, from intra- and interindividual variation)Jones et al. 2021, J NutrTable 2, row 'Triglycerides', column EDTA
whole-blood-heparin, BD Vacutainer lithium heparin, no gel (ref 368884)Before spinninguncentrifugedRT10 hstable for the longest time tested|mean % deviation 10 h vs baseline| > laboratory bias goal (Table II; mostly biological-variation based, some empirically modified)Henriksen et al. 2014, Scand J Clin Lab InvestTable II, row 'TGLY'
whole-blood-heparin, BD Vacutainer lithium heparin, no gel (ref 368884)Before spinninguncentrifugedRT10 hstable for the longest time testedCV of paired baseline/10 h results > goal CV (Table II)Henriksen et al. 2014, Scand J Clin Lab InvestTable II, row 'TGLY'
whole-blood-heparin, Sarstedt Monovette lithium heparin, trace-element freeBefore spinninguncentrifugedchilled in transit (insulated box, 2 frozen cool packs; temperature not recorded)24 hstable for the longest time tested% difference within desirable quality specification for bias (Fraser, from intra- and interindividual variation)Jones et al. 2021, J NutrTable 2, row 'Triglycerides', column LH
Troponin I (assay sensitivity not stated)2 findings
Specimen, tubeConditionTemperatureStable forCriterionStudy
plasma-li-heparin, BD Barricor 4.5 mL (lithium heparin, mechanical separator)Storagecentrifuged-on-gel2-8 °C24 hthen changed, changed by 7 daysmean % difference vs time 0 exceeds desirable imprecision based on biological variation (EFLM/Westgard), here 7.03%Mandić et al. 2020, Biochem Med (Zagreb)Tables 3 and 4, row 'TnI, ng/L', BD Barricor columns
serum, BD Vacutainer RST 5 mL (thrombin clot activator, gel)Storagecentrifuged-on-gel2-8 °C24 hthen changed, changed by 7 daysmean % difference vs time 0 exceeds desirable imprecision based on biological variation (EFLM/Westgard), here 7.03%Mandić et al. 2020, Biochem Med (Zagreb)Tables 3 and 4, row 'TnI, ng/L', BD RST columns
Troponin T (High-Sensitivity)2 findings
Specimen, tubeConditionTemperatureStable forCriterionStudy
plasma-li-heparin, BD Vacutainer lithium heparin 6 mL (regular)Storagenot-stated2-8 °C72 hstable for the longest time testedmean % difference vs T0 exceeds total change limit (TCL), here ±9.37%Dupuy et al. 2018, Biochem Med (Zagreb)Table 2, BD lithium heparin tubes, row 'hs-TnT, ng/L'
plasma-li-heparin, BD Barricor lithium heparin 5.5 mL (mechanical separator)Storagecentrifuged-on-gel2-8 °C72 hstable for the longest time testedmean % difference vs T0 exceeds total change limit (TCL), here ±9.37%Dupuy et al. 2018, Biochem Med (Zagreb)Table 2, BD Barricor lithium heparin tubes, row 'hs-TnT, ng/L'
Urea12 findings
Specimen, tubeConditionTemperatureStable forCriterionStudy
plasma-edta, BD Vacutainer EDTA, 4 mLFreeze–thawseparated−20 °C4 cyclesstable for the longest time testedsignificant change vs day 1 (repeated-measures ANOVA, Tukey HSD post hoc, p<0.05)Gojković et al. 2024, J Med BiochemResults paragraph 3 (EDTA -20°C, Figure 3)
plasma-edta, BD Vacutainer EDTA, 4 mLStorageseparated2-8 °C4 daysstable for the longest time testedsignificant change vs day 1 (repeated-measures ANOVA, Tukey HSD post hoc, p<0.05)Gojković et al. 2024, J Med BiochemResults paragraph 1 (EDTA+4°C, Figure 1)
plasma-fluoride, BD Vacutainer sodium fluoride/potassium oxalate, 4 mLFreeze–thawseparated−20 °C4 cyclesstable for the longest time testedsignificant change vs day 1 (repeated-measures ANOVA, Tukey HSD post hoc, p<0.05)Gojković et al. 2024, J Med BiochemResults paragraph 4 (NaF -20°C, Figure 4)
plasma-fluoride, BD Vacutainer sodium fluoride/potassium oxalate, 4 mLStorageseparated2-8 °C4 daysstable for the longest time testedsignificant change vs day 1 (repeated-measures ANOVA, Tukey HSD post hoc, p<0.05)Gojković et al. 2024, J Med BiochemResults paragraph 2 (NaF+4°C, Figure 2)
plasma-li-heparin, Sarstedt lithium heparin, no gel separator (primary tube)Storagecentrifuged-on-cells2-8 °C48 hstable for the longest time testedmean % difference vs time 0 exceeds total limit of change, TLC = ±√((1.65·CVa)² + (0.5·CVb)²) (EFLM BV), here 12.8%; stability limit from instability-equation regressionIllana et al. 2024, Biochem Med (Zagreb)Table 1, row 'Urea'
plasma-li-heparin, Sarstedt lithium heparin, no gel separator; plasma aliquotStorageseparated2-8 °C96 hstable for the longest time testedmean % difference vs time 0 exceeds total limit of change, TLC = ±√((1.65·CVa)² + (0.5·CVb)²) (EFLM BV), here 12.8%; stability limit from instability-equation regressionIllana et al. 2024, Biochem Med (Zagreb)Table 2, row 'Urea'
plasma-li-heparin, BD Vacutainer LH PST II gel separator, 3.0 mL (REF 367374)Storagecentrifuged-on-gel2-8 °C24 hstable for the longest time testedmean % difference smaller than TEa (Ricos: 0.25 x (CVi^2 + CVg^2)^0.5 + 1.65 x 0.5 x CVi)Nonkes et al. 2019, Biochem Med (Zagreb)Table 1, row 'Urea, mmol/L'
plasma-li-heparin, BD Barricor 4.5 mL (lithium heparin, mechanical separator)Storagecentrifuged-on-gel2-8 °C7 daysstable for the longest time testedmean % difference vs time 0 exceeds desirable imprecision based on biological variation (EFLM/Westgard), here 6.95%Mandić et al. 2020, Biochem Med (Zagreb)Tables 3 and 4, row 'Urea, mmol/L', BD Barricor columns
serum, BD Vacutainer serum (plain), 7.5 mLStorageseparated−20 °C30 daysstable for the longest time testedclinically significant change: mean outside significant change limit (SCL = T1d mean ± 3.0 usual SD from 2 months QC)Kachhawa et al. 2017, J Lab PhysiciansResults paragraph 1 (Table 2 image-only)
serum, BD Vacutainer RST 5 mL (thrombin clot activator, gel)Storagecentrifuged-on-gel2-8 °C7 daysstable for the longest time testedmean % difference vs time 0 exceeds desirable imprecision based on biological variation (EFLM/Westgard), here 6.95%Mandić et al. 2020, Biochem Med (Zagreb)Tables 3 and 4, row 'Urea, mmol/L', BD RST columns
whole-blood-heparin, BD Vacutainer lithium heparin, no gel (ref 368884)Before spinninguncentrifugedRT10 hstable for the longest time tested|mean % deviation 10 h vs baseline| > laboratory bias goal (Table II; mostly biological-variation based, some empirically modified)Henriksen et al. 2014, Scand J Clin Lab InvestTable II, row 'UREA'
whole-blood-heparin, BD Vacutainer lithium heparin, no gel (ref 368884)Before spinninguncentrifugedRT10 hstable for the longest time testedCV of paired baseline/10 h results > goal CV (Table II)Henriksen et al. 2014, Scand J Clin Lab InvestTable II, row 'UREA'
Uric acid16 findings
Specimen, tubeConditionTemperatureStable forCriterionStudy
plasma-li-heparin, Sarstedt lithium heparin, no gel separator (primary tube)Storagecentrifuged-on-cells2-8 °C48 hstable for the longest time testedmean % difference vs time 0 exceeds total limit of change, TLC = ±√((1.65·CVa)² + (0.5·CVb)²) (EFLM BV), here 8.3%; stability limit from instability-equation regressionIllana et al. 2024, Biochem Med (Zagreb)Table 1, row 'UA'
plasma-li-heparin, Sarstedt lithium heparin, no gel separator; plasma aliquotStorageseparated2-8 °C96 hstable for the longest time testedmean % difference vs time 0 exceeds total limit of change, TLC = ±√((1.65·CVa)² + (0.5·CVb)²) (EFLM BV), here 8.3%; stability limit from instability-equation regressionIllana et al. 2024, Biochem Med (Zagreb)Table 2, row 'UA'
plasma-li-heparin, BD Barricor 4.5 mL (lithium heparin, mechanical separator)Storagecentrifuged-on-gel2-8 °C7 daysstable for the longest time testedmean % difference vs time 0 exceeds desirable imprecision based on biological variation (EFLM/Westgard), here 4.3%Mandić et al. 2020, Biochem Med (Zagreb)Tables 3 and 4, row 'Uric acid, μmol/L', BD Barricor columns
serum, BD Vacutainer 8.5 mL plastic serum tube; pooled serum aliquoted into 1.5 mL Eppendorf tubesFreeze–thawseparated−20 °C3 cyclesthen changed, changed by 4 cyclesunstable only if both P < 0.005 (Bonferroni) and % change from T0 > desirable bias (Westgard/Ricos), here 4.9%Cuhadar et al. 2013, Biochem Med (Zagreb)Table 2, row 'Uric acid (μmol/L)', columns T1d-T10d
serum, BD Vacutainer 8.5 mL plastic serum tube; pooled serum aliquoted into 1.5 mL Eppendorf tubesStorageseparated−20 °C3 monthsstable for the longest time testedunstable only if both P < 0.013 (Bonferroni) and % change from T0 > desirable bias (Westgard/Ricos), here 4.9%Cuhadar et al. 2013, Biochem Med (Zagreb)Table 2, row 'Uric acid (μmol/L)', columns T1m-T3m
serum, BD Vacutainer serum (plain), 7.5 mLStorageseparated−20 °C30 daysstable for the longest time testedclinically significant change: mean outside significant change limit (SCL = T1d mean ± 3.0 usual SD from 2 months QC)Kachhawa et al. 2017, J Lab PhysiciansResults paragraph 1 (Table 2 image-only)
serum, Vacuette Tube Serum Gel Separator 8 mL (Greiner, 455071)Storagecentrifuged-on-gel2-8 °C30 hthen changed, changed by 36 h% bias of median vs baseline exceeds desirable bias (Westgard/Ricos BV), here 4.9%Cuhadar et al. 2012, Biochem Med (Zagreb)Table 3, row 'Uric acid', gel 4°C
serum, Vacuette Tube Serum No Gel 9 mL (Greiner, 455092), plainStoragecentrifuged-on-cells2-8 °C18 hthen changed, changed by 24 h% bias of median vs baseline exceeds desirable bias (Westgard/Ricos BV), here 4.9%Cuhadar et al. 2012, Biochem Med (Zagreb)Table 3, row 'Uric acid', w/o gel 4°C
serum, BD Vacutainer RST 5 mL (thrombin clot activator, gel)Storagecentrifuged-on-gel2-8 °C7 daysstable for the longest time testedmean % difference vs time 0 exceeds desirable imprecision based on biological variation (EFLM/Westgard), here 4.3%Mandić et al. 2020, Biochem Med (Zagreb)Tables 3 and 4, row 'Uric acid, μmol/L', BD RST columns
serum, Vacuette Tube Serum Gel Separator 8 mL (Greiner, 455071)Storagecentrifuged-on-gelRT6 hthen changed, changed by 12 h% bias of median vs baseline exceeds desirable bias (Westgard/Ricos BV), here 4.9%Cuhadar et al. 2012, Biochem Med (Zagreb)Table 3, row 'Uric acid', gel 24°C
serum, Vacuette Tube Serum No Gel 9 mL (Greiner, 455092), plainStoragecentrifuged-on-cellsRT6 hthen changed, changed by 12 h% bias of median vs baseline exceeds desirable bias (Westgard/Ricos BV), here 4.9%Cuhadar et al. 2012, Biochem Med (Zagreb)Table 3, row 'Uric acid', w/o gel 24°C
serum, BD Vacutainer SST (SKU 367986), gelStoragecentrifuged-on-gel30 °C (±1 °C)72 hstable for the longest time testedmean bias vs paired baseline within CLIA acceptance limits AND >2/3 of individual results within limitsIacovetti et al. 2026, Pract Lab MedResults, 'Analyte stability after storage at 30 °C', paragraph 1 (Table 1 image-only)
serum, BD SST Microtainer (SKU 395967), gel; filled from venous bloodStoragecentrifuged-on-gel0 °C72 hstable for the longest time testedmean bias vs paired baseline within CLIA acceptance limits AND >2/3 of individual results within limitsIacovetti et al. 2026, Pract Lab MedResults, 'Analyte stability after separation and storage in microvolume tubes at 0 °C' (Table 2 image-only)
whole-blood-clotted, BD Vacutainer SST (SKU 367986), gelBefore spinninguncentrifuged30 °C (±1 °C)72 hstable for the longest time testedmean bias vs paired baseline within CLIA acceptance limits AND >2/3 of individual results within limitsIacovetti et al. 2026, Pract Lab MedResults, 'Analyte stability after storage at 30 °C', paragraph 2 (Table 1 image-only)
whole-blood-heparin, BD Vacutainer lithium heparin, no gel (ref 368884)Before spinninguncentrifugedRT10 hstable for the longest time tested|mean % deviation 10 h vs baseline| > laboratory bias goal (Table II; mostly biological-variation based, some empirically modified)Henriksen et al. 2014, Scand J Clin Lab InvestTable II, row 'URATE'
whole-blood-heparin, BD Vacutainer lithium heparin, no gel (ref 368884)Before spinninguncentrifugedRT10 hstable for the longest time testedCV of paired baseline/10 h results > goal CV (Table II)Henriksen et al. 2014, Scand J Clin Lab InvestTable II, row 'URATE'
Urine Albumin6 findings
Specimen, tubeConditionTemperatureStable forCriterionStudy
urine, Sarstedt 2 ml polypropylene micro tube (P72.609.001), untreatedStoragenot-stated−20 °C0 monthschanged at the first time point, changed by 12 monthsWilcoxon signed-rank test, 12 months vs freshChapman et al. 2019, Pract Lab MedResults, paragraph 2
urine, alkalised (NaOH to pH > 8), Sarstedt 2 ml polypropylene micro tube (P72.609.001)Storagenot-stated−20 °C0 monthschanged at the first time point, changed by 12 monthsWilcoxon signed-rank test, 12 months vs freshChapman et al. 2019, Pract Lab MedResults, paragraph 2; Fig. 1A
urine, boric acid precoated container (Sarstedt P51.9923.820), then Sarstedt 2 ml polypropylene micro tube (P72.609.001)Storagenot-stated−20 °C0 monthschanged at the first time point, changed by 12 monthsWilcoxon signed-rank test, 12 months vs freshChapman et al. 2019, Pract Lab MedResults, paragraph 2; Fig. 1A
urine, pepstatin 1 μg/ml added, Sarstedt 2 ml polypropylene micro tube (P72.609.001)Storagenot-stated−20 °C0 monthschanged at the first time point, changed by 12 monthsWilcoxon signed-rank test, 12 months vs freshChapman et al. 2019, Pract Lab MedResults, paragraph 2; Fig. 1A
urine, Sarstedt 1.5 ml low protein binding micro tube (P72.706.600)Storagenot-stated−20 °C0 monthschanged at the first time point, changed by 12 monthsWilcoxon signed-rank test, 12 months vs freshChapman et al. 2019, Pract Lab MedResults, paragraph 2; Fig. 1A
urine, Sarstedt 2 ml polypropylene micro tube (P72.609.001), untreatedStoragenot-stated−80 °C12 monthsstable for the longest time testedWilcoxon signed-rank test, 12 months vs freshChapman et al. 2019, Pract Lab MedResults, paragraph 2
Urine Albumin/Creatinine Ratio6 findings
Specimen, tubeConditionTemperatureStable forCriterionStudy
urine, Sarstedt 2 ml polypropylene micro tube (P72.609.001), untreatedStoragenot-stated−20 °C0 monthschanged at the first time point, changed by 12 monthsWilcoxon signed-rank test, 12 months vs freshChapman et al. 2019, Pract Lab MedResults, paragraph 4
urine, alkalised (NaOH to pH > 8), Sarstedt 2 ml polypropylene micro tube (P72.609.001)Storagenot-stated−20 °C0 monthschanged at the first time point, changed by 12 monthsWilcoxon signed-rank test, 12 months vs freshChapman et al. 2019, Pract Lab MedResults, paragraph 4; Fig. 1C
urine, boric acid precoated container (Sarstedt P51.9923.820), then Sarstedt 2 ml polypropylene micro tube (P72.609.001)Storagenot-stated−20 °C0 monthschanged at the first time point, changed by 12 monthsWilcoxon signed-rank test, 12 months vs freshChapman et al. 2019, Pract Lab MedResults, paragraph 4; Fig. 1C
urine, pepstatin 1 μg/ml added, Sarstedt 2 ml polypropylene micro tube (P72.609.001)Storagenot-stated−20 °C0 monthschanged at the first time point, changed by 12 monthsWilcoxon signed-rank test, 12 months vs freshChapman et al. 2019, Pract Lab MedResults, paragraph 4; Fig. 1C
urine, Sarstedt 1.5 ml low protein binding micro tube (P72.706.600)Storagenot-stated−20 °C0 monthschanged at the first time point, changed by 12 monthsWilcoxon signed-rank test, 12 months vs freshChapman et al. 2019, Pract Lab MedResults, paragraph 4; Fig. 1C
urine, Sarstedt 2 ml polypropylene micro tube (P72.609.001), untreatedStoragenot-stated−80 °C12 monthsstable for the longest time testedWilcoxon signed-rank test, 12 months vs freshChapman et al. 2019, Pract Lab MedResults, paragraph 4
Urine Creatinine8 findings
Specimen, tubeConditionTemperatureStable forCriterionStudy
urine, Sarstedt 2 ml polypropylene micro tube (P72.609.001), untreatedStoragenot-stated−20 °C0 monthschanged at the first time point, changed by 12 monthsWilcoxon signed-rank test, 12 months vs freshChapman et al. 2019, Pract Lab MedResults, paragraph 3
urine, Sarstedt 2 ml polypropylene micro tube (P72.609.001), untreatedStoragenot-stated−20 °C12 monthsstable for the longest time testedchange < 10% not clinically significantChapman et al. 2019, Pract Lab MedResults, paragraph 3
urine, alkalised (NaOH to pH > 8), Sarstedt 2 ml polypropylene micro tube (P72.609.001)Storagenot-stated−20 °C0 monthschanged at the first time point, changed by 12 monthsWilcoxon signed-rank test, 12 months vs freshChapman et al. 2019, Pract Lab MedResults, paragraph 3
urine, alkalised (NaOH to pH > 8), Sarstedt 2 ml polypropylene micro tube (P72.609.001)Storagenot-stated−20 °C0 monthschanged at the first time point, changed by 12 monthschange < 10% not clinically significantChapman et al. 2019, Pract Lab MedResults, paragraph 3
urine, Sarstedt 2 ml polypropylene micro tube (P72.609.001), untreatedStoragenot-stated−80 °C0 monthschanged at the first time point, changed by 12 monthsWilcoxon signed-rank test, 12 months vs freshChapman et al. 2019, Pract Lab MedResults, paragraph 3
urine, Sarstedt 2 ml polypropylene micro tube (P72.609.001), untreatedStoragenot-stated−80 °C12 monthsstable for the longest time testedchange < 10% not clinically significantChapman et al. 2019, Pract Lab MedResults, paragraph 3
urine, alkalised (NaOH to pH > 8), Sarstedt 2 ml polypropylene micro tube (P72.609.001)Storagenot-stated−80 °C0 monthschanged at the first time point, changed by 12 monthsWilcoxon signed-rank test, 12 months vs freshChapman et al. 2019, Pract Lab MedResults, paragraph 3
urine, alkalised (NaOH to pH > 8), Sarstedt 2 ml polypropylene micro tube (P72.609.001)Storagenot-stated−80 °C12 monthsstable for the longest time testedchange < 10% not clinically significantChapman et al. 2019, Pract Lab MedResults, paragraph 3
Vitamin A (Retinol)5 findings
Specimen, tubeConditionTemperatureStable forCriterionStudy
whole-blood-clotted, BD Vacutainer SSTBefore spinninguncentrifuged> 30 °C (logger 38-40 °C)12 hthen changed, changed by 24 hmean % deviation from baseline within total change limit TCL = sqrt((2.77 CVa)^2 + (0.5 CVb)^2), CVb from Ricos (Table 2)Abraham et al. 2019, Biochem Med (Zagreb)Results, paragraph 1 (Figure 3a)
whole-blood-clotted, Sarstedt Monovette serumBefore spinninguncentrifugedchilled in transit (insulated box, 2 frozen cool packs; temperature not recorded)24 hstable for the longest time tested% difference within desirable quality specification for bias (Fraser, from intra- and interindividual variation)Jones et al. 2021, J NutrTable 2, row 'Retinol', column Serum
whole-blood-edta, Sarstedt Monovette EDTABefore spinninguncentrifugedchilled in transit (insulated box, 2 frozen cool packs; temperature not recorded)24 hstable for the longest time tested% difference within desirable quality specification for bias (Fraser, from intra- and interindividual variation)Jones et al. 2021, J NutrTable 2, row 'Retinol', column EDTA
whole-blood-heparin, BD Vacutainer lithium heparinBefore spinninguncentrifugedcold packs in insulated box; 9 (2) °C on arrival24 hstable for the longest time tested% difference within desirable quality specification (from intra- and interindividual variation)Jones et al. 2024, J NutrTable 2, row 'Retinol, μmol/L'; Table 1
whole-blood-heparin, Sarstedt Monovette lithium heparin, trace-element freeBefore spinninguncentrifugedchilled in transit (insulated box, 2 frozen cool packs; temperature not recorded)24 hstable for the longest time tested% difference within desirable quality specification for bias (Fraser, from intra- and interindividual variation)Jones et al. 2021, J NutrTable 2, row 'Retinol', column LH
Vitamin B12 (Cobalamin)11 findings
Specimen, tubeConditionTemperatureStable forCriterionStudy
serum, BD Vacutainer SSTFreeze–thawseparated−70 °C, 1 month; thawed at room temperature1 cyclestable for the longest time testedmean % change (bias) from fresh baseline within total allowable error (TEa, desirable bias)Abraham et al. 2021, J Lab PhysiciansTable 2 and Table 3, row 'Vitamin B12 (pg/mL)'
serum, serum (tube not stated)Storageseparated−70 °C for the first year, −80 °C thereafter0 yearschanged at the first time point, changed by 1 yeart-test (two-sample, equal variances) vs t=0, P < 0.05Jansen et al. 2018, J Nutr MetabResults, sections 3.1 and 3.3
whole-blood-clotted, BD Vacutainer serum, no gel (ref 368815)Before spinninguncentrifugedRT10 hstable for the longest time tested|mean % deviation 10 h vs baseline| > laboratory bias goal (Table II; mostly biological-variation based, some empirically modified)Henriksen et al. 2014, Scand J Clin Lab InvestTable II, row 'B12'
whole-blood-clotted, BD Vacutainer serum, no gel (ref 368815)Before spinninguncentrifugedRT10 hstable for the longest time testedCV of paired baseline/10 h results > goal CV (Table II)Henriksen et al. 2014, Scand J Clin Lab InvestTable II, row 'B12'
whole-blood-clotted, BD Vacutainer SSTBefore spinninguncentrifugedRT24 hstable for the longest time testedmean % deviation from baseline within total change limit TCL = sqrt((2.77 CVa)^2 + (0.5 CVb)^2), CVb from Ricos (Table 2)Abraham et al. 2019, Biochem Med (Zagreb)Discussion, paragraph 1 (Results paragraphs 1-3; Figure 3)
whole-blood-clotted, BD Vacutainer trace element-free serum tubeBefore spinninguncentrifugedcold packs in insulated box; 9 (2) °C on arrival24 hstable for the longest time tested% difference within desirable quality specification (from intra- and interindividual variation)Jones et al. 2024, J NutrTable 2, row 'Vitamin B12, ng/L'; Table 1
whole-blood-clotted, BD Vacutainer SSTBefore spinninguncentrifuged22-30 °C (logger 24-26 °C)24 hstable for the longest time testedmean % deviation from baseline within total change limit TCL = sqrt((2.77 CVa)^2 + (0.5 CVb)^2), CVb from Ricos (Table 2)Abraham et al. 2019, Biochem Med (Zagreb)Discussion, paragraph 1 (Results paragraphs 1-3; Figure 3)
whole-blood-clotted, BD Vacutainer SSTBefore spinninguncentrifuged> 30 °C (logger 38-40 °C)24 hstable for the longest time testedmean % deviation from baseline within total change limit TCL = sqrt((2.77 CVa)^2 + (0.5 CVb)^2), CVb from Ricos (Table 2)Abraham et al. 2019, Biochem Med (Zagreb)Discussion, paragraph 1 (Results paragraphs 1-3; Figure 3)
whole-blood-clotted, Sarstedt Monovette serumBefore spinninguncentrifugedchilled in transit (insulated box, 2 frozen cool packs; temperature not recorded)24 hstable for the longest time tested% difference within desirable quality specification for bias (Fraser, from intra- and interindividual variation)Jones et al. 2021, J NutrTable 2, row 'Vitamin B-12', column Serum
whole-blood-edta, Sarstedt Monovette EDTABefore spinninguncentrifugedchilled in transit (insulated box, 2 frozen cool packs; temperature not recorded)24 hstable for the longest time tested% difference within desirable quality specification for bias (Fraser, from intra- and interindividual variation)Jones et al. 2021, J NutrTable 2, row 'Vitamin B-12', column EDTA
whole-blood-heparin, Sarstedt Monovette lithium heparin, trace-element freeBefore spinninguncentrifugedchilled in transit (insulated box, 2 frozen cool packs; temperature not recorded)24 hstable for the longest time tested% difference within desirable quality specification for bias (Fraser, from intra- and interindividual variation)Jones et al. 2021, J NutrTable 2, row 'Vitamin B-12', column LH
Vitamin B6 (Pyridoxal-5-Phosphate)4 findings
Specimen, tubeConditionTemperatureStable forCriterionStudy
whole-blood-clotted, Sarstedt Monovette serumBefore spinninguncentrifugedchilled in transit (insulated box, 2 frozen cool packs; temperature not recorded)0 hchanged at the first time point, changed by 24 h% difference within desirable quality specification for bias (Fraser, from intra- and interindividual variation)Jones et al. 2021, J NutrTable 2, row 'PLP (vitamin B-6)', column Serum
whole-blood-edta, Sarstedt Monovette EDTABefore spinninguncentrifugedchilled in transit (insulated box, 2 frozen cool packs; temperature not recorded)24 hstable for the longest time tested% difference within desirable quality specification for bias (Fraser, from intra- and interindividual variation)Jones et al. 2021, J NutrTable 2, row 'PLP (vitamin B-6)', column EDTA
whole-blood-heparin, BD Vacutainer lithium heparinBefore spinninguncentrifugedcold packs in insulated box; 9 (2) °C on arrival24 hstable for the longest time tested% difference within desirable quality specification (from intra- and interindividual variation)Jones et al. 2024, J NutrTable 2, row 'Vitamin B6 (PLP), nmol/L'; Table 1
whole-blood-heparin, Sarstedt Monovette lithium heparin, trace-element freeBefore spinninguncentrifugedchilled in transit (insulated box, 2 frozen cool packs; temperature not recorded)24 hstable for the longest time tested% difference within desirable quality specification for bias (Fraser, from intra- and interindividual variation)Jones et al. 2021, J NutrTable 2, row 'PLP (vitamin B-6)', column LH
Vitamin C (Ascorbic Acid)8 findings
Specimen, tubeConditionTemperatureStable forCriterionStudy
plasma-li-heparin, BD lithium heparin vacutainer (cat. 367886); plasma in 1.7 mL microfuge tubes, non-stabilizedStorageseparated−80 °C26 weeksstable for the longest time testedPaired t-test P≤0.05, difference-plot 95% CI through zero, or Passing–Bablok slope CI excluding 1.00Collie et al. 2021, Ann Lab MedResults, storage experiments
plasma-li-heparin, BD lithium heparin vacutainer (cat. 367886); plasma in 1.7 mL microfuge tubes, non-stabilizedStorageseparated−80 °C26 weeksstable for the longest time testedRCPAQAP allowable performance specifications (±25% for results ≥36.0 µmol/L)Collie et al. 2021, Ann Lab MedResults, storage experiments; Fig. 2A
plasma-li-heparin, BD lithium heparin vacutainer (cat. 367886); plasma + PCA/DTPA 1:1 (v:v) in 1.7 mL microfuge tubesStorageseparated−80 °C26 weeksstable for the longest time testedPaired t-test P≤0.05, difference-plot 95% CI through zero, or Passing–Bablok slope CI excluding 1.00Collie et al. 2021, Ann Lab MedResults, storage experiments
plasma-li-heparin, BD lithium heparin vacutainer (cat. 367886); plasma + PCA/DTPA 1:1 (v:v) in 1.7 mL microfuge tubesStorageseparated−80 °C26 weeksstable for the longest time testedRCPAQAP allowable performance specifications (±25% for results ≥36.0 µmol/L)Collie et al. 2021, Ann Lab MedResults, storage experiments; Fig. 2B
whole-blood-clotted, Sarstedt Monovette serumBefore spinninguncentrifugedchilled in transit (insulated box, 2 frozen cool packs; temperature not recorded)24 hstable for the longest time tested% difference within desirable quality specification for bias (Fraser, from intra- and interindividual variation)Jones et al. 2021, J NutrTable 2, row 'Vitamin C', column Serum
whole-blood-edta, Sarstedt Monovette EDTABefore spinninguncentrifugedchilled in transit (insulated box, 2 frozen cool packs; temperature not recorded)0 hchanged at the first time point, changed by 24 h% difference within desirable quality specification for bias (Fraser, from intra- and interindividual variation)Jones et al. 2021, J NutrTable 2, row 'Vitamin C', column EDTA
whole-blood-heparin, BD Vacutainer lithium heparinBefore spinninguncentrifugedcold packs in insulated box; 9 (2) °C on arrival24 hstable for the longest time tested% difference within desirable quality specification (from intra- and interindividual variation)Jones et al. 2024, J NutrTable 2, row 'Vitamin C, μmol/L'; Table 1
whole-blood-heparin, Sarstedt Monovette lithium heparin, trace-element freeBefore spinninguncentrifugedchilled in transit (insulated box, 2 frozen cool packs; temperature not recorded)24 hstable for the longest time tested% difference within desirable quality specification for bias (Fraser, from intra- and interindividual variation)Jones et al. 2021, J NutrTable 2, row 'Vitamin C', column LH
Vitamin E (Alpha-Tocopherol)4 findings
Specimen, tubeConditionTemperatureStable forCriterionStudy
whole-blood-clotted, Sarstedt Monovette serumBefore spinninguncentrifugedchilled in transit (insulated box, 2 frozen cool packs; temperature not recorded)24 hstable for the longest time tested% difference within desirable quality specification for bias (Fraser, from intra- and interindividual variation)Jones et al. 2021, J NutrTable 2, row 'α-tocopherol', column Serum
whole-blood-edta, Sarstedt Monovette EDTABefore spinninguncentrifugedchilled in transit (insulated box, 2 frozen cool packs; temperature not recorded)24 hstable for the longest time tested% difference within desirable quality specification for bias (Fraser, from intra- and interindividual variation)Jones et al. 2021, J NutrTable 2, row 'α-tocopherol', column EDTA
whole-blood-heparin, BD Vacutainer lithium heparinBefore spinninguncentrifugedcold packs in insulated box; 9 (2) °C on arrival24 hstable for the longest time tested% difference within desirable quality specification (from intra- and interindividual variation)Jones et al. 2024, J NutrTable 2, row 'α-Tocopherol, μmol/L'; Table 1
whole-blood-heparin, Sarstedt Monovette lithium heparin, trace-element freeBefore spinninguncentrifugedchilled in transit (insulated box, 2 frozen cool packs; temperature not recorded)24 hstable for the longest time tested% difference within desirable quality specification for bias (Fraser, from intra- and interindividual variation)Jones et al. 2021, J NutrTable 2, row 'α-tocopherol', column LH
Von Willebrand factor activity5 findings
Specimen, tubeConditionTemperatureStable forCriterionStudy
plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquotsFreeze–thawseparated−20 °C1 cyclelimit stated by the authorsmax allowable mean deviation from t0: ±10%Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability'
plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquotsFreeze–thawseparated−80 °C3 cycleslimit stated by the authorsmax allowable mean deviation from t0: ±10%Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability'
plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquotsStorageseparated−20 °C12 weekslimit stated by the authorsmax allowable mean deviation from t0: ±10%Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability'
plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquotsStorageseparated−80 °C3 monthslimit stated by the authorsmax allowable mean deviation from t0: ±10%Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability'
plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquotsStorageseparatedRT24 hlimit stated by the authorsmax allowable mean deviation from t0: ±10%Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability'
White Blood Cell Count21 findings
Specimen, tubeConditionTemperatureStable forCriterionStudy
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C48 hstable for the longest time testedHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'WBC', XN-9000, 4 °C
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C48 hstable for the longest time testedBias% vs optimal bias (OP-Bias%, from biological variation; Table 1)Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'WBC', XN-9000, 4 °C
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C48 hstable for the longest time testedHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'WBC', BC-6800, 4 °C
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C48 hstable for the longest time testedBias% vs optimal bias (OP-Bias%, from biological variation; Table 1)Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'WBC', BC-6800, 4 °C
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C8 hstable for the longest time testedHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'WBC', XN-9000, 4 °C
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C8 hstable for the longest time testedBias% vs optimal bias (OP-Bias%, from biological variation; Table 1)Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'WBC', XN-9000, 4 °C
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C8 hstable for the longest time testedHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'WBC', BC-6800, 4 °C
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C4 hthen changed, changed by 8 hBias% vs optimal bias (OP-Bias%, from biological variation; Table 1)Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'WBC', BC-6800, 4 °C
whole-blood-edta, K3-EDTA (Becton Dickinson)Storageuncentrifuged2-8 °C8 hstable for the longest time testedBias% vs critical difference (CD%, reference change value; Table 1)Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'WBC', BC-6800, 4 °C
whole-blood-edta, K2-EDTA (BD Vacutainer)Storageuncentrifuged2-8 °C48 hstable for the longest time testedRepeated-measures one-way ANOVA with Tukey post hoc vs 0 h, p < 0.05 (* in table)Unalli et al. 2021, J Med BiochemResults paragraphs 1-2 and Discussion paragraph 1 (Figures 1-2; numbers in figures only)
whole-blood-edta, K3-EDTA (Becton Dickinson)StorageuncentrifugedRT48 hstable for the longest time testedHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'WBC', XN-9000, RT
whole-blood-edta, K3-EDTA (Becton Dickinson)StorageuncentrifugedRT48 hstable for the longest time testedBias% vs optimal bias (OP-Bias%, from biological variation; Table 1)Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'WBC', XN-9000, RT
whole-blood-edta, K3-EDTA (Becton Dickinson)StorageuncentrifugedRT48 hstable for the longest time testedHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'WBC', BC-6800, RT
whole-blood-edta, K3-EDTA (Becton Dickinson)StorageuncentrifugedRT48 hstable for the longest time testedBias% vs optimal bias (OP-Bias%, from biological variation; Table 1)Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'WBC', BC-6800, RT
whole-blood-edta, K3-EDTA (Becton Dickinson)StorageuncentrifugedRT8 hstable for the longest time testedHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'WBC', XN-9000, RT
whole-blood-edta, K3-EDTA (Becton Dickinson)StorageuncentrifugedRT8 hstable for the longest time testedBias% vs optimal bias (OP-Bias%, from biological variation; Table 1)Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'WBC', XN-9000, RT
whole-blood-edta, K3-EDTA (Becton Dickinson)StorageuncentrifugedRT8 hstable for the longest time testedHodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time')Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'WBC', BC-6800, RT
whole-blood-edta, K3-EDTA (Becton Dickinson)StorageuncentrifugedRT8 hstable for the longest time testedBias% vs optimal bias (OP-Bias%, from biological variation; Table 1)Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'WBC', BC-6800, RT
whole-blood-edta, K2-EDTA (BD Vacutainer)StorageuncentrifugedRT24 hthen changed, changed by 48 hRepeated-measures one-way ANOVA with Tukey post hoc vs 0 h, p < 0.05 (* in table)Unalli et al. 2021, J Med BiochemResults, paragraph 1 (Figure 1B)
whole-blood-edta, K2EDTA (Becton Dickinson)Storageuncentrifuged33 °C24 hthen changed, changed by 48 hUnacceptable if median % difference > Biasmax (biological variation) and Holm-adjusted Wilcoxon P < 0.05; stable if equivalence shown (two one-tailed Wilcoxon tests vs Biasmax, Pholm < 0.05)Jain et al. 2018, Biochem Med (Zagreb)Results, last paragraph (equivalence); Table 2, row 'WBC'
whole-blood-edta, K2-EDTA (BD Vacutainer)Storageuncentrifuged10 °C48 hstable for the longest time testedRepeated-measures one-way ANOVA with Tukey post hoc vs 0 h, p < 0.05 (* in table)Unalli et al. 2021, J Med BiochemResults paragraphs 1-2 and Discussion paragraph 1 (Figures 1-2; numbers in figures only)
Zinc5 findings
Specimen, tubeConditionTemperatureStable forCriterionStudy
whole-blood-clotted, BD Vacutainer metal-free SSTBefore spinninguncentrifugedRT12 hthen changed, changed by 24 hmean % deviation from baseline within total change limit TCL = sqrt((2.77 CVa)^2 + (0.5 CVb)^2), CVb from Ricos (Table 2)Abraham et al. 2019, Biochem Med (Zagreb)Results, paragraph 1 (Figure 3e)
whole-blood-clotted, BD Vacutainer trace element-free serum tubeBefore spinninguncentrifugedcold packs in insulated box; 9 (2) °C on arrival24 hstable for the longest time tested% difference within desirable quality specification (from intra- and interindividual variation)Jones et al. 2024, J NutrTable 2, row 'Zinc, μg/L'; Table 1
whole-blood-clotted, Sarstedt Monovette serumBefore spinninguncentrifugedchilled in transit (insulated box, 2 frozen cool packs; temperature not recorded)24 hstable for the longest time tested% difference within desirable quality specification for bias (Fraser, from intra- and interindividual variation)Jones et al. 2021, J NutrTable 2, row 'Zinc', column Serum
whole-blood-edta, Sarstedt Monovette EDTABefore spinninguncentrifugedchilled in transit (insulated box, 2 frozen cool packs; temperature not recorded)24 hstable for the longest time tested% difference within desirable quality specification for bias (Fraser, from intra- and interindividual variation)Jones et al. 2021, J NutrTable 2, row 'Zinc', column EDTA
whole-blood-heparin, Sarstedt Monovette lithium heparin, trace-element freeBefore spinninguncentrifugedchilled in transit (insulated box, 2 frozen cool packs; temperature not recorded)0 hchanged at the first time point, changed by 24 h% difference within desirable quality specification for bias (Fraser, from intra- and interindividual variation)Jones et al. 2021, J NutrTable 2, row 'Zinc', column LH

How it was built

Each row is one finding from a study read in full: specimen, tube, processing, temperature, the longest time the analyte met the study's stability criterion, and that criterion. Studies use different criteria, so a shorter time in one study and a longer one in another can both be right. Where studies disagree, all rows are kept; the summary file gives the shortest time reported for each combination and how many studies agree.

Download

Code, data and tests: github.com/caraulani/sample-stability

Archived on Zenodo: doi.org/10.5281/zenodo.24543300

Cite

LifeFrom (2026). How long a sample keeps, version 1.0.0. lifefrom.ai/research/open-data/sample-stability. https://doi.org/10.5281/zenodo.24543300