How long a blood sample keeps
How long a sample can wait before it's measured depends on the analyte, the tube, whether it was spun and separated, and the temperature. This collects what published studies found: 160 analytes, 1434 findings from 51 studies, each with the study's own test for “stable” and the table it came from.
25-Hydroxyvitamin D (25-OH D)34 findings
| Specimen, tube | Condition | Temperature | Stable for | Criterion | Study |
|---|---|---|---|---|---|
| plasma-edta, BD Vacutainer EDTA (2 mL) | Storageseparated | −20 °C | 7 daysstable for the longest time tested | Friedman repeated measures ANOVA vs zero time, P <= 0.05 | Colak et al. 2013, Biochem Med (Zagreb)Table 2 |
| plasma-edta, BD Vacutainer EDTA (2 mL) | Storageseparated | −80 °C | 3 monthsstable for the longest time tested | Friedman repeated measures ANOVA vs zero time, P <= 0.05 | Colak et al. 2013, Biochem Med (Zagreb)Table 2 |
| plasma-edta, BD Vacutainer EDTA (2 mL) | Storageseparated | 2-8 °C | 24 hstable for the longest time tested | Friedman repeated measures ANOVA vs zero time, P <= 0.05 | Colak et al. 2013, Biochem Med (Zagreb)Table 1 |
| plasma-edta, BD Vacutainer EDTA (2 mL) | Storageseparated | RT | 4 hstable for the longest time tested | Friedman repeated measures ANOVA vs zero time, P <= 0.05 | Colak et al. 2013, Biochem Med (Zagreb)Table 1 |
| serum, BD Vacutainer SST | Freeze–thawseparated | −70 °C, 1 month; thawed at room temperature | 1 cyclestable for the longest time tested | mean % change (bias) from fresh baseline within total allowable error (TEa, desirable bias) | Abraham et al. 2021, J Lab PhysiciansTable 2 and Table 3, row 'Vitamin D (ng/mL)' |
| serum, not stated | Storageseparated | −20 °C | 7 monthsstable for the longest time tested | Friedman test P < 0.05 with Bonferroni-adjusted post hoc vs baseline (#) | Ko et al. 2025, Sci RepTable 1, overall, −20 °C |
| serum, not stated | Storageseparated | −20 °C | 7 monthsstable for the longest time tested | Friedman test P < 0.05 with Bonferroni-adjusted post hoc vs baseline (#) | Ko et al. 2025, Sci RepTable 2, overall, −20 °C |
| serum, BD Vacutainer serum separator tube with clot activator (8.5 mL) | Storageseparated | −20 °C | 7 daysstable for the longest time tested | Friedman repeated measures ANOVA vs zero time, P <= 0.05 | Colak et al. 2013, Biochem Med (Zagreb)Table 2 |
| serum, BD plain (red top) tube | Storageseparated | −20 °C | 1 daysthen changed, changed by 2 days | paired t test vs day 0, P < .05 | Borai et al. 2020, J Clin Lab AnalTable 2, non-vortexed |
| serum, BD plain (red top) tube | Storageseparated | −20 °C | 0 dayschanged at the first time point, changed by 1 days | paired t test vs day 0, P < .05 | Borai et al. 2020, J Clin Lab AnalTable 2, vortexed |
| serum, BD plain (red top) tube | Storageseparated | −20 °C | 30 daysstable for the longest time tested | % difference vs day 0 > RCV 18% (Brescia, LIAISON) | Borai et al. 2020, J Clin Lab AnalDiscussion, paragraph 5; Table 2 |
| serum, BD plain (red top) tube | Storageseparated | −20 °C | 30 daysstable for the longest time tested | % difference vs day 0 > TEa 46% (DEQAS) | Borai et al. 2020, J Clin Lab AnalDiscussion, paragraph 6; Table 2 |
| serum, BD plain (red top) tube | Storageseparated | −20 °C | 30 daysstable for the longest time tested | % difference vs day 0 > RCV 18% (Brescia, LIAISON) | Borai et al. 2020, J Clin Lab AnalDiscussion, paragraph 5; Table 2 |
| serum, BD plain (red top) tube | Storageseparated | −20 °C | 30 daysstable for the longest time tested | % difference vs day 0 > TEa 46% (DEQAS) | Borai et al. 2020, J Clin Lab AnalDiscussion, paragraph 6; Table 2 |
| serum, not stated | Storageseparated | −80 °C | 2 weeksthen changed, changed by 7 months | Friedman test P < 0.05 with Bonferroni-adjusted post hoc vs baseline (#) | Ko et al. 2025, Sci RepTable 1, overall, −80 °C |
| serum, not stated | Storageseparated | −80 °C | 0 weekschanged at the first time point, changed by 2 weeks | Friedman test P < 0.05 with Bonferroni-adjusted post hoc vs baseline (#) | Ko et al. 2025, Sci RepTable 2, overall, −80 °C |
| serum, BD Vacutainer serum separator tube with clot activator (8.5 mL) | Storageseparated | −80 °C | 3 monthsstable for the longest time tested | Friedman repeated measures ANOVA vs zero time, P <= 0.05 | Colak et al. 2013, Biochem Med (Zagreb)Table 2 |
| serum, BD Vacutainer SST (5 mL, REF 366566) | Storageseparated | −80 °C | 48 hstable for the longest time tested | paired t-test vs immediate analysis, P < 0.05 | Yu et al. 2016, PLoS OneDiscussion, paragraph 3 |
| serum, VACUETTE no-additive (4 mL, REF 454001) | Storageseparated | −80 °C | 48 hstable for the longest time tested | paired t-test vs immediate analysis, P < 0.05 | Yu et al. 2016, PLoS OneDiscussion, paragraph 3 |
| serum, VACUETTE serum clot activator and gel (REF 454067) | Storageseparated | −80 °C | 48 hstable for the longest time tested | paired t-test vs immediate analysis, P < 0.05 | Yu et al. 2016, PLoS OneDiscussion, paragraph 3 |
| serum, IMPROVEVACUTER serum clot activator and gel (REF U222538) | Storageseparated | −80 °C | 0 hchanged at the first time point, changed by 48 h | paired t-test vs immediate analysis, P < 0.05 | Yu et al. 2016, PLoS OneResults, 'Evaluation of the effects of storage temperature', paragraph 1 |
| serum, IMPROVEVACUTER serum clot activator and gel (REF U222538) | Storageseparated | −80 °C | 48 hstable for the longest time tested | no significant difference vs immediate analysis (test not named for this set; P < 0.05) | Yu et al. 2016, PLoS OneDiscussion, paragraph 3 |
| serum, BD Vacutainer serum separator tube with clot activator (8.5 mL) | Storageseparated | 2-8 °C | 24 hstable for the longest time tested | Friedman repeated measures ANOVA vs zero time, P <= 0.05 | Colak et al. 2013, Biochem Med (Zagreb)Table 1 |
| serum, BD Vacutainer SST (5 mL, REF 366566) | Storageseparated | 2-8 °C | 48 hstable for the longest time tested | paired t-test vs immediate analysis, P < 0.05 | Yu et al. 2016, PLoS OneResults, 'Evaluation of the effects of storage temperature', paragraph 1 |
| serum, IMPROVEVACUTER serum clot activator and gel (REF U222538) | Storageseparated | 2-8 °C | 48 hstable for the longest time tested | paired t-test vs immediate analysis, P < 0.05 | Yu et al. 2016, PLoS OneResults, 'Evaluation of the effects of storage temperature', paragraph 1 |
| serum, VACUETTE no-additive (4 mL, REF 454001) | Storageseparated | 2-8 °C | 48 hstable for the longest time tested | paired t-test vs immediate analysis, P < 0.05 | Yu et al. 2016, PLoS OneResults, 'Evaluation of the effects of storage temperature', paragraph 1 |
| serum, VACUETTE serum clot activator and gel (REF 454067) | Storageseparated | 2-8 °C | 48 hstable for the longest time tested | paired t-test vs immediate analysis, P < 0.05 | Yu et al. 2016, PLoS OneResults, 'Evaluation of the effects of storage temperature', paragraph 1 |
| serum, IMPROVEVACUTER serum clot activator and gel (REF U222538) | Storageseparated | 2-8 °C | 48 hstable for the longest time tested | no significant difference vs immediate analysis (test not named for this set; P < 0.05) | Yu et al. 2016, PLoS OneDiscussion, paragraph 3 |
| serum, BD Vacutainer serum separator tube with clot activator (8.5 mL) | Storageseparated | RT | 4 hstable for the longest time tested | Friedman repeated measures ANOVA vs zero time, P <= 0.05 | Colak et al. 2013, Biochem Med (Zagreb)Table 1 |
| whole-blood-clotted, BD Vacutainer trace element-free serum tube | Before spinninguncentrifuged | cold packs in insulated box; 9 (2) °C on arrival | 24 hstable for the longest time tested | % difference within desirable quality specification (from intra- and interindividual variation) | Jones et al. 2024, J NutrTable 2, row '25-Hydroxyvitamin D, nmol/L'; Table 1 |
| whole-blood-clotted, BD Vacutainer SST | Before spinninguncentrifuged | > 30 °C (logger 38-40 °C) | 6 hthen changed, changed by 12 h | mean % deviation from baseline within total change limit TCL = sqrt((2.77 CVa)^2 + (0.5 CVb)^2), CVb from Ricos (Table 2) | Abraham et al. 2019, Biochem Med (Zagreb)Results, paragraph 1 (Figure 3b) |
| whole-blood-clotted, Sarstedt Monovette serum | Before spinninguncentrifuged | chilled in transit (insulated box, 2 frozen cool packs; temperature not recorded) | 24 hstable for the longest time tested | % difference within desirable quality specification for bias (Fraser, from intra- and interindividual variation) | Jones et al. 2021, J NutrTable 2, row '25-hydroxyvitamin D', column Serum |
| whole-blood-edta, Sarstedt Monovette EDTA | Before spinninguncentrifuged | chilled in transit (insulated box, 2 frozen cool packs; temperature not recorded) | 24 hstable for the longest time tested | % difference within desirable quality specification for bias (Fraser, from intra- and interindividual variation) | Jones et al. 2021, J NutrTable 2, row '25-hydroxyvitamin D', column EDTA |
| whole-blood-heparin, Sarstedt Monovette lithium heparin, trace-element free | Before spinninguncentrifuged | chilled in transit (insulated box, 2 frozen cool packs; temperature not recorded) | 24 hstable for the longest time tested | % difference within desirable quality specification for bias (Fraser, from intra- and interindividual variation) | Jones et al. 2021, J NutrTable 2, row '25-hydroxyvitamin D', column LH |
4-Pyridoxic acid4 findings
| Specimen, tube | Condition | Temperature | Stable for | Criterion | Study |
|---|---|---|---|---|---|
| whole-blood-clotted, Sarstedt Monovette serum | Before spinninguncentrifuged | chilled in transit (insulated box, 2 frozen cool packs; temperature not recorded) | 24 hstable for the longest time tested | % difference within desirable quality specification for bias (Fraser, from intra- and interindividual variation) | Jones et al. 2021, J NutrTable 2, row 'PA (vitamin B-6)', column Serum |
| whole-blood-edta, Sarstedt Monovette EDTA | Before spinninguncentrifuged | chilled in transit (insulated box, 2 frozen cool packs; temperature not recorded) | 24 hstable for the longest time tested | % difference within desirable quality specification for bias (Fraser, from intra- and interindividual variation) | Jones et al. 2021, J NutrTable 2, row 'PA (vitamin B-6)', column EDTA |
| whole-blood-heparin, BD Vacutainer lithium heparin | Before spinninguncentrifuged | cold packs in insulated box; 9 (2) °C on arrival | 24 hstable for the longest time tested | % difference within desirable quality specification (from intra- and interindividual variation) | Jones et al. 2024, J NutrTable 2, row 'Vitamin B6 (PA), nmol/L'; Table 1 |
| whole-blood-heparin, Sarstedt Monovette lithium heparin, trace-element free | Before spinninguncentrifuged | chilled in transit (insulated box, 2 frozen cool packs; temperature not recorded) | 24 hstable for the longest time tested | % difference within desirable quality specification for bias (Fraser, from intra- and interindividual variation) | Jones et al. 2021, J NutrTable 2, row 'PA (vitamin B-6)', column LH |
Activated Partial Thromboplastin Time10 findings
| Specimen, tube | Condition | Temperature | Stable for | Criterion | Study |
|---|---|---|---|---|---|
| plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquots | Freeze–thawseparated | −80 °C | 1 cyclelimit stated by the authors | max allowable mean deviation from t0: <100 s: ±10%; ≥100 s: ±20% | Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability' |
| plasma-citrate, BD Vacutainer 2.7 mL, 3.2% buffered sodium citrate; PPP aliquoted into microcentrifuge tubes | Storageseparated | −20 °C | 36 hstable for the longest time tested | change of more than ±10% of mean from baseline | Patil et al. 2022, CureusTable 2 |
| plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquots | Storageseparated | −80 °C | 11 monthslimit stated by the authors | max allowable mean deviation from t0: <100 s: ±10%; ≥100 s: ±20% | Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability' |
| plasma-citrate, BD 0.109 M sodium citrate (9:1); plasma aliquoted into nonactivating Eppendorf tubes | Storageseparated | 2-8 °C | 12 hthen changed, changed by 24 h | mean % change from baseline >10%, or >25% of samples with individual change >10% (van Geest-Daalderop) | Feng et al. 2014, Sci RepTable 2, 'Acceptable Time (h)' |
| plasma-citrate, BD Vacutainer 2.7 mL, 3.2% buffered sodium citrate; PPP aliquoted into microcentrifuge tubes | Storageseparated | 2-8 °C | 0 hchanged at the first time point, changed by 12 h | change of more than ±10% of mean from baseline | Patil et al. 2022, CureusTable 2 |
| plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquots | Storageseparated | RT | 2 hlimit stated by the authors | max allowable mean deviation from t0: <100 s: ±10%; ≥100 s: ±20% | Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability' |
| plasma-citrate, BD 0.109 M sodium citrate (9:1); plasma aliquoted into nonactivating Eppendorf tubes | Storageseparated | RT | 8 hthen changed, changed by 12 h | mean % change from baseline >10%, or >25% of samples with individual change >10% (van Geest-Daalderop) | Feng et al. 2014, Sci RepTable 2, 'Acceptable Time (h)' |
| plasma-citrate, BD Vacutainer 2.7 mL, 3.2% buffered sodium citrate; PPP aliquoted into microcentrifuge tubes | Storageseparated | RT | 0 hchanged at the first time point, changed by 12 h | change of more than ±10% of mean from baseline | Patil et al. 2022, CureusTable 2 |
| whole-blood-citrate, Greiner Vacuette sodium citrate, standard fill (blood–air interface) | Before spinninguncentrifuged | RT | 6 hthen changed, changed by 24 h | maximum acceptable bias = 0.375 × total biological variation (EFLM BV database, Model 2) | Storm et al. 2026, Adv HematolTable 2, 'Standard tubes' row 'APTT' |
| whole-blood-citrate, Greiner Vacuette sodium citrate, completely filled (no air) | Before spinninguncentrifuged | RT | 6 hthen changed, changed by 24 h | maximum acceptable bias = 0.375 × total biological variation (EFLM BV database, Model 2) | Storm et al. 2026, Adv HematolTable 2, 'Full tubes' row 'APTT' |
Adenosine deaminase2 findings
| Specimen, tube | Condition | Temperature | Stable for | Criterion | Study |
|---|---|---|---|---|---|
| plasma-li-heparin, Sarstedt lithium heparin, no gel separator (primary tube) | Storagecentrifuged-on-cells | 2-8 °C | 48 hstable for the longest time tested | mean % difference vs time 0 exceeds total limit of change, TLC = ±√((1.65·CVa)² + (0.5·CVb)²) (EFLM BV), here 25.1%; stability limit from instability-equation regression | Illana et al. 2024, Biochem Med (Zagreb)Table 1, row 'ADA' |
| plasma-li-heparin, Sarstedt lithium heparin, no gel separator; plasma aliquot | Storageseparated | 2-8 °C | 96 hstable for the longest time tested | mean % difference vs time 0 exceeds total limit of change, TLC = ±√((1.65·CVa)² + (0.5·CVb)²) (EFLM BV), here 25.1%; stability limit from instability-equation regression | Illana et al. 2024, Biochem Med (Zagreb)Table 2, row 'ADA' |
Adrenocorticotropic Hormone3 findings
| Specimen, tube | Condition | Temperature | Stable for | Criterion | Study |
|---|---|---|---|---|---|
| whole-blood-edta, BD K2-EDTA | Before spinninguncentrifuged | 2-8 °C | 8 hstable for the longest time tested | mean % deviation vs T0 > total change limit (TCL) 7.5% | Fraissinet et al. 2024, Biochem Med (Zagreb)Table 1; Results paragraph 1 |
| whole-blood-edta, BD K3-EDTA + aprotinin | Before spinninguncentrifuged | 2-8 °C | 4 hthen changed, changed by 6 h | mean % deviation vs T0 > total change limit (TCL) 7.5% | Fraissinet et al. 2024, Biochem Med (Zagreb)Table 1; Results paragraph 1 |
| whole-blood-edta, BD K3-EDTA + aprotinin | Before spinninguncentrifuged | RT | 2 hthen changed, changed by 4 h | mean % deviation vs T0 > total change limit (TCL) 7.5% | Fraissinet et al. 2024, Biochem Med (Zagreb)Table 1; Results paragraph 1 |
Alanine Aminotransferase (ALT)20 findings
| Specimen, tube | Condition | Temperature | Stable for | Criterion | Study |
|---|---|---|---|---|---|
| plasma-li-heparin, Sarstedt lithium heparin, no gel separator (primary tube) | Storagecentrifuged-on-cells | 2-8 °C | 48 hstable for the longest time tested | mean % difference vs time 0 exceeds total limit of change, TLC = ±√((1.65·CVa)² + (0.5·CVb)²) (EFLM BV), here 15.3%; stability limit from instability-equation regression | Illana et al. 2024, Biochem Med (Zagreb)Table 1, row 'ALT' |
| plasma-li-heparin, Sarstedt lithium heparin, no gel separator; plasma aliquot | Storageseparated | 2-8 °C | 96 hstable for the longest time tested | mean % difference vs time 0 exceeds total limit of change, TLC = ±√((1.65·CVa)² + (0.5·CVb)²) (EFLM BV), here 15.3%; stability limit from instability-equation regression | Illana et al. 2024, Biochem Med (Zagreb)Table 2, row 'ALT' |
| plasma-li-heparin, BD Vacutainer lithium heparin 6 mL (regular) | Storagenot-stated | 2-8 °C | 72 hstable for the longest time tested | mean % difference vs T0 exceeds total change limit (TCL), here ±21.75% | Dupuy et al. 2018, Biochem Med (Zagreb)Table 2, BD lithium heparin tubes, row 'ALT, U/L' |
| plasma-li-heparin, BD Barricor lithium heparin 5.5 mL (mechanical separator) | Storagecentrifuged-on-gel | 2-8 °C | 72 hstable for the longest time tested | mean % difference vs T0 exceeds total change limit (TCL), here ±21.75% | Dupuy et al. 2018, Biochem Med (Zagreb)Table 2, BD Barricor lithium heparin tubes, row 'ALT, U/L' |
| plasma-li-heparin, BD Barricor 4.5 mL (lithium heparin, mechanical separator) | Storagecentrifuged-on-gel | 2-8 °C | 7 daysstable for the longest time tested | mean % difference vs time 0 exceeds desirable imprecision based on biological variation (EFLM/Westgard), here 5.2% | Mandić et al. 2020, Biochem Med (Zagreb)Tables 3 and 4, row 'ALT, U/L', BD Barricor columns |
| serum, BD Vacutainer 8.5 mL plastic serum tube; pooled serum aliquoted into 1.5 mL Eppendorf tubes | Freeze–thawseparated | −20 °C | 10 cyclesstable for the longest time tested | unstable only if both P < 0.005 (Bonferroni) and % change from T0 > desirable bias (Westgard/Ricos), here 11.4% | Cuhadar et al. 2013, Biochem Med (Zagreb)Table 2, row 'ALT (U/L)', columns T1d-T10d |
| serum, BD Vacutainer 8.5 mL plastic serum tube; pooled serum aliquoted into 1.5 mL Eppendorf tubes | Storageseparated | −20 °C | 3 monthsstable for the longest time tested | unstable only if both P < 0.013 (Bonferroni) and % change from T0 > desirable bias (Westgard/Ricos), here 11.4% | Cuhadar et al. 2013, Biochem Med (Zagreb)Table 2, row 'ALT (U/L)', columns T1m-T3m |
| serum, BD Vacutainer serum (plain), 7.5 mL | Storageseparated | −20 °C | 30 daysstable for the longest time tested | clinically significant change: mean outside significant change limit (SCL = T1d mean ± 3.0 usual SD from 2 months QC) | Kachhawa et al. 2017, J Lab PhysiciansDiscussion, 'Enzymes' (Table 2 image-only) |
| serum, Vacuette Tube Serum Gel Separator 8 mL (Greiner, 455071) | Storagecentrifuged-on-gel | 2-8 °C | 1 weekstable for the longest time tested | % bias of median vs baseline exceeds desirable bias (Westgard/Ricos BV), here 12.0% | Cuhadar et al. 2012, Biochem Med (Zagreb)Table 3, row 'Alanine aminotransferase', gel 4°C |
| serum, Vacuette Tube Serum No Gel 9 mL (Greiner, 455092), plain | Storagecentrifuged-on-cells | 2-8 °C | 1 weekstable for the longest time tested | % bias of median vs baseline exceeds desirable bias (Westgard/Ricos BV), here 12.0% | Cuhadar et al. 2012, Biochem Med (Zagreb)Table 3, row 'Alanine aminotransferase', w/o gel 4°C |
| serum, BD Vacutainer RST 4.0 mL (thrombin clot activator, gel) | Storagecentrifuged-on-gel | 2-8 °C | 24 hstable for the longest time tested | clinically significant if mean bias vs t0 exceeds desirable bias (biological variation), here 3.0% | Kocijancic et al. 2014, Biochem Med (Zagreb)Table 3, row 'ALT' |
| serum, BD Vacutainer SST II Advance 5.0 mL (clot activator, gel) | Storagecentrifuged-on-gel | 2-8 °C | 24 hstable for the longest time tested | clinically significant if mean bias vs t0 exceeds desirable bias (biological variation), here 3.0% | Kocijancic et al. 2014, Biochem Med (Zagreb)Table 4, row 'ALT' |
| serum, BD Vacutainer RST 5 mL (thrombin clot activator, gel) | Storagecentrifuged-on-gel | 2-8 °C | 7 daysstable for the longest time tested | mean % difference vs time 0 exceeds desirable imprecision based on biological variation (EFLM/Westgard), here 5.2% | Mandić et al. 2020, Biochem Med (Zagreb)Tables 3 and 4, row 'ALT, U/L', BD RST columns |
| serum, Vacuette Tube Serum Gel Separator 8 mL (Greiner, 455071) | Storagecentrifuged-on-gel | RT | 1 weekstable for the longest time tested | % bias of median vs baseline exceeds desirable bias (Westgard/Ricos BV), here 12.0% | Cuhadar et al. 2012, Biochem Med (Zagreb)Table 3, row 'Alanine aminotransferase', gel 24°C |
| serum, Vacuette Tube Serum No Gel 9 mL (Greiner, 455092), plain | Storagecentrifuged-on-cells | RT | 72 hthen changed, changed by 1 week | % bias of median vs baseline exceeds desirable bias (Westgard/Ricos BV), here 12.0% | Cuhadar et al. 2012, Biochem Med (Zagreb)Table 3, row 'Alanine aminotransferase', w/o gel 24°C |
| serum, BD Vacutainer SST (SKU 367986), gel | Storagecentrifuged-on-gel | 30 °C (±1 °C) | 24 hthen changed, changed by 72 h | mean bias vs paired baseline within CLIA acceptance limits AND >2/3 of individual results within limits | Iacovetti et al. 2026, Pract Lab MedResults, 'Analyte stability after storage at 30 °C', paragraph 1 (Table 1 image-only) |
| serum, BD SST Microtainer (SKU 395967), gel; filled from venous blood | Storagecentrifuged-on-gel | 0 °C | 72 hstable for the longest time tested | mean bias vs paired baseline within CLIA acceptance limits AND >2/3 of individual results within limits | Iacovetti et al. 2026, Pract Lab MedResults, 'Analyte stability after separation and storage in microvolume tubes at 0 °C' (Table 2 image-only) |
| whole-blood-clotted, BD Vacutainer SST (SKU 367986), gel | Before spinninguncentrifuged | 30 °C (±1 °C) | 0 hchanged at the first time point, changed by 24 h | mean bias vs paired baseline within CLIA acceptance limits AND >2/3 of individual results within limits | Iacovetti et al. 2026, Pract Lab MedResults, 'Analyte stability after storage at 30 °C', paragraph 2 (Table 1 image-only) |
| whole-blood-heparin, BD Vacutainer lithium heparin, no gel (ref 368884) | Before spinninguncentrifuged | RT | 10 hstable for the longest time tested | |mean % deviation 10 h vs baseline| > laboratory bias goal (Table II; mostly biological-variation based, some empirically modified) | Henriksen et al. 2014, Scand J Clin Lab InvestTable II, row 'ALAT' |
| whole-blood-heparin, BD Vacutainer lithium heparin, no gel (ref 368884) | Before spinninguncentrifuged | RT | 10 hstable for the longest time tested | CV of paired baseline/10 h results > goal CV (Table II) | Henriksen et al. 2014, Scand J Clin Lab InvestTable II, row 'ALAT' |
Albumin24 findings
| Specimen, tube | Condition | Temperature | Stable for | Criterion | Study |
|---|---|---|---|---|---|
| plasma-edta, BD Vacutainer EDTA, 4 mL | Freeze–thawseparated | −20 °C | 0 cycleschanged at the first time point, changed by 1 cycles | significant change vs day 1 (repeated-measures ANOVA, Tukey HSD post hoc, p<0.05) | Gojković et al. 2024, J Med BiochemResults paragraph 3 (EDTA -20°C, Figure 3) |
| plasma-edta, BD Vacutainer EDTA, 4 mL | Storageseparated | 2-8 °C | 0 dayschanged at the first time point, changed by 1 days | significant change vs day 1 (repeated-measures ANOVA, Tukey HSD post hoc, p<0.05) | Gojković et al. 2024, J Med BiochemResults paragraph 1 (EDTA+4°C, Figure 1) |
| plasma-fluoride, BD Vacutainer sodium fluoride/potassium oxalate, 4 mL | Freeze–thawseparated | −20 °C | 0 cycleschanged at the first time point, changed by 1 cycles | significant change vs day 1 (repeated-measures ANOVA, Tukey HSD post hoc, p<0.05) | Gojković et al. 2024, J Med BiochemResults paragraph 4 (NaF -20°C, Figure 4) |
| plasma-fluoride, BD Vacutainer sodium fluoride/potassium oxalate, 4 mL | Storageseparated | 2-8 °C | 0 dayschanged at the first time point, changed by 1 days | significant change vs day 1 (repeated-measures ANOVA, Tukey HSD post hoc, p<0.05) | Gojković et al. 2024, J Med BiochemResults paragraph 2 (NaF+4°C, Figure 2) |
| plasma-li-heparin, Sarstedt lithium heparin, no gel separator (primary tube) | Storagecentrifuged-on-cells | 2-8 °C | 48 hstable for the longest time tested | mean % difference vs time 0 exceeds total limit of change, TLC = ±√((1.65·CVa)² + (0.5·CVb)²) (EFLM BV), here 4.7%; stability limit from instability-equation regression | Illana et al. 2024, Biochem Med (Zagreb)Table 1, row 'Alb' |
| plasma-li-heparin, Sarstedt lithium heparin, no gel separator; plasma aliquot | Storageseparated | 2-8 °C | 96 hstable for the longest time tested | mean % difference vs time 0 exceeds total limit of change, TLC = ±√((1.65·CVa)² + (0.5·CVb)²) (EFLM BV), here 4.7%; stability limit from instability-equation regression | Illana et al. 2024, Biochem Med (Zagreb)Table 2, row 'Alb' |
| plasma-li-heparin, BD Barricor 4.5 mL (lithium heparin, mechanical separator) | Storagecentrifuged-on-gel | 2-8 °C | 24 hthen changed, changed by 7 days | mean % difference vs time 0 exceeds desirable imprecision based on biological variation (EFLM/Westgard), here 1.3% | Mandić et al. 2020, Biochem Med (Zagreb)Tables 3 and 4, row 'Albumin, g/L', BD Barricor columns |
| serum, BD Vacutainer 8.5 mL plastic serum tube; pooled serum aliquoted into 1.5 mL Eppendorf tubes | Freeze–thawseparated | −20 °C | 7 cyclesthen changed, changed by 8 cycles | unstable only if both P < 0.005 (Bonferroni) and % change from T0 > desirable bias (Westgard/Ricos), here 1.3% | Cuhadar et al. 2013, Biochem Med (Zagreb)Table 2, row 'Albumin (g/L)', columns T1d-T10d |
| serum, BD Vacutainer SST | Freeze–thawseparated | −70 °C, 1 month; thawed at room temperature | 1 cyclestable for the longest time tested | mean % change (bias) from fresh baseline within total allowable error (TEa, desirable bias) | Abraham et al. 2021, J Lab PhysiciansTable 2 and Table 3, row 'Serum albumin (mg/dL)' |
| serum, BD Vacutainer 8.5 mL plastic serum tube; pooled serum aliquoted into 1.5 mL Eppendorf tubes | Storageseparated | −20 °C | 2 monthsthen changed, changed by 3 months | unstable only if both P < 0.013 (Bonferroni) and % change from T0 > desirable bias (Westgard/Ricos), here 1.3% | Cuhadar et al. 2013, Biochem Med (Zagreb)Table 2, row 'Albumin (g/L)', columns T1m-T3m |
| serum, BD Vacutainer serum (plain), 7.5 mL | Storageseparated | −20 °C | 30 daysstable for the longest time tested | clinically significant change: mean outside significant change limit (SCL = T1d mean ± 3.0 usual SD from 2 months QC) | Kachhawa et al. 2017, J Lab PhysiciansDiscussion, 'Protein' (Table 2 image-only) |
| serum, Vacuette Tube Serum Gel Separator 8 mL (Greiner, 455071) | Storagecentrifuged-on-gel | 2-8 °C | 24 hthen changed, changed by 30 h | % bias of median vs baseline exceeds desirable bias (Westgard/Ricos BV), here 1.3% | Cuhadar et al. 2012, Biochem Med (Zagreb)Table 3, row 'Albumin', gel 4°C |
| serum, Vacuette Tube Serum No Gel 9 mL (Greiner, 455092), plain | Storagecentrifuged-on-cells | 2-8 °C | 6 hthen changed, changed by 12 h | % bias of median vs baseline exceeds desirable bias (Westgard/Ricos BV), here 1.3% | Cuhadar et al. 2012, Biochem Med (Zagreb)Table 3, row 'Albumin', w/o gel 4°C |
| serum, BD Vacutainer RST 5 mL (thrombin clot activator, gel) | Storagecentrifuged-on-gel | 2-8 °C | 7 daysstable for the longest time tested | mean % difference vs time 0 exceeds desirable imprecision based on biological variation (EFLM/Westgard), here 1.3% | Mandić et al. 2020, Biochem Med (Zagreb)Tables 3 and 4, row 'Albumin, g/L', BD RST columns |
| serum, Vacuette Tube Serum Gel Separator 8 mL (Greiner, 455071) | Storagecentrifuged-on-gel | RT | 0 hchanged at the first time point, changed by 6 h | % bias of median vs baseline exceeds desirable bias (Westgard/Ricos BV), here 1.3% | Cuhadar et al. 2012, Biochem Med (Zagreb)Table 3, row 'Albumin', gel 24°C |
| serum, Vacuette Tube Serum No Gel 9 mL (Greiner, 455092), plain | Storagecentrifuged-on-cells | RT | 0 hchanged at the first time point, changed by 6 h | % bias of median vs baseline exceeds desirable bias (Westgard/Ricos BV), here 1.3% | Cuhadar et al. 2012, Biochem Med (Zagreb)Table 3, row 'Albumin', w/o gel 24°C |
| serum, BD Vacutainer SST (SKU 367986), gel | Storagecentrifuged-on-gel | 30 °C (±1 °C) | 72 hstable for the longest time tested | mean bias vs paired baseline within CLIA acceptance limits AND >2/3 of individual results within limits | Iacovetti et al. 2026, Pract Lab MedResults, 'Analyte stability after storage at 30 °C', paragraph 1 (Table 1 image-only) |
| serum, BD SST Microtainer (SKU 395967), gel; filled from venous blood | Storagecentrifuged-on-gel | 0 °C | 72 hstable for the longest time tested | mean bias vs paired baseline within CLIA acceptance limits AND >2/3 of individual results within limits | Iacovetti et al. 2026, Pract Lab MedResults, 'Analyte stability after separation and storage in microvolume tubes at 0 °C' (Table 2 image-only) |
| whole-blood-clotted, BD Vacutainer SST | Before spinninguncentrifuged | RT | 24 hstable for the longest time tested | mean % deviation from baseline within total change limit TCL = sqrt((2.77 CVa)^2 + (0.5 CVb)^2), CVb from Ricos (Table 2) | Abraham et al. 2019, Biochem Med (Zagreb)Discussion, paragraph 1 (Results paragraphs 1-3; Figure 3) |
| whole-blood-clotted, BD Vacutainer SST (SKU 367986), gel | Before spinninguncentrifuged | 30 °C (±1 °C) | 72 hstable for the longest time tested | mean bias vs paired baseline within CLIA acceptance limits AND >2/3 of individual results within limits | Iacovetti et al. 2026, Pract Lab MedResults, 'Analyte stability after storage at 30 °C', paragraph 2 (Table 1 image-only) |
| whole-blood-clotted, BD Vacutainer SST | Before spinninguncentrifuged | 22-30 °C (logger 24-26 °C) | 24 hstable for the longest time tested | mean % deviation from baseline within total change limit TCL = sqrt((2.77 CVa)^2 + (0.5 CVb)^2), CVb from Ricos (Table 2) | Abraham et al. 2019, Biochem Med (Zagreb)Discussion, paragraph 1 (Results paragraphs 1-3; Figure 3) |
| whole-blood-clotted, BD Vacutainer SST | Before spinninguncentrifuged | > 30 °C (logger 38-40 °C) | 24 hstable for the longest time tested | mean % deviation from baseline within total change limit TCL = sqrt((2.77 CVa)^2 + (0.5 CVb)^2), CVb from Ricos (Table 2) | Abraham et al. 2019, Biochem Med (Zagreb)Discussion, paragraph 1 (Results paragraphs 1-3; Figure 3) |
| whole-blood-heparin, BD Vacutainer lithium heparin, no gel (ref 368884) | Before spinninguncentrifuged | RT | 10 hstable for the longest time tested | |mean % deviation 10 h vs baseline| > laboratory bias goal (Table II; mostly biological-variation based, some empirically modified) | Henriksen et al. 2014, Scand J Clin Lab InvestTable II, row 'ALB' |
| whole-blood-heparin, BD Vacutainer lithium heparin, no gel (ref 368884) | Before spinninguncentrifuged | RT | 10 hstable for the longest time tested | CV of paired baseline/10 h results > goal CV (Table II) | Henriksen et al. 2014, Scand J Clin Lab InvestTable II, row 'ALB' |
Alkaline Phosphatase (ALP)16 findings
| Specimen, tube | Condition | Temperature | Stable for | Criterion | Study |
|---|---|---|---|---|---|
| plasma-li-heparin, Sarstedt lithium heparin, no gel separator (primary tube) | Storagecentrifuged-on-cells | 2-8 °C | 48 hstable for the longest time tested | mean % difference vs time 0 exceeds total limit of change, TLC = ±√((1.65·CVa)² + (0.5·CVb)²) (EFLM BV), here 10.1%; stability limit from instability-equation regression | Illana et al. 2024, Biochem Med (Zagreb)Table 1, row 'ALP' |
| plasma-li-heparin, Sarstedt lithium heparin, no gel separator; plasma aliquot | Storageseparated | 2-8 °C | 96 hstable for the longest time tested | mean % difference vs time 0 exceeds total limit of change, TLC = ±√((1.65·CVa)² + (0.5·CVb)²) (EFLM BV), here 10.1%; stability limit from instability-equation regression | Illana et al. 2024, Biochem Med (Zagreb)Table 2, row 'ALP' |
| plasma-li-heparin, BD Vacutainer lithium heparin 6 mL (regular) | Storagenot-stated | 2-8 °C | 72 hstable for the longest time tested | mean % difference vs T0 exceeds total change limit (TCL), here ±14.34% | Dupuy et al. 2018, Biochem Med (Zagreb)Table 2, BD lithium heparin tubes, row 'ALP, U/L' |
| plasma-li-heparin, BD Barricor lithium heparin 5.5 mL (mechanical separator) | Storagecentrifuged-on-gel | 2-8 °C | 72 hstable for the longest time tested | mean % difference vs T0 exceeds total change limit (TCL), here ±14.34% | Dupuy et al. 2018, Biochem Med (Zagreb)Table 2, BD Barricor lithium heparin tubes, row 'ALP, U/L' |
| plasma-li-heparin, BD Barricor 4.5 mL (lithium heparin, mechanical separator) | Storagecentrifuged-on-gel | 2-8 °C | 24 hthen changed, changed by 7 days | mean % difference vs time 0 exceeds desirable imprecision based on biological variation (EFLM/Westgard), here 2.7% | Mandić et al. 2020, Biochem Med (Zagreb)Tables 3 and 4, row 'ALP, U/L', BD Barricor columns |
| serum, BD Vacutainer serum (plain), 7.5 mL | Storageseparated | −20 °C | 30 daysstable for the longest time tested | clinically significant change: mean outside significant change limit (SCL = T1d mean ± 3.0 usual SD from 2 months QC) | Kachhawa et al. 2017, J Lab PhysiciansDiscussion, 'Enzymes' (Table 2 image-only) |
| serum, Vacuette Tube Serum Gel Separator 8 mL (Greiner, 455071) | Storagecentrifuged-on-gel | 2-8 °C | 1 weekstable for the longest time tested | % bias of median vs baseline exceeds desirable bias (Westgard/Ricos BV), here 6.4% | Cuhadar et al. 2012, Biochem Med (Zagreb)Table 3, row 'Alkaline phosphatase', gel 4°C |
| serum, Vacuette Tube Serum No Gel 9 mL (Greiner, 455092), plain | Storagecentrifuged-on-cells | 2-8 °C | 1 weekstable for the longest time tested | % bias of median vs baseline exceeds desirable bias (Westgard/Ricos BV), here 6.4% | Cuhadar et al. 2012, Biochem Med (Zagreb)Table 3, row 'Alkaline phosphatase', w/o gel 4°C |
| serum, BD Vacutainer RST 5 mL (thrombin clot activator, gel) | Storagecentrifuged-on-gel | 2-8 °C | 7 daysstable for the longest time tested | mean % difference vs time 0 exceeds desirable imprecision based on biological variation (EFLM/Westgard), here 2.7% | Mandić et al. 2020, Biochem Med (Zagreb)Tables 3 and 4, row 'ALP, U/L', BD RST columns |
| serum, Vacuette Tube Serum Gel Separator 8 mL (Greiner, 455071) | Storagecentrifuged-on-gel | RT | 1 weekstable for the longest time tested | % bias of median vs baseline exceeds desirable bias (Westgard/Ricos BV), here 6.4% | Cuhadar et al. 2012, Biochem Med (Zagreb)Table 3, row 'Alkaline phosphatase', gel 24°C |
| serum, Vacuette Tube Serum No Gel 9 mL (Greiner, 455092), plain | Storagecentrifuged-on-cells | RT | 1 weekstable for the longest time tested | % bias of median vs baseline exceeds desirable bias (Westgard/Ricos BV), here 6.4% | Cuhadar et al. 2012, Biochem Med (Zagreb)Table 3, row 'Alkaline phosphatase', w/o gel 24°C |
| serum, BD Vacutainer SST (SKU 367986), gel | Storagecentrifuged-on-gel | 30 °C (±1 °C) | 72 hstable for the longest time tested | mean bias vs paired baseline within CLIA acceptance limits AND >2/3 of individual results within limits | Iacovetti et al. 2026, Pract Lab MedResults, 'Analyte stability after storage at 30 °C', paragraph 1 (Table 1 image-only) |
| serum, BD SST Microtainer (SKU 395967), gel; filled from venous blood | Storagecentrifuged-on-gel | 0 °C | 72 hstable for the longest time tested | mean bias vs paired baseline within CLIA acceptance limits AND >2/3 of individual results within limits | Iacovetti et al. 2026, Pract Lab MedResults, 'Analyte stability after separation and storage in microvolume tubes at 0 °C' (Table 2 image-only) |
| whole-blood-clotted, BD Vacutainer SST (SKU 367986), gel | Before spinninguncentrifuged | 30 °C (±1 °C) | 72 hstable for the longest time tested | mean bias vs paired baseline within CLIA acceptance limits AND >2/3 of individual results within limits | Iacovetti et al. 2026, Pract Lab MedResults, 'Analyte stability after storage at 30 °C', paragraph 2 (Table 1 image-only) |
| whole-blood-heparin, BD Vacutainer lithium heparin, no gel (ref 368884) | Before spinninguncentrifuged | RT | 10 hstable for the longest time tested | |mean % deviation 10 h vs baseline| > laboratory bias goal (Table II; mostly biological-variation based, some empirically modified) | Henriksen et al. 2014, Scand J Clin Lab InvestTable II, row 'BASP' |
| whole-blood-heparin, BD Vacutainer lithium heparin, no gel (ref 368884) | Before spinninguncentrifuged | RT | 10 hstable for the longest time tested | CV of paired baseline/10 h results > goal CV (Table II) | Henriksen et al. 2014, Scand J Clin Lab InvestTable II, row 'BASP' |
Alpha-1 Antitrypsin (A1AT)2 findings
| Specimen, tube | Condition | Temperature | Stable for | Criterion | Study |
|---|---|---|---|---|---|
| whole-blood-heparin, BD Vacutainer lithium heparin, no gel (ref 368884) | Before spinninguncentrifuged | RT | 10 hstable for the longest time tested | |mean % deviation 10 h vs baseline| > laboratory bias goal (Table II; mostly biological-variation based, some empirically modified) | Henriksen et al. 2014, Scand J Clin Lab InvestTable II, row 'ATRYP' |
| whole-blood-heparin, BD Vacutainer lithium heparin, no gel (ref 368884) | Before spinninguncentrifuged | RT | 10 hstable for the longest time tested | CV of paired baseline/10 h results > goal CV (Table II) | Henriksen et al. 2014, Scand J Clin Lab InvestTable II, row 'ATRYP' |
Alpha-carotene3 findings
| Specimen, tube | Condition | Temperature | Stable for | Criterion | Study |
|---|---|---|---|---|---|
| whole-blood-clotted, Sarstedt Monovette serum | Before spinninguncentrifuged | chilled in transit (insulated box, 2 frozen cool packs; temperature not recorded) | 24 hstable for the longest time tested | % difference within desirable quality specification for bias (Fraser, from intra- and interindividual variation) | Jones et al. 2021, J NutrTable 2, row 'α-carotene', column Serum |
| whole-blood-edta, Sarstedt Monovette EDTA | Before spinninguncentrifuged | chilled in transit (insulated box, 2 frozen cool packs; temperature not recorded) | 24 hstable for the longest time tested | % difference within desirable quality specification for bias (Fraser, from intra- and interindividual variation) | Jones et al. 2021, J NutrTable 2, row 'α-carotene', column EDTA |
| whole-blood-heparin, Sarstedt Monovette lithium heparin, trace-element free | Before spinninguncentrifuged | chilled in transit (insulated box, 2 frozen cool packs; temperature not recorded) | 24 hstable for the longest time tested | % difference within desirable quality specification for bias (Fraser, from intra- and interindividual variation) | Jones et al. 2021, J NutrTable 2, row 'α-carotene', column LH |
Amylase7 findings
| Specimen, tube | Condition | Temperature | Stable for | Criterion | Study |
|---|---|---|---|---|---|
| plasma-li-heparin, Sarstedt lithium heparin, no gel separator (primary tube) | Storagecentrifuged-on-cells | 2-8 °C | 48 hstable for the longest time tested | mean % difference vs time 0 exceeds total limit of change, TLC = ±√((1.65·CVa)² + (0.5·CVb)²) (EFLM BV), here 9.1%; stability limit from instability-equation regression | Illana et al. 2024, Biochem Med (Zagreb)Table 1, row 'AMY' |
| plasma-li-heparin, Sarstedt lithium heparin, no gel separator; plasma aliquot | Storageseparated | 2-8 °C | 96 hstable for the longest time tested | mean % difference vs time 0 exceeds total limit of change, TLC = ±√((1.65·CVa)² + (0.5·CVb)²) (EFLM BV), here 9.1%; stability limit from instability-equation regression | Illana et al. 2024, Biochem Med (Zagreb)Table 2, row 'AMY' |
| plasma-li-heparin, BD Barricor 4.5 mL (lithium heparin, mechanical separator) | Storagecentrifuged-on-gel | 2-8 °C | 7 daysstable for the longest time tested | mean % difference vs time 0 exceeds desirable imprecision based on biological variation (EFLM/Westgard), here 3.3% | Mandić et al. 2020, Biochem Med (Zagreb)Tables 3 and 4, row 'Amylase, U/L', BD Barricor columns |
| serum, BD Vacutainer serum (plain), 7.5 mL | Storageseparated | −20 °C | 0 dayschanged at the first time point, changed by 7 days | clinically significant change: mean outside significant change limit (SCL = T1d mean ± 3.0 usual SD from 2 months QC) | Kachhawa et al. 2017, J Lab PhysiciansDiscussion, 'Enzymes' (Table 2 image-only) |
| serum, BD Vacutainer RST 5 mL (thrombin clot activator, gel) | Storagecentrifuged-on-gel | 2-8 °C | 7 daysstable for the longest time tested | mean % difference vs time 0 exceeds desirable imprecision based on biological variation (EFLM/Westgard), here 3.3% | Mandić et al. 2020, Biochem Med (Zagreb)Tables 3 and 4, row 'Amylase, U/L', BD RST columns |
| whole-blood-heparin, BD Vacutainer lithium heparin, no gel (ref 368884) | Before spinninguncentrifuged | RT | 10 hstable for the longest time tested | |mean % deviation 10 h vs baseline| > laboratory bias goal (Table II; mostly biological-variation based, some empirically modified) | Henriksen et al. 2014, Scand J Clin Lab InvestTable II, row 'AMYL' |
| whole-blood-heparin, BD Vacutainer lithium heparin, no gel (ref 368884) | Before spinninguncentrifuged | RT | 10 hstable for the longest time tested | CV of paired baseline/10 h results > goal CV (Table II) | Henriksen et al. 2014, Scand J Clin Lab InvestTable II, row 'AMYL' |
Anti-Müllerian Hormone12 findings
| Specimen, tube | Condition | Temperature | Stable for | Criterion | Study |
|---|---|---|---|---|---|
| plasma-li-heparin, VACUETTE lithium heparin Plasma Sep (gel) | Storageseparated | −20 °C | 30 daysstable for the longest time tested | equivalence test (TOST, paired) vs initial value, tolerance ±0.30 ng/mL | Vrzáková et al. 2023, Diagnostics (Basel)Results, paragraph 6; Table 1 |
| serum, VACUETTE Z Serum Sep (gel) | Freeze–thawseparated | −80 °C | 5 cyclesstable for the longest time tested | equivalence test (TOST, paired) vs initial value, tolerance ±0.30 ng/mL | Vrzáková et al. 2023, Diagnostics (Basel)Results, paragraph 3; Table 1 |
| serum, VACUETTE Z Serum Sep (gel) | Storageseparated | −20 °C | 30 daysstable for the longest time tested | equivalence test (TOST, paired) vs initial value, tolerance ±0.30 ng/mL | Vrzáková et al. 2023, Diagnostics (Basel)Results, paragraph 6; Table 1 |
| serum, V-Tube serum separator gel tube (AB Medical) | Storageseparated | −20 °C | 0 hchanged at the first time point, changed by 48 h | Wilcoxon signed-rank test vs fresh control, p < 0.05 | Lee et al. 2023, Clin Exp Reprod MedTable 3, Rev-Gen II-AMH |
| serum, V-Tube serum separator gel tube (AB Medical) | Storageseparated | −20 °C | 0 yearschanged at the first time point, changed by 2 years | Wilcoxon signed-rank test vs fresh control, p < 0.05 | Lee et al. 2023, Clin Exp Reprod MedTable 3, Rev-Gen II-AMH |
| serum, V-Tube serum separator gel tube (AB Medical) | Storageseparated | −20 °C | 48 hthen changed, changed by 1 week | Wilcoxon signed-rank test vs fresh control, p < 0.05 | Lee et al. 2023, Clin Exp Reprod MedTable 3, Access-AMH; Results paragraph 2 |
| serum, V-Tube serum separator gel tube (AB Medical) | Storageseparated | −20 °C | 0 yearschanged at the first time point, changed by 2 years | Wilcoxon signed-rank test vs fresh control, p < 0.05 | Lee et al. 2023, Clin Exp Reprod MedTable 3, Access-AMH |
| serum, VACUETTE Z Serum Sep (gel) | Storageseparated | 2-8 °C | 168 hstable for the longest time tested | equivalence test (TOST, paired) vs initial value, tolerance ±0.30 ng/mL | Vrzáková et al. 2023, Diagnostics (Basel)Results, paragraph 5; Table 1 |
| serum, V-Tube serum separator gel tube (AB Medical) | Storageseparated | 2-8 °C | 0 hchanged at the first time point, changed by 48 h | Wilcoxon signed-rank test vs fresh control, p < 0.05 | Lee et al. 2023, Clin Exp Reprod MedTable 3, Rev-Gen II-AMH |
| serum, V-Tube serum separator gel tube (AB Medical) | Storageseparated | 2-8 °C | 48 hthen changed, changed by 1 week | Wilcoxon signed-rank test vs fresh control, p < 0.05 | Lee et al. 2023, Clin Exp Reprod MedTable 3, Access-AMH; Results paragraph 2 |
| whole-blood-clotted, V-Tube serum separator gel tube (AB Medical) | Before spinninguncentrifuged | RT | 0 hchanged at the first time point, changed by 48 h | Wilcoxon signed-rank test vs fresh control, p < 0.05 | Lee et al. 2023, Clin Exp Reprod MedTable 4; Results paragraph 4 |
| whole-blood-clotted, V-Tube serum separator gel tube (AB Medical) | Before spinninguncentrifuged | RT | 0 hchanged at the first time point, changed by 48 h | Wilcoxon signed-rank test vs fresh control, p < 0.05 | Lee et al. 2023, Clin Exp Reprod MedTable 4; Results paragraph 4 |
Anti-Xa activity5 findings
| Specimen, tube | Condition | Temperature | Stable for | Criterion | Study |
|---|---|---|---|---|---|
| plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquots | Freeze–thawseparated | −20 °C | 1 cyclelimit stated by the authors | max allowable mean deviation from t0: ≤0.45 IU/mL: ≤0.045 IU/mL;>0.45 IU/mL: ≤10% | Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability' |
| plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquots | Freeze–thawseparated | −80 °C | 3 cycleslimit stated by the authors | max allowable mean deviation from t0: ≤0.45 IU/mL: ≤0.045 IU/mL;>0.45 IU/mL: ≤10% | Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability' |
| plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquots | Storageseparated | −20 °C | 14 dayslimit stated by the authors | max allowable mean deviation from t0: ≤0.45 IU/mL: ≤0.045 IU/mL;>0.45 IU/mL: ≤10% | Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability' |
| plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquots | Storageseparated | −80 °C | 11 weekslimit stated by the authors | max allowable mean deviation from t0: ≤0.45 IU/mL: ≤0.045 IU/mL;>0.45 IU/mL: ≤10% | Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability' |
| plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquots | Storageseparated | RT | 2 hlimit stated by the authors | max allowable mean deviation from t0: ≤0.45 IU/mL: ≤0.045 IU/mL;>0.45 IU/mL: ≤10% | Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability' |
Antithrombin III7 findings
| Specimen, tube | Condition | Temperature | Stable for | Criterion | Study |
|---|---|---|---|---|---|
| plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquots | Freeze–thawseparated | −20 °C | 1 cyclelimit stated by the authors | max allowable mean deviation from t0: ≤40.0%: ±4.0% (absolute);>40.0%: ±10% (relative) | Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability' |
| plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquots | Freeze–thawseparated | −80 °C | 3 cycleslimit stated by the authors | max allowable mean deviation from t0: ≤40.0%: ±4.0% (absolute);>40.0%: ±10% (relative) | Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability' |
| plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquots | Storageseparated | −20 °C | 28 dayslimit stated by the authors | max allowable mean deviation from t0: ≤40.0%: ±4.0% (absolute);>40.0%: ±10% (relative) | Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability' |
| plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquots | Storageseparated | −20 °C | 7 weekslimit stated by the authors | max allowable mean deviation from t0: ≤40.0%: ±4.0% (absolute);>40.0%: ±10% (relative) | Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability' |
| plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquots | Storageseparated | −80 °C | 23 monthslimit stated by the authors | max allowable mean deviation from t0: ≤40.0%: ±4.0% (absolute);>40.0%: ±10% (relative) | Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability' |
| plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquots | Storageseparated | 2-8 °C | 14 dayslimit stated by the authors | max allowable mean deviation from t0: ≤40.0%: ±4.0% (absolute);>40.0%: ±10% (relative) | Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability' |
| plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquots | Storageseparated | RT | 2 dayslimit stated by the authors | max allowable mean deviation from t0: ≤40.0%: ±4.0% (absolute);>40.0%: ±10% (relative) | Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability' |
aPTT (lupus-sensitive reagent)3 findings
| Specimen, tube | Condition | Temperature | Stable for | Criterion | Study |
|---|---|---|---|---|---|
| plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquots | Freeze–thawseparated | −80 °C | 1 cyclelimit stated by the authors | max allowable mean deviation from t0: <100 s: ±10%; ≥100 s: ±20% | Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability' |
| plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquots | Storageseparated | −80 °C | 5 monthslimit stated by the authors | max allowable mean deviation from t0: <100 s: ±10%; ≥100 s: ±20% | Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability' |
| plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquots | Storageseparated | RT | 2 hlimit stated by the authors | max allowable mean deviation from t0: <100 s: ±10%; ≥100 s: ±20% | Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability' |
aPTT screen3 findings
| Specimen, tube | Condition | Temperature | Stable for | Criterion | Study |
|---|---|---|---|---|---|
| plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquots | Freeze–thawseparated | −80 °C | 1 cyclelimit stated by the authors | max allowable mean deviation from t0: <100 s: ±10%; ≥100 s: ±20% | Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability' |
| plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquots | Storageseparated | −80 °C | 11 weekslimit stated by the authors | max allowable mean deviation from t0: <100 s: ±10%; ≥100 s: ±20% | Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability' |
| plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquots | Storageseparated | RT | 2 hlimit stated by the authors | max allowable mean deviation from t0: <100 s: ±10%; ≥100 s: ±20% | Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability' |
Aspartate Aminotransferase (AST)22 findings
| Specimen, tube | Condition | Temperature | Stable for | Criterion | Study |
|---|---|---|---|---|---|
| plasma-li-heparin, Sarstedt lithium heparin, no gel separator (primary tube) | Storagecentrifuged-on-cells | 2-8 °C | 7.9 hthen changed, changed by 8 h | mean % difference vs time 0 exceeds total limit of change, TLC = ±√((1.65·CVa)² + (0.5·CVb)²) (EFLM BV), here 8.7%; stability limit from instability-equation regression | Illana et al. 2024, Biochem Med (Zagreb)Table 1, row 'AST' |
| plasma-li-heparin, Sarstedt lithium heparin, no gel separator; plasma aliquot | Storageseparated | 2-8 °C | 96 hstable for the longest time tested | mean % difference vs time 0 exceeds total limit of change, TLC = ±√((1.65·CVa)² + (0.5·CVb)²) (EFLM BV), here 8.7%; stability limit from instability-equation regression | Illana et al. 2024, Biochem Med (Zagreb)Table 2, row 'AST' |
| plasma-li-heparin, BD Vacutainer lithium heparin 6 mL (regular) | Storagenot-stated | 2-8 °C | 24 hthen changed, changed by 72 h | mean % difference vs T0 exceeds total change limit (TCL), here ±10.83% | Dupuy et al. 2018, Biochem Med (Zagreb)Table 2, BD lithium heparin tubes, row 'AST, U/L' |
| plasma-li-heparin, BD Barricor lithium heparin 5.5 mL (mechanical separator) | Storagecentrifuged-on-gel | 2-8 °C | 24 hthen changed, changed by 72 h | mean % difference vs T0 exceeds total change limit (TCL), here ±10.83% | Dupuy et al. 2018, Biochem Med (Zagreb)Table 2, BD Barricor lithium heparin tubes, row 'AST, U/L' |
| plasma-li-heparin, BD Vacutainer LH PST II gel separator, 3.0 mL (REF 367374) | Storagecentrifuged-on-gel | 2-8 °C | 24 hstable for the longest time tested | mean % difference smaller than TEa (Ricos: 0.25 x (CVi^2 + CVg^2)^0.5 + 1.65 x 0.5 x CVi) | Nonkes et al. 2019, Biochem Med (Zagreb)Table 1, row 'AST, U/L' |
| plasma-li-heparin, BD Vacutainer Barricor LH 3 mL, mechanical separator | Storagecentrifuged-on-gel | 2-8 °C | 80 hstable for the longest time tested | bias vs aliquoted plasma reference tube > maximal allowable bias 5.4% (Ricos et al.); time to MAB from linear regression of bias on time to analysis | Gawria et al. 2020, J Clin Lab AnalTable 2, row 'AST', Barricor BD |
| plasma-li-heparin, BD Vacutainer PST LH 3 mL, gel | Storagecentrifuged-on-gel | 2-8 °C | 0 hchanged at the first time point, changed by 3 h | bias vs aliquoted plasma reference tube > maximal allowable bias 5.4% (Ricos et al.); time to MAB from linear regression of bias on time to analysis | Gawria et al. 2020, J Clin Lab AnalTable 2, row 'AST', Gel BD |
| plasma-li-heparin, Greiner Vacuette LH 3 mL Sep, gel | Storagecentrifuged-on-gel | 2-8 °C | 0 hchanged at the first time point, changed by 3 h | bias vs aliquoted plasma reference tube > maximal allowable bias 5.4% (Ricos et al.); time to MAB from linear regression of bias on time to analysis | Gawria et al. 2020, J Clin Lab AnalTable 2, row 'AST', Gel Greiner |
| plasma-li-heparin, BD Barricor 4.5 mL (lithium heparin, mechanical separator) | Storagecentrifuged-on-gel | 2-8 °C | 0 hchanged at the first time point, changed by 24 h | mean % difference vs time 0 exceeds desirable imprecision based on biological variation (EFLM/Westgard), here 4.75% | Mandić et al. 2020, Biochem Med (Zagreb)Tables 3 and 4, row 'AST, U/L', BD Barricor columns |
| serum, BD Vacutainer 8.5 mL plastic serum tube; pooled serum aliquoted into 1.5 mL Eppendorf tubes | Freeze–thawseparated | −20 °C | 10 cyclesstable for the longest time tested | unstable only if both P < 0.005 (Bonferroni) and % change from T0 > desirable bias (Westgard/Ricos), here 5.4% | Cuhadar et al. 2013, Biochem Med (Zagreb)Table 2, row 'AST (U/L)', columns T1d-T10d |
| serum, BD Vacutainer 8.5 mL plastic serum tube; pooled serum aliquoted into 1.5 mL Eppendorf tubes | Storageseparated | −20 °C | 3 monthsstable for the longest time tested | unstable only if both P < 0.013 (Bonferroni) and % change from T0 > desirable bias (Westgard/Ricos), here 5.4% | Cuhadar et al. 2013, Biochem Med (Zagreb)Table 2, row 'AST (U/L)', columns T1m-T3m |
| serum, BD Vacutainer serum (plain), 7.5 mL | Storageseparated | −20 °C | 30 daysstable for the longest time tested | clinically significant change: mean outside significant change limit (SCL = T1d mean ± 3.0 usual SD from 2 months QC) | Kachhawa et al. 2017, J Lab PhysiciansDiscussion, 'Enzymes' (Table 2 image-only) |
| serum, Vacuette Tube Serum Gel Separator 8 mL (Greiner, 455071) | Storagecentrifuged-on-gel | 2-8 °C | 0 hchanged at the first time point, changed by 6 h | % bias of median vs baseline exceeds desirable bias (Westgard/Ricos BV), here 5.4% | Cuhadar et al. 2012, Biochem Med (Zagreb)Table 3, row 'Aspartate aminotransferase', gel 4°C |
| serum, Vacuette Tube Serum No Gel 9 mL (Greiner, 455092), plain | Storagecentrifuged-on-cells | 2-8 °C | 6 hthen changed, changed by 12 h | % bias of median vs baseline exceeds desirable bias (Westgard/Ricos BV), here 5.4% | Cuhadar et al. 2012, Biochem Med (Zagreb)Table 3, row 'Aspartate aminotransferase', w/o gel 4°C |
| serum, BD Vacutainer RST 4.0 mL (thrombin clot activator, gel) | Storagecentrifuged-on-gel | 2-8 °C | 24 hstable for the longest time tested | clinically significant if mean bias vs t0 exceeds desirable bias (biological variation), here 2.0% | Kocijancic et al. 2014, Biochem Med (Zagreb)Table 3, row 'AST' |
| serum, BD Vacutainer SST II Advance 5.0 mL (clot activator, gel) | Storagecentrifuged-on-gel | 2-8 °C | 24 hstable for the longest time tested | clinically significant if mean bias vs t0 exceeds desirable bias (biological variation), here 2.0% | Kocijancic et al. 2014, Biochem Med (Zagreb)Table 4, row 'AST' |
| serum, BD Vacutainer RST 5 mL (thrombin clot activator, gel) | Storagecentrifuged-on-gel | 2-8 °C | 24 hthen changed, changed by 7 days | mean % difference vs time 0 exceeds desirable imprecision based on biological variation (EFLM/Westgard), here 4.75% | Mandić et al. 2020, Biochem Med (Zagreb)Tables 3 and 4, row 'AST, U/L', BD RST columns |
| serum, Vacuette Tube Serum Gel Separator 8 mL (Greiner, 455071) | Storagecentrifuged-on-gel | RT | 0 hchanged at the first time point, changed by 6 h | % bias of median vs baseline exceeds desirable bias (Westgard/Ricos BV), here 5.4% | Cuhadar et al. 2012, Biochem Med (Zagreb)Table 3, row 'Aspartate aminotransferase', gel 24°C |
| serum, Vacuette Tube Serum No Gel 9 mL (Greiner, 455092), plain | Storagecentrifuged-on-cells | RT | 6 hthen changed, changed by 12 h | % bias of median vs baseline exceeds desirable bias (Westgard/Ricos BV), here 5.4% | Cuhadar et al. 2012, Biochem Med (Zagreb)Table 3, row 'Aspartate aminotransferase', w/o gel 24°C |
| serum, BD Vacutainer SST (SKU 367986), gel | Storagecentrifuged-on-gel | 30 °C (±1 °C) | 72 hstable for the longest time tested | mean bias vs paired baseline within CLIA acceptance limits AND >2/3 of individual results within limits | Iacovetti et al. 2026, Pract Lab MedResults, 'Analyte stability after storage at 30 °C', paragraph 1 (Table 1 image-only) |
| serum, BD SST Microtainer (SKU 395967), gel; filled from venous blood | Storagecentrifuged-on-gel | 0 °C | 72 hstable for the longest time tested | mean bias vs paired baseline within CLIA acceptance limits AND >2/3 of individual results within limits | Iacovetti et al. 2026, Pract Lab MedResults, 'Analyte stability after separation and storage in microvolume tubes at 0 °C' (Table 2 image-only) |
| whole-blood-clotted, BD Vacutainer SST (SKU 367986), gel | Before spinninguncentrifuged | 30 °C (±1 °C) | 24 hthen changed, changed by 72 h | mean bias vs paired baseline within CLIA acceptance limits AND >2/3 of individual results within limits | Iacovetti et al. 2026, Pract Lab MedResults, 'Analyte stability after storage at 30 °C', paragraph 2 (Table 1 image-only) |
Basophils (Absolute)20 findings
| Specimen, tube | Condition | Temperature | Stable for | Criterion | Study |
|---|---|---|---|---|---|
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 48 hstable for the longest time tested | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 4, row 'BA', XN-9000, 4 °C |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 48 hstable for the longest time tested | Bias% vs optimal bias (OP-Bias%, from biological variation; Table 1) | Buoro et al. 2016, Rev Bras Hematol HemoterTable 4, row 'BA', XN-9000, 4 °C |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 48 hstable for the longest time tested | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 4, row 'BA', BC-6800, 4 °C |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 48 hstable for the longest time tested | Bias% vs optimal bias (OP-Bias%, from biological variation; Table 1) | Buoro et al. 2016, Rev Bras Hematol HemoterTable 4, row 'BA', BC-6800, 4 °C |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 8 hstable for the longest time tested | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 5, row 'BA', XN-9000, 4 °C |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 4 hthen changed, changed by 8 h | Bias% vs optimal bias (OP-Bias%, from biological variation; Table 1) | Buoro et al. 2016, Rev Bras Hematol HemoterTable 5, row 'BA', XN-9000, 4 °C |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 8 hstable for the longest time tested | Bias% vs critical difference (CD%, reference change value; Table 1) | Buoro et al. 2016, Rev Bras Hematol HemoterTable 5, row 'BA', XN-9000, 4 °C |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 8 hstable for the longest time tested | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 5, row 'BA', BC-6800, 4 °C |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 8 hstable for the longest time tested | Bias% vs optimal bias (OP-Bias%, from biological variation; Table 1) | Buoro et al. 2016, Rev Bras Hematol HemoterTable 5, row 'BA', BC-6800, 4 °C |
| whole-blood-edta, K2-EDTA (BD Vacutainer) | Storageuncentrifuged | 2-8 °C | 48 hstable for the longest time tested | Repeated-measures one-way ANOVA with Tukey post hoc vs 0 h, p < 0.05 (* in table) | Unalli et al. 2021, J Med BiochemTable 1, column '4 °C' (no * at any time point) |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | RT | 48 hstable for the longest time tested | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 4, row 'BA', XN-9000, RT |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | RT | 48 hstable for the longest time tested | Bias% vs optimal bias (OP-Bias%, from biological variation; Table 1) | Buoro et al. 2016, Rev Bras Hematol HemoterTable 4, row 'BA', XN-9000, RT |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | RT | 48 hstable for the longest time tested | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 4, row 'BA', BC-6800, RT |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | RT | 48 hstable for the longest time tested | Bias% vs optimal bias (OP-Bias%, from biological variation; Table 1) | Buoro et al. 2016, Rev Bras Hematol HemoterTable 4, row 'BA', BC-6800, RT |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | RT | 8 hstable for the longest time tested | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 5, row 'BA', XN-9000, RT |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | RT | 8 hstable for the longest time tested | Bias% vs optimal bias (OP-Bias%, from biological variation; Table 1) | Buoro et al. 2016, Rev Bras Hematol HemoterTable 5, row 'BA', XN-9000, RT |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | RT | 8 hstable for the longest time tested | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 5, row 'BA', BC-6800, RT |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | RT | 8 hstable for the longest time tested | Bias% vs optimal bias (OP-Bias%, from biological variation; Table 1) | Buoro et al. 2016, Rev Bras Hematol HemoterTable 5, row 'BA', BC-6800, RT |
| whole-blood-edta, K2-EDTA (BD Vacutainer) | Storageuncentrifuged | RT | 8 hthen changed, changed by 12 h | Repeated-measures one-way ANOVA with Tukey post hoc vs 0 h, p < 0.05 (* in table) | Unalli et al. 2021, J Med BiochemTable 1, block 'BASO', row '12 h', column '23 °C' |
| whole-blood-edta, K2-EDTA (BD Vacutainer) | Storageuncentrifuged | 10 °C | 48 hstable for the longest time tested | Repeated-measures one-way ANOVA with Tukey post hoc vs 0 h, p < 0.05 (* in table) | Unalli et al. 2021, J Med BiochemTable 1, column '10 °C' (no * at any time point) |
Beta-carotene4 findings
| Specimen, tube | Condition | Temperature | Stable for | Criterion | Study |
|---|---|---|---|---|---|
| whole-blood-clotted, Sarstedt Monovette serum | Before spinninguncentrifuged | chilled in transit (insulated box, 2 frozen cool packs; temperature not recorded) | 24 hstable for the longest time tested | % difference within desirable quality specification for bias (Fraser, from intra- and interindividual variation) | Jones et al. 2021, J NutrTable 2, row 'β -carotene', column Serum |
| whole-blood-edta, Sarstedt Monovette EDTA | Before spinninguncentrifuged | chilled in transit (insulated box, 2 frozen cool packs; temperature not recorded) | 24 hstable for the longest time tested | % difference within desirable quality specification for bias (Fraser, from intra- and interindividual variation) | Jones et al. 2021, J NutrTable 2, row 'β -carotene', column EDTA |
| whole-blood-heparin, BD Vacutainer lithium heparin | Before spinninguncentrifuged | cold packs in insulated box; 9 (2) °C on arrival | 24 hstable for the longest time tested | % difference within desirable quality specification (from intra- and interindividual variation) | Jones et al. 2024, J NutrTable 2, row 'β-Carotene, μmol/L'; Table 1 |
| whole-blood-heparin, Sarstedt Monovette lithium heparin, trace-element free | Before spinninguncentrifuged | chilled in transit (insulated box, 2 frozen cool packs; temperature not recorded) | 24 hstable for the longest time tested | % difference within desirable quality specification for bias (Fraser, from intra- and interindividual variation) | Jones et al. 2021, J NutrTable 2, row 'β -carotene', column LH |
Beta-cryptoxanthin3 findings
| Specimen, tube | Condition | Temperature | Stable for | Criterion | Study |
|---|---|---|---|---|---|
| whole-blood-clotted, Sarstedt Monovette serum | Before spinninguncentrifuged | chilled in transit (insulated box, 2 frozen cool packs; temperature not recorded) | 24 hstable for the longest time tested | % difference within desirable quality specification for bias (Fraser, from intra- and interindividual variation) | Jones et al. 2021, J NutrTable 2, row 'β-cryptoxanthin', column Serum |
| whole-blood-edta, Sarstedt Monovette EDTA | Before spinninguncentrifuged | chilled in transit (insulated box, 2 frozen cool packs; temperature not recorded) | 24 hstable for the longest time tested | % difference within desirable quality specification for bias (Fraser, from intra- and interindividual variation) | Jones et al. 2021, J NutrTable 2, row 'β-cryptoxanthin', column EDTA |
| whole-blood-heparin, Sarstedt Monovette lithium heparin, trace-element free | Before spinninguncentrifuged | chilled in transit (insulated box, 2 frozen cool packs; temperature not recorded) | 24 hstable for the longest time tested | % difference within desirable quality specification for bias (Fraser, from intra- and interindividual variation) | Jones et al. 2021, J NutrTable 2, row 'β-cryptoxanthin', column LH |
Bicarbonate (Total CO2)6 findings
| Specimen, tube | Condition | Temperature | Stable for | Criterion | Study |
|---|---|---|---|---|---|
| plasma-li-heparin, BD Vacutainer LH PST II gel separator, 3.0 mL (REF 367374) | Storagecentrifuged-on-gel | 2-8 °C | 24 hstable for the longest time tested | mean % difference smaller than TEa (Ricos: 0.25 x (CVi^2 + CVg^2)^0.5 + 1.65 x 0.5 x CVi) | Nonkes et al. 2019, Biochem Med (Zagreb)Table 1, row 'Bicarbonate, mmol/L' |
| serum, BD Vacutainer SST (SKU 367986), gel | Storagecentrifuged-on-gel | 30 °C (±1 °C) | 24 hthen changed, changed by 72 h | mean bias vs paired baseline within CLIA acceptance limits AND >2/3 of individual results within limits | Iacovetti et al. 2026, Pract Lab MedResults, 'Analyte stability after storage at 30 °C', paragraph 1 (Table 1 image-only) |
| serum, BD SST Microtainer (SKU 395967), gel; filled from venous blood | Storagecentrifuged-on-gel | 0 °C | 72 hstable for the longest time tested | mean bias vs paired baseline within CLIA acceptance limits AND >2/3 of individual results within limits | Iacovetti et al. 2026, Pract Lab MedResults, 'Analyte stability after separation and storage in microvolume tubes at 0 °C' (Table 2 image-only) |
| whole-blood-clotted, BD Vacutainer SST (SKU 367986), gel | Before spinninguncentrifuged | 30 °C (±1 °C) | 24 hthen changed, changed by 72 h | mean bias vs paired baseline within CLIA acceptance limits AND >2/3 of individual results within limits | Iacovetti et al. 2026, Pract Lab MedResults, 'Analyte stability after storage at 30 °C', paragraph 2 (Table 1 image-only) |
| whole-blood-heparin, BD Vacutainer lithium heparin, no gel (ref 368884) | Before spinninguncentrifuged | RT | 10 hstable for the longest time tested | |mean % deviation 10 h vs baseline| > laboratory bias goal (Table II; mostly biological-variation based, some empirically modified) | Henriksen et al. 2014, Scand J Clin Lab InvestTable II, row 'CO2' |
| whole-blood-heparin, BD Vacutainer lithium heparin, no gel (ref 368884) | Before spinninguncentrifuged | RT | 10 hstable for the longest time tested | CV of paired baseline/10 h results > goal CV (Table II) | Henriksen et al. 2014, Scand J Clin Lab InvestTable II, row 'CO2' |
Blood Urea Nitrogen9 findings
| Specimen, tube | Condition | Temperature | Stable for | Criterion | Study |
|---|---|---|---|---|---|
| serum, BD Vacutainer 8.5 mL plastic serum tube; pooled serum aliquoted into 1.5 mL Eppendorf tubes | Freeze–thawseparated | −20 °C | 3 cyclesthen changed, changed by 4 cycles | unstable only if both P < 0.005 (Bonferroni) and % change from T0 > desirable bias (Westgard/Ricos), here 5.5% | Cuhadar et al. 2013, Biochem Med (Zagreb)Table 2, row 'BUN (mmol/L)', columns T1d-T10d |
| serum, BD Vacutainer 8.5 mL plastic serum tube; pooled serum aliquoted into 1.5 mL Eppendorf tubes | Storageseparated | −20 °C | 0 monthschanged at the first time point, changed by 1 months | unstable only if both P < 0.013 (Bonferroni) and % change from T0 > desirable bias (Westgard/Ricos), here 5.5% | Cuhadar et al. 2013, Biochem Med (Zagreb)Table 2, row 'BUN (mmol/L)', columns T1m-T3m |
| serum, Vacuette Tube Serum Gel Separator 8 mL (Greiner, 455071) | Storagecentrifuged-on-gel | 2-8 °C | 30 hthen changed, changed by 36 h | % bias of median vs baseline exceeds desirable bias (Westgard/Ricos BV), here 5.5% | Cuhadar et al. 2012, Biochem Med (Zagreb)Table 3, row 'Urea nitrogen', gel 4°C |
| serum, Vacuette Tube Serum No Gel 9 mL (Greiner, 455092), plain | Storagecentrifuged-on-cells | 2-8 °C | 0 hchanged at the first time point, changed by 6 h | % bias of median vs baseline exceeds desirable bias (Westgard/Ricos BV), here 5.5% | Cuhadar et al. 2012, Biochem Med (Zagreb)Table 3, row 'Urea nitrogen', w/o gel 4°C |
| serum, Vacuette Tube Serum Gel Separator 8 mL (Greiner, 455071) | Storagecentrifuged-on-gel | RT | 0 hchanged at the first time point, changed by 6 h | % bias of median vs baseline exceeds desirable bias (Westgard/Ricos BV), here 5.5% | Cuhadar et al. 2012, Biochem Med (Zagreb)Table 3, row 'Urea nitrogen', gel 24°C |
| serum, Vacuette Tube Serum No Gel 9 mL (Greiner, 455092), plain | Storagecentrifuged-on-cells | RT | 12 hthen changed, changed by 18 h | % bias of median vs baseline exceeds desirable bias (Westgard/Ricos BV), here 5.5% | Cuhadar et al. 2012, Biochem Med (Zagreb)Table 3, row 'Urea nitrogen', w/o gel 24°C |
| serum, BD Vacutainer SST (SKU 367986), gel | Storagecentrifuged-on-gel | 30 °C (±1 °C) | 72 hstable for the longest time tested | mean bias vs paired baseline within CLIA acceptance limits AND >2/3 of individual results within limits | Iacovetti et al. 2026, Pract Lab MedResults, 'Analyte stability after storage at 30 °C', paragraph 1 (Table 1 image-only) |
| serum, BD SST Microtainer (SKU 395967), gel; filled from venous blood | Storagecentrifuged-on-gel | 0 °C | 72 hstable for the longest time tested | mean bias vs paired baseline within CLIA acceptance limits AND >2/3 of individual results within limits | Iacovetti et al. 2026, Pract Lab MedResults, 'Analyte stability after separation and storage in microvolume tubes at 0 °C' (Table 2 image-only) |
| whole-blood-clotted, BD Vacutainer SST (SKU 367986), gel | Before spinninguncentrifuged | 30 °C (±1 °C) | 24 hthen changed, changed by 72 h | mean bias vs paired baseline within CLIA acceptance limits AND >2/3 of individual results within limits | Iacovetti et al. 2026, Pract Lab MedResults, 'Analyte stability after storage at 30 °C', paragraph 2 (Table 1 image-only) |
C-Peptide10 findings
| Specimen, tube | Condition | Temperature | Stable for | Criterion | Study |
|---|---|---|---|---|---|
| plasma-edta, Sarstedt K+-EDTA 2.7 mL | Freeze–thawnot-stated | −80 °C | 5 cyclesstable for the longest time tested | mean % change from baseline > 10% considered clinically significant | McDonald et al. 2012, PLoS OneTable 1 |
| plasma-edta, Sarstedt K+-EDTA 2.7 mL | Storagecentrifuged-on-cells | RT | 24 hstable for the longest time tested | mean % change from baseline > 10% considered clinically significant | McDonald et al. 2012, PLoS OneResults 1.1, paragraph 4 |
| serum, Sarstedt serum gel 2.7 mL | Freeze–thawnot-stated | −80 °C | 5 cyclesstable for the longest time tested | mean % change from baseline > 10% considered clinically significant | McDonald et al. 2012, PLoS OneTable 1 |
| serum, Sarstedt serum gel 2.7 mL | Storagecentrifuged-on-gel | RT | 24 hstable for the longest time tested | mean % change from baseline > 10% considered clinically significant | McDonald et al. 2012, PLoS OneResults 1.1, paragraph 4 |
| whole-blood-clotted, Sarstedt serum gel 2.7 mL | Before spinninguncentrifuged | 2-8 °C | 24 hstable for the longest time tested | mean % change from baseline > 10% considered clinically significant | McDonald et al. 2012, PLoS OneResults 1.1, paragraph 5 |
| whole-blood-clotted, Sarstedt serum gel 2.7 mL | Before spinninguncentrifuged | 2-8 °C | 120 hstable for the longest time tested | mean % change from baseline > 10% considered clinically significant | McDonald et al. 2012, PLoS OneResults 1.2 |
| whole-blood-edta, Sarstedt K+-EDTA 2.7 mL | Before spinninguncentrifuged | 2-8 °C | 24 hstable for the longest time tested | mean % change from baseline > 10% considered clinically significant | McDonald et al. 2012, PLoS OneResults 1.1, paragraph 5 |
| whole-blood-edta, Sarstedt K+-EDTA 2.7 mL | Before spinninguncentrifuged | 2-8 °C | 120 hstable for the longest time tested | mean % change from baseline > 10% considered clinically significant | McDonald et al. 2012, PLoS OneResults 1.2 |
| whole-blood-edta, Sarstedt K+-EDTA 2.7 mL | Before spinninguncentrifuged | RT | 24 hstable for the longest time tested | mean % change from baseline > 10% considered clinically significant | McDonald et al. 2012, PLoS OneResults 1.1, paragraph 4 |
| whole-blood-edta, Sarstedt K+-EDTA 2.7 mL | Before spinninguncentrifuged | RT | 36 hthen changed, changed by 48 h | mean % change from baseline > 10% considered clinically significant | McDonald et al. 2012, PLoS OneResults 1.2; Discussion paragraph 2 |
C-Reactive Protein20 findings
| Specimen, tube | Condition | Temperature | Stable for | Criterion | Study |
|---|---|---|---|---|---|
| dbs, Whatman CF12 filter paper, zip-lock plastic bag; venous EDTA whole blood spotted | Storagedried | −20 °C | 31 daysstable for the longest time tested | samples remain in the same clinical CRP category (0-3, 3+, 10+ mg/L) as on day 1 | Schakelaar et al. 2023, Sci RepResults, 'CRP stability up to 31 days'; Figure 2; Methods 'Time and temperature stability' |
| dbs, Whatman CF12 filter paper, zip-lock plastic bag; venous EDTA whole blood spotted | Storagedried | 2-8 °C | 31 daysstable for the longest time tested | samples remain in the same clinical CRP category (0-3, 3+, 10+ mg/L) as on day 1 | Schakelaar et al. 2023, Sci RepResults, 'CRP stability up to 31 days'; Figure 2; Methods 'Time and temperature stability' |
| dbs, Whatman CF12 filter paper, zip-lock plastic bag; venous EDTA whole blood spotted | Storagedried | RT | 31 daysstable for the longest time tested | samples remain in the same clinical CRP category (0-3, 3+, 10+ mg/L) as on day 1 | Schakelaar et al. 2023, Sci RepResults, 'CRP stability up to 31 days'; Figure 2; Methods 'Time and temperature stability' |
| dbs, Whatman 903 card, mailed in foil pouch with desiccant | Storagedried | not stated; ambient postal shipping in the US (Apr-Dec 2016) | 8 daysstable for the longest time tested | regression coefficient for days before freezing vs 0-2 days significant (P < .05), Model 1 (DBS value, unadjusted) | Crimmins et al. 2020, Am J Hum BiolTable 3, rows 'Days before freezing', column 'CRP' |
| dbs, Whatman 903 card, mailed in foil pouch with desiccant | Storagedried | not stated; ambient postal shipping in the US (Apr-Dec 2016) | 8 daysstable for the longest time tested | regression coefficient for days before freezing vs 0-2 days significant (P < .05), Model 2 (DBS value adjusted for venous value) | Crimmins et al. 2020, Am J Hum BiolTable 3, rows 'Days before freezing', column 'CRP' |
| dbs, Whatman CF12 filter paper, zip-lock plastic bag; venous EDTA whole blood spotted | Storagedried | + 30 °C | 31 daysstable for the longest time tested | samples remain in the same clinical CRP category (0-3, 3+, 10+ mg/L) as on day 1 | Schakelaar et al. 2023, Sci RepResults, 'CRP stability up to 31 days'; Figure 2; Methods 'Time and temperature stability' |
| plasma-li-heparin, BD Vacutainer lithium heparin 6 mL (regular) | Storagenot-stated | 2-8 °C | 72 hstable for the longest time tested | mean % difference vs T0 exceeds total change limit (TCL), here ±38.72% | Dupuy et al. 2018, Biochem Med (Zagreb)Table 2, BD lithium heparin tubes, row 'CRP, mg/L' |
| plasma-li-heparin, BD Barricor lithium heparin 5.5 mL (mechanical separator) | Storagecentrifuged-on-gel | 2-8 °C | 72 hstable for the longest time tested | mean % difference vs T0 exceeds total change limit (TCL), here ±38.72% | Dupuy et al. 2018, Biochem Med (Zagreb)Table 2, BD Barricor lithium heparin tubes, row 'CRP, mg/L' |
| plasma-li-heparin, BD Barricor 4.5 mL (lithium heparin, mechanical separator) | Storagecentrifuged-on-gel | 2-8 °C | 7 daysstable for the longest time tested | mean % difference vs time 0 exceeds desirable imprecision based on biological variation (EFLM/Westgard), here 16.75% | Mandić et al. 2020, Biochem Med (Zagreb)Tables 3 and 4, row 'CRP, mg/L', BD Barricor columns |
| serum, BD Vacutainer SST | Freeze–thawseparated | −70 °C, 1 month; thawed at room temperature | 1 cyclestable for the longest time tested | mean % change (bias) from fresh baseline within total allowable error (TEa, desirable bias) | Abraham et al. 2021, J Lab PhysiciansTable 2 and Table 3, row 'CRP (mg/dL)' |
| serum, BD Vacutainer RST 5 mL (thrombin clot activator, gel) | Storagecentrifuged-on-gel | 2-8 °C | 7 daysstable for the longest time tested | mean % difference vs time 0 exceeds desirable imprecision based on biological variation (EFLM/Westgard), here 16.75% | Mandić et al. 2020, Biochem Med (Zagreb)Tables 3 and 4, row 'CRP, mg/L', BD RST columns |
| whole-blood-clotted, BD Vacutainer SST | Before spinninguncentrifuged | RT | 24 hstable for the longest time tested | mean % deviation from baseline within total change limit TCL = sqrt((2.77 CVa)^2 + (0.5 CVb)^2), CVb from Ricos (Table 2) | Abraham et al. 2019, Biochem Med (Zagreb)Discussion, paragraph 1 (Results paragraphs 1-3; Figure 3) |
| whole-blood-clotted, BD Vacutainer trace element-free serum tube | Before spinninguncentrifuged | cold packs in insulated box; 9 (2) °C on arrival | 24 hstable for the longest time tested | % difference within desirable quality specification (from intra- and interindividual variation) | Jones et al. 2024, J NutrTable 2, row 'C-reactive protein, mg/L'; Table 1 |
| whole-blood-clotted, BD Vacutainer SST | Before spinninguncentrifuged | 22-30 °C (logger 24-26 °C) | 24 hstable for the longest time tested | mean % deviation from baseline within total change limit TCL = sqrt((2.77 CVa)^2 + (0.5 CVb)^2), CVb from Ricos (Table 2) | Abraham et al. 2019, Biochem Med (Zagreb)Discussion, paragraph 1 (Results paragraphs 1-3; Figure 3) |
| whole-blood-clotted, BD Vacutainer SST | Before spinninguncentrifuged | > 30 °C (logger 38-40 °C) | 24 hstable for the longest time tested | mean % deviation from baseline within total change limit TCL = sqrt((2.77 CVa)^2 + (0.5 CVb)^2), CVb from Ricos (Table 2) | Abraham et al. 2019, Biochem Med (Zagreb)Discussion, paragraph 1 (Results paragraphs 1-3; Figure 3) |
| whole-blood-clotted, Sarstedt Monovette serum | Before spinninguncentrifuged | chilled in transit (insulated box, 2 frozen cool packs; temperature not recorded) | 24 hstable for the longest time tested | % difference within desirable quality specification for bias (Fraser, from intra- and interindividual variation) | Jones et al. 2021, J NutrTable 2, row 'C-reactive protein', column Serum |
| whole-blood-edta, Sarstedt Monovette EDTA | Before spinninguncentrifuged | chilled in transit (insulated box, 2 frozen cool packs; temperature not recorded) | 24 hstable for the longest time tested | % difference within desirable quality specification for bias (Fraser, from intra- and interindividual variation) | Jones et al. 2021, J NutrTable 2, row 'C-reactive protein', column EDTA |
| whole-blood-heparin, BD Vacutainer lithium heparin, no gel (ref 368884) | Before spinninguncentrifuged | RT | 10 hstable for the longest time tested | |mean % deviation 10 h vs baseline| > laboratory bias goal (Table II; mostly biological-variation based, some empirically modified) | Henriksen et al. 2014, Scand J Clin Lab InvestTable II, row 'CRP' |
| whole-blood-heparin, BD Vacutainer lithium heparin, no gel (ref 368884) | Before spinninguncentrifuged | RT | 10 hstable for the longest time tested | CV of paired baseline/10 h results > goal CV (Table II) | Henriksen et al. 2014, Scand J Clin Lab InvestTable II, row 'CRP' |
| whole-blood-heparin, Sarstedt Monovette lithium heparin, trace-element free | Before spinninguncentrifuged | chilled in transit (insulated box, 2 frozen cool packs; temperature not recorded) | 24 hstable for the longest time tested | % difference within desirable quality specification for bias (Fraser, from intra- and interindividual variation) | Jones et al. 2021, J NutrTable 2, row 'C-reactive protein', column LH |
Calcium, total15 findings
| Specimen, tube | Condition | Temperature | Stable for | Criterion | Study |
|---|---|---|---|---|---|
| plasma-li-heparin, Sarstedt lithium heparin, no gel separator (primary tube) | Storagecentrifuged-on-cells | 2-8 °C | 48 hstable for the longest time tested | mean % difference vs time 0 exceeds total limit of change, TLC = ±√((1.65·CVa)² + (0.5·CVb)²) (EFLM BV), here 4.6%; stability limit from instability-equation regression | Illana et al. 2024, Biochem Med (Zagreb)Table 1, row 'Ca' |
| plasma-li-heparin, Sarstedt lithium heparin, no gel separator; plasma aliquot | Storageseparated | 2-8 °C | 96 hstable for the longest time tested | mean % difference vs time 0 exceeds total limit of change, TLC = ±√((1.65·CVa)² + (0.5·CVb)²) (EFLM BV), here 4.6%; stability limit from instability-equation regression | Illana et al. 2024, Biochem Med (Zagreb)Table 2, row 'Ca' |
| plasma-li-heparin, BD Vacutainer LH PST II gel separator, 3.0 mL (REF 367374) | Storagecentrifuged-on-gel | 2-8 °C | 24 hstable for the longest time tested | mean % difference smaller than TEa (Ricos: 0.25 x (CVi^2 + CVg^2)^0.5 + 1.65 x 0.5 x CVi) | Nonkes et al. 2019, Biochem Med (Zagreb)Table 1, row 'Calcium, mmol/L' |
| plasma-li-heparin, BD Barricor 4.5 mL (lithium heparin, mechanical separator) | Storagecentrifuged-on-gel | 2-8 °C | 24 hthen changed, changed by 7 days | mean % difference vs time 0 exceeds desirable imprecision based on biological variation (EFLM/Westgard), here 1.05% | Mandić et al. 2020, Biochem Med (Zagreb)Tables 3 and 4, row 'Calcium, mmol/L', BD Barricor columns |
| serum, BD Vacutainer 8.5 mL plastic serum tube; pooled serum aliquoted into 1.5 mL Eppendorf tubes | Freeze–thawseparated | −20 °C | 3 cyclesthen changed, changed by 4 cycles | unstable only if both P < 0.005 (Bonferroni) and % change from T0 > desirable bias (Westgard/Ricos), here 0.8% | Cuhadar et al. 2013, Biochem Med (Zagreb)Table 2, row 'Calcium (mmol/L)', columns T1d-T10d |
| serum, BD Vacutainer 8.5 mL plastic serum tube; pooled serum aliquoted into 1.5 mL Eppendorf tubes | Storageseparated | −20 °C | 3 monthsstable for the longest time tested | unstable only if both P < 0.013 (Bonferroni) and % change from T0 > desirable bias (Westgard/Ricos), here 0.8% | Cuhadar et al. 2013, Biochem Med (Zagreb)Table 2, row 'Calcium (mmol/L)', columns T1m-T3m |
| serum, BD Vacutainer serum (plain), 7.5 mL | Storageseparated | −20 °C | 30 daysstable for the longest time tested | clinically significant change: mean outside significant change limit (SCL = T1d mean ± 3.0 usual SD from 2 months QC) | Kachhawa et al. 2017, J Lab PhysiciansResults paragraph 1 (Table 2 image-only) |
| serum, BD Vacutainer RST 4.0 mL (thrombin clot activator, gel) | Storagecentrifuged-on-gel | 2-8 °C | 24 hstable for the longest time tested | clinically significant if mean bias vs t0 exceeds desirable bias (biological variation), here 0.8% | Kocijancic et al. 2014, Biochem Med (Zagreb)Table 3, row 'Ca' |
| serum, BD Vacutainer SST II Advance 5.0 mL (clot activator, gel) | Storagecentrifuged-on-gel | 2-8 °C | 24 hstable for the longest time tested | clinically significant if mean bias vs t0 exceeds desirable bias (biological variation), here 0.8% | Kocijancic et al. 2014, Biochem Med (Zagreb)Table 4, row 'Ca' |
| serum, BD Vacutainer RST 5 mL (thrombin clot activator, gel) | Storagecentrifuged-on-gel | 2-8 °C | 24 hthen changed, changed by 7 days | mean % difference vs time 0 exceeds desirable imprecision based on biological variation (EFLM/Westgard), here 1.05% | Mandić et al. 2020, Biochem Med (Zagreb)Tables 3 and 4, row 'Calcium, mmol/L', BD RST columns |
| serum, BD Vacutainer SST (SKU 367986), gel | Storagecentrifuged-on-gel | 30 °C (±1 °C) | 72 hstable for the longest time tested | mean bias vs paired baseline within CLIA acceptance limits AND >2/3 of individual results within limits | Iacovetti et al. 2026, Pract Lab MedResults, 'Analyte stability after storage at 30 °C', paragraph 1 (Table 1 image-only) |
| serum, BD SST Microtainer (SKU 395967), gel; filled from venous blood | Storagecentrifuged-on-gel | 0 °C | 72 hstable for the longest time tested | mean bias vs paired baseline within CLIA acceptance limits AND >2/3 of individual results within limits | Iacovetti et al. 2026, Pract Lab MedResults, 'Analyte stability after separation and storage in microvolume tubes at 0 °C' (Table 2 image-only) |
| whole-blood-clotted, BD Vacutainer SST (SKU 367986), gel | Before spinninguncentrifuged | 30 °C (±1 °C) | 24 hthen changed, changed by 72 h | mean bias vs paired baseline within CLIA acceptance limits AND >2/3 of individual results within limits | Iacovetti et al. 2026, Pract Lab MedResults, 'Analyte stability after storage at 30 °C', paragraph 2 (Table 1 image-only) |
| whole-blood-heparin, BD Vacutainer lithium heparin, no gel (ref 368884) | Before spinninguncentrifuged | RT | 10 hstable for the longest time tested | |mean % deviation 10 h vs baseline| > laboratory bias goal (Table II; mostly biological-variation based, some empirically modified) | Henriksen et al. 2014, Scand J Clin Lab InvestTable II, row 'CA' |
| whole-blood-heparin, BD Vacutainer lithium heparin, no gel (ref 368884) | Before spinninguncentrifuged | RT | 10 hstable for the longest time tested | CV of paired baseline/10 h results > goal CV (Table II) | Henriksen et al. 2014, Scand J Clin Lab InvestTable II, row 'CA' |
Chloride11 findings
| Specimen, tube | Condition | Temperature | Stable for | Criterion | Study |
|---|---|---|---|---|---|
| plasma-li-heparin, Sarstedt lithium heparin, no gel separator (primary tube) | Storagecentrifuged-on-cells | 2-8 °C | 48 hstable for the longest time tested | mean % difference vs time 0 exceeds total limit of change, TLC = ±√((1.65·CVa)² + (0.5·CVb)²) (EFLM BV), here 4.8%; stability limit from instability-equation regression | Illana et al. 2024, Biochem Med (Zagreb)Table 1, row 'Cl' |
| plasma-li-heparin, Sarstedt lithium heparin, no gel separator; plasma aliquot | Storageseparated | 2-8 °C | 96 hstable for the longest time tested | mean % difference vs time 0 exceeds total limit of change, TLC = ±√((1.65·CVa)² + (0.5·CVb)²) (EFLM BV), here 4.8%; stability limit from instability-equation regression | Illana et al. 2024, Biochem Med (Zagreb)Table 2, row 'Cl' |
| plasma-li-heparin, BD Vacutainer PST II lithium heparin 3 mL (gel) | Storagecentrifuged-on-gel | 2-8 °C | 15 hstable for the longest time tested | clinically significant when the 95% CI of mean bias lies completely outside the SEQC maximum allowable bias (MAB = 1.65 x CVa), here 1% | Balbás et al. 2017, Biochem Med (Zagreb)Table 2, Cl- row 'PST II sealed, 4 şC' |
| plasma-li-heparin, BD Barricor lithium heparin 5.5 mL (mechanical separator) | Storagecentrifuged-on-gel | 2-8 °C | 15 hstable for the longest time tested | clinically significant when the 95% CI of mean bias lies completely outside the SEQC maximum allowable bias (MAB = 1.65 x CVa), here 1% | Balbás et al. 2017, Biochem Med (Zagreb)Table 2, Cl- row 'Barricor sealed, 4 şC' |
| plasma-li-heparin, BD Barricor 4.5 mL (lithium heparin, mechanical separator) | Storagecentrifuged-on-gel | 2-8 °C | 7 daysstable for the longest time tested | mean % difference vs time 0 exceeds desirable imprecision based on biological variation (EFLM/Westgard), here 0.55% | Mandić et al. 2020, Biochem Med (Zagreb)Tables 3 and 4, row 'Chloride, mmol/L', BD Barricor columns |
| plasma-li-heparin, BD Vacutainer PST II lithium heparin 3 mL (gel) | Storagecentrifuged-on-gel | RT | 3 hthen changed, changed by 6 h | clinically significant when the 95% CI of mean bias lies completely outside the SEQC maximum allowable bias (MAB = 1.65 x CVa), here 1% | Balbás et al. 2017, Biochem Med (Zagreb)Table 2, Cl- row 'PST II open, rt' |
| plasma-li-heparin, BD Barricor lithium heparin 5.5 mL (mechanical separator) | Storagecentrifuged-on-gel | RT | 0 hchanged at the first time point, changed by 3 h | clinically significant when the 95% CI of mean bias lies completely outside the SEQC maximum allowable bias (MAB = 1.65 x CVa), here 1% | Balbás et al. 2017, Biochem Med (Zagreb)Table 2, Cl- row 'Barricor open, rt' |
| serum, BD Vacutainer RST 5 mL (thrombin clot activator, gel) | Storagecentrifuged-on-gel | 2-8 °C | 7 daysstable for the longest time tested | mean % difference vs time 0 exceeds desirable imprecision based on biological variation (EFLM/Westgard), here 0.55% | Mandić et al. 2020, Biochem Med (Zagreb)Tables 3 and 4, row 'Chloride, mmol/L', BD RST columns |
| serum, BD Vacutainer SST (SKU 367986), gel | Storagecentrifuged-on-gel | 30 °C (±1 °C) | 72 hstable for the longest time tested | mean bias vs paired baseline within CLIA acceptance limits AND >2/3 of individual results within limits | Iacovetti et al. 2026, Pract Lab MedResults, 'Analyte stability after storage at 30 °C', paragraph 1 (Table 1 image-only) |
| serum, BD SST Microtainer (SKU 395967), gel; filled from venous blood | Storagecentrifuged-on-gel | 0 °C | 72 hstable for the longest time tested | mean bias vs paired baseline within CLIA acceptance limits AND >2/3 of individual results within limits | Iacovetti et al. 2026, Pract Lab MedResults, 'Analyte stability after separation and storage in microvolume tubes at 0 °C' (Table 2 image-only) |
| whole-blood-clotted, BD Vacutainer SST (SKU 367986), gel | Before spinninguncentrifuged | 30 °C (±1 °C) | 0 hchanged at the first time point, changed by 24 h | mean bias vs paired baseline within CLIA acceptance limits AND >2/3 of individual results within limits | Iacovetti et al. 2026, Pract Lab MedResults, 'Analyte stability after storage at 30 °C', paragraph 2 (Table 1 image-only) |
Coagulation factor II5 findings
| Specimen, tube | Condition | Temperature | Stable for | Criterion | Study |
|---|---|---|---|---|---|
| plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquots | Freeze–thawseparated | −20 °C | 1 cyclelimit stated by the authors | max allowable mean deviation from t0: ±15% | Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability' |
| plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquots | Freeze–thawseparated | −80 °C | 3 cycleslimit stated by the authors | max allowable mean deviation from t0: ±15% | Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability' |
| plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquots | Storageseparated | −20 °C | 28 dayslimit stated by the authors | max allowable mean deviation from t0: ±15% | Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability' |
| plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquots | Storageseparated | −80 °C | 12 monthslimit stated by the authors | max allowable mean deviation from t0: ±15% | Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability' |
| plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquots | Storageseparated | RT | 4 hlimit stated by the authors | max allowable mean deviation from t0: ±15% | Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability' |
Coagulation factor IX8 findings
| Specimen, tube | Condition | Temperature | Stable for | Criterion | Study |
|---|---|---|---|---|---|
| plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquots | Freeze–thawseparated | −20 °C | 1 cyclelimit stated by the authors | max allowable mean deviation from t0: ≤1.25%: ±0.250% (absolute); >1.25–≤30.0%: ±20.0% (relative); >30.0%: ±15.0% | Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability' |
| plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquots | Freeze–thawseparated | −80 °C | 3 cycleslimit stated by the authors | max allowable mean deviation from t0: ≤1.25%: ±0.250% (absolute); >1.25–≤30.0%: ±20.0% (relative); >30.0%: ±15.0% | Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability' |
| plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquots | Storageseparated | −20 °C | 14 dayslimit stated by the authors | max allowable mean deviation from t0: ≤1.25%: ±0.250% (absolute); >1.25–≤30.0%: ±20.0% (relative); >30.0%: ±15.0% | Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability' |
| plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquots | Storageseparated | −80 °C | 28 dayslimit stated by the authors | max allowable mean deviation from t0: ≤1.25%: ±0.250% (absolute); >1.25–≤30.0%: ±20.0% (relative); >30.0%: ±15.0% | Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability' |
| plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquots | Storageseparated | −80 °C | 24 monthslimit stated by the authors | max allowable mean deviation from t0: ≤1.25%: ±0.250% (absolute); >1.25–≤30.0%: ±20.0% (relative); >30.0%: ±15.0% | Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability' |
| plasma-citrate, BD 0.109 M sodium citrate (9:1); plasma aliquoted into nonactivating Eppendorf tubes | Storageseparated | 2-8 °C | 4 hthen changed, changed by 6 h | mean % change from baseline >10%, or >25% of samples with individual change >10% (van Geest-Daalderop) | Feng et al. 2014, Sci RepTable 2, 'Acceptable Time (h)' |
| plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquots | Storageseparated | RT | 4 hlimit stated by the authors | max allowable mean deviation from t0: ≤1.25%: ±0.250% (absolute); >1.25–≤30.0%: ±20.0% (relative); >30.0%: ±15.0% | Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability' |
| plasma-citrate, BD 0.109 M sodium citrate (9:1); plasma aliquoted into nonactivating Eppendorf tubes | Storageseparated | RT | 4 hthen changed, changed by 6 h | mean % change from baseline >10%, or >25% of samples with individual change >10% (van Geest-Daalderop) | Feng et al. 2014, Sci RepTable 2, 'Acceptable Time (h)' |
Coagulation factor V4 findings
| Specimen, tube | Condition | Temperature | Stable for | Criterion | Study |
|---|---|---|---|---|---|
| plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquots | Freeze–thawseparated | −80 °C | 1 cyclelimit stated by the authors | max allowable mean deviation from t0: ±15% | Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability' |
| plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquots | Storageseparated | −80 °C | 28 dayslimit stated by the authors | max allowable mean deviation from t0: ±15% | Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability' |
| plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquots | Storageseparated | −80 °C | 24 monthslimit stated by the authors | max allowable mean deviation from t0: ±15% | Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability' |
| plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquots | Storageseparated | RT | 4 hlimit stated by the authors | max allowable mean deviation from t0: ±15% | Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability' |
Coagulation factor VII5 findings
| Specimen, tube | Condition | Temperature | Stable for | Criterion | Study |
|---|---|---|---|---|---|
| plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquots | Freeze–thawseparated | −20 °C | 1 cyclelimit stated by the authors | max allowable mean deviation from t0: ±15% | Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability' |
| plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquots | Freeze–thawseparated | −80 °C | 3 cycleslimit stated by the authors | max allowable mean deviation from t0: ±15% | Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability' |
| plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquots | Storageseparated | −20 °C | 28 dayslimit stated by the authors | max allowable mean deviation from t0: ±15% | Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability' |
| plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquots | Storageseparated | −80 °C | 24 monthslimit stated by the authors | max allowable mean deviation from t0: ±15% | Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability' |
| plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquots | Storageseparated | RT | 4 hlimit stated by the authors | max allowable mean deviation from t0: ±15% | Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability' |
Coagulation factor VIII5 findings
| Specimen, tube | Condition | Temperature | Stable for | Criterion | Study |
|---|---|---|---|---|---|
| plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquots | Freeze–thawseparated | −20 °C | 1 cyclelimit stated by the authors | max allowable mean deviation from t0: ±15% | Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability' |
| plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquots | Storageseparated | −20 °C | 14 dayslimit stated by the authors | max allowable mean deviation from t0: ±15% | Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability' |
| plasma-citrate, BD 0.109 M sodium citrate (9:1); plasma aliquoted into nonactivating Eppendorf tubes | Storageseparated | 2-8 °C | 2 hthen changed, changed by 4 h | mean % change from baseline >10%, or >25% of samples with individual change >10% (van Geest-Daalderop) | Feng et al. 2014, Sci RepTable 2, 'Acceptable Time (h)' |
| plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquots | Storageseparated | RT | 2 hlimit stated by the authors | max allowable mean deviation from t0: ±15% | Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability' |
| plasma-citrate, BD 0.109 M sodium citrate (9:1); plasma aliquoted into nonactivating Eppendorf tubes | Storageseparated | RT | 2 hthen changed, changed by 4 h | mean % change from baseline >10%, or >25% of samples with individual change >10% (van Geest-Daalderop) | Feng et al. 2014, Sci RepTable 2, 'Acceptable Time (h)' |
Coagulation factor X5 findings
| Specimen, tube | Condition | Temperature | Stable for | Criterion | Study |
|---|---|---|---|---|---|
| plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquots | Freeze–thawseparated | −20 °C | 1 cyclelimit stated by the authors | max allowable mean deviation from t0: ±10% | Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability' |
| plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquots | Freeze–thawseparated | −80 °C | 1 cyclelimit stated by the authors | max allowable mean deviation from t0: ±10% | Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability' |
| plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquots | Storageseparated | −20 °C | 28 dayslimit stated by the authors | max allowable mean deviation from t0: ±10% | Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability' |
| plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquots | Storageseparated | −80 °C | 24 monthslimit stated by the authors | max allowable mean deviation from t0: ±10% | Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability' |
| plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquots | Storageseparated | RT | 4 hlimit stated by the authors | max allowable mean deviation from t0: ±10% | Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability' |
Coagulation factor XI5 findings
| Specimen, tube | Condition | Temperature | Stable for | Criterion | Study |
|---|---|---|---|---|---|
| plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquots | Freeze–thawseparated | −20 °C | 1 cyclelimit stated by the authors | max allowable mean deviation from t0: ±15% | Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability' |
| plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquots | Freeze–thawseparated | −80 °C | 3 cycleslimit stated by the authors | max allowable mean deviation from t0: ±15% | Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability' |
| plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquots | Storageseparated | −20 °C | 28 dayslimit stated by the authors | max allowable mean deviation from t0: ±15% | Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability' |
| plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquots | Storageseparated | −80 °C | 24 monthslimit stated by the authors | max allowable mean deviation from t0: ±15% | Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability' |
| plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquots | Storageseparated | RT | 4 hlimit stated by the authors | max allowable mean deviation from t0: ±15% | Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability' |
Coagulation factor XII6 findings
| Specimen, tube | Condition | Temperature | Stable for | Criterion | Study |
|---|---|---|---|---|---|
| plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquots | Freeze–thawseparated | −20 °C | 1 cyclelimit stated by the authors | max allowable mean deviation from t0: ±15% | Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability' |
| plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquots | Freeze–thawseparated | −80 °C | 3 cycleslimit stated by the authors | max allowable mean deviation from t0: ±15% | Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability' |
| plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquots | Storageseparated | −20 °C | 28 dayslimit stated by the authors | max allowable mean deviation from t0: ±15% | Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability' |
| plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquots | Storageseparated | −80 °C | 28 dayslimit stated by the authors | max allowable mean deviation from t0: ±15% | Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability' |
| plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquots | Storageseparated | −80 °C | 6 monthslimit stated by the authors | max allowable mean deviation from t0: ±15% | Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability' |
| plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquots | Storageseparated | RT | 4 hlimit stated by the authors | max allowable mean deviation from t0: ±15% | Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability' |
Coagulation factor XIII6 findings
| Specimen, tube | Condition | Temperature | Stable for | Criterion | Study |
|---|---|---|---|---|---|
| plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquots | Freeze–thawseparated | −20 °C | 1 cyclelimit stated by the authors | max allowable mean deviation from t0: 1.57–20.0% FXIII antigen potency: ≤±3.00% (absolute); >20.0–109% FXIII antigen potency: ≤±15.0% (relative) | Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability' |
| plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquots | Freeze–thawseparated | −80 °C | 3 cycleslimit stated by the authors | max allowable mean deviation from t0: 1.57–20.0% FXIII antigen potency: ≤±3.00% (absolute); >20.0–109% FXIII antigen potency: ≤±15.0% (relative) | Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability' |
| plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquots | Storageseparated | −20 °C | 28 dayslimit stated by the authors | max allowable mean deviation from t0: 1.57–20.0% FXIII antigen potency: ≤±3.00% (absolute); >20.0–109% FXIII antigen potency: ≤±15.0% (relative) | Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability' |
| plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquots | Storageseparated | −80 °C | 28 dayslimit stated by the authors | max allowable mean deviation from t0: 1.57–20.0% FXIII antigen potency: ≤±3.00% (absolute); >20.0–109% FXIII antigen potency: ≤±15.0% (relative) | Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability' |
| plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquots | Storageseparated | −80 °C | 24 monthslimit stated by the authors | max allowable mean deviation from t0: 1.57–20.0% FXIII antigen potency: ≤±3.00% (absolute); >20.0–109% FXIII antigen potency: ≤±15.0% (relative) | Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability' |
| plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquots | Storageseparated | RT | 4 hlimit stated by the authors | max allowable mean deviation from t0: 1.57–20.0% FXIII antigen potency: ≤±3.00% (absolute); >20.0–109% FXIII antigen potency: ≤±15.0% (relative) | Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability' |
Creatine Kinase MB Isoenzyme2 findings
| Specimen, tube | Condition | Temperature | Stable for | Criterion | Study |
|---|---|---|---|---|---|
| plasma-li-heparin, BD Barricor 4.5 mL (lithium heparin, mechanical separator) | Storagecentrifuged-on-gel | 2-8 °C | 24 hthen changed, changed by 7 days | mean % difference vs time 0 exceeds desirable imprecision based on biological variation (EFLM/Westgard), here 9.9% | Mandić et al. 2020, Biochem Med (Zagreb)Tables 3 and 4, row 'CK-MB, U/L', BD Barricor columns |
| serum, BD Vacutainer RST 5 mL (thrombin clot activator, gel) | Storagecentrifuged-on-gel | 2-8 °C | 24 hthen changed, changed by 7 days | mean % difference vs time 0 exceeds desirable imprecision based on biological variation (EFLM/Westgard), here 9.9% | Mandić et al. 2020, Biochem Med (Zagreb)Tables 3 and 4, row 'CK-MB, U/L', BD RST columns |
Creatine Kinase Total12 findings
| Specimen, tube | Condition | Temperature | Stable for | Criterion | Study |
|---|---|---|---|---|---|
| plasma-li-heparin, Sarstedt lithium heparin, no gel separator (primary tube) | Storagecentrifuged-on-cells | 2-8 °C | 48 hstable for the longest time tested | mean % difference vs time 0 exceeds total limit of change, TLC = ±√((1.65·CVa)² + (0.5·CVb)²) (EFLM BV), here 9.5%; stability limit from instability-equation regression | Illana et al. 2024, Biochem Med (Zagreb)Table 1, row 'CK' |
| plasma-li-heparin, Sarstedt lithium heparin, no gel separator; plasma aliquot | Storageseparated | 2-8 °C | 96 hstable for the longest time tested | mean % difference vs time 0 exceeds total limit of change, TLC = ±√((1.65·CVa)² + (0.5·CVb)²) (EFLM BV), here 9.5%; stability limit from instability-equation regression | Illana et al. 2024, Biochem Med (Zagreb)Table 2, row 'CK' |
| plasma-li-heparin, BD Barricor 4.5 mL (lithium heparin, mechanical separator) | Storagecentrifuged-on-gel | 2-8 °C | 7 daysstable for the longest time tested | mean % difference vs time 0 exceeds desirable imprecision based on biological variation (EFLM/Westgard), here 8.0% | Mandić et al. 2020, Biochem Med (Zagreb)Tables 3 and 4, row 'CK, U/L', BD Barricor columns |
| serum, BD Vacutainer 8.5 mL plastic serum tube; pooled serum aliquoted into 1.5 mL Eppendorf tubes | Freeze–thawseparated | −20 °C | 10 cyclesstable for the longest time tested | unstable only if both P < 0.005 (Bonferroni) and % change from T0 > desirable bias (Westgard/Ricos), here 11.5% | Cuhadar et al. 2013, Biochem Med (Zagreb)Table 2, row 'CK (U/L)', columns T1d-T10d |
| serum, BD Vacutainer 8.5 mL plastic serum tube; pooled serum aliquoted into 1.5 mL Eppendorf tubes | Storageseparated | −20 °C | 3 monthsstable for the longest time tested | unstable only if both P < 0.013 (Bonferroni) and % change from T0 > desirable bias (Westgard/Ricos), here 11.5% | Cuhadar et al. 2013, Biochem Med (Zagreb)Table 2, row 'CK (U/L)', columns T1m-T3m |
| serum, Vacuette Tube Serum Gel Separator 8 mL (Greiner, 455071) | Storagecentrifuged-on-gel | 2-8 °C | 1 weekstable for the longest time tested | % bias of median vs baseline exceeds desirable bias (Westgard/Ricos BV), here 11.5% | Cuhadar et al. 2012, Biochem Med (Zagreb)Table 3, row 'Creatine kinase', gel 4°C |
| serum, Vacuette Tube Serum No Gel 9 mL (Greiner, 455092), plain | Storagecentrifuged-on-cells | 2-8 °C | 1 weekstable for the longest time tested | % bias of median vs baseline exceeds desirable bias (Westgard/Ricos BV), here 11.5% | Cuhadar et al. 2012, Biochem Med (Zagreb)Table 3, row 'Creatine kinase', w/o gel 4°C |
| serum, BD Vacutainer RST 5 mL (thrombin clot activator, gel) | Storagecentrifuged-on-gel | 2-8 °C | 7 daysstable for the longest time tested | mean % difference vs time 0 exceeds desirable imprecision based on biological variation (EFLM/Westgard), here 8.0% | Mandić et al. 2020, Biochem Med (Zagreb)Tables 3 and 4, row 'CK, U/L', BD RST columns |
| serum, Vacuette Tube Serum Gel Separator 8 mL (Greiner, 455071) | Storagecentrifuged-on-gel | RT | 72 hthen changed, changed by 1 week | % bias of median vs baseline exceeds desirable bias (Westgard/Ricos BV), here 11.5% | Cuhadar et al. 2012, Biochem Med (Zagreb)Table 3, row 'Creatine kinase', gel 24°C |
| serum, Vacuette Tube Serum No Gel 9 mL (Greiner, 455092), plain | Storagecentrifuged-on-cells | RT | 72 hthen changed, changed by 1 week | % bias of median vs baseline exceeds desirable bias (Westgard/Ricos BV), here 11.5% | Cuhadar et al. 2012, Biochem Med (Zagreb)Table 3, row 'Creatine kinase', w/o gel 24°C |
| whole-blood-heparin, BD Vacutainer lithium heparin, no gel (ref 368884) | Before spinninguncentrifuged | RT | 10 hstable for the longest time tested | |mean % deviation 10 h vs baseline| > laboratory bias goal (Table II; mostly biological-variation based, some empirically modified) | Henriksen et al. 2014, Scand J Clin Lab InvestTable II, row 'CK' |
| whole-blood-heparin, BD Vacutainer lithium heparin, no gel (ref 368884) | Before spinninguncentrifuged | RT | 10 hstable for the longest time tested | CV of paired baseline/10 h results > goal CV (Table II) | Henriksen et al. 2014, Scand J Clin Lab InvestTable II, row 'CK' |
Creatinine22 findings
| Specimen, tube | Condition | Temperature | Stable for | Criterion | Study |
|---|---|---|---|---|---|
| other, BD Microtainer SST (capillary, lancet) | Before spinninguncentrifuged | RT | 3 daysstable for the longest time tested | Bland-Altman bias within total allowable error for creatinine (8%) and r > 0.95 | Woolley et al. 2023, Br J Biomed SciResults (Table 1, Figure 1) and Discussion paragraph 4 |
| other, BD Microtainer SST (capillary, lancet) | Before spinninguncentrifuged | 24 h at 2°C–8°C, 24 h at ambient (16°C–22°C) and 24 h 37°C | 3 daysstable for the longest time tested | correlation r > 0.95 and agreement within total allowable error for creatinine (8%) | Woolley et al. 2023, Br J Biomed SciTable 2 and Discussion paragraph 5 |
| plasma-li-heparin, Sarstedt lithium heparin, no gel separator (primary tube) | Storagecentrifuged-on-cells | 2-8 °C | 48 hstable for the longest time tested | mean % difference vs time 0 exceeds total limit of change, TLC = ±√((1.65·CVa)² + (0.5·CVb)²) (EFLM BV), here 10.1%; stability limit from instability-equation regression | Illana et al. 2024, Biochem Med (Zagreb)Table 1, row 'CREA' |
| plasma-li-heparin, Sarstedt lithium heparin, no gel separator; plasma aliquot | Storageseparated | 2-8 °C | 96 hstable for the longest time tested | mean % difference vs time 0 exceeds total limit of change, TLC = ±√((1.65·CVa)² + (0.5·CVb)²) (EFLM BV), here 10.1%; stability limit from instability-equation regression | Illana et al. 2024, Biochem Med (Zagreb)Table 2, row 'CREA' |
| plasma-li-heparin, BD Barricor 4.5 mL (lithium heparin, mechanical separator) | Storagecentrifuged-on-gel | 2-8 °C | 7 daysstable for the longest time tested | mean % difference vs time 0 exceeds desirable imprecision based on biological variation (EFLM/Westgard), here 2.25% | Mandić et al. 2020, Biochem Med (Zagreb)Tables 3 and 4, row 'Creatinine, μmol/L', BD Barricor columns |
| serum, BD Vacutainer 8.5 mL plastic serum tube; pooled serum aliquoted into 1.5 mL Eppendorf tubes | Freeze–thawseparated | −20 °C | 10 cyclesstable for the longest time tested | unstable only if both P < 0.005 (Bonferroni) and % change from T0 > desirable bias (Westgard/Ricos), here 4.0% | Cuhadar et al. 2013, Biochem Med (Zagreb)Table 2, row 'Creatinine (μmol/L)', columns T1d-T10d |
| serum, BD Vacutainer SST | Freeze–thawseparated | −70 °C, 1 month; thawed at room temperature | 1 cyclestable for the longest time tested | mean % change (bias) from fresh baseline within total allowable error (TEa, desirable bias) | Abraham et al. 2021, J Lab PhysiciansTable 2 and Table 3, row 'Serum creatinine (mg/dL)' |
| serum, BD Vacutainer 8.5 mL plastic serum tube; pooled serum aliquoted into 1.5 mL Eppendorf tubes | Storageseparated | −20 °C | 3 monthsstable for the longest time tested | unstable only if both P < 0.013 (Bonferroni) and % change from T0 > desirable bias (Westgard/Ricos), here 4.0% | Cuhadar et al. 2013, Biochem Med (Zagreb)Table 2, row 'Creatinine (μmol/L)', columns T1m-T3m |
| serum, BD Vacutainer serum (plain), 7.5 mL | Storageseparated | −20 °C | 30 daysstable for the longest time tested | clinically significant change: mean outside significant change limit (SCL = T1d mean ± 3.0 usual SD from 2 months QC) | Kachhawa et al. 2017, J Lab PhysiciansResults paragraph 1 (Table 2 image-only) |
| serum, Vacuette Tube Serum Gel Separator 8 mL (Greiner, 455071) | Storagecentrifuged-on-gel | 2-8 °C | 1 weekstable for the longest time tested | % bias of median vs baseline exceeds desirable bias (Westgard/Ricos BV), here 4.0% | Cuhadar et al. 2012, Biochem Med (Zagreb)Table 3, row 'Creatinine', gel 4°C |
| serum, Vacuette Tube Serum No Gel 9 mL (Greiner, 455092), plain | Storagecentrifuged-on-cells | 2-8 °C | 48 hthen changed, changed by 72 h | % bias of median vs baseline exceeds desirable bias (Westgard/Ricos BV), here 4.0% | Cuhadar et al. 2012, Biochem Med (Zagreb)Table 3, row 'Creatinine', w/o gel 4°C |
| serum, BD Vacutainer RST 5 mL (thrombin clot activator, gel) | Storagecentrifuged-on-gel | 2-8 °C | 7 daysstable for the longest time tested | mean % difference vs time 0 exceeds desirable imprecision based on biological variation (EFLM/Westgard), here 2.25% | Mandić et al. 2020, Biochem Med (Zagreb)Tables 3 and 4, row 'Creatinine, μmol/L', BD RST columns |
| serum, Vacuette Tube Serum Gel Separator 8 mL (Greiner, 455071) | Storagecentrifuged-on-gel | RT | 18 hthen changed, changed by 24 h | % bias of median vs baseline exceeds desirable bias (Westgard/Ricos BV), here 4.0% | Cuhadar et al. 2012, Biochem Med (Zagreb)Table 3, row 'Creatinine', gel 24°C |
| serum, Vacuette Tube Serum No Gel 9 mL (Greiner, 455092), plain | Storagecentrifuged-on-cells | RT | 1 weekstable for the longest time tested | % bias of median vs baseline exceeds desirable bias (Westgard/Ricos BV), here 4.0% | Cuhadar et al. 2012, Biochem Med (Zagreb)Table 3, row 'Creatinine', w/o gel 24°C |
| serum, BD Vacutainer SST (SKU 367986), gel | Storagecentrifuged-on-gel | 30 °C (±1 °C) | 24 hthen changed, changed by 72 h | mean bias vs paired baseline within CLIA acceptance limits AND >2/3 of individual results within limits | Iacovetti et al. 2026, Pract Lab MedResults, 'Analyte stability after storage at 30 °C', paragraph 1 (Table 1 image-only) |
| serum, BD SST Microtainer (SKU 395967), gel; filled from venous blood | Storagecentrifuged-on-gel | 0 °C | 72 hstable for the longest time tested | mean bias vs paired baseline within CLIA acceptance limits AND >2/3 of individual results within limits | Iacovetti et al. 2026, Pract Lab MedResults, 'Analyte stability after separation and storage in microvolume tubes at 0 °C' (Table 2 image-only) |
| whole-blood-clotted, BD Vacutainer SST (SKU 367986), gel | Before spinninguncentrifuged | 30 °C (±1 °C) | 0 hchanged at the first time point, changed by 24 h | mean bias vs paired baseline within CLIA acceptance limits AND >2/3 of individual results within limits | Iacovetti et al. 2026, Pract Lab MedResults, 'Analyte stability after storage at 30 °C', paragraph 2 (Table 1 image-only) |
| whole-blood-clotted, Sarstedt Monovette serum | Before spinninguncentrifuged | chilled in transit (insulated box, 2 frozen cool packs; temperature not recorded) | 24 hstable for the longest time tested | % difference within desirable quality specification for bias (Fraser, from intra- and interindividual variation) | Jones et al. 2021, J NutrTable 2, row 'Creatinine', column Serum |
| whole-blood-edta, Sarstedt Monovette EDTA | Before spinninguncentrifuged | chilled in transit (insulated box, 2 frozen cool packs; temperature not recorded) | 24 hstable for the longest time tested | % difference within desirable quality specification for bias (Fraser, from intra- and interindividual variation) | Jones et al. 2021, J NutrTable 2, row 'Creatinine', column EDTA |
| whole-blood-heparin, BD Vacutainer lithium heparin, no gel (ref 368884) | Before spinninguncentrifuged | RT | 10 hstable for the longest time tested | |mean % deviation 10 h vs baseline| > laboratory bias goal (Table II; mostly biological-variation based, some empirically modified) | Henriksen et al. 2014, Scand J Clin Lab InvestTable II, row 'CREA' |
| whole-blood-heparin, BD Vacutainer lithium heparin, no gel (ref 368884) | Before spinninguncentrifuged | RT | 10 hstable for the longest time tested | CV of paired baseline/10 h results > goal CV (Table II) | Henriksen et al. 2014, Scand J Clin Lab InvestTable II, row 'CREA' |
| whole-blood-heparin, Sarstedt Monovette lithium heparin, trace-element free | Before spinninguncentrifuged | chilled in transit (insulated box, 2 frozen cool packs; temperature not recorded) | 24 hstable for the longest time tested | % difference within desirable quality specification for bias (Fraser, from intra- and interindividual variation) | Jones et al. 2021, J NutrTable 2, row 'Creatinine', column LH |
Cystatin C2 findings
| Specimen, tube | Condition | Temperature | Stable for | Criterion | Study |
|---|---|---|---|---|---|
| dbs, Whatman 903 card, mailed in foil pouch with desiccant | Storagedried | not stated; ambient postal shipping in the US (Apr-Dec 2016) | 8 daysstable for the longest time tested | regression coefficient for days before freezing vs 0-2 days significant (P < .05), Model 1 (DBS value, unadjusted) | Crimmins et al. 2020, Am J Hum BiolTable 3, rows 'Days before freezing', column 'Cystatin C' |
| dbs, Whatman 903 card, mailed in foil pouch with desiccant | Storagedried | not stated; ambient postal shipping in the US (Apr-Dec 2016) | 7 daysthen changed, changed by 8 days | regression coefficient for days before freezing vs 0-2 days significant (P < .05), Model 2 (DBS value adjusted for venous value) | Crimmins et al. 2020, Am J Hum BiolTable 3, rows 'Days before freezing', column 'Cystatin C' |
D-Dimer7 findings
| Specimen, tube | Condition | Temperature | Stable for | Criterion | Study |
|---|---|---|---|---|---|
| plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquots | Freeze–thawseparated | −20 °C | 1 cyclelimit stated by the authors | max allowable mean deviation from t0: 0.2–0.5 μg FEU/mL: ±0.05 μg FEU/mL c ; >0.5 μg FEU/mL: ±10% | Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability' |
| plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquots | Freeze–thawseparated | −80 °C | 3 cycleslimit stated by the authors | max allowable mean deviation from t0: 0.2–0.5 μg FEU/mL: ±0.05 μg FEU/mL c ; >0.5 μg FEU/mL: ±10% | Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability' |
| plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquots | Storageseparated | −20 °C | 28 dayslimit stated by the authors | max allowable mean deviation from t0: 0.2–0.5 μg FEU/mL: ±0.05 μg FEU/mL c ; >0.5 μg FEU/mL: ±10% | Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability' |
| plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquots | Storageseparated | −20 °C | 11 weekslimit stated by the authors | max allowable mean deviation from t0: 0.2–0.5 μg FEU/mL: ±0.05 μg FEU/mL c ; >0.5 μg FEU/mL: ±10% | Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability' |
| plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquots | Storageseparated | −80 °C | 23 monthslimit stated by the authors | max allowable mean deviation from t0: 0.2–0.5 μg FEU/mL: ±0.05 μg FEU/mL c ; >0.5 μg FEU/mL: ±10% | Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability' |
| plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquots | Storageseparated | 2-8 °C | 4 dayslimit stated by the authors | max allowable mean deviation from t0: 0.2–0.5 μg FEU/mL: ±0.05 μg FEU/mL c ; >0.5 μg FEU/mL: ±10% | Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability' |
| plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquots | Storageseparated | RT | 8 hlimit stated by the authors | max allowable mean deviation from t0: 0.2–0.5 μg FEU/mL: ±0.05 μg FEU/mL c ; >0.5 μg FEU/mL: ±10% | Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability' |
Direct (Conjugated) Bilirubin5 findings
| Specimen, tube | Condition | Temperature | Stable for | Criterion | Study |
|---|---|---|---|---|---|
| plasma-li-heparin, Sarstedt lithium heparin, no gel separator (primary tube) | Storagecentrifuged-on-cells | 2-8 °C | 48 hstable for the longest time tested | mean % difference vs time 0 exceeds total limit of change, TLC = ±√((1.65·CVa)² + (0.5·CVb)²) (EFLM BV), here 20.8%; stability limit from instability-equation regression | Illana et al. 2024, Biochem Med (Zagreb)Table 1, row 'DBIL' |
| plasma-li-heparin, Sarstedt lithium heparin, no gel separator; plasma aliquot | Storageseparated | 2-8 °C | 96 hstable for the longest time tested | mean % difference vs time 0 exceeds total limit of change, TLC = ±√((1.65·CVa)² + (0.5·CVb)²) (EFLM BV), here 20.8%; stability limit from instability-equation regression | Illana et al. 2024, Biochem Med (Zagreb)Table 2, row 'DBIL' |
| serum, BD Vacutainer 8.5 mL plastic serum tube; pooled serum aliquoted into 1.5 mL Eppendorf tubes | Freeze–thawseparated | −20 °C | 10 cyclesstable for the longest time tested | unstable only if both P < 0.005 (Bonferroni) and % change from T0 > desirable bias (Westgard/Ricos), here 14.2% | Cuhadar et al. 2013, Biochem Med (Zagreb)Table 2, row 'Bilirubin, direct (μmol/L)', columns T1d-T10d |
| serum, BD Vacutainer 8.5 mL plastic serum tube; pooled serum aliquoted into 1.5 mL Eppendorf tubes | Storageseparated | −20 °C | 3 monthsstable for the longest time tested | unstable only if both P < 0.013 (Bonferroni) and % change from T0 > desirable bias (Westgard/Ricos), here 14.2% | Cuhadar et al. 2013, Biochem Med (Zagreb)Table 2, row 'Bilirubin, direct (μmol/L)', columns T1m-T3m |
| serum, BD Vacutainer serum (plain), 7.5 mL | Storageseparated | −20 °C | 30 daysstable for the longest time tested | clinically significant change: mean outside significant change limit (SCL = T1d mean ± 3.0 usual SD from 2 months QC) | Kachhawa et al. 2017, J Lab PhysiciansResults paragraph 1 (Table 2 image-only) |
Eosinophils (Absolute)21 findings
| Specimen, tube | Condition | Temperature | Stable for | Criterion | Study |
|---|---|---|---|---|---|
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 48 hstable for the longest time tested | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 4, row 'EO', XN-9000, 4 °C |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 48 hstable for the longest time tested | Bias% vs optimal bias (OP-Bias%, from biological variation; Table 1) | Buoro et al. 2016, Rev Bras Hematol HemoterTable 4, row 'EO', XN-9000, 4 °C |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 48 hstable for the longest time tested | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 4, row 'EO', BC-6800, 4 °C |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 48 hstable for the longest time tested | Bias% vs optimal bias (OP-Bias%, from biological variation; Table 1) | Buoro et al. 2016, Rev Bras Hematol HemoterTable 4, row 'EO', BC-6800, 4 °C |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 8 hstable for the longest time tested | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 5, row 'EO', XN-9000, 4 °C |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 4 hthen changed, changed by 8 h | Bias% vs optimal bias (OP-Bias%, from biological variation; Table 1) | Buoro et al. 2016, Rev Bras Hematol HemoterTable 5, row 'EO', XN-9000, 4 °C |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 8 hstable for the longest time tested | Bias% vs critical difference (CD%, reference change value; Table 1) | Buoro et al. 2016, Rev Bras Hematol HemoterTable 5, row 'EO', XN-9000, 4 °C |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 8 hstable for the longest time tested | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 5, row 'EO', BC-6800, 4 °C |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 8 hstable for the longest time tested | Bias% vs optimal bias (OP-Bias%, from biological variation; Table 1) | Buoro et al. 2016, Rev Bras Hematol HemoterTable 5, row 'EO', BC-6800, 4 °C |
| whole-blood-edta, K2-EDTA (BD Vacutainer) | Storageuncentrifuged | 2-8 °C | 48 hstable for the longest time tested | Repeated-measures one-way ANOVA with Tukey post hoc vs 0 h, p < 0.05 (* in table) | Unalli et al. 2021, J Med BiochemTable 1, column '4 °C' (no * at any time point) |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | RT | 48 hstable for the longest time tested | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 4, row 'EO', XN-9000, RT |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | RT | 48 hstable for the longest time tested | Bias% vs optimal bias (OP-Bias%, from biological variation; Table 1) | Buoro et al. 2016, Rev Bras Hematol HemoterTable 4, row 'EO', XN-9000, RT |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | RT | 48 hstable for the longest time tested | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 4, row 'EO', BC-6800, RT |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | RT | 48 hstable for the longest time tested | Bias% vs optimal bias (OP-Bias%, from biological variation; Table 1) | Buoro et al. 2016, Rev Bras Hematol HemoterTable 4, row 'EO', BC-6800, RT |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | RT | 8 hstable for the longest time tested | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 5, row 'EO', XN-9000, RT |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | RT | 8 hstable for the longest time tested | Bias% vs optimal bias (OP-Bias%, from biological variation; Table 1) | Buoro et al. 2016, Rev Bras Hematol HemoterTable 5, row 'EO', XN-9000, RT |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | RT | 8 hstable for the longest time tested | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 5, row 'EO', BC-6800, RT |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | RT | 4 hthen changed, changed by 8 h | Bias% vs optimal bias (OP-Bias%, from biological variation; Table 1) | Buoro et al. 2016, Rev Bras Hematol HemoterTable 5, row 'EO', BC-6800, RT |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | RT | 8 hstable for the longest time tested | Bias% vs critical difference (CD%, reference change value; Table 1) | Buoro et al. 2016, Rev Bras Hematol HemoterTable 5, row 'EO', BC-6800, RT |
| whole-blood-edta, K2-EDTA (BD Vacutainer) | Storageuncentrifuged | RT | 4 hthen changed, changed by 8 h | Repeated-measures one-way ANOVA with Tukey post hoc vs 0 h, p < 0.05 (* in table) | Unalli et al. 2021, J Med BiochemTable 1, block 'EOS', row '8 h', column '23 °C' |
| whole-blood-edta, K2-EDTA (BD Vacutainer) | Storageuncentrifuged | 10 °C | 48 hstable for the longest time tested | Repeated-measures one-way ANOVA with Tukey post hoc vs 0 h, p < 0.05 (* in table) | Unalli et al. 2021, J Med BiochemTable 1, column '10 °C' (no * at any time point) |
Erythrocyte glutathione reductase activation coefficient (EGRAC)3 findings
| Specimen, tube | Condition | Temperature | Stable for | Criterion | Study |
|---|---|---|---|---|---|
| whole-blood-edta, Sarstedt Monovette EDTA | Before spinninguncentrifuged | chilled in transit (insulated box, 2 frozen cool packs; temperature not recorded) | 24 hstable for the longest time tested | % difference within desirable quality specification for bias (Fraser, from intra- and interindividual variation) | Jones et al. 2021, J NutrTable 2, row 'EGRAC (riboflavin)', column EDTA |
| whole-blood-heparin, BD Vacutainer lithium heparin | Before spinninguncentrifuged | cold packs in insulated box; 9 (2) °C on arrival | 24 hstable for the longest time tested | % difference within desirable quality specification (from intra- and interindividual variation) | Jones et al. 2024, J NutrTable 2, row 'EGRAC (B2) [ratio]'; Table 1 |
| whole-blood-heparin, Sarstedt Monovette lithium heparin, trace-element free | Before spinninguncentrifuged | chilled in transit (insulated box, 2 frozen cool packs; temperature not recorded) | 24 hstable for the longest time tested | % difference within desirable quality specification for bias (Fraser, from intra- and interindividual variation) | Jones et al. 2021, J NutrTable 2, row 'EGRAC (riboflavin)', column LH |
Erythrocyte Sedimentation Rate5 findings
| Specimen, tube | Condition | Temperature | Stable for | Criterion | Study |
|---|---|---|---|---|---|
| whole-blood-edta, BD Vacutainer K2-EDTA 3.0 mL | Storageuncentrifuged | 2-8 °C | 36 hthen changed, changed by 48 h | statistically significant difference vs T0 (p < 0.05) | Lorubbio et al. 2024, Diagnostics (Basel)Results, paragraph 3 |
| whole-blood-edta, BD Vacutainer K2 EDTA | Storageuncentrifuged | 2-8 °C | 48 hstable for the longest time tested | Passing-Bablok vs baseline: 95% CI of slope includes 1.00, 95% CI of intercept includes 0.00, Spearman r >= 0.90 | Koshy et al. 2026, Diagnostics (Basel)Table 3 |
| whole-blood-edta, BD Vacutainer K2-EDTA 3.0 mL | Storageuncentrifuged | RT | 0 hchanged at the first time point, changed by 4 h | statistically significant difference vs T0 (p < 0.05) | Lorubbio et al. 2024, Diagnostics (Basel)Results, paragraph 2 |
| whole-blood-edta, BD Vacutainer K2 EDTA | Storageuncentrifuged | RT | 10 hthen changed, changed by 12 h | Passing-Bablok vs baseline: 95% CI of slope includes 1.00, 95% CI of intercept includes 0.00, Spearman r >= 0.90 | Koshy et al. 2026, Diagnostics (Basel)Table 3 |
| whole-blood-edta, BD Vacutainer K2 EDTA | Storageuncentrifuged | RT | 28 hthen changed, changed by 36 h | Passing-Bablok vs baseline: 95% CI of slope includes 1.00, 95% CI of intercept includes 0.00, Spearman r >= 0.90 | Koshy et al. 2026, Diagnostics (Basel)Table 3 |
Erythrocyte transketolase activation coefficient (ETKAC)3 findings
| Specimen, tube | Condition | Temperature | Stable for | Criterion | Study |
|---|---|---|---|---|---|
| whole-blood-edta, Sarstedt Monovette EDTA | Before spinninguncentrifuged | chilled in transit (insulated box, 2 frozen cool packs; temperature not recorded) | 24 hstable for the longest time tested | % difference within desirable quality specification for bias (Fraser, from intra- and interindividual variation) | Jones et al. 2021, J NutrTable 2, row 'ETKAC (thiamin)', column EDTA |
| whole-blood-heparin, BD Vacutainer lithium heparin | Before spinninguncentrifuged | cold packs in insulated box; 9 (2) °C on arrival | 24 hstable for the longest time tested | % difference within desirable quality specification (from intra- and interindividual variation) | Jones et al. 2024, J NutrTable 2, row 'ETKAC (B1) [ratio]'; Table 1 |
| whole-blood-heparin, Sarstedt Monovette lithium heparin, trace-element free | Before spinninguncentrifuged | chilled in transit (insulated box, 2 frozen cool packs; temperature not recorded) | 24 hstable for the longest time tested | % difference within desirable quality specification for bias (Fraser, from intra- and interindividual variation) | Jones et al. 2021, J NutrTable 2, row 'ETKAC (thiamin)', column LH |
Ferritin12 findings
| Specimen, tube | Condition | Temperature | Stable for | Criterion | Study |
|---|---|---|---|---|---|
| serum, VACUMED 16x100 mm, inert separator gel and clot activator | Freeze–thawseparated | −80 °C | 0 cycleschanged at the first time point, changed by 2 cycles | Student t test for paired samples vs fresh measurement (SF1), p < 0.05 implied | Sacri et al. 2017, PLoS OneResults, paragraph 3 |
| serum, BD Vacutainer SST | Freeze–thawseparated | −70 °C, 1 month; thawed at room temperature | 1 cyclestable for the longest time tested | mean % change (bias) from fresh baseline within total allowable error (TEa, desirable bias) | Abraham et al. 2021, J Lab PhysiciansTable 2 and Table 3, row 'Ferritin (ng/mL)' |
| serum, VACUMED 16x100 mm, inert separator gel and clot activator | Storageseparated | −80 °C | 0 yearschanged at the first time point, changed by 3 years | Student t test for paired samples vs fresh measurement (SF1), p < 0.05 implied | Sacri et al. 2017, PLoS OneResults, paragraph 2 |
| whole-blood-clotted, BD Vacutainer serum, no gel (ref 368815) | Before spinninguncentrifuged | RT | 10 hstable for the longest time tested | |mean % deviation 10 h vs baseline| > laboratory bias goal (Table II; mostly biological-variation based, some empirically modified) | Henriksen et al. 2014, Scand J Clin Lab InvestTable II, row 'FER' |
| whole-blood-clotted, BD Vacutainer serum, no gel (ref 368815) | Before spinninguncentrifuged | RT | 10 hstable for the longest time tested | CV of paired baseline/10 h results > goal CV (Table II) | Henriksen et al. 2014, Scand J Clin Lab InvestTable II, row 'FER' |
| whole-blood-clotted, BD Vacutainer SST | Before spinninguncentrifuged | RT | 24 hstable for the longest time tested | mean % deviation from baseline within total change limit TCL = sqrt((2.77 CVa)^2 + (0.5 CVb)^2), CVb from Ricos (Table 2) | Abraham et al. 2019, Biochem Med (Zagreb)Results, paragraph 2 (Figure 3f) |
| whole-blood-clotted, BD Vacutainer trace element-free serum tube | Before spinninguncentrifuged | cold packs in insulated box; 9 (2) °C on arrival | 0 hchanged at the first time point, changed by 24 h | % difference within desirable quality specification (from intra- and interindividual variation) | Jones et al. 2024, J NutrTable 2, row 'Ferritin, μg/L'; Table 1 |
| whole-blood-clotted, BD Vacutainer SST | Before spinninguncentrifuged | 22-30 °C (logger 24-26 °C) | 24 hstable for the longest time tested | mean % deviation from baseline within total change limit TCL = sqrt((2.77 CVa)^2 + (0.5 CVb)^2), CVb from Ricos (Table 2) | Abraham et al. 2019, Biochem Med (Zagreb)Results, paragraph 2 (Figure 3f) |
| whole-blood-clotted, BD Vacutainer SST | Before spinninguncentrifuged | > 30 °C (logger 38-40 °C) | 12 hthen changed, changed by 24 h | mean % deviation from baseline within total change limit TCL = sqrt((2.77 CVa)^2 + (0.5 CVb)^2), CVb from Ricos (Table 2) | Abraham et al. 2019, Biochem Med (Zagreb)Results, paragraph 2 (Figure 3f) |
| whole-blood-clotted, Sarstedt Monovette serum | Before spinninguncentrifuged | chilled in transit (insulated box, 2 frozen cool packs; temperature not recorded) | 24 hstable for the longest time tested | % difference within desirable quality specification for bias (Fraser, from intra- and interindividual variation) | Jones et al. 2021, J NutrTable 2, row 'Ferritin', column Serum |
| whole-blood-edta, Sarstedt Monovette EDTA | Before spinninguncentrifuged | chilled in transit (insulated box, 2 frozen cool packs; temperature not recorded) | 24 hstable for the longest time tested | % difference within desirable quality specification for bias (Fraser, from intra- and interindividual variation) | Jones et al. 2021, J NutrTable 2, row 'Ferritin', column EDTA |
| whole-blood-heparin, Sarstedt Monovette lithium heparin, trace-element free | Before spinninguncentrifuged | chilled in transit (insulated box, 2 frozen cool packs; temperature not recorded) | 24 hstable for the longest time tested | % difference within desirable quality specification for bias (Fraser, from intra- and interindividual variation) | Jones et al. 2021, J NutrTable 2, row 'Ferritin', column LH |
Fibrinogen7 findings
| Specimen, tube | Condition | Temperature | Stable for | Criterion | Study |
|---|---|---|---|---|---|
| plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquots | Freeze–thawseparated | −20 °C | 1 cyclelimit stated by the authors | max allowable mean deviation from t0: ±10% | Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability' |
| plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquots | Freeze–thawseparated | −80 °C | 1 cyclelimit stated by the authors | max allowable mean deviation from t0: ±10% | Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability' |
| plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquots | Storageseparated | −20 °C | 7 weekslimit stated by the authors | max allowable mean deviation from t0: ±10% | Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability' |
| plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquots | Storageseparated | −80 °C | 11 monthslimit stated by the authors | max allowable mean deviation from t0: ±10% | Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability' |
| plasma-citrate, BD 0.109 M sodium citrate (9:1); plasma aliquoted into nonactivating Eppendorf tubes | Storageseparated | 2-8 °C | 24 hstable for the longest time tested | mean % change from baseline >10%, or >25% of samples with individual change >10% (van Geest-Daalderop) | Feng et al. 2014, Sci RepTable 2, 'Acceptable Time (h)' |
| plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquots | Storageseparated | RT | 4 hlimit stated by the authors | max allowable mean deviation from t0: ±10% | Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability' |
| plasma-citrate, BD 0.109 M sodium citrate (9:1); plasma aliquoted into nonactivating Eppendorf tubes | Storageseparated | RT | 24 hstable for the longest time tested | mean % change from baseline >10%, or >25% of samples with individual change >10% (van Geest-Daalderop) | Feng et al. 2014, Sci RepTable 2, 'Acceptable Time (h)' |
Fibrinogen (PT-derived)5 findings
| Specimen, tube | Condition | Temperature | Stable for | Criterion | Study |
|---|---|---|---|---|---|
| plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquots | Freeze–thawseparated | −20 °C | 1 cyclelimit stated by the authors | max allowable mean deviation from t0: ±10% | Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability' |
| plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquots | Freeze–thawseparated | −80 °C | 1 cyclelimit stated by the authors | max allowable mean deviation from t0: ±10% | Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability' |
| plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquots | Storageseparated | −20 °C | 7 weekslimit stated by the authors | max allowable mean deviation from t0: ±10% | Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability' |
| plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquots | Storageseparated | −80 °C | 11 monthslimit stated by the authors | max allowable mean deviation from t0: ±10% | Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability' |
| plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquots | Storageseparated | RT | 12 hlimit stated by the authors | max allowable mean deviation from t0: ±10% | Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability' |
Folate (Red Blood Cell)1 finding
| Specimen, tube | Condition | Temperature | Stable for | Criterion | Study |
|---|---|---|---|---|---|
| other, BD Vacutainer EDTA; haemolysate in folate ascorbic acid | Storageuncentrifuged | > 30 °C (logger 38-40 °C) | 0 hchanged at the first time point, changed by 6 h | mean % deviation from baseline within total change limit TCL = sqrt((2.77 CVa)^2 + (0.5 CVb)^2), CVb from Ricos (Table 2) | Abraham et al. 2019, Biochem Med (Zagreb)Results, paragraph 1 (Figure 3d) |
Folate (serum/plasma)7 findings
| Specimen, tube | Condition | Temperature | Stable for | Criterion | Study |
|---|---|---|---|---|---|
| serum, serum (tube not stated) | Storageseparated | −70 °C for the first year, −80 °C thereafter | 0 yearschanged at the first time point, changed by 1 year | t-test (two-sample, equal variances) vs t=0, P < 0.05 | Jansen et al. 2018, J Nutr MetabResults, section 3.2 |
| whole-blood-clotted, BD Vacutainer serum, no gel (ref 368815) | Before spinninguncentrifuged | RT | 10 hstable for the longest time tested | |mean % deviation 10 h vs baseline| > laboratory bias goal (Table II; mostly biological-variation based, some empirically modified) | Henriksen et al. 2014, Scand J Clin Lab InvestTable II, row 'FOL' |
| whole-blood-clotted, BD Vacutainer serum, no gel (ref 368815) | Before spinninguncentrifuged | RT | 0 hchanged at the first time point, changed by 10 h | CV of paired baseline/10 h results > goal CV (Table II) | Henriksen et al. 2014, Scand J Clin Lab InvestTable II, row 'FOL' |
| whole-blood-clotted, BD Vacutainer trace element-free serum tube | Before spinninguncentrifuged | cold packs in insulated box; 9 (2) °C on arrival | 24 hstable for the longest time tested | % difference within desirable quality specification (from intra- and interindividual variation) | Jones et al. 2024, J NutrTable 2, row 'Total folate, nmol/L'; Table 1 |
| whole-blood-clotted, Sarstedt Monovette serum | Before spinninguncentrifuged | chilled in transit (insulated box, 2 frozen cool packs; temperature not recorded) | 24 hstable for the longest time tested | % difference within desirable quality specification for bias (Fraser, from intra- and interindividual variation) | Jones et al. 2021, J NutrTable 2, row 'Total folate', column Serum |
| whole-blood-edta, Sarstedt Monovette EDTA | Before spinninguncentrifuged | chilled in transit (insulated box, 2 frozen cool packs; temperature not recorded) | 24 hstable for the longest time tested | % difference within desirable quality specification for bias (Fraser, from intra- and interindividual variation) | Jones et al. 2021, J NutrTable 2, row 'Total folate', column EDTA |
| whole-blood-heparin, Sarstedt Monovette lithium heparin, trace-element free | Before spinninguncentrifuged | chilled in transit (insulated box, 2 frozen cool packs; temperature not recorded) | 24 hstable for the longest time tested | % difference within desirable quality specification for bias (Fraser, from intra- and interindividual variation) | Jones et al. 2021, J NutrTable 2, row 'Total folate', column LH |
Free Thyroxine6 findings
| Specimen, tube | Condition | Temperature | Stable for | Criterion | Study |
|---|---|---|---|---|---|
| plasma-li-heparin, Sarstedt lithium heparin, no gel separator (primary tube) | Storagecentrifuged-on-cells | 2-8 °C | 48 hstable for the longest time tested | mean % difference vs time 0 exceeds total limit of change, TLC = ±√((1.65·CVa)² + (0.5·CVb)²) (EFLM BV), here 18.9%; stability limit from instability-equation regression | Illana et al. 2024, Biochem Med (Zagreb)Table 1, row 'fT4' |
| plasma-li-heparin, Sarstedt lithium heparin, no gel separator; plasma aliquot | Storageseparated | 2-8 °C | 96 hstable for the longest time tested | mean % difference vs time 0 exceeds total limit of change, TLC = ±√((1.65·CVa)² + (0.5·CVb)²) (EFLM BV), here 18.9%; stability limit from instability-equation regression | Illana et al. 2024, Biochem Med (Zagreb)Table 2, row 'fT4' |
| plasma-li-heparin, BD Barricor 4.5 mL (lithium heparin, mechanical separator) | Storagecentrifuged-on-gel | 2-8 °C | 7 daysstable for the longest time tested | mean % difference vs time 0 exceeds desirable imprecision based on biological variation (EFLM/Westgard), here 3.85% | Mandić et al. 2020, Biochem Med (Zagreb)Tables 3 and 4, row 'fT4, pmol/L', BD Barricor columns |
| serum, BD Vacutainer RST 5 mL (thrombin clot activator, gel) | Storagecentrifuged-on-gel | 2-8 °C | 7 daysstable for the longest time tested | mean % difference vs time 0 exceeds desirable imprecision based on biological variation (EFLM/Westgard), here 3.85% | Mandić et al. 2020, Biochem Med (Zagreb)Tables 3 and 4, row 'fT4, pmol/L', BD RST columns |
| whole-blood-clotted, BD Vacutainer serum, no gel (ref 368815) | Before spinninguncentrifuged | RT | 10 hstable for the longest time tested | |mean % deviation 10 h vs baseline| > laboratory bias goal (Table II; mostly biological-variation based, some empirically modified) | Henriksen et al. 2014, Scand J Clin Lab InvestTable II, row 'F-T4' |
| whole-blood-clotted, BD Vacutainer serum, no gel (ref 368815) | Before spinninguncentrifuged | RT | 10 hstable for the longest time tested | CV of paired baseline/10 h results > goal CV (Table II) | Henriksen et al. 2014, Scand J Clin Lab InvestTable II, row 'F-T4' |
Free Triiodothyronine6 findings
| Specimen, tube | Condition | Temperature | Stable for | Criterion | Study |
|---|---|---|---|---|---|
| plasma-li-heparin, Sarstedt lithium heparin, no gel separator (primary tube) | Storagecentrifuged-on-cells | 2-8 °C | 48 hstable for the longest time tested | mean % difference vs time 0 exceeds total limit of change, TLC = ±√((1.65·CVa)² + (0.5·CVb)²) (EFLM BV), here 14.6%; stability limit from instability-equation regression | Illana et al. 2024, Biochem Med (Zagreb)Table 1, row 'fT3' |
| plasma-li-heparin, Sarstedt lithium heparin, no gel separator; plasma aliquot | Storageseparated | 2-8 °C | 96 hstable for the longest time tested | mean % difference vs time 0 exceeds total limit of change, TLC = ±√((1.65·CVa)² + (0.5·CVb)²) (EFLM BV), here 14.6%; stability limit from instability-equation regression | Illana et al. 2024, Biochem Med (Zagreb)Table 2, row 'fT3' |
| plasma-li-heparin, BD Barricor 4.5 mL (lithium heparin, mechanical separator) | Storagecentrifuged-on-gel | 2-8 °C | 7 daysstable for the longest time tested | mean % difference vs time 0 exceeds desirable imprecision based on biological variation (EFLM/Westgard), here 3.0% | Mandić et al. 2020, Biochem Med (Zagreb)Tables 3 and 4, row 'fT3, pmol/L', BD Barricor columns |
| serum, BD Vacutainer RST 5 mL (thrombin clot activator, gel) | Storagecentrifuged-on-gel | 2-8 °C | 0 hchanged at the first time point, changed by 24 h | mean % difference vs time 0 exceeds desirable imprecision based on biological variation (EFLM/Westgard), here 3.0% | Mandić et al. 2020, Biochem Med (Zagreb)Tables 3 and 4, row 'fT3, pmol/L', BD RST columns |
| whole-blood-clotted, BD Vacutainer serum, no gel (ref 368815) | Before spinninguncentrifuged | RT | 10 hstable for the longest time tested | |mean % deviation 10 h vs baseline| > laboratory bias goal (Table II; mostly biological-variation based, some empirically modified) | Henriksen et al. 2014, Scand J Clin Lab InvestTable II, row 'F-T3' |
| whole-blood-clotted, BD Vacutainer serum, no gel (ref 368815) | Before spinninguncentrifuged | RT | 10 hstable for the longest time tested | CV of paired baseline/10 h results > goal CV (Table II) | Henriksen et al. 2014, Scand J Clin Lab InvestTable II, row 'F-T3' |
Gamma-Glutamyl Transferase (GGT)12 findings
| Specimen, tube | Condition | Temperature | Stable for | Criterion | Study |
|---|---|---|---|---|---|
| plasma-li-heparin, Sarstedt lithium heparin, no gel separator (primary tube) | Storagecentrifuged-on-cells | 2-8 °C | 48 hstable for the longest time tested | mean % difference vs time 0 exceeds total limit of change, TLC = ±√((1.65·CVa)² + (0.5·CVb)²) (EFLM BV), here 9.7%; stability limit from instability-equation regression | Illana et al. 2024, Biochem Med (Zagreb)Table 1, row 'GGT' |
| plasma-li-heparin, Sarstedt lithium heparin, no gel separator; plasma aliquot | Storageseparated | 2-8 °C | 96 hstable for the longest time tested | mean % difference vs time 0 exceeds total limit of change, TLC = ±√((1.65·CVa)² + (0.5·CVb)²) (EFLM BV), here 9.7%; stability limit from instability-equation regression | Illana et al. 2024, Biochem Med (Zagreb)Table 2, row 'GGT' |
| plasma-li-heparin, BD Barricor 4.5 mL (lithium heparin, mechanical separator) | Storagecentrifuged-on-gel | 2-8 °C | 7 daysstable for the longest time tested | mean % difference vs time 0 exceeds desirable imprecision based on biological variation (EFLM/Westgard), here 4.4% | Mandić et al. 2020, Biochem Med (Zagreb)Tables 3 and 4, row 'GGT, U/L', BD Barricor columns |
| serum, BD Vacutainer 8.5 mL plastic serum tube; pooled serum aliquoted into 1.5 mL Eppendorf tubes | Freeze–thawseparated | −20 °C | 10 cyclesstable for the longest time tested | unstable only if both P < 0.005 (Bonferroni) and % change from T0 > desirable bias (Westgard/Ricos), here 10.8% | Cuhadar et al. 2013, Biochem Med (Zagreb)Table 2, row 'GGT (U/L)', columns T1d-T10d |
| serum, BD Vacutainer 8.5 mL plastic serum tube; pooled serum aliquoted into 1.5 mL Eppendorf tubes | Storageseparated | −20 °C | 3 monthsstable for the longest time tested | unstable only if both P < 0.013 (Bonferroni) and % change from T0 > desirable bias (Westgard/Ricos), here 10.8% | Cuhadar et al. 2013, Biochem Med (Zagreb)Table 2, row 'GGT (U/L)', columns T1m-T3m |
| serum, Vacuette Tube Serum Gel Separator 8 mL (Greiner, 455071) | Storagecentrifuged-on-gel | 2-8 °C | 1 weekstable for the longest time tested | % bias of median vs baseline exceeds desirable bias (Westgard/Ricos BV), here 10.8% | Cuhadar et al. 2012, Biochem Med (Zagreb)Table 3, row 'γ-Glutamyltransferase', gel 4°C |
| serum, Vacuette Tube Serum No Gel 9 mL (Greiner, 455092), plain | Storagecentrifuged-on-cells | 2-8 °C | 1 weekstable for the longest time tested | % bias of median vs baseline exceeds desirable bias (Westgard/Ricos BV), here 10.8% | Cuhadar et al. 2012, Biochem Med (Zagreb)Table 3, row 'γ-Glutamyltransferase', w/o gel 4°C |
| serum, BD Vacutainer RST 5 mL (thrombin clot activator, gel) | Storagecentrifuged-on-gel | 2-8 °C | 7 daysstable for the longest time tested | mean % difference vs time 0 exceeds desirable imprecision based on biological variation (EFLM/Westgard), here 4.4% | Mandić et al. 2020, Biochem Med (Zagreb)Tables 3 and 4, row 'GGT, U/L', BD RST columns |
| serum, Vacuette Tube Serum Gel Separator 8 mL (Greiner, 455071) | Storagecentrifuged-on-gel | RT | 1 weekstable for the longest time tested | % bias of median vs baseline exceeds desirable bias (Westgard/Ricos BV), here 10.8% | Cuhadar et al. 2012, Biochem Med (Zagreb)Table 3, row 'γ-Glutamyltransferase', gel 24°C |
| serum, Vacuette Tube Serum No Gel 9 mL (Greiner, 455092), plain | Storagecentrifuged-on-cells | RT | 1 weekstable for the longest time tested | % bias of median vs baseline exceeds desirable bias (Westgard/Ricos BV), here 10.8% | Cuhadar et al. 2012, Biochem Med (Zagreb)Table 3, row 'γ-Glutamyltransferase', w/o gel 24°C |
| whole-blood-heparin, BD Vacutainer lithium heparin, no gel (ref 368884) | Before spinninguncentrifuged | RT | 10 hstable for the longest time tested | |mean % deviation 10 h vs baseline| > laboratory bias goal (Table II; mostly biological-variation based, some empirically modified) | Henriksen et al. 2014, Scand J Clin Lab InvestTable II, row 'GGT' |
| whole-blood-heparin, BD Vacutainer lithium heparin, no gel (ref 368884) | Before spinninguncentrifuged | RT | 10 hstable for the longest time tested | CV of paired baseline/10 h results > goal CV (Table II) | Henriksen et al. 2014, Scand J Clin Lab InvestTable II, row 'GGT' |
Gamma-tocopherol4 findings
| Specimen, tube | Condition | Temperature | Stable for | Criterion | Study |
|---|---|---|---|---|---|
| whole-blood-clotted, Sarstedt Monovette serum | Before spinninguncentrifuged | chilled in transit (insulated box, 2 frozen cool packs; temperature not recorded) | 24 hstable for the longest time tested | % difference within desirable quality specification for bias (Fraser, from intra- and interindividual variation) | Jones et al. 2021, J NutrTable 2, row 'γ-tocopherol', column Serum |
| whole-blood-edta, Sarstedt Monovette EDTA | Before spinninguncentrifuged | chilled in transit (insulated box, 2 frozen cool packs; temperature not recorded) | 24 hstable for the longest time tested | % difference within desirable quality specification for bias (Fraser, from intra- and interindividual variation) | Jones et al. 2021, J NutrTable 2, row 'γ-tocopherol', column EDTA |
| whole-blood-heparin, BD Vacutainer lithium heparin | Before spinninguncentrifuged | cold packs in insulated box; 9 (2) °C on arrival | 24 hstable for the longest time tested | % difference within desirable quality specification (from intra- and interindividual variation) | Jones et al. 2024, J NutrTable 2, row 'γ-Tocopherol, μmol/L'; Table 1 |
| whole-blood-heparin, Sarstedt Monovette lithium heparin, trace-element free | Before spinninguncentrifuged | chilled in transit (insulated box, 2 frozen cool packs; temperature not recorded) | 24 hstable for the longest time tested | % difference within desirable quality specification for bias (Fraser, from intra- and interindividual variation) | Jones et al. 2021, J NutrTable 2, row 'γ-tocopherol', column LH |
Glucose30 findings
| Specimen, tube | Condition | Temperature | Stable for | Criterion | Study |
|---|---|---|---|---|---|
| plasma-edta, BD Vacutainer EDTA, 4 mL | Freeze–thawseparated | −20 °C | 4 cyclesstable for the longest time tested | significant change vs day 1 (repeated-measures ANOVA, Tukey HSD post hoc, p<0.05) | Gojković et al. 2024, J Med BiochemResults paragraph 3 (EDTA -20°C, Figure 3) |
| plasma-edta, BD Vacutainer EDTA, 4 mL | Storageseparated | 2-8 °C | 3 daysthen changed, changed by 4 days | significant change vs day 1 (repeated-measures ANOVA, Tukey HSD post hoc, p<0.05) | Gojković et al. 2024, J Med BiochemResults paragraph 1 (EDTA+4°C, Figure 1) |
| plasma-fluoride, BD Vacutainer sodium fluoride/potassium oxalate, 4 mL | Freeze–thawseparated | −20 °C | 3 cyclesthen changed, changed by 4 cycles | significant change vs day 1 (repeated-measures ANOVA, Tukey HSD post hoc, p<0.05) | Gojković et al. 2024, J Med BiochemResults paragraph 4 (NaF -20°C, Figure 4) |
| plasma-fluoride, BD Vacutainer fluoride | Freeze–thawseparated | −70 °C, 1 month; thawed at room temperature | 1 cyclestable for the longest time tested | mean % change (bias) from fresh baseline within total allowable error (TEa, desirable bias) | Abraham et al. 2021, J Lab PhysiciansTable 2 and Table 3, row 'Glucose (mg/dL)' |
| plasma-fluoride, sodium fluoride/potassium oxalate (NaF/KOx) | Storagenot-stated | 2-8 °C | 72 hstable for the longest time tested | no significant change in mixed effects model vs first analysis (overall p = 0.25) | Dibbasey et al. 2024, Pract Lab MedTable 2; Results |
| plasma-fluoride, BD Vacutainer sodium fluoride/potassium oxalate, 4 mL | Storageseparated | 2-8 °C | 4 daysstable for the longest time tested | significant change vs day 1 (repeated-measures ANOVA, Tukey HSD post hoc, p<0.05) | Gojković et al. 2024, J Med BiochemResults paragraph 2 (NaF+4°C, Figure 2) |
| plasma-li-heparin, Sarstedt lithium heparin, no gel separator (primary tube) | Storagecentrifuged-on-cells | 2-8 °C | 2.7 hthen changed, changed by 4 h | mean % difference vs time 0 exceeds total limit of change, TLC = ±√((1.65·CVa)² + (0.5·CVb)²) (EFLM BV), here 6.3%; stability limit from instability-equation regression | Illana et al. 2024, Biochem Med (Zagreb)Table 1, row 'Glc' |
| plasma-li-heparin, Sarstedt lithium heparin, no gel separator; plasma aliquot | Storageseparated | 2-8 °C | 96 hstable for the longest time tested | mean % difference vs time 0 exceeds total limit of change, TLC = ±√((1.65·CVa)² + (0.5·CVb)²) (EFLM BV), here 6.3%; stability limit from instability-equation regression | Illana et al. 2024, Biochem Med (Zagreb)Table 2, row 'Glc' |
| plasma-li-heparin, BD Vacutainer LH PST II gel separator, 3.0 mL (REF 367374) | Storagecentrifuged-on-gel | 2-8 °C | 24 hstable for the longest time tested | mean % difference smaller than TEa (Ricos: 0.25 x (CVi^2 + CVg^2)^0.5 + 1.65 x 0.5 x CVi) | Nonkes et al. 2019, Biochem Med (Zagreb)Table 1, row 'Glucose, mmol/L' |
| plasma-li-heparin, BD Vacutainer Barricor LH 3 mL, mechanical separator | Storagecentrifuged-on-gel | 2-8 °C | 18 hlimit stated by the authors | bias vs aliquoted plasma reference tube > maximal allowable bias 2.5% (Ricos et al.); time to MAB from linear regression of bias on time to analysis | Gawria et al. 2020, J Clin Lab AnalTable 2, row 'Glucose', Barricor BD |
| plasma-li-heparin, BD Vacutainer PST LH 3 mL, gel | Storagecentrifuged-on-gel | 2-8 °C | 28 hlimit stated by the authors | bias vs aliquoted plasma reference tube > maximal allowable bias 2.5% (Ricos et al.); time to MAB from linear regression of bias on time to analysis | Gawria et al. 2020, J Clin Lab AnalTable 2, row 'Glucose', Gel BD |
| plasma-li-heparin, Greiner Vacuette LH 3 mL Sep, gel | Storagecentrifuged-on-gel | 2-8 °C | 26 hlimit stated by the authors | bias vs aliquoted plasma reference tube > maximal allowable bias 2.5% (Ricos et al.); time to MAB from linear regression of bias on time to analysis | Gawria et al. 2020, J Clin Lab AnalTable 2, row 'Glucose', Gel Greiner |
| plasma-li-heparin, BD Barricor 4.5 mL (lithium heparin, mechanical separator) | Storagecentrifuged-on-gel | 2-8 °C | 0 hchanged at the first time point, changed by 24 h | mean % difference vs time 0 exceeds desirable imprecision based on biological variation (EFLM/Westgard), here 2.55% | Mandić et al. 2020, Biochem Med (Zagreb)Tables 3 and 4, row 'Glucose, mmol/L', BD Barricor columns |
| serum, BD Vacutainer 8.5 mL plastic serum tube; pooled serum aliquoted into 1.5 mL Eppendorf tubes | Freeze–thawseparated | −20 °C | 10 cyclesstable for the longest time tested | unstable only if both P < 0.005 (Bonferroni) and % change from T0 > desirable bias (Westgard/Ricos), here 2.2% | Cuhadar et al. 2013, Biochem Med (Zagreb)Table 2, row 'Glucose (mmol/L)', columns T1d-T10d |
| serum, BD Vacutainer 8.5 mL plastic serum tube; pooled serum aliquoted into 1.5 mL Eppendorf tubes | Storageseparated | −20 °C | 3 monthsstable for the longest time tested | unstable only if both P < 0.013 (Bonferroni) and % change from T0 > desirable bias (Westgard/Ricos), here 2.2% | Cuhadar et al. 2013, Biochem Med (Zagreb)Table 2, row 'Glucose (mmol/L)', columns T1m-T3m |
| serum, serum separator tube (SST) | Storagenot-stated | 2-8 °C | 72 hstable for the longest time tested | no significant change in mixed effects model vs first analysis (overall p = 0.25) | Dibbasey et al. 2024, Pract Lab MedTable 2; Results |
| serum, Vacuette Tube Serum Gel Separator 8 mL (Greiner, 455071) | Storagecentrifuged-on-gel | 2-8 °C | 36 hthen changed, changed by 48 h | % bias of median vs baseline exceeds desirable bias (Westgard/Ricos BV), here 2.2% | Cuhadar et al. 2012, Biochem Med (Zagreb)Table 3, row 'Glucose', gel 4°C |
| serum, Vacuette Tube Serum No Gel 9 mL (Greiner, 455092), plain | Storagecentrifuged-on-cells | 2-8 °C | 0 hchanged at the first time point, changed by 6 h | % bias of median vs baseline exceeds desirable bias (Westgard/Ricos BV), here 2.2% | Cuhadar et al. 2012, Biochem Med (Zagreb)Table 3, row 'Glucose', w/o gel 4°C |
| serum, BD Vacutainer RST 4.0 mL (thrombin clot activator, gel) | Storagecentrifuged-on-gel | 2-8 °C | 24 hstable for the longest time tested | clinically significant if mean bias vs t0 exceeds desirable bias (biological variation), here 1.5% | Kocijancic et al. 2014, Biochem Med (Zagreb)Table 3, row 'Glu' |
| serum, BD Vacutainer SST II Advance 5.0 mL (clot activator, gel) | Storagecentrifuged-on-gel | 2-8 °C | 24 hstable for the longest time tested | clinically significant if mean bias vs t0 exceeds desirable bias (biological variation), here 1.5% | Kocijancic et al. 2014, Biochem Med (Zagreb)Table 4, row 'Glu' |
| serum, BD Vacutainer RST 5 mL (thrombin clot activator, gel) | Storagecentrifuged-on-gel | 2-8 °C | 7 daysstable for the longest time tested | mean % difference vs time 0 exceeds desirable imprecision based on biological variation (EFLM/Westgard), here 2.55% | Mandić et al. 2020, Biochem Med (Zagreb)Tables 3 and 4, row 'Glucose, mmol/L', BD RST columns |
| serum, Vacuette Tube Serum Gel Separator 8 mL (Greiner, 455071) | Storagecentrifuged-on-gel | RT | 18 hthen changed, changed by 24 h | % bias of median vs baseline exceeds desirable bias (Westgard/Ricos BV), here 2.2% | Cuhadar et al. 2012, Biochem Med (Zagreb)Table 3, row 'Glucose', gel 24°C |
| serum, Vacuette Tube Serum No Gel 9 mL (Greiner, 455092), plain | Storagecentrifuged-on-cells | RT | 0 hchanged at the first time point, changed by 6 h | % bias of median vs baseline exceeds desirable bias (Westgard/Ricos BV), here 2.2% | Cuhadar et al. 2012, Biochem Med (Zagreb)Table 3, row 'Glucose', w/o gel 24°C |
| serum, BD Vacutainer SST (SKU 367986), gel | Storagecentrifuged-on-gel | 30 °C (±1 °C) | 72 hstable for the longest time tested | mean bias vs paired baseline within CLIA acceptance limits AND >2/3 of individual results within limits | Iacovetti et al. 2026, Pract Lab MedResults, 'Analyte stability after storage at 30 °C', paragraph 1 (Table 1 image-only) |
| serum, BD SST Microtainer (SKU 395967), gel; filled from venous blood | Storagecentrifuged-on-gel | 0 °C | 72 hstable for the longest time tested | mean bias vs paired baseline within CLIA acceptance limits AND >2/3 of individual results within limits | Iacovetti et al. 2026, Pract Lab MedResults, 'Analyte stability after separation and storage in microvolume tubes at 0 °C' (Table 2 image-only) |
| whole-blood-clotted, BD Vacutainer SST (SKU 367986), gel | Before spinninguncentrifuged | 30 °C (±1 °C) | 0 hchanged at the first time point, changed by 24 h | mean bias vs paired baseline within CLIA acceptance limits AND >2/3 of individual results within limits | Iacovetti et al. 2026, Pract Lab MedResults, 'Analyte stability after storage at 30 °C', paragraph 2 (Table 1 image-only) |
| whole-blood-fluoride, Greiner Vacuette FC-Mix NaF-EDTA-citrate (granulated), blood aliquoted into micro-tubes | Before spinninguncentrifuged | RT | 48 hthen changed, changed by 72 h | mean bias vs reference glucose within desirable bias (< ±1.95%, biological variation) | Bonetti et al. 2017, Biochem Med (Zagreb)Table 1 |
| whole-blood-fluoride, BD Vacutainer fluoride | Before spinninguncentrifuged | RT | 24 hstable for the longest time tested | mean % deviation from baseline within total change limit TCL = sqrt((2.77 CVa)^2 + (0.5 CVb)^2), CVb from Ricos (Table 2) | Abraham et al. 2019, Biochem Med (Zagreb)Discussion, paragraph 1 (Results paragraphs 1-3; Figure 3) |
| whole-blood-fluoride, BD Vacutainer fluoride | Before spinninguncentrifuged | 22-30 °C (logger 24-26 °C) | 24 hstable for the longest time tested | mean % deviation from baseline within total change limit TCL = sqrt((2.77 CVa)^2 + (0.5 CVb)^2), CVb from Ricos (Table 2) | Abraham et al. 2019, Biochem Med (Zagreb)Discussion, paragraph 1 (Results paragraphs 1-3; Figure 3) |
| whole-blood-fluoride, BD Vacutainer fluoride | Before spinninguncentrifuged | > 30 °C (logger 38-40 °C) | 24 hstable for the longest time tested | mean % deviation from baseline within total change limit TCL = sqrt((2.77 CVa)^2 + (0.5 CVb)^2), CVb from Ricos (Table 2) | Abraham et al. 2019, Biochem Med (Zagreb)Discussion, paragraph 1 (Results paragraphs 1-3; Figure 3) |
Haptoglobin2 findings
| Specimen, tube | Condition | Temperature | Stable for | Criterion | Study |
|---|---|---|---|---|---|
| whole-blood-heparin, BD Vacutainer lithium heparin, no gel (ref 368884) | Before spinninguncentrifuged | RT | 10 hstable for the longest time tested | |mean % deviation 10 h vs baseline| > laboratory bias goal (Table II; mostly biological-variation based, some empirically modified) | Henriksen et al. 2014, Scand J Clin Lab InvestTable II, row 'HAPT' |
| whole-blood-heparin, BD Vacutainer lithium heparin, no gel (ref 368884) | Before spinninguncentrifuged | RT | 10 hstable for the longest time tested | CV of paired baseline/10 h results > goal CV (Table II) | Henriksen et al. 2014, Scand J Clin Lab InvestTable II, row 'HAPT' |
HDL Cholesterol22 findings
| Specimen, tube | Condition | Temperature | Stable for | Criterion | Study |
|---|---|---|---|---|---|
| dbs, Whatman 903 card, mailed in foil pouch with desiccant | Storagedried | not stated; ambient postal shipping in the US (Apr-Dec 2016) | 3 daysthen changed, changed by 4 days | regression coefficient for days before freezing vs 0-2 days significant (P < .05), Model 1 (DBS value, unadjusted) | Crimmins et al. 2020, Am J Hum BiolTable 3, rows 'Days before freezing', column 'HDL cholesterol' |
| dbs, Whatman 903 card, mailed in foil pouch with desiccant | Storagedried | not stated; ambient postal shipping in the US (Apr-Dec 2016) | 7 daysthen changed, changed by 8 days | regression coefficient for days before freezing vs 0-2 days significant (P < .05), Model 2 (DBS value adjusted for venous value) | Crimmins et al. 2020, Am J Hum BiolTable 3, rows 'Days before freezing', column 'HDL cholesterol' |
| serum, BD Vacutainer 8.5 mL plastic serum tube; pooled serum aliquoted into 1.5 mL Eppendorf tubes | Freeze–thawseparated | −20 °C | 10 cyclesstable for the longest time tested | unstable only if both P < 0.005 (Bonferroni) and % change from T0 > desirable bias (Westgard/Ricos), here 5.2% | Cuhadar et al. 2013, Biochem Med (Zagreb)Table 2, row 'HDL (mmol/L)', columns T1d-T10d |
| serum, BD Vacutainer SST | Freeze–thawseparated | −70 °C, 1 month; thawed at room temperature | 1 cyclestable for the longest time tested | mean % change (bias) from fresh baseline within total allowable error (TEa, desirable bias) | Abraham et al. 2021, J Lab PhysiciansTable 2 and Table 3, row 'HDL (mg/dL)' |
| serum, serum (tube not stated) | Storageseparated | −20 °C | 2 monthsthen changed, changed by 6 months | t-test (two-sample, equal variances) vs t=0, 95% CI | Beekhof et al. 2012, Lipids Health DisResults, paragraph 4 |
| serum, BD Vacutainer 8.5 mL plastic serum tube; pooled serum aliquoted into 1.5 mL Eppendorf tubes | Storageseparated | −20 °C | 3 monthsstable for the longest time tested | unstable only if both P < 0.013 (Bonferroni) and % change from T0 > desirable bias (Westgard/Ricos), here 5.2% | Cuhadar et al. 2013, Biochem Med (Zagreb)Table 2, row 'HDL (mmol/L)', columns T1m-T3m |
| serum, Vacuette Tube Serum Gel Separator 8 mL (Greiner, 455071) | Storagecentrifuged-on-gel | 2-8 °C | 24 hthen changed, changed by 30 h | % bias of median vs baseline exceeds desirable bias (Westgard/Ricos BV), here 5.2% | Cuhadar et al. 2012, Biochem Med (Zagreb)Table 3, row 'HDL-cholesterol', gel 4°C |
| serum, Vacuette Tube Serum No Gel 9 mL (Greiner, 455092), plain | Storagecentrifuged-on-cells | 2-8 °C | 36 hthen changed, changed by 48 h | % bias of median vs baseline exceeds desirable bias (Westgard/Ricos BV), here 5.2% | Cuhadar et al. 2012, Biochem Med (Zagreb)Table 3, row 'HDL-cholesterol', w/o gel 4°C |
| serum, Vacuette Tube Serum Gel Separator 8 mL (Greiner, 455071) | Storagecentrifuged-on-gel | RT | 36 hthen changed, changed by 48 h | % bias of median vs baseline exceeds desirable bias (Westgard/Ricos BV), here 5.2% | Cuhadar et al. 2012, Biochem Med (Zagreb)Table 3, row 'HDL-cholesterol', gel 24°C |
| serum, Vacuette Tube Serum No Gel 9 mL (Greiner, 455092), plain | Storagecentrifuged-on-cells | RT | 6 hthen changed, changed by 12 h | % bias of median vs baseline exceeds desirable bias (Westgard/Ricos BV), here 5.2% | Cuhadar et al. 2012, Biochem Med (Zagreb)Table 3, row 'HDL-cholesterol', w/o gel 24°C |
| serum, BD Vacutainer SST (SKU 367986), gel | Storagecentrifuged-on-gel | 30 °C (±1 °C) | 72 hstable for the longest time tested | mean bias vs paired baseline within CLIA acceptance limits AND >2/3 of individual results within limits | Iacovetti et al. 2026, Pract Lab MedResults, 'Analyte stability after storage at 30 °C', paragraph 1 (Table 1 image-only) |
| serum, BD SST Microtainer (SKU 395967), gel; filled from venous blood | Storagecentrifuged-on-gel | 0 °C | 72 hstable for the longest time tested | mean bias vs paired baseline within CLIA acceptance limits AND >2/3 of individual results within limits | Iacovetti et al. 2026, Pract Lab MedResults, 'Analyte stability after separation and storage in microvolume tubes at 0 °C' (Table 2 image-only) |
| serum, serum (tube not stated) | Storageseparated | −70°C | 12 monthsstable for the longest time tested | t-test (two-sample, equal variances) vs t=0, 95% CI | Beekhof et al. 2012, Lipids Health DisResults, paragraph 4 |
| serum, serum (tube not stated) | Storageseparated | −196°C | 12 monthsstable for the longest time tested | t-test (two-sample, equal variances) vs t=0, 95% CI | Beekhof et al. 2012, Lipids Health DisResults, paragraph 4 |
| whole-blood-clotted, BD Vacutainer SST | Before spinninguncentrifuged | RT | 24 hstable for the longest time tested | mean % deviation from baseline within total change limit TCL = sqrt((2.77 CVa)^2 + (0.5 CVb)^2), CVb from Ricos (Table 2) | Abraham et al. 2019, Biochem Med (Zagreb)Discussion, paragraph 4 |
| whole-blood-clotted, BD Vacutainer trace element-free serum tube | Before spinninguncentrifuged | cold packs in insulated box; 9 (2) °C on arrival | 24 hstable for the longest time tested | % difference within desirable quality specification (from intra- and interindividual variation) | Jones et al. 2024, J NutrTable 2, row 'HDL cholesterol, mmol/L'; Table 1 |
| whole-blood-clotted, BD Vacutainer SST (SKU 367986), gel | Before spinninguncentrifuged | 30 °C (±1 °C) | 72 hstable for the longest time tested | mean bias vs paired baseline within CLIA acceptance limits AND >2/3 of individual results within limits | Iacovetti et al. 2026, Pract Lab MedResults, 'Analyte stability after storage at 30 °C', paragraph 2 (Table 1 image-only) |
| whole-blood-clotted, Sarstedt Monovette serum | Before spinninguncentrifuged | chilled in transit (insulated box, 2 frozen cool packs; temperature not recorded) | 24 hstable for the longest time tested | % difference within desirable quality specification for bias (Fraser, from intra- and interindividual variation) | Jones et al. 2021, J NutrTable 2, row 'HDL cholesterol', column Serum |
| whole-blood-edta, Sarstedt Monovette EDTA | Before spinninguncentrifuged | chilled in transit (insulated box, 2 frozen cool packs; temperature not recorded) | 24 hstable for the longest time tested | % difference within desirable quality specification for bias (Fraser, from intra- and interindividual variation) | Jones et al. 2021, J NutrTable 2, row 'HDL cholesterol', column EDTA |
| whole-blood-heparin, BD Vacutainer lithium heparin, no gel (ref 368884) | Before spinninguncentrifuged | RT | 10 hstable for the longest time tested | |mean % deviation 10 h vs baseline| > laboratory bias goal (Table II; mostly biological-variation based, some empirically modified) | Henriksen et al. 2014, Scand J Clin Lab InvestTable II, row 'HDL-C' |
| whole-blood-heparin, BD Vacutainer lithium heparin, no gel (ref 368884) | Before spinninguncentrifuged | RT | 10 hstable for the longest time tested | CV of paired baseline/10 h results > goal CV (Table II) | Henriksen et al. 2014, Scand J Clin Lab InvestTable II, row 'HDL-C' |
| whole-blood-heparin, Sarstedt Monovette lithium heparin, trace-element free | Before spinninguncentrifuged | chilled in transit (insulated box, 2 frozen cool packs; temperature not recorded) | 24 hstable for the longest time tested | % difference within desirable quality specification for bias (Fraser, from intra- and interindividual variation) | Jones et al. 2021, J NutrTable 2, row 'HDL cholesterol', column LH |
Hematocrit24 findings
| Specimen, tube | Condition | Temperature | Stable for | Criterion | Study |
|---|---|---|---|---|---|
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 48 hstable for the longest time tested | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'HT', XN-9000, 4 °C |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 24 hthen changed, changed by 36 h | Bias% vs optimal bias (OP-Bias%, from biological variation; Table 1) | Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'HT', XN-9000, 4 °C |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 48 hstable for the longest time tested | Bias% vs critical difference (CD%, reference change value; Table 1) | Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'HT', XN-9000, 4 °C |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 48 hstable for the longest time tested | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'HT', BC-6800, 4 °C |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 24 hthen changed, changed by 36 h | Bias% vs optimal bias (OP-Bias%, from biological variation; Table 1) | Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'HT', BC-6800, 4 °C |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 48 hstable for the longest time tested | Bias% vs critical difference (CD%, reference change value; Table 1) | Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'HT', BC-6800, 4 °C |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 8 hstable for the longest time tested | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'HT', XN-9000, 4 °C |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 8 hstable for the longest time tested | Bias% vs optimal bias (OP-Bias%, from biological variation; Table 1) | Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'HT', XN-9000, 4 °C |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 8 hstable for the longest time tested | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'HT', BC-6800, 4 °C |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 8 hstable for the longest time tested | Bias% vs optimal bias (OP-Bias%, from biological variation; Table 1) | Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'HT', BC-6800, 4 °C |
| whole-blood-edta, K2-EDTA (BD Vacutainer) | Storageuncentrifuged | 2-8 °C | 48 hstable for the longest time tested | Repeated-measures one-way ANOVA with Tukey post hoc vs 0 h, p < 0.05 (* in table) | Unalli et al. 2021, J Med BiochemResults paragraphs 1-2 and Discussion paragraph 1 (Figures 1-2; numbers in figures only) |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | RT | 36 hthen changed, changed by 48 h | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'HT', XN-9000, RT |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | RT | 8 hthen changed, changed by 24 h | Bias% vs optimal bias (OP-Bias%, from biological variation; Table 1) | Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'HT', XN-9000, RT |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | RT | 24 hthen changed, changed by 36 h | Bias% vs critical difference (CD%, reference change value; Table 1) | Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'HT', XN-9000, RT |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | RT | 36 hthen changed, changed by 48 h | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'HT', BC-6800, RT |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | RT | 8 hthen changed, changed by 24 h | Bias% vs optimal bias (OP-Bias%, from biological variation; Table 1) | Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'HT', BC-6800, RT |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | RT | 24 hthen changed, changed by 36 h | Bias% vs critical difference (CD%, reference change value; Table 1) | Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'HT', BC-6800, RT |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | RT | 8 hstable for the longest time tested | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'HT', XN-9000, RT |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | RT | 8 hstable for the longest time tested | Bias% vs optimal bias (OP-Bias%, from biological variation; Table 1) | Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'HT', XN-9000, RT |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | RT | 8 hstable for the longest time tested | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'HT', BC-6800, RT |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | RT | 8 hstable for the longest time tested | Bias% vs optimal bias (OP-Bias%, from biological variation; Table 1) | Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'HT', BC-6800, RT |
| whole-blood-edta, K2-EDTA (BD Vacutainer) | Storageuncentrifuged | RT | 24 hthen changed, changed by 48 h | Repeated-measures one-way ANOVA with Tukey post hoc vs 0 h, p < 0.05 (* in table) | Unalli et al. 2021, J Med BiochemResults, paragraph 2 (Figure 2B) |
| whole-blood-edta, K2EDTA (Becton Dickinson) | Storageuncentrifuged | 33 °C | 0 hchanged at the first time point, changed by 4 h | Unacceptable if median % difference > Biasmax (biological variation) and Holm-adjusted Wilcoxon P < 0.05; stable if equivalence shown (two one-tailed Wilcoxon tests vs Biasmax, Pholm < 0.05) | Jain et al. 2018, Biochem Med (Zagreb)Results paragraph 1; Table 2, row 'HCT' |
| whole-blood-edta, K2-EDTA (BD Vacutainer) | Storageuncentrifuged | 10 °C | 48 hstable for the longest time tested | Repeated-measures one-way ANOVA with Tukey post hoc vs 0 h, p < 0.05 (* in table) | Unalli et al. 2021, J Med BiochemResults paragraphs 1-2 and Discussion paragraph 1 (Figures 1-2; numbers in figures only) |
Hemoglobin24 findings
| Specimen, tube | Condition | Temperature | Stable for | Criterion | Study |
|---|---|---|---|---|---|
| whole-blood-edta, BD Vacutainer K2EDTA, 2 mL (spray-coated) | Storageuncentrifuged | 2-8 °C | 24 hstable for the longest time tested | Mean % change from baseline within analytical measurement uncertainty (MU ±1.2%) | Al-Najdawi et al. 2025, Med Pharm RepTable II, row 'Hb'; Results |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 48 hstable for the longest time tested | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'Hb', XN-9000, 4 °C |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 0 hchanged at the first time point, changed by 2 h | Bias% vs optimal bias (OP-Bias%, from biological variation; Table 1) | Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'Hb', XN-9000, 4 °C |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 48 hstable for the longest time tested | Bias% vs critical difference (CD%, reference change value; Table 1) | Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'Hb', XN-9000, 4 °C |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 48 hstable for the longest time tested | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'Hb', BC-6800, 4 °C |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 48 hstable for the longest time tested | Bias% vs optimal bias (OP-Bias%, from biological variation; Table 1) | Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'Hb', BC-6800, 4 °C |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 8 hstable for the longest time tested | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'Hb', XN-9000, 4 °C |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 8 hstable for the longest time tested | Bias% vs optimal bias (OP-Bias%, from biological variation; Table 1) | Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'Hb', XN-9000, 4 °C |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 8 hstable for the longest time tested | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'Hb', BC-6800, 4 °C |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 8 hstable for the longest time tested | Bias% vs optimal bias (OP-Bias%, from biological variation; Table 1) | Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'Hb', BC-6800, 4 °C |
| whole-blood-edta, K2-EDTA (BD Vacutainer) | Storageuncentrifuged | 2-8 °C | 48 hstable for the longest time tested | Repeated-measures one-way ANOVA with Tukey post hoc vs 0 h, p < 0.05 (* in table) | Unalli et al. 2021, J Med BiochemResults paragraphs 1-2 and Discussion paragraph 1 (Figures 1-2; numbers in figures only) |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | RT | 48 hstable for the longest time tested | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'Hb', XN-9000, RT |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | RT | 48 hstable for the longest time tested | Bias% vs optimal bias (OP-Bias%, from biological variation; Table 1) | Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'Hb', XN-9000, RT |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | RT | 48 hstable for the longest time tested | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'Hb', BC-6800, RT |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | RT | 48 hstable for the longest time tested | Bias% vs optimal bias (OP-Bias%, from biological variation; Table 1) | Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'Hb', BC-6800, RT |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | RT | 8 hstable for the longest time tested | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'Hb', XN-9000, RT |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | RT | 8 hstable for the longest time tested | Bias% vs optimal bias (OP-Bias%, from biological variation; Table 1) | Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'Hb', XN-9000, RT |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | RT | 8 hstable for the longest time tested | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'Hb', BC-6800, RT |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | RT | 8 hstable for the longest time tested | Bias% vs optimal bias (OP-Bias%, from biological variation; Table 1) | Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'Hb', BC-6800, RT |
| whole-blood-edta, K2-EDTA (BD Vacutainer) | Storageuncentrifuged | RT | 48 hstable for the longest time tested | Repeated-measures one-way ANOVA with Tukey post hoc vs 0 h, p < 0.05 (* in table) | Unalli et al. 2021, J Med BiochemResults paragraphs 1-2 and Discussion paragraph 1 (Figures 1-2; numbers in figures only) |
| whole-blood-edta, K2EDTA (Becton Dickinson) | Storageuncentrifuged | 33 °C | 24 hthen changed, changed by 48 h | Unacceptable if median % difference > Biasmax (biological variation) and Holm-adjusted Wilcoxon P < 0.05; stable if equivalence shown (two one-tailed Wilcoxon tests vs Biasmax, Pholm < 0.05) | Jain et al. 2018, Biochem Med (Zagreb)Results, last paragraph (equivalence); Table 2, row 'HGB' |
| whole-blood-edta, BD Vacutainer EDTA | Storageuncentrifuged | 22-30 °C (logger 24-26 °C) | 6 hthen changed, changed by 12 h | mean % deviation from baseline within total change limit TCL = sqrt((2.77 CVa)^2 + (0.5 CVb)^2), CVb from Ricos (Table 2) | Abraham et al. 2019, Biochem Med (Zagreb)Results, paragraph 2 (Figure 3g) |
| whole-blood-edta, BD Vacutainer EDTA | Storageuncentrifuged | > 30 °C (logger 38-40 °C) | 0 hchanged at the first time point, changed by 6 h | mean % deviation from baseline within total change limit TCL = sqrt((2.77 CVa)^2 + (0.5 CVb)^2), CVb from Ricos (Table 2) | Abraham et al. 2019, Biochem Med (Zagreb)Results, paragraph 2 (Figure 3g) |
| whole-blood-edta, K2-EDTA (BD Vacutainer) | Storageuncentrifuged | 10 °C | 48 hstable for the longest time tested | Repeated-measures one-way ANOVA with Tukey post hoc vs 0 h, p < 0.05 (* in table) | Unalli et al. 2021, J Med BiochemResults paragraphs 1-2 and Discussion paragraph 1 (Figures 1-2; numbers in figures only) |
Hemoglobin A1c7 findings
| Specimen, tube | Condition | Temperature | Stable for | Criterion | Study |
|---|---|---|---|---|---|
| dbs, Whatman 903 card, mailed in foil pouch with desiccant | Storagedried | not stated; ambient postal shipping in the US (Apr-Dec 2016) | 0 dayschanged at the first time point, changed by 3 days | regression coefficient for days before freezing vs 0-2 days significant (P < .05), Model 1 (DBS value, unadjusted) | Crimmins et al. 2020, Am J Hum BiolTable 3, rows 'Days before freezing', column 'HbA1c' |
| dbs, Whatman 903 card, mailed in foil pouch with desiccant | Storagedried | not stated; ambient postal shipping in the US (Apr-Dec 2016) | 0 dayschanged at the first time point, changed by 3 days | regression coefficient for days before freezing vs 0-2 days significant (P < .05), Model 2 (DBS value adjusted for venous value) | Crimmins et al. 2020, Am J Hum BiolTable 3, rows 'Days before freezing', column 'HbA1c' |
| dbs, HemaSpot device (Spot on Sciences), integrated desiccant; fingerprick blood | Storagedried | posted by participants (ambient), then stored at 4°C in the laboratory | 4 daysstable for the longest time tested | time between sampling and testing does not significantly change the DBS-venous calibration (nested model F-test, 5% level) | Hall et al. 2020, Diabet MedTable 2; Discussion paragraph 1 |
| other, ACD (Becton Dickinson) whole blood + RPMI 1640 + DMSO in cryotube; baseline fresh from K2-EDTA (Becton Dickinson) | Storageuncentrifuged | −196°C | 7 yearsstable for the longest time tested | difference of 0.50% HbA1c considered clinically significant; observed mean bias −0.04% | Abdullah et al. 2023, J Clin Lab AnalDiscussion, paragraph 1; Table 1 |
| whole-blood-edta, BD Vacutainer EDTA | Storageuncentrifuged | RT | 24 hstable for the longest time tested | mean % deviation from baseline within total change limit TCL = sqrt((2.77 CVa)^2 + (0.5 CVb)^2), CVb from Ricos (Table 2) | Abraham et al. 2019, Biochem Med (Zagreb)Discussion, paragraph 1 (Results paragraphs 1-3; Figure 3) |
| whole-blood-edta, BD Vacutainer EDTA | Storageuncentrifuged | 22-30 °C (logger 24-26 °C) | 24 hstable for the longest time tested | mean % deviation from baseline within total change limit TCL = sqrt((2.77 CVa)^2 + (0.5 CVb)^2), CVb from Ricos (Table 2) | Abraham et al. 2019, Biochem Med (Zagreb)Discussion, paragraph 1 (Results paragraphs 1-3; Figure 3) |
| whole-blood-edta, BD Vacutainer EDTA | Storageuncentrifuged | > 30 °C (logger 38-40 °C) | 24 hstable for the longest time tested | mean % deviation from baseline within total change limit TCL = sqrt((2.77 CVa)^2 + (0.5 CVb)^2), CVb from Ricos (Table 2) | Abraham et al. 2019, Biochem Med (Zagreb)Discussion, paragraph 1 (Results paragraphs 1-3; Figure 3) |
High-fluorescence cells (HFC)8 findings
| Specimen, tube | Condition | Temperature | Stable for | Criterion | Study |
|---|---|---|---|---|---|
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 48 hstable for the longest time tested | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 4, row 'HFC', XN-9000, 4 °C |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 48 hstable for the longest time tested | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 4, row 'HFC', BC-6800, 4 °C |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 8 hstable for the longest time tested | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 5, row 'HFC', XN-9000, 4 °C |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 8 hstable for the longest time tested | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 5, row 'HFC', BC-6800, 4 °C |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | RT | 48 hstable for the longest time tested | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 4, row 'HFC', XN-9000, RT |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | RT | 48 hstable for the longest time tested | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 4, row 'HFC', BC-6800, RT |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | RT | 8 hstable for the longest time tested | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 5, row 'HFC', XN-9000, RT |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | RT | 8 hstable for the longest time tested | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 5, row 'HFC', BC-6800, RT |
High-fluorescence reticulocyte ratio (HFR)8 findings
| Specimen, tube | Condition | Temperature | Stable for | Criterion | Study |
|---|---|---|---|---|---|
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 48 hstable for the longest time tested | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'HFR', XN-9000, 4 °C |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 48 hstable for the longest time tested | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'HFR', BC-6800, 4 °C |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 8 hstable for the longest time tested | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'HFR', XN-9000, 4 °C |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 8 hstable for the longest time tested | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'HFR', BC-6800, 4 °C |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | RT | 48 hstable for the longest time tested | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'HFR', XN-9000, RT |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | RT | 48 hstable for the longest time tested | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'HFR', BC-6800, RT |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | RT | 8 hstable for the longest time tested | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'HFR', XN-9000, RT |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | RT | 8 hstable for the longest time tested | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'HFR', BC-6800, RT |
High-Sensitivity C-Reactive Protein3 findings
| Specimen, tube | Condition | Temperature | Stable for | Criterion | Study |
|---|---|---|---|---|---|
| serum, BD Vacutainer SST (SKU 367986), gel | Storagecentrifuged-on-gel | 30 °C (±1 °C) | 72 hstable for the longest time tested | mean bias vs paired baseline within CLIA acceptance limits AND >2/3 of individual results within limits | Iacovetti et al. 2026, Pract Lab MedResults, 'Analyte stability after storage at 30 °C', paragraph 1 (Table 1 image-only) |
| serum, BD SST Microtainer (SKU 395967), gel; filled from venous blood | Storagecentrifuged-on-gel | 0 °C | 72 hstable for the longest time tested | mean bias vs paired baseline within CLIA acceptance limits AND >2/3 of individual results within limits | Iacovetti et al. 2026, Pract Lab MedResults, 'Analyte stability after separation and storage in microvolume tubes at 0 °C' (Table 2 image-only) |
| whole-blood-clotted, BD Vacutainer SST (SKU 367986), gel | Before spinninguncentrifuged | 30 °C (±1 °C) | 72 hstable for the longest time tested | mean bias vs paired baseline within CLIA acceptance limits AND >2/3 of individual results within limits | Iacovetti et al. 2026, Pract Lab MedResults, 'Analyte stability after storage at 30 °C', paragraph 2 (Table 1 image-only) |
Holotranscobalamin (Active B12)4 findings
| Specimen, tube | Condition | Temperature | Stable for | Criterion | Study |
|---|---|---|---|---|---|
| whole-blood-clotted, BD Vacutainer trace element-free serum tube | Before spinninguncentrifuged | cold packs in insulated box; 9 (2) °C on arrival | 24 hstable for the longest time tested | % difference within desirable quality specification (from intra- and interindividual variation) | Jones et al. 2024, J NutrTable 2, row 'Holotranscobalamin, pmol/L'; Table 1 |
| whole-blood-clotted, Sarstedt Monovette serum | Before spinninguncentrifuged | chilled in transit (insulated box, 2 frozen cool packs; temperature not recorded) | 24 hstable for the longest time tested | % difference within desirable quality specification for bias (Fraser, from intra- and interindividual variation) | Jones et al. 2021, J NutrTable 2, row 'Holotranscobalamin', column Serum |
| whole-blood-edta, Sarstedt Monovette EDTA | Before spinninguncentrifuged | chilled in transit (insulated box, 2 frozen cool packs; temperature not recorded) | 24 hstable for the longest time tested | % difference within desirable quality specification for bias (Fraser, from intra- and interindividual variation) | Jones et al. 2021, J NutrTable 2, row 'Holotranscobalamin', column EDTA |
| whole-blood-heparin, Sarstedt Monovette lithium heparin, trace-element free | Before spinninguncentrifuged | chilled in transit (insulated box, 2 frozen cool packs; temperature not recorded) | 24 hstable for the longest time tested | % difference within desirable quality specification for bias (Fraser, from intra- and interindividual variation) | Jones et al. 2021, J NutrTable 2, row 'Holotranscobalamin', column LH |
Homocysteine3 findings
| Specimen, tube | Condition | Temperature | Stable for | Criterion | Study |
|---|---|---|---|---|---|
| plasma-li-heparin, BD Vacutainer Barricor LH 3 mL, mechanical separator | Storagecentrifuged-on-gel | 2-8 °C | 80 hstable for the longest time tested | bias vs aliquoted plasma reference tube > maximal allowable bias 7.7% (Ricos et al.); time to MAB from linear regression of bias on time to analysis | Gawria et al. 2020, J Clin Lab AnalTable 2, row 'Homocysteine', Barricor BD |
| plasma-li-heparin, BD Vacutainer PST LH 3 mL, gel | Storagecentrifuged-on-gel | 2-8 °C | 80 hstable for the longest time tested | bias vs aliquoted plasma reference tube > maximal allowable bias 7.7% (Ricos et al.); time to MAB from linear regression of bias on time to analysis | Gawria et al. 2020, J Clin Lab AnalTable 2, row 'Homocysteine', Gel BD |
| plasma-li-heparin, Greiner Vacuette LH 3 mL Sep, gel | Storagecentrifuged-on-gel | 2-8 °C | 80 hstable for the longest time tested | bias vs aliquoted plasma reference tube > maximal allowable bias 7.7% (Ricos et al.); time to MAB from linear regression of bias on time to analysis | Gawria et al. 2020, J Clin Lab AnalTable 2, row 'Homocysteine', Gel Greiner |
Immature Reticulocyte Fraction8 findings
| Specimen, tube | Condition | Temperature | Stable for | Criterion | Study |
|---|---|---|---|---|---|
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 48 hstable for the longest time tested | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'IRF', XN-9000, 4 °C |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 48 hstable for the longest time tested | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'IRF', BC-6800, 4 °C |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 8 hstable for the longest time tested | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'IRF', XN-9000, 4 °C |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 8 hstable for the longest time tested | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'IRF', BC-6800, 4 °C |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | RT | 8 hthen changed, changed by 24 h | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'IRF', XN-9000, RT |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | RT | 24 hthen changed, changed by 36 h | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'IRF', BC-6800, RT |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | RT | 8 hstable for the longest time tested | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'IRF', XN-9000, RT |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | RT | 8 hstable for the longest time tested | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'IRF', BC-6800, RT |
Immunoglobulin A (IgA)2 findings
| Specimen, tube | Condition | Temperature | Stable for | Criterion | Study |
|---|---|---|---|---|---|
| whole-blood-heparin, BD Vacutainer lithium heparin, no gel (ref 368884) | Before spinninguncentrifuged | RT | 10 hstable for the longest time tested | |mean % deviation 10 h vs baseline| > laboratory bias goal (Table II; mostly biological-variation based, some empirically modified) | Henriksen et al. 2014, Scand J Clin Lab InvestTable II, row 'IgA' |
| whole-blood-heparin, BD Vacutainer lithium heparin, no gel (ref 368884) | Before spinninguncentrifuged | RT | 10 hstable for the longest time tested | CV of paired baseline/10 h results > goal CV (Table II) | Henriksen et al. 2014, Scand J Clin Lab InvestTable II, row 'IgA' |
Immunoglobulin G (IgG)2 findings
| Specimen, tube | Condition | Temperature | Stable for | Criterion | Study |
|---|---|---|---|---|---|
| whole-blood-heparin, BD Vacutainer lithium heparin, no gel (ref 368884) | Before spinninguncentrifuged | RT | 10 hstable for the longest time tested | |mean % deviation 10 h vs baseline| > laboratory bias goal (Table II; mostly biological-variation based, some empirically modified) | Henriksen et al. 2014, Scand J Clin Lab InvestTable II, row 'IgG' |
| whole-blood-heparin, BD Vacutainer lithium heparin, no gel (ref 368884) | Before spinninguncentrifuged | RT | 10 hstable for the longest time tested | CV of paired baseline/10 h results > goal CV (Table II) | Henriksen et al. 2014, Scand J Clin Lab InvestTable II, row 'IgG' |
Immunoglobulin M (IgM)2 findings
| Specimen, tube | Condition | Temperature | Stable for | Criterion | Study |
|---|---|---|---|---|---|
| whole-blood-heparin, BD Vacutainer lithium heparin, no gel (ref 368884) | Before spinninguncentrifuged | RT | 10 hstable for the longest time tested | |mean % deviation 10 h vs baseline| > laboratory bias goal (Table II; mostly biological-variation based, some empirically modified) | Henriksen et al. 2014, Scand J Clin Lab InvestTable II, row 'IgM' |
| whole-blood-heparin, BD Vacutainer lithium heparin, no gel (ref 368884) | Before spinninguncentrifuged | RT | 10 hstable for the longest time tested | CV of paired baseline/10 h results > goal CV (Table II) | Henriksen et al. 2014, Scand J Clin Lab InvestTable II, row 'IgM' |
Insulin8 findings
| Specimen, tube | Condition | Temperature | Stable for | Criterion | Study |
|---|---|---|---|---|---|
| plasma-edta, Sarstedt K+-EDTA 2.7 mL | Freeze–thawnot-stated | −80 °C | 5 cyclesstable for the longest time tested | mean % change from baseline > 10% considered clinically significant | McDonald et al. 2012, PLoS OneTable 1 |
| plasma-edta, Sarstedt K+-EDTA 2.7 mL | Storagecentrifuged-on-cells | RT | 24 hstable for the longest time tested | mean % change from baseline > 10% considered clinically significant | McDonald et al. 2012, PLoS OneResults 1.1, paragraph 2 |
| serum, Sarstedt serum gel 2.7 mL | Freeze–thawnot-stated | −80 °C | 5 cyclesstable for the longest time tested | mean % change from baseline > 10% considered clinically significant | McDonald et al. 2012, PLoS OneTable 1 |
| whole-blood-clotted, Sarstedt serum gel 2.7 mL | Before spinninguncentrifuged | 2-8 °C | 120 hstable for the longest time tested | mean % change from baseline > 10% considered clinically significant | McDonald et al. 2012, PLoS OneResults 1.2 |
| whole-blood-edta, Sarstedt K+-EDTA 2.7 mL | Before spinninguncentrifuged | 2-8 °C | 24 hstable for the longest time tested | mean % change from baseline > 10% considered clinically significant | McDonald et al. 2012, PLoS OneResults 1.1, paragraph 3 |
| whole-blood-edta, Sarstedt K+-EDTA 2.7 mL | Before spinninguncentrifuged | 2-8 °C | 120 hstable for the longest time tested | mean % change from baseline > 10% considered clinically significant | McDonald et al. 2012, PLoS OneResults 1.2 |
| whole-blood-edta, Sarstedt K+-EDTA 2.7 mL | Before spinninguncentrifuged | RT | 24 hstable for the longest time tested | mean % change from baseline > 10% considered clinically significant | McDonald et al. 2012, PLoS OneResults 1.1, paragraph 2 |
| whole-blood-edta, Sarstedt K+-EDTA 2.7 mL | Before spinninguncentrifuged | RT | 120 hstable for the longest time tested | mean % change from baseline > 10% considered clinically significant | McDonald et al. 2012, PLoS OneResults 1.2 |
International Normalized Ratio7 findings
| Specimen, tube | Condition | Temperature | Stable for | Criterion | Study |
|---|---|---|---|---|---|
| plasma-citrate, BD Vacutainer 2.7 mL, 3.2% buffered sodium citrate; PPP aliquoted into microcentrifuge tubes | Storageseparated | −20 °C | 36 hstable for the longest time tested | change of more than ±10% of mean from baseline | Patil et al. 2022, CureusTable 2 |
| plasma-citrate, BD 0.109 M sodium citrate (9:1); plasma aliquoted into nonactivating Eppendorf tubes | Storageseparated | 2-8 °C | 24 hstable for the longest time tested | mean % change from baseline >10%, or >25% of samples with individual change >10% (van Geest-Daalderop) | Feng et al. 2014, Sci RepTable 2, 'Acceptable Time (h)' |
| plasma-citrate, BD Vacutainer 2.7 mL, 3.2% buffered sodium citrate; PPP aliquoted into microcentrifuge tubes | Storageseparated | 2-8 °C | 24 hthen changed, changed by 36 h | change of more than ±10% of mean from baseline | Patil et al. 2022, CureusTable 2 |
| plasma-citrate, BD 0.109 M sodium citrate (9:1); plasma aliquoted into nonactivating Eppendorf tubes | Storageseparated | RT | 24 hstable for the longest time tested | mean % change from baseline >10%, or >25% of samples with individual change >10% (van Geest-Daalderop) | Feng et al. 2014, Sci RepTable 2, 'Acceptable Time (h)' |
| plasma-citrate, BD Vacutainer 2.7 mL, 3.2% buffered sodium citrate; PPP aliquoted into microcentrifuge tubes | Storageseparated | RT | 0 hchanged at the first time point, changed by 12 h | change of more than ±10% of mean from baseline | Patil et al. 2022, CureusTable 2 |
| whole-blood-citrate, Greiner Vacuette sodium citrate, standard fill (blood–air interface) | Before spinninguncentrifuged | RT | 0 hchanged at the first time point, changed by 2 h | maximum acceptable bias = 0.375 × total biological variation (EFLM BV database, Model 2) | Storm et al. 2026, Adv HematolTable 2, 'Standard tubes' row 'INR' |
| whole-blood-citrate, Greiner Vacuette sodium citrate, completely filled (no air) | Before spinninguncentrifuged | RT | 0 hchanged at the first time point, changed by 2 h | maximum acceptable bias = 0.375 × total biological variation (EFLM BV database, Model 2) | Storm et al. 2026, Adv HematolTable 2, 'Full tubes' row 'INR' |
Iron2 findings
| Specimen, tube | Condition | Temperature | Stable for | Criterion | Study |
|---|---|---|---|---|---|
| whole-blood-heparin, BD Vacutainer lithium heparin, no gel (ref 368884) | Before spinninguncentrifuged | RT | 10 hstable for the longest time tested | |mean % deviation 10 h vs baseline| > laboratory bias goal (Table II; mostly biological-variation based, some empirically modified) | Henriksen et al. 2014, Scand J Clin Lab InvestTable II, row 'Fe' |
| whole-blood-heparin, BD Vacutainer lithium heparin, no gel (ref 368884) | Before spinninguncentrifuged | RT | 10 hstable for the longest time tested | CV of paired baseline/10 h results > goal CV (Table II) | Henriksen et al. 2014, Scand J Clin Lab InvestTable II, row 'Fe' |
Lactate1 finding
| Specimen, tube | Condition | Temperature | Stable for | Criterion | Study |
|---|---|---|---|---|---|
| plasma-li-heparin, BD Vacutainer LH PST II gel separator, 3.0 mL (REF 367374) | Storagecentrifuged-on-gel | 2-8 °C | 24 hstable for the longest time tested | mean % difference smaller than TEa (Ricos: 0.25 x (CVi^2 + CVg^2)^0.5 + 1.65 x 0.5 x CVi) | Nonkes et al. 2019, Biochem Med (Zagreb)Table 1, row 'Lactate, mmol/L' |
Lactate Dehydrogenase (LDH)20 findings
| Specimen, tube | Condition | Temperature | Stable for | Criterion | Study |
|---|---|---|---|---|---|
| plasma-li-heparin, Sarstedt lithium heparin, no gel separator (primary tube) | Storagecentrifuged-on-cells | 2-8 °C | 2.9 hthen changed, changed by 6 h | mean % difference vs time 0 exceeds total limit of change, TLC = ±√((1.65·CVa)² + (0.5·CVb)²) (EFLM BV), here 11.5%; stability limit from instability-equation regression | Illana et al. 2024, Biochem Med (Zagreb)Table 1, row 'LD' |
| plasma-li-heparin, Sarstedt lithium heparin, no gel separator; plasma aliquot | Storageseparated | 2-8 °C | 63 hthen changed, changed by 72 h | mean % difference vs time 0 exceeds total limit of change, TLC = ±√((1.65·CVa)² + (0.5·CVb)²) (EFLM BV), here 11.5%; stability limit from instability-equation regression | Illana et al. 2024, Biochem Med (Zagreb)Table 2, row 'LD' |
| plasma-li-heparin, BD Vacutainer lithium heparin 6 mL (regular) | Storagenot-stated | 2-8 °C | 4 hthen changed, changed by 24 h | mean % difference vs T0 exceeds total change limit (TCL), here ±8.82% | Dupuy et al. 2018, Biochem Med (Zagreb)Table 2, BD lithium heparin tubes, row 'LD, U/L' |
| plasma-li-heparin, BD Barricor lithium heparin 5.5 mL (mechanical separator) | Storagecentrifuged-on-gel | 2-8 °C | 72 hstable for the longest time tested | mean % difference vs T0 exceeds total change limit (TCL), here ±8.82% | Dupuy et al. 2018, Biochem Med (Zagreb)Table 2, BD Barricor lithium heparin tubes, row 'LD, U/L' |
| plasma-li-heparin, BD Vacutainer LH PST II gel separator, 3.0 mL (REF 367374) | Storagecentrifuged-on-gel | 2-8 °C | 24 hstable for the longest time tested | mean % difference smaller than TEa (Ricos: 0.25 x (CVi^2 + CVg^2)^0.5 + 1.65 x 0.5 x CVi) | Nonkes et al. 2019, Biochem Med (Zagreb)Table 1, row 'LD, U/L' |
| plasma-li-heparin, BD Vacutainer Barricor LH 3 mL, mechanical separator | Storagecentrifuged-on-gel | 2-8 °C | 0 hchanged at the first time point, changed by 3 h | bias vs aliquoted plasma reference tube > maximal allowable bias 4.3% (Ricos et al.); time to MAB from linear regression of bias on time to analysis | Gawria et al. 2020, J Clin Lab AnalTable 2, row 'LDH', Barricor BD |
| plasma-li-heparin, BD Vacutainer PST LH 3 mL, gel | Storagecentrifuged-on-gel | 2-8 °C | 0 hchanged at the first time point, changed by 3 h | bias vs aliquoted plasma reference tube > maximal allowable bias 4.3% (Ricos et al.); time to MAB from linear regression of bias on time to analysis | Gawria et al. 2020, J Clin Lab AnalTable 2, row 'LDH', Gel BD |
| plasma-li-heparin, Greiner Vacuette LH 3 mL Sep, gel | Storagecentrifuged-on-gel | 2-8 °C | 0 hchanged at the first time point, changed by 3 h | bias vs aliquoted plasma reference tube > maximal allowable bias 4.3% (Ricos et al.); time to MAB from linear regression of bias on time to analysis | Gawria et al. 2020, J Clin Lab AnalTable 2, row 'LDH', Gel Greiner |
| plasma-li-heparin, BD Barricor 4.5 mL (lithium heparin, mechanical separator) | Storagecentrifuged-on-gel | 2-8 °C | 0 hchanged at the first time point, changed by 24 h | mean % difference vs time 0 exceeds desirable imprecision based on biological variation (EFLM/Westgard), here 2.6% | Mandić et al. 2020, Biochem Med (Zagreb)Tables 3 and 4, row 'LD, U/L', BD Barricor columns |
| serum, BD Vacutainer 8.5 mL plastic serum tube; pooled serum aliquoted into 1.5 mL Eppendorf tubes | Freeze–thawseparated | −20 °C | 5 cyclesthen changed, changed by 6 cycles | unstable only if both P < 0.005 (Bonferroni) and % change from T0 > desirable bias (Westgard/Ricos), here 4.3% | Cuhadar et al. 2013, Biochem Med (Zagreb)Table 2, row 'LD (U/L)', columns T1d-T10d |
| serum, BD Vacutainer 8.5 mL plastic serum tube; pooled serum aliquoted into 1.5 mL Eppendorf tubes | Storageseparated | −20 °C | 3 monthsstable for the longest time tested | unstable only if both P < 0.013 (Bonferroni) and % change from T0 > desirable bias (Westgard/Ricos), here 4.3% | Cuhadar et al. 2013, Biochem Med (Zagreb)Table 2, row 'LD (U/L)', columns T1m-T3m |
| serum, Vacuette Tube Serum Gel Separator 8 mL (Greiner, 455071) | Storagecentrifuged-on-gel | 2-8 °C | 1 weekstable for the longest time tested | % bias of median vs baseline exceeds desirable bias (Westgard/Ricos BV), here 4.3% | Cuhadar et al. 2012, Biochem Med (Zagreb)Table 3, row 'Lactate dehydrogenase', gel 4°C |
| serum, Vacuette Tube Serum No Gel 9 mL (Greiner, 455092), plain | Storagecentrifuged-on-cells | 2-8 °C | 24 hthen changed, changed by 30 h | % bias of median vs baseline exceeds desirable bias (Westgard/Ricos BV), here 4.3% | Cuhadar et al. 2012, Biochem Med (Zagreb)Table 3, row 'Lactate dehydrogenase', w/o gel 4°C |
| serum, BD Vacutainer RST 4.0 mL (thrombin clot activator, gel) | Storagecentrifuged-on-gel | 2-8 °C | 24 hstable for the longest time tested | clinically significant if mean bias vs t0 exceeds desirable bias (biological variation), here 2.0% | Kocijancic et al. 2014, Biochem Med (Zagreb)Table 3, row 'LD' |
| serum, BD Vacutainer SST II Advance 5.0 mL (clot activator, gel) | Storagecentrifuged-on-gel | 2-8 °C | 24 hstable for the longest time tested | clinically significant if mean bias vs t0 exceeds desirable bias (biological variation), here 2.0% | Kocijancic et al. 2014, Biochem Med (Zagreb)Table 4, row 'LD' |
| serum, BD Vacutainer RST 5 mL (thrombin clot activator, gel) | Storagecentrifuged-on-gel | 2-8 °C | 24 hthen changed, changed by 7 days | mean % difference vs time 0 exceeds desirable imprecision based on biological variation (EFLM/Westgard), here 2.6% | Mandić et al. 2020, Biochem Med (Zagreb)Tables 3 and 4, row 'LD, U/L', BD RST columns |
| serum, Vacuette Tube Serum Gel Separator 8 mL (Greiner, 455071) | Storagecentrifuged-on-gel | RT | 1 weekstable for the longest time tested | % bias of median vs baseline exceeds desirable bias (Westgard/Ricos BV), here 4.3% | Cuhadar et al. 2012, Biochem Med (Zagreb)Table 3, row 'Lactate dehydrogenase', gel 24°C |
| serum, Vacuette Tube Serum No Gel 9 mL (Greiner, 455092), plain | Storagecentrifuged-on-cells | RT | 72 hthen changed, changed by 1 week | % bias of median vs baseline exceeds desirable bias (Westgard/Ricos BV), here 4.3% | Cuhadar et al. 2012, Biochem Med (Zagreb)Table 3, row 'Lactate dehydrogenase', w/o gel 24°C |
| whole-blood-heparin, BD Vacutainer lithium heparin, no gel (ref 368884) | Before spinninguncentrifuged | RT | 10 hstable for the longest time tested | |mean % deviation 10 h vs baseline| > laboratory bias goal (Table II; mostly biological-variation based, some empirically modified) | Henriksen et al. 2014, Scand J Clin Lab InvestTable II, row 'LD' |
| whole-blood-heparin, BD Vacutainer lithium heparin, no gel (ref 368884) | Before spinninguncentrifuged | RT | 10 hstable for the longest time tested | CV of paired baseline/10 h results > goal CV (Table II) | Henriksen et al. 2014, Scand J Clin Lab InvestTable II, row 'LD' |
LDL Cholesterol4 findings
| Specimen, tube | Condition | Temperature | Stable for | Criterion | Study |
|---|---|---|---|---|---|
| serum, BD Vacutainer SST | Freeze–thawseparated | −70 °C, 1 month; thawed at room temperature | 1 cyclestable for the longest time tested | mean % change (bias) from fresh baseline within total allowable error (TEa, desirable bias) | Abraham et al. 2021, J Lab PhysiciansTable 2 and Table 3, row 'LDL (mg/dL)' |
| whole-blood-clotted, BD Vacutainer SST | Before spinninguncentrifuged | RT | 24 hstable for the longest time tested | mean % deviation from baseline within total change limit TCL = sqrt((2.77 CVa)^2 + (0.5 CVb)^2), CVb from Ricos (Table 2) | Abraham et al. 2019, Biochem Med (Zagreb)Discussion, paragraph 1 (Results paragraphs 1-3; Figure 3) |
| whole-blood-clotted, BD Vacutainer SST | Before spinninguncentrifuged | 22-30 °C (logger 24-26 °C) | 24 hstable for the longest time tested | mean % deviation from baseline within total change limit TCL = sqrt((2.77 CVa)^2 + (0.5 CVb)^2), CVb from Ricos (Table 2) | Abraham et al. 2019, Biochem Med (Zagreb)Discussion, paragraph 1 (Results paragraphs 1-3; Figure 3) |
| whole-blood-clotted, BD Vacutainer SST | Before spinninguncentrifuged | > 30 °C (logger 38-40 °C) | 24 hstable for the longest time tested | mean % deviation from baseline within total change limit TCL = sqrt((2.77 CVa)^2 + (0.5 CVb)^2), CVb from Ricos (Table 2) | Abraham et al. 2019, Biochem Med (Zagreb)Discussion, paragraph 1 (Results paragraphs 1-3; Figure 3) |
Lipase2 findings
| Specimen, tube | Condition | Temperature | Stable for | Criterion | Study |
|---|---|---|---|---|---|
| plasma-li-heparin, Sarstedt lithium heparin, no gel separator (primary tube) | Storagecentrifuged-on-cells | 2-8 °C | 48 hstable for the longest time tested | mean % difference vs time 0 exceeds total limit of change, TLC = ±√((1.65·CVa)² + (0.5·CVb)²) (EFLM BV), here 17.2%; stability limit from instability-equation regression | Illana et al. 2024, Biochem Med (Zagreb)Table 1, row 'Lip' |
| plasma-li-heparin, Sarstedt lithium heparin, no gel separator; plasma aliquot | Storageseparated | 2-8 °C | 96 hstable for the longest time tested | mean % difference vs time 0 exceeds total limit of change, TLC = ±√((1.65·CVa)² + (0.5·CVb)²) (EFLM BV), here 17.2%; stability limit from instability-equation regression | Illana et al. 2024, Biochem Med (Zagreb)Table 2, row 'Lip' |
Low-fluorescence reticulocyte ratio (LFR)8 findings
| Specimen, tube | Condition | Temperature | Stable for | Criterion | Study |
|---|---|---|---|---|---|
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 48 hstable for the longest time tested | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'LFR', XN-9000, 4 °C |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 48 hstable for the longest time tested | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'LFR', BC-6800, 4 °C |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 8 hstable for the longest time tested | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'LFR', XN-9000, 4 °C |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 8 hstable for the longest time tested | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'LFR', BC-6800, 4 °C |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | RT | 8 hthen changed, changed by 24 h | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'LFR', XN-9000, RT |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | RT | 24 hthen changed, changed by 36 h | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'LFR', BC-6800, RT |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | RT | 8 hstable for the longest time tested | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'LFR', XN-9000, RT |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | RT | 8 hstable for the longest time tested | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'LFR', BC-6800, RT |
Lupus anticoagulant screen/confirm ratio3 findings
| Specimen, tube | Condition | Temperature | Stable for | Criterion | Study |
|---|---|---|---|---|---|
| plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquots | Freeze–thawseparated | −80 °C | 1 cyclelimit stated by the authors | max allowable mean deviation from t0: ±10.0% up to INR 1.5;±20.0% for INR ≥1.5 based on seconds | Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability' |
| plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquots | Storageseparated | −80 °C | 12 weekslimit stated by the authors | max allowable mean deviation from t0: ±10.0% up to INR 1.5;±20.0% for INR ≥1.5 based on seconds | Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability' |
| plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquots | Storageseparated | RT | 4 hlimit stated by the authors | max allowable mean deviation from t0: ±10.0% up to INR 1.5;±20.0% for INR ≥1.5 based on seconds | Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability' |
Lutein4 findings
| Specimen, tube | Condition | Temperature | Stable for | Criterion | Study |
|---|---|---|---|---|---|
| whole-blood-clotted, Sarstedt Monovette serum | Before spinninguncentrifuged | chilled in transit (insulated box, 2 frozen cool packs; temperature not recorded) | 24 hstable for the longest time tested | % difference within desirable quality specification for bias (Fraser, from intra- and interindividual variation) | Jones et al. 2021, J NutrTable 2, row 'Lutein', column Serum |
| whole-blood-edta, Sarstedt Monovette EDTA | Before spinninguncentrifuged | chilled in transit (insulated box, 2 frozen cool packs; temperature not recorded) | 24 hstable for the longest time tested | % difference within desirable quality specification for bias (Fraser, from intra- and interindividual variation) | Jones et al. 2021, J NutrTable 2, row 'Lutein', column EDTA |
| whole-blood-heparin, BD Vacutainer lithium heparin | Before spinninguncentrifuged | cold packs in insulated box; 9 (2) °C on arrival | 24 hstable for the longest time tested | % difference within desirable quality specification (from intra- and interindividual variation) | Jones et al. 2024, J NutrTable 2, row 'Lutein, μmol/L'; Table 1 |
| whole-blood-heparin, Sarstedt Monovette lithium heparin, trace-element free | Before spinninguncentrifuged | chilled in transit (insulated box, 2 frozen cool packs; temperature not recorded) | 24 hstable for the longest time tested | % difference within desirable quality specification for bias (Fraser, from intra- and interindividual variation) | Jones et al. 2021, J NutrTable 2, row 'Lutein', column LH |
LY-WX6 findings
| Specimen, tube | Condition | Temperature | Stable for | Criterion | Study |
|---|---|---|---|---|---|
| whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mL | Storageuncentrifuged | 2-8 °C | 24 hstable for the longest time tested | Approach 2: absolute % difference between medians (Tx vs T0) must be < 10% | Roccaforte et al. 2026, Biochem Med (Zagreb)Table 1, row 'LY-WX', 4 °C, column 'Approach 2 (% difference between medians)' |
| whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mL | Storageuncentrifuged | 2-8 °C | 24 hstable for the longest time tested | EFLM regression: time at which fitted % difference line reaches MPE 10% | Roccaforte et al. 2026, Biochem Med (Zagreb)Table 3, row 'LY-WX', 4 °C |
| whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mL | Storageuncentrifuged | RT | 24 hstable for the longest time tested | Approach 2: absolute % difference between medians (Tx vs T0) must be < 10% | Roccaforte et al. 2026, Biochem Med (Zagreb)Table 1, row 'LY-WX', RT, column 'Approach 2 (% difference between medians)' |
| whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mL | Storageuncentrifuged | RT | 24 hstable for the longest time tested | EFLM regression: time at which fitted % difference line reaches MPE 10% | Roccaforte et al. 2026, Biochem Med (Zagreb)Table 3, row 'LY-WX', RT |
| whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mL | Storageuncentrifuged | 37 °C | 12 hthen changed, changed by 24 h | Approach 2: absolute % difference between medians (Tx vs T0) must be < 10% | Roccaforte et al. 2026, Biochem Med (Zagreb)Table 2, row 'LY-WX', 37 °C (blank in Table 1), column 'Approach 2 (% difference between medians)' |
| whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mL | Storageuncentrifuged | 37 °C | 24 hstable for the longest time tested | EFLM regression: time at which fitted % difference line reaches MPE 10% | Roccaforte et al. 2026, Biochem Med (Zagreb)Table 3, row 'LY-WX', 37 °C |
LY-WY4 findings
| Specimen, tube | Condition | Temperature | Stable for | Criterion | Study |
|---|---|---|---|---|---|
| whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mL | Storageuncentrifuged | 2-8 °C | 24 hstable for the longest time tested | Approach 2: absolute % difference between medians (Tx vs T0) must be < 10% | Roccaforte et al. 2026, Biochem Med (Zagreb)Table 1, row 'LY-WY', 4 °C, column 'Approach 2 (% difference between medians)' |
| whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mL | Storageuncentrifuged | 2-8 °C | 24 hstable for the longest time tested | EFLM regression: time at which fitted % difference line reaches MPE 10% | Roccaforte et al. 2026, Biochem Med (Zagreb)Table 3, row 'LY-WY', 4 °C |
| whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mL | Storageuncentrifuged | RT | 24 hstable for the longest time tested | Approach 2: absolute % difference between medians (Tx vs T0) must be < 10% | Roccaforte et al. 2026, Biochem Med (Zagreb)Table 1, row 'LY-WY', RT, column 'Approach 2 (% difference between medians)' |
| whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mL | Storageuncentrifuged | 37 °C | 13 hlimit stated by the authors | EFLM regression: time at which fitted % difference line reaches MPE 10% | Roccaforte et al. 2026, Biochem Med (Zagreb)Table 3, row 'LY-WY', 37 °C |
LY-WZ5 findings
| Specimen, tube | Condition | Temperature | Stable for | Criterion | Study |
|---|---|---|---|---|---|
| whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mL | Storageuncentrifuged | 2-8 °C | 24 hstable for the longest time tested | Approach 2: absolute % difference between medians (Tx vs T0) must be < 10% | Roccaforte et al. 2026, Biochem Med (Zagreb)Table 1, row 'LY-WZ', 4 °C, column 'Approach 2 (% difference between medians)' |
| whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mL | Storageuncentrifuged | 2-8 °C | 24 hstable for the longest time tested | EFLM regression: time at which fitted % difference line reaches MPE 10% | Roccaforte et al. 2026, Biochem Med (Zagreb)Table 3, row 'LYWZ', 4 °C |
| whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mL | Storageuncentrifuged | RT | 24 hstable for the longest time tested | Approach 2: absolute % difference between medians (Tx vs T0) must be < 10% | Roccaforte et al. 2026, Biochem Med (Zagreb)Table 1, row 'LY-WZ', RT, column 'Approach 2 (% difference between medians)' |
| whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mL | Storageuncentrifuged | RT | 24 hstable for the longest time tested | EFLM regression: time at which fitted % difference line reaches MPE 10% | Roccaforte et al. 2026, Biochem Med (Zagreb)Table 3, row 'LYWZ', RT |
| whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mL | Storageuncentrifuged | 37 °C | 7 hlimit stated by the authors | EFLM regression: time at which fitted % difference line reaches MPE 10% | Roccaforte et al. 2026, Biochem Med (Zagreb)Table 3, row 'LYWZ', 37 °C |
LY-X9 findings
| Specimen, tube | Condition | Temperature | Stable for | Criterion | Study |
|---|---|---|---|---|---|
| whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mL | Storageuncentrifuged | 2-8 °C | 24 hstable for the longest time tested | Approach 2: absolute % difference between medians (Tx vs T0) must be < 10% | Roccaforte et al. 2026, Biochem Med (Zagreb)Table 1, row 'LY-X', 4 °C, column 'Approach 2 (% difference between medians)' |
| whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mL | Storageuncentrifuged | 2-8 °C | 24 hstable for the longest time tested | Approach 3: fewer than 2 of 25 samples with absolute % change from baseline > 10% | Roccaforte et al. 2026, Biochem Med (Zagreb)Table 1, row 'LY-X', 4 °C, column 'Approach 3 (per-sample % difference)' |
| whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mL | Storageuncentrifuged | 2-8 °C | 24 hstable for the longest time tested | EFLM regression: time at which fitted % difference line reaches MPE 10% | Roccaforte et al. 2026, Biochem Med (Zagreb)Table 3, row 'LY-X', 4 °C |
| whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mL | Storageuncentrifuged | RT | 24 hstable for the longest time tested | Approach 2: absolute % difference between medians (Tx vs T0) must be < 10% | Roccaforte et al. 2026, Biochem Med (Zagreb)Table 1, row 'LY-X', RT, column 'Approach 2 (% difference between medians)' |
| whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mL | Storageuncentrifuged | RT | 24 hstable for the longest time tested | Approach 3: fewer than 2 of 25 samples with absolute % change from baseline > 10% | Roccaforte et al. 2026, Biochem Med (Zagreb)Table 1, row 'LY-X', RT, column 'Approach 3 (per-sample % difference)' |
| whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mL | Storageuncentrifuged | RT | 24 hstable for the longest time tested | EFLM regression: time at which fitted % difference line reaches MPE 10% | Roccaforte et al. 2026, Biochem Med (Zagreb)Table 3, row 'LY-X', RT |
| whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mL | Storageuncentrifuged | 37 °C | 24 hstable for the longest time tested | Approach 2: absolute % difference between medians (Tx vs T0) must be < 10% | Roccaforte et al. 2026, Biochem Med (Zagreb)Table 1, row 'LY-X', 37 °C, column 'Approach 2 (% difference between medians)' |
| whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mL | Storageuncentrifuged | 37 °C | 12 hthen changed, changed by 24 h | Approach 3: fewer than 2 of 25 samples with absolute % change from baseline > 10% | Roccaforte et al. 2026, Biochem Med (Zagreb)Table 2, row 'LY-X', 37 °C (blank in Table 1), column 'Approach 3 (per-sample % difference)' |
| whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mL | Storageuncentrifuged | 37 °C | 24 hstable for the longest time tested | EFLM regression: time at which fitted % difference line reaches MPE 10% | Roccaforte et al. 2026, Biochem Med (Zagreb)Table 3, row 'LY-X', 37 °C |
LY-Y8 findings
| Specimen, tube | Condition | Temperature | Stable for | Criterion | Study |
|---|---|---|---|---|---|
| whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mL | Storageuncentrifuged | 2-8 °C | 24 hstable for the longest time tested | Approach 2: absolute % difference between medians (Tx vs T0) must be < 10% | Roccaforte et al. 2026, Biochem Med (Zagreb)Table 1, row 'LY-Y', 4 °C, column 'Approach 2 (% difference between medians)' |
| whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mL | Storageuncentrifuged | 2-8 °C | 24 hstable for the longest time tested | Approach 3: fewer than 2 of 25 samples with absolute % change from baseline > 10% | Roccaforte et al. 2026, Biochem Med (Zagreb)Table 1, row 'LY-Y', 4 °C, column 'Approach 3 (per-sample % difference)' |
| whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mL | Storageuncentrifuged | 2-8 °C | 24 hstable for the longest time tested | EFLM regression: time at which fitted % difference line reaches MPE 10% | Roccaforte et al. 2026, Biochem Med (Zagreb)Table 3, row 'LY-Y', 4 °C |
| whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mL | Storageuncentrifuged | RT | 24 hstable for the longest time tested | Approach 2: absolute % difference between medians (Tx vs T0) must be < 10% | Roccaforte et al. 2026, Biochem Med (Zagreb)Table 1, row 'LY-Y', RT, column 'Approach 2 (% difference between medians)' |
| whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mL | Storageuncentrifuged | RT | 24 hstable for the longest time tested | Approach 3: fewer than 2 of 25 samples with absolute % change from baseline > 10% | Roccaforte et al. 2026, Biochem Med (Zagreb)Table 1, row 'LY-Y', RT, column 'Approach 3 (per-sample % difference)' |
| whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mL | Storageuncentrifuged | RT | 24 hstable for the longest time tested | EFLM regression: time at which fitted % difference line reaches MPE 10% | Roccaforte et al. 2026, Biochem Med (Zagreb)Table 3, row 'LY-Y', RT |
| whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mL | Storageuncentrifuged | 37 °C | 12 hthen changed, changed by 24 h | Approach 2: absolute % difference between medians (Tx vs T0) must be < 10% | Roccaforte et al. 2026, Biochem Med (Zagreb)Table 2, row 'LY-Y', 37 °C (blank in Table 1), column 'Approach 2 (% difference between medians)' |
| whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mL | Storageuncentrifuged | 37 °C | 8 hlimit stated by the authors | EFLM regression: time at which fitted % difference line reaches MPE 10% | Roccaforte et al. 2026, Biochem Med (Zagreb)Table 3, row 'LY-Y', 37 °C |
LY-Z6 findings
| Specimen, tube | Condition | Temperature | Stable for | Criterion | Study |
|---|---|---|---|---|---|
| whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mL | Storageuncentrifuged | 2-8 °C | 17 hlimit stated by the authors | EFLM regression: time at which fitted % difference line reaches MPE 10% | Roccaforte et al. 2026, Biochem Med (Zagreb)Table 3, row 'LY-Z', 4 °C |
| whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mL | Storageuncentrifuged | RT | 24 hstable for the longest time tested | Approach 2: absolute % difference between medians (Tx vs T0) must be < 10% | Roccaforte et al. 2026, Biochem Med (Zagreb)Table 1, row 'LY-Z', RT, column 'Approach 2 (% difference between medians)' |
| whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mL | Storageuncentrifuged | RT | 24 hstable for the longest time tested | Approach 3: fewer than 2 of 25 samples with absolute % change from baseline > 10% | Roccaforte et al. 2026, Biochem Med (Zagreb)Table 1, row 'LY-Z', RT, column 'Approach 3 (per-sample % difference)' |
| whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mL | Storageuncentrifuged | RT | 24 hstable for the longest time tested | EFLM regression: time at which fitted % difference line reaches MPE 10% | Roccaforte et al. 2026, Biochem Med (Zagreb)Table 3, row 'LY-Z', RT |
| whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mL | Storageuncentrifuged | 37 °C | 12 hthen changed, changed by 24 h | Approach 2: absolute % difference between medians (Tx vs T0) must be < 10% | Roccaforte et al. 2026, Biochem Med (Zagreb)Table 2, row 'LY-Z', 37 °C (blank in Table 1), column 'Approach 2 (% difference between medians)' |
| whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mL | Storageuncentrifuged | 37 °C | 14 hlimit stated by the authors | EFLM regression: time at which fitted % difference line reaches MPE 10% | Roccaforte et al. 2026, Biochem Med (Zagreb)Table 3, row 'LY-Z', 37 °C |
Lycopene4 findings
| Specimen, tube | Condition | Temperature | Stable for | Criterion | Study |
|---|---|---|---|---|---|
| whole-blood-clotted, Sarstedt Monovette serum | Before spinninguncentrifuged | chilled in transit (insulated box, 2 frozen cool packs; temperature not recorded) | 24 hstable for the longest time tested | % difference within desirable quality specification for bias (Fraser, from intra- and interindividual variation) | Jones et al. 2021, J NutrTable 2, row 'Lycopene', column Serum |
| whole-blood-edta, Sarstedt Monovette EDTA | Before spinninguncentrifuged | chilled in transit (insulated box, 2 frozen cool packs; temperature not recorded) | 24 hstable for the longest time tested | % difference within desirable quality specification for bias (Fraser, from intra- and interindividual variation) | Jones et al. 2021, J NutrTable 2, row 'Lycopene', column EDTA |
| whole-blood-heparin, BD Vacutainer lithium heparin | Before spinninguncentrifuged | cold packs in insulated box; 9 (2) °C on arrival | 24 hstable for the longest time tested | % difference within desirable quality specification (from intra- and interindividual variation) | Jones et al. 2024, J NutrTable 2, row 'Lycopene, μmol/L'; Table 1 |
| whole-blood-heparin, Sarstedt Monovette lithium heparin, trace-element free | Before spinninguncentrifuged | chilled in transit (insulated box, 2 frozen cool packs; temperature not recorded) | 24 hstable for the longest time tested | % difference within desirable quality specification for bias (Fraser, from intra- and interindividual variation) | Jones et al. 2021, J NutrTable 2, row 'Lycopene', column LH |
Lymphocytes (Absolute)21 findings
| Specimen, tube | Condition | Temperature | Stable for | Criterion | Study |
|---|---|---|---|---|---|
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 48 hstable for the longest time tested | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 4, row 'LY', XN-9000, 4 °C |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 48 hstable for the longest time tested | Bias% vs optimal bias (OP-Bias%, from biological variation; Table 1) | Buoro et al. 2016, Rev Bras Hematol HemoterTable 4, row 'LY', XN-9000, 4 °C |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 48 hstable for the longest time tested | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 4, row 'LY', BC-6800, 4 °C |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 8 hthen changed, changed by 24 h | Bias% vs optimal bias (OP-Bias%, from biological variation; Table 1) | Buoro et al. 2016, Rev Bras Hematol HemoterTable 4, row 'LY', BC-6800, 4 °C |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 48 hstable for the longest time tested | Bias% vs critical difference (CD%, reference change value; Table 1) | Buoro et al. 2016, Rev Bras Hematol HemoterTable 4, row 'LY', BC-6800, 4 °C |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 8 hstable for the longest time tested | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 5, row 'LY', XN-9000, 4 °C |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 8 hstable for the longest time tested | Bias% vs optimal bias (OP-Bias%, from biological variation; Table 1) | Buoro et al. 2016, Rev Bras Hematol HemoterTable 5, row 'LY', XN-9000, 4 °C |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 8 hstable for the longest time tested | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 5, row 'LY', BC-6800, 4 °C |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 4 hthen changed, changed by 8 h | Bias% vs optimal bias (OP-Bias%, from biological variation; Table 1) | Buoro et al. 2016, Rev Bras Hematol HemoterTable 5, row 'LY', BC-6800, 4 °C |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 8 hstable for the longest time tested | Bias% vs critical difference (CD%, reference change value; Table 1) | Buoro et al. 2016, Rev Bras Hematol HemoterTable 5, row 'LY', BC-6800, 4 °C |
| whole-blood-edta, K2-EDTA (BD Vacutainer) | Storageuncentrifuged | 2-8 °C | 48 hstable for the longest time tested | Repeated-measures one-way ANOVA with Tukey post hoc vs 0 h, p < 0.05 (* in table) | Unalli et al. 2021, J Med BiochemTable 1, column '4 °C' (no * at any time point) |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | RT | 48 hstable for the longest time tested | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 4, row 'LY', XN-9000, RT |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | RT | 48 hstable for the longest time tested | Bias% vs optimal bias (OP-Bias%, from biological variation; Table 1) | Buoro et al. 2016, Rev Bras Hematol HemoterTable 4, row 'LY', XN-9000, RT |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | RT | 48 hstable for the longest time tested | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 4, row 'LY', BC-6800, RT |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | RT | 48 hstable for the longest time tested | Bias% vs optimal bias (OP-Bias%, from biological variation; Table 1) | Buoro et al. 2016, Rev Bras Hematol HemoterTable 4, row 'LY', BC-6800, RT |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | RT | 8 hstable for the longest time tested | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 5, row 'LY', XN-9000, RT |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | RT | 8 hstable for the longest time tested | Bias% vs optimal bias (OP-Bias%, from biological variation; Table 1) | Buoro et al. 2016, Rev Bras Hematol HemoterTable 5, row 'LY', XN-9000, RT |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | RT | 8 hstable for the longest time tested | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 5, row 'LY', BC-6800, RT |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | RT | 8 hstable for the longest time tested | Bias% vs optimal bias (OP-Bias%, from biological variation; Table 1) | Buoro et al. 2016, Rev Bras Hematol HemoterTable 5, row 'LY', BC-6800, RT |
| whole-blood-edta, K2-EDTA (BD Vacutainer) | Storageuncentrifuged | RT | 8 hthen changed, changed by 12 h | Repeated-measures one-way ANOVA with Tukey post hoc vs 0 h, p < 0.05 (* in table) | Unalli et al. 2021, J Med BiochemTable 1, block 'LYM', row '12 h', column '23 °C' |
| whole-blood-edta, K2-EDTA (BD Vacutainer) | Storageuncentrifuged | 10 °C | 24 hthen changed, changed by 48 h | Repeated-measures one-way ANOVA with Tukey post hoc vs 0 h, p < 0.05 (* in table) | Unalli et al. 2021, J Med BiochemTable 1, block 'LYM', row '48 h', column '10 °C' |
Magnesium5 findings
| Specimen, tube | Condition | Temperature | Stable for | Criterion | Study |
|---|---|---|---|---|---|
| plasma-li-heparin, Sarstedt lithium heparin, no gel separator (primary tube) | Storagecentrifuged-on-cells | 2-8 °C | 48 hstable for the longest time tested | mean % difference vs time 0 exceeds total limit of change, TLC = ±√((1.65·CVa)² + (0.5·CVb)²) (EFLM BV), here 11.0%; stability limit from instability-equation regression | Illana et al. 2024, Biochem Med (Zagreb)Table 1, row 'Mg' |
| plasma-li-heparin, Sarstedt lithium heparin, no gel separator; plasma aliquot | Storageseparated | 2-8 °C | 96 hstable for the longest time tested | mean % difference vs time 0 exceeds total limit of change, TLC = ±√((1.65·CVa)² + (0.5·CVb)²) (EFLM BV), here 11.0%; stability limit from instability-equation regression | Illana et al. 2024, Biochem Med (Zagreb)Table 2, row 'Mg' |
| plasma-li-heparin, BD Vacutainer Barricor LH 3 mL, mechanical separator | Storagecentrifuged-on-gel | 2-8 °C | 75 hlimit stated by the authors | bias vs aliquoted plasma reference tube > maximal allowable bias 1.8% (Ricos et al.); time to MAB from linear regression of bias on time to analysis | Gawria et al. 2020, J Clin Lab AnalTable 2, row 'Magnesium', Barricor BD |
| plasma-li-heparin, BD Vacutainer PST LH 3 mL, gel | Storagecentrifuged-on-gel | 2-8 °C | 80 hstable for the longest time tested | bias vs aliquoted plasma reference tube > maximal allowable bias 1.8% (Ricos et al.); time to MAB from linear regression of bias on time to analysis | Gawria et al. 2020, J Clin Lab AnalTable 2, row 'Magnesium', Gel BD |
| plasma-li-heparin, Greiner Vacuette LH 3 mL Sep, gel | Storagecentrifuged-on-gel | 2-8 °C | 0 hchanged at the first time point, changed by 3 h | bias vs aliquoted plasma reference tube > maximal allowable bias 1.8% (Ricos et al.); time to MAB from linear regression of bias on time to analysis | Gawria et al. 2020, J Clin Lab AnalTable 2, row 'Magnesium', Gel Greiner |
Mean Corpuscular Hemoglobin21 findings
| Specimen, tube | Condition | Temperature | Stable for | Criterion | Study |
|---|---|---|---|---|---|
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 48 hstable for the longest time tested | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'MCH', XN-9000, 4 °C |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 48 hstable for the longest time tested | Bias% vs optimal bias (OP-Bias%, from biological variation; Table 1) | Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'MCH', XN-9000, 4 °C |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 48 hstable for the longest time tested | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'MCH', BC-6800, 4 °C |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 48 hstable for the longest time tested | Bias% vs optimal bias (OP-Bias%, from biological variation; Table 1) | Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'MCH', BC-6800, 4 °C |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 8 hstable for the longest time tested | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'MCH', XN-9000, 4 °C |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 8 hstable for the longest time tested | Bias% vs optimal bias (OP-Bias%, from biological variation; Table 1) | Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'MCH', XN-9000, 4 °C |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 8 hstable for the longest time tested | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'MCH', BC-6800, 4 °C |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 4 hthen changed, changed by 8 h | Bias% vs optimal bias (OP-Bias%, from biological variation; Table 1) | Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'MCH', BC-6800, 4 °C |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 8 hstable for the longest time tested | Bias% vs critical difference (CD%, reference change value; Table 1) | Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'MCH', BC-6800, 4 °C |
| whole-blood-edta, K2-EDTA (BD Vacutainer) | Storageuncentrifuged | 2-8 °C | 48 hstable for the longest time tested | Repeated-measures one-way ANOVA with Tukey post hoc vs 0 h, p < 0.05 (* in table) | Unalli et al. 2021, J Med BiochemTable 2, column '4 °C' (no * at any time point) |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | RT | 48 hstable for the longest time tested | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'MCH', XN-9000, RT |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | RT | 48 hstable for the longest time tested | Bias% vs optimal bias (OP-Bias%, from biological variation; Table 1) | Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'MCH', XN-9000, RT |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | RT | 48 hstable for the longest time tested | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'MCH', BC-6800, RT |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | RT | 48 hstable for the longest time tested | Bias% vs optimal bias (OP-Bias%, from biological variation; Table 1) | Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'MCH', BC-6800, RT |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | RT | 8 hstable for the longest time tested | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'MCH', XN-9000, RT |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | RT | 8 hstable for the longest time tested | Bias% vs optimal bias (OP-Bias%, from biological variation; Table 1) | Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'MCH', XN-9000, RT |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | RT | 8 hstable for the longest time tested | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'MCH', BC-6800, RT |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | RT | 8 hstable for the longest time tested | Bias% vs optimal bias (OP-Bias%, from biological variation; Table 1) | Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'MCH', BC-6800, RT |
| whole-blood-edta, K2-EDTA (BD Vacutainer) | Storageuncentrifuged | RT | 48 hstable for the longest time tested | Repeated-measures one-way ANOVA with Tukey post hoc vs 0 h, p < 0.05 (* in table) | Unalli et al. 2021, J Med BiochemTable 2, column '23 °C' (no * at any time point) |
| whole-blood-edta, K2EDTA (Becton Dickinson) | Storageuncentrifuged | 33 °C | 24 hthen changed, changed by 48 h | Unacceptable if median % difference > Biasmax (biological variation) and Holm-adjusted Wilcoxon P < 0.05; stable if equivalence shown (two one-tailed Wilcoxon tests vs Biasmax, Pholm < 0.05) | Jain et al. 2018, Biochem Med (Zagreb)Results, last paragraph (equivalence); Table 2, row 'MCH' |
| whole-blood-edta, K2-EDTA (BD Vacutainer) | Storageuncentrifuged | 10 °C | 48 hstable for the longest time tested | Repeated-measures one-way ANOVA with Tukey post hoc vs 0 h, p < 0.05 (* in table) | Unalli et al. 2021, J Med BiochemTable 2, column '10 °C' (no * at any time point) |
Mean Corpuscular Hemoglobin Concentration25 findings
| Specimen, tube | Condition | Temperature | Stable for | Criterion | Study |
|---|---|---|---|---|---|
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 0 hchanged at the first time point, changed by 2 h | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'MCHC', XN-9000, 4 °C |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 0 hchanged at the first time point, changed by 2 h | Bias% vs optimal bias (OP-Bias%, from biological variation; Table 1) | Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'MCHC', XN-9000, 4 °C |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 48 hstable for the longest time tested | Bias% vs critical difference (CD%, reference change value; Table 1) | Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'MCHC', XN-9000, 4 °C |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 36 hthen changed, changed by 48 h | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'MCHC', BC-6800, 4 °C |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 24 hthen changed, changed by 36 h | Bias% vs optimal bias (OP-Bias%, from biological variation; Table 1) | Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'MCHC', BC-6800, 4 °C |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 48 hstable for the longest time tested | Bias% vs critical difference (CD%, reference change value; Table 1) | Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'MCHC', BC-6800, 4 °C |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 8 hstable for the longest time tested | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'MCHC', XN-9000, 4 °C |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 8 hstable for the longest time tested | Bias% vs optimal bias (OP-Bias%, from biological variation; Table 1) | Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'MCHC', XN-9000, 4 °C |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 4 hthen changed, changed by 8 h | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'MCHC', BC-6800, 4 °C |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 4 hthen changed, changed by 8 h | Bias% vs optimal bias (OP-Bias%, from biological variation; Table 1) | Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'MCHC', BC-6800, 4 °C |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 8 hstable for the longest time tested | Bias% vs critical difference (CD%, reference change value; Table 1) | Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'MCHC', BC-6800, 4 °C |
| whole-blood-edta, K2-EDTA (BD Vacutainer) | Storageuncentrifuged | 2-8 °C | 48 hstable for the longest time tested | Repeated-measures one-way ANOVA with Tukey post hoc vs 0 h, p < 0.05 (* in table) | Unalli et al. 2021, J Med BiochemTable 2, column '4 °C' (no * at any time point) |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | RT | 2 hthen changed, changed by 4 h | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'MCHC', XN-9000, RT |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | RT | 0 hchanged at the first time point, changed by 2 h | Bias% vs optimal bias (OP-Bias%, from biological variation; Table 1) | Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'MCHC', XN-9000, RT |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | RT | 8 hthen changed, changed by 24 h | Bias% vs critical difference (CD%, reference change value; Table 1) | Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'MCHC', XN-9000, RT |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | RT | 8 hthen changed, changed by 24 h | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'MCHC', BC-6800, RT |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | RT | 8 hthen changed, changed by 24 h | Bias% vs optimal bias (OP-Bias%, from biological variation; Table 1) | Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'MCHC', BC-6800, RT |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | RT | 8 hthen changed, changed by 24 h | Bias% vs critical difference (CD%, reference change value; Table 1) | Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'MCHC', BC-6800, RT |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | RT | 8 hstable for the longest time tested | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'MCHC', XN-9000, RT |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | RT | 4 hthen changed, changed by 8 h | Bias% vs optimal bias (OP-Bias%, from biological variation; Table 1) | Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'MCHC', XN-9000, RT |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | RT | 8 hstable for the longest time tested | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'MCHC', BC-6800, RT |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | RT | 8 hstable for the longest time tested | Bias% vs optimal bias (OP-Bias%, from biological variation; Table 1) | Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'MCHC', BC-6800, RT |
| whole-blood-edta, K2-EDTA (BD Vacutainer) | Storageuncentrifuged | RT | 8 hthen changed, changed by 12 h | Repeated-measures one-way ANOVA with Tukey post hoc vs 0 h, p < 0.05 (* in table) | Unalli et al. 2021, J Med BiochemTable 2, block 'MCHC', row '12 h', column '23 °C' |
| whole-blood-edta, K2EDTA (Becton Dickinson) | Storageuncentrifuged | 33 °C | 0 hchanged at the first time point, changed by 4 h | Unacceptable if median % difference > Biasmax (biological variation) and Holm-adjusted Wilcoxon P < 0.05; stable if equivalence shown (two one-tailed Wilcoxon tests vs Biasmax, Pholm < 0.05) | Jain et al. 2018, Biochem Med (Zagreb)Results paragraph 1; Table 2, row 'MCHC' |
| whole-blood-edta, K2-EDTA (BD Vacutainer) | Storageuncentrifuged | 10 °C | 48 hstable for the longest time tested | Repeated-measures one-way ANOVA with Tukey post hoc vs 0 h, p < 0.05 (* in table) | Unalli et al. 2021, J Med BiochemTable 2, column '10 °C' (no * at any time point) |
Mean Corpuscular Volume25 findings
| Specimen, tube | Condition | Temperature | Stable for | Criterion | Study |
|---|---|---|---|---|---|
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 48 hstable for the longest time tested | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'MCV', XN-9000, 4 °C |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 24 hthen changed, changed by 36 h | Bias% vs optimal bias (OP-Bias%, from biological variation; Table 1) | Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'MCV', XN-9000, 4 °C |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 48 hstable for the longest time tested | Bias% vs critical difference (CD%, reference change value; Table 1) | Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'MCV', XN-9000, 4 °C |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 48 hstable for the longest time tested | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'MCV', BC-6800, 4 °C |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 36 hthen changed, changed by 48 h | Bias% vs optimal bias (OP-Bias%, from biological variation; Table 1) | Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'MCV', BC-6800, 4 °C |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 48 hstable for the longest time tested | Bias% vs critical difference (CD%, reference change value; Table 1) | Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'MCV', BC-6800, 4 °C |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 8 hstable for the longest time tested | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'MCV', XN-9000, 4 °C |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 8 hstable for the longest time tested | Bias% vs optimal bias (OP-Bias%, from biological variation; Table 1) | Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'MCV', XN-9000, 4 °C |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 8 hstable for the longest time tested | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'MCV', BC-6800, 4 °C |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 8 hstable for the longest time tested | Bias% vs optimal bias (OP-Bias%, from biological variation; Table 1) | Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'MCV', BC-6800, 4 °C |
| whole-blood-edta, K2-EDTA (BD Vacutainer) | Storageuncentrifuged | 2-8 °C | 48 hstable for the longest time tested | Repeated-measures one-way ANOVA with Tukey post hoc vs 0 h, p < 0.05 (* in table) | Unalli et al. 2021, J Med BiochemTable 2, column '4 °C' (no * at any time point) |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | RT | 8 hthen changed, changed by 24 h | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'MCV', XN-9000, RT |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | RT | 2 hthen changed, changed by 4 h | Bias% vs optimal bias (OP-Bias%, from biological variation; Table 1) | Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'MCV', XN-9000, RT |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | RT | 8 hthen changed, changed by 24 h | Bias% vs critical difference (CD%, reference change value; Table 1) | Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'MCV', XN-9000, RT |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | RT | 8 hthen changed, changed by 24 h | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'MCV', BC-6800, RT |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | RT | 8 hthen changed, changed by 24 h | Bias% vs optimal bias (OP-Bias%, from biological variation; Table 1) | Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'MCV', BC-6800, RT |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | RT | 8 hthen changed, changed by 24 h | Bias% vs critical difference (CD%, reference change value; Table 1) | Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'MCV', BC-6800, RT |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | RT | 8 hstable for the longest time tested | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'MCV', XN-9000, RT |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | RT | 8 hstable for the longest time tested | Bias% vs optimal bias (OP-Bias%, from biological variation; Table 1) | Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'MCV', XN-9000, RT |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | RT | 8 hstable for the longest time tested | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'MCV', BC-6800, RT |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | RT | 4 hthen changed, changed by 8 h | Bias% vs optimal bias (OP-Bias%, from biological variation; Table 1) | Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'MCV', BC-6800, RT |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | RT | 8 hstable for the longest time tested | Bias% vs critical difference (CD%, reference change value; Table 1) | Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'MCV', BC-6800, RT |
| whole-blood-edta, K2-EDTA (BD Vacutainer) | Storageuncentrifuged | RT | 24 hthen changed, changed by 48 h | Repeated-measures one-way ANOVA with Tukey post hoc vs 0 h, p < 0.05 (* in table) | Unalli et al. 2021, J Med BiochemTable 2, block 'MCV', row '48 h', column '23 °C' |
| whole-blood-edta, K2EDTA (Becton Dickinson) | Storageuncentrifuged | 33 °C | 0 hchanged at the first time point, changed by 4 h | Unacceptable if median % difference > Biasmax (biological variation) and Holm-adjusted Wilcoxon P < 0.05; stable if equivalence shown (two one-tailed Wilcoxon tests vs Biasmax, Pholm < 0.05) | Jain et al. 2018, Biochem Med (Zagreb)Results paragraph 1; Table 2, row 'MCV' |
| whole-blood-edta, K2-EDTA (BD Vacutainer) | Storageuncentrifuged | 10 °C | 48 hstable for the longest time tested | Repeated-measures one-way ANOVA with Tukey post hoc vs 0 h, p < 0.05 (* in table) | Unalli et al. 2021, J Med BiochemTable 2, column '10 °C' (no * at any time point) |
Mean Platelet Volume27 findings
| Specimen, tube | Condition | Temperature | Stable for | Criterion | Study |
|---|---|---|---|---|---|
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 36 hthen changed, changed by 48 h | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 4, row 'MPV', XN-9000, 4 °C |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 8 hthen changed, changed by 24 h | Bias% vs optimal bias (OP-Bias%, from biological variation; Table 1) | Buoro et al. 2016, Rev Bras Hematol HemoterTable 4, row 'MPV', XN-9000, 4 °C |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 48 hstable for the longest time tested | Bias% vs critical difference (CD%, reference change value; Table 1) | Buoro et al. 2016, Rev Bras Hematol HemoterTable 4, row 'MPV', XN-9000, 4 °C |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 8 hthen changed, changed by 24 h | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 4, row 'MPV', BC-6800, 4 °C |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 48 hstable for the longest time tested | Bias% vs optimal bias (OP-Bias%, from biological variation; Table 1) | Buoro et al. 2016, Rev Bras Hematol HemoterTable 4, row 'MPV', BC-6800, 4 °C |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 48 hstable for the longest time tested | Bias% vs critical difference (CD%, reference change value; Table 1) | Buoro et al. 2016, Rev Bras Hematol HemoterTable 4, row 'MPV', BC-6800, 4 °C |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 8 hstable for the longest time tested | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 5, row 'MPV', XN-9000, 4 °C |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 4 hthen changed, changed by 8 h | Bias% vs optimal bias (OP-Bias%, from biological variation; Table 1) | Buoro et al. 2016, Rev Bras Hematol HemoterTable 5, row 'MPV', XN-9000, 4 °C |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 8 hstable for the longest time tested | Bias% vs critical difference (CD%, reference change value; Table 1) | Buoro et al. 2016, Rev Bras Hematol HemoterTable 5, row 'MPV', XN-9000, 4 °C |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 8 hstable for the longest time tested | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 5, row 'MPV', BC-6800, 4 °C |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 0 hchanged at the first time point, changed by 4 h | Bias% vs optimal bias (OP-Bias%, from biological variation; Table 1) | Buoro et al. 2016, Rev Bras Hematol HemoterTable 5, row 'MPV', BC-6800, 4 °C |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 8 hstable for the longest time tested | Bias% vs critical difference (CD%, reference change value; Table 1) | Buoro et al. 2016, Rev Bras Hematol HemoterTable 5, row 'MPV', BC-6800, 4 °C |
| whole-blood-edta, K2-EDTA (BD Vacutainer) | Storageuncentrifuged | 2-8 °C | 16 hthen changed, changed by 20 h | Repeated-measures one-way ANOVA with Tukey post hoc vs 0 h, p < 0.05 (* in table) | Unalli et al. 2021, J Med BiochemTable 2, block 'MPV', row '20 h', column '4 °C' |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | RT | 8 hthen changed, changed by 24 h | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 4, row 'MPV', XN-9000, RT |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | RT | 0 hchanged at the first time point, changed by 2 h | Bias% vs optimal bias (OP-Bias%, from biological variation; Table 1) | Buoro et al. 2016, Rev Bras Hematol HemoterTable 4, row 'MPV', XN-9000, RT |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | RT | 48 hstable for the longest time tested | Bias% vs critical difference (CD%, reference change value; Table 1) | Buoro et al. 2016, Rev Bras Hematol HemoterTable 4, row 'MPV', XN-9000, RT |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | RT | 2 hthen changed, changed by 4 h | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 4, row 'MPV', BC-6800, RT |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | RT | 0 hchanged at the first time point, changed by 2 h | Bias% vs optimal bias (OP-Bias%, from biological variation; Table 1) | Buoro et al. 2016, Rev Bras Hematol HemoterTable 4, row 'MPV', BC-6800, RT |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | RT | 24 hthen changed, changed by 36 h | Bias% vs critical difference (CD%, reference change value; Table 1) | Buoro et al. 2016, Rev Bras Hematol HemoterTable 4, row 'MPV', BC-6800, RT |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | RT | 8 hstable for the longest time tested | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 5, row 'MPV', XN-9000, RT |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | RT | 8 hstable for the longest time tested | Bias% vs optimal bias (OP-Bias%, from biological variation; Table 1) | Buoro et al. 2016, Rev Bras Hematol HemoterTable 5, row 'MPV', XN-9000, RT |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | RT | 8 hstable for the longest time tested | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 5, row 'MPV', BC-6800, RT |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | RT | 0 hchanged at the first time point, changed by 4 h | Bias% vs optimal bias (OP-Bias%, from biological variation; Table 1) | Buoro et al. 2016, Rev Bras Hematol HemoterTable 5, row 'MPV', BC-6800, RT |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | RT | 8 hstable for the longest time tested | Bias% vs critical difference (CD%, reference change value; Table 1) | Buoro et al. 2016, Rev Bras Hematol HemoterTable 5, row 'MPV', BC-6800, RT |
| whole-blood-edta, K2-EDTA (BD Vacutainer) | Storageuncentrifuged | RT | 20 hthen changed, changed by 24 h | Repeated-measures one-way ANOVA with Tukey post hoc vs 0 h, p < 0.05 (* in table) | Unalli et al. 2021, J Med BiochemTable 2, block 'MPV', row '24 h', column '23 °C' |
| whole-blood-edta, K2EDTA (Becton Dickinson) | Storageuncentrifuged | 33 °C | 0 hchanged at the first time point, changed by 4 h | Unacceptable if median % difference > Biasmax (biological variation) and Holm-adjusted Wilcoxon P < 0.05; stable if equivalence shown (two one-tailed Wilcoxon tests vs Biasmax, Pholm < 0.05) | Jain et al. 2018, Biochem Med (Zagreb)Results paragraph 1; Table 2, row 'MPV' |
| whole-blood-edta, K2-EDTA (BD Vacutainer) | Storageuncentrifuged | 10 °C | 16 hthen changed, changed by 20 h | Repeated-measures one-way ANOVA with Tukey post hoc vs 0 h, p < 0.05 (* in table) | Unalli et al. 2021, J Med BiochemTable 2, block 'MPV', row '20 h', column '10 °C' |
Medium-fluorescence reticulocyte ratio (MFR)8 findings
| Specimen, tube | Condition | Temperature | Stable for | Criterion | Study |
|---|---|---|---|---|---|
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 48 hstable for the longest time tested | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'MFR', XN-9000, 4 °C |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 48 hstable for the longest time tested | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'MFR', BC-6800, 4 °C |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 8 hstable for the longest time tested | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'MFR', XN-9000, 4 °C |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 8 hstable for the longest time tested | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'MFR', BC-6800, 4 °C |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | RT | 8 hthen changed, changed by 24 h | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'MFR', XN-9000, RT |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | RT | 24 hthen changed, changed by 36 h | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'MFR', BC-6800, RT |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | RT | 8 hstable for the longest time tested | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'MFR', XN-9000, RT |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | RT | 8 hstable for the longest time tested | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'MFR', BC-6800, RT |
MO-WX5 findings
| Specimen, tube | Condition | Temperature | Stable for | Criterion | Study |
|---|---|---|---|---|---|
| whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mL | Storageuncentrifuged | 2-8 °C | 24 hstable for the longest time tested | Approach 2: absolute % difference between medians (Tx vs T0) must be < 10% | Roccaforte et al. 2026, Biochem Med (Zagreb)Table 1, row 'MO-WX', 4 °C, column 'Approach 2 (% difference between medians)' |
| whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mL | Storageuncentrifuged | 2-8 °C | 24 hstable for the longest time tested | EFLM regression: time at which fitted % difference line reaches MPE 10% | Roccaforte et al. 2026, Biochem Med (Zagreb)Table 3, row 'MO-WX', 4 °C |
| whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mL | Storageuncentrifuged | RT | 24 hstable for the longest time tested | Approach 2: absolute % difference between medians (Tx vs T0) must be < 10% | Roccaforte et al. 2026, Biochem Med (Zagreb)Table 1, row 'MO-WX', RT, column 'Approach 2 (% difference between medians)' |
| whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mL | Storageuncentrifuged | RT | 24 hstable for the longest time tested | EFLM regression: time at which fitted % difference line reaches MPE 10% | Roccaforte et al. 2026, Biochem Med (Zagreb)Table 3, row 'MO-WX', RT |
| whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mL | Storageuncentrifuged | 37 °C | 4 hlimit stated by the authors | EFLM regression: time at which fitted % difference line reaches MPE 10% | Roccaforte et al. 2026, Biochem Med (Zagreb)Table 3, row 'MO-WX', 37 °C |
MO-WY4 findings
| Specimen, tube | Condition | Temperature | Stable for | Criterion | Study |
|---|---|---|---|---|---|
| whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mL | Storageuncentrifuged | 2-8 °C | 24 hstable for the longest time tested | EFLM regression: time at which fitted % difference line reaches MPE 10% | Roccaforte et al. 2026, Biochem Med (Zagreb)Table 3, row 'MO-WY', 4 °C |
| whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mL | Storageuncentrifuged | RT | 12 hthen changed, changed by 24 h | Approach 2: absolute % difference between medians (Tx vs T0) must be < 10% | Roccaforte et al. 2026, Biochem Med (Zagreb)Table 2, row 'MO-WY', RT (blank in Table 1), column 'Approach 2 (% difference between medians)' |
| whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mL | Storageuncentrifuged | RT | 17 hlimit stated by the authors | EFLM regression: time at which fitted % difference line reaches MPE 10% | Roccaforte et al. 2026, Biochem Med (Zagreb)Table 3, row 'MO-WY', RT |
| whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mL | Storageuncentrifuged | 37 °C | 5 hlimit stated by the authors | EFLM regression: time at which fitted % difference line reaches MPE 10% | Roccaforte et al. 2026, Biochem Med (Zagreb)Table 3, row 'MO-WY', 37 °C |
MO-WZ5 findings
| Specimen, tube | Condition | Temperature | Stable for | Criterion | Study |
|---|---|---|---|---|---|
| whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mL | Storageuncentrifuged | 2-8 °C | 24 hstable for the longest time tested | Approach 2: absolute % difference between medians (Tx vs T0) must be < 10% | Roccaforte et al. 2026, Biochem Med (Zagreb)Table 1, row 'MO-WZ', 4 °C, column 'Approach 2 (% difference between medians)' |
| whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mL | Storageuncentrifuged | 2-8 °C | 24 hstable for the longest time tested | EFLM regression: time at which fitted % difference line reaches MPE 10% | Roccaforte et al. 2026, Biochem Med (Zagreb)Table 3, row 'MO-WZ', 4 °C |
| whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mL | Storageuncentrifuged | RT | 24 hstable for the longest time tested | Approach 2: absolute % difference between medians (Tx vs T0) must be < 10% | Roccaforte et al. 2026, Biochem Med (Zagreb)Table 1, row 'MO-WZ', RT, column 'Approach 2 (% difference between medians)' |
| whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mL | Storageuncentrifuged | RT | 24 hstable for the longest time tested | EFLM regression: time at which fitted % difference line reaches MPE 10% | Roccaforte et al. 2026, Biochem Med (Zagreb)Table 3, row 'MO-WZ', RT |
| whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mL | Storageuncentrifuged | 37 °C | 8 hlimit stated by the authors | EFLM regression: time at which fitted % difference line reaches MPE 10% | Roccaforte et al. 2026, Biochem Med (Zagreb)Table 3, row 'MO-WZ', 37 °C |
MO-X9 findings
| Specimen, tube | Condition | Temperature | Stable for | Criterion | Study |
|---|---|---|---|---|---|
| whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mL | Storageuncentrifuged | 2-8 °C | 24 hstable for the longest time tested | Approach 2: absolute % difference between medians (Tx vs T0) must be < 10% | Roccaforte et al. 2026, Biochem Med (Zagreb)Table 1, row 'MO-X', 4 °C, column 'Approach 2 (% difference between medians)' |
| whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mL | Storageuncentrifuged | 2-8 °C | 24 hstable for the longest time tested | Approach 3: fewer than 2 of 25 samples with absolute % change from baseline > 10% | Roccaforte et al. 2026, Biochem Med (Zagreb)Table 1, row 'MO-X', 4 °C, column 'Approach 3 (per-sample % difference)' |
| whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mL | Storageuncentrifuged | 2-8 °C | 24 hstable for the longest time tested | EFLM regression: time at which fitted % difference line reaches MPE 10% | Roccaforte et al. 2026, Biochem Med (Zagreb)Table 3, row 'MO-X', 4 °C |
| whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mL | Storageuncentrifuged | RT | 24 hstable for the longest time tested | Approach 2: absolute % difference between medians (Tx vs T0) must be < 10% | Roccaforte et al. 2026, Biochem Med (Zagreb)Table 1, row 'MO-X', RT, column 'Approach 2 (% difference between medians)' |
| whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mL | Storageuncentrifuged | RT | 24 hstable for the longest time tested | Approach 3: fewer than 2 of 25 samples with absolute % change from baseline > 10% | Roccaforte et al. 2026, Biochem Med (Zagreb)Table 1, row 'MO-X', RT, column 'Approach 3 (per-sample % difference)' |
| whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mL | Storageuncentrifuged | RT | 24 hstable for the longest time tested | EFLM regression: time at which fitted % difference line reaches MPE 10% | Roccaforte et al. 2026, Biochem Med (Zagreb)Table 3, row 'MO-X', RT |
| whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mL | Storageuncentrifuged | 37 °C | 24 hstable for the longest time tested | Approach 2: absolute % difference between medians (Tx vs T0) must be < 10% | Roccaforte et al. 2026, Biochem Med (Zagreb)Table 1, row 'MO-X', 37 °C, column 'Approach 2 (% difference between medians)' |
| whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mL | Storageuncentrifuged | 37 °C | 24 hstable for the longest time tested | Approach 3: fewer than 2 of 25 samples with absolute % change from baseline > 10% | Roccaforte et al. 2026, Biochem Med (Zagreb)Table 1, row 'MO-X', 37 °C, column 'Approach 3 (per-sample % difference)' |
| whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mL | Storageuncentrifuged | 37 °C | 24 hstable for the longest time tested | EFLM regression: time at which fitted % difference line reaches MPE 10% | Roccaforte et al. 2026, Biochem Med (Zagreb)Table 3, row 'MO-X', 37 °C |
MO-Y6 findings
| Specimen, tube | Condition | Temperature | Stable for | Criterion | Study |
|---|---|---|---|---|---|
| whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mL | Storageuncentrifuged | 2-8 °C | 24 hstable for the longest time tested | Approach 2: absolute % difference between medians (Tx vs T0) must be < 10% | Roccaforte et al. 2026, Biochem Med (Zagreb)Table 1, row 'MO-Y', 4 °C, column 'Approach 2 (% difference between medians)' |
| whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mL | Storageuncentrifuged | 2-8 °C | 24 hstable for the longest time tested | EFLM regression: time at which fitted % difference line reaches MPE 10% | Roccaforte et al. 2026, Biochem Med (Zagreb)Table 3, row 'MO-Y', 4 °C |
| whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mL | Storageuncentrifuged | RT | 24 hstable for the longest time tested | Approach 2: absolute % difference between medians (Tx vs T0) must be < 10% | Roccaforte et al. 2026, Biochem Med (Zagreb)Table 1, row 'MO-Y', RT, column 'Approach 2 (% difference between medians)' |
| whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mL | Storageuncentrifuged | RT | 24 hstable for the longest time tested | EFLM regression: time at which fitted % difference line reaches MPE 10% | Roccaforte et al. 2026, Biochem Med (Zagreb)Table 3, row 'MO-Y', RT |
| whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mL | Storageuncentrifuged | 37 °C | 12 hthen changed, changed by 24 h | Approach 2: absolute % difference between medians (Tx vs T0) must be < 10% | Roccaforte et al. 2026, Biochem Med (Zagreb)Table 2, row 'MO-Y', 37 °C (blank in Table 1), column 'Approach 2 (% difference between medians)' |
| whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mL | Storageuncentrifuged | 37 °C | 24 hstable for the longest time tested | EFLM regression: time at which fitted % difference line reaches MPE 10% | Roccaforte et al. 2026, Biochem Med (Zagreb)Table 3, row 'MO-Y', 37 °C |
MO-Z5 findings
| Specimen, tube | Condition | Temperature | Stable for | Criterion | Study |
|---|---|---|---|---|---|
| whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mL | Storageuncentrifuged | 2-8 °C | 15 hlimit stated by the authors | EFLM regression: time at which fitted % difference line reaches MPE 10% | Roccaforte et al. 2026, Biochem Med (Zagreb)Table 3, row 'MO-Z', 4 °C |
| whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mL | Storageuncentrifuged | RT | 24 hstable for the longest time tested | Approach 2: absolute % difference between medians (Tx vs T0) must be < 10% | Roccaforte et al. 2026, Biochem Med (Zagreb)Table 1, row 'MO-Z', RT, column 'Approach 2 (% difference between medians)' |
| whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mL | Storageuncentrifuged | RT | 24 hstable for the longest time tested | EFLM regression: time at which fitted % difference line reaches MPE 10% | Roccaforte et al. 2026, Biochem Med (Zagreb)Table 3, row 'MO-Z', RT |
| whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mL | Storageuncentrifuged | 37 °C | 12 hthen changed, changed by 24 h | Approach 2: absolute % difference between medians (Tx vs T0) must be < 10% | Roccaforte et al. 2026, Biochem Med (Zagreb)Table 2, row 'MO-Z', 37 °C (blank in Table 1), column 'Approach 2 (% difference between medians)' |
| whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mL | Storageuncentrifuged | 37 °C | 11 hlimit stated by the authors | EFLM regression: time at which fitted % difference line reaches MPE 10% | Roccaforte et al. 2026, Biochem Med (Zagreb)Table 3, row 'MO-Z', 37 °C |
Monocytes (Absolute)23 findings
| Specimen, tube | Condition | Temperature | Stable for | Criterion | Study |
|---|---|---|---|---|---|
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 48 hstable for the longest time tested | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 4, row 'MO', XN-9000, 4 °C |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 24 hthen changed, changed by 36 h | Bias% vs optimal bias (OP-Bias%, from biological variation; Table 1) | Buoro et al. 2016, Rev Bras Hematol HemoterTable 4, row 'MO', XN-9000, 4 °C |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 48 hstable for the longest time tested | Bias% vs critical difference (CD%, reference change value; Table 1) | Buoro et al. 2016, Rev Bras Hematol HemoterTable 4, row 'MO', XN-9000, 4 °C |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 36 hthen changed, changed by 48 h | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 4, row 'MO', BC-6800, 4 °C |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 24 hthen changed, changed by 36 h | Bias% vs optimal bias (OP-Bias%, from biological variation; Table 1) | Buoro et al. 2016, Rev Bras Hematol HemoterTable 4, row 'MO', BC-6800, 4 °C |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 48 hstable for the longest time tested | Bias% vs critical difference (CD%, reference change value; Table 1) | Buoro et al. 2016, Rev Bras Hematol HemoterTable 4, row 'MO', BC-6800, 4 °C |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 8 hstable for the longest time tested | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 5, row 'MO', XN-9000, 4 °C |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 8 hstable for the longest time tested | Bias% vs optimal bias (OP-Bias%, from biological variation; Table 1) | Buoro et al. 2016, Rev Bras Hematol HemoterTable 5, row 'MO', XN-9000, 4 °C |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 8 hstable for the longest time tested | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 5, row 'MO', BC-6800, 4 °C |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 4 hthen changed, changed by 8 h | Bias% vs optimal bias (OP-Bias%, from biological variation; Table 1) | Buoro et al. 2016, Rev Bras Hematol HemoterTable 5, row 'MO', BC-6800, 4 °C |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 8 hstable for the longest time tested | Bias% vs critical difference (CD%, reference change value; Table 1) | Buoro et al. 2016, Rev Bras Hematol HemoterTable 5, row 'MO', BC-6800, 4 °C |
| whole-blood-edta, K2-EDTA (BD Vacutainer) | Storageuncentrifuged | 2-8 °C | 48 hstable for the longest time tested | Repeated-measures one-way ANOVA with Tukey post hoc vs 0 h, p < 0.05 (* in table) | Unalli et al. 2021, J Med BiochemTable 1, column '4 °C' (no * at any time point) |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | RT | 48 hstable for the longest time tested | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 4, row 'MO', XN-9000, RT |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | RT | 48 hstable for the longest time tested | Bias% vs optimal bias (OP-Bias%, from biological variation; Table 1) | Buoro et al. 2016, Rev Bras Hematol HemoterTable 4, row 'MO', XN-9000, RT |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | RT | 36 hthen changed, changed by 48 h | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 4, row 'MO', BC-6800, RT |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | RT | 36 hthen changed, changed by 48 h | Bias% vs optimal bias (OP-Bias%, from biological variation; Table 1) | Buoro et al. 2016, Rev Bras Hematol HemoterTable 4, row 'MO', BC-6800, RT |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | RT | 48 hstable for the longest time tested | Bias% vs critical difference (CD%, reference change value; Table 1) | Buoro et al. 2016, Rev Bras Hematol HemoterTable 4, row 'MO', BC-6800, RT |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | RT | 8 hstable for the longest time tested | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 5, row 'MO', XN-9000, RT |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | RT | 8 hstable for the longest time tested | Bias% vs optimal bias (OP-Bias%, from biological variation; Table 1) | Buoro et al. 2016, Rev Bras Hematol HemoterTable 5, row 'MO', XN-9000, RT |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | RT | 8 hstable for the longest time tested | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 5, row 'MO', BC-6800, RT |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | RT | 8 hstable for the longest time tested | Bias% vs optimal bias (OP-Bias%, from biological variation; Table 1) | Buoro et al. 2016, Rev Bras Hematol HemoterTable 5, row 'MO', BC-6800, RT |
| whole-blood-edta, K2-EDTA (BD Vacutainer) | Storageuncentrifuged | RT | 24 hthen changed, changed by 48 h | Repeated-measures one-way ANOVA with Tukey post hoc vs 0 h, p < 0.05 (* in table) | Unalli et al. 2021, J Med BiochemTable 1, block 'MONO', row '48 h', column '23 °C' |
| whole-blood-edta, K2-EDTA (BD Vacutainer) | Storageuncentrifuged | 10 °C | 48 hstable for the longest time tested | Repeated-measures one-way ANOVA with Tukey post hoc vs 0 h, p < 0.05 (* in table) | Unalli et al. 2021, J Med BiochemTable 1, column '10 °C' (no * at any time point) |
Myoglobin4 findings
| Specimen, tube | Condition | Temperature | Stable for | Criterion | Study |
|---|---|---|---|---|---|
| plasma-li-heparin, Sarstedt lithium heparin, no gel separator (primary tube) | Storagecentrifuged-on-cells | 2-8 °C | 48 hstable for the longest time tested | mean % difference vs time 0 exceeds total limit of change, TLC = ±√((1.65·CVa)² + (0.5·CVb)²) (EFLM BV), here 24.9%; stability limit from instability-equation regression | Illana et al. 2024, Biochem Med (Zagreb)Table 1, row 'Myoglobin' |
| plasma-li-heparin, Sarstedt lithium heparin, no gel separator; plasma aliquot | Storageseparated | 2-8 °C | 96 hstable for the longest time tested | mean % difference vs time 0 exceeds total limit of change, TLC = ±√((1.65·CVa)² + (0.5·CVb)²) (EFLM BV), here 24.9%; stability limit from instability-equation regression | Illana et al. 2024, Biochem Med (Zagreb)Table 2, row 'Myoglobin' |
| plasma-li-heparin, BD Barricor 4.5 mL (lithium heparin, mechanical separator) | Storagecentrifuged-on-gel | 2-8 °C | 7 daysstable for the longest time tested | mean % difference vs time 0 exceeds desirable imprecision based on biological variation (EFLM/Westgard), here 7.0% | Mandić et al. 2020, Biochem Med (Zagreb)Tables 3 and 4, row 'Myoglobin, μg/L', BD Barricor columns |
| serum, BD Vacutainer RST 5 mL (thrombin clot activator, gel) | Storagecentrifuged-on-gel | 2-8 °C | 7 daysstable for the longest time tested | mean % difference vs time 0 exceeds desirable imprecision based on biological variation (EFLM/Westgard), here 7.0% | Mandić et al. 2020, Biochem Med (Zagreb)Tables 3 and 4, row 'Myoglobin, μg/L', BD RST columns |
N-Terminal Pro-BNP2 findings
| Specimen, tube | Condition | Temperature | Stable for | Criterion | Study |
|---|---|---|---|---|---|
| plasma-li-heparin, BD Vacutainer lithium heparin 6 mL (regular) | Storagenot-stated | 2-8 °C | 72 hstable for the longest time tested | mean % difference vs T0 exceeds total change limit (TCL), here ±18.35% | Dupuy et al. 2018, Biochem Med (Zagreb)Table 2, BD lithium heparin tubes, row 'NT-proBNP, ng/L' |
| plasma-li-heparin, BD Barricor lithium heparin 5.5 mL (mechanical separator) | Storagecentrifuged-on-gel | 2-8 °C | 72 hstable for the longest time tested | mean % difference vs T0 exceeds total change limit (TCL), here ±18.35% | Dupuy et al. 2018, Biochem Med (Zagreb)Table 2, BD Barricor lithium heparin tubes, row 'NT-proBNP, ng/L' |
NE-FSC7 findings
| Specimen, tube | Condition | Temperature | Stable for | Criterion | Study |
|---|---|---|---|---|---|
| whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mL | Storageuncentrifuged | 2-8 °C | 24 hstable for the longest time tested | EFLM regression: time at which fitted % difference line reaches MPE 10% | Roccaforte et al. 2026, Biochem Med (Zagreb)Table 3, row 'NE-FSC', 4 °C |
| whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mL | Storageuncentrifuged | RT | 24 hstable for the longest time tested | Approach 2: absolute % difference between medians (Tx vs T0) must be < 10% | Roccaforte et al. 2026, Biochem Med (Zagreb)Table 1, row 'NE-FSC', RT, column 'Approach 2 (% difference between medians)' |
| whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mL | Storageuncentrifuged | RT | 24 hstable for the longest time tested | Approach 3: fewer than 2 of 25 samples with absolute % change from baseline > 10% | Roccaforte et al. 2026, Biochem Med (Zagreb)Table 1, row 'NE-FSC', RT, column 'Approach 3 (per-sample % difference)' |
| whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mL | Storageuncentrifuged | RT | 24 hstable for the longest time tested | EFLM regression: time at which fitted % difference line reaches MPE 10% | Roccaforte et al. 2026, Biochem Med (Zagreb)Table 3, row 'NE-FSC', RT |
| whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mL | Storageuncentrifuged | 37 °C | 24 hstable for the longest time tested | Approach 2: absolute % difference between medians (Tx vs T0) must be < 10% | Roccaforte et al. 2026, Biochem Med (Zagreb)Table 1, row 'NE-FSC', 37 °C, column 'Approach 2 (% difference between medians)' |
| whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mL | Storageuncentrifuged | 37 °C | 24 hstable for the longest time tested | Approach 3: fewer than 2 of 25 samples with absolute % change from baseline > 10% | Roccaforte et al. 2026, Biochem Med (Zagreb)Table 1, row 'NE-FSC', 37 °C, column 'Approach 3 (per-sample % difference)' |
| whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mL | Storageuncentrifuged | 37 °C | 24 hstable for the longest time tested | EFLM regression: time at which fitted % difference line reaches MPE 10% | Roccaforte et al. 2026, Biochem Med (Zagreb)Table 3, row 'NE-FSC', 37 °C |
NE-SFL6 findings
| Specimen, tube | Condition | Temperature | Stable for | Criterion | Study |
|---|---|---|---|---|---|
| whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mL | Storageuncentrifuged | 2-8 °C | 24 hstable for the longest time tested | Approach 2: absolute % difference between medians (Tx vs T0) must be < 10% | Roccaforte et al. 2026, Biochem Med (Zagreb)Table 1, row 'NE-SFL', 4 °C, column 'Approach 2 (% difference between medians)' |
| whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mL | Storageuncentrifuged | 2-8 °C | 19 hlimit stated by the authors | EFLM regression: time at which fitted % difference line reaches MPE 10% | Roccaforte et al. 2026, Biochem Med (Zagreb)Table 3, row 'NE-SFL', 4 °C |
| whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mL | Storageuncentrifuged | RT | 24 hstable for the longest time tested | Approach 2: absolute % difference between medians (Tx vs T0) must be < 10% | Roccaforte et al. 2026, Biochem Med (Zagreb)Table 1, row 'NE-SFL', RT, column 'Approach 2 (% difference between medians)' |
| whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mL | Storageuncentrifuged | RT | 24 hstable for the longest time tested | EFLM regression: time at which fitted % difference line reaches MPE 10% | Roccaforte et al. 2026, Biochem Med (Zagreb)Table 3, row 'NE-SFL', RT |
| whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mL | Storageuncentrifuged | 37 °C | 12 hthen changed, changed by 24 h | Approach 2: absolute % difference between medians (Tx vs T0) must be < 10% | Roccaforte et al. 2026, Biochem Med (Zagreb)Table 2, row 'NE-SFL', 37 °C (blank in Table 1), column 'Approach 2 (% difference between medians)' |
| whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mL | Storageuncentrifuged | 37 °C | 19 hlimit stated by the authors | EFLM regression: time at which fitted % difference line reaches MPE 10% | Roccaforte et al. 2026, Biochem Med (Zagreb)Table 3, row 'NE-SFL', 37 °C |
NE-SSC9 findings
| Specimen, tube | Condition | Temperature | Stable for | Criterion | Study |
|---|---|---|---|---|---|
| whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mL | Storageuncentrifuged | 2-8 °C | 24 hstable for the longest time tested | Approach 2: absolute % difference between medians (Tx vs T0) must be < 10% | Roccaforte et al. 2026, Biochem Med (Zagreb)Table 1, row 'NE-SSC', 4 °C, column 'Approach 2 (% difference between medians)' |
| whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mL | Storageuncentrifuged | 2-8 °C | 24 hstable for the longest time tested | Approach 3: fewer than 2 of 25 samples with absolute % change from baseline > 10% | Roccaforte et al. 2026, Biochem Med (Zagreb)Table 1, row 'NE-SSC', 4 °C, column 'Approach 3 (per-sample % difference)' |
| whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mL | Storageuncentrifuged | 2-8 °C | 24 hstable for the longest time tested | EFLM regression: time at which fitted % difference line reaches MPE 10% | Roccaforte et al. 2026, Biochem Med (Zagreb)Table 3, row 'NE-SSC', 4 °C |
| whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mL | Storageuncentrifuged | RT | 24 hstable for the longest time tested | Approach 2: absolute % difference between medians (Tx vs T0) must be < 10% | Roccaforte et al. 2026, Biochem Med (Zagreb)Table 1, row 'NE-SSC', RT, column 'Approach 2 (% difference between medians)' |
| whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mL | Storageuncentrifuged | RT | 24 hstable for the longest time tested | Approach 3: fewer than 2 of 25 samples with absolute % change from baseline > 10% | Roccaforte et al. 2026, Biochem Med (Zagreb)Table 1, row 'NE-SSC', RT, column 'Approach 3 (per-sample % difference)' |
| whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mL | Storageuncentrifuged | RT | 24 hstable for the longest time tested | EFLM regression: time at which fitted % difference line reaches MPE 10% | Roccaforte et al. 2026, Biochem Med (Zagreb)Table 3, row 'NE-SSC', RT |
| whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mL | Storageuncentrifuged | 37 °C | 24 hstable for the longest time tested | Approach 2: absolute % difference between medians (Tx vs T0) must be < 10% | Roccaforte et al. 2026, Biochem Med (Zagreb)Table 1, row 'NE-SSC', 37 °C, column 'Approach 2 (% difference between medians)' |
| whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mL | Storageuncentrifuged | 37 °C | 24 hstable for the longest time tested | Approach 3: fewer than 2 of 25 samples with absolute % change from baseline > 10% | Roccaforte et al. 2026, Biochem Med (Zagreb)Table 1, row 'NE-SSC', 37 °C, column 'Approach 3 (per-sample % difference)' |
| whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mL | Storageuncentrifuged | 37 °C | 24 hstable for the longest time tested | EFLM regression: time at which fitted % difference line reaches MPE 10% | Roccaforte et al. 2026, Biochem Med (Zagreb)Table 3, row 'NE-SSC', 37 °C |
NE-WX6 findings
| Specimen, tube | Condition | Temperature | Stable for | Criterion | Study |
|---|---|---|---|---|---|
| whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mL | Storageuncentrifuged | 2-8 °C | 24 hstable for the longest time tested | Approach 2: absolute % difference between medians (Tx vs T0) must be < 10% | Roccaforte et al. 2026, Biochem Med (Zagreb)Table 1, row 'NE-WX', 4 °C, column 'Approach 2 (% difference between medians)' |
| whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mL | Storageuncentrifuged | 2-8 °C | 23 hlimit stated by the authors | EFLM regression: time at which fitted % difference line reaches MPE 10% | Roccaforte et al. 2026, Biochem Med (Zagreb)Table 3, row 'NE-WX', 4 °C |
| whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mL | Storageuncentrifuged | RT | 12 hthen changed, changed by 24 h | Approach 2: absolute % difference between medians (Tx vs T0) must be < 10% | Roccaforte et al. 2026, Biochem Med (Zagreb)Table 2, row 'NE-WX', RT (blank in Table 1), column 'Approach 2 (% difference between medians)' |
| whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mL | Storageuncentrifuged | RT | 22 hlimit stated by the authors | EFLM regression: time at which fitted % difference line reaches MPE 10% | Roccaforte et al. 2026, Biochem Med (Zagreb)Table 3, row 'NE-WX', RT |
| whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mL | Storageuncentrifuged | 37 °C | 12 hthen changed, changed by 24 h | Approach 2: absolute % difference between medians (Tx vs T0) must be < 10% | Roccaforte et al. 2026, Biochem Med (Zagreb)Table 2, row 'NE-WX', 37 °C (blank in Table 1), column 'Approach 2 (% difference between medians)' |
| whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mL | Storageuncentrifuged | 37 °C | 11 hlimit stated by the authors | EFLM regression: time at which fitted % difference line reaches MPE 10% | Roccaforte et al. 2026, Biochem Med (Zagreb)Table 3, row 'NE-WX', 37 °C |
NE-WY6 findings
| Specimen, tube | Condition | Temperature | Stable for | Criterion | Study |
|---|---|---|---|---|---|
| whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mL | Storageuncentrifuged | 2-8 °C | 24 hstable for the longest time tested | Approach 2: absolute % difference between medians (Tx vs T0) must be < 10% | Roccaforte et al. 2026, Biochem Med (Zagreb)Table 1, row 'NE-WY', 4 °C, column 'Approach 2 (% difference between medians)' |
| whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mL | Storageuncentrifuged | 2-8 °C | 24 hstable for the longest time tested | EFLM regression: time at which fitted % difference line reaches MPE 10% | Roccaforte et al. 2026, Biochem Med (Zagreb)Table 3, row 'NEWY', 4 °C |
| whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mL | Storageuncentrifuged | RT | 24 hstable for the longest time tested | Approach 2: absolute % difference between medians (Tx vs T0) must be < 10% | Roccaforte et al. 2026, Biochem Med (Zagreb)Table 1, row 'NE-WY', RT, column 'Approach 2 (% difference between medians)' |
| whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mL | Storageuncentrifuged | RT | 24 hstable for the longest time tested | EFLM regression: time at which fitted % difference line reaches MPE 10% | Roccaforte et al. 2026, Biochem Med (Zagreb)Table 3, row 'NEWY', RT |
| whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mL | Storageuncentrifuged | 37 °C | 12 hthen changed, changed by 24 h | Approach 2: absolute % difference between medians (Tx vs T0) must be < 10% | Roccaforte et al. 2026, Biochem Med (Zagreb)Table 2, row 'NE-WY', 37 °C (blank in Table 1), column 'Approach 2 (% difference between medians)' |
| whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mL | Storageuncentrifuged | 37 °C | 5 hlimit stated by the authors | EFLM regression: time at which fitted % difference line reaches MPE 10% | Roccaforte et al. 2026, Biochem Med (Zagreb)Table 3, row 'NEWY', 37 °C |
NE-WZ5 findings
| Specimen, tube | Condition | Temperature | Stable for | Criterion | Study |
|---|---|---|---|---|---|
| whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mL | Storageuncentrifuged | 2-8 °C | 24 hstable for the longest time tested | Approach 2: absolute % difference between medians (Tx vs T0) must be < 10% | Roccaforte et al. 2026, Biochem Med (Zagreb)Table 1, row 'NE-WZ', 4 °C, column 'Approach 2 (% difference between medians)' |
| whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mL | Storageuncentrifuged | 2-8 °C | 24 hstable for the longest time tested | EFLM regression: time at which fitted % difference line reaches MPE 10% | Roccaforte et al. 2026, Biochem Med (Zagreb)Table 3, row 'NE-WZ', 4 °C |
| whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mL | Storageuncentrifuged | RT | 12 hthen changed, changed by 24 h | Approach 2: absolute % difference between medians (Tx vs T0) must be < 10% | Roccaforte et al. 2026, Biochem Med (Zagreb)Table 2, row 'NE-WZ', RT (blank in Table 1), column 'Approach 2 (% difference between medians)' |
| whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mL | Storageuncentrifuged | RT | 13 hlimit stated by the authors | EFLM regression: time at which fitted % difference line reaches MPE 10% | Roccaforte et al. 2026, Biochem Med (Zagreb)Table 3, row 'NE-WZ', RT |
| whole-blood-edta, K2EDTA (Greiner Bio-One Vacuette), 2 mL | Storageuncentrifuged | 37 °C | 9 hlimit stated by the authors | EFLM regression: time at which fitted % difference line reaches MPE 10% | Roccaforte et al. 2026, Biochem Med (Zagreb)Table 3, row 'NE-WZ', 37 °C |
Neutrophils (Absolute)20 findings
| Specimen, tube | Condition | Temperature | Stable for | Criterion | Study |
|---|---|---|---|---|---|
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 48 hstable for the longest time tested | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 4, row 'NE', XN-9000, 4 °C |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 48 hstable for the longest time tested | Bias% vs optimal bias (OP-Bias%, from biological variation; Table 1) | Buoro et al. 2016, Rev Bras Hematol HemoterTable 4, row 'NE', XN-9000, 4 °C |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 48 hstable for the longest time tested | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 4, row 'NE', BC-6800, 4 °C |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 24 hthen changed, changed by 36 h | Bias% vs optimal bias (OP-Bias%, from biological variation; Table 1) | Buoro et al. 2016, Rev Bras Hematol HemoterTable 4, row 'NE', BC-6800, 4 °C |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 8 hstable for the longest time tested | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 5, row 'NE', XN-9000, 4 °C |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 8 hstable for the longest time tested | Bias% vs optimal bias (OP-Bias%, from biological variation; Table 1) | Buoro et al. 2016, Rev Bras Hematol HemoterTable 5, row 'NE', XN-9000, 4 °C |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 8 hstable for the longest time tested | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 5, row 'NE', BC-6800, 4 °C |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 4 hthen changed, changed by 8 h | Bias% vs optimal bias (OP-Bias%, from biological variation; Table 1) | Buoro et al. 2016, Rev Bras Hematol HemoterTable 5, row 'NE', BC-6800, 4 °C |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 8 hstable for the longest time tested | Bias% vs critical difference (CD%, reference change value; Table 1) | Buoro et al. 2016, Rev Bras Hematol HemoterTable 5, row 'NE', BC-6800, 4 °C |
| whole-blood-edta, K2-EDTA (BD Vacutainer) | Storageuncentrifuged | 2-8 °C | 48 hstable for the longest time tested | Repeated-measures one-way ANOVA with Tukey post hoc vs 0 h, p < 0.05 (* in table) | Unalli et al. 2021, J Med BiochemTable 1, column '4 °C' (no * at any time point) |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | RT | 48 hstable for the longest time tested | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 4, row 'NE', XN-9000, RT |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | RT | 48 hstable for the longest time tested | Bias% vs optimal bias (OP-Bias%, from biological variation; Table 1) | Buoro et al. 2016, Rev Bras Hematol HemoterTable 4, row 'NE', XN-9000, RT |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | RT | 48 hstable for the longest time tested | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 4, row 'NE', BC-6800, RT |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | RT | 48 hstable for the longest time tested | Bias% vs optimal bias (OP-Bias%, from biological variation; Table 1) | Buoro et al. 2016, Rev Bras Hematol HemoterTable 4, row 'NE', BC-6800, RT |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | RT | 8 hstable for the longest time tested | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 5, row 'NE', XN-9000, RT |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | RT | 8 hstable for the longest time tested | Bias% vs optimal bias (OP-Bias%, from biological variation; Table 1) | Buoro et al. 2016, Rev Bras Hematol HemoterTable 5, row 'NE', XN-9000, RT |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | RT | 8 hstable for the longest time tested | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 5, row 'NE', BC-6800, RT |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | RT | 8 hstable for the longest time tested | Bias% vs optimal bias (OP-Bias%, from biological variation; Table 1) | Buoro et al. 2016, Rev Bras Hematol HemoterTable 5, row 'NE', BC-6800, RT |
| whole-blood-edta, K2-EDTA (BD Vacutainer) | Storageuncentrifuged | RT | 8 hthen changed, changed by 12 h | Repeated-measures one-way ANOVA with Tukey post hoc vs 0 h, p < 0.05 (* in table) | Unalli et al. 2021, J Med BiochemTable 1, block 'NEU', row '12 h', column '23 °C' |
| whole-blood-edta, K2-EDTA (BD Vacutainer) | Storageuncentrifuged | 10 °C | 48 hstable for the longest time tested | Repeated-measures one-way ANOVA with Tukey post hoc vs 0 h, p < 0.05 (* in table) | Unalli et al. 2021, J Med BiochemTable 1, column '10 °C' (no * at any time point) |
Nucleated Red Blood Cells (Absolute)8 findings
| Specimen, tube | Condition | Temperature | Stable for | Criterion | Study |
|---|---|---|---|---|---|
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 48 hstable for the longest time tested | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'NRBC', XN-9000, 4 °C |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 48 hstable for the longest time tested | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'NRBC', BC-6800, 4 °C |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 0 hchanged at the first time point, changed by 4 h | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'NRBC', XN-9000, 4 °C |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 4 hthen changed, changed by 8 h | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'NRBC', BC-6800, 4 °C |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | RT | 48 hstable for the longest time tested | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'NRBC', XN-9000, RT |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | RT | 48 hstable for the longest time tested | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'NRBC', BC-6800, RT |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | RT | 4 hthen changed, changed by 8 h | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'NRBC', XN-9000, RT |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | RT | 8 hstable for the longest time tested | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'NRBC', BC-6800, RT |
Orosomucoid (alpha-1-acid glycoprotein)2 findings
| Specimen, tube | Condition | Temperature | Stable for | Criterion | Study |
|---|---|---|---|---|---|
| whole-blood-heparin, BD Vacutainer lithium heparin, no gel (ref 368884) | Before spinninguncentrifuged | RT | 10 hstable for the longest time tested | |mean % deviation 10 h vs baseline| > laboratory bias goal (Table II; mostly biological-variation based, some empirically modified) | Henriksen et al. 2014, Scand J Clin Lab InvestTable II, row 'OROS' |
| whole-blood-heparin, BD Vacutainer lithium heparin, no gel (ref 368884) | Before spinninguncentrifuged | RT | 10 hstable for the longest time tested | CV of paired baseline/10 h results > goal CV (Table II) | Henriksen et al. 2014, Scand J Clin Lab InvestTable II, row 'OROS' |
Paraoxonase-1 (PON1) activity3 findings
| Specimen, tube | Condition | Temperature | Stable for | Criterion | Study |
|---|---|---|---|---|---|
| serum, serum (tube not stated) | Storageseparated | −20 °C | 2 monthsthen changed, changed by 6 months | t-test (two-sample, equal variances) vs t=0, 95% CI | Beekhof et al. 2012, Lipids Health DisResults, paragraph 3 |
| serum, serum (tube not stated) | Storageseparated | −70°C | 12 monthsstable for the longest time tested | t-test (two-sample, equal variances) vs t=0, 95% CI | Beekhof et al. 2012, Lipids Health DisResults, paragraph 3; Discussion |
| serum, serum (tube not stated) | Storageseparated | −196°C | 12 monthsstable for the longest time tested | t-test (two-sample, equal variances) vs t=0, 95% CI | Beekhof et al. 2012, Lipids Health DisResults, paragraph 3; Discussion |
Parathyroid Hormone (PTH)16 findings
| Specimen, tube | Condition | Temperature | Stable for | Criterion | Study |
|---|---|---|---|---|---|
| plasma-edta, BD Vacutainer EDTA-K2 | Storageseparated | −20 °C | 8 daysstable for the longest time tested | TEo (% change from initial) > TEa 30% (CLIA) | Khalil et al. 2021, J Clin Lab AnalTable 2 |
| plasma-edta, BD Vacutainer EDTA-K2 | Storageseparated | −20 °C | 0 dayschanged at the first time point, changed by 2 days | LSD post hoc vs initial, p < 0.001 (repeated-measures ANOVA) | Khalil et al. 2021, J Clin Lab AnalTable 2 |
| plasma-edta, BD Vacutainer EDTA-K2 | Storageseparated | 2-8 °C | 8 daysstable for the longest time tested | TEo (% change from initial) > TEa 30% (CLIA) | Khalil et al. 2021, J Clin Lab AnalTable 2 |
| plasma-edta, BD Vacutainer EDTA-K2 | Storageseparated | 2-8 °C | 0 dayschanged at the first time point, changed by 2 days | LSD post hoc vs initial, p < 0.001 (repeated-measures ANOVA) | Khalil et al. 2021, J Clin Lab AnalTable 2 |
| plasma-edta, BD Vacutainer EDTA-K2 | Storageseparated | RT | 4 daysthen changed, changed by 8 days | TEo (% change from initial) > TEa 30% (CLIA) | Khalil et al. 2021, J Clin Lab AnalTable 2 |
| plasma-edta, BD Vacutainer EDTA-K2 | Storageseparated | RT | 0 dayschanged at the first time point, changed by 2 days | LSD post hoc vs initial, p < 0.001 (repeated-measures ANOVA) | Khalil et al. 2021, J Clin Lab AnalTable 2 |
| serum, BD Vacutainer plain tube (no gel, no additive) | Storageseparated | −20 °C | 8 daysstable for the longest time tested | TEo (% change from initial) > TEa 30% (CLIA) | Khalil et al. 2021, J Clin Lab AnalTable 2 |
| serum, BD Vacutainer plain tube (no gel, no additive) | Storageseparated | −20 °C | 0 dayschanged at the first time point, changed by 2 days | LSD post hoc vs initial, p < 0.001 (repeated-measures ANOVA) | Khalil et al. 2021, J Clin Lab AnalTable 2 |
| serum, BD Vacutainer plain tube (no gel, no additive) | Storageseparated | 2-8 °C | 8 daysstable for the longest time tested | TEo (% change from initial) > TEa 30% (CLIA) | Khalil et al. 2021, J Clin Lab AnalTable 2 |
| serum, BD Vacutainer plain tube (no gel, no additive) | Storageseparated | 2-8 °C | 0 dayschanged at the first time point, changed by 2 days | LSD post hoc vs initial, p < 0.001 (repeated-measures ANOVA) | Khalil et al. 2021, J Clin Lab AnalTable 2 |
| serum, BD Vacutainer plain tube (no gel, no additive) | Storageseparated | RT | 0 dayschanged at the first time point, changed by 2 days | TEo (% change from initial) > TEa 30% (CLIA) | Khalil et al. 2021, J Clin Lab AnalTable 2 |
| serum, BD Vacutainer plain tube (no gel, no additive) | Storageseparated | RT | 0 dayschanged at the first time point, changed by 2 days | LSD post hoc vs initial, p < 0.001 (repeated-measures ANOVA) | Khalil et al. 2021, J Clin Lab AnalTable 2 |
| whole-blood-clotted, BD Vacutainer plain tube (no gel, no additive) | Before spinninguncentrifuged | not stated | 2 hstable for the longest time tested | TEo (% change from initial) > TEa 30% (CLIA) | Khalil et al. 2021, J Clin Lab AnalResults, 'Effects of a two-h delay in separation' |
| whole-blood-clotted, BD Vacutainer plain tube (no gel, no additive) | Before spinninguncentrifuged | not stated | 0 hchanged at the first time point, changed by 2 h | paired t test vs immediate separation, p < 0.05 | Khalil et al. 2021, J Clin Lab AnalResults, 'Effects of a two-h delay in separation' |
| whole-blood-edta, BD Vacutainer EDTA-K2 | Before spinninguncentrifuged | not stated | 2 hstable for the longest time tested | TEo (% change from initial) > TEa 30% (CLIA) | Khalil et al. 2021, J Clin Lab AnalResults, 'Effects of a two-h delay in separation' |
| whole-blood-edta, BD Vacutainer EDTA-K2 | Before spinninguncentrifuged | not stated | 0 hchanged at the first time point, changed by 2 h | paired t test vs immediate separation, p < 0.05 | Khalil et al. 2021, J Clin Lab AnalResults, 'Effects of a two-h delay in separation' |
Phosphate (inorganic phosphorus)12 findings
| Specimen, tube | Condition | Temperature | Stable for | Criterion | Study |
|---|---|---|---|---|---|
| plasma-li-heparin, Sarstedt lithium heparin, no gel separator (primary tube) | Storagecentrifuged-on-cells | 2-8 °C | 6.2 hthen changed, changed by 8 h | mean % difference vs time 0 exceeds total limit of change, TLC = ±√((1.65·CVa)² + (0.5·CVb)²) (EFLM BV), here 7.4%; stability limit from instability-equation regression | Illana et al. 2024, Biochem Med (Zagreb)Table 1, row 'Phos' |
| plasma-li-heparin, Sarstedt lithium heparin, no gel separator; plasma aliquot | Storageseparated | 2-8 °C | 96 hstable for the longest time tested | mean % difference vs time 0 exceeds total limit of change, TLC = ±√((1.65·CVa)² + (0.5·CVb)²) (EFLM BV), here 7.4%; stability limit from instability-equation regression | Illana et al. 2024, Biochem Med (Zagreb)Table 2, row 'Phos' |
| plasma-li-heparin, BD Vacutainer LH PST II gel separator, 3.0 mL (REF 367374) | Storagecentrifuged-on-gel | 2-8 °C | 24 hstable for the longest time tested | mean % difference smaller than TEa (Ricos: 0.25 x (CVi^2 + CVg^2)^0.5 + 1.65 x 0.5 x CVi) | Nonkes et al. 2019, Biochem Med (Zagreb)Table 1, row 'Phosphate, mmol/L' |
| plasma-li-heparin, BD Vacutainer Barricor LH 3 mL, mechanical separator | Storagecentrifuged-on-gel | 2-8 °C | 68 hlimit stated by the authors | bias vs aliquoted plasma reference tube > maximal allowable bias 3.2% (Ricos et al.); time to MAB from linear regression of bias on time to analysis | Gawria et al. 2020, J Clin Lab AnalTable 2, row 'Phosphate', Barricor BD |
| plasma-li-heparin, BD Vacutainer PST LH 3 mL, gel | Storagecentrifuged-on-gel | 2-8 °C | 29 hlimit stated by the authors | bias vs aliquoted plasma reference tube > maximal allowable bias 3.2% (Ricos et al.); time to MAB from linear regression of bias on time to analysis | Gawria et al. 2020, J Clin Lab AnalTable 2, row 'Phosphate', Gel BD |
| plasma-li-heparin, Greiner Vacuette LH 3 mL Sep, gel | Storagecentrifuged-on-gel | 2-8 °C | 35 hlimit stated by the authors | bias vs aliquoted plasma reference tube > maximal allowable bias 3.2% (Ricos et al.); time to MAB from linear regression of bias on time to analysis | Gawria et al. 2020, J Clin Lab AnalTable 2, row 'Phosphate', Gel Greiner |
| serum, BD Vacutainer serum (plain), 7.5 mL | Storageseparated | −20 °C | 30 daysstable for the longest time tested | clinically significant change: mean outside significant change limit (SCL = T1d mean ± 3.0 usual SD from 2 months QC) | Kachhawa et al. 2017, J Lab PhysiciansResults paragraph 1 (Table 2 image-only) |
| serum, BD Vacutainer SST (SKU 367986), gel | Storagecentrifuged-on-gel | 30 °C (±1 °C) | 72 hstable for the longest time tested | mean bias vs paired baseline within CLIA acceptance limits AND >2/3 of individual results within limits | Iacovetti et al. 2026, Pract Lab MedResults, 'Analyte stability after storage at 30 °C', paragraph 1 (Table 1 image-only) |
| serum, BD SST Microtainer (SKU 395967), gel; filled from venous blood | Storagecentrifuged-on-gel | 0 °C | 72 hstable for the longest time tested | mean bias vs paired baseline within CLIA acceptance limits AND >2/3 of individual results within limits | Iacovetti et al. 2026, Pract Lab MedResults, 'Analyte stability after separation and storage in microvolume tubes at 0 °C' (Table 2 image-only) |
| whole-blood-clotted, BD Vacutainer SST (SKU 367986), gel | Before spinninguncentrifuged | 30 °C (±1 °C) | 0 hchanged at the first time point, changed by 24 h | mean bias vs paired baseline within CLIA acceptance limits AND >2/3 of individual results within limits | Iacovetti et al. 2026, Pract Lab MedResults, 'Analyte stability after storage at 30 °C', paragraph 2 (Table 1 image-only) |
| whole-blood-heparin, BD Vacutainer lithium heparin, no gel (ref 368884) | Before spinninguncentrifuged | RT | 0 hchanged at the first time point, changed by 10 h | |mean % deviation 10 h vs baseline| > laboratory bias goal (Table II; mostly biological-variation based, some empirically modified) | Henriksen et al. 2014, Scand J Clin Lab InvestTable II, row 'P' |
| whole-blood-heparin, BD Vacutainer lithium heparin, no gel (ref 368884) | Before spinninguncentrifuged | RT | 0 hchanged at the first time point, changed by 10 h | CV of paired baseline/10 h results > goal CV (Table II) | Henriksen et al. 2014, Scand J Clin Lab InvestTable II, row 'P' |
Platelet Count21 findings
| Specimen, tube | Condition | Temperature | Stable for | Criterion | Study |
|---|---|---|---|---|---|
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 48 hstable for the longest time tested | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 4, row 'PLT', XN-9000, 4 °C |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 0 hchanged at the first time point, changed by 2 h | Bias% vs optimal bias (OP-Bias%, from biological variation; Table 1) | Buoro et al. 2016, Rev Bras Hematol HemoterTable 4, row 'PLT', XN-9000, 4 °C |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 48 hstable for the longest time tested | Bias% vs critical difference (CD%, reference change value; Table 1) | Buoro et al. 2016, Rev Bras Hematol HemoterTable 4, row 'PLT', XN-9000, 4 °C |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 48 hstable for the longest time tested | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 4, row 'PLT', BC-6800, 4 °C |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 48 hstable for the longest time tested | Bias% vs optimal bias (OP-Bias%, from biological variation; Table 1) | Buoro et al. 2016, Rev Bras Hematol HemoterTable 4, row 'PLT', BC-6800, 4 °C |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 8 hstable for the longest time tested | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 5, row 'PLT', XN-9000, 4 °C |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 8 hstable for the longest time tested | Bias% vs optimal bias (OP-Bias%, from biological variation; Table 1) | Buoro et al. 2016, Rev Bras Hematol HemoterTable 5, row 'PLT', XN-9000, 4 °C |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 8 hstable for the longest time tested | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 5, row 'PLT', BC-6800, 4 °C |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 8 hstable for the longest time tested | Bias% vs optimal bias (OP-Bias%, from biological variation; Table 1) | Buoro et al. 2016, Rev Bras Hematol HemoterTable 5, row 'PLT', BC-6800, 4 °C |
| whole-blood-edta, K2-EDTA (BD Vacutainer) | Storageuncentrifuged | 2-8 °C | 48 hstable for the longest time tested | Repeated-measures one-way ANOVA with Tukey post hoc vs 0 h, p < 0.05 (* in table) | Unalli et al. 2021, J Med BiochemResults paragraphs 1-2 and Discussion paragraph 1 (Figures 1-2; numbers in figures only) |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | RT | 48 hstable for the longest time tested | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 4, row 'PLT', XN-9000, RT |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | RT | 48 hstable for the longest time tested | Bias% vs optimal bias (OP-Bias%, from biological variation; Table 1) | Buoro et al. 2016, Rev Bras Hematol HemoterTable 4, row 'PLT', XN-9000, RT |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | RT | 48 hstable for the longest time tested | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 4, row 'PLT', BC-6800, RT |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | RT | 48 hstable for the longest time tested | Bias% vs optimal bias (OP-Bias%, from biological variation; Table 1) | Buoro et al. 2016, Rev Bras Hematol HemoterTable 4, row 'PLT', BC-6800, RT |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | RT | 8 hstable for the longest time tested | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 5, row 'PLT', XN-9000, RT |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | RT | 8 hstable for the longest time tested | Bias% vs optimal bias (OP-Bias%, from biological variation; Table 1) | Buoro et al. 2016, Rev Bras Hematol HemoterTable 5, row 'PLT', XN-9000, RT |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | RT | 8 hstable for the longest time tested | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 5, row 'PLT', BC-6800, RT |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | RT | 8 hstable for the longest time tested | Bias% vs optimal bias (OP-Bias%, from biological variation; Table 1) | Buoro et al. 2016, Rev Bras Hematol HemoterTable 5, row 'PLT', BC-6800, RT |
| whole-blood-edta, K2-EDTA (BD Vacutainer) | Storageuncentrifuged | RT | 48 hstable for the longest time tested | Repeated-measures one-way ANOVA with Tukey post hoc vs 0 h, p < 0.05 (* in table) | Unalli et al. 2021, J Med BiochemResults paragraphs 1-2 and Discussion paragraph 1 (Figures 1-2; numbers in figures only) |
| whole-blood-edta, K2EDTA (Becton Dickinson) | Storageuncentrifuged | 33 °C | 0 hchanged at the first time point, changed by 4 h | Unacceptable if median % difference > Biasmax (biological variation) and Holm-adjusted Wilcoxon P < 0.05; stable if equivalence shown (two one-tailed Wilcoxon tests vs Biasmax, Pholm < 0.05) | Jain et al. 2018, Biochem Med (Zagreb)Results paragraph 1; Table 2, row 'PLT' |
| whole-blood-edta, K2-EDTA (BD Vacutainer) | Storageuncentrifuged | 10 °C | 48 hstable for the longest time tested | Repeated-measures one-way ANOVA with Tukey post hoc vs 0 h, p < 0.05 (* in table) | Unalli et al. 2021, J Med BiochemResults paragraphs 1-2 and Discussion paragraph 1 (Figures 1-2; numbers in figures only) |
Platelet Distribution Width8 findings
| Specimen, tube | Condition | Temperature | Stable for | Criterion | Study |
|---|---|---|---|---|---|
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 36 hthen changed, changed by 48 h | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 4, row 'PDW', XN-9000, 4 °C |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 8 hthen changed, changed by 24 h | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 4, row 'PDW', BC-6800, 4 °C |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 8 hstable for the longest time tested | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 5, row 'PDW', XN-9000, 4 °C |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 8 hstable for the longest time tested | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 5, row 'PDW', BC-6800, 4 °C |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | RT | 8 hthen changed, changed by 24 h | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 4, row 'PDW', XN-9000, RT |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | RT | 8 hthen changed, changed by 24 h | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 4, row 'PDW', BC-6800, RT |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | RT | 8 hstable for the longest time tested | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 5, row 'PDW', XN-9000, RT |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | RT | 8 hstable for the longest time tested | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 5, row 'PDW', BC-6800, RT |
Platelet Large Cell Ratio8 findings
| Specimen, tube | Condition | Temperature | Stable for | Criterion | Study |
|---|---|---|---|---|---|
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 36 hthen changed, changed by 48 h | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 4, row 'P-LCR', XN-9000, 4 °C |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 8 hthen changed, changed by 24 h | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 4, row 'P-LCR', BC-6800, 4 °C |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 8 hstable for the longest time tested | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 5, row 'P-LCR', XN-9000, 4 °C |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 8 hstable for the longest time tested | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 5, row 'P-LCR', BC-6800, 4 °C |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | RT | 8 hthen changed, changed by 24 h | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 4, row 'P-LCR', XN-9000, RT |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | RT | 4 hthen changed, changed by 6 h | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 4, row 'P-LCR', BC-6800, RT |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | RT | 8 hstable for the longest time tested | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 5, row 'P-LCR', XN-9000, RT |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | RT | 8 hstable for the longest time tested | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 5, row 'P-LCR', BC-6800, RT |
Plateletcrit8 findings
| Specimen, tube | Condition | Temperature | Stable for | Criterion | Study |
|---|---|---|---|---|---|
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 48 hstable for the longest time tested | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 4, row 'PCT', XN-9000, 4 °C |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 48 hstable for the longest time tested | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 4, row 'PCT', BC-6800, 4 °C |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 8 hstable for the longest time tested | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 5, row 'PCT', XN-9000, 4 °C |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 8 hstable for the longest time tested | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 5, row 'PCT', BC-6800, 4 °C |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | RT | 48 hstable for the longest time tested | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 4, row 'PCT', XN-9000, RT |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | RT | 48 hstable for the longest time tested | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 4, row 'PCT', BC-6800, RT |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | RT | 8 hstable for the longest time tested | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 5, row 'PCT', XN-9000, RT |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | RT | 8 hstable for the longest time tested | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 5, row 'PCT', BC-6800, RT |
Potassium25 findings
| Specimen, tube | Condition | Temperature | Stable for | Criterion | Study |
|---|---|---|---|---|---|
| plasma-li-heparin, Sarstedt lithium heparin, no gel separator (primary tube) | Storagecentrifuged-on-cells | 2-8 °C | 4.7 hthen changed, changed by 6 h | mean % difference vs time 0 exceeds total limit of change, TLC = ±√((1.65·CVa)² + (0.5·CVb)²) (EFLM BV), here 4.7%; stability limit from instability-equation regression | Illana et al. 2024, Biochem Med (Zagreb)Table 1, row 'K' |
| plasma-li-heparin, Sarstedt lithium heparin, no gel separator; plasma aliquot | Storageseparated | 2-8 °C | 96 hstable for the longest time tested | mean % difference vs time 0 exceeds total limit of change, TLC = ±√((1.65·CVa)² + (0.5·CVb)²) (EFLM BV), here 4.7%; stability limit from instability-equation regression | Illana et al. 2024, Biochem Med (Zagreb)Table 2, row 'K' |
| plasma-li-heparin, BD Vacutainer PST II lithium heparin 3 mL (gel) | Storagecentrifuged-on-gel | 2-8 °C | 15 hstable for the longest time tested | clinically significant when the 95% CI of mean bias lies completely outside the SEQC maximum allowable bias (MAB = 1.65 x CVa), here 4% | Balbás et al. 2017, Biochem Med (Zagreb)Table 2, K+ row 'PST II sealed, 4 şC' |
| plasma-li-heparin, BD Barricor lithium heparin 5.5 mL (mechanical separator) | Storagecentrifuged-on-gel | 2-8 °C | 15 hstable for the longest time tested | clinically significant when the 95% CI of mean bias lies completely outside the SEQC maximum allowable bias (MAB = 1.65 x CVa), here 4% | Balbás et al. 2017, Biochem Med (Zagreb)Table 2, K+ row 'Barricor sealed, 4 şC' |
| plasma-li-heparin, BD Vacutainer lithium heparin 6 mL (regular) | Storagenot-stated | 2-8 °C | 4 hthen changed, changed by 24 h | mean % difference vs T0 exceeds total change limit (TCL), here ±6.21% | Dupuy et al. 2018, Biochem Med (Zagreb)Table 2, BD lithium heparin tubes, row 'K, mmol/L' |
| plasma-li-heparin, BD Barricor lithium heparin 5.5 mL (mechanical separator) | Storagecentrifuged-on-gel | 2-8 °C | 72 hstable for the longest time tested | mean % difference vs T0 exceeds total change limit (TCL), here ±6.21% | Dupuy et al. 2018, Biochem Med (Zagreb)Table 2, BD Barricor lithium heparin tubes, row 'K, mmol/L' |
| plasma-li-heparin, BD Vacutainer LH PST II gel separator, 3.0 mL (REF 367374) | Storagecentrifuged-on-gel | 2-8 °C | 24 hstable for the longest time tested | mean % difference smaller than TEa (Ricos: 0.25 x (CVi^2 + CVg^2)^0.5 + 1.65 x 0.5 x CVi) | Nonkes et al. 2019, Biochem Med (Zagreb)Table 1, row 'Potassium, mmol/L' |
| plasma-li-heparin, BD Vacutainer Barricor LH 3 mL, mechanical separator | Storagecentrifuged-on-gel | 2-8 °C | 40 hlimit stated by the authors | bias vs aliquoted plasma reference tube > maximal allowable bias 1.8% (Ricos et al.); time to MAB from linear regression of bias on time to analysis | Gawria et al. 2020, J Clin Lab AnalTable 2, row 'Potassium', Barricor BD |
| plasma-li-heparin, BD Vacutainer PST LH 3 mL, gel | Storagecentrifuged-on-gel | 2-8 °C | 8.7 hlimit stated by the authors | bias vs aliquoted plasma reference tube > maximal allowable bias 1.8% (Ricos et al.); time to MAB from linear regression of bias on time to analysis | Gawria et al. 2020, J Clin Lab AnalTable 2, row 'Potassium', Gel BD |
| plasma-li-heparin, Greiner Vacuette LH 3 mL Sep, gel | Storagecentrifuged-on-gel | 2-8 °C | 12 hlimit stated by the authors | bias vs aliquoted plasma reference tube > maximal allowable bias 1.8% (Ricos et al.); time to MAB from linear regression of bias on time to analysis | Gawria et al. 2020, J Clin Lab AnalTable 2, row 'Potassium', Gel Greiner |
| plasma-li-heparin, BD Vacutainer PST LH 3 mL, gel | Storagecentrifuged-on-gel | 2-8 °C | 32 hlimit stated by the authors | bias vs aliquoted reference > 4.0% (higher quality goal used by other authors) | Gawria et al. 2020, J Clin Lab AnalDiscussion, paragraph on potassium |
| plasma-li-heparin, Greiner Vacuette LH 3 mL Sep, gel | Storagecentrifuged-on-gel | 2-8 °C | 34 hlimit stated by the authors | bias vs aliquoted reference > 4.0% (higher quality goal used by other authors) | Gawria et al. 2020, J Clin Lab AnalDiscussion, paragraph on potassium |
| plasma-li-heparin, BD Vacutainer Barricor LH 3 mL, mechanical separator | Storagecentrifuged-on-gel | 2-8 °C | 80 hstable for the longest time tested | bias vs aliquoted reference > 4.0% (higher quality goal used by other authors) | Gawria et al. 2020, J Clin Lab AnalDiscussion, paragraph on potassium |
| plasma-li-heparin, BD Barricor 4.5 mL (lithium heparin, mechanical separator) | Storagecentrifuged-on-gel | 2-8 °C | 24 hthen changed, changed by 7 days | mean % difference vs time 0 exceeds desirable imprecision based on biological variation (EFLM/Westgard), here 2.05% | Mandić et al. 2020, Biochem Med (Zagreb)Tables 3 and 4, row 'Potassium, mmol/L', BD Barricor columns |
| plasma-li-heparin, BD Vacutainer PST II lithium heparin 3 mL (gel) | Storagecentrifuged-on-gel | RT | 6 hthen changed, changed by 9 h | clinically significant when the 95% CI of mean bias lies completely outside the SEQC maximum allowable bias (MAB = 1.65 x CVa), here 4% | Balbás et al. 2017, Biochem Med (Zagreb)Table 2, K+ row 'PST II open, rt' |
| plasma-li-heparin, BD Barricor lithium heparin 5.5 mL (mechanical separator) | Storagecentrifuged-on-gel | RT | 6 hthen changed, changed by 9 h | clinically significant when the 95% CI of mean bias lies completely outside the SEQC maximum allowable bias (MAB = 1.65 x CVa), here 4% | Balbás et al. 2017, Biochem Med (Zagreb)Table 2, K+ row 'Barricor open, rt' |
| serum, BD Vacutainer serum (plain), 7.5 mL | Storageseparated | −20 °C | 30 daysstable for the longest time tested | clinically significant change: mean outside significant change limit (SCL = T1d mean ± 3.0 usual SD from 2 months QC) | Kachhawa et al. 2017, J Lab PhysiciansResults paragraph 1 (Table 2 image-only) |
| serum, BD Vacutainer RST 4.0 mL (thrombin clot activator, gel) | Storagecentrifuged-on-gel | 2-8 °C | 24 hstable for the longest time tested | clinically significant if mean bias vs t0 exceeds desirable bias (biological variation), here 1.5% | Kocijancic et al. 2014, Biochem Med (Zagreb)Table 3, row 'K' |
| serum, BD Vacutainer SST II Advance 5.0 mL (clot activator, gel) | Storagecentrifuged-on-gel | 2-8 °C | 24 hstable for the longest time tested | clinically significant if mean bias vs t0 exceeds desirable bias (biological variation), here 1.5% | Kocijancic et al. 2014, Biochem Med (Zagreb)Table 4, row 'K' |
| serum, BD Vacutainer RST 5 mL (thrombin clot activator, gel) | Storagecentrifuged-on-gel | 2-8 °C | 24 hthen changed, changed by 7 days | mean % difference vs time 0 exceeds desirable imprecision based on biological variation (EFLM/Westgard), here 2.05% | Mandić et al. 2020, Biochem Med (Zagreb)Tables 3 and 4, row 'Potassium, mmol/L', BD RST columns |
| serum, BD Vacutainer SST (SKU 367986), gel | Storagecentrifuged-on-gel | 30 °C (±1 °C) | 72 hstable for the longest time tested | mean bias vs paired baseline within CLIA acceptance limits AND >2/3 of individual results within limits | Iacovetti et al. 2026, Pract Lab MedResults, 'Analyte stability after storage at 30 °C', paragraph 1 (Table 1 image-only) |
| serum, BD SST Microtainer (SKU 395967), gel; filled from venous blood | Storagecentrifuged-on-gel | 0 °C | 0 hchanged at the first time point, changed by 24 h | mean bias vs paired baseline within CLIA acceptance limits AND >2/3 of individual results within limits | Iacovetti et al. 2026, Pract Lab MedResults, 'Analyte stability after separation and storage in microvolume tubes at 0 °C' (Table 2 image-only) |
| whole-blood-clotted, BD Vacutainer SST (SKU 367986), gel | Before spinninguncentrifuged | 30 °C (±1 °C) | 0 hchanged at the first time point, changed by 24 h | mean bias vs paired baseline within CLIA acceptance limits AND >2/3 of individual results within limits | Iacovetti et al. 2026, Pract Lab MedResults, 'Analyte stability after storage at 30 °C', paragraph 2 (Table 1 image-only) |
| whole-blood-heparin, BD Vacutainer lithium heparin, no gel (ref 368884) | Before spinninguncentrifuged | RT | 10 hstable for the longest time tested | |mean % deviation 10 h vs baseline| > laboratory bias goal (Table II; mostly biological-variation based, some empirically modified) | Henriksen et al. 2014, Scand J Clin Lab InvestTable II, row 'K' |
| whole-blood-heparin, BD Vacutainer lithium heparin, no gel (ref 368884) | Before spinninguncentrifuged | RT | 0 hchanged at the first time point, changed by 10 h | CV of paired baseline/10 h results > goal CV (Table II) | Henriksen et al. 2014, Scand J Clin Lab InvestTable II, row 'K' |
Prostate-Specific Antigen, Total (PSA)2 findings
| Specimen, tube | Condition | Temperature | Stable for | Criterion | Study |
|---|---|---|---|---|---|
| whole-blood-clotted, BD Vacutainer serum, no gel (ref 368815) | Before spinninguncentrifuged | RT | 10 hstable for the longest time tested | |mean % deviation 10 h vs baseline| > laboratory bias goal (Table II; mostly biological-variation based, some empirically modified) | Henriksen et al. 2014, Scand J Clin Lab InvestTable II, row 'PSA' |
| whole-blood-clotted, BD Vacutainer serum, no gel (ref 368815) | Before spinninguncentrifuged | RT | 10 hstable for the longest time tested | CV of paired baseline/10 h results > goal CV (Table II) | Henriksen et al. 2014, Scand J Clin Lab InvestTable II, row 'PSA' |
Protein C5 findings
| Specimen, tube | Condition | Temperature | Stable for | Criterion | Study |
|---|---|---|---|---|---|
| plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquots | Freeze–thawseparated | −20 °C | 1 cyclelimit stated by the authors | max allowable mean deviation from t0: 4.00–8.00%: ±0.800% (absolute);>8.00%: ±10.0% (relative) | Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability' |
| plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquots | Freeze–thawseparated | −80 °C | 3 cycleslimit stated by the authors | max allowable mean deviation from t0: 4.00–8.00%: ±0.800% (absolute);>8.00%: ±10.0% (relative) | Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability' |
| plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquots | Storageseparated | −20 °C | 28 dayslimit stated by the authors | max allowable mean deviation from t0: 4.00–8.00%: ±0.800% (absolute);>8.00%: ±10.0% (relative) | Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability' |
| plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquots | Storageseparated | −80 °C | 10 monthslimit stated by the authors | max allowable mean deviation from t0: 4.00–8.00%: ±0.800% (absolute);>8.00%: ±10.0% (relative) | Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability' |
| plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquots | Storageseparated | RT | 4 hlimit stated by the authors | max allowable mean deviation from t0: 4.00–8.00%: ±0.800% (absolute);>8.00%: ±10.0% (relative) | Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability' |
Protein S5 findings
| Specimen, tube | Condition | Temperature | Stable for | Criterion | Study |
|---|---|---|---|---|---|
| plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquots | Freeze–thawseparated | −20 °C | 1 cyclelimit stated by the authors | max allowable mean deviation from t0: <20.0%: ≤±2.00% (absolute);≥20.0%: ≤±10.0% (relative) | Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability' |
| plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquots | Freeze–thawseparated | −80 °C | 3 cycleslimit stated by the authors | max allowable mean deviation from t0: <20.0%: ≤±2.00% (absolute);≥20.0%: ≤±10.0% (relative) | Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability' |
| plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquots | Storageseparated | −20 °C | 12 weekslimit stated by the authors | max allowable mean deviation from t0: <20.0%: ≤±2.00% (absolute);≥20.0%: ≤±10.0% (relative) | Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability' |
| plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquots | Storageseparated | −80 °C | 24 monthslimit stated by the authors | max allowable mean deviation from t0: <20.0%: ≤±2.00% (absolute);≥20.0%: ≤±10.0% (relative) | Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability' |
| plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquots | Storageseparated | RT | 24 hlimit stated by the authors | max allowable mean deviation from t0: <20.0%: ≤±2.00% (absolute);≥20.0%: ≤±10.0% (relative) | Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability' |
Prothrombin Time15 findings
| Specimen, tube | Condition | Temperature | Stable for | Criterion | Study |
|---|---|---|---|---|---|
| plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquots | Freeze–thawseparated | −20 °C | 1 cyclelimit stated by the authors | max allowable mean deviation from t0: ±10% | Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability' |
| plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquots | Freeze–thawseparated | −20 °C | 1 cyclelimit stated by the authors | max allowable mean deviation from t0: ≤4.5 INR: 100% ± 10%;>4.5 INR: 100% ± 15% | Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability' |
| plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquots | Freeze–thawseparated | −80 °C | 1 cyclelimit stated by the authors | max allowable mean deviation from t0: ±10% | Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability' |
| plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquots | Freeze–thawseparated | −80 °C | 1 cyclelimit stated by the authors | max allowable mean deviation from t0: ≤4.5 INR: 100% ± 10%;>4.5 INR: 100% ± 15% | Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability' |
| plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquots | Storageseparated | −20 °C | 7 weekslimit stated by the authors | max allowable mean deviation from t0: ±10% | Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability' |
| plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquots | Storageseparated | −20 °C | 6 weekslimit stated by the authors | max allowable mean deviation from t0: ≤4.5 INR: 100% ± 10%;>4.5 INR: 100% ± 15% | Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability' |
| plasma-citrate, BD Vacutainer 2.7 mL, 3.2% buffered sodium citrate; PPP aliquoted into microcentrifuge tubes | Storageseparated | −20 °C | 36 hstable for the longest time tested | change of more than ±10% of mean from baseline | Patil et al. 2022, CureusTable 2 |
| plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquots | Storageseparated | −80 °C | 11 monthslimit stated by the authors | max allowable mean deviation from t0: ±10% | Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability' |
| plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquots | Storageseparated | −80 °C | 3 monthslimit stated by the authors | max allowable mean deviation from t0: ≤4.5 INR: 100% ± 10%;>4.5 INR: 100% ± 15% | Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability' |
| plasma-citrate, BD 0.109 M sodium citrate (9:1); plasma aliquoted into nonactivating Eppendorf tubes | Storageseparated | 2-8 °C | 24 hstable for the longest time tested | mean % change from baseline >10%, or >25% of samples with individual change >10% (van Geest-Daalderop) | Feng et al. 2014, Sci RepTable 2, 'Acceptable Time (h)' |
| plasma-citrate, BD Vacutainer 2.7 mL, 3.2% buffered sodium citrate; PPP aliquoted into microcentrifuge tubes | Storageseparated | 2-8 °C | 24 hthen changed, changed by 36 h | change of more than ±10% of mean from baseline | Patil et al. 2022, CureusTable 2 |
| plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquots | Storageseparated | RT | 12 hlimit stated by the authors | max allowable mean deviation from t0: ±10% | Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability' |
| plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquots | Storageseparated | RT | 24 hlimit stated by the authors | max allowable mean deviation from t0: ≤4.5 INR: 100% ± 10%;>4.5 INR: 100% ± 15% | Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability' |
| plasma-citrate, BD 0.109 M sodium citrate (9:1); plasma aliquoted into nonactivating Eppendorf tubes | Storageseparated | RT | 24 hstable for the longest time tested | mean % change from baseline >10%, or >25% of samples with individual change >10% (van Geest-Daalderop) | Feng et al. 2014, Sci RepTable 2, 'Acceptable Time (h)' |
| plasma-citrate, BD Vacutainer 2.7 mL, 3.2% buffered sodium citrate; PPP aliquoted into microcentrifuge tubes | Storageseparated | RT | 0 hchanged at the first time point, changed by 12 h | change of more than ±10% of mean from baseline | Patil et al. 2022, CureusTable 2 |
Red Blood Cell Count21 findings
| Specimen, tube | Condition | Temperature | Stable for | Criterion | Study |
|---|---|---|---|---|---|
| whole-blood-edta, BD Vacutainer K2EDTA, 2 mL (spray-coated) | Storageuncentrifuged | 2-8 °C | 24 hstable for the longest time tested | Mean % change from baseline within analytical measurement uncertainty (MU ±1.5%) | Al-Najdawi et al. 2025, Med Pharm RepTable II, row 'RBC'; Recommendations |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 48 hstable for the longest time tested | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'RBC', XN-9000, 4 °C |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 48 hstable for the longest time tested | Bias% vs optimal bias (OP-Bias%, from biological variation; Table 1) | Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'RBC', XN-9000, 4 °C |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 48 hstable for the longest time tested | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'RBC', BC-6800, 4 °C |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 24 hthen changed, changed by 36 h | Bias% vs optimal bias (OP-Bias%, from biological variation; Table 1) | Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'RBC', BC-6800, 4 °C |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 8 hstable for the longest time tested | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'RBC', XN-9000, 4 °C |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 8 hstable for the longest time tested | Bias% vs optimal bias (OP-Bias%, from biological variation; Table 1) | Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'RBC', XN-9000, 4 °C |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 8 hstable for the longest time tested | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'RBC', BC-6800, 4 °C |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 8 hstable for the longest time tested | Bias% vs optimal bias (OP-Bias%, from biological variation; Table 1) | Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'RBC', BC-6800, 4 °C |
| whole-blood-edta, K2-EDTA (BD Vacutainer) | Storageuncentrifuged | 2-8 °C | 48 hstable for the longest time tested | Repeated-measures one-way ANOVA with Tukey post hoc vs 0 h, p < 0.05 (* in table) | Unalli et al. 2021, J Med BiochemResults paragraphs 1-2 and Discussion paragraph 1 (Figures 1-2; numbers in figures only) |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | RT | 48 hstable for the longest time tested | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'RBC', XN-9000, RT |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | RT | 48 hstable for the longest time tested | Bias% vs optimal bias (OP-Bias%, from biological variation; Table 1) | Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'RBC', XN-9000, RT |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | RT | 48 hstable for the longest time tested | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'RBC', BC-6800, RT |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | RT | 48 hstable for the longest time tested | Bias% vs optimal bias (OP-Bias%, from biological variation; Table 1) | Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'RBC', BC-6800, RT |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | RT | 8 hstable for the longest time tested | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'RBC', XN-9000, RT |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | RT | 8 hstable for the longest time tested | Bias% vs optimal bias (OP-Bias%, from biological variation; Table 1) | Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'RBC', XN-9000, RT |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | RT | 8 hstable for the longest time tested | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'RBC', BC-6800, RT |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | RT | 8 hstable for the longest time tested | Bias% vs optimal bias (OP-Bias%, from biological variation; Table 1) | Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'RBC', BC-6800, RT |
| whole-blood-edta, K2-EDTA (BD Vacutainer) | Storageuncentrifuged | RT | 48 hstable for the longest time tested | Repeated-measures one-way ANOVA with Tukey post hoc vs 0 h, p < 0.05 (* in table) | Unalli et al. 2021, J Med BiochemResults paragraphs 1-2 and Discussion paragraph 1 (Figures 1-2; numbers in figures only) |
| whole-blood-edta, K2EDTA (Becton Dickinson) | Storageuncentrifuged | 33 °C | 72 hstable for the longest time tested | Unacceptable if median % difference > Biasmax (biological variation) and Holm-adjusted Wilcoxon P < 0.05; stable if equivalence shown (two one-tailed Wilcoxon tests vs Biasmax, Pholm < 0.05) | Jain et al. 2018, Biochem Med (Zagreb)Results, last paragraph (equivalence); Table 2, row 'RBC' |
| whole-blood-edta, K2-EDTA (BD Vacutainer) | Storageuncentrifuged | 10 °C | 48 hstable for the longest time tested | Repeated-measures one-way ANOVA with Tukey post hoc vs 0 h, p < 0.05 (* in table) | Unalli et al. 2021, J Med BiochemResults paragraphs 1-2 and Discussion paragraph 1 (Figures 1-2; numbers in figures only) |
Red cell distribution width (CV or SD not stated)4 findings
| Specimen, tube | Condition | Temperature | Stable for | Criterion | Study |
|---|---|---|---|---|---|
| whole-blood-edta, K2-EDTA (BD Vacutainer) | Storageuncentrifuged | 2-8 °C | 48 hstable for the longest time tested | Repeated-measures one-way ANOVA with Tukey post hoc vs 0 h, p < 0.05 (* in table) | Unalli et al. 2021, J Med BiochemTable 2, column '4 °C' (no * at any time point) |
| whole-blood-edta, K2-EDTA (BD Vacutainer) | Storageuncentrifuged | RT | 8 hthen changed, changed by 12 h | Repeated-measures one-way ANOVA with Tukey post hoc vs 0 h, p < 0.05 (* in table) | Unalli et al. 2021, J Med BiochemTable 2, block 'RDW', row '12 h', column '23 °C' |
| whole-blood-edta, K2EDTA (Becton Dickinson) | Storageuncentrifuged | 33 °C | 0 hchanged at the first time point, changed by 4 h | Unacceptable if median % difference > Biasmax (biological variation) and Holm-adjusted Wilcoxon P < 0.05; stable if equivalence shown (two one-tailed Wilcoxon tests vs Biasmax, Pholm < 0.05) | Jain et al. 2018, Biochem Med (Zagreb)Results paragraph 1; Table 2, row 'RDW' |
| whole-blood-edta, K2-EDTA (BD Vacutainer) | Storageuncentrifuged | 10 °C | 48 hstable for the longest time tested | Repeated-measures one-way ANOVA with Tukey post hoc vs 0 h, p < 0.05 (* in table) | Unalli et al. 2021, J Med BiochemTable 2, column '10 °C' (no * at any time point) |
Red Cell Distribution Width (CV)21 findings
| Specimen, tube | Condition | Temperature | Stable for | Criterion | Study |
|---|---|---|---|---|---|
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 8 hthen changed, changed by 24 h | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'RDW-CV', XN-9000, 4 °C |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 8 hthen changed, changed by 24 h | Bias% vs optimal bias (OP-Bias%, from biological variation; Table 1) | Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'RDW-CV', XN-9000, 4 °C |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 48 hstable for the longest time tested | Bias% vs critical difference (CD%, reference change value; Table 1) | Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'RDW-CV', XN-9000, 4 °C |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 48 hstable for the longest time tested | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'RDW-CV', BC-6800, 4 °C |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 48 hstable for the longest time tested | Bias% vs optimal bias (OP-Bias%, from biological variation; Table 1) | Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'RDW-CV', BC-6800, 4 °C |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 8 hstable for the longest time tested | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'RDW-CV', XN-9000, 4 °C |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 8 hstable for the longest time tested | Bias% vs optimal bias (OP-Bias%, from biological variation; Table 1) | Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'RDW-CV', XN-9000, 4 °C |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 8 hstable for the longest time tested | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'RDW-CV', BC-6800, 4 °C |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 4 hthen changed, changed by 8 h | Bias% vs optimal bias (OP-Bias%, from biological variation; Table 1) | Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'RDW-CV', BC-6800, 4 °C |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 8 hstable for the longest time tested | Bias% vs critical difference (CD%, reference change value; Table 1) | Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'RDW-CV', BC-6800, 4 °C |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | RT | 8 hthen changed, changed by 24 h | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'RDW-CV', XN-9000, RT |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | RT | 8 hthen changed, changed by 24 h | Bias% vs optimal bias (OP-Bias%, from biological variation; Table 1) | Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'RDW-CV', XN-9000, RT |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | RT | 48 hstable for the longest time tested | Bias% vs critical difference (CD%, reference change value; Table 1) | Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'RDW-CV', XN-9000, RT |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | RT | 8 hthen changed, changed by 24 h | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'RDW-CV', BC-6800, RT |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | RT | 2 hthen changed, changed by 4 h | Bias% vs optimal bias (OP-Bias%, from biological variation; Table 1) | Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'RDW-CV', BC-6800, RT |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | RT | 8 hthen changed, changed by 24 h | Bias% vs critical difference (CD%, reference change value; Table 1) | Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'RDW-CV', BC-6800, RT |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | RT | 8 hstable for the longest time tested | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'RDW-CV', XN-9000, RT |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | RT | 8 hstable for the longest time tested | Bias% vs optimal bias (OP-Bias%, from biological variation; Table 1) | Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'RDW-CV', XN-9000, RT |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | RT | 8 hstable for the longest time tested | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'RDW-CV', BC-6800, RT |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | RT | 0 hchanged at the first time point, changed by 4 h | Bias% vs optimal bias (OP-Bias%, from biological variation; Table 1) | Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'RDW-CV', BC-6800, RT |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | RT | 8 hstable for the longest time tested | Bias% vs critical difference (CD%, reference change value; Table 1) | Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'RDW-CV', BC-6800, RT |
Red Cell Distribution Width (SD)8 findings
| Specimen, tube | Condition | Temperature | Stable for | Criterion | Study |
|---|---|---|---|---|---|
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 48 hstable for the longest time tested | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'RDW-SD', XN-9000, 4 °C |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 48 hstable for the longest time tested | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'RDW-SD', BC-6800, 4 °C |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 8 hstable for the longest time tested | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'RDW-SD', XN-9000, 4 °C |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 8 hstable for the longest time tested | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'RDW-SD', BC-6800, 4 °C |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | RT | 8 hthen changed, changed by 24 h | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'RDW-SD', XN-9000, RT |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | RT | 8 hthen changed, changed by 24 h | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'RDW-SD', BC-6800, RT |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | RT | 8 hstable for the longest time tested | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'RDW-SD', XN-9000, RT |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | RT | 4 hthen changed, changed by 8 h | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'RDW-SD', BC-6800, RT |
Reticulocyte Count (Absolute)18 findings
| Specimen, tube | Condition | Temperature | Stable for | Criterion | Study |
|---|---|---|---|---|---|
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 48 hstable for the longest time tested | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'RET', XN-9000, 4 °C |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 48 hstable for the longest time tested | Bias% vs optimal bias (OP-Bias%, from biological variation; Table 1) | Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'RET', XN-9000, 4 °C |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 48 hstable for the longest time tested | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'RET', BC-6800, 4 °C |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 36 hthen changed, changed by 48 h | Bias% vs optimal bias (OP-Bias%, from biological variation; Table 1) | Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'RET', BC-6800, 4 °C |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 48 hstable for the longest time tested | Bias% vs critical difference (CD%, reference change value; Table 1) | Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'RET', BC-6800, 4 °C |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 8 hstable for the longest time tested | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'RET', XN-9000, 4 °C |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 8 hstable for the longest time tested | Bias% vs optimal bias (OP-Bias%, from biological variation; Table 1) | Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'RET', XN-9000, 4 °C |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 8 hstable for the longest time tested | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'RET', BC-6800, 4 °C |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 8 hstable for the longest time tested | Bias% vs optimal bias (OP-Bias%, from biological variation; Table 1) | Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'RET', BC-6800, 4 °C |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | RT | 48 hstable for the longest time tested | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'RET', XN-9000, RT |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | RT | 48 hstable for the longest time tested | Bias% vs optimal bias (OP-Bias%, from biological variation; Table 1) | Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'RET', XN-9000, RT |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | RT | 48 hstable for the longest time tested | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'RET', BC-6800, RT |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | RT | 36 hthen changed, changed by 48 h | Bias% vs optimal bias (OP-Bias%, from biological variation; Table 1) | Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'RET', BC-6800, RT |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | RT | 48 hstable for the longest time tested | Bias% vs critical difference (CD%, reference change value; Table 1) | Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'RET', BC-6800, RT |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | RT | 8 hstable for the longest time tested | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'RET', XN-9000, RT |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | RT | 8 hstable for the longest time tested | Bias% vs optimal bias (OP-Bias%, from biological variation; Table 1) | Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'RET', XN-9000, RT |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | RT | 8 hstable for the longest time tested | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'RET', BC-6800, RT |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | RT | 8 hstable for the longest time tested | Bias% vs optimal bias (OP-Bias%, from biological variation; Table 1) | Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'RET', BC-6800, RT |
ROTEM EXTEM clot formation time2 findings
| Specimen, tube | Condition | Temperature | Stable for | Criterion | Study |
|---|---|---|---|---|---|
| whole-blood-citrate, Greiner Vacuette sodium citrate, standard fill (blood–air interface) | Storageuncentrifuged | RT | 0 hchanged at the first time point, changed by 2 h | maximum acceptable bias = 0.375 × total biological variation (EFLM BV database, Model 2) | Storm et al. 2026, Adv HematolTable 1, 'EXTEM, standard tubes' row 'CFT (s)' |
| whole-blood-citrate, Greiner Vacuette sodium citrate, completely filled (no air) | Storageuncentrifuged | RT | 2 hthen changed, changed by 4 h | maximum acceptable bias = 0.375 × total biological variation (EFLM BV database, Model 2) | Storm et al. 2026, Adv HematolTable 1, 'EXTEM, full tubes' row 'CFT (s)' |
ROTEM EXTEM clotting time2 findings
| Specimen, tube | Condition | Temperature | Stable for | Criterion | Study |
|---|---|---|---|---|---|
| whole-blood-citrate, Greiner Vacuette sodium citrate, standard fill (blood–air interface) | Storageuncentrifuged | RT | 24 hthen changed, changed by 48 h | maximum acceptable bias = 0.375 × total biological variation (EFLM BV database, Model 2) | Storm et al. 2026, Adv HematolTable 1, 'EXTEM, standard tubes' row 'CT (s)' |
| whole-blood-citrate, Greiner Vacuette sodium citrate, completely filled (no air) | Storageuncentrifuged | RT | 4 hthen changed, changed by 6 h | maximum acceptable bias = 0.375 × total biological variation (EFLM BV database, Model 2) | Storm et al. 2026, Adv HematolTable 1, 'EXTEM, full tubes' row 'CT (s)' |
ROTEM EXTEM maximum clot firmness2 findings
| Specimen, tube | Condition | Temperature | Stable for | Criterion | Study |
|---|---|---|---|---|---|
| whole-blood-citrate, Greiner Vacuette sodium citrate, standard fill (blood–air interface) | Storageuncentrifuged | RT | 6 hthen changed, changed by 24 h | maximum acceptable bias = 0.375 × total biological variation (EFLM BV database, Model 2) | Storm et al. 2026, Adv HematolTable 1, 'EXTEM, standard tubes' row 'MCF (millimeters)' |
| whole-blood-citrate, Greiner Vacuette sodium citrate, completely filled (no air) | Storageuncentrifuged | RT | 2 hthen changed, changed by 4 h | maximum acceptable bias = 0.375 × total biological variation (EFLM BV database, Model 2) | Storm et al. 2026, Adv HematolTable 1, 'EXTEM, full tubes' row 'MCF (millimeters)' |
ROTEM INTEM clot formation time2 findings
| Specimen, tube | Condition | Temperature | Stable for | Criterion | Study |
|---|---|---|---|---|---|
| whole-blood-citrate, Greiner Vacuette sodium citrate, standard fill (blood–air interface) | Storageuncentrifuged | RT | 48 hstable for the longest time tested | maximum acceptable bias = 0.375 × total biological variation (EFLM BV database, Model 2) | Storm et al. 2026, Adv HematolTable 1, 'INTEM, standard tubes' row 'CFT (s)' |
| whole-blood-citrate, Greiner Vacuette sodium citrate, completely filled (no air) | Storageuncentrifuged | RT | 24 hthen changed, changed by 48 h | maximum acceptable bias = 0.375 × total biological variation (EFLM BV database, Model 2) | Storm et al. 2026, Adv HematolTable 1, 'INTEM, full tubes' row 'CFT (s)' |
ROTEM INTEM clotting time2 findings
| Specimen, tube | Condition | Temperature | Stable for | Criterion | Study |
|---|---|---|---|---|---|
| whole-blood-citrate, Greiner Vacuette sodium citrate, standard fill (blood–air interface) | Storageuncentrifuged | RT | 48 hstable for the longest time tested | maximum acceptable bias = 0.375 × total biological variation (EFLM BV database, Model 2) | Storm et al. 2026, Adv HematolTable 1, 'INTEM, standard tubes' row 'CT (s)' |
| whole-blood-citrate, Greiner Vacuette sodium citrate, completely filled (no air) | Storageuncentrifuged | RT | 48 hstable for the longest time tested | maximum acceptable bias = 0.375 × total biological variation (EFLM BV database, Model 2) | Storm et al. 2026, Adv HematolTable 1, 'INTEM, full tubes' row 'CT (s)' |
Selenium4 findings
| Specimen, tube | Condition | Temperature | Stable for | Criterion | Study |
|---|---|---|---|---|---|
| whole-blood-clotted, BD Vacutainer trace element-free serum tube | Before spinninguncentrifuged | cold packs in insulated box; 9 (2) °C on arrival | 24 hstable for the longest time tested | % difference within desirable quality specification (from intra- and interindividual variation) | Jones et al. 2024, J NutrTable 2, row 'Selenium, μg/L'; Table 1 |
| whole-blood-clotted, Sarstedt Monovette serum | Before spinninguncentrifuged | chilled in transit (insulated box, 2 frozen cool packs; temperature not recorded) | 24 hstable for the longest time tested | % difference within desirable quality specification for bias (Fraser, from intra- and interindividual variation) | Jones et al. 2021, J NutrTable 2, row 'Selenium', column Serum |
| whole-blood-edta, Sarstedt Monovette EDTA | Before spinninguncentrifuged | chilled in transit (insulated box, 2 frozen cool packs; temperature not recorded) | 24 hstable for the longest time tested | % difference within desirable quality specification for bias (Fraser, from intra- and interindividual variation) | Jones et al. 2021, J NutrTable 2, row 'Selenium', column EDTA |
| whole-blood-heparin, Sarstedt Monovette lithium heparin, trace-element free | Before spinninguncentrifuged | chilled in transit (insulated box, 2 frozen cool packs; temperature not recorded) | 24 hstable for the longest time tested | % difference within desirable quality specification for bias (Fraser, from intra- and interindividual variation) | Jones et al. 2021, J NutrTable 2, row 'Selenium', column LH |
Sodium16 findings
| Specimen, tube | Condition | Temperature | Stable for | Criterion | Study |
|---|---|---|---|---|---|
| plasma-li-heparin, Sarstedt lithium heparin, no gel separator (primary tube) | Storagecentrifuged-on-cells | 2-8 °C | 48 hstable for the longest time tested | mean % difference vs time 0 exceeds total limit of change, TLC = ±√((1.65·CVa)² + (0.5·CVb)²) (EFLM BV), here 3.0%; stability limit from instability-equation regression | Illana et al. 2024, Biochem Med (Zagreb)Table 1, row 'Na' |
| plasma-li-heparin, Sarstedt lithium heparin, no gel separator; plasma aliquot | Storageseparated | 2-8 °C | 96 hstable for the longest time tested | mean % difference vs time 0 exceeds total limit of change, TLC = ±√((1.65·CVa)² + (0.5·CVb)²) (EFLM BV), here 3.0%; stability limit from instability-equation regression | Illana et al. 2024, Biochem Med (Zagreb)Table 2, row 'Na' |
| plasma-li-heparin, BD Vacutainer PST II lithium heparin 3 mL (gel) | Storagecentrifuged-on-gel | 2-8 °C | 15 hstable for the longest time tested | clinically significant when the 95% CI of mean bias lies completely outside the SEQC maximum allowable bias (MAB = 1.65 x CVa), here 0.6% | Balbás et al. 2017, Biochem Med (Zagreb)Table 2, Na+ row 'PST II sealed, 4 şC' |
| plasma-li-heparin, BD Barricor lithium heparin 5.5 mL (mechanical separator) | Storagecentrifuged-on-gel | 2-8 °C | 12 hthen changed, changed by 15 h | clinically significant when the 95% CI of mean bias lies completely outside the SEQC maximum allowable bias (MAB = 1.65 x CVa), here 0.6% | Balbás et al. 2017, Biochem Med (Zagreb)Table 2, Na+ row 'Barricor sealed, 4 şC' |
| plasma-li-heparin, BD Vacutainer lithium heparin 6 mL (regular) | Storagenot-stated | 2-8 °C | 72 hstable for the longest time tested | mean % difference vs T0 exceeds total change limit (TCL), here ±3.64% | Dupuy et al. 2018, Biochem Med (Zagreb)Table 2, BD lithium heparin tubes, row 'Na, mmol/L' |
| plasma-li-heparin, BD Barricor lithium heparin 5.5 mL (mechanical separator) | Storagecentrifuged-on-gel | 2-8 °C | 72 hstable for the longest time tested | mean % difference vs T0 exceeds total change limit (TCL), here ±3.64% | Dupuy et al. 2018, Biochem Med (Zagreb)Table 2, BD Barricor lithium heparin tubes, row 'Na, mmol/L' |
| plasma-li-heparin, BD Barricor 4.5 mL (lithium heparin, mechanical separator) | Storagecentrifuged-on-gel | 2-8 °C | 24 hthen changed, changed by 7 days | mean % difference vs time 0 exceeds desirable imprecision based on biological variation (EFLM/Westgard), here 0.25% | Mandić et al. 2020, Biochem Med (Zagreb)Tables 3 and 4, row 'Sodium, mmol/L', BD Barricor columns |
| plasma-li-heparin, BD Vacutainer PST II lithium heparin 3 mL (gel) | Storagecentrifuged-on-gel | RT | 0 hchanged at the first time point, changed by 3 h | clinically significant when the 95% CI of mean bias lies completely outside the SEQC maximum allowable bias (MAB = 1.65 x CVa), here 0.6% | Balbás et al. 2017, Biochem Med (Zagreb)Table 2, Na+ row 'PST II open, rt' |
| plasma-li-heparin, BD Barricor lithium heparin 5.5 mL (mechanical separator) | Storagecentrifuged-on-gel | RT | 0 hchanged at the first time point, changed by 3 h | clinically significant when the 95% CI of mean bias lies completely outside the SEQC maximum allowable bias (MAB = 1.65 x CVa), here 0.6% | Balbás et al. 2017, Biochem Med (Zagreb)Table 2, Na+ row 'Barricor open, rt' |
| serum, BD Vacutainer serum (plain), 7.5 mL | Storageseparated | −20 °C | 30 daysstable for the longest time tested | clinically significant change: mean outside significant change limit (SCL = T1d mean ± 3.0 usual SD from 2 months QC) | Kachhawa et al. 2017, J Lab PhysiciansResults paragraph 1; Discussion 'Metabolites' (Table 2 image-only) |
| serum, BD Vacutainer RST 5 mL (thrombin clot activator, gel) | Storagecentrifuged-on-gel | 2-8 °C | 24 hthen changed, changed by 7 days | mean % difference vs time 0 exceeds desirable imprecision based on biological variation (EFLM/Westgard), here 0.25% | Mandić et al. 2020, Biochem Med (Zagreb)Tables 3 and 4, row 'Sodium, mmol/L', BD RST columns |
| serum, BD Vacutainer SST (SKU 367986), gel | Storagecentrifuged-on-gel | 30 °C (±1 °C) | 24 hthen changed, changed by 72 h | mean bias vs paired baseline within CLIA acceptance limits AND >2/3 of individual results within limits | Iacovetti et al. 2026, Pract Lab MedResults, 'Analyte stability after storage at 30 °C', paragraph 1 (Table 1 image-only) |
| serum, BD SST Microtainer (SKU 395967), gel; filled from venous blood | Storagecentrifuged-on-gel | 0 °C | 72 hstable for the longest time tested | mean bias vs paired baseline within CLIA acceptance limits AND >2/3 of individual results within limits | Iacovetti et al. 2026, Pract Lab MedResults, 'Analyte stability after separation and storage in microvolume tubes at 0 °C' (Table 2 image-only) |
| whole-blood-clotted, BD Vacutainer SST (SKU 367986), gel | Before spinninguncentrifuged | 30 °C (±1 °C) | 24 hthen changed, changed by 72 h | mean bias vs paired baseline within CLIA acceptance limits AND >2/3 of individual results within limits | Iacovetti et al. 2026, Pract Lab MedResults, 'Analyte stability after storage at 30 °C', paragraph 2 (Table 1 image-only) |
| whole-blood-heparin, BD Vacutainer lithium heparin, no gel (ref 368884) | Before spinninguncentrifuged | RT | 10 hstable for the longest time tested | |mean % deviation 10 h vs baseline| > laboratory bias goal (Table II; mostly biological-variation based, some empirically modified) | Henriksen et al. 2014, Scand J Clin Lab InvestTable II, row 'Na' |
| whole-blood-heparin, BD Vacutainer lithium heparin, no gel (ref 368884) | Before spinninguncentrifuged | RT | 10 hstable for the longest time tested | CV of paired baseline/10 h results > goal CV (Table II) | Henriksen et al. 2014, Scand J Clin Lab InvestTable II, row 'Na' |
Soluble P-selectin2 findings
| Specimen, tube | Condition | Temperature | Stable for | Criterion | Study |
|---|---|---|---|---|---|
| whole-blood-edta, Greiner Vacuette K2E EDTA, standard fill | Before spinninguncentrifuged | RT | 2 hthen changed, changed by 4 h | maximum acceptable bias = 0.375 × total biological variation (EFLM BV database, Model 2) | Storm et al. 2026, Adv HematolTable 2, 'Standard tubes' row 'sP-selectin' |
| whole-blood-edta, Greiner Vacuette K2E EDTA, completely filled (no air) | Before spinninguncentrifuged | RT | 0 hchanged at the first time point, changed by 2 h | maximum acceptable bias = 0.375 × total biological variation (EFLM BV database, Model 2) | Storm et al. 2026, Adv HematolTable 2, 'Full tubes' row 'sP-selectin' |
Soluble Transferrin Receptor3 findings
| Specimen, tube | Condition | Temperature | Stable for | Criterion | Study |
|---|---|---|---|---|---|
| whole-blood-clotted, BD Vacutainer SST | Before spinninguncentrifuged | RT | 24 hstable for the longest time tested | mean % deviation from baseline within total change limit TCL = sqrt((2.77 CVa)^2 + (0.5 CVb)^2), CVb from Ricos (Table 2) | Abraham et al. 2019, Biochem Med (Zagreb)Discussion, paragraph 1 (Results paragraphs 1-3; Figure 3) |
| whole-blood-clotted, BD Vacutainer SST | Before spinninguncentrifuged | 22-30 °C (logger 24-26 °C) | 24 hstable for the longest time tested | mean % deviation from baseline within total change limit TCL = sqrt((2.77 CVa)^2 + (0.5 CVb)^2), CVb from Ricos (Table 2) | Abraham et al. 2019, Biochem Med (Zagreb)Discussion, paragraph 1 (Results paragraphs 1-3; Figure 3) |
| whole-blood-clotted, BD Vacutainer SST | Before spinninguncentrifuged | > 30 °C (logger 38-40 °C) | 24 hstable for the longest time tested | mean % deviation from baseline within total change limit TCL = sqrt((2.77 CVa)^2 + (0.5 CVb)^2), CVb from Ricos (Table 2) | Abraham et al. 2019, Biochem Med (Zagreb)Discussion, paragraph 1 (Results paragraphs 1-3; Figure 3) |
Testosterone, Total2 findings
| Specimen, tube | Condition | Temperature | Stable for | Criterion | Study |
|---|---|---|---|---|---|
| dbs, PerkinElmer 226 Spot Saver RUO card | Storagedried | 2-8 °C | 30 daysstable for the longest time tested | authors' verdict vs baseline; no numeric stability limit stated | Gruenstein 2024, Anal Sci AdvResults i (Stability of the DBS); Table 4, row 'Refrigerated'; Table 5 |
| dbs, PerkinElmer 226 Spot Saver RUO card | Storagedried | RT | 30 daysstable for the longest time tested | authors' verdict vs baseline; no numeric stability limit stated | Gruenstein 2024, Anal Sci AdvResults i (Stability of the DBS); Table 4, row 'Room temperature'; Table 5 |
Thrombin Time3 findings
| Specimen, tube | Condition | Temperature | Stable for | Criterion | Study |
|---|---|---|---|---|---|
| plasma-citrate, BD 0.109 M sodium citrate (9:1); plasma aliquoted into nonactivating Eppendorf tubes | Storageseparated | 2-8 °C | 24 hstable for the longest time tested | mean % change from baseline >10%, or >25% of samples with individual change >10% (van Geest-Daalderop) | Feng et al. 2014, Sci RepTable 2, 'Acceptable Time (h)' |
| plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquots | Storageseparated | RT | 4 hlimit stated by the authors | max allowable mean deviation from t0: ±10%; ±20% for heparin samples | Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability' |
| plasma-citrate, BD 0.109 M sodium citrate (9:1); plasma aliquoted into nonactivating Eppendorf tubes | Storageseparated | RT | 24 hstable for the longest time tested | mean % change from baseline >10%, or >25% of samples with individual change >10% (van Geest-Daalderop) | Feng et al. 2014, Sci RepTable 2, 'Acceptable Time (h)' |
Thyroid Stimulating Hormone12 findings
| Specimen, tube | Condition | Temperature | Stable for | Criterion | Study |
|---|---|---|---|---|---|
| dbs, Whatman 903 Neonatal Screening Cards; heparinized venous blood (50 µL), wrapped in aluminium foil | Storagedried | −20 °C | 60 daysthen changed, changed by 180 days | significant change of log-TSH vs D0 (mixed effects model, P < 0.05) | Magalhães et al. 2018, Arch Endocrinol MetabTable 1; Results paragraph 2 |
| dbs, Whatman 903 Neonatal Screening Cards; heparinized venous blood (50 µL), wrapped in aluminium foil | Storagedried | −20 °C | 60 daysthen changed, changed by 180 days | significant change of log-TSH vs D0 (mixed effects model, P < 0.05) | Magalhães et al. 2018, Arch Endocrinol MetabTable 1; Results paragraph 2 |
| dbs, Whatman 903 Neonatal Screening Cards; heparinized venous blood (50 µL), wrapped in aluminium foil | Storagedried | 2-8 °C | 60 daysthen changed, changed by 180 days | significant change of log-TSH vs D0 (mixed effects model, P < 0.05) | Magalhães et al. 2018, Arch Endocrinol MetabTable 1; Results paragraph 2 |
| dbs, Whatman 903 Neonatal Screening Cards; heparinized venous blood (50 µL), wrapped in aluminium foil | Storagedried | 2-8 °C | 60 daysthen changed, changed by 180 days | significant change of log-TSH vs D0 (mixed effects model, P < 0.05) | Magalhães et al. 2018, Arch Endocrinol MetabTable 1; Results paragraph 2 |
| dbs, Whatman 903 Neonatal Screening Cards; heparinized venous blood (50 µL), wrapped in aluminium foil | Storagedried | RT | 30 daysthen changed, changed by 60 days | significant change of log-TSH vs D0 (mixed effects model, P < 0.05) | Magalhães et al. 2018, Arch Endocrinol MetabTable 1; Results paragraph 2 |
| dbs, Whatman 903 Neonatal Screening Cards; heparinized venous blood (50 µL), wrapped in aluminium foil | Storagedried | RT | 30 daysthen changed, changed by 60 days | significant change of log-TSH vs D0 (mixed effects model, P < 0.05) | Magalhães et al. 2018, Arch Endocrinol MetabTable 1; Results paragraph 2 |
| plasma-li-heparin, Sarstedt lithium heparin, no gel separator (primary tube) | Storagecentrifuged-on-cells | 2-8 °C | 48 hstable for the longest time tested | mean % difference vs time 0 exceeds total limit of change, TLC = ±√((1.65·CVa)² + (0.5·CVb)²) (EFLM BV), here 15.3%; stability limit from instability-equation regression | Illana et al. 2024, Biochem Med (Zagreb)Table 1, row 'TSH' |
| plasma-li-heparin, Sarstedt lithium heparin, no gel separator; plasma aliquot | Storageseparated | 2-8 °C | 96 hstable for the longest time tested | mean % difference vs time 0 exceeds total limit of change, TLC = ±√((1.65·CVa)² + (0.5·CVb)²) (EFLM BV), here 15.3%; stability limit from instability-equation regression | Illana et al. 2024, Biochem Med (Zagreb)Table 2, row 'TSH' |
| plasma-li-heparin, BD Barricor 4.5 mL (lithium heparin, mechanical separator) | Storagecentrifuged-on-gel | 2-8 °C | 7 daysstable for the longest time tested | mean % difference vs time 0 exceeds desirable imprecision based on biological variation (EFLM/Westgard), here 7.95% | Mandić et al. 2020, Biochem Med (Zagreb)Tables 3 and 4, row 'TSH, mIU/L', BD Barricor columns |
| serum, BD Vacutainer RST 5 mL (thrombin clot activator, gel) | Storagecentrifuged-on-gel | 2-8 °C | 7 daysstable for the longest time tested | mean % difference vs time 0 exceeds desirable imprecision based on biological variation (EFLM/Westgard), here 7.95% | Mandić et al. 2020, Biochem Med (Zagreb)Tables 3 and 4, row 'TSH, mIU/L', BD RST columns |
| whole-blood-clotted, BD Vacutainer serum, no gel (ref 368815) | Before spinninguncentrifuged | RT | 10 hstable for the longest time tested | |mean % deviation 10 h vs baseline| > laboratory bias goal (Table II; mostly biological-variation based, some empirically modified) | Henriksen et al. 2014, Scand J Clin Lab InvestTable II, row 'TSH' |
| whole-blood-clotted, BD Vacutainer serum, no gel (ref 368815) | Before spinninguncentrifuged | RT | 10 hstable for the longest time tested | CV of paired baseline/10 h results > goal CV (Table II) | Henriksen et al. 2014, Scand J Clin Lab InvestTable II, row 'TSH' |
Total Bilirubin16 findings
| Specimen, tube | Condition | Temperature | Stable for | Criterion | Study |
|---|---|---|---|---|---|
| plasma-li-heparin, Sarstedt lithium heparin, no gel separator (primary tube) | Storagecentrifuged-on-cells | 2-8 °C | 48 hstable for the longest time tested | mean % difference vs time 0 exceeds total limit of change, TLC = ±√((1.65·CVa)² + (0.5·CVb)²) (EFLM BV), here 21.4%; stability limit from instability-equation regression | Illana et al. 2024, Biochem Med (Zagreb)Table 1, row 'TBIL' |
| plasma-li-heparin, Sarstedt lithium heparin, no gel separator; plasma aliquot | Storageseparated | 2-8 °C | 96 hstable for the longest time tested | mean % difference vs time 0 exceeds total limit of change, TLC = ±√((1.65·CVa)² + (0.5·CVb)²) (EFLM BV), here 21.4%; stability limit from instability-equation regression | Illana et al. 2024, Biochem Med (Zagreb)Table 2, row 'TBIL' |
| plasma-li-heparin, BD Barricor 4.5 mL (lithium heparin, mechanical separator) | Storagecentrifuged-on-gel | 2-8 °C | 7 daysstable for the longest time tested | mean % difference vs time 0 exceeds desirable imprecision based on biological variation (EFLM/Westgard), here 10.9% | Mandić et al. 2020, Biochem Med (Zagreb)Tables 3 and 4, row 'Bilirubin, μmol/L', BD Barricor columns |
| serum, BD Vacutainer 8.5 mL plastic serum tube; pooled serum aliquoted into 1.5 mL Eppendorf tubes | Freeze–thawseparated | −20 °C | 1 cyclesthen changed, changed by 2 cycles | unstable only if both P < 0.005 (Bonferroni) and % change from T0 > desirable bias (Westgard/Ricos), here 11.4% | Cuhadar et al. 2013, Biochem Med (Zagreb)Table 2, row 'Bilirubin, total (μmol/L)', columns T1d-T10d |
| serum, BD Vacutainer 8.5 mL plastic serum tube; pooled serum aliquoted into 1.5 mL Eppendorf tubes | Storageseparated | −20 °C | 3 monthsstable for the longest time tested | unstable only if both P < 0.013 (Bonferroni) and % change from T0 > desirable bias (Westgard/Ricos), here 11.4% | Cuhadar et al. 2013, Biochem Med (Zagreb)Table 2, row 'Bilirubin, total (μmol/L)', columns T1m-T3m |
| serum, BD Vacutainer serum (plain), 7.5 mL | Storageseparated | −20 °C | 30 daysstable for the longest time tested | clinically significant change: mean outside significant change limit (SCL = T1d mean ± 3.0 usual SD from 2 months QC) | Kachhawa et al. 2017, J Lab PhysiciansResults paragraph 1 (Table 2 image-only) |
| serum, Vacuette Tube Serum Gel Separator 8 mL (Greiner, 455071) | Storagecentrifuged-on-gel | 2-8 °C | 1 weekstable for the longest time tested | % bias of median vs baseline exceeds desirable bias (Westgard/Ricos BV), here 11.4% | Cuhadar et al. 2012, Biochem Med (Zagreb)Table 3, row 'Bilirubin (total)', gel 4°C |
| serum, Vacuette Tube Serum No Gel 9 mL (Greiner, 455092), plain | Storagecentrifuged-on-cells | 2-8 °C | 12 hthen changed, changed by 18 h | % bias of median vs baseline exceeds desirable bias (Westgard/Ricos BV), here 11.4% | Cuhadar et al. 2012, Biochem Med (Zagreb)Table 3, row 'Bilirubin (total)', w/o gel 4°C |
| serum, BD Vacutainer RST 5 mL (thrombin clot activator, gel) | Storagecentrifuged-on-gel | 2-8 °C | 7 daysstable for the longest time tested | mean % difference vs time 0 exceeds desirable imprecision based on biological variation (EFLM/Westgard), here 10.9% | Mandić et al. 2020, Biochem Med (Zagreb)Tables 3 and 4, row 'Bilirubin, μmol/L', BD RST columns |
| serum, Vacuette Tube Serum Gel Separator 8 mL (Greiner, 455071) | Storagecentrifuged-on-gel | RT | 1 weekstable for the longest time tested | % bias of median vs baseline exceeds desirable bias (Westgard/Ricos BV), here 11.4% | Cuhadar et al. 2012, Biochem Med (Zagreb)Table 3, row 'Bilirubin (total)', gel 24°C |
| serum, Vacuette Tube Serum No Gel 9 mL (Greiner, 455092), plain | Storagecentrifuged-on-cells | RT | 12 hthen changed, changed by 18 h | % bias of median vs baseline exceeds desirable bias (Westgard/Ricos BV), here 11.4% | Cuhadar et al. 2012, Biochem Med (Zagreb)Table 3, row 'Bilirubin (total)', w/o gel 24°C |
| serum, BD Vacutainer SST (SKU 367986), gel | Storagecentrifuged-on-gel | 30 °C (±1 °C) | 24 hthen changed, changed by 72 h | mean bias vs paired baseline within CLIA acceptance limits AND >2/3 of individual results within limits | Iacovetti et al. 2026, Pract Lab MedResults, 'Analyte stability after storage at 30 °C', paragraph 1 (Table 1 image-only) |
| serum, BD SST Microtainer (SKU 395967), gel; filled from venous blood | Storagecentrifuged-on-gel | 0 °C | 72 hstable for the longest time tested | mean bias vs paired baseline within CLIA acceptance limits AND >2/3 of individual results within limits | Iacovetti et al. 2026, Pract Lab MedResults, 'Analyte stability after separation and storage in microvolume tubes at 0 °C' (Table 2 image-only) |
| whole-blood-clotted, BD Vacutainer SST (SKU 367986), gel | Before spinninguncentrifuged | 30 °C (±1 °C) | 72 hstable for the longest time tested | mean bias vs paired baseline within CLIA acceptance limits AND >2/3 of individual results within limits | Iacovetti et al. 2026, Pract Lab MedResults, 'Analyte stability after storage at 30 °C', paragraph 2 (Table 1 image-only) |
| whole-blood-heparin, BD Vacutainer lithium heparin, no gel (ref 368884) | Before spinninguncentrifuged | RT | 10 hstable for the longest time tested | |mean % deviation 10 h vs baseline| > laboratory bias goal (Table II; mostly biological-variation based, some empirically modified) | Henriksen et al. 2014, Scand J Clin Lab InvestTable II, row 'BILI' |
| whole-blood-heparin, BD Vacutainer lithium heparin, no gel (ref 368884) | Before spinninguncentrifuged | RT | 10 hstable for the longest time tested | CV of paired baseline/10 h results > goal CV (Table II) | Henriksen et al. 2014, Scand J Clin Lab InvestTable II, row 'BILI' |
Total Cholesterol28 findings
| Specimen, tube | Condition | Temperature | Stable for | Criterion | Study |
|---|---|---|---|---|---|
| dbs, Whatman 903 card, mailed in foil pouch with desiccant | Storagedried | not stated; ambient postal shipping in the US (Apr-Dec 2016) | 5 daysthen changed, changed by 6 days | regression coefficient for days before freezing vs 0-2 days significant (P < .05), Model 1 (DBS value, unadjusted) | Crimmins et al. 2020, Am J Hum BiolTable 3, rows 'Days before freezing', column 'Total cholesterol' |
| dbs, Whatman 903 card, mailed in foil pouch with desiccant | Storagedried | not stated; ambient postal shipping in the US (Apr-Dec 2016) | 5 daysthen changed, changed by 6 days | regression coefficient for days before freezing vs 0-2 days significant (P < .05), Model 2 (DBS value adjusted for venous value) | Crimmins et al. 2020, Am J Hum BiolTable 3, rows 'Days before freezing', column 'Total cholesterol' |
| plasma-edta, BD Vacutainer EDTA, 4 mL | Freeze–thawseparated | −20 °C | 2 cyclesthen changed, changed by 3 cycles | significant change vs day 1 (repeated-measures ANOVA, Tukey HSD post hoc, p<0.05) | Gojković et al. 2024, J Med BiochemResults paragraph 3 (EDTA -20°C, Figure 3) |
| plasma-edta, BD Vacutainer EDTA, 4 mL | Storageseparated | 2-8 °C | 2 daysthen changed, changed by 3 days | significant change vs day 1 (repeated-measures ANOVA, Tukey HSD post hoc, p<0.05) | Gojković et al. 2024, J Med BiochemResults paragraph 1 (EDTA+4°C, Figure 1) |
| plasma-fluoride, BD Vacutainer sodium fluoride/potassium oxalate, 4 mL | Freeze–thawseparated | −20 °C | 0 cycleschanged at the first time point, changed by 1 cycles | significant change vs day 1 (repeated-measures ANOVA, Tukey HSD post hoc, p<0.05) | Gojković et al. 2024, J Med BiochemResults paragraph 4 (NaF -20°C, Figure 4) |
| plasma-fluoride, BD Vacutainer sodium fluoride/potassium oxalate, 4 mL | Storageseparated | 2-8 °C | 4 daysstable for the longest time tested | significant change vs day 1 (repeated-measures ANOVA, Tukey HSD post hoc, p<0.05) | Gojković et al. 2024, J Med BiochemResults paragraph 2 (NaF+4°C, Figure 2) |
| plasma-li-heparin, Sarstedt lithium heparin, no gel separator (primary tube) | Storagecentrifuged-on-cells | 2-8 °C | 48 hstable for the longest time tested | mean % difference vs time 0 exceeds total limit of change, TLC = ±√((1.65·CVa)² + (0.5·CVb)²) (EFLM BV), here 4.7%; stability limit from instability-equation regression | Illana et al. 2024, Biochem Med (Zagreb)Table 1, row 'CHOL' |
| plasma-li-heparin, Sarstedt lithium heparin, no gel separator; plasma aliquot | Storageseparated | 2-8 °C | 96 hstable for the longest time tested | mean % difference vs time 0 exceeds total limit of change, TLC = ±√((1.65·CVa)² + (0.5·CVb)²) (EFLM BV), here 4.7%; stability limit from instability-equation regression | Illana et al. 2024, Biochem Med (Zagreb)Table 2, row 'CHOL' |
| serum, BD Vacutainer 8.5 mL plastic serum tube; pooled serum aliquoted into 1.5 mL Eppendorf tubes | Freeze–thawseparated | −20 °C | 10 cyclesstable for the longest time tested | unstable only if both P < 0.005 (Bonferroni) and % change from T0 > desirable bias (Westgard/Ricos), here 4.0% | Cuhadar et al. 2013, Biochem Med (Zagreb)Table 2, row 'Cholesterol (mmol/L)', columns T1d-T10d |
| serum, BD Vacutainer SST | Freeze–thawseparated | −70 °C, 1 month; thawed at room temperature | 0 cycleschanged at the first time point, changed by 1 cycle | mean % change (bias) from fresh baseline within total allowable error (TEa, desirable bias) | Abraham et al. 2021, J Lab PhysiciansTable 2 and Table 3, row 'Cholesterol (mg/dL)' |
| serum, BD Vacutainer 8.5 mL plastic serum tube; pooled serum aliquoted into 1.5 mL Eppendorf tubes | Storageseparated | −20 °C | 3 monthsstable for the longest time tested | unstable only if both P < 0.013 (Bonferroni) and % change from T0 > desirable bias (Westgard/Ricos), here 4.0% | Cuhadar et al. 2013, Biochem Med (Zagreb)Table 2, row 'Cholesterol (mmol/L)', columns T1m-T3m |
| serum, BD Vacutainer serum (plain), 7.5 mL | Storageseparated | −20 °C | 30 daysstable for the longest time tested | clinically significant change: mean outside significant change limit (SCL = T1d mean ± 3.0 usual SD from 2 months QC) | Kachhawa et al. 2017, J Lab PhysiciansDiscussion, 'Lipid' (Table 2 image-only) |
| serum, Vacuette Tube Serum Gel Separator 8 mL (Greiner, 455071) | Storagecentrifuged-on-gel | 2-8 °C | 1 weekstable for the longest time tested | % bias of median vs baseline exceeds desirable bias (Westgard/Ricos BV), here 4.0% | Cuhadar et al. 2012, Biochem Med (Zagreb)Table 3, row 'Cholesterol', gel 4°C |
| serum, Vacuette Tube Serum No Gel 9 mL (Greiner, 455092), plain | Storagecentrifuged-on-cells | 2-8 °C | 1 weekstable for the longest time tested | % bias of median vs baseline exceeds desirable bias (Westgard/Ricos BV), here 4.0% | Cuhadar et al. 2012, Biochem Med (Zagreb)Table 3, row 'Cholesterol', w/o gel 4°C |
| serum, Vacuette Tube Serum Gel Separator 8 mL (Greiner, 455071) | Storagecentrifuged-on-gel | RT | 1 weekstable for the longest time tested | % bias of median vs baseline exceeds desirable bias (Westgard/Ricos BV), here 4.0% | Cuhadar et al. 2012, Biochem Med (Zagreb)Table 3, row 'Cholesterol', gel 24°C |
| serum, Vacuette Tube Serum No Gel 9 mL (Greiner, 455092), plain | Storagecentrifuged-on-cells | RT | 1 weekstable for the longest time tested | % bias of median vs baseline exceeds desirable bias (Westgard/Ricos BV), here 4.0% | Cuhadar et al. 2012, Biochem Med (Zagreb)Table 3, row 'Cholesterol', w/o gel 24°C |
| serum, BD Vacutainer SST (SKU 367986), gel | Storagecentrifuged-on-gel | 30 °C (±1 °C) | 72 hstable for the longest time tested | mean bias vs paired baseline within CLIA acceptance limits AND >2/3 of individual results within limits | Iacovetti et al. 2026, Pract Lab MedResults, 'Analyte stability after storage at 30 °C', paragraph 1 (Table 1 image-only) |
| serum, BD SST Microtainer (SKU 395967), gel; filled from venous blood | Storagecentrifuged-on-gel | 0 °C | 72 hstable for the longest time tested | mean bias vs paired baseline within CLIA acceptance limits AND >2/3 of individual results within limits | Iacovetti et al. 2026, Pract Lab MedResults, 'Analyte stability after separation and storage in microvolume tubes at 0 °C' (Table 2 image-only) |
| whole-blood-clotted, BD Vacutainer SST | Before spinninguncentrifuged | RT | 24 hstable for the longest time tested | mean % deviation from baseline within total change limit TCL = sqrt((2.77 CVa)^2 + (0.5 CVb)^2), CVb from Ricos (Table 2) | Abraham et al. 2019, Biochem Med (Zagreb)Discussion, paragraph 1 (Results paragraphs 1-3; Figure 3) |
| whole-blood-clotted, BD Vacutainer trace element-free serum tube | Before spinninguncentrifuged | cold packs in insulated box; 9 (2) °C on arrival | 24 hstable for the longest time tested | % difference within desirable quality specification (from intra- and interindividual variation) | Jones et al. 2024, J NutrTable 2, row 'Total cholesterol, mmol/L'; Table 1 |
| whole-blood-clotted, BD Vacutainer SST (SKU 367986), gel | Before spinninguncentrifuged | 30 °C (±1 °C) | 72 hstable for the longest time tested | mean bias vs paired baseline within CLIA acceptance limits AND >2/3 of individual results within limits | Iacovetti et al. 2026, Pract Lab MedResults, 'Analyte stability after storage at 30 °C', paragraph 2 (Table 1 image-only) |
| whole-blood-clotted, BD Vacutainer SST | Before spinninguncentrifuged | 22-30 °C (logger 24-26 °C) | 24 hstable for the longest time tested | mean % deviation from baseline within total change limit TCL = sqrt((2.77 CVa)^2 + (0.5 CVb)^2), CVb from Ricos (Table 2) | Abraham et al. 2019, Biochem Med (Zagreb)Discussion, paragraph 1 (Results paragraphs 1-3; Figure 3) |
| whole-blood-clotted, BD Vacutainer SST | Before spinninguncentrifuged | > 30 °C (logger 38-40 °C) | 24 hstable for the longest time tested | mean % deviation from baseline within total change limit TCL = sqrt((2.77 CVa)^2 + (0.5 CVb)^2), CVb from Ricos (Table 2) | Abraham et al. 2019, Biochem Med (Zagreb)Discussion, paragraph 1 (Results paragraphs 1-3; Figure 3) |
| whole-blood-clotted, Sarstedt Monovette serum | Before spinninguncentrifuged | chilled in transit (insulated box, 2 frozen cool packs; temperature not recorded) | 24 hstable for the longest time tested | % difference within desirable quality specification for bias (Fraser, from intra- and interindividual variation) | Jones et al. 2021, J NutrTable 2, row 'Total cholesterol', column Serum |
| whole-blood-edta, Sarstedt Monovette EDTA | Before spinninguncentrifuged | chilled in transit (insulated box, 2 frozen cool packs; temperature not recorded) | 24 hstable for the longest time tested | % difference within desirable quality specification for bias (Fraser, from intra- and interindividual variation) | Jones et al. 2021, J NutrTable 2, row 'Total cholesterol', column EDTA |
| whole-blood-heparin, BD Vacutainer lithium heparin, no gel (ref 368884) | Before spinninguncentrifuged | RT | 10 hstable for the longest time tested | |mean % deviation 10 h vs baseline| > laboratory bias goal (Table II; mostly biological-variation based, some empirically modified) | Henriksen et al. 2014, Scand J Clin Lab InvestTable II, row 'CHOL' |
| whole-blood-heparin, BD Vacutainer lithium heparin, no gel (ref 368884) | Before spinninguncentrifuged | RT | 10 hstable for the longest time tested | CV of paired baseline/10 h results > goal CV (Table II) | Henriksen et al. 2014, Scand J Clin Lab InvestTable II, row 'CHOL' |
| whole-blood-heparin, Sarstedt Monovette lithium heparin, trace-element free | Before spinninguncentrifuged | chilled in transit (insulated box, 2 frozen cool packs; temperature not recorded) | 24 hstable for the longest time tested | % difference within desirable quality specification for bias (Fraser, from intra- and interindividual variation) | Jones et al. 2021, J NutrTable 2, row 'Total cholesterol', column LH |
Total protein15 findings
| Specimen, tube | Condition | Temperature | Stable for | Criterion | Study |
|---|---|---|---|---|---|
| plasma-li-heparin, Sarstedt lithium heparin, no gel separator (primary tube) | Storagecentrifuged-on-cells | 2-8 °C | 48 hstable for the longest time tested | mean % difference vs time 0 exceeds total limit of change, TLC = ±√((1.65·CVa)² + (0.5·CVb)²) (EFLM BV), here 4.0%; stability limit from instability-equation regression | Illana et al. 2024, Biochem Med (Zagreb)Table 1, row 'TP' |
| plasma-li-heparin, Sarstedt lithium heparin, no gel separator; plasma aliquot | Storageseparated | 2-8 °C | 96 hstable for the longest time tested | mean % difference vs time 0 exceeds total limit of change, TLC = ±√((1.65·CVa)² + (0.5·CVb)²) (EFLM BV), here 4.0%; stability limit from instability-equation regression | Illana et al. 2024, Biochem Med (Zagreb)Table 2, row 'TP' |
| plasma-li-heparin, BD Barricor 4.5 mL (lithium heparin, mechanical separator) | Storagecentrifuged-on-gel | 2-8 °C | 7 daysstable for the longest time tested | mean % difference vs time 0 exceeds desirable imprecision based on biological variation (EFLM/Westgard), here 1.3% | Mandić et al. 2020, Biochem Med (Zagreb)Tables 3 and 4, row 'Total protein, g/L', BD Barricor columns |
| serum, BD Vacutainer 8.5 mL plastic serum tube; pooled serum aliquoted into 1.5 mL Eppendorf tubes | Freeze–thawseparated | −20 °C | 0 cycleschanged at the first time point, changed by 1 cycles | unstable only if both P < 0.005 (Bonferroni) and % change from T0 > desirable bias (Westgard/Ricos), here 1.2% | Cuhadar et al. 2013, Biochem Med (Zagreb)Table 2, row 'Protein, total (g/L)', columns T1d-T10d |
| serum, BD Vacutainer SST | Freeze–thawseparated | −70 °C, 1 month; thawed at room temperature | 1 cyclestable for the longest time tested | mean % change (bias) from fresh baseline within total allowable error (TEa, desirable bias) | Abraham et al. 2021, J Lab PhysiciansTable 2 and Table 3, row 'Total protein (g/dL)' |
| serum, BD Vacutainer 8.5 mL plastic serum tube; pooled serum aliquoted into 1.5 mL Eppendorf tubes | Storageseparated | −20 °C | 2 monthsthen changed, changed by 3 months | unstable only if both P < 0.013 (Bonferroni) and % change from T0 > desirable bias (Westgard/Ricos), here 1.2% | Cuhadar et al. 2013, Biochem Med (Zagreb)Table 2, row 'Protein, total (g/L)', columns T1m-T3m |
| serum, BD Vacutainer serum (plain), 7.5 mL | Storageseparated | −20 °C | 30 daysstable for the longest time tested | clinically significant change: mean outside significant change limit (SCL = T1d mean ± 3.0 usual SD from 2 months QC) | Kachhawa et al. 2017, J Lab PhysiciansDiscussion, 'Protein' (Table 2 image-only) |
| serum, Vacuette Tube Serum Gel Separator 8 mL (Greiner, 455071) | Storagecentrifuged-on-gel | 2-8 °C | 0 hchanged at the first time point, changed by 6 h | % bias of median vs baseline exceeds desirable bias (Westgard/Ricos BV), here 1.2% | Cuhadar et al. 2012, Biochem Med (Zagreb)Table 3, row 'Protein (total)', gel 4°C |
| serum, Vacuette Tube Serum No Gel 9 mL (Greiner, 455092), plain | Storagecentrifuged-on-cells | 2-8 °C | 0 hchanged at the first time point, changed by 6 h | % bias of median vs baseline exceeds desirable bias (Westgard/Ricos BV), here 1.2% | Cuhadar et al. 2012, Biochem Med (Zagreb)Table 3, row 'Protein (total)', w/o gel 4°C |
| serum, BD Vacutainer RST 5 mL (thrombin clot activator, gel) | Storagecentrifuged-on-gel | 2-8 °C | 7 daysstable for the longest time tested | mean % difference vs time 0 exceeds desirable imprecision based on biological variation (EFLM/Westgard), here 1.3% | Mandić et al. 2020, Biochem Med (Zagreb)Tables 3 and 4, row 'Total protein, g/L', BD RST columns |
| serum, Vacuette Tube Serum Gel Separator 8 mL (Greiner, 455071) | Storagecentrifuged-on-gel | RT | 0 hchanged at the first time point, changed by 6 h | % bias of median vs baseline exceeds desirable bias (Westgard/Ricos BV), here 1.2% | Cuhadar et al. 2012, Biochem Med (Zagreb)Table 3, row 'Protein (total)', gel 24°C |
| serum, Vacuette Tube Serum No Gel 9 mL (Greiner, 455092), plain | Storagecentrifuged-on-cells | RT | 12 hthen changed, changed by 18 h | % bias of median vs baseline exceeds desirable bias (Westgard/Ricos BV), here 1.2% | Cuhadar et al. 2012, Biochem Med (Zagreb)Table 3, row 'Protein (total)', w/o gel 24°C |
| serum, BD Vacutainer SST (SKU 367986), gel | Storagecentrifuged-on-gel | 30 °C (±1 °C) | 72 hstable for the longest time tested | mean bias vs paired baseline within CLIA acceptance limits AND >2/3 of individual results within limits | Iacovetti et al. 2026, Pract Lab MedResults, 'Analyte stability after storage at 30 °C', paragraph 1 (Table 1 image-only) |
| serum, BD SST Microtainer (SKU 395967), gel; filled from venous blood | Storagecentrifuged-on-gel | 0 °C | 72 hstable for the longest time tested | mean bias vs paired baseline within CLIA acceptance limits AND >2/3 of individual results within limits | Iacovetti et al. 2026, Pract Lab MedResults, 'Analyte stability after separation and storage in microvolume tubes at 0 °C' (Table 2 image-only) |
| whole-blood-clotted, BD Vacutainer SST (SKU 367986), gel | Before spinninguncentrifuged | 30 °C (±1 °C) | 72 hstable for the longest time tested | mean bias vs paired baseline within CLIA acceptance limits AND >2/3 of individual results within limits | Iacovetti et al. 2026, Pract Lab MedResults, 'Analyte stability after storage at 30 °C', paragraph 2 (Table 1 image-only) |
Transferrin2 findings
| Specimen, tube | Condition | Temperature | Stable for | Criterion | Study |
|---|---|---|---|---|---|
| whole-blood-heparin, BD Vacutainer lithium heparin, no gel (ref 368884) | Before spinninguncentrifuged | RT | 10 hstable for the longest time tested | |mean % deviation 10 h vs baseline| > laboratory bias goal (Table II; mostly biological-variation based, some empirically modified) | Henriksen et al. 2014, Scand J Clin Lab InvestTable II, row 'TRANS' |
| whole-blood-heparin, BD Vacutainer lithium heparin, no gel (ref 368884) | Before spinninguncentrifuged | RT | 10 hstable for the longest time tested | CV of paired baseline/10 h results > goal CV (Table II) | Henriksen et al. 2014, Scand J Clin Lab InvestTable II, row 'TRANS' |
Triglycerides26 findings
| Specimen, tube | Condition | Temperature | Stable for | Criterion | Study |
|---|---|---|---|---|---|
| plasma-edta, BD Vacutainer EDTA, 4 mL | Freeze–thawseparated | −20 °C | 1 cyclesthen changed, changed by 2 cycles | significant change vs day 1 (repeated-measures ANOVA, Tukey HSD post hoc, p<0.05) | Gojković et al. 2024, J Med BiochemResults paragraph 3 (EDTA -20°C, Figure 3) |
| plasma-edta, BD Vacutainer EDTA, 4 mL | Storageseparated | 2-8 °C | 4 daysstable for the longest time tested | significant change vs day 1 (repeated-measures ANOVA, Tukey HSD post hoc, p<0.05) | Gojković et al. 2024, J Med BiochemResults paragraph 1 (EDTA+4°C, Figure 1) |
| plasma-fluoride, BD Vacutainer sodium fluoride/potassium oxalate, 4 mL | Freeze–thawseparated | −20 °C | 4 cyclesstable for the longest time tested | significant change vs day 1 (repeated-measures ANOVA, Tukey HSD post hoc, p<0.05) | Gojković et al. 2024, J Med BiochemResults paragraph 4 (NaF -20°C, Figure 4) |
| plasma-fluoride, BD Vacutainer sodium fluoride/potassium oxalate, 4 mL | Storageseparated | 2-8 °C | 4 daysstable for the longest time tested | significant change vs day 1 (repeated-measures ANOVA, Tukey HSD post hoc, p<0.05) | Gojković et al. 2024, J Med BiochemResults paragraph 2 (NaF+4°C, Figure 2) |
| plasma-li-heparin, Sarstedt lithium heparin, no gel separator (primary tube) | Storagecentrifuged-on-cells | 2-8 °C | 48 hstable for the longest time tested | mean % difference vs time 0 exceeds total limit of change, TLC = ±√((1.65·CVa)² + (0.5·CVb)²) (EFLM BV), here 11.4%; stability limit from instability-equation regression | Illana et al. 2024, Biochem Med (Zagreb)Table 1, row 'TG' |
| plasma-li-heparin, Sarstedt lithium heparin, no gel separator; plasma aliquot | Storageseparated | 2-8 °C | 96 hstable for the longest time tested | mean % difference vs time 0 exceeds total limit of change, TLC = ±√((1.65·CVa)² + (0.5·CVb)²) (EFLM BV), here 11.4%; stability limit from instability-equation regression | Illana et al. 2024, Biochem Med (Zagreb)Table 2, row 'TG' |
| serum, BD Vacutainer 8.5 mL plastic serum tube; pooled serum aliquoted into 1.5 mL Eppendorf tubes | Freeze–thawseparated | −20 °C | 10 cyclesstable for the longest time tested | unstable only if both P < 0.005 (Bonferroni) and % change from T0 > desirable bias (Westgard/Ricos), here 10.7% | Cuhadar et al. 2013, Biochem Med (Zagreb)Table 2, row 'Triglyceride (mmol/L)', columns T1d-T10d |
| serum, BD Vacutainer SST | Freeze–thawseparated | −70 °C, 1 month; thawed at room temperature | 1 cyclestable for the longest time tested | mean % change (bias) from fresh baseline within total allowable error (TEa, desirable bias) | Abraham et al. 2021, J Lab PhysiciansTable 2 and Table 3, row 'TAG (mg/dL)' |
| serum, BD Vacutainer 8.5 mL plastic serum tube; pooled serum aliquoted into 1.5 mL Eppendorf tubes | Storageseparated | −20 °C | 3 monthsstable for the longest time tested | unstable only if both P < 0.013 (Bonferroni) and % change from T0 > desirable bias (Westgard/Ricos), here 10.7% | Cuhadar et al. 2013, Biochem Med (Zagreb)Table 2, row 'Triglyceride (mmol/L)', columns T1m-T3m |
| serum, BD Vacutainer serum (plain), 7.5 mL | Storageseparated | −20 °C | 30 daysstable for the longest time tested | clinically significant change: mean outside significant change limit (SCL = T1d mean ± 3.0 usual SD from 2 months QC) | Kachhawa et al. 2017, J Lab PhysiciansDiscussion, 'Lipid' (Table 2 image-only) |
| serum, Vacuette Tube Serum Gel Separator 8 mL (Greiner, 455071) | Storagecentrifuged-on-gel | 2-8 °C | 1 weekstable for the longest time tested | % bias of median vs baseline exceeds desirable bias (Westgard/Ricos BV), here 10.7% | Cuhadar et al. 2012, Biochem Med (Zagreb)Table 3, row 'Triglycerides', gel 4°C |
| serum, Vacuette Tube Serum No Gel 9 mL (Greiner, 455092), plain | Storagecentrifuged-on-cells | 2-8 °C | 1 weekstable for the longest time tested | % bias of median vs baseline exceeds desirable bias (Westgard/Ricos BV), here 10.7% | Cuhadar et al. 2012, Biochem Med (Zagreb)Table 3, row 'Triglycerides', w/o gel 4°C |
| serum, Vacuette Tube Serum Gel Separator 8 mL (Greiner, 455071) | Storagecentrifuged-on-gel | RT | 1 weekstable for the longest time tested | % bias of median vs baseline exceeds desirable bias (Westgard/Ricos BV), here 10.7% | Cuhadar et al. 2012, Biochem Med (Zagreb)Table 3, row 'Triglycerides', gel 24°C |
| serum, Vacuette Tube Serum No Gel 9 mL (Greiner, 455092), plain | Storagecentrifuged-on-cells | RT | 1 weekstable for the longest time tested | % bias of median vs baseline exceeds desirable bias (Westgard/Ricos BV), here 10.7% | Cuhadar et al. 2012, Biochem Med (Zagreb)Table 3, row 'Triglycerides', w/o gel 24°C |
| serum, BD Vacutainer SST (SKU 367986), gel | Storagecentrifuged-on-gel | 30 °C (±1 °C) | 72 hstable for the longest time tested | mean bias vs paired baseline within CLIA acceptance limits AND >2/3 of individual results within limits | Iacovetti et al. 2026, Pract Lab MedResults, 'Analyte stability after storage at 30 °C', paragraph 1 (Table 1 image-only) |
| serum, BD SST Microtainer (SKU 395967), gel; filled from venous blood | Storagecentrifuged-on-gel | 0 °C | 72 hstable for the longest time tested | mean bias vs paired baseline within CLIA acceptance limits AND >2/3 of individual results within limits | Iacovetti et al. 2026, Pract Lab MedResults, 'Analyte stability after separation and storage in microvolume tubes at 0 °C' (Table 2 image-only) |
| whole-blood-clotted, BD Vacutainer SST | Before spinninguncentrifuged | RT | 24 hstable for the longest time tested | mean % deviation from baseline within total change limit TCL = sqrt((2.77 CVa)^2 + (0.5 CVb)^2), CVb from Ricos (Table 2) | Abraham et al. 2019, Biochem Med (Zagreb)Discussion, paragraph 1 (Results paragraphs 1-3; Figure 3) |
| whole-blood-clotted, BD Vacutainer trace element-free serum tube | Before spinninguncentrifuged | cold packs in insulated box; 9 (2) °C on arrival | 24 hstable for the longest time tested | % difference within desirable quality specification (from intra- and interindividual variation) | Jones et al. 2024, J NutrTable 2, row 'Triglycerides, mmol/L'; Table 1 |
| whole-blood-clotted, BD Vacutainer SST (SKU 367986), gel | Before spinninguncentrifuged | 30 °C (±1 °C) | 72 hstable for the longest time tested | mean bias vs paired baseline within CLIA acceptance limits AND >2/3 of individual results within limits | Iacovetti et al. 2026, Pract Lab MedResults, 'Analyte stability after storage at 30 °C', paragraph 2 (Table 1 image-only) |
| whole-blood-clotted, BD Vacutainer SST | Before spinninguncentrifuged | 22-30 °C (logger 24-26 °C) | 24 hstable for the longest time tested | mean % deviation from baseline within total change limit TCL = sqrt((2.77 CVa)^2 + (0.5 CVb)^2), CVb from Ricos (Table 2) | Abraham et al. 2019, Biochem Med (Zagreb)Discussion, paragraph 1 (Results paragraphs 1-3; Figure 3) |
| whole-blood-clotted, BD Vacutainer SST | Before spinninguncentrifuged | > 30 °C (logger 38-40 °C) | 24 hstable for the longest time tested | mean % deviation from baseline within total change limit TCL = sqrt((2.77 CVa)^2 + (0.5 CVb)^2), CVb from Ricos (Table 2) | Abraham et al. 2019, Biochem Med (Zagreb)Discussion, paragraph 1 (Results paragraphs 1-3; Figure 3) |
| whole-blood-clotted, Sarstedt Monovette serum | Before spinninguncentrifuged | chilled in transit (insulated box, 2 frozen cool packs; temperature not recorded) | 24 hstable for the longest time tested | % difference within desirable quality specification for bias (Fraser, from intra- and interindividual variation) | Jones et al. 2021, J NutrTable 2, row 'Triglycerides', column Serum |
| whole-blood-edta, Sarstedt Monovette EDTA | Before spinninguncentrifuged | chilled in transit (insulated box, 2 frozen cool packs; temperature not recorded) | 24 hstable for the longest time tested | % difference within desirable quality specification for bias (Fraser, from intra- and interindividual variation) | Jones et al. 2021, J NutrTable 2, row 'Triglycerides', column EDTA |
| whole-blood-heparin, BD Vacutainer lithium heparin, no gel (ref 368884) | Before spinninguncentrifuged | RT | 10 hstable for the longest time tested | |mean % deviation 10 h vs baseline| > laboratory bias goal (Table II; mostly biological-variation based, some empirically modified) | Henriksen et al. 2014, Scand J Clin Lab InvestTable II, row 'TGLY' |
| whole-blood-heparin, BD Vacutainer lithium heparin, no gel (ref 368884) | Before spinninguncentrifuged | RT | 10 hstable for the longest time tested | CV of paired baseline/10 h results > goal CV (Table II) | Henriksen et al. 2014, Scand J Clin Lab InvestTable II, row 'TGLY' |
| whole-blood-heparin, Sarstedt Monovette lithium heparin, trace-element free | Before spinninguncentrifuged | chilled in transit (insulated box, 2 frozen cool packs; temperature not recorded) | 24 hstable for the longest time tested | % difference within desirable quality specification for bias (Fraser, from intra- and interindividual variation) | Jones et al. 2021, J NutrTable 2, row 'Triglycerides', column LH |
Troponin I (assay sensitivity not stated)2 findings
| Specimen, tube | Condition | Temperature | Stable for | Criterion | Study |
|---|---|---|---|---|---|
| plasma-li-heparin, BD Barricor 4.5 mL (lithium heparin, mechanical separator) | Storagecentrifuged-on-gel | 2-8 °C | 24 hthen changed, changed by 7 days | mean % difference vs time 0 exceeds desirable imprecision based on biological variation (EFLM/Westgard), here 7.03% | Mandić et al. 2020, Biochem Med (Zagreb)Tables 3 and 4, row 'TnI, ng/L', BD Barricor columns |
| serum, BD Vacutainer RST 5 mL (thrombin clot activator, gel) | Storagecentrifuged-on-gel | 2-8 °C | 24 hthen changed, changed by 7 days | mean % difference vs time 0 exceeds desirable imprecision based on biological variation (EFLM/Westgard), here 7.03% | Mandić et al. 2020, Biochem Med (Zagreb)Tables 3 and 4, row 'TnI, ng/L', BD RST columns |
Troponin T (High-Sensitivity)2 findings
| Specimen, tube | Condition | Temperature | Stable for | Criterion | Study |
|---|---|---|---|---|---|
| plasma-li-heparin, BD Vacutainer lithium heparin 6 mL (regular) | Storagenot-stated | 2-8 °C | 72 hstable for the longest time tested | mean % difference vs T0 exceeds total change limit (TCL), here ±9.37% | Dupuy et al. 2018, Biochem Med (Zagreb)Table 2, BD lithium heparin tubes, row 'hs-TnT, ng/L' |
| plasma-li-heparin, BD Barricor lithium heparin 5.5 mL (mechanical separator) | Storagecentrifuged-on-gel | 2-8 °C | 72 hstable for the longest time tested | mean % difference vs T0 exceeds total change limit (TCL), here ±9.37% | Dupuy et al. 2018, Biochem Med (Zagreb)Table 2, BD Barricor lithium heparin tubes, row 'hs-TnT, ng/L' |
Urea12 findings
| Specimen, tube | Condition | Temperature | Stable for | Criterion | Study |
|---|---|---|---|---|---|
| plasma-edta, BD Vacutainer EDTA, 4 mL | Freeze–thawseparated | −20 °C | 4 cyclesstable for the longest time tested | significant change vs day 1 (repeated-measures ANOVA, Tukey HSD post hoc, p<0.05) | Gojković et al. 2024, J Med BiochemResults paragraph 3 (EDTA -20°C, Figure 3) |
| plasma-edta, BD Vacutainer EDTA, 4 mL | Storageseparated | 2-8 °C | 4 daysstable for the longest time tested | significant change vs day 1 (repeated-measures ANOVA, Tukey HSD post hoc, p<0.05) | Gojković et al. 2024, J Med BiochemResults paragraph 1 (EDTA+4°C, Figure 1) |
| plasma-fluoride, BD Vacutainer sodium fluoride/potassium oxalate, 4 mL | Freeze–thawseparated | −20 °C | 4 cyclesstable for the longest time tested | significant change vs day 1 (repeated-measures ANOVA, Tukey HSD post hoc, p<0.05) | Gojković et al. 2024, J Med BiochemResults paragraph 4 (NaF -20°C, Figure 4) |
| plasma-fluoride, BD Vacutainer sodium fluoride/potassium oxalate, 4 mL | Storageseparated | 2-8 °C | 4 daysstable for the longest time tested | significant change vs day 1 (repeated-measures ANOVA, Tukey HSD post hoc, p<0.05) | Gojković et al. 2024, J Med BiochemResults paragraph 2 (NaF+4°C, Figure 2) |
| plasma-li-heparin, Sarstedt lithium heparin, no gel separator (primary tube) | Storagecentrifuged-on-cells | 2-8 °C | 48 hstable for the longest time tested | mean % difference vs time 0 exceeds total limit of change, TLC = ±√((1.65·CVa)² + (0.5·CVb)²) (EFLM BV), here 12.8%; stability limit from instability-equation regression | Illana et al. 2024, Biochem Med (Zagreb)Table 1, row 'Urea' |
| plasma-li-heparin, Sarstedt lithium heparin, no gel separator; plasma aliquot | Storageseparated | 2-8 °C | 96 hstable for the longest time tested | mean % difference vs time 0 exceeds total limit of change, TLC = ±√((1.65·CVa)² + (0.5·CVb)²) (EFLM BV), here 12.8%; stability limit from instability-equation regression | Illana et al. 2024, Biochem Med (Zagreb)Table 2, row 'Urea' |
| plasma-li-heparin, BD Vacutainer LH PST II gel separator, 3.0 mL (REF 367374) | Storagecentrifuged-on-gel | 2-8 °C | 24 hstable for the longest time tested | mean % difference smaller than TEa (Ricos: 0.25 x (CVi^2 + CVg^2)^0.5 + 1.65 x 0.5 x CVi) | Nonkes et al. 2019, Biochem Med (Zagreb)Table 1, row 'Urea, mmol/L' |
| plasma-li-heparin, BD Barricor 4.5 mL (lithium heparin, mechanical separator) | Storagecentrifuged-on-gel | 2-8 °C | 7 daysstable for the longest time tested | mean % difference vs time 0 exceeds desirable imprecision based on biological variation (EFLM/Westgard), here 6.95% | Mandić et al. 2020, Biochem Med (Zagreb)Tables 3 and 4, row 'Urea, mmol/L', BD Barricor columns |
| serum, BD Vacutainer serum (plain), 7.5 mL | Storageseparated | −20 °C | 30 daysstable for the longest time tested | clinically significant change: mean outside significant change limit (SCL = T1d mean ± 3.0 usual SD from 2 months QC) | Kachhawa et al. 2017, J Lab PhysiciansResults paragraph 1 (Table 2 image-only) |
| serum, BD Vacutainer RST 5 mL (thrombin clot activator, gel) | Storagecentrifuged-on-gel | 2-8 °C | 7 daysstable for the longest time tested | mean % difference vs time 0 exceeds desirable imprecision based on biological variation (EFLM/Westgard), here 6.95% | Mandić et al. 2020, Biochem Med (Zagreb)Tables 3 and 4, row 'Urea, mmol/L', BD RST columns |
| whole-blood-heparin, BD Vacutainer lithium heparin, no gel (ref 368884) | Before spinninguncentrifuged | RT | 10 hstable for the longest time tested | |mean % deviation 10 h vs baseline| > laboratory bias goal (Table II; mostly biological-variation based, some empirically modified) | Henriksen et al. 2014, Scand J Clin Lab InvestTable II, row 'UREA' |
| whole-blood-heparin, BD Vacutainer lithium heparin, no gel (ref 368884) | Before spinninguncentrifuged | RT | 10 hstable for the longest time tested | CV of paired baseline/10 h results > goal CV (Table II) | Henriksen et al. 2014, Scand J Clin Lab InvestTable II, row 'UREA' |
Uric acid16 findings
| Specimen, tube | Condition | Temperature | Stable for | Criterion | Study |
|---|---|---|---|---|---|
| plasma-li-heparin, Sarstedt lithium heparin, no gel separator (primary tube) | Storagecentrifuged-on-cells | 2-8 °C | 48 hstable for the longest time tested | mean % difference vs time 0 exceeds total limit of change, TLC = ±√((1.65·CVa)² + (0.5·CVb)²) (EFLM BV), here 8.3%; stability limit from instability-equation regression | Illana et al. 2024, Biochem Med (Zagreb)Table 1, row 'UA' |
| plasma-li-heparin, Sarstedt lithium heparin, no gel separator; plasma aliquot | Storageseparated | 2-8 °C | 96 hstable for the longest time tested | mean % difference vs time 0 exceeds total limit of change, TLC = ±√((1.65·CVa)² + (0.5·CVb)²) (EFLM BV), here 8.3%; stability limit from instability-equation regression | Illana et al. 2024, Biochem Med (Zagreb)Table 2, row 'UA' |
| plasma-li-heparin, BD Barricor 4.5 mL (lithium heparin, mechanical separator) | Storagecentrifuged-on-gel | 2-8 °C | 7 daysstable for the longest time tested | mean % difference vs time 0 exceeds desirable imprecision based on biological variation (EFLM/Westgard), here 4.3% | Mandić et al. 2020, Biochem Med (Zagreb)Tables 3 and 4, row 'Uric acid, μmol/L', BD Barricor columns |
| serum, BD Vacutainer 8.5 mL plastic serum tube; pooled serum aliquoted into 1.5 mL Eppendorf tubes | Freeze–thawseparated | −20 °C | 3 cyclesthen changed, changed by 4 cycles | unstable only if both P < 0.005 (Bonferroni) and % change from T0 > desirable bias (Westgard/Ricos), here 4.9% | Cuhadar et al. 2013, Biochem Med (Zagreb)Table 2, row 'Uric acid (μmol/L)', columns T1d-T10d |
| serum, BD Vacutainer 8.5 mL plastic serum tube; pooled serum aliquoted into 1.5 mL Eppendorf tubes | Storageseparated | −20 °C | 3 monthsstable for the longest time tested | unstable only if both P < 0.013 (Bonferroni) and % change from T0 > desirable bias (Westgard/Ricos), here 4.9% | Cuhadar et al. 2013, Biochem Med (Zagreb)Table 2, row 'Uric acid (μmol/L)', columns T1m-T3m |
| serum, BD Vacutainer serum (plain), 7.5 mL | Storageseparated | −20 °C | 30 daysstable for the longest time tested | clinically significant change: mean outside significant change limit (SCL = T1d mean ± 3.0 usual SD from 2 months QC) | Kachhawa et al. 2017, J Lab PhysiciansResults paragraph 1 (Table 2 image-only) |
| serum, Vacuette Tube Serum Gel Separator 8 mL (Greiner, 455071) | Storagecentrifuged-on-gel | 2-8 °C | 30 hthen changed, changed by 36 h | % bias of median vs baseline exceeds desirable bias (Westgard/Ricos BV), here 4.9% | Cuhadar et al. 2012, Biochem Med (Zagreb)Table 3, row 'Uric acid', gel 4°C |
| serum, Vacuette Tube Serum No Gel 9 mL (Greiner, 455092), plain | Storagecentrifuged-on-cells | 2-8 °C | 18 hthen changed, changed by 24 h | % bias of median vs baseline exceeds desirable bias (Westgard/Ricos BV), here 4.9% | Cuhadar et al. 2012, Biochem Med (Zagreb)Table 3, row 'Uric acid', w/o gel 4°C |
| serum, BD Vacutainer RST 5 mL (thrombin clot activator, gel) | Storagecentrifuged-on-gel | 2-8 °C | 7 daysstable for the longest time tested | mean % difference vs time 0 exceeds desirable imprecision based on biological variation (EFLM/Westgard), here 4.3% | Mandić et al. 2020, Biochem Med (Zagreb)Tables 3 and 4, row 'Uric acid, μmol/L', BD RST columns |
| serum, Vacuette Tube Serum Gel Separator 8 mL (Greiner, 455071) | Storagecentrifuged-on-gel | RT | 6 hthen changed, changed by 12 h | % bias of median vs baseline exceeds desirable bias (Westgard/Ricos BV), here 4.9% | Cuhadar et al. 2012, Biochem Med (Zagreb)Table 3, row 'Uric acid', gel 24°C |
| serum, Vacuette Tube Serum No Gel 9 mL (Greiner, 455092), plain | Storagecentrifuged-on-cells | RT | 6 hthen changed, changed by 12 h | % bias of median vs baseline exceeds desirable bias (Westgard/Ricos BV), here 4.9% | Cuhadar et al. 2012, Biochem Med (Zagreb)Table 3, row 'Uric acid', w/o gel 24°C |
| serum, BD Vacutainer SST (SKU 367986), gel | Storagecentrifuged-on-gel | 30 °C (±1 °C) | 72 hstable for the longest time tested | mean bias vs paired baseline within CLIA acceptance limits AND >2/3 of individual results within limits | Iacovetti et al. 2026, Pract Lab MedResults, 'Analyte stability after storage at 30 °C', paragraph 1 (Table 1 image-only) |
| serum, BD SST Microtainer (SKU 395967), gel; filled from venous blood | Storagecentrifuged-on-gel | 0 °C | 72 hstable for the longest time tested | mean bias vs paired baseline within CLIA acceptance limits AND >2/3 of individual results within limits | Iacovetti et al. 2026, Pract Lab MedResults, 'Analyte stability after separation and storage in microvolume tubes at 0 °C' (Table 2 image-only) |
| whole-blood-clotted, BD Vacutainer SST (SKU 367986), gel | Before spinninguncentrifuged | 30 °C (±1 °C) | 72 hstable for the longest time tested | mean bias vs paired baseline within CLIA acceptance limits AND >2/3 of individual results within limits | Iacovetti et al. 2026, Pract Lab MedResults, 'Analyte stability after storage at 30 °C', paragraph 2 (Table 1 image-only) |
| whole-blood-heparin, BD Vacutainer lithium heparin, no gel (ref 368884) | Before spinninguncentrifuged | RT | 10 hstable for the longest time tested | |mean % deviation 10 h vs baseline| > laboratory bias goal (Table II; mostly biological-variation based, some empirically modified) | Henriksen et al. 2014, Scand J Clin Lab InvestTable II, row 'URATE' |
| whole-blood-heparin, BD Vacutainer lithium heparin, no gel (ref 368884) | Before spinninguncentrifuged | RT | 10 hstable for the longest time tested | CV of paired baseline/10 h results > goal CV (Table II) | Henriksen et al. 2014, Scand J Clin Lab InvestTable II, row 'URATE' |
Urine Albumin6 findings
| Specimen, tube | Condition | Temperature | Stable for | Criterion | Study |
|---|---|---|---|---|---|
| urine, Sarstedt 2 ml polypropylene micro tube (P72.609.001), untreated | Storagenot-stated | −20 °C | 0 monthschanged at the first time point, changed by 12 months | Wilcoxon signed-rank test, 12 months vs fresh | Chapman et al. 2019, Pract Lab MedResults, paragraph 2 |
| urine, alkalised (NaOH to pH > 8), Sarstedt 2 ml polypropylene micro tube (P72.609.001) | Storagenot-stated | −20 °C | 0 monthschanged at the first time point, changed by 12 months | Wilcoxon signed-rank test, 12 months vs fresh | Chapman et al. 2019, Pract Lab MedResults, paragraph 2; Fig. 1A |
| urine, boric acid precoated container (Sarstedt P51.9923.820), then Sarstedt 2 ml polypropylene micro tube (P72.609.001) | Storagenot-stated | −20 °C | 0 monthschanged at the first time point, changed by 12 months | Wilcoxon signed-rank test, 12 months vs fresh | Chapman et al. 2019, Pract Lab MedResults, paragraph 2; Fig. 1A |
| urine, pepstatin 1 μg/ml added, Sarstedt 2 ml polypropylene micro tube (P72.609.001) | Storagenot-stated | −20 °C | 0 monthschanged at the first time point, changed by 12 months | Wilcoxon signed-rank test, 12 months vs fresh | Chapman et al. 2019, Pract Lab MedResults, paragraph 2; Fig. 1A |
| urine, Sarstedt 1.5 ml low protein binding micro tube (P72.706.600) | Storagenot-stated | −20 °C | 0 monthschanged at the first time point, changed by 12 months | Wilcoxon signed-rank test, 12 months vs fresh | Chapman et al. 2019, Pract Lab MedResults, paragraph 2; Fig. 1A |
| urine, Sarstedt 2 ml polypropylene micro tube (P72.609.001), untreated | Storagenot-stated | −80 °C | 12 monthsstable for the longest time tested | Wilcoxon signed-rank test, 12 months vs fresh | Chapman et al. 2019, Pract Lab MedResults, paragraph 2 |
Urine Albumin/Creatinine Ratio6 findings
| Specimen, tube | Condition | Temperature | Stable for | Criterion | Study |
|---|---|---|---|---|---|
| urine, Sarstedt 2 ml polypropylene micro tube (P72.609.001), untreated | Storagenot-stated | −20 °C | 0 monthschanged at the first time point, changed by 12 months | Wilcoxon signed-rank test, 12 months vs fresh | Chapman et al. 2019, Pract Lab MedResults, paragraph 4 |
| urine, alkalised (NaOH to pH > 8), Sarstedt 2 ml polypropylene micro tube (P72.609.001) | Storagenot-stated | −20 °C | 0 monthschanged at the first time point, changed by 12 months | Wilcoxon signed-rank test, 12 months vs fresh | Chapman et al. 2019, Pract Lab MedResults, paragraph 4; Fig. 1C |
| urine, boric acid precoated container (Sarstedt P51.9923.820), then Sarstedt 2 ml polypropylene micro tube (P72.609.001) | Storagenot-stated | −20 °C | 0 monthschanged at the first time point, changed by 12 months | Wilcoxon signed-rank test, 12 months vs fresh | Chapman et al. 2019, Pract Lab MedResults, paragraph 4; Fig. 1C |
| urine, pepstatin 1 μg/ml added, Sarstedt 2 ml polypropylene micro tube (P72.609.001) | Storagenot-stated | −20 °C | 0 monthschanged at the first time point, changed by 12 months | Wilcoxon signed-rank test, 12 months vs fresh | Chapman et al. 2019, Pract Lab MedResults, paragraph 4; Fig. 1C |
| urine, Sarstedt 1.5 ml low protein binding micro tube (P72.706.600) | Storagenot-stated | −20 °C | 0 monthschanged at the first time point, changed by 12 months | Wilcoxon signed-rank test, 12 months vs fresh | Chapman et al. 2019, Pract Lab MedResults, paragraph 4; Fig. 1C |
| urine, Sarstedt 2 ml polypropylene micro tube (P72.609.001), untreated | Storagenot-stated | −80 °C | 12 monthsstable for the longest time tested | Wilcoxon signed-rank test, 12 months vs fresh | Chapman et al. 2019, Pract Lab MedResults, paragraph 4 |
Urine Creatinine8 findings
| Specimen, tube | Condition | Temperature | Stable for | Criterion | Study |
|---|---|---|---|---|---|
| urine, Sarstedt 2 ml polypropylene micro tube (P72.609.001), untreated | Storagenot-stated | −20 °C | 0 monthschanged at the first time point, changed by 12 months | Wilcoxon signed-rank test, 12 months vs fresh | Chapman et al. 2019, Pract Lab MedResults, paragraph 3 |
| urine, Sarstedt 2 ml polypropylene micro tube (P72.609.001), untreated | Storagenot-stated | −20 °C | 12 monthsstable for the longest time tested | change < 10% not clinically significant | Chapman et al. 2019, Pract Lab MedResults, paragraph 3 |
| urine, alkalised (NaOH to pH > 8), Sarstedt 2 ml polypropylene micro tube (P72.609.001) | Storagenot-stated | −20 °C | 0 monthschanged at the first time point, changed by 12 months | Wilcoxon signed-rank test, 12 months vs fresh | Chapman et al. 2019, Pract Lab MedResults, paragraph 3 |
| urine, alkalised (NaOH to pH > 8), Sarstedt 2 ml polypropylene micro tube (P72.609.001) | Storagenot-stated | −20 °C | 0 monthschanged at the first time point, changed by 12 months | change < 10% not clinically significant | Chapman et al. 2019, Pract Lab MedResults, paragraph 3 |
| urine, Sarstedt 2 ml polypropylene micro tube (P72.609.001), untreated | Storagenot-stated | −80 °C | 0 monthschanged at the first time point, changed by 12 months | Wilcoxon signed-rank test, 12 months vs fresh | Chapman et al. 2019, Pract Lab MedResults, paragraph 3 |
| urine, Sarstedt 2 ml polypropylene micro tube (P72.609.001), untreated | Storagenot-stated | −80 °C | 12 monthsstable for the longest time tested | change < 10% not clinically significant | Chapman et al. 2019, Pract Lab MedResults, paragraph 3 |
| urine, alkalised (NaOH to pH > 8), Sarstedt 2 ml polypropylene micro tube (P72.609.001) | Storagenot-stated | −80 °C | 0 monthschanged at the first time point, changed by 12 months | Wilcoxon signed-rank test, 12 months vs fresh | Chapman et al. 2019, Pract Lab MedResults, paragraph 3 |
| urine, alkalised (NaOH to pH > 8), Sarstedt 2 ml polypropylene micro tube (P72.609.001) | Storagenot-stated | −80 °C | 12 monthsstable for the longest time tested | change < 10% not clinically significant | Chapman et al. 2019, Pract Lab MedResults, paragraph 3 |
Vitamin A (Retinol)5 findings
| Specimen, tube | Condition | Temperature | Stable for | Criterion | Study |
|---|---|---|---|---|---|
| whole-blood-clotted, BD Vacutainer SST | Before spinninguncentrifuged | > 30 °C (logger 38-40 °C) | 12 hthen changed, changed by 24 h | mean % deviation from baseline within total change limit TCL = sqrt((2.77 CVa)^2 + (0.5 CVb)^2), CVb from Ricos (Table 2) | Abraham et al. 2019, Biochem Med (Zagreb)Results, paragraph 1 (Figure 3a) |
| whole-blood-clotted, Sarstedt Monovette serum | Before spinninguncentrifuged | chilled in transit (insulated box, 2 frozen cool packs; temperature not recorded) | 24 hstable for the longest time tested | % difference within desirable quality specification for bias (Fraser, from intra- and interindividual variation) | Jones et al. 2021, J NutrTable 2, row 'Retinol', column Serum |
| whole-blood-edta, Sarstedt Monovette EDTA | Before spinninguncentrifuged | chilled in transit (insulated box, 2 frozen cool packs; temperature not recorded) | 24 hstable for the longest time tested | % difference within desirable quality specification for bias (Fraser, from intra- and interindividual variation) | Jones et al. 2021, J NutrTable 2, row 'Retinol', column EDTA |
| whole-blood-heparin, BD Vacutainer lithium heparin | Before spinninguncentrifuged | cold packs in insulated box; 9 (2) °C on arrival | 24 hstable for the longest time tested | % difference within desirable quality specification (from intra- and interindividual variation) | Jones et al. 2024, J NutrTable 2, row 'Retinol, μmol/L'; Table 1 |
| whole-blood-heparin, Sarstedt Monovette lithium heparin, trace-element free | Before spinninguncentrifuged | chilled in transit (insulated box, 2 frozen cool packs; temperature not recorded) | 24 hstable for the longest time tested | % difference within desirable quality specification for bias (Fraser, from intra- and interindividual variation) | Jones et al. 2021, J NutrTable 2, row 'Retinol', column LH |
Vitamin B12 (Cobalamin)11 findings
| Specimen, tube | Condition | Temperature | Stable for | Criterion | Study |
|---|---|---|---|---|---|
| serum, BD Vacutainer SST | Freeze–thawseparated | −70 °C, 1 month; thawed at room temperature | 1 cyclestable for the longest time tested | mean % change (bias) from fresh baseline within total allowable error (TEa, desirable bias) | Abraham et al. 2021, J Lab PhysiciansTable 2 and Table 3, row 'Vitamin B12 (pg/mL)' |
| serum, serum (tube not stated) | Storageseparated | −70 °C for the first year, −80 °C thereafter | 0 yearschanged at the first time point, changed by 1 year | t-test (two-sample, equal variances) vs t=0, P < 0.05 | Jansen et al. 2018, J Nutr MetabResults, sections 3.1 and 3.3 |
| whole-blood-clotted, BD Vacutainer serum, no gel (ref 368815) | Before spinninguncentrifuged | RT | 10 hstable for the longest time tested | |mean % deviation 10 h vs baseline| > laboratory bias goal (Table II; mostly biological-variation based, some empirically modified) | Henriksen et al. 2014, Scand J Clin Lab InvestTable II, row 'B12' |
| whole-blood-clotted, BD Vacutainer serum, no gel (ref 368815) | Before spinninguncentrifuged | RT | 10 hstable for the longest time tested | CV of paired baseline/10 h results > goal CV (Table II) | Henriksen et al. 2014, Scand J Clin Lab InvestTable II, row 'B12' |
| whole-blood-clotted, BD Vacutainer SST | Before spinninguncentrifuged | RT | 24 hstable for the longest time tested | mean % deviation from baseline within total change limit TCL = sqrt((2.77 CVa)^2 + (0.5 CVb)^2), CVb from Ricos (Table 2) | Abraham et al. 2019, Biochem Med (Zagreb)Discussion, paragraph 1 (Results paragraphs 1-3; Figure 3) |
| whole-blood-clotted, BD Vacutainer trace element-free serum tube | Before spinninguncentrifuged | cold packs in insulated box; 9 (2) °C on arrival | 24 hstable for the longest time tested | % difference within desirable quality specification (from intra- and interindividual variation) | Jones et al. 2024, J NutrTable 2, row 'Vitamin B12, ng/L'; Table 1 |
| whole-blood-clotted, BD Vacutainer SST | Before spinninguncentrifuged | 22-30 °C (logger 24-26 °C) | 24 hstable for the longest time tested | mean % deviation from baseline within total change limit TCL = sqrt((2.77 CVa)^2 + (0.5 CVb)^2), CVb from Ricos (Table 2) | Abraham et al. 2019, Biochem Med (Zagreb)Discussion, paragraph 1 (Results paragraphs 1-3; Figure 3) |
| whole-blood-clotted, BD Vacutainer SST | Before spinninguncentrifuged | > 30 °C (logger 38-40 °C) | 24 hstable for the longest time tested | mean % deviation from baseline within total change limit TCL = sqrt((2.77 CVa)^2 + (0.5 CVb)^2), CVb from Ricos (Table 2) | Abraham et al. 2019, Biochem Med (Zagreb)Discussion, paragraph 1 (Results paragraphs 1-3; Figure 3) |
| whole-blood-clotted, Sarstedt Monovette serum | Before spinninguncentrifuged | chilled in transit (insulated box, 2 frozen cool packs; temperature not recorded) | 24 hstable for the longest time tested | % difference within desirable quality specification for bias (Fraser, from intra- and interindividual variation) | Jones et al. 2021, J NutrTable 2, row 'Vitamin B-12', column Serum |
| whole-blood-edta, Sarstedt Monovette EDTA | Before spinninguncentrifuged | chilled in transit (insulated box, 2 frozen cool packs; temperature not recorded) | 24 hstable for the longest time tested | % difference within desirable quality specification for bias (Fraser, from intra- and interindividual variation) | Jones et al. 2021, J NutrTable 2, row 'Vitamin B-12', column EDTA |
| whole-blood-heparin, Sarstedt Monovette lithium heparin, trace-element free | Before spinninguncentrifuged | chilled in transit (insulated box, 2 frozen cool packs; temperature not recorded) | 24 hstable for the longest time tested | % difference within desirable quality specification for bias (Fraser, from intra- and interindividual variation) | Jones et al. 2021, J NutrTable 2, row 'Vitamin B-12', column LH |
Vitamin B6 (Pyridoxal-5-Phosphate)4 findings
| Specimen, tube | Condition | Temperature | Stable for | Criterion | Study |
|---|---|---|---|---|---|
| whole-blood-clotted, Sarstedt Monovette serum | Before spinninguncentrifuged | chilled in transit (insulated box, 2 frozen cool packs; temperature not recorded) | 0 hchanged at the first time point, changed by 24 h | % difference within desirable quality specification for bias (Fraser, from intra- and interindividual variation) | Jones et al. 2021, J NutrTable 2, row 'PLP (vitamin B-6)', column Serum |
| whole-blood-edta, Sarstedt Monovette EDTA | Before spinninguncentrifuged | chilled in transit (insulated box, 2 frozen cool packs; temperature not recorded) | 24 hstable for the longest time tested | % difference within desirable quality specification for bias (Fraser, from intra- and interindividual variation) | Jones et al. 2021, J NutrTable 2, row 'PLP (vitamin B-6)', column EDTA |
| whole-blood-heparin, BD Vacutainer lithium heparin | Before spinninguncentrifuged | cold packs in insulated box; 9 (2) °C on arrival | 24 hstable for the longest time tested | % difference within desirable quality specification (from intra- and interindividual variation) | Jones et al. 2024, J NutrTable 2, row 'Vitamin B6 (PLP), nmol/L'; Table 1 |
| whole-blood-heparin, Sarstedt Monovette lithium heparin, trace-element free | Before spinninguncentrifuged | chilled in transit (insulated box, 2 frozen cool packs; temperature not recorded) | 24 hstable for the longest time tested | % difference within desirable quality specification for bias (Fraser, from intra- and interindividual variation) | Jones et al. 2021, J NutrTable 2, row 'PLP (vitamin B-6)', column LH |
Vitamin C (Ascorbic Acid)8 findings
| Specimen, tube | Condition | Temperature | Stable for | Criterion | Study |
|---|---|---|---|---|---|
| plasma-li-heparin, BD lithium heparin vacutainer (cat. 367886); plasma in 1.7 mL microfuge tubes, non-stabilized | Storageseparated | −80 °C | 26 weeksstable for the longest time tested | Paired t-test P≤0.05, difference-plot 95% CI through zero, or Passing–Bablok slope CI excluding 1.00 | Collie et al. 2021, Ann Lab MedResults, storage experiments |
| plasma-li-heparin, BD lithium heparin vacutainer (cat. 367886); plasma in 1.7 mL microfuge tubes, non-stabilized | Storageseparated | −80 °C | 26 weeksstable for the longest time tested | RCPAQAP allowable performance specifications (±25% for results ≥36.0 µmol/L) | Collie et al. 2021, Ann Lab MedResults, storage experiments; Fig. 2A |
| plasma-li-heparin, BD lithium heparin vacutainer (cat. 367886); plasma + PCA/DTPA 1:1 (v:v) in 1.7 mL microfuge tubes | Storageseparated | −80 °C | 26 weeksstable for the longest time tested | Paired t-test P≤0.05, difference-plot 95% CI through zero, or Passing–Bablok slope CI excluding 1.00 | Collie et al. 2021, Ann Lab MedResults, storage experiments |
| plasma-li-heparin, BD lithium heparin vacutainer (cat. 367886); plasma + PCA/DTPA 1:1 (v:v) in 1.7 mL microfuge tubes | Storageseparated | −80 °C | 26 weeksstable for the longest time tested | RCPAQAP allowable performance specifications (±25% for results ≥36.0 µmol/L) | Collie et al. 2021, Ann Lab MedResults, storage experiments; Fig. 2B |
| whole-blood-clotted, Sarstedt Monovette serum | Before spinninguncentrifuged | chilled in transit (insulated box, 2 frozen cool packs; temperature not recorded) | 24 hstable for the longest time tested | % difference within desirable quality specification for bias (Fraser, from intra- and interindividual variation) | Jones et al. 2021, J NutrTable 2, row 'Vitamin C', column Serum |
| whole-blood-edta, Sarstedt Monovette EDTA | Before spinninguncentrifuged | chilled in transit (insulated box, 2 frozen cool packs; temperature not recorded) | 0 hchanged at the first time point, changed by 24 h | % difference within desirable quality specification for bias (Fraser, from intra- and interindividual variation) | Jones et al. 2021, J NutrTable 2, row 'Vitamin C', column EDTA |
| whole-blood-heparin, BD Vacutainer lithium heparin | Before spinninguncentrifuged | cold packs in insulated box; 9 (2) °C on arrival | 24 hstable for the longest time tested | % difference within desirable quality specification (from intra- and interindividual variation) | Jones et al. 2024, J NutrTable 2, row 'Vitamin C, μmol/L'; Table 1 |
| whole-blood-heparin, Sarstedt Monovette lithium heparin, trace-element free | Before spinninguncentrifuged | chilled in transit (insulated box, 2 frozen cool packs; temperature not recorded) | 24 hstable for the longest time tested | % difference within desirable quality specification for bias (Fraser, from intra- and interindividual variation) | Jones et al. 2021, J NutrTable 2, row 'Vitamin C', column LH |
Vitamin E (Alpha-Tocopherol)4 findings
| Specimen, tube | Condition | Temperature | Stable for | Criterion | Study |
|---|---|---|---|---|---|
| whole-blood-clotted, Sarstedt Monovette serum | Before spinninguncentrifuged | chilled in transit (insulated box, 2 frozen cool packs; temperature not recorded) | 24 hstable for the longest time tested | % difference within desirable quality specification for bias (Fraser, from intra- and interindividual variation) | Jones et al. 2021, J NutrTable 2, row 'α-tocopherol', column Serum |
| whole-blood-edta, Sarstedt Monovette EDTA | Before spinninguncentrifuged | chilled in transit (insulated box, 2 frozen cool packs; temperature not recorded) | 24 hstable for the longest time tested | % difference within desirable quality specification for bias (Fraser, from intra- and interindividual variation) | Jones et al. 2021, J NutrTable 2, row 'α-tocopherol', column EDTA |
| whole-blood-heparin, BD Vacutainer lithium heparin | Before spinninguncentrifuged | cold packs in insulated box; 9 (2) °C on arrival | 24 hstable for the longest time tested | % difference within desirable quality specification (from intra- and interindividual variation) | Jones et al. 2024, J NutrTable 2, row 'α-Tocopherol, μmol/L'; Table 1 |
| whole-blood-heparin, Sarstedt Monovette lithium heparin, trace-element free | Before spinninguncentrifuged | chilled in transit (insulated box, 2 frozen cool packs; temperature not recorded) | 24 hstable for the longest time tested | % difference within desirable quality specification for bias (Fraser, from intra- and interindividual variation) | Jones et al. 2021, J NutrTable 2, row 'α-tocopherol', column LH |
Von Willebrand factor activity5 findings
| Specimen, tube | Condition | Temperature | Stable for | Criterion | Study |
|---|---|---|---|---|---|
| plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquots | Freeze–thawseparated | −20 °C | 1 cyclelimit stated by the authors | max allowable mean deviation from t0: ±10% | Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability' |
| plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquots | Freeze–thawseparated | −80 °C | 3 cycleslimit stated by the authors | max allowable mean deviation from t0: ±10% | Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability' |
| plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquots | Storageseparated | −20 °C | 12 weekslimit stated by the authors | max allowable mean deviation from t0: ±10% | Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability' |
| plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquots | Storageseparated | −80 °C | 3 monthslimit stated by the authors | max allowable mean deviation from t0: ±10% | Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability' |
| plasma-citrate, 3.2% sodium citrate; platelet-poor plasma (double centrifuged), 500 µL aliquots | Storageseparated | RT | 24 hlimit stated by the authors | max allowable mean deviation from t0: ±10% | Hoffmann et al. 2025, Front Mol BiosciTable 2, 'Determined stability' |
White Blood Cell Count21 findings
| Specimen, tube | Condition | Temperature | Stable for | Criterion | Study |
|---|---|---|---|---|---|
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 48 hstable for the longest time tested | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'WBC', XN-9000, 4 °C |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 48 hstable for the longest time tested | Bias% vs optimal bias (OP-Bias%, from biological variation; Table 1) | Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'WBC', XN-9000, 4 °C |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 48 hstable for the longest time tested | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'WBC', BC-6800, 4 °C |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 48 hstable for the longest time tested | Bias% vs optimal bias (OP-Bias%, from biological variation; Table 1) | Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'WBC', BC-6800, 4 °C |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 8 hstable for the longest time tested | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'WBC', XN-9000, 4 °C |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 8 hstable for the longest time tested | Bias% vs optimal bias (OP-Bias%, from biological variation; Table 1) | Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'WBC', XN-9000, 4 °C |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 8 hstable for the longest time tested | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'WBC', BC-6800, 4 °C |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 4 hthen changed, changed by 8 h | Bias% vs optimal bias (OP-Bias%, from biological variation; Table 1) | Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'WBC', BC-6800, 4 °C |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | 2-8 °C | 8 hstable for the longest time tested | Bias% vs critical difference (CD%, reference change value; Table 1) | Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'WBC', BC-6800, 4 °C |
| whole-blood-edta, K2-EDTA (BD Vacutainer) | Storageuncentrifuged | 2-8 °C | 48 hstable for the longest time tested | Repeated-measures one-way ANOVA with Tukey post hoc vs 0 h, p < 0.05 (* in table) | Unalli et al. 2021, J Med BiochemResults paragraphs 1-2 and Discussion paragraph 1 (Figures 1-2; numbers in figures only) |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | RT | 48 hstable for the longest time tested | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'WBC', XN-9000, RT |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | RT | 48 hstable for the longest time tested | Bias% vs optimal bias (OP-Bias%, from biological variation; Table 1) | Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'WBC', XN-9000, RT |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | RT | 48 hstable for the longest time tested | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'WBC', BC-6800, RT |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | RT | 48 hstable for the longest time tested | Bias% vs optimal bias (OP-Bias%, from biological variation; Table 1) | Buoro et al. 2016, Rev Bras Hematol HemoterTable 2, row 'WBC', BC-6800, RT |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | RT | 8 hstable for the longest time tested | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'WBC', XN-9000, RT |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | RT | 8 hstable for the longest time tested | Bias% vs optimal bias (OP-Bias%, from biological variation; Table 1) | Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'WBC', XN-9000, RT |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | RT | 8 hstable for the longest time tested | Hodges-Lehmann shift from baseline significant at p < 0.0001 (column 'ΔX with p-value < 0.0001 at the time') | Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'WBC', BC-6800, RT |
| whole-blood-edta, K3-EDTA (Becton Dickinson) | Storageuncentrifuged | RT | 8 hstable for the longest time tested | Bias% vs optimal bias (OP-Bias%, from biological variation; Table 1) | Buoro et al. 2016, Rev Bras Hematol HemoterTable 3, row 'WBC', BC-6800, RT |
| whole-blood-edta, K2-EDTA (BD Vacutainer) | Storageuncentrifuged | RT | 24 hthen changed, changed by 48 h | Repeated-measures one-way ANOVA with Tukey post hoc vs 0 h, p < 0.05 (* in table) | Unalli et al. 2021, J Med BiochemResults, paragraph 1 (Figure 1B) |
| whole-blood-edta, K2EDTA (Becton Dickinson) | Storageuncentrifuged | 33 °C | 24 hthen changed, changed by 48 h | Unacceptable if median % difference > Biasmax (biological variation) and Holm-adjusted Wilcoxon P < 0.05; stable if equivalence shown (two one-tailed Wilcoxon tests vs Biasmax, Pholm < 0.05) | Jain et al. 2018, Biochem Med (Zagreb)Results, last paragraph (equivalence); Table 2, row 'WBC' |
| whole-blood-edta, K2-EDTA (BD Vacutainer) | Storageuncentrifuged | 10 °C | 48 hstable for the longest time tested | Repeated-measures one-way ANOVA with Tukey post hoc vs 0 h, p < 0.05 (* in table) | Unalli et al. 2021, J Med BiochemResults paragraphs 1-2 and Discussion paragraph 1 (Figures 1-2; numbers in figures only) |
Zinc5 findings
| Specimen, tube | Condition | Temperature | Stable for | Criterion | Study |
|---|---|---|---|---|---|
| whole-blood-clotted, BD Vacutainer metal-free SST | Before spinninguncentrifuged | RT | 12 hthen changed, changed by 24 h | mean % deviation from baseline within total change limit TCL = sqrt((2.77 CVa)^2 + (0.5 CVb)^2), CVb from Ricos (Table 2) | Abraham et al. 2019, Biochem Med (Zagreb)Results, paragraph 1 (Figure 3e) |
| whole-blood-clotted, BD Vacutainer trace element-free serum tube | Before spinninguncentrifuged | cold packs in insulated box; 9 (2) °C on arrival | 24 hstable for the longest time tested | % difference within desirable quality specification (from intra- and interindividual variation) | Jones et al. 2024, J NutrTable 2, row 'Zinc, μg/L'; Table 1 |
| whole-blood-clotted, Sarstedt Monovette serum | Before spinninguncentrifuged | chilled in transit (insulated box, 2 frozen cool packs; temperature not recorded) | 24 hstable for the longest time tested | % difference within desirable quality specification for bias (Fraser, from intra- and interindividual variation) | Jones et al. 2021, J NutrTable 2, row 'Zinc', column Serum |
| whole-blood-edta, Sarstedt Monovette EDTA | Before spinninguncentrifuged | chilled in transit (insulated box, 2 frozen cool packs; temperature not recorded) | 24 hstable for the longest time tested | % difference within desirable quality specification for bias (Fraser, from intra- and interindividual variation) | Jones et al. 2021, J NutrTable 2, row 'Zinc', column EDTA |
| whole-blood-heparin, Sarstedt Monovette lithium heparin, trace-element free | Before spinninguncentrifuged | chilled in transit (insulated box, 2 frozen cool packs; temperature not recorded) | 0 hchanged at the first time point, changed by 24 h | % difference within desirable quality specification for bias (Fraser, from intra- and interindividual variation) | Jones et al. 2021, J NutrTable 2, row 'Zinc', column LH |
How it was built
Each row is one finding from a study read in full: specimen, tube, processing, temperature, the longest time the analyte met the study's stability criterion, and that criterion. Studies use different criteria, so a shorter time in one study and a longer one in another can both be right. Where studies disagree, all rows are kept; the summary file gives the shortest time reported for each combination and how many studies agree.
Download
Code, data and tests: github.com/caraulani/sample-stability
Archived on Zenodo: doi.org/10.5281/zenodo.24543300
- excluded_sources.csv (7 KB)